Delta MDS-400APB24-AA Medical ac-dc open frame power supply Datasheet

Medical AC-DC Open Frame Power Supply
24V/400W 3”x5” High Density MDS-400APB24 AA
MDS-400APB24 AA
Highlights & Features
– 3" x 5" x 1.4" Package
– Safety Approvals to IEC 60601-1 3.1rd ed. & IEC 60950-1
– Compliant with IEC 60601-1-2 4th Ed. Requirements
– Up to 200W convection, 400W force air
– IT & Medical Safety Approvals
Safety Standards
CB Certified for worldwide use
Model Number:
MDS-400APB24 AA
Unit Weight:
405g
Dimensions (W x L x H): 76.2x127x35.6mm
3x5x1.4 inch
General Description
The MDS-400APB24 of internal open frame power supplies come with universal AC input range from 90Vac to 264Vac.
Other features include low leakage, Type BF Patient Access Leakage Currents, and electric shock protection compliance
with 2 x MOPP requirements. The MDS series is certified for EMC standards according to EN 55011for industrial,
scientific and medical (ISM) radio-frequency equipment; and, EN 55022 for Industrial Technology Equipment (ITE) radiofrequency equipment. In addition, only recognized Japanese capacitors are used to ensure long product life.
The MDS-400A series comes with both medical and ITE safety approvals, including UL/CE/CCC (5000meters), and CB
certification. Designs are compliant with RoHS Directive 2011/65/EU for environmental protection.
Model Information
Medical AC-DC Open Frame
Model Number
Main Output Main Output Standby
Voltage
Current
Output
Voltage
Standby
Output
Current
Fan Output
Voltage
Fan Output
Current
MDS-400APB24 AA
24Vdc
0-2A
12Vdc
0.05-0.6A
0-16.66A
5Vdc
1)
Total Max Output Power
2)
400W
1) Fan output will be presenting when 24V main output is available
2)With 10CFM force air
Model Numbering
MDS
400
A
P
B
24
Delta Medical
power Supply
Max wattage in
the product
Series.
Maybe lower at
some voltage.
060→60W
150 →150W
1K2 → 1,200W
Family Code
Product Type
Input Type Code
Output Voltage
A~ Z
P: Open Frame
B: 3pin ClassI
24 for24V
1
All parameters are specified at 25°C ambient unless otherwise noted.
www.DeltaPSU.com (November 2016, Rev. 08)
A
Revision Code
A
Revision code
Medical AC-DC Open Frame Power Supply
24V/400W 3”x5” High Density MDS-400APB24 AA
Specifications
Input Ratings / Characteristics
Nominal Input Voltage
100-240Vac
Input Voltage Range
90-264Vac
Nominal Input Frequency
50-60Hz
Input Frequency Range
47-63Hz
Input Current (max)
5.5A
Input Surge Voltage (max)
300Vac for 100ms
Full load Efficiency (typ.)
92% @ 115Vac/60Hz
93% @ 230Vac/50Hz, Reference Fig.1
Standby Power (max)
0.5W (only standby working with Inhibit signal high)
@ 115Vac/60Hz, 230Vac/50Hz
Inrush Current (max)
40A@230Vac, cold start
Input-PE(protective earth) leakage current(max)
0.1mA @ NC, 0.3mA @ SFC1)
Output-PE(protective earth) leakage current for Type BF application (max)
0.1mA @ NC, 0.5mA @ SFC 1)
Power Factor (min)
0.95 @ 115V/50Hz, 230V/50Hz, full load
1) NC: normal condition, SFC: single fault condition
Leakage Current
Input-PE Leakage Current
100Vac/60Hz(Typ)
264Vac/60Hz(Typ) Delta Limit
IEC60601-1 Limit
Normal Condition 18.3uA
44.6uA
100uA max
5000uA max
Single Fault Condition 33.7uA
91.9uA
300uA max
10000uA max
87.3uA
100uA max
100uA max
130.5uA
500uA max
500uA max
Output-PE Leakage Current forType BF application
Normal Condition 29.4uA
Single Fault Condition 43.5uA
100.000
90.000
Efficiency (%)
80.000
115Vac
70.000
230Vac
60.000
50.000
25%
50%
Output Load
Fig.1 Efficiency versus output load
2
All parameters are specified at 25°C ambient unless otherwise noted.
www.DeltaPSU.com (November 2016, Rev. 08)
75%
100%
Medical AC-DC Open Frame Power Supply
24V/400W 3”x5” High Density MDS-400APB24 AA
Output Ratings / Characteristics
Nominal Output Voltage (Vrated)
24V
Output Voltage Tolerance
±3%
Output Power
400W max with 10CFM air cooling
Line Regulation (max)
±0.5%
Load Regulation (max)
±1%
Ripple& Noise(typ.)
