Keysight DSOX4154A Infiniivision 4000 x-series oscilloscope Datasheet

Keysight Technologies
InfiniiVision 4000 X-Series Oscilloscopes
Data Sheet
02 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Oscilloscope Experience Redefined:
Experience the Speed, Usability, and Integration
Imagine an oscilloscope that sees everything, triggers on
anything, has the ease-of-use of a tablet device…and grows with
your projects.
The Keysight Technologies, Inc. 4000 X-Series oscilloscopes are
engineered for next-generation performance, delivering waveform
update rates 20 times faster than the competition to display
the most signal detail. An industry-leading 12.1-inch capacitive
touch screen with innovative hardware-based InfiniiScan Zone
touch triggering provides the most intuitive interface to get you
answers faster. The 4000 X-Series provides maximum investment
protection with fully upgradable 5-instruments-in-1.
Experience the speed
Anomalies and elusive events are the toughest to debug. The
4000 X-Series oscilloscope redefines your debugging experience
with MegaZoom IV smart memory technology. The industryleading 1-million-waveforms-per-second update rate, means you
see more of your signal behavior and can feel more confident in
your design.
4000 X-Series – Oscilloscope experience
redefined
Key features:
–– Experience the speed:
–– One million waveforms per second update rate
–– MegaZoom IV smart memory technology
–– Standard segmented memory
–– Experience the usability:
–– Industry’s first capacitive touch screen
–– Industry’s largest 12-inch display
–– InfiniiScan Zone touch trigger
–– Experience the integration:
–– Industry’s first 5 instruments in 1
–– Industry’s first fully upgradable including bandwidth
to 1.5 GHz
–– Industry’s leading application solutions
Experience the usability
You may be surprised just how easy it is to use the InfiniiVision
4000 X-Series. A 12.1-inch capacitive touch screen − the
industry’s largest − works just like your favorite tablet or smart
phone, so debugging your devices is faster than ever before.
Innovative InfiniiScan Zone touch triggering makes triggering on
anything a snap. Just draw a box around signals of interest and
the oscilloscope triggers on them. So, if you can see it, you can
trigger on it.
Experience the integration
The 4000 X-Series further redefines your oscilloscope experience
by integrating the capabilities of five instruments in one:
oscilloscope channels, logic channels, digital voltmeter (DVM),
dual-channel WaveGen function/arbitrary waveform generator,
and serial protocol analyzer including USB. All are upgradable,
including bandwidth, for the ultimate investment protection.
Figure 1. MegaZoom IV smart memory technology enables the speed,
usability, and integration.
Figure 2. Take advantage of a new oscilloscope application bundle that
will enable ALL software applications (including DVM and WaveGen) for
one low price (Option DSOX4APPBNDL).
03 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Oscilloscope Experience Redefined:
Experience the Speed, Usability, and Integration (Continued)
Overview of Keysight InfiniiVision X-Series oscilloscopes
InfiniiVision
Analog channels
Digital channels
Bandwidth
6000 X-Series
2 or 4
16 (MSO models or upgrade)
1, 2.5, 4, 6 GHz
Max sample rate
Max memory depth
Max waveform update
rate
Display
Zone touch trigger
Voice control
WaveGen 20-MHz
function/arbitrary
waveform generator
Integrated DVM
Integrated hardware
counter
Search and navigate
Segment memory
Mask/limit test
Serial protocol analysis
options
Advanced analysis
options
Color grade
Histogram
FFT
Multi-domain analysis
Jitter analysis with clock
recovery
Realtime eye diagram
Advanced math
Connectivity
20 GSa/s
4 Mpts
4000 X-Series
2 or 4
16 (MSO models or upgrade)
200, 350, 500 MHz, 1,
1.5 GHz
5 GSa/s
4 Mpts
3000T X-Series
2 or 4
16 (MSO models or upgrade)
100, 200, 350, 500 MHz,
1 GHz
5 GSa/s
4 Mpts
2000 X-Series
2 or 4
8 (MSO models or upgrade)
70, 100, 200 MHz
> 450,000 wfms/sec
> 1,000,000 wfms/sec
> 1,000,000 wfms/sec
12.1-inch, capacitive touch
Standard
Standard
Dual-channel AWG (option)
12.1-inch, capacitive touch
Standard
No
Dual-channel AWG (option)
8.5-inch, capacitive touch
Standard
No
Single-channel AWG (option)
8.5-inch
No
No
Single-channel function
(option)
Option
5 digits (standard), 10 digits
+ totalizer (option)
Standard, lister supported
Standard
Option
I2C/SPI, UART/RS232, CAN/
CAN-dbc/CAN-FD/LIN/LIN
symbolic, SENT, FlexRay, I2S,
MIL-STD1553, ARINC429,
USB 2.0
Power analysis, USB 2.0
signal quality test, HDTV
analysis, FPGA
Standard
Standard
Standard enhanced FFT
Gated FFT
Option
Option
5 digits (standard)
Option
5 digits (standard)
Standard, lister supported
Standard
Option
I2C/SPI, UART/RS232, CAN/
CAN-dbc/CAN-FD/LIN/LIN
symbolic, SENT, FlexRay,
I2S, MIL-STD1553, CXPI,
ARINC429, USB 2.0
Power analysis, USB 2.0
signal quality test, HDTV
analysis, FPGA
No
No
Standard enhanced FFT
Gated FFT
No
Option
5 digits (standard), 8 digits +
totalizer (option)
Standard, lister supported
Standard
Option
I2C/SPI, UART/RS232, CAN/
CAN-dbc/CAN-FD/LIN/LIN
symbolic, SENT, FlexRay,
I2S, MIL-STD1553, CXPI,
ARINC429
Power analysis, HDTV
analysis
Serial only
Option
Option
I2C/SPI, UART, CAN/LIN (will
not operate simultaneously
with digital channels)
No
No
Standard enhanced FFT
Gated FFT
No
No
No
Standard
No
No
Option
Standard, display four
functions simultaneously
Standard USB 2.0, LAN,
video (GPIB option), USB
mouse and keyboard support
No
Standard, display one
function
Standard USB 2.0, LAN,
video (GPIB option), USB
mouse and keyboard support
No
Standard, display one
function
Standard USB 2.0 (LAN/
video/GPIB option), USB
mouse and keyboard support
No
No
2 GSa/s
100 kpts (standard)
1 Mpts (option)
> 50,000 wfms/sec
No
Standard USB 2.0 (LAN/
video/GPIB option), USB
keyboard support
Need more bandwidth, sampling rate, and analysis?
Consider the Infiniium 6000 X-Series
–– 1, 2.5, 4, and 6 GHz
–– 20 GSa/s
–– Ultra-low noise at 1 mV/div
–– 12.1-inch multi-touch capacitive display with gesture support
–– Standard color grade, histogram, and enhanced FFT
–– Optional jitter and real-time eye diagram analysis
See www.keysight.com/find/6000X-Series for more details.
04 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Oscilloscope Experience Redefined:
Experience the Speed
One million waveforms per second update rate
If you can’t see the problem, it is hard to troubleshoot it. With an
industry-leading one million waveforms per second update rate,
the InfiniiVision 4000 X-Series gives you the highest probability of
capturing random and infrequent events that you would miss on
an oscilloscope with a lower waveform update rate.
Powered by MegaZoom IV smart memory technology,
the InfiniiVision 4000 X-Series not only lets you see more
waveforms, but it has the uncompromised ability to find the most
difficult problems in your design. Unlike other oscilloscopes,
uncompromised ability means:
–– Always-fast, responsive operation
–– No slowdown with logic channels on
–– No slowdown with protocol decoding on
–– No slowdown with math functions turned on
–– No slowdown with measurements turned on
Figure 3. The 4000 X-Series captures a glitch occurring once in a million
waveform cycles.
What is waveform update rate and why is it important?
As oscilloscopes acquire data, process it, and plot it to the screen, there is inevitable “dead time,” or the time oscilloscopes miss
signals completely. In general, the faster the waveform update rate, the shorter the dead time. The shorter the dead time, the more
likely an oscilloscope is to capture anomalies and infrequent events. This is why it is critical to select an oscilloscope with a fast
waveform update rate.
Figure 4. Other vendor’s oscilloscope with 50,000 waveforms/second. A long dead time decreases your chances of
capturing infrequent events.
Figure 5. InfiniiVision 4000 X-Series with 1,000,000 waveforms/second. A short dead time increases your chances of
capturing infrequent events.
05 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Oscilloscope Experience Redefined:
Experience the Speed (Continued)
Keysight achieves this industry-leading waveform update rate with
MegaZoom IV smart memory technology
Traditionally, CPU processing was the major bottleneck for oscilloscope waveform
update rate and responsiveness. Typically, the CPU handles interpolations, logic channel
plotting, serial bus decoding, measurements and more, and the waveform update rate
drops dramatically as these features are turned on.
The InfiniiVision 4000 X-Series requires minimum support from a CPU, as most core
operations are handled by Keysight proprietary technology, the MegaZoom IV smart
memory ASIC. MegaZoom includes hardware serial decoders and hardware mask/
limit testing capability, plots analog and digital data directly to the display, supports
GUI operation, and integrates additional instruments like the dual-channel WaveGen
function/arbitrary waveform generator.
MegaZoom IV ASIC
Handware
accelerated
Analog
signals
(Measurements, Search)
Acquisition
memory manager
Plotter
(Segmented memory)
CPU
(Math, I/O, GUI)
16 digital
signals
DRAM
(Acquisition memory)
DRAM
GUI
Serial
decoders
Trigger
(InfiniiScan and
traditional)
Mask
WaveGen
Synthesis
Figure 6. The 4000 X-Series oscilloscopes’ uncompromised responsiveness, speed and waveform update rate is enabled by the MegaZoom IV, smart
memory ASIC. The CPU is not used for core waveform operations.
