MPD MB1048DMRW-15RI4 En 60601 approved compact, 2:1 input, 10w dc/dc con vert er Datasheet

MB1000MRWRI4
EN 60601 Approved
Compact, 2:1 Input, 10W
DC/DC Converters
Key Features:
Electrical Specifications
Specifications typical @ +25°C, nominal input voltage & rated output current, unless otherwise noted. Specifications subject to change without notice.
• EN 60601 3RD Ed. Approved
Input
• 10W Output Power
Parameter
• 4.2 kVrms Isolation
• Reinforced Insulation
Input Start Voltage
• 1 x MOPP & 2 x MOOP per
EN 60601-1 3RD Edition & Under Voltage Shutdown
ANSI/AAMI ES 60601-1
• 10 µA Max Leakage Current
• Wide 2:1 Input Range
• Compact 1 x 2 In Case
• Single & Dual Outputs
• 1.0 MH MTBF
Input Filter
Short Circuit Input Power
Conditions
Min.
12 VDC Input
7.0
24 VDC Input 13.0
48 VDC Input 30.0
12 VDC Input
24 VDC Input
48 VDC Input
␲ (Pi) Filter
Typ.
8.0
15.0
33.0
Max.
9.0
18.0
36.0
8.5
16.0
34.0
Units
3,000
mW
Max.
±1.0
±2.0
±0.5
±1.0
±1.2
100
150
Units
%
%
%
VDC
VDC
Output
Parameter
Output Voltage Accuracy
Output Voltage Balance
Line Regulation
Conditions
Load Regulation
Ripple & Noise (20 MHz), See Note 2
Output Power Protection
Transient Recovery Time, See Note 3
Transient Response Deviation
Temperature Coefficient
Output Short Circuit
Min.
Dual Output , Balanced Loads
VIN = Min to Max
IOUT = 15% to 100%
IOUT = 5% to 100%
5V Output
All Other Outputs
Typ.
±0.5
±0.3
±0.5
±0.6
120
150
300
±3.0
±0.02
Continuous (Autorecovery)
25% Load Step Change
600
±5.0
±0.05
%
mV P - P
%
µSec
%
%/ºC
General
Parameter
Isolation Voltage, Rated
Reinforced Insulation Working Voltage
Leakage Current
Isolation Resistance
Isolation Capacitance
Switching Frequency
Conditions
Min.
60 Seconds 4,200
300 Vrms
240 VAC, 60 Hz
500 VDC
10
100 kHz, 1V
120
Typ.
Max.
Units
Vrms
10
60
150
80
180
µA
G⍀
pF
kHz
Typ.
Max.
+75
+95
+125
Units
ºC
ºC
ºC
95
4,000
%
m
Environmental
Parameter
Conditions
Operating Temperature Range
Storage Temperature Range
Cooling
Humidity
Altitude
Ambient
Case
Min.
-40
-50
Free Air Convection
RH, Non-condensing
Physical
MicroPower Direct
292 Page Street
Suite D
Stoughton, MA 02072
USA
T: (781) 344-8226
F: (781) 344-8481
E: [email protected]
W: www.micropowerdirect.com
Case Size
Case Material
Weight
See Mechanical Diagram (Page 3)
Non-Conductive Black Plastic (UL94-V0)
0.86 Oz (24.5g)
Reliability Specifications
Parameter
MTBF
Conditions
Min.
Typ.
Max.
Units
MIL HDBK 217F, 25ºC, Gnd Benign
1.0
MHours
IEC/EN 60601-1, EN 60601-1 3RD Edition, 1xMOPP & 2xMOOP
ANSI/AAMI ES 60601-1 1xMOPP & 2xMOOP Recognition, (UL Certificate)
ANSI/AAMI ES 60601-1, CAN/CSA-C22.2 No.60601-1
Safety Standards
Absolute Maximum Ratings
Parameter
Input Voltage Surge (1 Sec)
Lead Temperature
Conditions
12 VDC Input
24 VDC Input
48 VDC Input
1.5 mm From Case For 10 Sec
Min.
Typ.
Max.
25.0
50.0
100.0
260
Caution: Exceeding Absolute Maximum Ratings may damage the module. These are not continuous operating ratings.
www.micropowerdirect.com
Units
VDC
ºC
www.micropowerdirect.com
Model Selection Guide
Input
Model
Number
Voltage (VDC)
Current (mA)
Reflected
Ripple
Current Voltage
Full-Load No-Load (mA, Typ)
(VDC)
Output
Current
Current
(mA, Max)
(mA, Min)
(µF Max)
(mA)
74
1,000
3,000
80
470
3,000
0.0
80
220
3,000
±333
0.0
81
220
3,000
2,000
0.0
75
1,000
1,500
835
0.0
81
470
1,500
±417
0.0
81
220
1,500
82
220
1,500
75
1,000
750
81
470
750
0.0
81
220
750
0.0
82
220
750
Nominal
Range
MB1012SMRW-05RI4
12
9.0 - 18.0
907
30
100
5.0
1,600
0.0
MB1012SMRW-12RI4
12
9.0 - 18.0
1,044
30
100
12.0
835
0.0
MB1012DMRW-12RI4
12
9.0 - 18.0
1,042
30
100
±12.0
±417
MB1012DMRW-15RI4
12
9.0 - 18.0
1,028
30
100
±15.0
MB1024SMRW-05RI4
24
18.0 - 36.0
559
20
50
5.0
MB1024SMRW-12RI4
24
18.0 - 36.0
516
20
50
12.0
MB1024DMRW-12RI4
24
18.0 - 36.0
516
20
50
±12.0
MB1024DMRW-15RI4
24
18.0 - 36.0
508
20
50
±15.0
±333
0.0
MB1048SMRW-05RI4
48
36.0 - 75.0
280
10
25
5.0
2,000
0.0
MB1048SMRW-12RI4
48
36.0 - 75.0
258
10
25
12.0
835
0.0
MB1048DMRW-12RI4
48
36.0 - 75.0
258
10
25
±12.0
±417
MB1048DMRW-15RI4
48
36.0 - 75.0
254
10
25
±15.0
±333
Notes:
