ETC2 DPO3FLEX Mixed signal oscilloscope Datasheet

Mixed Signal Oscilloscopes
MSO3000 Series, DPO3000 Series Datasheet
Mixed signal design and analysis (MSO series)
Automated triggering, decode, and search on parallel buses
Multichannel setup and hold triggering
MagniVu™ high-speed acquisition provides 121.2 ps fine timing
resolution on digital channels
Optional serial triggering and analysis - automated serial triggering,
decode, and search options for I2C, SPI, CAN, LIN, RS-232/422/485/
UART, and I2S/LJ/RJ/TDM
TekVPI® probe interface supports active, differential, and current
probes for automatic scaling and units
9 in. (229 mm) WVGA widescreen color display
With the MSO/DPO3000 mixed signal oscilloscope series, you can analyze
up to 20 analog and digital signals with a single instrument to quickly find
and diagnose problems in complex designs. Bandwidths up to 500 MHz
and a minimum of 5x oversampling on all channels ensure you have the
performance you need for many of today's mainstream applications. To
capture long windows of signal activity while maintaining fine timing
resolution, the MSO/DPO3000 offers a deep record length of 5 M points
standard on all channels.
Key performance specifications
500, 300, 100 MHz bandwidth models
Bandwidth is upgradable (up to 500 MHz)
2 and 4 analog channel models
2.5 GS/s sample rate on all channels
Small footprint and lightweight – Only 5.8 in. (147 mm) deep and 9 lb.
(4 kg)
Connectivity
USB 2.0 host port on both the front panel and rear panel for quick and
easy data storage, printing, and connecting a USB keyboard
USB 2.0 device port on the rear panel for easy connection to a PC or
direct printing to a PictBridge®-compatible printer
Integrated 10/100 Ethernet port for network connection and video out
port to export the oscilloscope display to a monitor or projector
Optional application support
Power analysis
HDTV and custom video analysis
5 mega-point record length on all channels
>50,000 wfm/s maximum waveform capture rate
Suite of advanced triggers
Key features
Wave Inspector® Controls provide easy navigation and automated
search of waveform data
29 automated measurements, and FFT analysis for simplified
waveform analysis
16 digital channels (MSO series)
www.tektronix.com 1
Datasheet
Feature-rich tools for debugging mixed
signal designs
With Wave Inspector® controls for rapid waveform navigation, automated
serial and parallel bus analysis, and automated power analysis – your
Tektronix oscilloscope provides the feature-rich tools you need to simplify
and speed debug of your complex design.
Comprehensive features speed every stage
of debug
These oscilloscopes offer a robust set of features to speed every stage of
debugging your design – from quickly discovering an anomaly and
capturing it, to searching your waveform record for the event and analyzing
its characteristics and your device’s behavior.
Discover
To debug a design problem, first you must know it exists. Every design
engineer spends time looking for problems in their design, a timeconsuming and frustrating task without the right debug tools.
The industry’s most complete visualization of signals provides fast insight
into the real operation of your device. A fast waveform capture rate –
greater than 50,000 waveforms per second – enables you to see glitches
and other infrequent transients within seconds, revealing the true nature of
device faults. A digital phosphor display with intensity grading shows the
history of a signal’s activity by intensifying areas of the signal that occur
more frequently, providing a visual display of just how often anomalies
occur.
Capture
Discovering a device fault is only the first step. Next, you must capture the
event of interest to identify root cause.
The MSO/DPO3000 Series provides a complete set of triggers – including
runt, logic, pulse width/glitch, setup/hold violation, serial packet, and
parallel data – to help quickly find your event. With up to a 5 Mpoint record
length, you can capture many events of interest, even thousands of serial
packets, in a single acquisition for further analysis while maintaining high
resolution to zoom in on fine signal details.
From triggering on specific packet content to automatic decode in multiple
data formats, the MSO/DPO3000 Series provides integrated support for the
industry’s broadest range of serial buses – I2C, SPI, MIL-STD-1553, CAN,
LIN, FlexRay, RS-232/422/485/UART, and I2S/LJ/RJ/TDM. The ability to
decode up to two serial and/or parallel buses simultaneously means you
gain insight into system-level problems quickly.
To further help troubleshoot system-level interactions in complex
embedded systems, the oscilloscope offers 16 digital channels in addition
to its analog channels. Since the digital channels are fully integrated into
the oscilloscope, you can trigger across all input channels, automatically
time-correlating all analog, digital, and serial signals. The MagniVu™ highspeed acquisition enables you to acquire fine signal detail around the
trigger point for precision measurements. MagniVu is essential for making
accurate timing measurements for setup and hold measurements, clock
delay, signal skew, and glitch characterization.
Capture ‒ Triggering on a specific transmit data packet going across an RS-232 bus. A
complete set of triggers, including triggers for specific serial packet content, ensures you
quickly capture your event of interest.
Discover ‒ Fast waveform capture rate - over 50,000 wfm/s - maximizes the probability
of capturing elusive glitches and other infrequent events.
2 www.tektronix.com
Search
Finding your event of interest in a long waveform record can be time
consuming without the right search tools. With today’s record lengths
pushing beyond a million data points, locating your event can mean
scrolling through thousands of screens of signal activity.
MSO3000 Series and DPO3000 Series Mixed Signal Oscilloscopes
The innovative Wave Inspector® controls give you the industry’s most
comprehensive search and waveform navigation capability. These controls
speed panning and zooming through your record. With a unique forcefeedback system, you can move from one end of your record to the other in
just seconds. User marks allow you to mark any location that you may want
to reference later for further investigation. Or, automatically search your
record for criteria you define. Wave Inspector will instantly search your
entire record, including analog, digital, and serial bus data. Along the way it
will automatically mark every occurrence of your defined event so you can
quickly move between events.
Analyze ‒ FFT analysis of a pulsed signal. A comprehensive set of integrated analysis
tools speeds verification of your design's performance.
Wave Inspector® navigation and search
With long record lengths, a single acquisition can include thousands of
screens of waveform data. Wave Inspector®, the industry’s best tool for
navigation and search, enables you to find events of interest in seconds.
Search ‒ I2C decode showing results from a Wave Inspector search for Address value
50. Wave Inspector controls provide unprecedented efficiency in viewing and navigating
waveform data.
Analyze
Verifying that your prototype’s performance matches simulations and meets
the project’s design goals requires analyzing its behavior. Tasks can range
from simple checks of rise times and pulse widths to sophisticated power
loss analysis and investigation of noise sources.
