XP Power AHM85PS12C2 Ip22 environmental rating Datasheet

AHM85 Series
GREEN
POWER
•
Medical & IT Safety Approvals
•
Energy Star Level V
•
CEC 2008 & EISA 2007 Compliant
•
IP22 Environmental Rating
•
Compact Format 5.90”x 2.52”x 1.45”
•
<0.5 W Standby Power
•
85 W - Convection Cooled Ratings
•
Class I & Class II Models
•
0 ºC to +60 ºC Operation
•
Very Low Earth Leakage Current
•
3 Year Warranty
The AHM85 series of medical external power supplies is fully approved to international medical safety standards. It has been designed
with very high efficiency and low standby power, enabling it to meet the latest envirnomental legislation. The unit has a fully sealed
enclosure complying with IP22 and a smooth surface finish making it easier to wipe down in a clinical setting. With approvals for class
I and II the product is suitable for hospital, home healthcare and portable medial device applications.
xppower.com
Models and Ratings - Convection-cooled
Output Power
85 W
85 W
85 W
85 W
85 W
85 W
85 W
85 W
Output Voltage V1
12.0 VDC
15.0 VDC
19.0 VDC
24.0 VDC
12.0 VDC
15.0 VDC
19.0 VDC
24.0 VDC
Max Output Current
7.08 A
5.67 A
4.47 A
3.54 A
7.08 A
5.67 A
4.47 A
3.54 A
Notes:
1. Models with suffix ‘C2’ have a Class II equipment protection classification.
Peak(2)
Model Number(1)
AHM85PS12
AHM85PS15
AHM85PS19
AHM85PS24
AHM85PS12C2
AHM85PS15C2
AHM85PS19C2
AHM85PS24C2
2. For optional input connector retention clip, add suffix ‘-A’ to the model number
e.g. AHM85PS24C2-A.
Input Characteristics
Characteristic
Minimum
Input Voltage - Operating
80
Input Frequency
Power Factor
Input Current - No Load
Input Current - Full Load
Inrush Current
No Load Input Power
47
Earth Leakage Current
Input Protection
Typical
115/230
Maximum
264
50/60
63
>0.9
0.05/0.09
0.8/0.4
60-80
120
0.3/0.4
0.5
50/85
180
0.3/0.6
T2.5A/250 V internal fuse in both lines
Units
Notes & Conditions
VAC
Full power 90 VAC to 264 VAC. Derate output power linearly by
20% from 90 VAC to 80 VAC.
Hz
A
A
A
W
µA
mA
EN61000-3-2 class A & D compliant & Energy Star Compliant
115/230 VAC
115/230 VAC
230 VAC cold start, 25 ºC
115/230 VAC
115 V 60 Hz/230 V 50 Hz (Typ.), 264 VAC/60 Hz (Max.)
115/230 VAC/400 Hz
Output Characteristics
Characteristic
Output Voltage - V1
Minimum
Typical
12
Maximum
Units
48
VDC
Output Voltage Adjustment
Minimum Load
%
0
See Models and Ratings table
No user adjustment
A
Start Up Delay
200/100
ms
Hold Up Time
15
ms
115/230 VAC full load (see fig.1)
115/230 VAC full load (see fig.2)
Drift
±0.2
%
After 20 min warm up
Line Regulation
±0.5
%
90-264 VAC (50% load)
Load Regulation
±3
%
0f50g100% load.
%
Recovery within 1% in less than 500 µs for a 50-75% and 7550% load step
Transient Response - V1
5
Over/Undershoot - V1
3
Ripple & Noise
<1
%
1.5
125
Overvoltage Protection
Overload Protection
% pk-pk
%
Temperature Coefficient
Overtemperature Protection
20 MHz bandwidth with external circuit (see fig.3-6)
Vnom, Recycle AC to reset
13.2
18
AHM85PS12 & C2
16.5
22
AHM85PS15 & C2
21.0
28
26.4
33
115.0
175
8.5
12.0
AHM85PS12 & C2
6.8
9.6
AHM85PS15 & C2
4.9
7.6
4.2
6.0
VDC
AHM85PS19 & C2
AHM85PS24 & C2
%
A
I nom, Auto reset
AHM85PS19 & C2
AHM85PS24 & C2
Short Circuit Protection
2
Notes & Conditions
Continuous, trip & restart (hiccup mode)
0.05
%/˚C
°C
Connected to transformer. Auto reset.
xppower.com
Start Up Delay From AC Turn On
Figure 1
Start up example from AC turn on
(230 VAC, 79 ms)
Hold Up Time From Loss of AC
Figure 2
Hold up example at 85 W load
with 230 VAC input (19 ms)
3
xppower.com
Ripple & Noise
Figure 3
AHM85PS12
Ripple & noise example at 85 W load
with 230 VAC input (75 mV)
Figure 4
AHM85PS24
Ripple & noise example at 85 W load
with 230 VAC input (77 mV)
Figure 6
Ripple & noise measurement circuit
10 µF
Electrolytic
4
0.1 µF
Ceramic
xppower.com
Xxxxxxxxxx
General
Specifications
Characteristic
Efficiency
Isolation: Input to Output
Input to Ground
Output to Ground
Minimum
Typical
92
Maximum
4000
1500
500
45
90
Switching Frequency
Units
%
VAC
VAC
VAC
200
110
3.9
Power Density
Mean Time Between Failure
Weight
kHz
W/in3
kHrs
lb (g)
172
0.9 (400)
Notes & Conditions
Full load (see fig.7-8)
Class I Models
Class I Models
PFC stage
DC-DC stage
MIL-HDBK-217F, Notice 2 +25 °C GB
Average Active Efficiency
Characteristic
Average Active Efficiency
115 V / 60 Hz
230 VAC / 50 Hz
AHM85PS12 & C2
90.41
89.92
AHM85PS15 & C2
89.37
88.99
AHM85PS19 & C2
90.36
89.93
AHM85PS24 & C2
91.67
91.23
Units
Notes & Conditions
%
As per Energy Star Level V test procedure
Efficiency Versus Load
100%
90%
80%
Efficiency
70%
60%
50%
40%
30%
20%
10%
0%
10
20
30
40
50
60
70
80
90
100
Load (%)
Figure 7 - AHM85PS12
115Vac input
230Vac input
100%
90%
80%
Efficiency
70%
60%
50%
40%
30%
20%
10%
0%
10
20
30
40
50
60
70
80
90
100
Load (%)
Figure 8 - AHM85PS24
5
xppower.com
Xxxxxxxxxx
Environmental
Characteristic
Minimum
Operating Temperature
Typical
Maximum
0
Units
Notes & Conditions
+60
ºC
Derate linearly to 60% load at 60 ºC from
+40 °C. (See fig.9)
71
ºC
100% Load, with TAMB +40 ºC
Case Temperature
(IEC60601 3rd Edition)
60
90% Load Maximum, with TAMB +40 ºC
48
Storage Temperature
10% Load Maximum, with TAMB +40 ºC
-40
+85
ºC
5
95
%RH
3000
m
Cooling
Convection cooled, see fig.9
Humidity
Operating Altitude
Ingress Protection
Non-condensing
IP22
Shock
3 x 30 g/11 ms shocks in both +ve & -ve
directions along the 3 orthogonal axis,
total 18 shocks.