1. The specified maximum capacitive load is for each output.
2. When measuring output ripple, it is recommended that an external 0.47 µF ceramic capacitor
be placed from the +VOUT pin to the -VOUT pin for single output units and from each output to
common for dual output units. For noise sensitive applications, the use of 3.3 µF capacitors
will reduce the output ripple.
3. Transient recovery is measured to within a 1% error band for a load step change of 75% to
100%.
3. Dual output units may be connected to provide a 24 VDC or 30 VDC output. To do this, connect
the load across the positive (+VOUT) and negative (-VOUT) outputs and float the output common.
4. The converter should be connected to a low ac-impedance source. An input source with a
highly inductive impedance may affect the stability of the converter. In applications where the
converter output loading is high and input power is supplied over long lines, it may be necessary
to use a capacitor on the input to insure start-up.
In this case, it is recommended that a low ESR (ESR <1.0⍀ at 100 kHz) capacitor be mounted
close to the converter. For 12V input units a 10.0 µF is recommended, for 24V a 4.7 µF and
for 48V units a 2.2 µF.
5. It is recommended that a fuse be used on the input of a power supply for protection. See the
table above for the correct rating.
Fuse Rating
Efficiency Capacitive
Load
Slow-Blow
(%, Typ)
Derating Curve
Typical Connection
EMI Characteristics
Parameter
Conducted Emissions
Radiated Emissions
These converters are specified for operation without external components.
Standard
EN 55011, EN 55022
EN 55024
Criteria
ESD
EN 61000-4-2
A
RS
EFT, See Note at right
Surge, See Note at right
CS
PFMF
EN 61000-4-3
EN 61000-4-4
EN 61000-4-5
EN 61000-4-6
EN 61000-4-8
A
A
A
A
A
Page 2
Level
Class A
Class A
±15 kV Air
±8 kV Contact
10V/m
±2 kV
±1 kV
10 Vrms
30A/m
However, in some applications the addition of input/output capacitors, as shown
in the typical connection diagram above, will enhance stability and reduce output
ripple. This simple connection includes a low ESR (<1ม at 100 kHz) capacitor
connected across the input (C1). It is recommended that a 10 µF be used
for 12V input models, a 4.7 µF for 24V and a 2.2 µF for 48V input units.
To improve the output ripple performance, a 3.3 µF is connected across
the output. For dual output units, a 3.3 µF capacitor should be connected from
each output to common.
To meet the specified EN 61000-4-4 and EN 61000-4-5 limits, an external
capacitor must be connected across the input pins of the module (C1). A
330 µF/100V capacitor is recommended. This capacitor should be mounted as
close to the module as possible.
We Power Your Success - For Less
Mechanical Dimensions
www.micropowerdirect.com
Pin Connections
Pin Single
1
+VIN
2
-VIN
3
+VOUT
Dual
+VIN
-VIN
+VOUT
Pin
4
5
Single
No Pin
-VOUT
Dual
Comm.
-VOUT
Mechanical Notes:
• All dimensions are typical in inches (mm)
• Tolerance x.xx = ±0.01 (±0.25)
Related Products
Medical Approved DC/DC’s
MA300MRURI4 Series
MB2000MRWRI4 Series
ML200MRRI4 Series
• 3W Output Power
• 20W Output Power
• 2W Output Power
• 4 kVAC Isolation
• 4.2 kVAC Isolation
• 4 kVAC Isolation
• Reinforced Insulation
• Reinforced Insulation
• Reinforced Insulation
• Meets 1xMOPP & 2xMOOP
• 2 µA Leakage Current Max
• Wide 4:1 Input Range
• Compact DIP Case
• Single & Dual Outputs
• 1.0 MH MTBF
• EN 60601 Approved
• Meets 1xMOPP & 2xMOOP
• 5 µA Leakage Current Max
• Wide 2:1 Input Range
• Compact 1 x 2 In Case
• Single & Dual Outputs
• 1.08 MH MTBF
• EN 60601 Approved
• Meets 1xMOPP & 2xMOOP
• 2 µA Leakage Current Max
• Compact SMT Case
• Available on Tape/Reel
• 2.0 MH MTBF
• EN 60601 Approved
Thousands of standard power products ranging from 0.5W to 500W are
available from MPD in a wide variety of packages and pin-outs. This
includes many more families with EN 60601 medical approval
MicroPower Direct
We Power Your Success - For Less!
292 Page Street Ste D Stoughton, MA 02072
•
TEL: (781) 344-8226
•
FAX: (781) 344-8481
•
E-Mail: [email protected]
Similar pages