The oscilloscope offers a comprehensive set of integrated analysis tools
including waveform- and screen-based cursors, automated measurements,
advanced waveform math including arbitrary equation editing, FFT
analysis, and trend plots for visually determining how a measurement is
changing over time. Specialized application support for serial bus analysis,
power supply design, and video design and development is also available.
For extended analysis, National Instrument’s LabVIEW SignalExpress®
Tektronix Edition provides over 200 built-in functions including time and
frequency domain analysis, limit testing, data logging, and customizable
reports.
Wave Inspector controls provide unprecedented efficiency in viewing, navigating, and
analyzing waveform data. Zip through your long record by turning the outer pan control
(1). Get details from the beginning to end in seconds. See something of interest and want
to see more details? Just turn the inner zoom control (2).
Zoom and pan
A dedicated, two-tier front-panel control provides intuitive control of both
zooming and panning. The inner control adjusts the zoom factor (or zoom
scale); turning it clockwise activates zoom and goes to progressively higher
zoom factors, while turning it counterclockwise results in lower zoom
factors and eventually turning zoom off. No longer do you need to navigate
through multiple menus to adjust your zoom view. The outer control pans
the zoom box across the waveform to quickly get to the portion of waveform
you are interested in. The outer control also utilizes force-feedback to
determine how fast to pan on the waveform. The farther you turn the outer
control, the faster the zoom box moves. Pan direction is changed by simply
turning the control the other way.
www.tektronix.com 3
Datasheet
Play/Pause
A dedicated Play/Pause front-panel button scrolls the waveform across the
display automatically while you look for anomalies or an event of interest.
Playback speed and direction are controlled using the intuitive pan control.
Once again, turning the control further makes the waveform scroll faster
and changing direction is as simple as turning the control the other way.
User marks
Press the Set Mark front-panel button to place one or more marks on the
waveform. Navigating between marks is as simple as pressing the
Previous (←) and Next (→) buttons on the front panel.
Search marks
The Search button allows you to automatically search through your long
acquisition looking for user-defined events. All occurrences of the event are
highlighted with search marks and are easily navigated to, using the frontpanel Previous (←) and Next (→) buttons. Search types include edge,
pulse width/glitch, timeout, runt, logic, setup and hold, rise/fall time, parallel
bus, and I2C, SPI, CAN, LIN, FlexRay, RS-232/422/485/UART, MILSTD-1553, and I2S/LJ/RJ/TDM packet content.
Search step 2: Wave Inspector automatically searches through the record and marks
each event with a hollow white triangle. You can then use the Previous and Next buttons
to jump from one event to the next.
Digital phosphor technology
Digital phosphor technology provides you with fast insight into the real
operation of your device. Its fast waveform capture rate – greater than
50,000 wfm/s – gives you a high probability of quickly seeing the infrequent
problems common in digital systems: runt pulses, glitches, timing issues,
and more.
Waveforms are superimposed with one another and waveform points that
occur more frequently are intensified. This quickly highlights the events that
over time occur more often or, in the case of infrequent anomalies, occur
less often.
You can choose infinite persistence or variable persistence, determining
how long the previous waveform acquisitions stay on-screen. This allows
you to determine how often an anomaly is occurring.
Search step 1: You define what you would like to find.
Digital phosphor technology enables greater than 50,000 wfms/s waveform capture rate
and real-time intensity grading.
4 www.tektronix.com
MSO3000 Series and DPO3000 Series Mixed Signal Oscilloscopes
Mixed signal design and analysis (MSO
series)
The MSO models provide 16 digital channels which are tightly integrated
into the oscilloscope's user interface. This simplifies operation and makes it
possible to solve mixed-signal issues easily.
White edges indicate additional information is available by zooming in. As shown here,
zooming in on the white edge reveals a hidden glitch.
You can group digital waveforms and enter waveform labels by using a
USB keyboard. By simply placing digital waveforms next to each other, they
form a group.
The MSO Series provides 16 integrated digital channels enabling you to view and
analyze time-correlated analog and digital signals.
Color-coded digital waveform display
This oscilloscope has redefined the way you view digital waveforms. One
common problem shared by both logic analyzers and mixed-signal
oscilloscopes is determining if data is a one or a zero when zoomed in far
enough that the digital trace stays flat all the way across the display. Colorcoded digital traces display ones in green and zeros in blue.
The multiple transition detection hardware shows you a white edge on the
display when the system detects multiple transitions. White edges indicate
that more information is available by zooming in or acquiring at faster
sampling rates. In most cases zooming in will reveal the pulse that was not
viewable with the previous settings. If the white edge is still present after
zooming in as far as possible, this indicates that increasing the sample rate
on the next acquisition will reveal higher frequency information than the
previous settings could acquire.
With color-coded digital waveform display, groups are created by simply placing digital
channels together on the screen, allowing digital channels to be moved as a group. You
can set threshold values for each channel, enabling support for up to 16 different logic
families.
Once a group is formed, you can position all the channels contained in that
group collectively. This greatly reduces the normal setup time associated
with positioning channels individually.
www.tektronix.com 5
Datasheet
MagniVu™ high-speed acquisition
Serial triggering and analysis (optional)
The main digital acquisition mode on the MSO3000 Series will capture up
to 5 Mpoints at 500 MS/s (2 ns resolution). In addition to the main record,
the MSO3000 provides an ultra high-resolution record called MagniVu
which acquires 10,000 points at up to 8.25 GS/s (121.2 ps resolution). Both
main and MagniVu waveforms are acquired on every trigger and can be
switched between in the display at any time, running or stopped. MagniVu
provides significantly finer timing resolution than comparable MSOs on the
market, instilling confidence when making critical timing measurements on
digital waveforms.
On a serial bus, a single signal often includes address, control, data, and
clock information. This can make isolating events of interest difficult.
Automatic trigger, decode, and search on bus events and conditions gives
you a robust set of tools for debugging serial buses.
Triggering on a specific data packet going across an I2C bus. The yellow waveform is
clock and the blue waveform is the data. A bus waveform provides decoded packet
content including Start, Address, Read/Write, Data, and Stop.
Serial triggering
The MagniVu high-resolution record provides 121.2 ps timing resolution, enabling you to
take critical timing measurements on your digital waveforms.
