Bel MBC75-1058L 3 x 2 x 1 inch form factor Datasheet

The MBC75 Series of open frame medical power supplies feature a
wide universal AC input range of 85 V – 264 VAC, offering 75 W of
output power in a compact footprint, with a variety of isolated single
output voltages.
The MBC series is designed and approved to the latest Medical
standards (EN/IEC 60601-1), providing 2 x MOPP isolation for Class I
& Class II applications.
These power supplies are ideal for medical, telecom, datacom,
industrial equipment and other applications.
•
3 x 2 x 1 Inch Form Factor
•
75 Watts with Convection Cooling
•
Approved to EN/IEC 60601-1
•
Efficiencies up to 93%
•
-40 To 70°C Operating Temperature
•
Dual Fusing
•
Thermal Shut-Down Feature
•
2 Million Hours, Telcordia -SR332-Issue 3
•
Standby Power < 0.3 W
•
Suitable for BF Applications
•
Class II Option Available
•
RoHS Compliant
•
CE Marked
•
Diagnostic
•
Home Health Care
•
Drug Pump
•
Monitoring
•
Dialysis
•
Portable Equipment
MBC75 Series
2
MBC75-1T12L
MBC75-1012L
MBC75-1T15L
MBC75-1015L
MBC75-1T24L
MBC75-1024L
MBC75-1T30L
MBC75-1030L
MBC75-1T48L
MBC75-1048L
MBC75-1T58L
MBC75-1058L
COVER-75-XBC3
Screw Terminal
Molex Header
Screw Terminal
Molex Header
Screw Terminal
Molex Header
Screw Terminal
Molex Header
Screw Terminal
Molex Header
Screw Terminal
Molex Header
12 V
6.25 A
0.0 A
1%
15 V
5.00 A
0.0 A
1%
24 V
3.12 A
0.0 A
1%
30 V
2.50 A
0.0 A
1%
48 V
1.56 A
0.0 A
1%
58 V
1.29 A
0.0 A
1%
metal cover kit accessory
Specifications are for nominal input voltage, 25°C unless otherwise stated.
Input Voltage
Universal
(Derate from 75 W @ 100 VAC to 65 W @ 85 VAC)
Input Frequency
85-264 VAC / 390 VDC4
47-63 Hz
Input Current
115 VAC:
230 VAC:
1 A max.
0.5 A max.
No Load Power
Typical
< 0.3 W
115 VAC:
230 VAC:
264 VAC:
Typical (N.A. For Class II Option)
Touch current
25 A
45 A
75 A
300 uA
<100 uA
Power Factor
@ Full Load, Active PFC
> 0.95
Switching Frequency
Typical
60 kHz
Inrush Current
Leakage Current
1
For Class II Option (without input Earth pin) add suffix: -2 (e.g.: MBC75-1012L-2).
Ripple is peak to peak with 20 MHz bandwidth and 10 μF (Tantalum capacitor) in parallel with a 0.1 μF capacitor at rated line voltage
and load ranges.
3
When used in Cover Kit, de-rate output power to 70 % under all operating conditions
4
Functional, not approved.
2
[email protected]
MBC75 Series
3
Output Power
Convection:
75 W
Efficiency
48 V, 58 V:
24 V, 30 V:
12 V, 15 V:
93%
91%
90%
Hold-up Time
Typical
>16 ms
Line Regulation
+/-0.5%
Load Regulation
Transient Response
+/-1%
25% step load change, at 0.1A/uS slew rate,
50% duty cycle, 50 Hz = 4%
Voltage Adjustment
Rise Time
+/-3%
Typical
55 ms
Set Point Tolerance
+/-1%
Over Current Protection
> 110%
Over Voltage Protection
Latch type (AC recycling required)
Short Circuit Protection
Hiccup mode
Operating Temperature5
recovery time < 5 ms
110 to 140%
Startup is guaranteed with spec deviation, see Fig 1.
