Wall DTMPU15x-102 15 watt single output for medical equipment ac/dc desktop power supply Datasheet

DTMPU15 Series
15 Watt Single Output
For Medical Equipment
AC/DC Desktop Power Supply
Rev. B
FEATURES
● Splash Proof
● Single Output
● 3 Year Warranty
● Optional Output Connectors
● Over Temperature Detection
● A & C Types: Class I Insulation
● 3 Types of Inlet Connectors Available
● Over Voltage and Over Load Protection
● Wide Input Voltage 90 to 264VAC, 47~63Hz
● B Type: Class II Insulation (Double Insulation)
● Output Voltages Available from 5VDC to 36VDC
SPECIFICATIONS: DTMPU15 Series
Type A
Type B
Type C
o
All specifications are based on 25 C, Nominal Input Voltage, and Maximum Output Current unless otherwise noted.
We reserve the right to change specifications based on technological advances.
SPECIFICATION
INPUT (Vin)
Operating Voltage Range
Input Frequency
Input Current (Low Line)
Input Current (High Line)
Inrush Current (Low Line)
Inrush Current (High Line)
Safety Ground Leakage Current
Start-Up Time
OUTPUT (Vo)
Output Voltage Range
Load Regulation
Line Regulation
Output Power Range
Output Current Range
Ripple & Noise (peak to peak)
Transient Response
Hold-Up Time
PROTECTION
(1)
Over Voltage Protection
(1)
Over Current Protection
GENERAL
Efficiency
Dielectric Withstanding Voltage
For Primary to Secondary
Dielectric Withstanding Voltage
For Primary to Ground (A & C Types Only)
Isolation Resistance
Power Consumption (No Load)
ENVIRONMENTAL
Operating Temperature
Storage Temperature
Relative Humidity
Temperature Coefficient
Thermal Shutdown by Junction
(1)
Temperature Controller
MTBF
PHYSICAL
Weight
TEST CONDITIONS
Nom
Max
Unit
0.25
0.17
14
28
264
63
0.33
0.18
16
30
0.3
1
VAC
Hz
A
A
A
A
mA
s
90
47
Io = Full Load, Vin = 115VAC
Io = Full Load, Vin = 230VAC
Io = Full Load, 25°C, Cool Start, Vin = 115VAC
Io = Full Load, 25°C, Cool Start, Vin = 230VAC
Io = Full Load, Vin = 240VAC
Io = Full Load, Vin = 100VAC
Vin = 230VAC
Io = Full Load
Vin = 90 to 264VAC
Full Load, Vin = 90VAC
Io = Full Load to Half Load, Vin = 100VAC
Io = Full Load, Vin = 110VAC
Io = Full Load, Vin = 230VAC
0.5
See Rating Chart
3
5
0.5
1
0
15
See Rating Chart
0.5
1
4
10
16
1
VDC
%
%
W
A
%
ms
ms
112
110
132
150
%
%
73
85
%
Primary to Secondary
5600
VDC
Primary to Ground
2800
VDC
Test Voltage = 500VDC
No Load, Vin = 230VAC
Derate linearly from 100% Load at 50°C to 50% load at 70°C
All Outputs
The parameter is not subject to production test-verified by
design/characterization of integrated controller.
Operating temperature at 25°C, calculated per MIL-HDBK-217F
50
0
0
-40
5
-0.04
0.4
0.5
MΩ
W
50
70
85
95
+0.04
°C
°C
%
%/°C
+125
°C
-20
100,000
hours
Approximately 170 grams
A Type: 4.11 x 1.65 x 1.22 inches
B & C Types: 3.90 x 1.65 x 1.22 inches
IEC-320-C14, IEC-320-C8, IEC-320-C6
3
years
Dimensions
Input Inlet
Warranty
SAFETY
EMI Requirements for EN55022
EMI Requirements for FCC Part-15, FCC
Part 18
Min
Vin = 230VAC, 50Hz
B
Class
Vin = 120VAC, 60Hz
B
Class
(1) When the power system interruption is isolated, the product would re-start after recovering by hand.
Wall Industries, Inc. 5 Watson Brook Road Exeter, NH 03833
603-778-2300 www.wallindustries.com Fax 603-778-9797
DTMPU15 Series
15 Watt Single Output
For Medical Equipment
AC/DC Desktop Power Supply
Rev. B
OUTPUT VOLTAGE/ CURRENT RATING CHART
Model Number
Output Voltage
Output Current
Total Regulation (2)
Maximum Output Power
DTMPU15x-102
5 VDC
2.60A
5%
13W
DTMPU15x-103
7 VDC
1.85A
5%
13W
DTMPU15x-104
9 VDC
1.66A
5%
15W
DTMPU15x-105
12 VDC
1.25A
5%
15W
DTMPU15x-106
16 VDC
0.93A
5%
15W
DTMPU15x-107
18 VDC
0.83A
5%
15W
DTMPU15x-108
24 VDC
0.62A
3%
15W
DTMPU15x-109
30 VDC
0.50A
3%
15W
DTMPU15x-110
36 VDC
0.41A
3%
15W
NOTES
1. The DTMPU15 Series is designated as DTMPU15x-y where x represents the type of AC input inlet connector which can
either be A (IEC-320-C14), B (IEC-320-C8), or C (IEC-320-C6); y can be 102, 103, 104, 105, 106, 107, 108, 109, or
110 for output voltage.
2. Models DTMPU15-102, 103, and 105 needs to use AWG#16/4FT output cable in order to meet the total regulation.
Models DTMPU15-104, 106~110 needs to use AWG#18/4FT output cable in order to meet the total regulation.
The regulation and efficiency will change if a different output cable is used.
3. Optional output connectors are available (see “DC Output Plug Selector List” link located at the bottom of the
“Desktop” category page).
DERATING CURVE
Wall Industries, Inc. 5 Watson Brook Road Exeter, NH 03833
603-778-2300 www.wallindustries.com Fax 603-778-9797
Rev. B
MECHANICAL DRAWINGS
TYPE A
Unit: inches [mm]
TYPES B & C
Wall Industries, Inc. 5 Watson Brook Road Exeter, NH 03833
603-778-2300 www.wallindustries.com Fax 603-778-9797
DTMPU15 Series
15 Watt Single Output
For Medical Equipment
AC/DC Desktop Power Supply
Similar pages