06 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Oscilloscope Experience Redefined:
Experience the Speed (Continued)
Segmented memory: A smart and efficient way to
capture waveforms
Acquisition memory size is an essential oscilloscope specification
because it determines the amount of data you can capture in a
single acquisition. In general, longer memory is better. However,
no memory is always long enough to capture all the signals you
need, especially when capturing infrequent anomalies, data
bursts, or multiple serial bus packets. Segmented memory
acquisition lets you selectively capture and store important signal
activity without capturing unimportant signal idle time with the
time stamp of each segment relative to the first trigger event.
Segmented memory comes standard in the 4000 X-Series.
Figure 7 shows segmented memory successfully capturing 1,000
events in 3.27274 seconds. Traditional memory architecture
would require 2.7 Gpts of memory to accomplish the same result.
This memory is not available on any scope in the market.
Figure 7. Segmented memory efficiently manages the memory to capture
up to 1,000 segments of interest to you, making it an effective ultradeep memory oscilloscope that can easily capture infrequent events and
anomalies.
Segmented memory + serial decode
Segmented memory works in conjunction with serial protocol
decode. For example, by setting the trigger condition to “CAN
serial bus error,” segmented memory captures and stores only
CAN error packets and stitches together each segment for easy
viewing. You can quickly compare time tags in the event lister to
discover time intervals between errors.
Figure 8. Segmented memory being used in conjunction with serial
decode resulting in maximum insight into serial bus.
07 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Oscilloscope Experience Redefined:
Experience the Speed (Continued)
Mask/limit testing (option)
Whether you are performing pass/fail tests to specified
standards in manufacturing or testing for infrequent signal
anomalies, mask/limit testing can be a valuable productivity tool
(DSOX4MASK). The 4000 X-Series features powerful hardwarebased mask testing and can perform up to 270,000 tests per
second. You can select multiple test criteria, including the ability
to run tests for a specific number of acquisitions, a specified time,
or until detection of a failure.
Figure 9. Mask testing evaluated > 22 M waveforms in just 2 minutes.
Search and navigation
The parametric and serial bus search and navigation feature
comes standard on the 4000 X-Series oscilloscopes. When you
are capturing long, complex waveforms using an oscilloscope’s
deep acquisition memory, manually scrolling through stored
waveform data to find specific events of interest can be slow and
cumbersome. With automatic search and navigation capability,
you can easily set up specific search criteria and then quickly
navigate to “found and marked” events. Available search criteria
include edges, pulse width (time-qualified), rise/fall times (timequalified), runt pulses (time-and level-qualified), frequency peaks
(FFT function, threshold and excursion qualified), and serial bus
frames, packets, and errors.
Figure 10a. The 4000 X-Series was set up to capture data signals with
various rise time edges. Using the search and navigation capability, the
oscilloscope was able to find, mark (white triangles), and quickly navigate
to 16 occurrences of “out of compliance” rise-time edges.
Figure 11. Using the error condition search, the 4000 X-Series quickly
found 5 places with a missing acknowledgment in an I²C serial bus. The
navigation feature moves between the errors and zooms automatically to
show the error packet.
Figure 10b. The 4000 X-Series was set up to capture clock signals for FFT
analysis. Using the search and navigation capability, the scope found,
marked (white triangles) and quickly navigated to the first 11 frequency
peaks occurrences. You can sort it in the order of frequency or amplitude.
8 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Oscilloscope Experience Redefined:
Experience the Usability
Industry’s largest 12.1-inch display
From the start of product development, we designed every aspect
of this oscilloscope for a touch interface. Large, easily touchable
targets on the industry’s largest 12.1-inch display with capacitive
touch screen technology mean operation is quick and natural, just
like your favorite tablet devices.
Capacitive touch screen technology
Capacitive touch screen technology provides enhanced
productivity. Use the alphanumeric pad for quick annotation,
place waveforms or cursors in exact positions and drag docking
panels across the screen to see more measurement information.
The 4000 X-Series offers three ways to access key menus
and features: touch GUI for those that prefer tablet or smart
phone touch interfaces, front panel keypads for the traditional
oscilloscope users, and Keysight pull down menu for users who
prefer Windows-like operations. The 4000 X-Series also offers a
“touch off” button as well as USB mouse and keyboard support.
Figure 12. The industry’s largest 12.1-inch display and capacitive touch
screen technology with large, touchable targets.
Redefine your remote Web control oscilloscope experience. The
4000 X-Series not only supports traditional control via a PC Web
browser, but also supports remote control through popular tablet
devices.
Figure 13. See 10 measurements, cursor information, and the DVM
simultaneously by dragging the desired docking panel to any open area.
Figure 14. Use the Keysight pull-down menu for Windows-like operation.
9 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Oscilloscope Experience Redefined:
Experience the Usability (Continued)
InfiniiScan Zone touch trigger
One of the biggest challenges of using an oscilloscope is setting
up an advanced trigger to isolate a signal of interest. While
advanced triggers are powerful features, InfiniiScan Zone touch
trigger provides a turnkey trigger solution.
You simply observe the signal of interest on the display, and
draw a zone (box) around it. What used to be hours of work can
now take just a few seconds. If you want to move your zones to
another location, just drag them over. The 4000 X-Series can be
set up to easily trigger on one or two zone boxes simultaneously
with either must intersect or must not intersect conditions.
InfiniiScan Zone triggering does not compromise the waveform
update rate; the 4000 X-Series will still maintain an ultra-fast
200,000 waveforms per second or more, even with additional
features turned on. In other words, the oscilloscope that sees
everything can easily trigger on anything.
InfiniiScan Zone touch trigger + segment memory:
a whole new experience
The combination of the industry’s only hardware-based
InfiniiScan Zone touch trigger with the 4000 X-Series’ segment
memory simplifies and enhances your debugging experience.
In Figure 16, the 4000 X-Series has isolated and captured
1,000 metastable signals, showing the critical bit errors over a
32-second time span at 5 GS/s sampling rate in the segment
memory. The segment memory also allows you to overlay all
segments to identify the worst-case signal.
Figure 15. When you see anomalies, all you have to do is
draw a zone box to trigger on them.
Figure 16. Combination of the InfiniiScan Zone touch trigger
and segmented memory.
10 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Oscilloscope Experience Redefined:
Experience the Integration
Investment protection through a fully-upgradable 5-in-1 instrument
The InfiniiVision 4000 X-Series redefines the
oscilloscope experience with unprecedented
integration. This 5-in-1 instrument provides:
–– Oscilloscope
–– 16 digital channels
–– Serial protocol analyzer
–– Dual-channel WaveGen 20 MHz function/
arbitrary waveform generator
–– 3-digit voltmeter
Serial protocol
analysis
Oscilloscope
(analog)
channels
Logic channels
Digital
voltmeter
Dual-channel
WaveGen
Figure 17. The 4000 X-Series provides the capabilities of five instruments
seamlessly integrated into one.
Multi-domain analysis: Time-correlate analog,
digital, and frequency domain signals
Viewing the frequency content of waveforms is greatly simplified
by a touch screen operation. Pop up keypads make inputting
start, stop, span and center frequency easy. And the new problem
solving feature called “gated FFT” lets you time correlate the
analog, digital, and frequency domain to aid in analysis and
debug. In addition, there are new capabilities for peak searching,
max and min hold and averaging of FFTs to increase dynamic
range.
When gated FFT is on, the oscilloscope goes into zoom mode.
The FFT analysis shown in the zoomed (bottom) window is taken
from the period of time indicated by the zoom box in the main
(top) window. In the gated FFT mode, touch and flick the zoom
box through the acquisition to investigate how the FFT analysis
changes over time, correlating the RF phenomenon with the
analog and digital phenomenon.
Figure 18. Gated FFT successfully correlated the hopping of the FSK
modulation with the analog and I2C control command.
11 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Oscilloscope Experience Redefined:
Experience the Integration (Continued)
Mixed signal oscilloscope (MSO): Integrated
16 digital channels
With an additional 16 integrated digital channels, you now have
up to 20 channels of time-correlated triggering, acquisition and
viewing on the same instrument. This is especially important
in today’s embedded designs with sophisticated digital control
circuitry. Unlike other oscilloscopes in this class, you can buy a
2- or 4-channel DSO and enable the 16 digital channels already in
the instrument at any time to make it an MSO. (DSOXPERFMSO)
Figure 19. Digital channels are captured and displayed time-correlated
with analog channels in MSOs or upgraded DSOs.
Serial protocol analysis: Hardware-based serial
protocol decode and triggering
Keysight InfiniiVision Series, including the new 4000 X-Series,
are the only oscilloscopes to use hardware-based serial protocol
decoding. Other vendors’ oscilloscopes use software postprocessing techniques to decode serial packets ⁄ frames, and
therefore have slow waveform and decode capture rates and
could miss critical events and errors due to a long dead-time.
Faster decoding with hardware-based technology enhances the
probability of capturing infrequent serial communication errors.
After capturing serial bus communication, you can easily
perform a search operation based on specific criteria and then
quickly navigate to bytes/frames of serial data that satisfy that
search criteria. The 4000 X-Series can decode two serial buses
simultaneously using hardware-based decoding, and display the
captured data in a time interleaved “lister”display.
Figure 20. Dual serial bus CAN and LIN decode and interleaved “lister“
display.
Serial protocol decoding can be used simultaneously with
segmented memory and InfiniiScan Zone touch triggering.
