XP Power JHM1512S05 Dc-dc converter Datasheet

JHM15 Series
DC-DC Converter
15 Watts
xxx Series
• International Medical Approvals
• 4000 VAC Reinforced Insulation
• Medical Approval, IEC60601-1, 3rd Edition
• 2 µA Patient Leakage Current
• Compact 1 x 1.6” Footprint
• EN55011 Level A With No External Components
• 3 Year Warranty
Dimensions:
JHM15:
1.60 x 1.00 x 0.40” (40.60 x 25.40 x 10.20 mm)
Models & Ratings
Input Voltage
Output Voltage
9-18 V
18-36 V
Input Current
No Load(1)
Full Load(2)
Output Current
Maximum
Capacitive Load(3)
Efficiency(4)
Model Number
5.0 V
3000 mA
9.2 mA
1930 mA
3000 µF
87%
JHM1512S05
12.0 V
1250 mA
6.5 mA
1938 mA
1330 µF
86%
JHM1512S12
15.0 V
1000 mA
8.0 mA
1944 mA
1000 µF
86%
JHM1512S15
±5.0 V
±1500 mA
6.6 mA
1955 mA
±1470 µF
84%
JHM1512D05
±12.0 V
±625 mA
11.2 mA
1911 mA
±660 µF
87%
JHM1512D12
±15.0 V
±500 mA
11.0 mA
1879 mA
±550 µF
88%
JHM1512D15
5.0 V
3000 mA
5.6 mA
972 mA
3000 µF
86%
JHM1524S05
12.0 V
1250 mA
6.1 mA
968 mA
1830 µF
85%
JHM1524S12
15.0 V
1000 mA
6.4 mA
966 mA
1000 µF
87%
JHM1524S15
±5.0 V
±1500 mA
5.4 mA
981 mA
±1470 µF
83%
JHM1524D05
±12.0 V
±625 mA
7.3 mA
954 mA
±660 µF
87%
JHM1524D12
±15.0 V
±500 mA
8.5 mA
943 mA
±550 µF
87%
JHM1524D15
Notes
1. Input current measured at nominal input voltage.
2. Input current measured at lowest input voltage.
3. Maximum capacitive load is per output.
4. Typical values.
Mechanical Details
0.400 (10.16)
0.100 (2.54)
1.60 (40.6)
1.40 (35.56)
6
ø 0.040 (1.02)
0.40 (10.2)
0.22 (5.6) min
2
1
BOTTOM VIEW
5
4
3
0.400 (10.16)
1.00 (25.4)
1
Pin
Single
Dual
1
+Vin
+Vin
2
-Vin
-Vin
3
+Vout
+Vout
Common
4
-Vout
5
Trim
-Vout
6
No Pin
No Pin
Notes
0.100 (2.54)
0.400 (10.16)
Pin Connections
0.200 (5.08)
0.200 (5.08)