Vibration
Three axis 5-500 Hz at 2 g x 10 sweeps
Derating Curve
Figure 9
Output Power (%)
100%
60%
0
20
Ambient Temperature (ºC)
10
60
40
Electromagnetic Compatibility - Immunity
Phenomenon
Standard
Test Level
Criteria
EN61204-3
High severity level
as below
Harmonic Current
EN61000-3-2
Class A
ESD
EN61000-4-2
3
A
Radiated
EN61000-4-3
3
A
EFT
EN61000-4-4
3
A
Surges
EN61000-4-5
Installation class 3
A
Conducted
EN61000-4-6
3
A
Magnetic Field
EN61000-4-8
3
Low Voltage PSU EMC
EN61000-4-11
Dips and Interruptions
EN60601-1-2
6
Notes & Conditions
A
Dip:
30% 500 ms
A
Dip:
60% 200 ms
B
Dip:
80% 5000 ms
A
Dip:
100% 5000 ms
B
Dip:
30% 25 AC Cycles
A
230 VAC 100% load, 100 VAC 80% load
Dip:
60% 5 AC Cycles
A
230 VAC 100% load, 100 VAC 15% load
Dip:
100% 0.5 AC Cycles
A
Int.:
>95% 5000 ms
B
xppower.com
Xxxxxxxxxx
Electromagnetic
Compatibility - Emissions
Phenomenon
Standard
Test Level
EN55011/22
Class B
Radiated
EN55011/22
Class B
Voltage Fluctuations
EN61000-3-3
Conducted
Criteria
Notes & Conditions
Safety Agency Approvals
Safety Agency
Safety Standard
CB Report
Certificate # US/15675/UL, IEC60950-1:2005 Ed 2
Category
Information Technology
UL
UL File # E139109-A57-UL-1, UL60950-1 Ed 2 (2007), CSA 22.2 No.60950-1-07 Ed 2
Information Technology
TUV
TUV Certificate # Z1A 10 12 57396 085, EN60950-1:2006
Information Technology
Denan Japan
PSE Certificate
CE
LVD
Safety Agency
Safety Standard
Category
CB Report
Certificate #US/16953/UL, IEC60601-1 Ed 3 Including Risk Management
Medical
UL
UL File # E146893-V1-S8, ANSI/AAMI ES 60601-1:2005 & CSA C22.2 No. 60601-1:08
Medical
TUV
TUV Certificate # B11 06 57396 102, EN60601-1:2006
Medical
Primary to Secondary
2 x MOPP (Means of Patient Protection)
Primary to Earth
1 x MOPP (Means of Patient Protection)
Secondary to Earth
1 x MOPP (Means of Patient Protection)
Equipment Protection Class
Safety Standard
Notes & Conditions
IEC60950-1:2005 Ed 2 & IEC60601-1 Ed 3
See safety agency conditions of
acceptibility for details
Means of Protection
Class I & Class II
Category
IEC60601-1 Ed 3
Environmental Legislation
Authority
Location
Date
EISA
US
2007
CEC
California, US
2008
Notes & Conditions
Energy Star
US
2008
Level V
ErP Directive
Europe
2011
Regulation No. 278/2009
7
xppower.com
Xxxxxxxxxx Details
Mechanical
Weight: 0.9 lbs (400 g)
Dimensions shown in inches (mm).
5.9 (150.0)
36.4 (900.0)
0.79 (Ø20)
2.52
(64.0)
1.4 (36.0)
5.12 (130.0)
IEC320-C14
1.45
(37.0)
3.02 (76.9)
1.57 (40.0)
1.55
(39.4)
5.11 (130.0)
Shell
Optional Input Connector Retention (-A)
2
1
4
Standard (Class I)
C2 Models (Class II)
C2-8 Models (Class II)
3
1.40 (35.7)
Output Connector equivalent
to KPPX-4P (Non Locking)
IEC 320-C14
IEC 320-C18
IEC 320-C8
Pin 1
Pin 2
Pin 3
Pin 4
Outer Shell
Output +
Output +
Return
Return
GND*
Outer Shell
C2 Models
Floating
* Functional earth.
8
29-Sept-14
AHM100 Series
GREEN
POWER
•
Medical & IT Safety Approvals
•
Energy Star Level V
•
CEC 2008 & EISA 2007 Compliant
•
IP22 Environmental Rating
•
Compact Format 6.50” x 2.52” x 1.46”
•
<0.5 W Standby Power
•
100 W - Convection Cooled Ratings
•
Class I & Class II Models
•
0 ºC to +60 ºC Operation
•
Very Low Earth Leakage Current
•
3 Year Warranty
The AHM100 series of medical external power supplies is fully approved to international medical safety standards. It has been
designed with very high efficiency and low standby power, enabling it to meet the latest envirnomental legislation. The unit has a fully
sealed enclosure complying with IP22 and a smooth surface finish making it easier to wipe down in a clinical setting. With approvals
for class I and II the product is suitable for hospital, home healthcare and portable medial device applications.
xppower.com
Models and Ratings - Convection-cooled
Output Power
100 W
100 W
100 W
90 W
100 W
100 W
100 W
100 W
100 W
90 W
100 W
100 W
Output Voltage V1
12.0 VDC
15.0 VDC
19.0 VDC
24.0 VDC
24.0 VDC
48.0 VDC
12.0 VDC
15.0 VDC
19.0 VDC
24.0 VDC
24.0 VDC
48.0 VDC
Max Output Current
8.33 A
6.67 A
5.26 A
3.75 A
4.16 A
2.08 A
8.33 A
6.67 A
5.26 A
3.75 A
4.16 A
2.08 A
Notes:
1. Models with suffix ‘C2’ have a Class II equipment protection classification.
2. Maximum peak duration 300 ms, average power must not exceed 90 W.
Peak(2)
7A
7A
Model Number(1)
AHM100PS12
AHM100PS15
AHM100PS19
AHM100PS24P(3)
AHM100PS24
AHM100PS48
AHM100PS12C2
AHM100PS15C2
AHM100PS19C2
AHM100PS24C2P(3)
AHM100PS24C2
AHM100PS48C2
3. Peak models are not standard product. Contact sales for details and availability.
4. For optional input connector retention clip, add suffix ‘-A’ to the model number
e.g. AHM100PS24C2-A.
Input Characteristics
Characteristic
Minimum
Input Voltage - Operating
80
Input Frequency
Power Factor
Input Current - No Load
Input Current - Full Load
Inrush Current
No Load Input Power
47
Earth Leakage Current
Input Protection
Typical
115/230
Maximum
264
50/60
63
>0.9
0.05/0.09
1.0/0.5
60-80
120
0.3/0.4
0.5
50/85
180
0.3/0.6
T2.5A/250 V internal fuse in both lines
Units
Notes & Conditions
VAC
Full power 90 VAC to 264 VAC. Derate output power linearly by
20% from 90 VAC to 80 VAC.
Hz
A
A
A
W
µA
mA
EN61000-3-2 class A & D compliant & Energy Star Compliant
115/230 VAC
115/230 VAC
230 VAC cold start, 25 ºC
115/230 VAC
115 V 60 Hz/230 V 50 Hz (Typ.), 264 VAC/60 Hz (Max.)
115/230 VAC/400 Hz
Output Characteristics
Characteristic
Output Voltage - V1
Minimum
Typical
12
Maximum
Units
48
VDC
Output Voltage Adjustment
Minimum Load
%
0
See Models and Ratings table
No user adjustment
A
Start Up Delay
200/100
ms
Hold Up Time
20
ms
115/230 VAC full load (see fig.1)
115/230 VAC full load (see fig.2)
Drift
±0.2
%
After 20 min warm up
Line Regulation
±0.5
%
90-264 VAC (50% load)
Load Regulation
±3
%
0f50g100% load.