The 4000 X-Series has: SENT, I2C, SPI, USB 2.0, RS232/UART,
CAN, CAN FD, LIN, FlexRay, CXPI, MIL-STD 1553, ARINC 429, and
I2S. (See page 22)
Figure 21. USB 2.0 trigger, decode and “lister” display.
12 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Oscilloscope Experience Redefined:
Experience the Integration (Continued)
Dual-channel WaveGen 20-MHz function/arbitrary
waveform generator: Industry-exclusive
The 4000 X-Series offers the industry’s only dual-channel,
integrated 20-MHz function/arbitrary waveform generator.
(DSOX4WAVEGEN2) The integrated generator provides stimulus
output of sine, square, ramp, pulse, DC, noise, sine cardinal
(sinc), exponential rise, exponential fall, cardiac, Gaussian pulse
and arbitrary waveforms (AWG) to your device under test. Signal
modulation capability is also available.
With AWG functionality, you can store waveforms from analog
channels or reference memory to the arbitrary memory and
output from WaveGen. Easily create and edit the waveform using
the built-in editor or Keysight’s free BenchLink Waveform Builder
Basic software: www.keysight.com/find/33503.
Figure 22. WaveGen sine wave output with and without added AM
modulation.
With dual channels, you can generate differential signals to:
output arbitrary clock and data signals to simulate serial buses,
create complex modulations (more than the standard modulation
feature), output IQ signals and more. The two channels can be
tracked together as well (identical frequency, amplitude, offset
and duty cycle).
3-digit voltmeter
The 4000 X-Series offers an integrated 3-digit voltmeter
(DVM) and 5-digit frequency counter inside the oscilloscope
(DSOXDVM). The voltmeter operates through the same probes as
the oscilloscope channels. However, the DVM measurements are
de-coupled from the oscilloscope triggering system so that both
the DVM and triggered oscilloscope waveform capture can be
made with the same connection. The voltmeter results are always
displayed, keeping these quick characterization measurements at
your fingertips.
Figure 25. DVM 3-digit
voltage and 5-digit frequency
measurements always at your
fingertips.
Figure 23. WaveGen arbitrary waveform editing screen.
Figure 24. Dual channel WaveGen output of differential arbitrary signals.
Common mode is shown as a math function.
13 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Oscilloscope Experience Redefined:
Other Key Productivity Tools
Power measurements and analysis
When you are working with switching power supplies and power
devices, the power measurements application (DSOX4PWR)
provides a full suite of power measurements and analysis in the
oscilloscope.
Included with the DSOX4PWR is a license for the U1881A
PC-based power analysis software package, which provides
additional offline measurements and report generation.
See www.keysight.com/find/DSOX4PWR for more information.
Figure 26. Power quality measurement, one of many in the power
measurements application.
HDTV video triggering and analysis
Whether you are debugging consumer electronics with HDTV
or characterizing a design, the HDTV measurement application
(DSOX4VID) provides support for a variety of HDTV standards for
triggering and analysis.
See www.keysight.com/find/DSOX4VID for more information.
Figure 27. Triggering on 1080p HDTV signal analysis.
USB 2.0 signal quality analysis
With the USB 2.0 signal quality test option (DSOX4USBSQ),
designers of systems with USB interfaces can now perform
automated signal quality testing. This option supports low-speed,
full-speed, and hi-speed applications (hi-speed tests require
1.5 GHz models). The USB 2.0 signal quality test with HTML pass/
fail report generation includes eye-diagram mask testing, jitter
analysis, EOP bit-width, signaling rate, edge monotonicity, and
rise/fall times; all based on official USB-IF algorithms embedded
in the oscilloscope.
See www.keysight.com/find/DSOX4USBSQ for more information.
Figure 28. Perform automatic signal quality testing on USB 2.0 lowspeed, full-speed, and hi-speed signals.
14 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Oscilloscope Experience Redefined:
Other Key Productivity Tools (Continued)
The Xilinx FPGA dynamic probe
The Xilinx FPGA Dynamic Probe for the 4000 X-Series
(DSOX4FPGAX) allows internal FPGA signal activity to be
correlated to external signal activity. Running on an external
PC, it allows the user to switch between up to 64 signals inside
the FPGA for each debug pin on the outside of the FPGA in
seconds, all while automatically mapping internal signal names
to the oscilloscope channel labels. The DSOX4FPGAX supports
the Virtex-6 series, Virtex-5 series, Virtex-4 series, Virtex-II Pro
series, Virtex-II series, and Spartan-3 series devices.
Figure 29. FPGA dynamic probe routing out internal FPGA signals, to be
captured by the 4000 X-Series, correlated with this external analog sine
wave in the design.
Educator’s oscillocope training kit
Teach your students what an oscilloscope is and how to perform
basic measurements with the Educator’s Oscilloscope Training Kit
(DSOXEDK). This kit includes training tools created specifically
for electrical engineering and physics undergraduate students
and professors. It contains an array of built-in training signals,
a comprehensive oscilloscope lab guide and tutorial written
specifically for undergraduate students and an oscilloscope
fundamentals PowerPoint ® slide set for professors and lab
assistants. Also available is an advanced triggering guide to help
even the most experienced oscilloscope users to get the most out
of their 4000 X-Series oscilloscope.
See www.keysight.com/find/dsoxedk for more information.
Figure 30. DSOXEDK gets both students and experienced users
efficiently using the 4000 X-Series.
15 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Oscilloscope Experience Redefined:
Other Key Productivity Tools (Continued)
Advanced math analysis provides a variety of additional math
functions and comes standard on the 4000 X-Series. Additionally,
math functions can be nested to provide additional insight into
your designs. You can create up to four math functions, with one
resultant math function displayed at a time.
Operators
–– Add, subtract, multiply, divide
Transforms
–– Differentiate, integrate
–– FFT
–– Ax + B
–– Squared, square root
–– Absolute value
–– Common logarithm, natural logarithm
–– Exponential, base 10 exponential
Figure 31. A variety of advanced math functions are standard in the 4000
X-Series.
Filters
–– Low-pass filter, high-pass filter
–– Averaged value, smoothing, envelope
Visualizations
–– Magnify
–– Max hold, min hold
–– Measurement trend
–– Chart logic bus timing, chart logic bus state
Figure 32. Four math functions can be created and nested with one
resultant math function.
36 automatic measurements
Automatic measurements are the essential tool of an
oscilloscope. In order to make quick and efficient measurements,
the 4000 X-Series provides 36 powerful automatic measurements
and can display up to 10 at a time. Measurements can be gated
by auto select, main window, zoom window, or cursors.
Figure 33. Up to 10 automated measurements displayed simultaneously.
Measurements can be gated by cursors.
16 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Oscilloscope Experience Redefined:
Other Key Productivity Tools (Continued)
Reference waveforms
Store up to four waveforms in the scope’s non-volatile reference
waveform memory. Compare reference waveforms with live
waveforms, and perform post analysis and measurements on
stored data. You can also store waveforms on a removable
USB memory device in *.h5 format and recall them back into
oscilloscope’s reference waveform memory later. Save and/
or transfer waveforms to a PC as XY data pairs in a commaseparated values format (*.csv) or store bitmap images and
transfer them to a PC for documentation purposes in a variety of
image formats.
Figure 34. Store and recall up to four reference waveforms.
Powerful probe solutions and compatibility
Get the most out of your 4000 X-Series scope, by using
Keysight’s complete family of innovative probes and accessories
for your application. The 4000 X-Series supports up to four active
probes simultaneously with its full AutoProbe interface. 1
All 4000 X-Series scopes come standard with a 700 MHz
bandwidth, 10 MΩ input passive probe per each channel and
gives you 700 MHz system bandwidth when used in conjunction
with the 4000 X-Series 1 GHz/1.5 GHz models.Also available
is the N2750A InfiniiMode differential probe and N2795A/96A
single-ended active probe for high signal fidelity measurements
without the high price. For ultra low current measurements,
the N2820A Series high-sensitivity current probes are the
best solution in the industry. For power rail measurement, the
N7020A Power Rail Probe provides the unmatched measurement
accuracy.
For the most up-to-date and complete information about
Keysight’s probes and accessories, visit our Web site at
www.keysight.com/find/scope_probes or refer to the InfiniiVision
Probes and Accessories data sheet with the Keysight literature
number 5968-8153EN.
1.
Figure 35. N7020A Power Rail Probe is the industry’s only probe
designed and developed to solve your toughest power integrity problems.
Some restriction may apply. Contact Keysight for more details.
Localized front panel, GUI and help
Operate the oscilloscope in the language most familiar to you.
The graphical user interface, built-in help system, front panel
overlays, and user’s manual are available in 11 languages. During
operation, access the built-in help system just by pressing and
holding any button.
Figure 36. The 4000 X-Series and N2820A Series high-sensitivity current
probe measuring > 500 mA and < 1 mA current simultaneously.
17 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Oscilloscope Experience Redefined:
Other Key Productivity Tools (Continued)
Connectivity and LXI compatibility
Standard USB 2.0 hi-speed host (two on front, one on back) and
device (one on back) ports make PC connectivity easy. Operate
the scope from your PC and save/recall stored waveforms and
setup files via standard LAN (LXI IPv6 Extended Function).
Connect your projector or external monitor through VGA output,
standard with the 4000 X-Series, when sharing and presenting
screen information. An optional external GPIB-to-LAN adapter is
also available (N4865A).
BV0000A BenchVue lets you visualize the 4000 X-Series and
multiple measurements simultaneously. Save time with the ability
to export measurement data to Excel, Word and MATLAB in three
clicks. Monitor and control your 4000 X-Series with a mobile
device from anywhere. Learn more at
www.keysight.com/find/BenchVue.
Figure 37. BV0000A BenchVue.