www.xppower.com
1. All dimensions are in inches (mm)
2. Weight: 0.04 lbs (20 g) approx.
3. Pin diameter: 0.02 ±0.002 (0.5 ±0.05)
4. Pin pitch tolerance: ±0.01 (±0.25)
5. Case tolerance: ±0.02 (±0.5)
JHM15 Series
DC-DC Converter
Input
Characteristic
Input Voltage Range
Minimum
Maximum
Units
9
Typical
18
VDC
12 V nominal
18
36
VDC
24 V nominal
20
A
2
µA
Input Current
See Models and Ratings table
Inrush Current
Input Filter
at 36 V
Pi type
Patient Leakage Current
Undervoltage Lockout
Notes & Conditions
On at >8.8 V. Off <8.3 V
12 V models
On at >17.5 V. Off <17.0 V
24 V models
Input Surge
25
VDC
12 V models for 3 s
50
VDC
24 V models for 3 s
Maximum
Units
15
V
See Models and Ratings table
±10
%
Via external resistors, see Application Notes
±1
%
on V1
±2
%
on V2 of dual output models
A
No minimum load required
Output
Characteristic
Minimum
Output Voltage
5
Typical
Output Voltage Trim
Initial Set Accuracy
Minimum Load
0
Start Up Delay
5
ms
Start Up Rise Time
2
ms
Line Regulation
Load Regulation
Notes & Conditions
±0.3
%
±2
%
±1
%
10 - 100% load
On dual output models with one output set to 50% load and
the other varied from 10% to 100% load (D05 20% to 100%)
0 - 10% load
Cross Regulation
±4
%
Transient Response
4
% deviation
Ripple & Noise
1
% pk-pk
Short Circuit Protection
Recovery to within 1% in <500 µs for a 50% load change
at 0.25 A/µs rate
20 MHz bandwidth
Trip & Restart (hiccup mode), auto recovery
Overload Protection
120
200
%
Trip & Restart (hiccup mode)
Overvoltage Protection
115
140
%
Non latching, auto recovery
0.03
%/ ˚C
Maximum
Units
Temperature Coefficient
General
Characteristic
Minimum
Efficiency
Typical
86
%
Isolation
4000
Input to Output Capacitance
20
Switching Frequency
250
23
Weight
See Models and Ratings table
VAC
For 1 min. Double/reinforced with a working voltage of 250
VAC. Meets 2 x MOPP per 3rd edition of IEC60601-1 5000
VAC for 10 ms in accordance with IEC60664-1
pF
kHz
Power Density
Mean Time Between Failure
Notes & Conditions
W/in3
>1
MHrs
0.04 (20.0)
lb (g)
MIL-HDBK-217F, +25 °C GB
Environmental
Characteristic
Minimum
Operating Temperature
-40
Storage Temperature
-55
Case Temperature
Humidity
5
Typical
Maximum
Units
Notes & Conditions
+80
°C
See derating curve
+100
°C
+100
°C
90
%RH
Cooling
Non-condensing
Natural convection
Shock
±3 shocks in each plane, total 18 shocks of 30 g : 11 ms halfsine. Conforms to EN60068-2-27 & EN60068-2-47
Vibration
10-500 Hz at 2 g sweep and endurance at resonance in all 3 planes. Conforms to EN60068-2-6
www.xppower.com
2
JHM15 Series
DC-DC Converter
EMC: Emissions
Standard
Test Level
Conducted
EN55011 & EN55022
Level A
Radiated
EN55011 & EN55022
Level A
Phenomenon
Notes & Conditions
EMC: Immunity
Phenomenon
Immunity
Test Level
Standard
Notes & Conditions
Criteria
IEC60601-1-2, EN61204-3
ESD Immunity
EN61000-4-2
2
A
Radiated Immunity
EN61000-4-3
10 V/m
A
EFT/Burst
EN61000-4-4
2
A
Surges
EN61000-4-5
1
A
Conducted Immunity
EN61000-4-6
10 Vm
A
Magnetic Fields
EN61000-4-8
3 A/m
A
Safety Approvals
ANSI/AMMI ES60601-1 3rd Edition, CSA-22.2 No.60601-1:2008, IEC60601-1 3rd Edition
Safety Approvals
Safety Agency
Safety Standard
Notes & Conditions
CB Report
IEC60601-1 Ed 3 Including Risk Management
Medical
UL
ANSI/AAMI ES60601-1 3rd Ed. & CSA C22.2, No.60601-1:2008
Medical
Application Notes
Derating Curve
100
Load (%)
80
60
40
20
0
-40 -30
-20
-10
0
10
20
30
40
50
60
70
80
90
Ambient Temperature (°C)
External Output Trim
Trim
+ Vout
Trim Up
-Vout
3
Trim Down
Trim
For 5 V output:
Trim +10%, R = 3.4 k typical
Trim -10%, R = 1.1 k typical
For ±5 V output:
Trim +10%, R = 12.0 k typical
Trim -10%, R = 8.0 k typical
For 12 V output:
Trim +10%, R = 5.9 k typical
Trim -10%, R = 11.3 k typical
For ±12 V output:
Trim +10%, R = 12.8 k typical
Trim -10%, R = 9.5 k typical
For 15 V output:
Trim +10%, R = 8.4 k typical
Trim -10%, R = 10.4 k typical
For ±15 V output:
Trim +10%, R = 18 k typical
Trim -10%, R = 14.8 k typical
www.xppower.com
07 July 15
Similar pages