5
%
Recovery within 1% in less than 500 µs for a 50-75% and 7550% load step
1.5
% pk-pk
Transient Response - V1
Over/Undershoot - V1
3
Ripple & Noise
<1
%
125
Overvoltage Protection
Overload Protection
%
Temperature Coefficient
Overtemperature Protection
20 MHz bandwidth with external circuit (see fig.3-6)
Vnom, Recycle AC to reset
13.2
18
AHM100PS12 & C2
16.5
22
AHM100PS15 & C2
21.0
28
26.4
33
52.8
59
115.0
175
10.0
14.0
8.0
11.4
6.3
9.0
5.0
7.1
AHM100PS24 & C2
2.5
3.6
AHM100PS48 & C2
VDC
AHM100PS19 & C2
AHM100PS24 & C2
AHM100PS48 & C2
%
I nom, Auto reset
AHM100PS12 & C2
AHM100PS15 & C2
A
Short Circuit Protection
2
Notes & Conditions
AHM100PS19 & C2
Continuous, trip & restart (hiccup mode)
0.05
%/˚C
°C
Connected to transformer. Auto reset.
xppower.com
Start Up Delay From AC Turn On
Figure 1
Start up example from AC turn on
(230 VAC, 95 ms)
Hold Up Time From Loss of AC
Figure 2
Hold up example at 100 W load
with 230 VAC input (25 ms)
3
xppower.com
Ripple & Noise
Figure 3
AHM100PS12
Ripple & noise example at 100 W load
with 230 VAC input (58 mV)
Figure 4
AHM100PS24
Ripple & noise example at 100 W load
with 230 VAC input (258 mV)
Figure 5
AHM100PS48
Ripple & noise example at 100 W load
with 230 VAC input (318 mV)
Figure 6
Ripple & noise measurement circuit
10 µF
Electrolytic
4
0.1 µF
Ceramic
xppower.com
Xxxxxxxxxx
General
Specifications
Characteristic
Efficiency
Isolation: Input to Output
Input to Ground
Output to Ground
Minimum
Typical
92
Maximum
4000
1500
500
45
90
Switching Frequency
Units
%
VAC
VAC
VAC
200
110
4.4
Power Density
Mean Time Between Failure
Weight
Notes & Conditions
Full load (see fig.7-9)
Class I Models
Class I Models
PFC stage
DC-DC stage
kHz
W/in3
kHrs
lb (g)
153
1.1 (500)
MIL-HDBK-217F, Notice 2 +25 °C GB
Average Active Efficiency
GREEN
Characteristic
AHM100PS12
AHM100PS15
AHM100PS19
AHM100PS24
AHM100PS48
POWER
Average Active Efficiency
115 V / 60 Hz
230 VAC / 50 Hz
90.52
90.32
89.88
89.64
89.99
89.74
91.67
91.43
91.80
91.63
& C2
& C2
& C2
& C2
& C2
Units
Notes & Conditions
%
As per Energy Star Level V test procedure
Efficiency Versus Load
100%
90%
80%
Efficiency
70%
60%
50%
40%
30%
20%
10%
0%
10
20
30
40
50
60
70
80
90
100
Load (%)
100%
100%
90%
90%
80%
80%
70%
70%
60%
60%
Efficiency
Efficiency
Figure 7 - AHM100PS12
50%
40%
50%
40%
30%
30%
20%
20%
10%
10%
0%
0%
10
20
30
40
50
60
70
80
90
100
10
Load (%)
20
30
40
50
60
70
80
90
100
Load (%)
Figure 8 - AHM100PS24
Figure 9 - AHM100PS48
115Vac input
230Vac input
5
xppower.com
Xxxxxxxxxx
Environmental
Characteristic
Minimum
Operating Temperature
Typical
Maximum
0
Units
Notes & Conditions
+60
ºC
Derate linearly to 60% load at 60 ºC from
+40 °C. (See fig.10)
71
ºC
100% Load, with TAMB +40 ºC
Case Temperature
(IEC60601 3rd Edition)
60
80% Load Maximum, with TAMB +40 ºC
48
Storage Temperature
10% Load Maximum, with TAMB +40 ºC
-40
+85
ºC
5
95
%RH
3000
m
Cooling
Convection cooled, see fig.10
Humidity
Operating Altitude
Ingress Protection
Non-condensing
IP22
Shock
3 x 30 g/11 ms shocks in both +ve & -ve
directions along the 3 orthogonal axis,
total 18 shocks.
Vibration
Three axis 5-500 Hz at 2 g x 10 sweeps
Derating Curve
Figure 10
Output Power (%)
100%
60%
0
20
Ambient Temperature (ºC)
10
60
40
Electromagnetic Compatibility - Immunity
Phenomenon
Standard
Test Level
Criteria
EN61204-3
High severity level
as below
Harmonic Current
EN61000-3-2
Class A
ESD
EN61000-4-2
3
A
Radiated
EN61000-4-3
3
A
EFT
EN61000-4-4
3
A
Surges
EN61000-4-5
Installation class 3
A
Conducted
EN61000-4-6
3
A
Magnetic Field
EN61000-4-8
3
Low Voltage PSU EMC
EN61000-4-11
Dips and Interruptions
EN60601-1-2
6
Notes & Conditions
A
Dip:
30% 500 ms
A
Dip:
60% 200 ms
B
Dip:
80% 5000 ms
A
Dip:
100% 5000 ms
B
Dip:
30% 25 AC Cycles
A
230 VAC 100% load, 100 VAC 80% load
Dip:
60% 5 AC Cycles
A
230 VAC 100% load, 100 VAC 15% load
Dip:
100% 0.5 AC Cycles
A
Int.:
>95% 5000 ms
B
xppower.com
Xxxxxxxxxx
Electromagnetic
Compatibility - Emissions
Phenomenon
Standard
Test Level
EN55011/22
Class B
Radiated
EN55011/22
Class B
Voltage Fluctuations
EN61000-3-3
Conducted
Criteria
Notes & Conditions
Safety Agency Approvals
Safety Agency
Safety Standard
CB Report
Certificate # US/15053A/UL, IEC60950-1:2005 Ed 2
Category
Information Technology
UL
UL File # E139109-A45-UL-1, UL60950-1 Ed 2 (2007), CSA 22.2 No.60950-1-07 Ed 2
Information Technology
TUV
TUV Certificate # Z1A 10 07 57396 078, EN60950-1:2006
Information Technology
Denan Japan
PSE Certificate
CE
LVD
Safety Agency
Safety Standard
Category
CB Report
Certificate #US/18097/UL, IEC60601-1 Ed 3 Including Risk Management
Medical
UL
UL File # E146893, ANSI/AAMI ES 60601-1:2005 & CSA C22.2 No. 60601-1:08
Medical
TUV
TUV Certificate # Z11 12 57396 122, EN60601-1:2006
Medical
Primary to Secondary
2 x MOPP (Means of Patient Protection)
Primary to Earth
1 x MOPP (Means of Patient Protection)
Secondary to Earth
1 x MOPP (Means of Patient Protection)
Equipment Protection Class
Safety Standard
Notes & Conditions
IEC60950-1:2005 Ed 2 & IEC60601-1 Ed 3
See safety agency conditions of
acceptibility for details
Means of Protection
Class I & Class II
Category
IEC60601-1 Ed 3
Environmental Legislation
Authority
Location
Date
EISA
US
2007
CEC
California, US
2008
Notes & Conditions
Energy Star
US
2008
Level V
ErP Directive
Europe
2011
Regulation No. 278/2009
7
xppower.com
Xxxxxxxxxx Details
Mechanical
Weight: 1.1 lbs (500 g)
Dimensions shown in inches (mm).