Virtual front panel
The 4000 X-Series’ innovative capacitive touch screen matches
perfectly with the latest tablet technologies. In addition to the
traditional virtual front panel remote operation through your
favorite PC Web browser, the 4000 X-Series supports remote
oscilloscope control from your tablet devices (and smart phones
with enough resolution). The tablet virtual front panel is identical
to the 4000 X-Series’ touch GUI so you can touch icons, draw
InfiniiScan Zone Touch trigger zones and drag slide panels as if
you are sitting in front of the actual oscilloscope.
Figure 38. Tablet virtual front panel control.
Documentation and e-mail
Annotation becomes a simple task. Bring up the annotation menu
and start editing it using the keypad, and then drag it to the
desired location.
Quick e-mail allows you to e-mail the data you want instantly
to your inbox. Send out the screenshot, waveform data, or even
a USB signal quality test report. This removes the hassle of
connecting your PC to your oscilloscope.
Figure 39(a). Annotation and keypad.
Figure 39(b). E-mail configuration screen.
18 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Oscilloscope Experience Redefined:
Other Key Productivity Tools (Continued)
Infinium Offline oscilloscope analysis software
Keysight’s Infiniium Offline PC-based oscilloscope analysis
software (N8900A) allows you to do additional signal viewing,
analysis and documentation tasks away from your oscilloscope.
Capture waveforms, save to a file, and recall the waveforms into
Infiniium Offline. The application supports a variety of popular
waveform formats from multiple oscilloscope vendors and
includes the following features: navigate, view, measurements,
analyze, view windows, documentation, and optional analysis
upgrades.
For more information, go to: www.keysight.com/find/N8900A.
Figure 40. Infiniium Offline enables a variety of advanced signal analysis
while providing extensive, yet intuitive, waveform documentation.
Keysight spectrum visualizer (AKSV software)
KSV PC-based software package (64997A) connects to the
oscilloscope via USB or Ethernet connection. KSV provides
affordable advanced FFT frequency-domain analysis and
spectrum and spectrogram analysis with an intuitive user
interface familiar to RF engineers.
For more information go to: www.keysight.com/find/KSV_
InfiniiVision.
Figure 41. Waterfall view for KSV spectrogram measurements.
3-year standard warranty, 2-year calibration
interval
Through improved quality processes and rigorous testing, the
Keysight InfiniiVision 4000 X-Series oscilloscopes are able
to perform at guaranteed specification for two years without
calibration, thereby reducing your cost of ownership. These
oscilloscopes come with a standard factory 3-year warranty.
Secure erase
The secure erase feature comes standard with all 4000 X-Series
models. At the press of a button, internal non-volatile memory
is clear of all setup, reference waveforms, and user preferences,
ensuring the highest level of security in compliance with National
Industrial Security Program Operation Manual (NISPOM) Chapter
8 requirements.
19 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Oscilloscope Experience Redefined:
Other Key Productivity Tools (Continued)
High-resolution mode for viewing signal details
To build more confidence in your designs, sometimes you need to
look into more signal detail than you can see with the standard
8-bit vertical resolution of the 4000 X-Series.
High-resolution mode offers additional resolution and insight into
the signal, without requiring a repetitive signal. Using real-time
boxcar averaging, high-resolution mode reduces random noise
and effectively increases vertical resolution, up to 12 bits. For
example, it achieves the 113 uVrms noise floor at 1 mV/div,
100 us/div setting.
Figure 42. Getting 113 uVrms noise floor at 1 mV/div with the high
resolution mode.
Advanced parametric triggering
With today’s more complex signals, you often need to trigger
on complex signal conditions to synchronize the oscilloscope’s
acquisition on specific events. The 4000 X-Series oscilloscope
can trigger on the following conditions: edge, edge then edge,
pulse width (time-qualified), pattern, or, rise/fall time, Nth edge
burst, runt, setup and hold, video, and various serial buses
(optional).
Figure 43. Wide array of advanced parametric trigger modes.
Freeze display
Perhaps you need to share with others an infrequent event you
found. With the “freeze display” feature, you can keep intensity
information on the screen while the oscilloscope is stopped or
before saving a screen shot.
Figure 43:
Figure 44. The “freeze screen” feature keeps the intensity-grading
information while stopping the waveform acquisition.
20 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Oscilloscope Experience Redefined
“Designed for touch.” Industry’s first and
largest 12.1 inch capacitive touch screen
to redefine your oscilloscope experience.
The way an oscilloscope was meant to
be driven, with a designed-for-touch
interface.
The class leading 1.5 GHz upgradeable
bandwidth expands your application
coverage, including USB 2.0 hi-speed
signal integrity testing.
The new InfiniiScan Zone touch trigger, if
you can see it, you can trigger on it by just
drawing a box.
5-in-1 instruments redefines the
integration experiences: oscilloscope
channels, digital channels, serial protocol
analysis, dual-channel WaveGen, and
DVM. All features are fully upgradeable,
including bandwidth.
Industry-leading coverage of serial protocol
including USB 2.0 trigger and decode.
Industry’s first dual-channel WaveGen
function/arbitrary generator now allows
you to generate differential, clock and
data, two channel modulation, and IQ
signals. Modulation of any signal is also
included.
Both USB keyboard and mouse are
supported for additional ease of use.
21 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Industry-leading 1 million waveform per
second update rate minimizes the dead-time
for maximum probability of capturing infrequent
events and anomalies.
Docking panels with the capacitive touch screen add
a new dimension of usability. See setup summary,
automatic measurements, cursor info, DVM, and
navigation pane in any combination, anywhere on the
screen.
Standard advanced math and four cascade-able
math functions enable even the most sophisticated
signal analysis.
Display up to 10 measurements simultaneously,
without compromising other key info. 35 automatic
measurements can be gated by cursors.
Not a touch screen fan? Turn off the touch screen
from a front panel button if desired.
Independent knobs per channel for fast operation.
All front panel knobs are push-able for access to
common controls.
Standard segmented memory powered by
MegaZoom IV smart memory technology provides
intelligent capture of just the signal of interest.
Industry’s first integrated DVM.
Asynchronous from the 4 analog
triggered waveforms.
Simultaneous 1GHz bandwidth
across all 4 channels.
Four AutoProbe (active or current probes)
are supported simultaneously for demanding
applications.
22 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Oscilloscope Experience Redefined:
Configuring Your InfiniiVision 4000 X-Series Oscilloscope
Step 1. Choose your bandwidth and number of channels
InfiniiVision 4000 X-Series scopes oscilloscopes
4022A
200 MHz
Bandwidth 1 (–3 dB)
Calculated rise time (10 to 90%)
≤ 1.75 ns
Input channels
DSOX
2
MSOX
2 + 16
1.
4024A
4
4 + 16
4032A
350 MHz
≤ 1 ns
2
2 + 16
4034A
4
4 + 16
4052A
500 MHz
≤ 700 ps
2
2 + 16
4054A
4
4 + 16
4104A
1 GHz
≤ 450 ps
4
4 + 16
4154A
1.5 GHz
≤ 300 ps
4
4 + 16
For example, if you chose 1 GHz, 4+16 channels, the model number will be MSOX4104A.
Step 2. Tailor your oscilloscope with integrated capabilities and measurement applications to save
time and money. After purchase upgrade model numbers are listed below (values in parentheses are
factory-installed option numbers)
Description
Enables all software options listed below
Serial protocols
Embedded serial triggering and decode (I²C, SPI)
Computer serial triggering and decode (RS232/UART)
USB 2.0 Full/Low Speed serial triggering and decode
USB 2.0 Hi-Speed serial triggering and decode
Automotive serial triggering and decode CAN/CAN-FD/CAN-dbc/LIN/LIN symbolic
FlexRay serial triggering and decode
Single edge nibble transmission triggering and decode (SENT)
CXPI serial triggering and decode
Audio serial triggering and decode (I²S)
Aerospace and defence serial triggering and decode (MIL-STD 1553, ARINC 429)
Measurement applications
Dual-channel WaveGen 20 MHz arbitrary/function generator
3-digit voltmeter (DVM)
Power analysis application
Mask limit testing
Enhanced video/TV application package
FPGA dynamic probe option for Xilinx
USB 2.0 signal quality test option
Productivity tools
Education training kit
Infiniium Offline oscilloscope analysis software
Keysight spectrum visualizer (KSV)
Vector signal analyzer software (version 16 and higher)
BenchVue Software
1. DSOX4USBH is only available for 1 GHz and 1.5 GHz models.
2. USB 2.0 hi-speed tests require 1.5 GHz models.
See page 33 for more detailed upgradability and installation process information.
