JHM15 Series DC-DC Converter 15 Watts xxx Series • International Medical Approvals • 4000 VAC Reinforced Insulation • Medical Approval, IEC60601-1, 3rd Edition • 2 µA Patient Leakage Current • Compact 1 x 1.6” Footprint • EN55011 Level A With No External Components • 3 Year Warranty Dimensions: JHM15: 1.60 x 1.00 x 0.40” (40.60 x 25.40 x 10.20 mm) Models & Ratings Input Voltage Output Voltage 9-18 V 18-36 V Input Current No Load(1) Full Load(2) Output Current Maximum Capacitive Load(3) Efficiency(4) Model Number 5.0 V 3000 mA 9.2 mA 1930 mA 3000 µF 87% JHM1512S05 12.0 V 1250 mA 6.5 mA 1938 mA 1330 µF 86% JHM1512S12 15.0 V 1000 mA 8.0 mA 1944 mA 1000 µF 86% JHM1512S15 ±5.0 V ±1500 mA 6.6 mA 1955 mA ±1470 µF 84% JHM1512D05 ±12.0 V ±625 mA 11.2 mA 1911 mA ±660 µF 87% JHM1512D12 ±15.0 V ±500 mA 11.0 mA 1879 mA ±550 µF 88% JHM1512D15 5.0 V 3000 mA 5.6 mA 972 mA 3000 µF 86% JHM1524S05 12.0 V 1250 mA 6.1 mA 968 mA 1830 µF 85% JHM1524S12 15.0 V 1000 mA 6.4 mA 966 mA 1000 µF 87% JHM1524S15 ±5.0 V ±1500 mA 5.4 mA 981 mA ±1470 µF 83% JHM1524D05 ±12.0 V ±625 mA 7.3 mA 954 mA ±660 µF 87% JHM1524D12 ±15.0 V ±500 mA 8.5 mA 943 mA ±550 µF 87% JHM1524D15 Notes 1. Input current measured at nominal input voltage. 2. Input current measured at lowest input voltage. 3. Maximum capacitive load is per output. 4. Typical values. Mechanical Details 0.400 (10.16) 0.100 (2.54) 1.60 (40.6) 1.40 (35.56) 6 ø 0.040 (1.02) 0.40 (10.2) 0.22 (5.6) min 2 1 BOTTOM VIEW 5 4 3 0.400 (10.16) 1.00 (25.4) 1 Pin Single Dual 1 +Vin +Vin 2 -Vin -Vin 3 +Vout +Vout Common 4 -Vout 5 Trim -Vout 6 No Pin No Pin Notes 0.100 (2.54) 0.400 (10.16) Pin Connections 0.200 (5.08) 0.200 (5.08) www.xppower.com 1. All dimensions are in inches (mm) 2. Weight: 0.04 lbs (20 g) approx. 3. Pin diameter: 0.02 ±0.002 (0.5 ±0.05) 4. Pin pitch tolerance: ±0.01 (±0.25) 5. Case tolerance: ±0.02 (±0.5) JHM15 Series DC-DC Converter Input Characteristic Input Voltage Range Minimum Maximum Units 9 Typical 18 VDC 12 V nominal 18 36 VDC 24 V nominal 20 A 2 µA Input Current See Models and Ratings table Inrush Current Input Filter at 36 V Pi type Patient Leakage Current Undervoltage Lockout Notes & Conditions On at >8.8 V. Off <8.3 V 12 V models On at >17.5 V. Off <17.0 V 24 V models Input Surge 25 VDC 12 V models for 3 s 50 VDC 24 V models for 3 s Maximum Units 15 V See Models and Ratings table ±10 % Via external resistors, see Application Notes ±1 % on V1 ±2 % on V2 of dual output models A No minimum load required Output Characteristic Minimum Output Voltage 5 Typical Output Voltage Trim Initial Set Accuracy Minimum Load 0 Start Up Delay 5 ms Start Up Rise Time 2 ms Line Regulation Load Regulation Notes & Conditions ±0.3 % ±2 % ±1 % 10 - 100% load On dual output models with one output set to 50% load and the other varied from 10% to 