Fujikura FGM-307PGSR Sensing element gauge Datasheet

FGM Data sheet
Sensing element/Gauge
■Features
Surface mount package
■Ordering Information
FGM-6 07P G SR
Terminal leads configuration
Surface mount package
Model
Length of the pressure port
3:3mm
6:6mm
Measurable pressure
range(kPa)
-13.79 to 13.79
-34.47 to 34.47
-48.26 to 48.26
-98.07 to 103.4
-98.07 to 206.8
-98.07 to 344.7
-98.07 to 482.6
-98.07 to 827.4
Pressure type
Rated Pressure
(PSI)
G:Gauge Pressure
Weight
(grams)
Weight
(grams)
0.3
0.35
Part number
3mm pressure port
FGM-302PGSR
FGM-305PGSR
FGM-307PGSR
FGM-315PGSR
FGM-330PGSR
FGM-350PGSR
FGM-370PGSR
FGM-3120PGSR
6mm pressure port
FGM-602PGSR
FGM-605PGSR
FGM-607PGSR
FGM-615PGSR
FGM-630PGSR
FGM-650PGSR
FGM-670PGSR
FGM-6120PGSR
RoHS compliance
■Specifications
Model
02PG
Recommended operating conditions
Pressure type
Rated pressure
Measurable pressure range
Temperature range
Pressure media
Excitation current (Constant)
13.79
-13.79
to 13.79
05PG
07PG
34.47
-34.47
to 34.47
48.26
-48.26
to 48.26
15PG
30PG
50PG
Gauge pressure
103.4
206.8
344.7
-98.07
-98.07
-98.07
to 103.4 to 206.8 to 344.7
0 to 50
Non-corrosive gases only (No liquid)
1.5
Absolute maximum rating
Maximum load pressure
Twice of rated pressure
3.0
-20 to 100
-40 to 120
30 to 80 (Non dew condition)
Electric characteristics (Drive Current 1.5mA constant ,ambient temperature Ta=25deg.C)
Output span voltage
60 to 140 (at 0kPa to rated pressure)
Offset voltage
+/-20 (at 0kPa)
Bridge resistance
4000 to 6000
Response time
2 (for the reference)
Accuracy
TSO*
+/-5
TCS*
2.5
Linearity
+/-0.5
+/-0.3
Pressure hysteresis
+/-0.4
+/-0.2
70PG
120PG
482.6
-98.07
to 482.6
827.4
-98.07
to 827.4
1
kPa
kPa
deg.C
mADC
1.5 times
of rating
pressure
mADC
deg.C
deg.C
%RH
Maximum excitation current
Operating temperature
Storage temperature
Operating humidity
*TSO : Temperature sensitivity of offset voltage(Temperature range from 0-50 deg.C)
*TCS : Temperature coefficient of output span voltage(Temperature range from 0-50 deg.C)
Unit
mV
mV
Ω
msec.
%FS/050deg.C
+/-0.5
+/-0.6
+/-0.4
%FS
%FS
FGM Data sheet
Sensing element/Gauge
P1= 0 kPa
P2= 1/2 x P3 (kPa)
P3= reted pressure (kPa)
T1= 0 deg.C
T2= 25 deg.C
T3= 50 deg.C
Offset voltage (mV)
Voff = V(P1,T)
Voff(0) = V(P1,T1)
Voff(25) = V(P1,T2)
Voff(50) = V(P1,T3)
Output voltage at full scale (mV)
Vfs = V(P3,T)
Vfs(0) = V(P3,T1)
Vfs(25) = V(P3,T2)
Vfs(50) = V(P3,T3)
Output span voltage (mV)
SV = Vfs - Voff
SV(0) = Vfs(0) - Voff(0)
SV(25) = Vfs(25) - Voff(25)
SV(50) = Vfs(50) - Voff(50)
Temperature sensitivity of offset voltage (%FS)
TSO = {LARGER ONE} / SV(25) x 100
LARGER ONE = larger absolute value which of {Voff(0)-Voff(25)} and {Voff(50)-Voff(25)}
Temperature coefficient of output span voltage (%FS)
TCS = {max[SV(0),SV(25),SV(50)] - min[SV(0),SV(25),SV(50)] } / SV(25) x 100
Linearity (%FS)
NL = {V(P2,T2) - [Voff(25)+Vfs(25)]/2 } / SV(25) x 100
Pressure hysteresis (%FS)
Phys = {Voff’(25) - Voff(25) } / SV(25) x 100
Voff’(25) : Output voltage against P1 after stressing by P3 pressure.
2
FGM Data sheet
Sensing element/Gauge
■Outline dimensions■
2-φ1.1
3
FGM Data sheet
Sensing element/Gauge
■Connection diagram■
■Output characteristics■
<302PG~307PG>
<602PG~607PG>
<315PG~3120PG>
<615PG~6120PG>
↑Output voltage(mV)
↑Output voltage(mV)
Output span
voltage
Offset voltage
-Rated pressure
Offset voltage
-98.07kPa
0kPa
0kPa
Rated pressure
→Pressure(kPa)
Rated pressure
→Pressure(kPa)
Atomospheric
pressure
Note ;
Output span
voltage
Atomospheric
pressure
Please read instruction “Notes” before using the sensor.
Fujikura reserves the right to change specifications without notice.
Please keep the sensors sealed using static shielding bags on storage.The pins of the sensor are plated by Ag.
If the sensors expose to an atmosphere, the pins will be black by sulfuration.
Please set Zero-calibration function up your products. The offset voltage may be shifted some mechanical stress
such as mounting, installation and etc. over longtime using.
4
FGM Data sheet
Sensing element/Gauge
Reflow Soldering process recommendation profile
Package
Surface
Temperature
(Degree C)
230
210
150
Room
Temp.
Time
a
a: Rump up rate
b: Pre-heating
c: Rump up rate
d: Heating
e: Rump down rate
Note ;
b
c
d
e
1 or 2 deg.C/sec.
150 to 180 deg.C,within 60 to120sec.
1 to 2 deg.C/sec.
max.230 deg.C,max.10sec. 210 deg.C,within 30sec.
1 or 2 deg.C/sec.
1 ) Temperature means Surface temperature of the sensor package.
2 ) Reflow process max. 2 times.
3 ) Do not wash the sensor.
4 ) Do not put the solder and flux on the sensor package.
If you have any questions regarding technical issues or specifications, please contact us.
Fujikura Ltd. Sensor Department
5-1 Kiba 1-chome, Koto-ku, Tokyo 135-8512, Japan
Phone +81-(0)3-5606-1072
E-mail : [email protected]
5
Similar pages