Model number
DSOX4APPBNDL
DSOX4EMBD(-EMB)
DSOX4COMP(-CMP)
DSOX4USBFL(-USF)
DSOX4USBH(-U2H) 1
DSOX4AUTO(-AMS)
DSOX4FLEX(-FLX)
DSOX4SENSOR(-SEN)
DSOX4CXPI(-CXP)
DSOX4AUDIO(-SND)
DSOX4AERO(-AER)
DSOX4WAVEGEN2(-WAV)
DSOXDVM(-DVM)
DSOX4PWR(-PWR)
DSOX4MASK(-MSK)
DSOX4VID(-VID)
DSOX4FPGAX(-FPX)
DSOX4USBSQ(-U2Q) 2
DSOXEDK(-EDK)
N8900A
64997A
89601B
BV0000A
23 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Oscilloscope Experience Redefined:
Configuring Your InfiniiVision 4000 X-Series Oscilloscope (Continued)
Step 3. Choose your probes – For a complete list of compatible probes, visit
www.keysight.com/find/scope_probes
Probes
N2894A passive probe 700 MHz, 10:1, 10 MΩ
N6450-60001 or N2756A 16 digital channel MSO cable
10076B high-voltage passive probe 250 MHz 4 kV
N2795A active single-ended probe 1-GHz 1-pF 1-MΩ with AutoProbe
N2796A active single-ended probe 2-GHz 1-pF 1-MΩ with AutoProbe
N2750A InfiniiMode differential probe 1.5-GHz 700-fF 200-kΩ with AutoProbe
N2797A extreme temperature active probe 1.5-GHz 1-pF 1-MΩ with AutoProbe
N2790A differential active probe 100 MHz, ± 1.4 kV with AutoProbe
N2791A differential active probe 25 MHz, ± 700 V
N2792A differential active probe 200 MHz, ± 20 V
N2793A differential active probe 800 MHz, ± 15 V
1147B AC/DC current probe 50 MHz 15 A with AutoProbe
N2893A AC/DC current probe 100 MHz 15 A with AutoProbe
N2820A 2-channel high-sensitivity current probe 50 uA to 5 A
N7020A power rail probe 2-GHz, 1:1, 50 kΩ, ± 24 V offset range
N2805A high voltage differential probe, 200 MHz, ± 100 V (DC + peak AC), 50:1, 4-MΩ, 4 pF
N2804A high voltage differential probe, 300 MHz, ± 300 V (DC + peak AC), 100:1, 4-MΩ, 4 pF
4000 X-Series
Included standard. 1 per channel
Included on MSOX models and DSOXPERFMSO
Optional
Optional
Optional
Optional
Optional
Optional
Optional
Optional
Optional
Optional
Optional
Optional
Optional
Optional
Optional
Step 4. Choose your accessories
Recommended accessories
GPIB-to-LAN external adapter
Rack mount kit
Soft carrying case
Hard copy manual
4000 X-Series
N4865A
N2763A
N2733B
N6455A
Step 5. Calibration and warranty plans
Calibration and warranties
D/MSOX3000T-A6J
ANSI Z540-1-1994 calibration
D/MSOX3000T-AMG ISO17025 compliant calibration with accreditation
R-51B-001-5F
Warranty Assurance Plan - Return to Keysight - 5 years
Optional
Optional
Optional
24 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Oscilloscope Experience Redefined:
InfiniiVision 4000 X-Series Performance Characteristics
DSO and MSO 4000 X-Series oscilloscopes
4000 X-Series specification overview
4022A
4024A
4032A
4034A
4052A
4054A
4104A
4154A
200 MHz
350 MHz
500 MHz
1 GHz
1.5 GHz 3
Bandwidth 1 (–3 dB)
All-channel real-time bandwidth
200 MHz
350 MHz
500 MHz
1 GHz
1 GHz
Calculated rise time (10 to 90%)
≤ 1.75 ns
≤ 1 ns
≤ 700 ps
≤ 450 ps
≤ 300 ps
Input channels
DSOX
2
4
2
4
2
4
4
4
MSOX
2 + 16
4 + 16
2 + 16
4 + 16
2 + 16
4 + 16
4 + 16
4 + 16
Maximum sample rate
5 GSa/s half channel, 2.5 GSa/s all channel
Maximum memory depth
Standard 4 Mpts, standard segmented memory
Display size and type
12.1-inch high-definition capacitive touch display
Waveform update rate
> 1 million waveforms per second
System bandwidth with N2894A standard 200 MHz
200 MHz
350 MHz
350 MHz
500 MHz
500 MHz
700 MHz
700 MHz
passive probe
Vertical system analog channels
Hardware bandwidth limits
Approximately 20 MHz (selectable)
Input coupling
AC, DC
Input impedance
Selectable: 1 MΩ ± 1% (16 pF), 50 Ω ± 1.5%
Input sensitivity range
200 MHz ~ 500 MHz models: 1 mV/div to 5 V/div 2 (1 MΩ and 50 Ω)
1 and 1.5 GHz models: 1 mV/div to 5 V/div 2 (1 MΩ), 1 mV/div to 1 V/div (50 Ω)
Vertical resolution
8 bits (measurement resolution is 12 bits with averaging)
Maximum input voltage
1 MΩ
300 Vrms, 400 Vpk; transient overvoltage 1.6 kVpk. Frequency derating (assumes sine
wave input): 400 Vpk until 40 kHz. Then derates at 20 db/dec until 6 Vpk
50 Ω
50 Ω: ≤ 5 Vrms max
± 2.0% full scale 2
DC vertical gain accuracy 1
DC vertical offset accuracy
± 0.1 div ± 2 mV ± 1% of offset setting
Channel-to-channel isolation
200 MHz~1 GHz
≥ 40 dB from DC to maximum specified bandwidth of each model
1.5 GHz
≥ 40 dB from DC to 1 GHz, ≥ 35 dB from 1 to 1.5 GHz
Offset range
± 5 V (< 10 mV/div), ± 20 V (10 to 200 mV/div), ± 75 V (> 200 mV/div)
Vertical system digital channels
Digital input channels
16 digital (D0 to D15. Pod 1: D7 ~ D0, Pod 2: D15 ~ D8)
Thresholds
Threshold per pod
Threshold selections
TTL (+1.4 V), 5 V CMOS (+2.5 V), ECL (–1.3 V), user-defined (selectable by pod)
User-defined threshold range
± 8.0 V in 10 mV steps
Maximum input voltage
± 40 V peak CAT I; transient overvoltage 800 Vpk
± (100 mV + 3% of threshold setting)
Threshold accuracy 1
Maximum input dynamic range
± 10 V about threshold
Minimum voltage swing
500 mVpp
Input impedance
100 kΩ ± 2% at probe tip
Input capacitance
~8 pF
Vertical resolution
1 bit
1.
Denotes warranted specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ± 10 °C from firmware calibration
temperature.
2. 1 mV/div and 2 mV/div is a magnification of 4 mV/div setting. For vertical accuracy calculations, use full scale of 32 mV for 1 mV/div and 2 mV/div sensitivity
setting.
3. 1.5 GHz real time bandwidth in half-channel mode or full channel equivalent time mode.
25 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Oscilloscope Experience Redefined:
InfiniiVision 4000 X-Series Performance Characteristics (Continued)
Horizontal system analog channels
4022A
Time base range
Time base accuracy 1
Time base delay time range
Pre-trigger
Post-trigger
Channel-to-channel deskew range
Δ Time accuracy (using cursors)
Modes
XY
Horizontal system digital channels
Minimum detectable pulse width
Channel-to-channel skew
Acquisition system
4034A
4052A
4054A
1 ns/div to 50 s/div
4104A
4154A
500 ps/div to 50 s/div
± 10 ppm
Greater of 1 screen width or 200 μs (400 μs in interleaving mode)
1 to 500 s
± 100 ns
± 0.001% of reading ± 0.16% screen width ± 30pS
Main, zoom, roll, XY
On channels 1 and 2 only. Z Blanking on Ext Trigger Input, 1.4 V threshold
Bandwidth: Maximum bandwidth. Phase error at 1 MHz: < 0.5 degree
Time base: 200 ns/div to 50 ms/div
2 ns
2 ns (typical); 3 ns (maximum)
Maximum analog channels sample rate
Analog channels equivalent sample rate
Maximum analog channels record length
Maximum digital channels sample rate
Maximum digital channels record length
Modes
Normal
Peak detect
Averaging
High
resolution
Segmented
Roll
Equivalent
time
1.
4024A
4032A
2 ns/div to 50 s/div
4022A
4024A
4032A
4034A
4052A
4054A
4104A
4154A
5 GSa/s half channel interleaved, 2.5 GSa/s all channels
N/A
128 Gsa/s
4 Mpts half channel interleaved, 2 Mpts all channel
1.25 GSa/s
2 Mpts (with digital channels only)
Default mode
Capture glitches as narrow as 200 ps at all time base settings
Selectable from 2, 4, 8, 16, 64, ... to 65,536
Real-time boxcar averaging reduces random noise and effectively increases vertical resolution
–– 12 bits: ≥ 50 μs/div
–– 11 bits: ≥ 20 μs/div
–– 10 bits: ≥ 10 us/div
–– 9 bits: ≥ 5 us/div
Segmented memory optimizes available memory for data streams that have long dead times
between activity. Maximum segments = 1000. Re-arm time = 1 μs (minimum time between trigger
events). Re-arm time when used with the InfiniiScan Zone = 65 us or faster (typical)
Displays the waveform moving across the screen from right to left. Available at the time base
50 ms/div or slower
1 GHz and 1.5 GHz models only. 7.8 ps fine interpolator resolution yields a maximum effective
sample rate of 128 GSa/s
Denotes warranted specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ± 10 °C from firmware calibration
temperature.
26 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Oscilloscope Experience Redefined:
InfiniiVision 4000 X-Series Performance Characteristics (Continued)
Trigger system
Trigger sources
Trigger modes
Trigger coupling
Trigger holdoff range
Trigger sensitivity (internal) 1
Trigger sensitivity (external) 1
Trigger level range
1.