6.29 (160.0)
36.4 (900.0)
0.2 (5.0)
0.79 (Ø20)
2.52
(64.0)
5.12 (130.0)
1.4 (36.0)
1.45
(37.0)
3.02 (76.9)
1.57 (40.0)
1.55
(39.4)
5.11 (130.0)
Shell
Optional Input Connector Retention (-A)
2
1
4
Standard (Class I)
C2 Models (Class II)
C2-8 Models (Class II)
3
1.40 (35.7)
Output Connector equivalent
to KPPX-4P (Non Locking)
IEC 320-C14
IEC 320-C18
IEC 320-C8
Pin 1
Pin 2
Pin 3
Pin 4
Outer Shell
Output +
Output +
Return
Return
GND*
Outer Shell
C2 Models
Floating
* Functional earth.
8
29-Sept-14
AHM150 Series
GREEN
POWER
•
Medical & IT Safety Approvals
•
Energy Star Level V
•
CEC 2008 & EISA 2007 Compliant
•
IP22 Environmental Rating
•
Compact Format 7.80” x 3.15” x 1.45”
•
<0.5 W Standby Power
•
150 W - Convection Cooled Ratings
•
Class I & Class II Models
•
0 ºC to +60 ºC Operation
•
Very Low Earth Leakage Current
•
3 Year Warranty
The AHM150 series of medical external power supplies is fully approved to international medical safety standards. It has been
designed with very high efficiency and low standby power, enabling it to meet the latest envirnomental legislation. The unit has a fully
sealed enclosure complying with IP22 and a smooth surface finish making it easier to wipe down in a clinical setting. With both
medical & IT approvals in class I and class II formats the product is suitable for hospital, home healthcare, portable medial device
applications and a wide range of IT applications.
xppower.com
Models and Ratings - Convection-cooled
Output Power
150 W
150 W
150 W
150 W
150 W
150 W
150 W
150 W
150 W
150 W
Output Voltage V1
12.0 VDC
15.0 VDC
19.0 VDC
24.0 VDC
48.0 VDC
12.0 VDC
15.0 VDC
19.0 VDC
24.0 VDC
48.0 VDC
Max Output Current
12.50 A
10.00 A
7.89 A
6.25 A
3.13 A
12.50 A
10.00 A
7.89 A
6.25 A
3.13 A
Model Number(1)
AHM150PS12
AHM150PS15
AHM150PS19
AHM150PS24
AHM150PS48
AHM150PS12C2
AHM150PS15C2
AHM150PS19C2
AHM150PS24C2
AHM150PS48C2
Notes:
1. For optional input connector retention clip, add suffix -A’
‘ to the model number e.g. AHM150PS24-A. Models with suffix C2’
‘ have a class II equipment protection
classification. For IEC320-C8 input connector with class II models, add suffix -8’
‘ to the model number, e.g. AHM150PS24C2-8.
Input Characteristics
Characteristic
Minimum
Input Voltage - Operating
80
Input Frequency
Power Factor
Input Current - No Load
Input Current - Full Load
Inrush Current
No Load Input Power
47
Earth Leakage Current
Input Protection
Typical
115/230
Maximum
264
50/60
63
>0.9
0.06/0.04
1.4/0.7
60-80
120
0.3/0.4
0.5
50/100
200
0.3/0.6
T4.0A/250 V internal fuse in both lines
Units
Notes & Conditions
VAC
Full power 90 VAC to 264 VAC. Derate output power linearly by
20% from 90 VAC to 80 VAC.
Hz
A
A
A
W
µA
mA
EN61000-3-2 class A & D compliant & Energy Star Compliant
115/230 VAC
115/230 VAC
230 VAC cold start, 25 ºC
115/230 VAC
115 V 60 Hz/230 V 50 Hz (Typ.), 264 VAC/60 Hz (Max.)
115/230 VAC/400 Hz
Output Characteristics
Characteristic
Output Voltage - V1
Minimum
Typical
12
Maximum
Units
48
VDC
Output Voltage Adjustment
Minimum Load
%
0
See Models and Ratings table
No user adjustment
A
Start Up Delay
200/100
ms
Hold Up Time
10
ms
115/230 VAC full load (see fig.1)
115/230 VAC full load (see fig.2)
Drift
±0.2
%
After 20 min warm up
Line Regulation
±0.5
%
90-264 VAC (50% load)
Load Regulation
±4
%
0f50g100% load.
5
%
Recovery within 1% in less than 500 µs for a 50-75% and 7550% load step
1.5
% pk-pk
Transient Response - V1
Over/Undershoot - V1
3
Ripple & Noise
<1
%
125
Overvoltage Protection
Overload Protection
%
Temperature Coefficient
Overtemperature Protection
20 MHz bandwidth with external circuit (see fig.3-6)
Vnom, Recycle AC to reset
13.2
18
AHM150PS12 & C2
16.5
22
AHM150PS15 & C2
20.9
28
26.4
33
52.8
59
115.0
175
15.0
21.3
12.0
17.0
8.7
13.4
6.8
10.6
AHM150PS24 & C2
3.7
5.3
AHM150PS48 & C2
VDC
AHM150PS19 & C2
AHM150PS24 & C2
AHM150PS48 & C2
%
I nom, Auto reset
AHM150PS12 & C2
AHM150PS15 & C2
A
Short Circuit Protection
2
Notes & Conditions
AHM150PS19 & C2
Continuous, trip & restart (hiccup mode)
0.05
%/˚C
°C
Connected to transformer. Auto reset.