1%pk-pkVrated@ Full load, Reference Fig. 2
Start-up Time(max)
2000ms@115Vac
Hold-up Time (min)
10ms@115Vac, with 400W load
Dynamic Response (Overshoot & Undershoot O/P Voltage)
±5% @50-100% load
Capacitive load (max)
1500uF
Rise time (max)
100ms
Remote Sense
Compensate up to 500mV lead drop with remote sense
Short and reverse connection protected. PSU can work
normally with remote sense pins left open.
(a) 115V (measured value=106mV)
(b) 230V(measured value=104mV)
Fig.2Ripple & Noise example, 20MHz BW
Ripple & Noise measurement circuit
Load
PSU
5cm
0.47uF Ceramic
10uF Electrolytic
Test Fixture
3
All parameters are specified at 25°C ambient unless otherwise noted.
www.DeltaPSU.com (November 2016, Rev. 08)
Scope
Medical AC-DC Open Frame Power Supply
24V/400W 3”x5” High Density MDS-400APB24 AA
Mechanical
Case Cover
NA
Dimensions (W x L x H typ.)
76.2x127x35.6mm (3x5x1.4inch)
Unit Weight
405g
Indicator
NA
Cooling System
NA
Environment
Surrounding Air Temperature
Operating
Storage
Absolute Max
-10°C to+70°C, supported power linearly de-rate from
50°C to 50% rated up to 70°C for forced air.
Convection power de-rate from 200W @ 40°C to 100W
@ 60°C
Note: see power de-rating curve
-40°C to+85°C
Operating Humidity
5-95% RH (Non-Condensing)
Operating Altitude
5,000 meters(16402 fts)
Shock Test (Non-Operating)
50G, 11ms, 3 shocks for each direction
Vibration (Operating)
5-500Hz, 2Grms, 15 minute for each three axis
4
All parameters are specified at 25°C ambient unless otherwise noted.
www.DeltaPSU.com (November 2016, Rev. 08)
Medical AC-DC Open Frame Power Supply
24V/400W 3”x5” High Density MDS-400APB24 AA
Power De-rating curve
Output Power
400W
10 CFM forced air
200W
Convection
100W
Ambient
Temperature
-10°C
5
40°C 50°C
All parameters are specified at 25°C ambient unless otherwise noted.
www.DeltaPSU.com (November 2016, Rev. 08)
60°C 70°C
Medical AC-DC Open Frame Power Supply
24V/400W 3”x5” High Density MDS-400APB24 AA
Protections
Overvoltage (max)
135% of rated voltage, Latch Mode
Overload / Overcurrent (max)
Main output 160% of ratedcurrent
Standby 3A max
Hiccup Mode(Non-Latching, Auto-Recovery)
Over Temperature
Latch Mode
Short Circuit
Hiccup Mode,
(Non-Latching, Auto-Recovery)
Reliability
MTBF(Minimum) at 115Vac, 400W, 35°C,
10CFM Air Flow
Operating life at 115Vac, 400W, ambient 25 °C,10CFM Air Flow
500 kHrs based on Telecordia SR-332
26,280Hrs
Safety Standards / Directives
nd
rd
rd
Medical Safety
IEC60601-12 and 3 , and 3 +A1 edition CB report
TUV EN60601-1:2006
UL60601-1+CAN/CSA 60601-1: (Ed.3.2005)
ITE Safety
IEC60950-1 CB report
TUV60950-1
UL60950-1+CAN/CSA60950-1
GB4943.1-2011, GB9254-2008, GB17625.1-2003
CE
MDD Directive 93/42/EEC
Environmental
RoHS Directive 2011/65/EU Compliant
Galvanic Isolation
1)
6
Input to/Output (2XMOPP) 4000 Vac
1)
Input to/Ground(1XMOPP) 1500Vac
Output to/Ground(1XMOPP) 1500Vac ( Type BF application rated)
PSU can support PoE applications with Primary to FG 2500Vac test.