Analog channel (1 ~ 4), digital channel (D0 ~ D15), line, external, WaveGen (1, 2, or Mod)
(FM/FSK)
Normal
Requires trigger event for oscilloscope to trigger
Auto
Triggers automatically in absence of trigger event
Single
Front panel button that triggers only once on a trigger event. Press [Single] button again
for oscilloscope to find another trigger event, or press [Run] front-panel button to trigger
continuously in either auto or normal mode
Force
Front panel button that forces a trigger
DC
DC coupled trigger
AC
AC coupled trigger, cutoff frequency: < 10 Hz (internal); < 50 Hz (external)
HF reject
High-frequency reject, cutoff frequency ~ 50 kHz
LF reject
Low-frequency reject, cutoff frequency ~ 50 kHz
Noise reject
Adds hysteresis to the trigger circuitry. Selectable OFF or ON, decreases sensitivity 2x
40 ns to 10.00 s
200 MHz ~ 1 GHz < 10 mV/div: greater of 1 div or 5 mV; ≥ 10 mV/div: 0.6 div
1.5 GHz
DC to 1 GHz: < 10 mV/div: Greater of 1 div or 5 mV; ≥ 10 mV/div: 0.6 div
1 to 1.5 GHz: < 10 mV/div: Greater of 1.5 div or 5 mV; ≥ 10 mV/div: 1.0 div
± 1.6 V
40 mVpp DC to 100 MHz, 70 mVpp 100 to 200 MHz
±8V
200 mVpp DC to 100 MHz, 350 mVpp 100 to 200 MHz
Any channel
± 6 div from center screen
External
8 V range = ± 8 V, 1.6 V range = ± 1.6 V
Denotes warranted specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ± 10 °C from firmware calibration
temperature.
27 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Oscilloscope Experience Redefined:
InfiniiVision 4000 X-Series Performance Characteristics (Continued)
Trigger type selections
InfiniiScan Zone (HW zone qualifier)
Edge
Edge then edge (B trigger)
Pulse width
Pattern
Or
Rise/fall time
Nth edge burst
Runt
Setup and hold
Video
Enhanced video (HDTV) (Option)
ARINC429 (Option)
CAN (Option)
FlexRay (Option)
I²C (Option)
I²S (Option)
LIN (Option)
Trigger on user-defined zones drawn on the display. Applies to one analog channel at a time. Specify zones as
either “must intersect” or “must not intersect.” Up to two zones. > 200,000 wfm/sec update rate.
Supported modes: normal, peak detect, high resolution.
Also works simultaneously with the serial decodes and mask/limit test.
Trigger on a rising, falling, alternating, or either edge of analog channels, digital channels, or an external signal.
Trigger on a rising or falling of a line signal.
Arm on a selected edge, wait a specified time, then trigger on a specified count of another selected edge.
Minimum 4 ns
Trigger on a pulse on a selected channel, whose time duration is less than a value, greater than a value, or inside a
time range
–– Minimum duration setting: 2 ns (500 MHz, 1 GHz, 1.5 GHz), 4 ns (350 MHz), 6 ns (200 MHz)
–– Maximum duration setting: 10 s
–– Range minimum: 10 ns
Trigger when a specified pattern of high, low, and don’t-care levels on any combination of analog, digital, or
trigger channels is [entered | exited]. Pattern must have stabilized for a minimum of 2 ns to qualify as a valid
trigger condition.
–– Minimum duration setting: 2 ns (500 MHz, 1 GHz, 1.5 GHz), 4 ns (350 MHz), 6 ns (200 MHz)
–– Maximum duration setting: 10 s
Trigger on any selected edges from available sources (analog and digital channels only up to 500 MHz)
Trigger on rise-time or fall-time edge speed violations (< or >) based on user-selectable threshold. Select from
(< or >) and time settings range between
–– Minimum: 1 ns (500 MHz, 1 GHz, 1.5 GHz model), 2 ns (350 MHz model), 3 ns (200 MHz model)
–– Maximum: 10 s
Trigger on the Nth (1 to 65535) edge of a pulse burst. Specify idle time (10 ns to 10 s) for framing
Trigger on a positive runt pulse that fails to exceed a high-level threshold. Trigger on a negative runt pulse that
fails to exceed a low-level threshold. Trigger on either polarity runt pulse based on two threshold settings. Runt
triggering can also be time-qualified (< or >) with a minimum time setting of 2 ~ 6 ns and maximum time setting of
10 s
–– Minimum time setting: 2 ns (500 MHz, 1 GHz, 1.5 GHz), 4 ns (350 MHz), 6 ns (200 MHz)
Trigger on setup/hold violations. Setup time can be set from –7 to 10 s. Hold time can be set from 0 s to 10 ns.
Minimum window (setup time + hold time) must be 3 ns or greater
Trigger on all lines or individual lines, odd/even or all fields from composite video, or broadcast standards (NTSC,
PAL, SECAM, PAM-M)
Trigger on lines and fields of enhanced and HDTV standards (480p/60, 567p/50, 720p/50, 720p/60, 1080p/24,
1080p/25, 1080p/30, 1080p/50, 1080p/60, 1080i/50, 1080i/60)
Trigger and decode on ARINC429 data. Trigger on word start/stop, label, label + bits, label range, error conditions
(parity, word, gap, word or gap, all), all bits (eye), all 0 bits, all 1 bits
Trigger on CAN (controller area network) version 2.0A,2.0B, and CAN-FD (Flexible Data-rate) signals. Trigger on
the start of frame (SOF), the end of frame (EOF), data frame ID, data frame ID and data (non-FD), data frame ID
and data (FD), remote frame ID, remote or data frame ID, error frame, acknowledge error, from error, stuff error,
CRC error, spec error (ack or form or stuff or CRC), all errors, BRS Bit (FD), CRC delimiter bit (FD), ESI bit active
(FD), ESI bit passive (FD), overload frame., message, message and signal (non-FD), message and signal (FD, first
8 bytes only)
Trigger on frame ID or specific error condition, along with cycle-base and repetition-cycle filtering. Can also
trigger on specific events such as BSS, TSS, FES, and wake up
Trigger at a start/stop condition or user defined frame with address and/or data values. Also trigger on missing
acknowledge, address with no acq, restart, EEPROM read, and 10-bit write
Trigger on 2’s complement data of audio left channel or right channel (=, ≠, <, >, > <, < >, increasing value, or
decreasing value)
Trigger on LIN (local interconnect network) sync break, sync frame ID, frame ID and data, parity error, or
checksum error
28 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Oscilloscope Experience Redefined:
InfiniiVision 4000 X-Series Performance Characteristics (Continued)
Trigger type selections
CXPI (optional)
MIL-STD1553 (Option)
SPI (Option)
UART/RS232/422/485 (Option)
USB (Option)
SENT (optional)
Trigger on the start of frame (SOF), the end of frame (EOF), PTYPE, frame ID, data and info frame ID, data and info
frame ID (long frame), CRC field error, parity error, inter-byte space error, inter-frame space error, framing error,
data length error, sample error, all errors, sleep frame, wakeup pulse
Trigger on MIL-STD 1553 signals on data word start/stop, command/status start/stop, RTA, RTA + 11 bits, and
error conditions (parity, sync, Manchester)
Trigger on SPI (serial peripheral interface) data pattern during a specific framing period. Supports positive and
negative chip select framing as well as clock Idle framing and user-specified number of bits per frame. Supports
MOSI and MISO data
Trigger on Rx or Tx start bit, stop bit, data content, or parity error
Trigger on start of packet (SOP), end of packet (EOP), suspend 3, resume 3, reset 3, packets (token, data,
handshake, or special), and errors (PID, CRC5, CRC16, glitch, bit stuff 3, SE1 3). Supports USB 2.0 low speed, full
speed, and hi-speed (hi-speed is supported on 1 GHz and 1.5 GHz models only)
Trigger and decode on SENT bus. start of fast channel message, start of slow channel message, fast channel SC
and data, slow channel message ID, slow channel message ID and data, tolerance violation, fast channel CRC
error, slow channel CRC error, all CRC errors, pulse period error, successive sync pulses error (1/64)
Waveform measurements and cursors
DC vertical accuracy/cursors 2
Cursors
Automatic measurements
Voltage
Time
Count
Mixed
Automatic measurement logging
Counter
Source
Resolution
Max frequency
Mask limit test option
Waveform math
Number of math functions
Arithmetic
FFT
1.
Single cursor accuracy: ± [DC vertical gain accuracy + DC vertical offset accuracy + 0.21% full
scale]
Dual cursor accuracy: + [DC vertical gain accuracy + 0.42% full scale] 1
2 pairs of XY cursors
Automatic measurement of positions, ∆X, 1/∆X, ∆Y, and ∆Y/∆X
Measurements continuously updated with statistics. Cursors track last selected measurement.
Select up to four measurements from the list below:
Peak-to-peak, maximum, minimum, amplitude, top, base, overshoot, pre-shoot, average- N
cycles, average- full screen, DC RMS- N cycles, DC RMS- full screen, AC RMS- N cycles, AC
RMS- full screen (standard deviation), ratio (RMS1/RMS2)
Period, frequency, counter, + width, - width, burst width, duty cycle, bit rate, rise time, fall time,
delay, phase, X at min Y, X at max Y
Positive pulse count, negative pulse count, rising edge count, falling edge count
Area- N cycles, area- full screen
Available via BenchVue
Built-in frequency counter
Any analog or digital channel
5 digits. Up to 8 digits with an external reference 10 MHz input
Bandwidth of oscilloscope
Standard mask/limit test capability that provides easy, fast pass/fail comparison of a signal
under test to a predefined mask template or auto-mask template. Predefined mask templates or
edits to an auto-mask template can be made via a simple text editor. > 270,000 mask tests per
second (waveform update rate)
Four (display one at a time)
Add, subtract, multiply, divide, differentiate, integrate, FFT, Ax + B, squared, square root, absolute, common log,
natural log, exponential, base 10 exponential, LP filter, HP filter, averaged value, smoothing, envelope, magnify,
max hold, min hold, measurement trend, chart logic bus (timing or state)
Up to 64 kpts resolution. FFT window types: Hanning, flat top, rectangular, Blackman-Harris
Denotes warranted specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ± 10 °C from firmware calibration
temperature.
2. 1 mV/div and 2 mV/div is a magnification of 4 mV/div setting. For vertical accuracy calculations, use full scale of 32 mV for 1 mV/div and 2 mV/div sensitivity
setting.