Storage Temperature
-40 to +70°C
-40 to 0°C
-40 to +85°C
Cooling
With natural convection cooling at 100 to 264 VAC
75 W
Relative Humidity
Noncondensing
5% to 95%
Altitude
Operating:
Nonoperating:
16,000 ft
40,000 ft.
Reliability
MTBF according to Telcordia -SR332-Issue 3
2.00 million hours
Conducted Emissions
Input Current Harmonics
EN 55011-B, CISPR22-B, FCC PART15-B
EN 55011 A;
with external core (King core K5B RC 25x12x15-M in input cable)
EN 61000-3-2
Pass
Pass
Level B
Class D
Voltage Fluctuation and Flicker
EN 61000-3-3
Pass
ESD Immunity
EN 61000-4-2
Level 4, Criterion A
Radiated Field Immunity
EN 61000-4-3
Level 3, Criterion A
Electrical Fast Transient Immunity
EN 61000-4-4
Level 3, Criterion A
Surge Immunity
EN 61000-4-5
Level 3, Criterion A
Conducted Immunity
EN 61000-4-6
Level 3, Criterion A
Magnetic Field Immunity
EN 61000-4-8
Level 4, Criterion A
Voltage Dips, Interruptions
EN 61000-4-11
Criterion B
Radiated Emissions
5
Output ripple can be more than 10% of the output voltage.
Europe, Middle East
+353 61 225 977
North America
+1 408 785 5200
© 2017 Bel Power Solutions & Protection
BCD.00665_AC
Asia-Pacific
+86 755 298 85888
MBC75 Series
4
Agency Approvals
Input to Output: (For medical applications)
Input to GND: (Not Applicable For Class II Option)
Output to GND: for type BF
for type B (N/A For Class II Option)
Primary to Secondary:
Primary to Earth:
Secondary to Earth:
Approved to the latest edition of the following standards:
CSA/UL60601-1, EN60601-1 and IEC60601-1.
Nemko, cULus, CB
CE mark
Complies with LVD Directive
Isolation Voltage
Protection Level
Safety Standard(s)
4000 VAC
1500 VAC
1500 VAC
500 VAC
2 MOPP
1 MOPP
1 MOPP
De-rate linearly from 100% at 50°C
to 50% at 70°C
Figure 1. Derating Curve
AC Input Connector
DC Output Connector
J1
J2
Pin 1
Pin 2
Pin 3
AC Line
Not Fitted
AC Neutral
Pin 1, 2
Pin 3, 4
V1 -VE
V1 +VE
Screw Terminal (Option 1)
Molex Header (Option 2)
Screw Terminal (Option 1)
Molex Header (Option 2)
Weight
150 g
Dimensions
76.2 x 50.8 x 25.4 mm (3 x 2 x 1 inch)
Molex: 39357-0003
Tyco-2-1776112-3
Molex: 1722861103
(Mating conn: Molex 1722561003)
Molex: 39357-0004
Tyco-2-1776112-4
Molex: 1722861104
(Mating conn: Molex 1722561004)
[email protected]
MBC75 Series
5
Figure 2. Mechanical Drawing - Screw Terminal (Option 1)
Figure 3. Mechanical Drawing - Molex Header (Option 2)
NOTES: In case the PCB is mounted in a metal enclosure, using metal hardware ensure the following:
1
Stand off, used to mount PCB has OD of 5.4 mm max.
2
Screws, used to fix PCB on stand off, have head dia of 6.0 mm max.
3
Washer, if used, to have dia of 6.5 mm max.
NUCLEAR AND MEDICAL APPLICATIONS - Products are not designed or intended for use as critical components in life support systems,
equipment used in hazardous environments, or nuclear control systems.
TECHNICAL REVISIONS - The appearance of products, including safety agency certifications pictured on labels, may change depending on the
date manufactured. Specifications are subject to change without notice.
Europe, Middle East
+353 61 225 977
North America
+1 408 785 5200
© 2017 Bel Power Solutions & Protection
BCD.00665_AC
Asia-Pacific
+86 755 298 85888
Similar pages