100% load (D05 20% to 100%) 0 - 10% load Cross Regulation ±4 % Transient Response 4 % deviation Ripple & Noise 1 % pk-pk Short Circuit Protection Recovery to within 1% in <500 µs for a 50% load change at 0.25 A/µs rate 20 MHz bandwidth Trip & Restart (hiccup mode), auto recovery Overload Protection 120 200 % Trip & Restart (hiccup mode) Overvoltage Protection 115 140 % Non latching, auto recovery 0.03 %/ ˚C Maximum Units Temperature Coefficient General Characteristic Minimum Efficiency Typical 86 % Isolation 4000 Input to Output Capacitance 20 Switching Frequency 250 23 Weight See Models and Ratings table VAC For 1 min. Double/reinforced with a working voltage of 250 VAC. Meets 2 x MOPP per 3rd edition of IEC60601-1 5000 VAC for 10 ms in accordance with IEC60664-1 pF kHz Power Density Mean Time Between Failure Notes & Conditions W/in3 >1 MHrs 0.04 (20.0) lb (g) MIL-HDBK-217F, +25 °C GB Environmental Characteristic Minimum Operating Temperature -40 Storage Temperature -55 Case Temperature Humidity 5 Typical Maximum Units Notes & Conditions +80 °C See derating curve +100 °C +100 °C 90 %RH Cooling Non-condensing Natural convection Shock ±3 shocks in each plane, total 18 shocks of 30 g : 11 ms halfsine. Conforms to EN60068-2-27 & EN60068-2-47 Vibration 10-500 Hz at 2 g sweep and endurance at resonance in all 3 planes. Conforms to EN60068-2-6 www.xppower.com 2 JHM15 Series DC-DC Converter EMC: Emissions Standard Test Level Conducted EN55011 & EN55022 Level A Radiated EN55011 & EN55022 Level A Phenomenon Notes & Conditions EMC: Immunity Phenomenon Immunity Test Level Standard Notes & Conditions Criteria IEC60601-1-2, EN61204-3 ESD Immunity EN61000-4-2 2 A Radiated Immunity EN61000-4-3 10 V/m A EFT/Burst EN61000-4-4 2 A Surges EN61000-4-5 1 A Conducted Immunity EN61000-4-6 10 Vm A Magnetic Fields EN61000-4-8 3 A/m A Safety Approvals ANSI/AMMI ES60601-1 3rd Edition, CSA-22.2 No.60601-1:2008, IEC60601-1 3rd Edition Safety Approvals Safety Agency Safety Standard Notes & Conditions CB Report IEC60601-1 Ed 3 Including Risk Management Medical UL ANSI/AAMI ES60601-1 3rd Ed. & CSA C22.2, No.60601-1:2008 Medical Application Notes Derating Curve 100 Load (%) 80 60 40 20 0 -40 -30 -20 -10 0 10 20 30 40 50 60 70 80 90 Ambient Temperature (°C) External Output Trim Trim + Vout Trim Up -Vout 3 Trim Down Trim For 5 V output: Trim +10%, R = 3.4 k typical Trim -10%, R = 1.1 k typical For ±5 V output: Trim +10%, R = 12.0 k typical Trim -10%, R = 8.0 k typical For 12 V output: Trim +10%, R = 5.9 k typical Trim -10%, R = 11.3 k typical For ±12 V output: Trim +10%, R = 12.8 k typical Trim -10%, R = 9.5 k typical For 15 V output: Trim +10%, R = 8.4 k typical Trim -10%, R = 10.4 k typical For ±15 V output: Trim +10%, R = 18 k typical Trim -10%, R = 14.8 k typical www.xppower.com 07 July 15