ALM120 Series AC-DC Power Supplies 120 Watts • Medical & IT Safety Approvals • Energy Efficiency Level VI & EU CoC Tier 2 Compliant • 4th Edition Medical EMC • IP32 Environmental Rating • Class I and Class II Versions • <0.15 W Standby Power • 0 °C to 60 °C Operation • Low Earth Leakage Current • 3 Year Warranty Dimensions: The ALM120 series of medical external power supplies is fully approved to international medical & IT safety standards. It has been designed with very high efficiency and low standby power, enabling it to meet the latest environmental legislation. The unit has a fully sealed enclosure complying with IP32 and a smooth surface finish making it easier to wipe down in a clinical setting. ALM120: 6.732 x 2.677 x 1.496” (171.0 x 68.0 x 38.0 mm) Models & Ratings Output Power Output Voltage Output Current Total Regulation Model Number(1,2,3,4) 10.00 A 8.00 A 6.32 A 5.00 A ±5% ALM120PS12 ALM120PS15 ALM120PS19 ALM120PS24 12.0 V 15.0 V 19.0 V 24.0 V 120 W Notes 1. For class II versions, add suffix 'C2-8' to the end of the part number e.g. ALM120PS24C2-8. 2. For optional input connector retention clip add suffix ‘-A’ to the model number, e.g. ALM120PS24-A (not available for C2 versions) 3. For optional output connector, DC barrel jack, add suffix -B5 to the model number, e.g. ALM120PS24-B5 (not available for 12 & 15 V models) 4. Power de-rated <100 VAC for 12 & 15 V models, refer to input specifications. Mechanical Details 0.195 (4.9) 6.73 (171.0) 1.41 ±0.079 (36.0 ±2.0) 2.67 (68.0) 35.42 ±0.79 (900.0 ±20.0) KPPX-4P K (Non-Locking) 0.787 ±0.079 (ø20.0 ±2.0) Shell 1.96 ±0.787 (50.0 ±20.0) AC Input IEC 320-C8 Pin 1 2 3 4 Connection +Vout +Vout RTN RTN Standard Models: Shell Functional Earth C2 Models: Floating 1.49 (38.0) 0.039 (1.0) 3.95 (100.3) AC Input IEC 320-C14 Optional -B5 Versions 1.73 (44.0) 1.59 (40.3) Return (-) 5.51 (140.0) Output (+) Notes 1. All dimensions shown in inches (mm). Tolerance is 0.02 (0.5) maximum, except output cable length. 2. Weight: 1.1 lbs (0.5 kg) approx. 3. For European mains lead order part EU-MAINS-IEC for C14 versions, or EU-MAINS-8 for C8 versions. 4. For UK mains lead order part UK-MAINS-IEC for C14 versions, or UK-MAINS-8 for C8 versions. 1 5. For US mains lead order part US-MAINS-IEC for C14 versions, or US-MAINS-8 for C8 versions. 6. Output connector: 4 pin power din with pin 1 & 2 - positive and pin 3 & 4 - return, equivalent to KPPX-4P (non-locking). Optional for 19 & 24 models: DC barrel jack; 5.5 outer diameter, 2.5 mm inner diameter with centre positive. www.xppower.com ALM120 Series AC-DC Power Supplies Input Characteristic Minimum Input Voltage 80 Input Frequency Input Current Inrush Current Power Factor Earth Leakage Current No Load Input Power Input Protection 47 Typical Units Notes & Conditions 264 VAC 19 & 24 V models: Derate linearly from 100% load at 90 VAC to 80% load at 80 VAC, 300 VAC/5 s maximum. 12 & 15 V models: Derate linearly from 100% load at 100 VAC to 80% load at 80 VAC, 300 VAC/5 s maximum. 63 Hz A A Maximum 1.2/0.6 60/120 >0.9 160 250 0.15 T3.15A/250 VAC internal fuse in both line & neutral Measured at 115/230 VAC 115/230 VAC, cold start at 25 °C EN61000-3-2 Class A 264 VAC, 60 Hz µA W Output Characteristic Output Voltage Initial Set Accuracy Minimum Load Start Up Delay Start Up Rise Time Hold Up Time Line Regulation Total Regulation Min. 12 20 Typ. Max. 24 ±2 Units VDC % 1 2 50 ±0.5 ±5 s ms ms % % 4 % 1.5 % pk-pk 10 175 150 % % % 95 %/°C µA 264 VAC, 60 Hz Maximum +60 Units °C Notes & Conditions Derate from 100% load at 40 °C to 30% load at 60 °C 30 Transient Response Ripple and Noise Overshoot Overload Protection Overvoltage Protection Short Circuit Protection Thermal Protection Temperature Coefficient Patient Leakage Current 5 115 Trip and restart (hiccup), auto resetting Measured internally, auto resetting 