GLD140 Gold Performance Medical Switchers 140 Watt Multiple Output FEATURES: • • • • • Compact size (4.5" x 8.5" x 1.95") Power factor corrected to IEC 1000-3-2 Class A Documented FMEA, WCA and EMC test results EMI compliance to CISPR11, FCC Class B, IEC601-1-2 Approved to UL2601, IEC601-1 and CSA 22.2 No. 601.1-M90 and EN60601: 1990 • marked to LVD SPECIFICATIONS: Ac Input 90-264 Vac, 47-63 Hz single phase. Output Power Continuous output power 140 W with unrestricted convection cooling, 180 W with 150 LFM of air. Input Current Active Power Factor Correction circuitry assures compliance with IEC1000-3-2, Class A. Maximum input current at minimum input voltage and full load is 3.0 A. Efficiency 73 – 85 % at full rated load, nominal input voltage, depending on model and load distribution. Hold-up Time 26 ms total. 16 ms minimum from loss of ac input at full load until activation of the power fail signal. Output voltages remain within specified regulation limits for an additional 9 ms minimum after power fail activates. Overload Protection Fully protected against short circuit and output overload. Short circuit protection is cycling type power limit. Output Noise 0.5% rms, 1% pk-pk, 20 MHz Bandwidth, differential mode. Measured with noise probe directly across output terminals of the power supply. Transient Response 500 µs typical response time for return to within 0.5% of final value for a 45% load step from any load greater than 10% of full rated load, ∆i/∆t< 0.2 A µs. Maximum voltage deviation is 3.5%. Remote Sense Provided as a standard feature on V1 and V2. Overvoltage Protection Built in on V1, 2 and 3 outputs. Voltage Adjustment Factory set to specified voltage with user adjustable potentiometer on V1, 2 and 3. User can adjust V1, 2 and 3 at least ±5% of nominal output voltage. Output Regulation Regulation for all outputs is the maximum deviation from initial set point under all line and load conditions. Initial set tolerance is measured with all outputs at 50 % of full rated load. No Load Turn-On/Standby A minimum load of 3 A on V1 is required for proper regulation. If not met, no degradation of reliability will occur. Temperature Coefficient 0.03% / °C typical on all outputs. Input Protection Internal ac fuses provided on both lines on all units. Inrush Current Inrush is limited by internal thermistors. The inrush at 240 Vac, averaged over the first ac half-cycle under cold start conditions will not exceed 37 A. 58 Thermal Shutdown Provided as a standard feature. Designed to protect unit from prolonged over temperature. Power Fail TTL / CMOS compatible output goes low (<0.5 V) 10 ms before output voltage drops more than 4% below nominal voltage upon loss of ac power. Power Good TTL / CMOS compatible output rises high 100 to 300 ms after V1 reaches regulation and should assure that sufficient energy is stored in the input section to provide normal power fail / shutdown. Inhibit Inhibit signal when pulled to the V1 output common will inhibit all output voltages. Fan Output An additional thermally controlled 12 Vdc, 250 mA output suitable for powering a dc fan is included in all models (Factory installed cover with fan makes this output unavailable). EMI/EMC Compliance All models include built-in EMI filtering to meet the following requirements: EMC SPECIFICATION COMPLIANCE LEVEL Conducted Emissions EN55011, Class B; FCC Class B Static Discharge EN61000-4-2, 6 kV contact 8 kV air RF Field Susceptibility EN61000-4-3, 3V/meter Fast Transients / Bursts EN61000-4-4, 2 kV, 5 kHz Surge Susceptibility EN61000-4-5, 1 kV diff., 2 kV com. Conducted RF Susceptibility EN61000-4-6, 3V Voltage Sags & Surges EN61000-4-11 Line Frequency Harmonics EN61000-3-2 Class A EMC Compliance for Gold Series Magnetic Emissions - Emissions will not exceed the limits of the Army curve in MIL-STD-461D, RE101 when measured at 7 