xppower.com
Start Up Delay From AC Turn On
Figure 1
Start up example from AC turn on
(230 VAC, 196 ms)
Hold Up Time From Loss of AC
Figure 2
Hold up example at 150 W load
with 230 VAC input (16 ms)
3
xppower.com
Ripple & Noise
Figure 3
AHM150PS12
Ripple & noise example at 150 W load
with 230 VAC input (60 mV)
Figure 4
AHM150PS24
Ripple & noise example at 150 W load
with 230 VAC input (137 mV)
Figure 5
AHM150PS48
Ripple & noise example at 150 W load
with 230 VAC input (296 mV)
Figure 6
Ripple & noise measurement circuit
10 µF
Electrolytic
4
0.1 µF
Ceramic
xppower.com
Xxxxxxxxxx
General
Specifications
Characteristic
Efficiency
Isolation: Input to Output
Input to Ground
Output to Ground
Minimum
Typical
92
Maximum
4000
1500
500
30
90
Switching Frequency
Units
%
VAC
VAC
VAC
200
110
4.2
Power Density
Mean Time Between Failure
Weight
163
1.3 (600)
Notes & Conditions
Full load (see fig.7-9)
kHz
PFC stage
DC-DC stage
W/in3
kHrs
lb (g)
MIL-HDBK-217F, Notice 2 +25 °C GB
Average Active Efficiency
GREEN
POWER
Average Active Efficiency
115 V / 60 Hz
230 VAC / 50 Hz
90.31
91.74
91.44
91.78
92.00
92.52
91.46
92.99
92.38
92.85
Characteristic
AHM150PS12
AHM150PS15
AHM150PS19
AHM150PS24
AHM150PS48
Units
Notes & Conditions
%
As per Energy Star Level V test procedure
Efficiency Versus Load
100%
90%
80%
70%
60%
50%
40%
30%
20%
10%
0%
10
20
30
40
50
60
70
80
90
100
Load (%)
Figure 7 - AHM150PS12
100%
100%
90%
90%
80%
80%
70%
70%
60%
60%
50%
50%
40%
40%
30%
30%
20%
20%
10%
10%
0%
0%
10
20
30
40
50
60
70
80
90
10
100
20
30
40
50
60
70
80
90
100
Load (%)
Load (%)
Figure 8 - AHM150PS24
Figure 9 - AHM150PS48
115Vac input
230Vac input
5
xppower.com
Xxxxxxxxxx
Environmental
Characteristic
Minimum
Operating Temperature
Typical
Maximum
0
Units
Notes & Conditions
+60
ºC
Derate linearly to 60% load at 60 ºC from
+40 °C. (See fig.10)
71
ºC
100% Load, with TAMB +40 ºC
Case Temperature
(IEC60601 3rd Edition)
60
70% Load Maximum, with TAMB +40 ºC
48
Storage Temperature
10% Load Maximum, with TAMB +40 ºC
-40
+85
ºC
5
95
%RH
3000
m
Cooling
Convection cooled, see fig.10
Humidity
Operating Altitude
Ingress Protection
Non-condensing
IP22
Shock
3 x 30 g/11 ms shocks in both +ve & -ve
directions along the 3 orthogonal axis,
total 18 shocks.
Vibration
Three axis 5-500 Hz at 2 g x 10 sweeps
Derating Curve
Figure 10
Output Power (%)
100%
60%
0
20
Ambient Temperature (ºC)
10
60
40
Electromagnetic Compatibility - Immunity
Phenomenon
Standard
Test Level
Criteria
EN61204-3
High severity level
as below
Harmonic Current
EN61000-3-2
Class A
ESD
EN61000-4-2
3
A
Radiated
EN61000-4-3
3
A
EFT
EN61000-4-4
3
A
Surges
EN61000-4-5
Installation class 3
A
Conducted
EN61000-4-6
3
A
Magnetic Field
EN61000-4-8
3
Low Voltage PSU EMC
EN61000-4-11
Dips and Interruptions
EN60601-1-2
6
Notes & Conditions
A
Dip:
30% 500 ms
A
Dip:
60% 200 ms
B
Dip:
80% 5000 ms
A
Dip:
100% 5000 ms
B
Dip:
30% 25 AC Cycles
A
230 VAC 100% load, 100 VAC 80% load
Dip:
60% 5 AC Cycles
A
230 VAC 100% load, 100 VAC 15% load
Dip:
100% 0.5 AC Cycles
A
Int.:
>95% 5000 ms
B
xppower.com
Xxxxxxxxxx
Electromagnetic
Compatibility - Emissions
Phenomenon
Standard
Test Level
EN55011/22
Class B
Radiated
EN55011/22
Class B
Voltage Fluctuations
EN61000-3-3
Conducted
Criteria
Notes & Conditions
Safety Agency Approvals
Safety Agency
Safety Standard
CB Report
Certificate # US/16498/UL IEC60950-1:2005 Ed 2
Information Technology
UL
UL File #E139109-A62-UL UL60950-1 (2007), CSA 22.2 No.60950-1-07 Ed 2
Information Technology
TUV
TUV Certificate # Z1A 11 02 57396 093, EN60950-1:2006
Information Technology
Denan Japan
PSE Certificate
CE
LVD
Safety Agency
Category
Safety Standard
Category
CB Report
Certificate #US/17207/UL, IEC60601-1 Ed 3 Including Risk Management
Medical
UL
UL File # E146893-V1-S9, ANSI/AAMI ES 60601-1:2005 & CSA C22.2 No. 60601-1:08
Medical
TUV
TUV Certificate # B11 07 57396 105, EN60601-1:2006
Medical
Means of Protection
Category
Primary to Secondary
2 x MOPP (Means of Patient Protection)
Primary to Earth
1 x MOPP (Means of Patient Protection)
Secondary to Earth
1 x MOPP (Means of Patient Protection)
Equipment Protection Class
Safety Standard
Notes & Conditions
IEC60950-1:2005 Ed 2 & IEC60601-1 Ed 3
See safety agency conditions of
acceptibility for details
Class I & Class II
IEC60601-1 Ed 3
Environmental Legislation
Authority
Location
Date
EISA
US
2007
CEC
California, US
2008
Notes & Conditions
Energy Star
US
2008
Level V
ErP Directive
Europe
2011
Regulation No. 278/2009
7
xppower.com
Xxxxxxxxxx Details
Mechanical
Weight: 1.3 lbs (600 g)
Dimensions shown in inches (mm).
7.8 (200.0)
36.4 (900.0)
0.19 (4.9)
0.79 (Ø20)
3.15
(80.0)
1.4 (36.0)
5.12 (130.0)
IEC320-C14
1.45
(37.0)
3.94 (100.3)
2.02
(51.5)
1.58
(40.30)
6.8 (173.0)
Shell
Optional Input Connector Retention (-A)
2
Standard (Class I)
C2 Models (Class II)
C2-8 Models (Class II)
1.40 (35.7)
1
4
3
Output Connector equivalent
to KPPX-4P (Non Locking)
IEC 320-C14
IEC 320-C18
IEC 320-C8
Pin 1
Pin 2
Pin 3
Pin 4
Outer Shell
Output +
Output +
Return
Return
GND*
Outer Shell
C2 Models
Floating
* Functional earth.
8
29-Sept-14
AHM180 Series
GREEN
POWER
•
Medical & IT Safety Approvals
•
Energy Star Level V
•
CEC 2008 & EISA 2007 Compliant
•
IP22 Environmental Rating
•
Compact Format 7.90” x 3.15” x 1.61”
•
<0.5 W Standby Power
•
180 W - Convection Cooled Ratings
•
Class I & Class II Models
•
0 ºC to +60 ºC Operation
•
Very Low Earth Leakage Current
•
3 Year Warranty
The AHM180 series of medical external power supplies is fully approved to international medical safety standards. It has been
designed with very high efficiency and low standby power, enabling it to meet the latest envirnomental legislation. The unit has a fully
sealed enclosure complying with IP22 and a smooth surface finish making it easier to wipe down in a clinical setting. With both
medical & IT approvals the product is suitable for hospital, home healthcare, portable medial device applications and a wide range of
IT applications.
xppower.com
Models and Ratings - Convection-cooled
Output Power
165 W
180 W
180 W
180 W
180 W
165 W
180 W
180 W
180 W
180 W
Output Voltage V1
12.0 VDC
15.0 VDC
19.0 VDC
24.0 VDC
48.0 VDC
12.0 VDC
15.0 VDC
19.0 VDC
24.0 VDC
48.0 VDC
Max Output Current
13.75 A
12.00 A
9.47 A
7.50 A
3.75 A
13.50 A
12.00 A
9.47 A
7.50 A
3.75 A
Model Number(1)
AHM180PS12
AHM180PS15
AHM180PS19
AHM180PS24
AHM180PS48
AHM180PS12C2
AHM180PS15C2
AHM180PS19C2
AHM180PS24C2
AHM180PS48C2
Notes:
1. For optional input connector retention clip, add suffix -A’
‘ to the model number e.g. AHM180PS24-A. Models with suffix C2’
‘ have a class II equipment protection
classification. For IEC320-C8 input connector with class II models, add suffix -8’
‘ to the model number, e.g. AHM180PS24C2-8.