0.02 Notes & Conditions See Models and Ratings table At 50% load No minimum load required Full load and 115/230 VAC Maximum deviation, recovering to less than 1% within 500 µs for 50-75-50% load change 20 MHz bandwidth, measured with 20 MHz Bandwidth and 10 µF electrolytic in parallel with 0.1 µF ceramic capacitor. At turn on / turn off Recycle mains to reset Environmental Characteristic Operating Temperature Cooling Operating Humidity Storage Temperature Operating Altitude Shock Vibration Minimum 0 Natural convection 5 -20 Typical 95 %RH +80 °C 5000 m IEC68-2-27, 30 g, 11 ms half sine, 3 times in each of 6 axes IEC68-2-6, 10-500 Hz, 2 g 10 mins/sweep, 60 mins for each of 3 axes Derating Curve Non-condensing Output Power (%) 100% 80% 60% 40% 20% 0 20 Ambient Temperature (ºC) www.xppower.com 40 60 2 ALM120 Series AC-DC Power Supplies General Characteristic Minimum Typical Efficiency Maximum Units % 91 Isolation: Input to Output Input to Ground Output to Ground 45 85 Switching Frequency Power Density Weight See Models and Ratings table and curves. 4000 VAC 2 x MOPP 1500 VAC 1 x MOPP (Class I versions only) 500 140 190 VAC Class I versions only PFC Main Converter kHz W/in3 4.45 Mean Time Between Failure Notes & Conditions >300 kHrs 1.1 (500) lb (g) MIL-HDBK-217F at 25 °C GB Efficiency Curves 100% ALM120PS12 Efficiency 90% 115V 80% 230V 70% 60% 10% 20% 30% 40% 50% 60% 70% 80% 90% 100% Load ALM120PS24 100% Efficiency 90% 115V 80% 230V 70% 60% 10% 20% 30% 6 40% 50% 60% 70% 80% 90% 100% Load EMC: Emissions Phenomenon Emissions Harmonic Current Voltage Flicker Standard EN55032 EN61000-3-2 EN61000-3-3 Test Level Level B Class A www.xppower.com Notes & Conditions Conducted & Radiated 3 ALM120 Series AC-DC Power Supplies EMC: Immunity Phenomenon ESD Radiated EFT/Burst Surge Conducted Magnetic Fields Test Level 3 10 V/m 3 Installation Class 3 10 V 4 Dip: 30% 500 ms Dip: 60% 200 ms Dip: 80% 5000 ms Int: 100% 10 ms Int: 100% 20 ms Int: 100% 5000 ms Dip: 30% 25 AC Cycles Dip: 60% 5 AC Cycles Int: 100% 0.5 AC Cycles Int: 100% 1.0 AC Cycles Int: 100% 250 AC Cycles Standard EN61000-4-2 EN61000-4-3 EN61000-4-4 EN61000-4-5 EN61000-4-6 EN61000-4-8 EN61000-4-11 Dips and Interruptions EN60601-1-2 Criteria A A A A A A A A/B B A A B A A A A B Notes & Conditions ±15 kV air/±8 kV contact 80-2700 MHz. IEC60601-1-2 Ed.4 at other frequencies High Line/Low Line 230 VAC 100% load, 100 VAC 25% load Safety Approvals Safety Agency UL TUV CB CE UL CSA TUV CB Others Safety Standard UL62368-1 EN62368-1 IEC60950-1 & IEC62368-1 LVD ANSI/AAMI ES 60601-1 CSA C22.2 No. 60601 EN60601-1 IEC60601-1 CCC, PSE, KC & RCM Notes & Conditions Information Technology Medical May require addtional importer information Mechanical Details 0.195 (4.9) 6.73 (171.0) 1.41 ±0.079 (36.0 ±2.0) 2.67 (68.0) KPPX-4P (Non-Locking) K 35.42 ±0.79 (900.0 ±20.0) Shell 0.787 ±0.079 (ø20.0 ±2.0) 1.96 ±0.787 (50.0 ±20.0) Pin 1 2 3 4 Connection +Vout +Vout RTN RTN Standard Models: Shell Functional Earth C2 Models: Floating AC Input IEC 320-C8 1.49 (38.0) 0.039 (1.0) 3.95 (100.3) Optional -B5 Versions AC Input IEC 320-C14 Return (-) 1.73 (44.0) 1.59 (40.3) 5.51 (140.0) Output (+) Notes 1. All dimensions shown in inches (mm). Tolerance is 0.02 (0.5) maximum, except output cable length. 2. Weight: 1.1 lbs (0.5 kg) approx. 3. For European mains lead order part EU-MAINS-IEC for C14 versions, or EU-MAINS-8 for C8 versions. 4. For UK mains lead order part UK-MAINS-IEC for C14 versions, or UK-MAINS-8 for C8 versions. 5. For US mains lead order part US-MAINS-IEC for C14 versions, or US-MAINS-8 for C8 versions. 6. Output connector: 4 pin power din with pin 1 & 2 - positive and pin 3 & 4 - return, equivalent to KPPX-4P (non-locking). Optional for 19 & 24 models: DC barrel jack; 5.5 outer diameter, 2.5 mm inner diameter with centre positive. www.xppower.com 22 Jun 17