All parameters are specified at 25°C ambient unless otherwise noted.
www.DeltaPSU.com (November 2016, Rev. 08)
Medical AC-DC Open Frame Power Supply
24V/400W 3”x5” High Density MDS-400APB24 AA
EMC (Compliant with IEC 60601-1-2 4th Ed. Requirements)
EMC / Emissions
Harmonic Current Emissions
EN55011/EN55022,FCC Title 47:Class B
IEC61000-3-2
Meet Class D limit
Immunity to
Voltage Flicker
IEC61000-3-3
Electrostatic Discharge
IEC61000-4-2
Level 4 Criteria A
Air Discharge: 15kV
Contact Discharge: 8kV
Radiated Field
IEC61000-4-3
Electrical Fast Transient / Burst
IEC61000-4-4
Criteria A
80MHz-2700MHz, 10V/m AM modulation
385MHz-5785MHz, 28V/m Pulse mode and other
modulation
1)
Level 3 Criteria A :2kV
Surge
IEC61000-4-5
Level 3 Criteria A
3)
Common Mode : 2kV
4)
Differential Mode : 1kV
Conducted
IEC61000-4-6
Level 2 Criteria A
150kHz-80MHz, 3Vrms, 6Vrms at ISM bands and
1)
1)
1)
1)
Amateur radio bands
Power Frequency Magnetic Fields
Voltage Dips
IEC61000-4-8
IEC61000-4-11
1)
Criteria A
Magnetic field strength 30A/m
1)
Criteria A
0% UT, 0.5 cycle (10ms)
o
o
o
o
o
o
o
o
o
, 0 /45 /90 /135 /180 /225 /270 /315 /360
2)
Criteria B
o
0% UT, 1 cycle (20ms), 0
2)
Criteria B
o
70% UT, 25 cycle (500ms), 0
2)
Criteria B
o
0% UT, 250 cycle (5000ms), 0
1) Criteria A: Normal performance within the specification limits
2) Criteria B: Output out of regulation, or shuts down during test. Automatically restore to normal operation after test.
3) Asymmetrical: Common mode (Line to earth)
4) Symmetrical: Differential mode (Line to line)
7
All parameters are specified at 25°C ambient unless otherwise noted.
www.DeltaPSU.com (November 2016, Rev. 08)
Medical AC-DC Open Frame Power Supply
24V/400W 3”x5” High Density MDS-400APB24 AA
Dimensions
W x L x H:76.2 x 127.0 x 35.6 mm
Mechanical drawing (3Pin input type)
Suggested air flow
direction, draw air
from input side to
output side
Notes
―Dimensions are in mm(inches)
Input Connector CN1
Pin 1
Neutral
Pin 2
Line
CN1 mates with Molex housing
26033031 and Molex series 6838
crimp terminals.
Output Connector
CN102
Vo
CN103
DC RTN
Pin 1
Pin 2
Pin 3
Pin 4
Pin 5
Pin 6
Pin 7
Pin 8
Pin 9
Pin 10
Control and STANDBY connector CN101
Remote sense +
Remote sense –
Power Good +
Power Good -(DC RTN)
Remote ON_OFF/INHIBIT +
Remote ON_OFF/INHIBIT -(DC RTN)
5V Standby
DC RTN
5V Standby
DC RTN
Fan Connector CN202
CN101 mates with Molex housing 1041421000 and
Pin1
DC RTN
Molex series 104539 crimp terminals.
Pin2
12V Fan
CN202 mates with JST housing PHR-2 and JST SPH-002T-P0.5S terminals.
Two mounting points in mechanical drawing need to be connected to system earth case together, Protective
bonding conductor from the end product protective earth terminal (if any) can be tied to CN2 for open frame
model.
8
All parameters are specified at 25°C ambient unless otherwise noted.
www.DeltaPSU.com (November 2016, Rev. 08)
Medical AC-DC Open Frame Power Supply
24V/400W 3”x5” High Density MDS-400APB24 AA
Functions
Start-up Time
The time required for the output voltage to reach 90% of its set value, after the input voltage is applied.
Rise Time
The time required for the output voltage to change from 10% to 90% of its set value.
Hold-up Time
Hold up time is the time when the AC input collapses and output voltage retains regulation for a certain period of time. The time
required for the output to reach 90% of its set value, after the input voltage is removed.
■Graph illustrating the Start-up Time, Rise Time, and Hold-up Time
9
All parameters are specified at 25°C ambient unless otherwise noted.
www.DeltaPSU.com (November 2016, Rev. 08)
Medical AC-DC Open Frame Power Supply
24V/400W 3”x5” High Density MDS-400APB24 AA
Dynamic Response
Short Circuit Protection
The power supply output voltage will remain within ±5% of its
steady state value, when subjected to a dynamic load 50 to 100%
of its rated current.