3. Suspend, resume, reset, bit stuff error, and SE1 error are USB 2.0 low and full speed only.
29 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Oscilloscope Experience Redefined:
InfiniiVision 4000 X-Series Performance Characteristics (Continued)
Display characteristics
Display
Resolution
Graticules
Format
Maximum waveform update rate
Persistence
Intensity gradation
Connectivity
USB 2.0 hi-speed host port
USB 2.0 hi-speed device port
LAN port
Web remote control
Video out port
GPIB port
10 MHz out/in
Trigger out
12.1-inch capacitive touch/gesture enabled color TFT LCD
800 (H) x 600 (V) pixel format (screen area)
8 vertical divisions by 10 horizontal divisions with intensity controls
YT, XY and Roll
> 1,000,000 wfm/s
Off, infinite, variable persistence (100 ms to 60 s)
64 intensity levels
USB 2.0 hi-speed host ports x3, two front and one real panel. Supports memory devices, printers, keyboards and mice
One USB 2.0 hi-speed device port on rear panel. USB Test and Measurement Class (USBTMC) compatible
10/100Base-T port on rear panel. LXI IPv6 extended function
VNC Web interface (via major Web browsers)
SVGA out on rear panel. Connect oscilloscope display to an external monitor or projector
N4865A GPIB-to-LAN adapter (optional)
BNC connector on the rear panel. Supported modes: Off, 10 MHz out, or reference signal mode (10 MHz in)
BNC connector on the rear panel. Supported modes: triggers, mask, waveform generator 1 sync pulse, and waveform
generator 2 sync pulse
Dual-channel WaveGen built-in function/arbitary waveform generator (specifications are typical)
WaveGen outputs
Two (front-panel BNC connectors)
Both waveform generator outputs can be frequency tracked, amplitude tracked, or completely tracked.* A generator’s
output can be inverted to create a differential signal
Waveforms
Sine, square, ramp, pulse, DC, noise, sine cardinal (sinc), exponential rise, exponential fall, cardiac, Gaussian pulse, and
arbitrary
Modulation
Modulation is available on channel 1 only. Modulation is not available when tracking mode is enabled.
Modulation types: AM, FM, FSK
Carrier waveforms: Sine, Ramp, Sine Cardinal, Exponential Rise, Exponential Fall, and Cardiac.Modulation Source:
Internal (no external modulation capability)
AM:
Modulation: Sine, Square, Ramp
Modulation frequency: 1 Hz to 20 kHz
Depth: 0% to 100%
FM:
Modulation: Sine, Square, Ramp
Modulation frequency: 1 Hz to 20 kHz
Minimum carrier frequency: 10 Hz
Deviation: 1 Hz to carrier frequency or (2e12 /carrier frequency), whichever is smaller
FSK:
Modulation: 50% duty cycle square wave
FSK rate: 1 Hz to 20 kHz
Hop frequency: 2 x FSK rate to 10 MHz
1.
Only the following combination of wave shapes can be frequency tracked or completely tracked:
1) Sine, ramp, sine cardinal, cardiac, and Gaussian pulse.
2) Square wave and pulse.
3) Exponential rise and exponetial fall.
4) Arbitrary.
30 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Oscilloscope Experience Redefined:
InfiniiVision 4000 X-Series Performance Characteristics (Continued)
Dual-channel WaveGen built-in function/arbitary waveform generator (specifications are typical)
Sine
Frequency range
0.1 Hz to 20 MHz
Amplitude flatness
± 0.5 dB (relative to 1 kHz)
Harmonic distortion
–40 dBc
Spurious (non harmonics)
–40 dBc
Total harmonic distortion
1%
SNR (50-Ω load, 500-MHz BW)
40 dB (Vpp ≥ 0.1 V); 30 dB (Vpp < 0.1 V)
Square wave/pulse
Frequency range
0.1 Hz to 20 MHz
Duty cycle
20 to 80%
Duty cycle resolution
Larger of 1% or 10 ns
Rise/fall time
19 ns (10 to 90%)
Overshoot
< 2%
Asymmetry (at 50% DC)
± 1% ± 5ns
Jitter (TIE RMS)
500 ps
Ramp/triangle wave
Frequency range
0.1 Hz to 200 kHz
Linearity
1%
Variable symmetry
0 to 100%
Symmetry resolution
1%
Pulse
Frequency range
0.1 Hz to 10 MHz
Pulse width
20 ns minimum
Pulse with resolution
10 ns
Edge time
Fixed at 19 ns (not variable)
Overshoot
< 2%
Noise
Bandwidth
20 MHz typical
Sine cardinal (Sinc)
Frequency range
0.1 Hz to 1.0 MHz
Exponential rise/fall
Frequency range
0.1 Hz to 5.0 MHz
Cardiac
Frequency range
0.1 Hz to 200.0 kHz
Gaussian pulse
Frequency range
0.1 Hz to 5.0 MHz
Arbitrary
Waveform length
1 to 8,192 points
Amplitude resolution
10 bits (including sign bit) 2
Repetition rate
0.1 Hz to 12 MHz
Sample rate
100 MSa/s
Filter bandwidth
20 MHz
Frequency
Sine wave and ramp accuracy
130 ppm (frequency < 10 kHz)
50 ppm (frequency > 10 kHz)
Square wave and pulse accuracy
[50 + frequency/200] ppm (frequency < 25 kHz)
50 ppm (frequency ≥ 25 kHz)
Resolution
0.1 Hz or 4 digits, whichever is larger
Amplitude
Range: Minimum
20 mVpp if |offset| ≤ 0.5 Vpp into Hi-Z 1
10 mVpp if |offset| ≤ 0.5 Vpp into 50 Ω 1
Range: Maximum
10 Vpp except, 9 Vpp if Sinc or Cardiac, 7.5 Vpp if Gaussian
pulse into Hi-Z; 5 Vpp/4.5 Vpp into 50 Ω
Resolution
100 μV or 3 digits, whichever is higher
Accuracy
1.5% (frequency = 1 kHz)
1.
2.
Sinc, cardiac and Gaussian pulse: ± 1.25 V into Hi-Z; ± 625 mV into 50 Ω.
Full resolution is not available at output due to internal attenuator stepping.
31 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Oscilloscope Experience Redefined:
InfiniiVision 4000 X-Series Performance Characteristics (Continued)
Dual channel WaveGen - built-in function/arbitary waveform generator
DC offset
Range
Main output
Trigger output
Resolution
Accuracy (waveform modes)
Accuracy (DC mode)
Impedance
Isolation
Protection
Trigger output available on trig-out BNC
± 5 V into Hi-Z, except ± 4 V if sine wave, ± 2.5 V if sinc, cardiac, or
Gaussian pulse into Hi-Z
± 2.5 V into Hi-Z, except ± 2 V if sine wave, ± 1.25 V if sinc, cardiac, or
Gaussian pulse into 50 Ω
Larger of 250 uV or 3 digits
± 1.5% of offset setting ± 1% of amplitude ± 1 mV
± 1.5% of offset setting ± 3 mV
50 Ω typical
Not available, main output BNC is grounded
Overload automatically disables output
Digital voltmeter (specifications are typical)
Source
Analog channels only (1-4)
Functions
ACrms, DC, DCrms, frequency
Resolution
ACV/DCV: 3 digits
Counter frequency: 5.5 digits
Measuring rate
100 times/second
Autoranging
Automatic adjustment of vertical amplification to maximize the dynamic range of measurements
Range meter
Graphical display of most recent measurement, plus extrema over the previous 3 seconds
General and environmental characteristics
Power line consumption
Power voltage range
Temperature
Operating
Non-operating
Humidity
Operating
Non-operating
Altitude
Operating
Non-operating
Electromagnetic compatibility
Safety
Vibration
Maximum 120 W
100 to 120 V, 50/60/400 Hz; 100 to 240 V, 50/60 Hz ± 10% auto ranging
0 to +50 °C
–30 to +70 °C
50% to 95% RH at 40 °C for 5 days
90% RH at 65 °C for 24 hr
Up to 3,000 m
Up to 15,300 m
Meets EMC Directive (2004/108/EC), meets or exceeds IEC 61326-1:2005/EN
61326-1:2006 Group 1 Class A requirement
CISPR 11/EN 55011
IEC 61000-4-2/EN 61000-4-2
IEC 61000-4-3/EN 61000-4-3
IEC 61000-4-4/EN 61000-4-4
IEC 61000-4-5/EN 61000-4-5
IEC 61000-4-6/EN 61000-4-6
IEC 61000-4-11/EN 61000-4-11
Canada: ICES-001:2004
Australia/New Zealand: AS/NZS
UL61010-1 2nd edition, CAN/CSA22.2 No. 61010-1-04
Meets IEC60068-2-6 and MIL-PRF-28800; class 3 random
32 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Oscilloscope Experience Redefined:
InfiniiVision 4000 X-Series Performance Characteristics (Continued)
General and environmental characteristics
Shock
Meets IEC 60068-2-27 and MIL-PRF-28800; class 3 random; (operating 30 g, ½ sine. 11 ms duration,
3 shocks/axis along major axis, total of 18 shocks)
Dimensions (W x H x D)
454 mm x 275 mm x 156 mm
Weight
Net: 6.3 kg (13.9 lbs), shipping: 11.4 kg (25 lbs)
Kensington style lock
Rear-panel security slot connects to standard Kensington-style lock
Nonvolatile storage
Reference waveform display
Save formats
Max USB flash drive size
Set ups without USB flash drive
Set ups with USB flash drive
Included standard with oscilloscope
Factory warranty
Calibration
Probes
Localized interface
Power cord
Front panel protection
Documentation
4 internal waveforms or USB thumb drive
Setup (*.scp), 8- or 24-bit bitmap image (*.bmp), PNG 24-bit image (*.png), CSV data (*.csv), ASCII XY data
(*.csv), binary data (*.bin), lister data (*.csv), reference waveform data (*.h5), multi-channel waveform data (*.h5),
mask (*.mask), arbitrary waveform data (*.csv), power harmonics data (*.csv), USB signal quality (*.html & *.bmp)
Supports industry-standard flash drives
10 internal setups
Limited by size of USB drive
3-year warranty (90 days for unserialized accessories such as passive probes)
Certificate of Calibration, 2-year calibration interval
One per channel N2894A 700 MHz passive probe (10:1 attenuation)
N6450-60001 16-digital-channel MSO cable (1 per oscilloscope included on all MSO models and
DSOXPERFMSO)
English, Chinese (simplfied and traditional), French, German, Italian, Japanese, Korean, Portuguese, Russian, and
Spanish localized front panel overlays, interface, and built-in help system
Localized power cord
Front panel cover
CD containing localized user’s guide, service guide, and programmer’s manual
For MET/CAL procedures, visit the Cal Lab Solutions website: http://www.callabsolutions.com/products/Keysight/.