P6316 MSO probe
This unique probe design offers two eight-channel pods, simplifying the
process of connecting to the device-under-test. When connecting to square
pins, the P6316 can connect directly to 8×2 square pin headers spaced on
tenth-inch centers. When more attachment flexibility is required, you can
use the included flying lead sets and grabbers to clip onto surface mount
devices or test points. The P6316 offers outstanding electrical
characteristics applying only 8 pF of capacitive loading with 101 kΩ input
impedance.
Trigger on packet content such as start of packet, specific addresses,
specific data content, unique identifiers, etc. on popular serial interfaces
such as I2C, SPI, MIL-STD-1553, CAN, LIN, FlexRay, RS-232/422/485/
UART, and I2S/LJ/RJ/TDM.
Bus display
Provides a higher-level, combined view of the individual signals (clock,
data, chip enable, etc.) that make up your bus, making it easy to identify
where packets begin and end and identifying sub-packet components such
as address, data, identifier, CRC, etc.
Bus decoding
Tired of having to visually inspect the waveform to count clocks, determine
if each bit is a 1 or a 0, combine bits into bytes, and determine the hex
value? Let the oscilloscope do it for you! Once you’ve set up a bus, the
MSO/DPO3000 Series will decode each packet on the bus, and display the
value in hex, binary, decimal (LIN, FlexRay, and MIL-STD-1553 only),
signed decimal (I2S/LJ/RJ/TDM only), or ASCII (MIL-STD-1553 and
RS-232/422/485/UART only) in the bus waveform.
The P6316 MSO probe offers two eight-channel pods to simplify connecting to your
device.
6 www.tektronix.com
MSO3000 Series and DPO3000 Series Mixed Signal Oscilloscopes
Capabilities of the serial triggering application modules
Technology
Trigger
Bus decode
Event table
Search
Order product
I2C
■
■
■
■
DPO3EMBD
SPI
■
■
■
■
DPO3EMBD
Computer
RS232/422/485, UART ■
■
■
■
DPO3COMP
Automotive
CAN
■
■
■
■
DPO3AUTO
LIN
■
■
■
■
DPO3AUTO
FlexRay
■
■
■
■
DPO3FLEX
■
■
■
■
DPO3AERO
I2S
■
■
■
■
DPO3AUDIO
LJ, RJ
■
■
■
■
DPO3AUDIO
TDM
■
■
■
■
DPO3AUDIO
Embedded
Military and aerospace MIL-STD-1553
Audio
Event table
Power analysis (optional)
In addition to seeing decoded packet data on the bus waveform itself, you
can view all captured packets in a tabular view much like you would see in
a software listing. Packets are time stamped and listed consecutively with
columns for each component (Address, Data, etc.). You can save the event
table data in .csv format.
Ever increasing consumer demand for longer battery-life devices and for
green solutions that consume less power require power-supply designers to
characterize and minimize switching losses to improve efficiency. In
addition, the supply’s power levels, output purity, and harmonic feedback
into the power line must be characterized to comply with national and
regional power quality standards. Historically, making these and many
other power measurements on an oscilloscope has been a long, manual,
and tedious process. The optional power analysis tools greatly simplify
these tasks, enabling quick and accurate analysis of power quality,
switching loss, harmonics, safe operating area (SOA), modulation, ripple,
and slew rate (di/dt, dv/dt). Completely integrated into the oscilloscope, the
power analysis tools provide automated, repeatable power measurements
with a touch of a button; no external PC or complex software setup is
required.
Event table showing decoded identifier, DLC, DATA, and CRC for every CAN packet in a
long acquisition.
Search (serial triggering)
Serial triggering is very useful for isolating the event of interest, but once
you’ve captured it and need to analyze the surrounding data, what do you
do? In the past, users had to manually scroll through the waveform
counting and converting bits and looking for what caused the event. You
can have the oscilloscope automatically search through the acquired data
for user-defined criteria including serial packet content. Each occurrence is
highlighted by a search mark. Rapid navigation between marks is as simple
as pressing the Previous (←) and Next (→) buttons on the front panel.
Safe operating area measurement. Automated power measurements enable quick and
accurate analysis of common power parameters.
www.tektronix.com 7
Datasheet
Video design and development
Large high-resolution display
Many video engineers have remained loyal to analog oscilloscopes,
believing the intensity gradations on an analog display are the only way to
see certain video waveform details. The fast waveform capture rate,
coupled with its intensity-graded view of the signal, provides the same
information-rich display as an analog oscilloscope, but with much more
detail and all the benefits of digital scopes.
The MSO/DPO3000 Series features a 9 inch (229 mm) wide-screen, highresolution (800 × 480 WVGA) display for seeing intricate signal details.
Standard features such as IRE and mV graticules, holdoff by fields, video
polarity, and an Autoset smart enough to detect video signals, make these
the easiest to use oscilloscopes on the market for video applications. And
with high bandwidth, four analog inputs, and a built-in 75 Ω input
termination, the oscilloscope provides ample performance for analog and
digital video use.
The video functionality is further extended with the optional video
application module, which provides the industry's most complete suite of
HDTV and custom (nonstandard) video triggers.
Dedicated front-panel controls
Per-channel vertical controls provide simple and intuitive operation. No
longer do you need to share one set of vertical controls across all four
channels.
Connectivity
A USB host port on the front panel enables easy transfer of screen shots,
instrument settings, and waveform data to a USB mass storage device. The
rear panel contains a second USB host port and a USB device port for
controlling the oscilloscope remotely from a PC or for connecting a USB
keyboard. The USB device port can also be used to print directly to a
PictBridge®-compatible printer. An integrated 10/100 Ethernet port enables
easy connection to networks and a Video Out port allows the oscilloscope
display to be exported to an external monitor or projector.
Compact form factor
With the compact, portable form factor, you can easily move the
oscilloscope between labs. And with a depth of just 5.8 inches (147 mm), it
saves you valuable space on your test bench.
Viewing an NTSC video signal. Notice the intensity-graded view provided by the DPO's
ability to represent time, amplitude, and distribution over time.
Designed to make your work easier
The MSO/DPO3000 Series compact form factor frees up valuable space on your bench
or desktop.
The MSO/DPO3000 is designed to make your work easier. The large, high-resolution
display shows intricate signal details. Dedicated front-panel controls simplify operation. A
USB host port on the front panel allows you to easily transfer screen shots, instrument
settings, and waveform data to a USB mass storage device.