Input Characteristics
Characteristic
Minimum
Input Voltage - Operating
80
Input Frequency
Power Factor
Input Current - No Load
Input Current - Full Load
Inrush Current
No Load Input Power
47
Earth Leakage Current
Input Protection
Typical
115/230
Maximum
264
50/60
63
>0.9
0.08/0.05
1.7/0.9
60-80
120
0.3/0.4
0.5
50/100
200
0.3/0.6
T4.0A/250 V internal fuse in both lines
Units
Notes & Conditions
VAC
Full power 90 VAC to 264 VAC. Derate output power linearly by
20% from 90 VAC to 80 VAC.
Hz
A
A
A
W
µA
mA
EN61000-3-2 class A & D compliant & Energy Star Compliant
115/230 VAC
115/230 VAC
230 VAC cold start, 25 ºC
115/230 VAC
115 V 60 Hz/230 V 50 Hz (Typ.), 264 VAC/60 Hz (Max.)
115/230 VAC/400 Hz
Output Characteristics
Characteristic
Output Voltage - V1
Minimum
Typical
12
Maximum
Units
48
VDC
Output Voltage Adjustment
Minimum Load
%
0
See Models and Ratings table
No user adjustment
A
Start Up Delay
200/100
ms
Hold Up Time
15
ms
115/230 VAC full load (see fig.1)
115/230 VAC full load (see fig.2)
Drift
±0.2
%
After 20 min warm up
Line Regulation
±0.5
%
90-264 VAC (50% load)
Load Regulation
±4
%
0f50g100% load.
5
%
Recovery within 1% in less than 500 µs for a 50-75% and 7550% load step
1.5
% pk-pk
Transient Response - V1
Over/Undershoot - V1
3
Ripple & Noise
<1
%
125
Overvoltage Protection
Overload Protection
%
Temperature Coefficient
Overtemperature Protection
20 MHz bandwidth with external circuit (see fig.3-6)
Vnom, Recycle AC to reset
13.2
18
AHM180PS12 & C2
16.5
22
AHM180PS15 & C2
20.9
28
26.4
33
52.8
59
115.0
175
16.5
23.4
14.4
20.4
10.4
16.1
8.3
12.8
AHM180PS24 & C2
4.5
6.4
AHM180PS48 & C2
VDC
AHM180PS19 & C2
AHM180PS24 & C2
AHM180PS48 & C2
%
I nom, Auto reset
AHM180PS12 & C2
AHM180PS15 & C2
A
Short Circuit Protection
2
Notes & Conditions
AHM180PS19 & C2
Continuous, trip & restart (hiccup mode)
0.05
%/˚C
°C
Connected to transformer. Auto reset.
xppower.com
Start Up Delay From AC Turn On
Figure 1
Start up example from AC turn on
(230 VAC, 196 ms)
Hold Up Time From Loss of AC
Figure 2
Hold up example at 180 W load
with 230 VAC input (23 ms)
3
xppower.com
Ripple & Noise
Figure 3
AHM180PS12
Ripple & noise example at 165 W load
with 230 VAC input (50 mV)
Figure 4
AHM180PS24
Ripple & noise example at 180 W load
with 230 VAC input (228 mV)
Figure 5
AHM180PS48
Ripple & noise example at 180 W load
with 230 VAC input (360 mV)
Figure 6
Ripple & noise measurement circuit
10 µF
Electrolytic
4
0.1 µF
Ceramic
xppower.com
Xxxxxxxxxx
General
Specifications
Characteristic
Efficiency
Isolation: Input to Output
Input to Ground
Output to Ground
Minimum
Typical
92
Maximum
4000
1500
500
30
90
Switching Frequency
Units
%
VAC
VAC
VAC
200
110
4.4
Power Density
Mean Time Between Failure
Weight
156
1.4 (620)
Average Active Efficiency
GREEN
AHM180PS12
AHM180PS15
AHM180PS19
AHM180PS24
AHM180PS48
kHz
PFC stage
DC-DC stage
W/in3
kHrs
lb (g)
MIL-HDBK-217F, Notice 2 +25 °C GB
POWER
Average Active Efficiency
115 V / 60 Hz
230 VAC / 50 Hz
90.19
91.25
91.13
92.26
91.28
92.44
91.67
92.83
92.11
93.12
Characteristic
Notes & Conditions
Full load (see fig.7-9)
Units
Notes & Conditions
%
As per Energy Star Level V test procedure
Efficiency Versus Load
100%
90%
80%
70%
60%
50%
40%
30%
20%
10%
0%
10
20
30
40
50
60
70
80
90
100
Load (%)
Figure 7 - AHM180PS12
100%
100%
90%
90%
80%
80%
70%
70%
60%
60%
50%
50%
40%
40%
30%
30%
20%
20%
10%
10%
0%
0%
10
20
30
40
50
60
70
80
90
10
100
20
30
40
50
60
70
80
90
100
Load (%)
Load (%)
Figure 8 - AHM180PS24
Figure 9 - AHM180PS48
115Vac input
230Vac input
5
xppower.com
Xxxxxxxxxx
Environmental
Characteristic
Minimum
Operating Temperature
Typical
Maximum
0
Units
Notes & Conditions
+60
ºC
Derate linearly to 60% load at 60 ºC from
+40 °C. (See fig.10)
71
ºC
100% Load, with TAMB +40 ºC
Case Temperature
(IEC60601 3rd Edition)
60
60% Load Maximum, with TAMB +40 ºC
48
Storage Temperature
5% Load Maximum, with TAMB +40 ºC
-40
+85
ºC
5
95
%RH
3000
m
Cooling
Convection cooled, see fig.10
Humidity
Operating Altitude
Ingress Protection
Non-condensing
IP22
Shock
3 x 30 g/11 ms shocks in both +ve & -ve
directions along the 3 orthogonal axis,
total 18 shocks.