cm from 30 Hz - 100 kHz over all nominal inputs and at full rated load. Leakage Current 70 µA under normal conditions 120 Vac @ 60 Hz Single fault conditions 280 µA, 264 Vac @ 50 Hz. SAFETY AGENCY APPROVALS: All models are Certified to be in compliance with the applicable requirements of UL2601-1, CSA-C22.2 No. 601.1, IEC 601-1/60601-1. DESIGN VERIFICATION DOCUMENTS The “Gold” series has undergone rigorous review and design analysis. The following product documentation is available upon request: 1. Failure Mode and Effects analysis (FMEA), 2. DVT Data, 3. EMC / Susceptibility test results. Condor D.C. Power Supplies, Inc., 2311 Statham Parkway, Oxnard, CA 93033 800-235-5929 • 805-486-4565 • FAX 805-487-8911 • www.condorpower.com GLD140 Medical Switchers 140 Watt Multiple Output Medical Model Output N o. Output Vo l t a g e Output Current (A) Output Current (B) Set To l e r a n c e Line Load OV P Tr i p GLD140A 1 2 3 4 +5.0 V +12 V 12 V (C) -12 V 16 A 6A 4A 0.75 A 20 A 8A 5A 1.2 A 1% 1% 1% 2.2% 1% 1% 1% 1% 6.2 ± 0.3 V 14 ± 0.7 V 14 ± 1.2 V ------- GLD140B 1 2 3 4 1 2 3 4 +5.0 V +12 V 5 V (C) -12 V +5.0 V +12 V 15 V (C) -15 V 16 A 6A 4A 0.75 16 A 6A 4A 0.75 A 20 A 8A 5A 1.2 A 20 A 8A 5A 1.2 A 1% 1% 1% 2.2% 1% 1% 1% 2.2% 1% 1% 1% 1% 1% 1% 1% 1% 6.2 ± 0.3 V 14 ± 0.7 V 6.3 ± .5 V ------6.2 ± 0.3 V 14 ± 0.7 V 18 ± 1.2 V ------- 1% 1% 1% 1% 1% 1% 1% 1% 1% 1% 1% 1% 1 2 3 4 +5.0 V +24 V 12 V (C) -12 V 16 A 4A 4A 0.75 20 A 6A 5A 1.2 A 1% 1% 1% 2.2% 1% 1% 1% 1% 6.2 ± 0.3 V 27 ± 1 V 14 ± 1.2 V ------- 1% 1% 1% 1% 1 2 3 4 +5.0 V +24 V 15 V (C) -15 V 16 A 4A 4A 0.75 A 20 A 6A 5A 1.2 A 1% 1% 1% 2.2% 1% 1% 1% 1% 6.2 ±0.3 V 27 ± 1 V 18 ± 1.2 V ------- 1% 1% 1% 1% 1 2 3 4 +5.0 V +15 V 12V (C) -12 V 16 A 5A 4A 0.75 A 20 A 7A 5A 1.2 A 1% 1% 1% 2.2% 1% 1% 1% 1% 6.2 ± 0.3 V 18 ± 1.2 V 14 ± 1.2 V ------- 1% 1% 1% 1% 1 2 3 4 +5.0 V +3.3 V 12 V (C) -12 V 16 A 10 A 4A 0.75 A 20 A 12 A 5A 1.2 A 1% 1% 1% 2.2% 1% 1% 1% 1% 6.2 ± 0.3 V 4.2 ± 0.4 V 14 ± 1.2 V ------- 1% 1% 1% 1% GLD140C GLD140D GLD140E GLD140F GLD140G Ripple and Noise Output Current for Individual Outputs: A. Output current for unrestricted natural convection. B. Output current with 150 LFM forced air convection or peak current rating. C. Isolated (floating) output may be referenced positive or negative. GLD140 MECHANICAL SPECIFICATIONS INPUT: JI MOLEX P.C.B. HEADER P/N 39-30-2056 PIN 1) AC GROUND PIN 2) N/C PIN 3) AC NEUTRAL PIN 4) N/C PIN 5) AC LINE MATING CONNECTOR MOLEX P/N: HOUSING 39-01-4051, CONTACT 39-00-0164 SIGNALS: J2 AMP P.C.B. HEADER P/N: 641215-6 PIN 1) +SENSE PIN 2) -SENSE PIN 3) POWER FAIL PIN 4) POWER GOOD PIN 5) COMMON PIN 6) INHIBIT MATING CONNECTOR AMP P/N: HOUSING 770602-6, CONTACT 770666-2 OUTPUT: J3 MOLEX P.C.B. HEADER P/N: 39-29-9206 PINS 1) +V3out PINS 2) -V4out PINS 3,12,13) COMMON PINS 4) +V2 SENSE PINS 5,14,15) +V2out PINS 6-8, 16) COMMON PINS 9,10,19,20) +V1out PINS 11) +V3 RTN PINS 17,18) COMMON MATING CONNECTOR MOLEX P/N: HOUSING 39-01-2200, CONTACT 39-00-0164 FAN: J5 AMP P.C.B. HEADER P/N 641215-2 PINS 1) FAN RTN PINS 2) +FAN MATING CONNECTOR AMP P/N: HOUSING 770602-2, CONTACT 770666-2 Environmental Specification Operating Non-operating Temperature (A) 0 to 50oC -40 to +85oC 0 to 95% RH 0 to 95% RH Humidity (A) Shock (B) Altitude Vibration (C) 20 gpk 40 gpk -500 to 10,000 ft -500 to 40,000 ft 1.5 grms, 0.003 g2/Hz 5 grms, 0.026 g2/Hz A. Units should be allowed to warm up/operate under non-condensing conditions before application of power. B. Random vibration—10 to 2000Hz, 6dB/octave roll-off from 350 to 2000Hz, 3 orthogonal axes. Tested for 10 min./axis operating and 1 hr./axis non-operating. C. Shock testing—half-sinusoidal, 10 ± 3 ms duration, ± direction, 3 orthogonal axes, total 6 shocks. Condor D.C. Power Supplies, Inc., 2311 Statham Parkway, Oxnard, CA 93033 800-235-5929 • 805-486-4565 • FAX 805-487-8911 • www.condorpower.com 59