8 www.tektronix.com
MSO3000 Series and DPO3000 Series Mixed Signal Oscilloscopes
TekVPI® probe interface
The TekVPI probe interface sets the standard for ease of use in probing. In
addition to the secure, reliable connection that the interface provides,
TekVPI probes feature status indicators and controls, as well as a probe
menu button right on the comp box itself. This button brings up a probe
menu on the oscilloscope display with all relevant settings and controls for
the probe. The TekVPI interface enables direct attachment of current
probes without requiring a separate power supply. TekVPI probes can be
controlled remotely through USB, GPIB, or LAN, enabling more versatile
solutions in ATE environments.
NI LabVIEW SignalExpress™ Tektronix Edition is a fully interactive measurement and
analysis software package developed jointly with National Instruments and optimized for
the MSO/DPO Series.
TekVPI probe interface simplifies connecting your probes to the oscilloscope.
Extended analysis
Exporting data and measurements is as simple as connecting a USB cable
from the oscilloscope to your PC. Key software applications – NI LabVIEW
SignalExpress™ Tektronix Edition LE, OpenChoice® Desktop, and
Microsoft Excel and Word toolbars – are included standard with each
oscilloscope to enable fast and easy direct communication with your
Windows PC.
NI LabVIEW SignalExpress Tektronix Edition LE enables you to instantly
acquire, generate, analyze, compare, import, and save measurement data
and signals using an intuitive drag-and-drop user interface that does not
require any programming. The optional Professional Version offers over
200 built-in functions that provide additional signal processing, advanced
analysis, sweeping, limit testing and user-defined step capabilities.
OpenChoice® Desktop software enables seamless connection between the oscilloscope
and your PC.
For simple tasks, the included OpenChoice Desktop enables fast and easy
communication between the oscilloscope and your PC through USB or LAN
for transferring settings, waveforms, and screen images.
www.tektronix.com 9
Datasheet
Specifications
All specifications apply to all models unless noted otherwise.
Model overview
MSO3012 and
DPO3012
MSO3014 and
DPO3014
MSO3032 and
DPO3032
MSO3034 and
DPO3034
DPO3052
MSO3054 and
DPO3054
Analog channels
2
4
2
4
2
4
Analog bandwidth (-3dB)
100 MHz
100 MHz
300 MHz
300 MHz
500 MHz
500 MHz
Calculated rise time at 5 mv/div,
typical
3.5 ns
3.5 ns
1.17 ns
1.17 ns
700 ps
700 ps
Sample rate (analog channels)
2.5 GS/s
2.5 GS/s
2.5 GS/s
2.5 GS/s
2.5 GS/s
2.5 GS/s
Record length
5 M points
5 M points
5 M points
5 M points
5 M points
5 M points
Digital channels
MSO models add 16 digital channels to the corresponding DPO model
Vertical system analog channels
Hardware bandwidth limits
≥300 MHz models
20 MHz and 150 MHz
100 MHz models
20 MHz
Input coupling
AC, DC, GND
Input impedance
1 MΩ ±1%, 75 Ω ± 1%, 50 Ω ±1%
Input sensitivity range
1 MΩ
1 mV/div to 10 V/div
75 Ω, 50 Ω
1 mV/div to 1 V/div
Vertical resolution
8 bits (11 bits with Hi Res)
Maximum input voltage
1 MΩ
300 VRMS with peaks ≤ ±450 V
75 Ω, 50 Ω
5 VRMS with peaks ≤ ±20 V
DC gain accuracy
±1.5% for 5 mV/div and above
±2.0% for 2 mV/div
±2.5% for 1 mV/div
Channel-to-channel isolation
Offset range
10 www.tektronix.com
Any two channels at equal vertical scale ≥100:1 at ≤100 MHz and ≥30:1 at >100 MHz up to the rated bandwidth
Volts/div setting
Offset range
1 MΩ input
50 Ω and 75 Ω input
1 mV/div to 99.5 mV/div
±1 V
±1 V
100 mV/div to 995 mV/div
±10 V
±5 V
1 V/div to 10 V/div
±100 V
±5 V
MSO3000 Series and DPO3000 Series Mixed Signal Oscilloscopes
Vertical system digital channels
Thresholds
Threshold per set of 8 channels
Threshold selections
TTL, CMOS, ECL, PECL, User-defined
User-defined threshold range
–15 V to +25 V
Maximum input voltage
–20 V to +30 V
Threshold accuracy
±[100 mV + 3% of threshold setting]
Maximum input dynamic range
50 Vp-p (threshold setting dependent)
Minimum voltage swing
500 mVp-p
Input resistance
101 kΩ
Probe loading
8 pF
Vertical resolution
1 bit
Horizontal system analog channels
Maximum duration of time
captured at highest sample rate
(all channels)
2 ms
Seconds/division range
1 ns/div to 1000 s/div
Time-base delay time range
-10 divisions to 5000 s
Channel-to-channel deskew range
±100 ns
Time base accuracy
±10 ppm over any ≥1 ms interval
Horizontal system digital channels
Seconds/division range
1 ns/div to 1000 s/div
Maximum record length (main, all
channels)
5 M points
Maximum sample rate (MagniVu,
all channels)
8.25 GS/s (121.2 ps resolution)
Maximum record length (MagniVu,
all channels)
10 k points centered on the trigger
Minimum detectible pulse width
2.0 ns
Channel-to-channel skew
500 ps, typical, digital channel to digital channel only This is the propagation path skew and ignores skew contributions due to
bandpass distortion, threshold inaccuracies (see Threshold accuracy), and sample binning (see Digital channel timing resolution)
www.tektronix.com 11
Datasheet
Trigger system
Trigger modes
Auto, Normal, and Single
Trigger coupling
DC, AC, HF reject (attenuates >50 kHz), LF reject (attenuates <50 kHz), noise reject (reduces sensitivity)
Trigger holdoff range
20 ns to 8 s
Trigger sensitivity, typical
Edge type, DC coupled
Trigger Source
Sensitivity
Any input channel
0.50 div from DC to 50 MHz, increasing to 1 div at oscilloscope
bandwidth
Aux input (external trigger)
200 mV from DC to 50 MHz, increasing to 500 mV at 250 MHz
Line
Fixed
Trigger level ranges
Any input channel
±8 divisions from center of screen, ±8 divisions from 0 V when vertical LF reject trigger coupling is selected
Aux Input (external trigger)
±8 V
Line
The line trigger level is fixed at about 50% of the line voltage.