Vibration
Three axis 5-500 Hz at 2 g x 10 sweeps
Derating Curve
Figure 10
Output Power (%)
100%
60%
0
20
Ambient Temperature (ºC)
10
60
40
Electromagnetic Compatibility - Immunity
Phenomenon
Standard
Test Level
Criteria
EN61204-3
High severity level
as below
Harmonic Current
EN61000-3-2
Class A
ESD
EN61000-4-2
3
A
Radiated
EN61000-4-3
3
A
EFT
EN61000-4-4
3
A
Surges
EN61000-4-5
Installation class 3
A
Conducted
EN61000-4-6
3
A
Magnetic Field
EN61000-4-8
3
Low Voltage PSU EMC
EN61000-4-11
Dips and Interruptions
EN60601-1-2
6
Notes & Conditions
A
Dip:
30% 500 ms
A
Dip:
60% 200 ms
B
Dip:
80% 5000 ms
A
Dip:
100% 5000 ms
B
Dip:
30% 25 AC Cycles
A
230 VAC 100% load, 100 VAC 80% load
Dip:
60% 5 AC Cycles
A
230 VAC 100% load, 100 VAC 80% load
Dip:
100% 0.5 AC Cycles
A
Int.:
>95% 5000 ms
B
xppower.com
Xxxxxxxxxx
Electromagnetic
Compatibility - Emissions
Phenomenon
Standard
Test Level
EN55011/22
Class B
Radiated
EN55011/22
Class B
Voltage Fluctuations
EN61000-3-3
Conducted
Criteria
Notes & Conditions
Safety Agency Approvals
Safety Agency
Safety Standard
CB Report
Certificate # US/16502/UL EC60950-1:2005 Ed 2
Category
Information Technology
UL
UL File # E139109-A48-UL UL60950-1 (2007), CSA 22.2 No.60950-1-07 Ed 2
Information Technology
TUV
TUV Certificate # Z1A 11 02 57396 094, EN60950-1:2006
Information Technology
Denan Japan
PSE Certificate
CE
LVD
Safety Agency
Safety Standard
Category
CB Report
Certificate #US/17220/UL, IEC60601-1 Ed 3 Including Risk Management
Medical
UL
UL File # E146893-V1-S10, ANSI/AAMI ES 60601-1:2005 & CSA C22.2 No. 60601-1:08
Medical
TUV
TUV Certificate # B11 07 57396 104, EN60601-1:2006
Medical
Primary to Secondary
2 x MOPP (Means of Patient Protection)
Primary to Earth
1 x MOPP (Means of Patient Protection)
Secondary to Earth
1 x MOPP (Means of Patient Protection)
Equipment Protection Class
Safety Standard
Notes & Conditions
IEC60950-1:2005 Ed 2 & IEC60601-1 Ed 2
See safety agency conditions of
acceptibility for details
Means of Protection
Class I & Class II
Category
IEC60601-1 Ed 3
Environmental Legislation
Authority
Location
Date
EISA
US
2007
CEC
Notes & Conditions
California, US
2008
Energy Star
US
2008
Level V
ErP Directive
Europe
2011
Regulation No. 278/2009
7
xppower.com
Xxxxxxxxxx Details
Mechanical
Weight: 1.1 lbs (500 g)
Dimensions shown in inches (mm).
7.87 (200.0)
0.193
(4.9)
35.43 (900.0)
0.79 (Ø20.0)
3.15
(80.0)
1.41 (36.0)
5.12 (130.0)
1.61
(41.0)
3.94 (100.3)
2.02
(51.5) 1.58
(40.3)
6.81 (173.0)
Shell
Optional Input Connector Retention (-A)
Standard (Class I)
C2 Models (Class II)
C2-8 Models (Class II)
1.40 (35.7)
2
4
1
3
Output Connector equivalent
to KPPX-4P (Non Locking)
IEC 320-C14
IEC 320-C18
IEC 320-C8
Pin 1
Pin 2
Pin 3
Pin 4
Outer Shell
Output +
Output +
Return
Return
GND*
Outer Shell
C2 Models
Floating
* Functional earth.
8
29-Sept-14
AHM250 Series
GREEN
POWER
•
Medical & IT Safety Approvals
•
Energy Star Level V
•
CEC 2008 & EISA 2007 Compliant
•
IP22 Environmental Rating
•
Compact Format 8.60” x 3.50” x 1.46”
•
<0.5 W Standby Power
•
250 W - Convection Cooled Ratings
•
IEC Input Cable Retention (Optional)
•
0 ºC to +60 ºC Operation
•
Very Low Earth Leakage Current
•
3 Year Warranty
The AHM250 series of medical external power supplies is fully approved to international medical safety standards. It has been
designed with very high efficiency and low standby power, enabling it to meet the latest envirnomental legislation. The unit has a fully
sealed enclosure complying with IP22 and a smooth surface finish making it easier to wipe down in a clinical setting. With both
medical & IT approvals the product is suitable for hospital, home healthcare, portable medial device applications and a wide range of
IT applications.
xppower.com
Models and Ratings - Convection-cooled
Output Power
210 W
220 W
240 W
250 W
250 W
Output Voltage V1
12.0 VDC
15.0 VDC
19.0 VDC
24.0 VDC
48.0 VDC
Max Output Current
17.50 A
14.66 A
12.63 A
10.41 A
5.21 A
Model Number
AHM250PS12T
AHM250PS15T
AHM250PS19T
AHM250PS24T
AHM250PS48T
Notes:
1. For optional input connector retention clip, add suffix ‘-A’ to the model number e.g. AHM250PS24T-A.
2. For 6 pin DIN connector, remove ‘T’ from the end of the model number e.g. AHM250PS24 (DIN connector for medical applications only).
Input Characteristics
Characteristic
Input Voltage - Operating
Input Frequency
Power Factor
Input Current - No Load
Input Current - Full Load
Inrush Current
No Load Input Power
Earth Leakage Current
Input Protection
Minimum
80
47
Typical
Maximum
115/230
264
50/60
63
>0.9
0.09/0.10
2.3/1.2
60-80
120
0.3/0.35
0.5
60/120
200
0.4/0.8
T6.3A/250 V internal fuse in both lines
Units
VAC
Hz
A
A
A
W
μA
mA
Notes & Conditions
Derate <100 VAC (see fig. 1)
EN61000-3-2 class A & D compliant & Energy Star Compliant
115/230 VAC
115/230 VAC
230 VAC cold start, 25 şC
115/230 VAC
115 V 60 Hz/230 V 50 Hz (Typ.), 264 VAC/60 Hz (Max.)
115/230 VAC/400 Hz
Output Characteristics
Characteristic
Output Voltage - V1
Minimum
Typical
12
Maximum
Units
Notes & Conditions
48
VDC
See Models and Ratings table
Output Voltage Adjustment
Minimum Load
%
0
Start Up Delay
300/200
ms
Hold Up Time
15
ms
115/230 VAC full load (see fig.2)
115/230 VAC full load (see fig.3)
Drift
±0.2
%
After 20 min warm up
Line Regulation
±0.5
%
90-264 VAC (50% load)
Load Regulation
±4
%
0f50g100% load.
Transient Response - V1
5
%
Recovery within 1% in less than 500 μs for a 50-75% and 7550% load step
1.5
% pk-pk
Over/Undershoot - V1
3
Ripple & Noise
<1
%
125
Overvoltage Protection
Overload Protection
%
Temperature Coefficient
Overtemperature Protection
20 MHz bandwidth with external circuit (see fig.4-7)
Vnom, Recycle AC to reset
13.2
18
AHM250PS12
16.5
22
AHM250PS15
21.0
28
26.4
33
52.8
59
115.0
175
21.0
29.8
17.6
26.3
15.2
21.5
11.5
17.7
AHM250PS24
5.7
8.9
AHM250PS48
VDC
AHM250PS19
AHM250PS24
AHM250PS48
%
I nom, Auto reset
AHM250PS12
AHM250PS15
A
Short Circuit Protection
2
No user adjustment
A
AHM250PS19
Continuous, trip & restart (hiccup mode)
0.05
%/˚C
°C
Connected to transformer. Auto reset.