Related literature
Publication title
Oscilloscope Memory Architectures – Why All Acquisition Memory is Not Created Equal - Application Note
Triggering on Infrequent Anomalies and Complex Signals using Zone Trigger - Application Note
Switch Mode Power Supply Measurements - Application Note
Characterizing Hi-Speed USB 2.0 Serial Buses In Embedded Designs - Data Sheet
Publication number
5991-1024EN
5991-1107EN
5991-1117EN
5991-1148EN
33 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Oscilloscope Experience Redefined:
License-only Bandwidth Upgrades and Measurement Applications
Model number
DSOX4EMBD
DSOX4COMP
4000 X-Series bandwidth upgrade models
Model number
Description
DSOX4B3T52U
350 MHz ~ 500 MHz, 2 ch
DSOX4B3T54U
350 MHz ~ 500 MHz, 4 ch
Type
License only
License only
Description
Embedded serial triggering and analysis (I²C, SPI)
Computer serial triggering and analysis
(RS232/UART)
DSOX4USBFL
USB 2.0 Full/Low Speed serial decode and
triggering
DSOX4USBH 1
USB 2.0 Hi-Speed serial decode and triggering
DSOX4USBSQ
USB 2.0 signal quality test
DSOX4AUTO
Automotive serial triggering and analysis
(CAN/CAN-dbc/LIN)
DSOX4FLEX
FlexRay serial triggering and analysis
DSOX4CXPI
CXPI serial triggering and decode
DSOX4AUDIO
Audio serial triggering and analysis (I²S)
DSOX4AERO
Aerospace and defence serial triggering and decode
(MIL-STD 1553 and ARINC 429)
DSOX4WAVEGEN2 Dual-channel WaveGen 20 MHz arbitrary/function
generator
DSOXDVM
3-digit voltmeter (DVM)
DSOX4PWR
Power analysis application
DSOX4MASK
Mask Limit Testing
DSOX4VID
Enhanced video/TV application package
DSOX4FPGAX
FPGA dynamic probe option for Xilinx
DSOXEDK
Education and training kit
DSOXPERFMSO
4000 X-Series oscilloscope MSO upgrade
(N6450-60001 MSO cable is shipped separately)
1.
DSOX4USBH is available only for 1 GHz and 1.5 GHz models.
Process description
1
Place order for a license-only upgrade or measurement application product with a Keysight sales partner.
For measurement applications, you will receive a paper or electronic .pdf entitlement certificate. For bandwidth upgrades only, you will receive a
2
stick-on label document indicating upgraded bandwidth specification in addition to a paper entitlement certificate.
3
Use entitlement certificate containing instructions and certificate number needed to generate a license file for a particular 4000 X-Series
oscilloscope model number and serial number unit.
Receive the licensed file and installation instructions via email.
4
5
Copy license file (.lic extension) from email to a USB drive and follow instructions in email to install the purchased bandwidth upgrade or
measurement application on the oscilloscope.
6
For bandwidth upgrades only, attach bandwidth upgraded stick-on labels to front and rear panels of the oscilloscope. Model number and serial
number of the oscilloscope do not change.
34 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Oscilloscope Experience Redefined:
Return-to-Keysight Service Center Bandwidth Upgrades
4000 X-Series bandwidth upgrade models
Model number
Description
DSOX4B2T32U
200 MHz ~ 350 MHz , 2 ch
DSOX4B2T34U
200 MHz ~ 350 MHz , 4 ch
DSOX4B5T104U
500 MHz ~ 1 GHz , 4 ch
DSOX4B1T154U
1 GHz ~ 1.5 GHz , 4 ch
DSOX4B2T52U
200 MHz ~ 500 MHz , 2 ch
DSOX4B2T54U
200 MHz ~ 500 MHz , 4 ch
DSOX4B2T104U
200 MHz ~ 1 GHz , 4 ch
DSOX4B2T154U
200 MHz ~ 1.5 GHz , 4 ch
DSOX4B3T104U
350 MHz ~ 1 GHz , 4 ch
DSOX4B3T154U
350 MHz ~ 1.5 GHz , 4 ch
DSOX4B5T154U
500 MHz ~ 1.5 GHz , 4 ch
Type
Service center
Service center
Service center
Service center
Service center
Service center
Service center
Service center
Service center
Service center
Service center
Process description
1
Place order for a return-to-Keysight service center bandwidth upgrade product to a Keysight sales partner. Serivce Center installation,
calibration, shipment costs are in addition to bandwidth upgrade product price.
2
Keysight Business Center will contact you regarding process and timing of the service center installation. Continue to use oscilloscope until
contacted again later when parts are available at service center.
3
Ship the oscilloscope per provided instructions to service center.
4
Service center ships back upgraded oscilloscope with stick-on labels applied to front and rear panels indicating upgraded bandwidth
specification. Model number and serial number of the oscilloscope do not change.
35 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Keysight Oscilloscopes
Multiple form factors from 20 MHz to > 90 GHz | Industry-leading specs | Powerful applications
36 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
www.axiestandard.org
AdvancedTCA® Extensions for Instrumentation and Test (AXIe) is an
open standard that extends the AdvancedTCA for general purpose and
semiconductor test. The business that became Keysight was a founding
member of the AXIe consortium. ATCA®, AdvancedTCA®, and the ATCA logo
are registered US trademarks of the PCI Industrial Computer Manufacturers
Group.
www.lxistandard.org
LAN eXtensions for Instruments puts the power of Ethernet and the Web
inside your test systems. The business that became Keysight was a founding
member of the LXI consortium.
www.pxisa.org
PCI eXtensions for Instrumentation (PXI) modular instrumentation delivers a
rugged, PC-based high-performance measurement and automation system.
37 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Evolving
Our unique combination of hardware, software, support, and people can help
you reach your next breakthrough. We are unlocking the future of technology.
From Hewlett-Packard to Agilent to Keysight
myKeysight
www.keysight.com/find/mykeysight
A personalized view into the information most relevant to you.
Keysight Services
www.keysight.com/find/service
Our deep offering in design, test, and measurement services deploys an
industry-leading array of people, processes, and tools. The result? We help
you implement new technologies and engineer improved processes that
lower costs.
Three-Year Warranty
www.keysight.com/find/ThreeYearWarranty
Keysight’s committed to superior product quality and lower total cost
of ownership. Keysight is the only test and measurement company with
three-year warranty standard on all instruments, worldwide. And, we provide
a one-year warranty on many accessories, calibration devices, systems and
custom products.
Keysight Assurance Plans
www.keysight.com/find/AssurancePlans
Up to ten years of protection and no budgetary surprises to ensure your
instruments are operating to specification, so you can rely on accurate
measurements.
Keysight Channel Partners
www.keysight.com/find/channelpartners
Get the best of both worlds: Keysight’s measurement expertise and product
breadth, combined with channel partner convenience.
For more information on Keysight
Technologies’ products, applications or
services, please contact your local Keysight
office. The complete list is available at:
www.keysight.com/find/contactus
Americas
Canada
Brazil
Mexico
United States
(877) 894 4414
55 11 3351 7010
001 800 254 2440
(800) 829 4444
Asia Pacific
Australia
China
Hong Kong
India
Japan
Korea
Malaysia
Singapore
Taiwan
Other AP Countries
1 800 629 485
800 810 0189
800 938 693
1 800 11 2626
0120 (421) 345
080 769 0800
1 800 888 848
1 800 375 8100
0800 047 866
(65) 6375 8100
Europe & Middle East
Austria
Belgium
Finland
France
Germany
Ireland
Israel
Italy
Luxembourg
Netherlands
Russia
Spain
Sweden
Switzerland
United Kingdom
0800 001122
0800 58580
0800 523252
0805 980333
0800 6270999
1800 832700
1 809 343051
800 599100
+32 800 58580
0800 0233200
8800 5009286
800 000154
0200 882255
0800 805353
Opt. 1 (DE)
Opt. 2 (FR)
Opt. 3 (IT)
0800 0260637
For other unlisted countries:
www.keysight.com/find/contactus
(BP-09-23-16)
www.keysight.com/find/4000X-Series
DEKRA Certified
ISO9001 Quality Management System
www.keysight.com/go/quality
Keysight Technologies, Inc.
DEKRA Certified ISO 9001:2015
Quality Management System
This information is subject to change without notice.
© Keysight Technologies, 2013 - 2016
Published in USA, October 5, 2016
5991-1103EN
www.keysight.com
Similar pages