Trigger types
Edge
Positive or negative slope on any channel or front-panel auxiliary input. Coupling includes DC, AC, HF reject, LF reject, and noise
reject.
Sequence (B-trigger)
Trigger Delay by Time: 8 ns to 8 s. Or Trigger Delay by Events: 1 to 9,999,999 events.
Pulse Width
Trigger on width of positive or negative pulses that are >, <, =, ≠, or inside/outside a specified period of time.
Timeout
Trigger on an event which remains high, low, or either, for a specified time period (4 ns to 8 s).
Runt
Trigger on a pulse that crosses one threshold but fails to cross a second threshold before crossing the first again.
Logic
Trigger when any logical pattern of channels goes false or stays true for specified period of time. Any input can be used as a clock
to look for the pattern on a clock edge. Pattern (AND, OR, NAND, NOR) specified for all input channels defined as High, Low, or
Don’t Care.
Setup and Hold
Trigger on violations of both setup time and hold time between clock and data present on any of the analog and digital input
channels.
Rise/Fall Time
Trigger on pulse edge rates that are faster or slower than specified. Slope may be positive, negative, or either.
Video
Trigger on all lines, odd, even, or all fields on NTSC, PAL, and SECAM video signals.
Extended Video (optional)
Trigger on 480p/60, 576p/50, 720p/30, 720p/50, 720p/60, 875i/60, 1080i/50, 1080i/60, 1080p/24, 1080p/24sF, 1080p/25, 1080p/
30, 1080p/50, 1080p/60, and custom bi-level and tri-level sync video standards.
I2C (optional)
Trigger on Start, Repeated Start, Stop, Missing ACK, Address (7 or 10 bit), Data, or Address and Data on I2C buses up to 10 Mb/s.
SPI (optional)
Trigger on SS, MOSI, MISO, or MOSI and MISO on SPI buses up to 10.0 Mb/s.
RS-232/422/485/UART
(optional)
Trigger on Tx Start Bit, Rx Start Bit, Tx End of Packet, Rx End of Packet, Tx Data, Rx Data, Tx Parity Error, and Rx Parity Error up
to 10 Mb/s.
CAN (optional)
Trigger on Start of Frame, Frame Type (data, remote, error, overload), Identifier (standard or extended), Data, Identifier and Data,
End of Frame, Missing ACK, or Bit Stuffing Error on CAN signals up to 1 Mb/s. Data can be further specified to trigger on ≤, <, =,
>, ≥, or ≠ a specific data value. User-adjustable sample point is set to 50% by default.
LIN (optional)
Trigger on Sync, Identifier, Data, Identifier and Data, Wakeup Frame, Sleep Frame, Errors such as Sync, Parity, or Checksum
Errors up to 100 kb/s (by LIN definition, 20 kb/s).
FlexRay (optional)
Trigger on Start of Frame, Type of Frame (Normal, Payload, Null, Sync, Startup), Identifier, Cycle Count, Complete Header Field,
Data, Identifier and Data, End of Frame or Errors such as Header CRC, Trailer CRC, Null Frame, Sync Frame, or Startup Frame
Errors up to 100 Mb/s.
12 www.tektronix.com
MSO3000 Series and DPO3000 Series Mixed Signal Oscilloscopes
Trigger system
MIL-STD-1553 (optional)
Trigger on Sync, Word Type 1 (Command, Status, Data), Command Word (set RT Address, T/R, Sub-address/Mode, Data Word
Count/Mode Code, and Parity individually), Status Word (set RT Address, Message Error, Instrumentation, Service Request Bit,
Broadcast Command Received, Busy, Subsystem Flag, Dynamic Bus Control Acceptance (DBCA), Terminal Flag, and Parity
individually), Data Word (user-specified 16-bit data value), Error (Sync, Parity, Manchester, Non-contiguous data), Idle Time
(minimum time selectable from 2 µs to 100 µs; maximum time selectable from 2 µs to 100 µs; trigger on < minimum, > maximum,
inside range, outside range). RT Address can be further specified to trigger on =, ≠, <, >, ≤, ≥ a particular value, or inside or
outside of a range.
I2S/LJ/RJ/TDM (optional)
Trigger on Word Select, Frame Sync, or Data. Data can be further specified to trigger on ≤, <, =, >, ≥, ≠ a specific data value, or
inside or outside of a range. Maximum data rate for I2S/LJ/RJ is 12.5 Mb/s. Maximum data rate for TDM is 25 Mb/s.
Parallel (available on MSO
models only)
Trigger on a parallel bus data value. Parallel bus can be from 1 to 16 bits (from the digital channels) plus 2 or 4 bits (from the
analog channels) in size. Binary and Hex radices are supported.
Acquisition system
Acquisition Modes
Sample
Acquire sampled values.
Peak Detect
Captures glitches as narrow as 2 ns at all sweep speeds.
Averaging
From 2 to 512 waveforms included in average.
Envelope
Min-max envelope reflecting Peak Detect data over multiple acquisitions.
Hi Res
Real-time boxcar averaging reduces random noise and increases vertical resolution.
Roll
Scrolls waveforms right to left across the screen at sweep speeds slower than or equal to 40 ms/div.
Waveform measurements
Cursors
Waveform and Screen.
Automatic measurements
29, of which up to four can be displayed on-screen at any one time. Measurements include: Period, Frequency, Delay, Rise Time,
Fall Time, Positive Duty Cycle, Negative Duty Cycle, Positive Pulse Width, Negative Pulse Width, Burst Width, Phase, Positive
Overshoot, Negative Overshoot, Peak to Peak, Amplitude, High, Low, Max, Min, Mean, Cycle Mean, RMS, Cycle RMS, Positive
Pulse Count, Negative Pulse Count, Rising Edge Count, Falling Edge Count, Area and Cycle Area.
Measurement statistics
Mean, Min, Max, Standard Deviation.
Reference levels
User-definable reference levels for automatic measurements can be specified in either percent or units.
Gating
Isolate the specific occurrence within an acquisition to take measurements on, using either the screen, or waveform cursors.
Power measurements (optional)
Power Quality Measurements
VRMS, VCrest Factor, Frequency, IRMS, ICrest Factor, True Power, Apparent Power, Reactive Power, Power Factor, Phase Angle.
Switching loss measurements
1
Power loss
Ton, Toff, Conduction, Total.
Energy loss
Ton, Toff, Conduction, Total.