■ 50 to 100% Load
The power supply’s output OLP/OCP function also provides
protection against short circuits. When a short circuit is applied,
the output current will operate in “Hiccup mode”, as shown in the
illustration in the OLP/OCP section on this page. The power
supply will return to normal operation after the short circuit is
removed.
Inrush Current
Inrush current is the peak, instantaneous, input current measured
and, occurs when the input voltage is first applied. For AC input
voltages, the maximum peak value of inrush current will occur
during the first half cycle of the applied AC voltage. This peak
value decreases exponentially during subsequent cycles of AC
voltage.
Overload & Overcurrent Protections
The power supply’s Overload (OLP) and Overcurrent (OCP)
Protections will be activated before output current under 160% of
I O (Max load). Upon such occurrence, V O will start to drop. Once
the power supply has reached its maximum power limit, the
protection will be activated and the power supply will go into
“Hiccup mode” (Auto-Recovery). The power supply will recover
once the fault condition causing the OLP and OCP is removed
and I O is back within the specified limit.
Additionally, if the Iout is <160% but >100% for a prolong period
of time (depending on the load), the Over Temperature
Protection (OTP) will be activated due to high temperature on
critical components. The power supply will then go into
latchmode.
Over Temperature Protection
Overvoltage Protection
The power supply’s overvoltage circuit will be activated when its
internal feedback circuit fails. The output voltage shall not exceed
its specifications defined on Page 4 under “Protections”. Power
supply will latch off, and require removal/re-application of input
AC voltage in order to restart.
10
As mentioned above, the power supply also has Over
Temperature Protection (OTP). This is activated when the
overload condition persists for an extended duration and the
output current is below the overload trigger point but >100% load.
In the event of a higher operating temperature condition at 100%
load, the power supply will run into OTP when the surrounding air
temperature is higher than the operating temperature. When
activated, the output voltage will go into latch mode until the input
voltage is removed; then, reapplied, and the surrounding air
temperature drops to its normal operating temperature.
All parameters are specified at 25°C ambient unless otherwise noted.
www.DeltaPSU.com (November 2016, Rev. 08)
Medical AC-DC Open Frame Power Supply
24V/400W 3”x5” High Density MDS-400APB24 AA
Power Good
Power Good+ pin is an open collector transistor (40V/600mA
rating). A resistor (suggested value 10Kohm, 1/8W) can be added
between 5V STANDBY pin (or, other available pull-up voltage
that is no greater than 30V) and the Power Good+ pin (refer to
figure below). Value of pull-up resistor may have to be adjusted,
depending on voltage used, and other end-use conditions of the
Power Good+ pin connection to the product. When AC input is
on, Power Good+ pin will be high. When AC input is off, Power
Good+ pin will be low. There will be a minimum of 3 milliseconds
between the time the power good goes to low level, and the time
when the output reaches 90% of its rated value.
11
Remote On_Off/Inhibit
Remote ON_OFF/INHIBIT can be used to enable or disable only
the main output. When the main output is disabled, the +5V
Standby output will continue to operate. This signal can be pulled
down to a low level of 0.3 volts, or shorted to DC-Return, in order
for the main output to be enabled; and, floated (no connection to
the signal), or pulled up to a value greater than or equal to 3 volts,
in order to disable the main output.
Remote Sense
Remote sense feature can be used to compensate for the extra
voltage drop on output wires that are connected from the main
output terminals, to the load. With wires connected from the
remote sense pins, at the same locations as the wires from the
main output, the remote sense function can compensate up to
500mV voltage drop. The power supply will not be damaged if the
remote sense pins are shorted, or if a reverse/inverted polarity
connection is made to the load.
All parameters are specified at 25°C ambient unless otherwise noted.
www.DeltaPSU.com (November 2016, Rev. 08)
Medical AC-DC Open Frame Power Supply
24V/400W 3”x5” High Density MDS-400APB24 AA
Certificate
All Delta Medical Power products conform to the European directive 2011/65/EU.
RoHS is the abbreviation for “Restriction of the use of certain hazardous
substances
Delta has been certified as meeting the requirement of ISO 13485: 2003 and EN
ISO 13485:2012 for the design and manufacture of switching power supply and
adaptor for medical device.
In addition to a UL Total Certification Program (TCP) approved client laboratory for
IEC60950 and IEC60065. Delta also has participated UL Client Test Data Program
(CDTP) for IEC 60601
12
All parameters are specified at 25°C ambient unless otherwise noted.
www.DeltaPSU.com (November 2016, Rev. 08)