xppower.com
Derating Curve
Figure 1
100
Output Power (%)
90
80
70
60
50
40
80
90
100
110
240
250
264
Start Up Delay From AC Turn On
Figure 2
Start up example from AC turn on
(230 VAC, 210 ms)
Hold Up Time From Loss of AC
Figure 3
Hold up example at 250 W load
with 230 VAC input (15 ms)
3
xppower.com
Ripple & Noise
Figure 4
AHM250PS12
Ripple & noise example at 210 W load
with 230 VAC input (50 mV)
Figure 5
AHM250PS24
Ripple & noise example at 250 W load
with 230 VAC input (100 mV)
Figure 6
AHM250PS48
Ripple & noise example at 250 W load
with 230 VAC input (180 mV)
Figure 7
Ripple & noise measurement circuit
10 µF
Electrolytic
4
0.1 µF
Ceramic
xppower.com
Xxxxxxxxxx
General
Specifications
Characteristic
Efficiency
Isolation: Input to Output
Input to Ground
Output to Ground
Minimum
Typical
92
Maximum
4000
1500
500
40
80
Switching Frequency
Units
%
VAC
VAC
VAC
220
150
5.7
Power Density
Mean Time Between Failure
Weight
151
2.1 (1000)
Notes & Conditions
Full load (see fig.8-10)
kHz
PFC stage
DC-DC stage
W/in3
kHrs
lb (g)
MIL-HDBK-217F, Notice 2 +25 °C GB
Average Active Efficiency
GREEN
POWER
Average Active Efficiency
115 V / 60 Hz
230 VAC / 50 Hz
89.16
89.67
90.05
90.71
89.65
90.76
91.85
92.20
91.32
92.46
Characteristic
AHM250PS12
AHM250PS15
AHM250PS19
AHM250PS24
AHM250PS48
Units
Notes & Conditions
%
As per Energy Star Level V test procedure
Efficiency Versus Load
100%
90%
Efficiency
80%
70%
60%
50%
40%
30%
20%
10%
0%
10
20
30
40
50
60
70
80
90
100
Load (%)
Figure 8 - AHM250PS12
100%
100%
90%
90%
80%
80%
70%
70%
60%
60%
50%
50%
40%
40%
30%
30%
20%
20%
10%
10%
0%
0%
10
20
30
40
50
60
70
80
90
100
10
Load (%)
115Vac input
20
30
40
50
60
70
80
90
100
Load (%)
Figure 10 - AHM250PS48
Figure 9 - AHM250PS24
230Vac input
5
xppower.com
Xxxxxxxxxx
Environmental
Characteristic
Minimum
Operating Temperature
Typical
Maximum
0
Units
+60
°C
86
71
°C
Case Temperature
(IEC60601 3rd Edition)
Notes & Conditions
Derate linearly to 60% load at 60 °C from
+40 °C. (See fig.11)
100% load with TAMB +40 °C
80% Load, with TAMB +40 °C
60
Storage Temperature
60% Load Maximum, with TAMB +40 °C
48
°C
-40
+85
°C
5% Load Maximum, with TAMB +40 °C
5
95
%RH
3000
m
Cooling
Convection cooled, see fig.11
Humidity
Operating Altitude
Ingress Protection
Non-condensing
IP22
Shock
3 x 30 g/11 ms shocks in both +ve & -ve
directions along the 3 orthogonal axis,
total 18 shocks.
Vibration
Three axis 5-500 Hz at 2 g x 10 sweeps
Derating Curve
100%
Output Power (%)
Figure 11
60%
0
10
20
Ambient Temperature (ºC)
60
40
Electromagnetic Compatibility - Immunity
Phenomenon
Standard
Test Level
Criteria
EN61204-3
High severity level
as below
Harmonic Current
EN61000-3-2
Class A
ESD
EN61000-4-2
3
A
Radiated
EN61000-4-3
3
A
EFT
EN61000-4-4
3
A
Surges
EN61000-4-5
Installation class 3
A
Conducted
EN61000-4-6
3
A
Magnetic Field
EN61000-4-8
3
A
Low Voltage PSU EMC
EN55024
(EN61000-4-11)
Dips and Interruptions
EN60601-1-2
6
Notes & Conditions
Dip:
30% 500 ms
Dip:
>95% 10 ms
A
A
Int:
>95% 5000 ms
B
Dip:
30% 500 ms
A
230 VAC 100% load, 100 VAC 60% load
Dip:
60% 100 ms
A
230 VAC 100% load, 100 VAC 15% load
Dip:
>95% 10 ms
A
Int.:
>95% 5000 ms
B
xppower.com
Xxxxxxxxxx
Electromagnetic
Compatibility - Emissions
Phenomenon
Conducted
Radiated
Voltage Fluctuations
Standard
Test Level
EN55011/22
Class B
EN55011/22
Class B
Criteria
Notes & Conditions
EN61000-3-3
Safety Agency Approvals
Safety Agency
Safety Standard
Category
CB Report
UL TBA IEC60950-1:2005 Ed 2
Information Technology
UL
UL File TBA UL60950-1 (2007), CSA 22.2 No.60950-1-07 Ed 2
Information Technology
TUV
TUV Certificate # TBA, EN60950-1:2006
Information Technology
Denan Japan
PSE Certificate
CE
LVD
Safety Agency
Safety Standard
Category
CB Report
Certificate #US/18020/UL, IEC60601-1 Ed 3 Including Risk Management
Medical
UL
UL File # E146893, ANSI/AAMI ES 60601-1:2005 & CSA C22.2 No. 60601-1:08
Medical
TUV
TUV Certificate # Z11 11 57396 116, EN60601-1:2006
Medical
Primary to Secondary
2 x MOPP (Means of Patient Protection)
Means of Protection
Category
Primary to Earth
1 x MOPP (Means of Patient Protection)
Secondary to Earth
1 x MOPP (Means of Patient Protection)
Equipment Protection Class
Safety Standard
Notes & Conditions
IEC60950-1:2005 Ed 2 & IEC60601-1 Ed 3
See safety agency conditions of
acceptibility for details
Class I
IEC60601-1 Ed 3
Environmental Legislation
Authority
Location
Date
EISA
US
2007
CEC
Notes & Conditions
California, US
2008
Energy Star
US
2008
Level V
ErP Directive
Europe
2011
Regulation No. 278/2009
7
xppower.com
Xxxxxxxxxx Details
Mechanical
Weight: 2.2 lbs (1000 g)
Dimensions shown in inches (mm).
0.2 (5.0)
Standard Molex 6 pin Connector
8.8 (223.0)
8.58 (218.0)
35.43 (900.0)
0.79 (Ø 20.0)
3.484
(88.50)
1.417 (36.0)
Optional 6 pin DIN connector shown
(for medical applications only)
IEC 320-C14
0.79 (Ø 20.0)
1.457 (37.0)
0.063 (1.60)
4.134 (105.0)
1.417 (36.0)
7.41 (188.40)
2.37 (60.20)
1.97 (50.20)
5.12 (130.20)
Optional Input Connector Retention (-A)
5
4
6
1.40 (35.7)
OUTPUT CONNECTOR
1
2
3
Shell
IEC 320-C14
1
2
3
6
5
4
Pin No.
Molex Type(3)
DIN Type(4)
Pin 1
Return
Output +
Pin 2
Return
Return
Pin 3
Return
Return
Pin 4
Output +
Return
Pin 5
Output +
Output +
Pin 6
Ouput +
Ouput +
Outer Shell
GND*
Notes
1. Dimensions shown in inches (mm). Tolerance is 0.02
(0.5) maximum, except output cable length.
2. Weight 2.2 lbs (1000 g).
3. Molex part no. 50-36-1672 mates with Molex part no.
39-30-1062 or equivalent.
4. Equivalent to DIN45322 (6 pin at 60°) Male.
* Functional earth.
8
28-Oct-13
Similar pages