Harmonics
THD-F, THD-R, RMS measurements. Graphical and table displays of harmonics. Test to IEC61000-3-2 Class A and MILSTD-1399, Section 300A.
Ripple measurements
VRipple and IRipple.
Modulation Analysis
Graphical display of +Pulse Width, –Pulse Width, Period, Frequency, +Duty Cycle, and –Duty Cycle modulation types.
Trigger selection of Command Word will trigger on Command and ambiguous Command/Status words. Trigger selection of Status Word will trigger on Status and ambiguous Command/Status words.
www.tektronix.com 13
Datasheet
Power measurements (optional)
Safe operating area
Graphical display and mask testing of switching device safe operating area measurements.
dV/dt and dI/dt measurements
Cursor measurements of slew rate.
Waveform math
Arithmetic
Add, subtract, multiply, and divide waveforms.
Math functions
Integrate, Differentiate, FFT.
FFT
Spectral magnitude. Set FFT Vertical Scale to Linear RMS or dBV RMS, and FFT Window to Rectangular, Hamming, Hanning, or
Blackman-Harris.
Advanced math
Define extensive algebraic expressions including waveforms, reference waveforms, math functions (FFT, Intg, Diff, Log, Exp, Sqrt,
Sine, Cosine, Tangent), scalars, up to two user-adjustable variables and results of parametric measurements (Period, Freq, Delay,
Rise, Fall, PosWidth, NegWidth, BurstWidth, Phase, PosDutyCycle, NegDutyCycle, PosOverShoot, NegOverShoot, PeakPeak,
Amplitude, RMS, CycleRMS, High, Low, Max, Min, Mean, CycleMean, Area, CycleArea, and trend plots), e.g.,(Intg(Ch1 Mean(Ch1)) × 1.414 × VAR1).
Software
NI LabVIEW SignalExpress™
Tektronix Edition
A fully interactive measurement software environment optimized for your Tektronix oscilloscope, enables you to instantly acquire,
generate, analyze, compare, import, and save measurement data and signals using an intuitive drag-and-drop user interface that
does not require any programming.
Standard support for acquiring, controlling, viewing, and exporting your live analog-channel signal data is permanently available
through the software. The full version (SIGEXPTE) adds additional signal processing, advanced analysis, mixed signal, sweeping,
limit testing, and user-defined step capabilities and is available for a 30-day trial period standard with each instrument.
OpenChoice® Desktop
Enables fast and easy communication between a Windows PC and your oscilloscope using USB or LAN. Transfer and save
settings, waveforms, measurements, and screen images. Included Word and Excel toolbars automate the transfer of acquisition
data and screen images from the oscilloscope into Word and Excel for quick reporting or further analysis.
IVI driver
Provides a standard instrument programming interface for common applications such as LabVIEW, LabWindows/CVI,
Microsoft .NET, and MATLAB.
e*Scope® Web-based remote
control
Enables control of the oscilloscope over a network connection through a standard web browser. Simply enter the IP address or
network name of the oscilloscope and a web page will be served to the browser.
Display system
Display type
9 inch (228.6 mm) wide format liquid crystal TFT color display.
Display resolution
800 horizontal × 480 vertical pixels (WVGA).
Waveform styles
Vectors, Dots, Variable Persistence, Infinite Persistence.
Graticules
Full, Grid, Cross Hair, Frame, IRE and mV.
Format
YT and XY.
Maximum waveform capture rate
>50,000 wfm/s.
14 www.tektronix.com
MSO3000 Series and DPO3000 Series Mixed Signal Oscilloscopes
Input/output ports
USB 2.0 high-speed host port
Supports USB mass storage devices, printers, and keyboards. One port available on rear panel and one on front panel.
LAN port (Ethernet)
RJ-45 connector, supports 10/100BASE-T.
GPIB interface
Available as an optional accessory that connects to the USB Device and USB Host port with the TEK-USB-488 GPIB to USB
Adapter. The control interface is incorporated in the instrument user interface.
Video out port
DB-15 female connector, connect to show the oscilloscope display on an external monitor or projector. SVGA resolution.
Auxiliary input
Front-panel BNC connector. Input Impedance 1 MΩ. Max input 300 VRMS CAT II with peaks ≤ ±425 V.
Probe compensator output voltage Front-panel pins
and frequency
Amplitude
0 to 2.5 V
Frequency
1 kHz
Trigger out
Rear-panel BNC connector, provides a negative-polarity pulse when the oscilloscope triggers.
Kensington-style lock
Rear-panel security slot connects to standard Kensington-style lock.
Power source
Power source voltage
85 to 265 V ±10%
Power source frequency
45 to 440 Hz (85 to 265 V)
Power consumption
120 W maximum
Optional TekVPI® external power
supply 2
Output voltage
12 V
Output current
5A
Power consumption
50 W
Physical characteristics
Dimensions
mm
in.
Height
203.2
8
Width
416.6
16.4
Depth
147.3
5.8
kg
lb.
Net
4.17
9.2
Shipping
8.62
19
Weight
2
Rackmount configuration
5U
Cooling clearance
2 in. (51 mm) required on left side and rear of instrument
Required when total oscilloscope probe power usage exceeds 20 W.
www.tektronix.com 15
Datasheet
EMC, environment, and safety
Temperature
Operating
0 ºC to +50 ºC (+32 ºF to 122 ºF)
Nonoperating
-40 ºC to +71 ºC (-40 ºF to 160 ºF)
Humidity
Operating
High: 30 ºC to 50 ºC, 5% to 45% relative humidity Low: 0 ºC to 30 ºC, 5% to 95% relative humidity
Nonoperating
High: 30 ºC to 50 ºC, 5%to 45% relative humidity Low: 0 ºC to 30 ºC, 5% to 95% relative humidity
Altitude
Operating
3,000 meters (9,843 feet)
Nonoperating
12,000 meters (39,370 feet)
Random vibration
Operating
0.31 GRMS from 5 to 500 Hz, 10 minutes each axis, 3 axes, 30 minutes total
Nonoperating
2.46 GRMS from 5 to 500 Hz, 10 minutes each axis, 3 axes, 30 minutes total
Regulatory
Electromagnetic compatibility
EC Council Directive 2004/108/EC
Safety
UL61010-1:2004, CAN/CSA-C22.2 No. 61010.1: 2004, Low Voltage Directive 2006/95/EC and EN61010-1:2001,
IEC 61010-1:2001, ANSI 61010-1-2004, ISA 82.02.01
Ordering information
DPO3000 and MSO3000 models
DPO3012
100 MHz, 2.5 GS/s, 5 M record length, 2-channel digital phosphor oscilloscope
DPO3014
100 MHz, 2.5 GS/s, 5 M record length, 4-channel digital phosphor oscilloscope
DPO3032
300 MHz, 2.5 GS/s, 5 M record length, 2-channel digital phosphor oscilloscope
DPO3034
300 MHz, 2.5 GS/s, 5 M record length, 4-channel digital phosphor oscilloscope
DPO3052
500 MHz, 2.5 GS/s, 5 M record length, 2-channel digital phosphor oscilloscope
DPO3054
500 MHz, 2.5 GS/s, 5 M record length, 4-channel digital phosphor oscilloscope
MSO3012
100 MHz, 2.5 GS/s, 5 M record length, 2+16 channel mixed-signal oscilloscope
MSO3014
100 MHz, 2.5 GS/s, 5 M record length, 4+16 channel mixed-signal oscilloscope
MSO3032
300 MHz, 2.5 GS/s, 5 M record length, 2+16 channel mixed-signal oscilloscope
MSO3034
300 MHz, 2.5 GS/s, 5 M record length, 4+16 channel mixed-signal oscilloscope
MSO3054
500 MHz, 2.5 GS/s, 5 M record length, 4+16 channel mixed-signal oscilloscope
16 www.tektronix.com
MSO3000 Series and DPO3000 Series Mixed Signal Oscilloscopes
Standard accessories
Probes
P6139A
500 MHz, 10x passive probe (one per analog channel)
P6316 (MSO models only)
16-channel logic probe and accessory kit (one per instrument)
Accessories
Please specify power plug and manual language version when ordering.
200-5052-xx
Front cover
063-4104-xx
Documentation CD
016-2008-xx
Accessory pouch
—
User manual
—
Power cord
—
OpenChoice® Desktop software
—
NI LabVIEW SignalExpress® Tektronix Edition LE software
—
Calibration certificate documenting traceability to National Metrology Institute(s) and ISO9001 quality system registration
Warranty
Three-year warranty covering all parts and labor, excluding probes.
Application modules
Application modules have licenses which can be transferred between an application module and an oscilloscope. The license may be contained in the module; allowing the
module to be moved from one instrument to another. Or, the license can be contained in the oscilloscope; allowing the module to be removed and stored for safekeeping.
Transferring the license to an oscilloscope and removing the module permits the use of more than 4 applications simultaneously.
DPO3AERO
Aerospace Serial Triggering and Analysis Module. Enables triggering on packet-level information on MIL-STD-1553 buses as well
as analytical tools such as digital views of the signal, bus views, packet decoding, search tools, and packet decode tables with
time-stamp information.
Signal Inputs - Any Ch1 - Ch4, Math, Ref1 - Ref4
Recommended Probing - Differential or single ended (only one single-ended signal required)
DPO3AUDIO
Audio Serial Triggering and Analysis Module. Enables triggering on packet-level information on I2S, Left Justified, Right Justified,
TDM, and custom audio buses, as well as analytical tools such as digital views of the signal, bus views, packet decoding, search
tools, and packet decode tables with time stamp information.
Signal Inputs - Any Ch1 - Ch4 (and any D0 - D15 on MSO models)
Recommended Probing - Single ended
DPO3AUTO
Automotive Serial Triggering and Analysis Module. Enables triggering on packet-level information on CAN bus and LIN bus as well
as analytical tools such as digital views of the signal, bus views, packet decoding, search tools, and packet decode tables with
time stamp information.
Signal Inputs - LIN: Any Ch1 - Ch4 (and any D0 - D15 on MSO models); CAN: Any Ch1 - Ch4 (and any D0 - D15 on MSO models;
single-ended probing only)
Recommended Probing - LIN: Single ended; CAN: Single ended or differential
www.tektronix.com 17
Datasheet
DPO3COMP
Computer Serial Triggering and Analysis Module. Enables triggering on packet-level information on RS-232/422/485/UART buses
as well as analytical tools such as digital views of the signal, bus views, packet decoding, search tools, and packet decode tables
with time stamp information.
Signal Inputs - Any Ch1 - Ch4 (and any D0 - D15 on MSO models; single-ended probing only)
Recommended Probing - RS-232/UART: Single ended; RS-422/485: Differential
DPO3EMBD
Embedded Serial Triggering and Analysis Module. Enables triggering on packet-level information on I2C and SPI buses as well as
analytical tools such as digital views of the signal, bus views, packet decoding, search tools, and packet decode tables with time
stamp information. Only two-wire SPI support available on DPO3012, DPO3032, and DPO3052 models.
Signal Inputs - I2C: Any Ch1 - Ch4 (and any D0 - D15 on MSO models); SPI: Any Ch1 - Ch4 (and any D0 - D15 on MSO models)
Recommended Probing - I2C, SPI: Single ended
DPO3FLEX
FlexRay Serial Triggering and Analysis Module. Enables triggering on packet-level information on FlexRay buses as well as
analytical tools such as digital views of the signal, bus views, packet decoding, search tools, packet decode tables with time-stamp
information.
Signal Inputs - Any Ch1 - Ch4 (and any D0 - D15 on MSO models; single-ended probing only)
Recommended Probing - Single ended or differential
DPO3PWR
Power Analysis Module. Enables quick and accurate analysis of power quality, switching loss, harmonics, safe operating area
(SOA), modulation, ripple, and slew rate (dI/dt, dV/dt).
DPO3VID
HDTV and Custom (nonstandard) Video Triggering Module.
Instrument options
Bandwidth upgrades
Bandwidth is fully upgradable (up to 500 MHz) on any MSO/DPO3000 Series oscilloscope. Instruments with serial numbers starting with C02 or B02 are upgraded with a
software option key. Instruments with serial numbers starting with C01 or B01 must be upgraded at an authorized Tektronix service center.
DPO3BW1T32
Bandwidth upgrade from 100 MHz to 300 MHz for an MSO/DPO3012
DPO3BW1T34
Bandwidth upgrade from 100 MHz to 300 MHz for an MSO/DPO3014
DPO3BW1T52
Bandwidth upgrade from 100 MHz to 500 MHz for an MSO/DPO3012
DPO3BW1T54
Bandwidth upgrade from 100 MHz to 500 MHz for an MSO/DPO3014
DPO3BW3T52
Bandwidth upgrade from 300 MHz to 500 MHz for an MSO/DPO3012 or an MSO/DPO3032
DPO3BW3T54
Bandwidth upgrade from 300 MHz to 500 MHz for an MSO/DPO3014 or an MSO/DPO3034
Power cord and plug options
Opt. A0
North America power plug (115 V, 60 Hz)
Opt. A1
Universal Euro power plug (220 V, 50 Hz)
Opt. A2
United Kingdom power plug (240 V, 50 Hz)
Opt. A3
Australia power plug (240 V, 50 Hz)
Opt. A5
Switzerland power plug (220 V, 50 Hz)
Opt. A6
Japan power plug (100 V, 110/120 V, 60 Hz)
Opt. A10
China power plug (50 Hz)
Opt. A11
India power plug (50 Hz)
Opt. A12
Brazil power plug (60 Hz)
Opt. A99
No power cord
18 www.tektronix.com
MSO3000 Series and DPO3000 Series Mixed Signal Oscilloscopes
Language options
Opt. L0
English manual
Opt. L1
French manual
Opt. L2
Italian manual
Opt. L3
German manual
Opt. L4
Spanish manual
Opt. L5
Japanese manual
Opt. L6
Portuguese manual
Opt. L7
Simplified Chinese manual
Opt. L8
Traditional Chinese manual
Opt. L9
Korean manual
Opt. L10
Russian manual
Opt. L99
No manual
Language options include translated front-panel overlay for the selected language(s).
Service options
Opt. CA1
Single Calibration or Functional Verification
Opt. D1
Calibration Data Report
Opt. R5
Repair Service 5 Years (including warranty)
Opt. SILV400
Standard warranty extended to 5 years
Probes and accessories are not covered by the oscilloscope warranty and service offerings. Refer to the datasheet of each probe and accessory model for its unique warranty
and calibration terms.
Recommended accessories
Probes
TAP1500
1.5 GHz TekVPI® active voltage probe
TAP1500X2
Bundle of two 1.5 GHz active probes, single ended with TekVPI® interface
TCP0030
120 MHz TekVPI® 30 Ampere AC/DC current probe
TCP0150
20 MHz TekVPI® 150 Ampere AC/DC current probe
TCPA300/400
3
3
Current measurement systems amplifier
TDP0500
500 MHz TekVPI® differential voltage probe with ±42 V differential input voltage
TDP1000
1 GHz TekVPI® differential voltage probe with ±42 V differential input voltage
TMDP0200
±750 V, 200 MHz high-voltage differential probe
THDP0200
±1.5 kV, 200 MHz high-voltage differential probe
THDP0100
±6 kV, 100 MHz high-voltage differential probe
P5100A
2.5 kV, 100X high-voltage passive probe
Requires TekVPI® to TekProbe® BNC adapter (TPA-BNC).
www.tektronix.com 19
Datasheet
P5200
1.3 kV, 25 MHz high-voltage differential probe
P5205
1.3 kV, 100 MHz high-voltage differential probe
P5210
5.6 kV, 50 MHz high-voltage differential probe
ADA400A 3
100X, 10X, 1X, 0.1X high-gain differential amplifier
NEX-HD2HEADER
Mictor connector breakout to 0.1 in. header pins
Accessories
071-2667-xx
Service manual (English only)
TPA-BNC
TekVPI® to TekProbe® BNC adapter
TEK-DPG
TekVPI® Deskew pulse generator signal source
067-1686-xx
Power measurement deskew and calibration fixture
119-7465-xx
TekVPI® external power supply 4
SIGEXPTE
National Instruments LabVIEW SignalExpress® Tektronix Edition software – full version
FPGAView-xx
MSO support for Altera and Xilinx FPGAs
TEK-USB-488
GPIB-to-USB adapter
ACD4000
Soft transit case
HCTEK4321
Hard transit case (requires ACD4000)
RMD3000
Rackmount kit
Tektronix is registered to ISO 9001 and ISO 14001 by SRI Quality System Registrar.
Product(s) complies with IEEE Standard 488.1-1987, RS-232-C, and with Tektronix Standard Codes and Formats.
4
Required when total oscilloscope probe power usage exceeds 20 W. Power cord not included.
20 www.tektronix.com
MSO3000 Series and DPO3000 Series Mixed Signal Oscilloscopes
www.tektronix.com 21
Datasheet
ASEAN / Australasia (65) 6356 3900
Belgium 00800 2255 4835*
Central East Europe and the Baltics +41 52 675 3777
Finland +41 52 675 3777
Hong Kong 400 820 5835
Japan 81 (3) 6714 3010
Middle East, Asia, and North Africa +41 52 675 3777
People's Republic of China 400 820 5835
Republic of Korea 001 800 8255 2835
Spain 00800 2255 4835*
Taiwan 886 (2) 2722 9622
* European toll-free number. If not accessible, call: +41 52 675 3777
Austria 00800 2255 4835*
Brazil +55 (11) 3759 7627
Central Europe & Greece +41 52 675 3777
France 00800 2255 4835*
India 000 800 650 1835
Luxembourg +41 52 675 3777
The Netherlands 00800 2255 4835*
Poland +41 52 675 3777
Russia & CIS +7 (495) 6647564
Sweden 00800 2255 4835*
United Kingdom & Ireland 00800 2255 4835*
Balkans, Israel, South Africa and other ISE Countries +41 52 675 3777
Canada 1 800 833 9200
Denmark +45 80 88 1401
Germany 00800 2255 4835*
Italy 00800 2255 4835*
Mexico, Central/South America & Caribbean 52 (55) 56 04 50 90
Norway 800 16098
Portugal 80 08 12370
South Africa +41 52 675 3777
Switzerland 00800 2255 4835*
USA 1 800 833 9200
Updated 10 April 2013
For Further Information. Tektronix maintains a comprehensive, constantly expanding collection of application notes, technical briefs and other resources to help engineers working on the cutting edge of technology. Please visit www.tektronix.com.
Copyright © Tektronix, Inc. All rights reserved. Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supersedes that in all previously published material. Specification and price
change privileges reserved. TEKTRONIX and TEK are registered trademarks of Tektronix, Inc. All other trade names referenced are the service marks, trademarks, or registered trademarks of their respective companies.
16 Jul 2013
www.tektronix.com
3GW-21364-11
Similar pages