IMM01 Series DC-DC Converter 1 Watt xxx Series • World Wide Medical Approvals • Single and Dual Outputs • 2 µA Patient Leakage Current • SIP7 Package • -20 °C to +100 °C Operation • Full Load at 60 °C Ambient • 1500 VAC Isolation, 1 MOPP • MTBF 1 MHrs • 3 Year Warranty Dimensions: IMM01: 0.76 x 0.36 x 0.44” (19.5 x 9.2 x 11.1 mm) Models & Ratings Input Voltage Output Voltage 4.5-9.0 V 9.0-18.0 V Output Current Input current No Load Full Load Maximum capacitive load Model Number 3V3 303 mA 75 mA 337 mA 2200 µF IMM0105S3V3 5V 200 mA 75 mA 337 mA 2200 µF IMM0105S05 12 V 83 mA 75 mA 327 mA 470 µF IMM0105S12 15 V 67 mA 75 mA 327 mA 470 µF IMM0105S15 ±3.3 V ±150 mA 55 mA 337 mA ±1000 µF IMM0105D03 ±5 V ±100 mA 55 mA 337 mA ±1000 µF IMM0105D05 ±12 V ±42 mA 75 mA 327 mA ±220 µF IMM0105D12 ±15 V ±33 mA 75 mA 327 mA ±220 µF IMM0105D15 3V3 303 mA 25 mA 163 mA 2200 µF IMM0112S3V3 5V 200 mA 25 mA 163 mA 2200 µF IMM0112S05 12 V 83 mA 25 mA 150 mA 470 µF IMM0112S12 15 V 67 mA 25 mA 150 mA 470 µF IMM0112S15 ±3.3 V ±150 mA 55 mA 163 mA ±1000 µF IMM0112D03 ±5 V ±100 mA 55 mA 163 mA ±1000 µF IMM0112D05 ±12 V ±42 mA 30 mA 150 mA ±220 µF IMM0112D12 ±15 V ±33 mA 30 mA 150 mA ±220 µF IMM0112D15 Notes Input currents measured at low input voltage. Mechanical Details Pin Connections 0.76 (19.5) 0.02 (0.50) 0.36 (9.20) Printed Face 0.42 (10.6) 0.12 (3.0) 1 2 3 5 6 0.44 (11.1) Bottom View 0.36 (9.20) 0.13 (3.20) 0.02 (0.50) 7 0.02 (0.50) Single -Vin +Vin Dual -Vin +Vin 3 Remote On/Off Remote On/Off 5 6 7 +Vout -Vout No Pin +Vout Common -Vout Notes 1. All dimensions are in inches (mm) 2. Weight: 0.008 lbs (3.6 g) approx. 3. Pin diameter: 0.02±0.002 (0.5±0.05) 4. Pin pitch tolerance: ±0.014 (±0.35) 5. Case tolerance: ±0.02 (±0.5) 0.10 (2.54) 0.6 (15.24) 1 0.01 (0.25) Pin 1 2 www.xppower.com IMM01 Series DC-DC Converter Input Characteristic Input Voltage Range Minimum Typical 4.5 9 Inrush Current Input Reflected Ripple Current Maximum Units Notes & Conditions 9 VDC 5 V nominal 12 V nominal 18 VDC 0.05 A2s 16 VDC for 100 ms 5 V models 25 VDC for 100 ms 12 V models Maximum Units Notes & Conditions 30 VDC See Models and Ratings table 2 µA 30 mA pk-pk Input Surge Through 12 µH inductor and 47 µF capacitor Output Characteristic Minimum Output Voltage 3.3 Typical Patient Leakage Current ±2 Initial Set Accuracy Minimum Load ±1 0 3.3 V & 5.0 V at full load % 12.0 V & 15.0 V at full load % No minimum load required Line Regulation ±0.5 % From minimum to maximum input Load Regulation 2.0 % From 10% to full load (2% from 0% to 10% load) Cross Regulation ±5 % On dual output models, when one output is at 25% load and other is varied from 10% load to full load Ripple & Noise 1/2 % pk-pk For 12 V & 15 V/ 3.3 V & 5 V models. 20 MHz bandwidth. Measured using 0.1 µF ceramic capacitor Short Circuit Protection Continuous fold-back mode, with auto recovery Maximum Capacitive Load See Models and Ratings table Temperature Coefficient Overload Protection Remote On/Off 190 250 0.03 %/ ˚C 310 % Of nominal output current at nominal input voltage Output is on if remote on/off (pin 3) is open Output turns off if remote on/off (pin 3) is high (4.0-5.5 VDC max) or if 2-4 mA is applied. General Characteristic Efficiency Minimum Typical Maximum Units 64 78 81 % Isolation: Input to Output 1500 Switching Frequency 175 Isolation Resistance 109 Isolation Capacitance Power Density Mean Time Between Failure Typical value is for IMM0112S12 VAC At 250 VAC working voltage, 1 MOPP 1000 kHz May enter burst mode frequency of 12-28 kHz at light load 27 pF W/in3 8.31 1 Weight Notes & Conditions Ω MHrs 0.008 (3.6) MIL-HDBK-217F, +25 °C GB lb (g) Environmental Characteristic Maximum Units Operating Altitude Minimum 3048 M Transport Altitude 10,000 Operating Temperature -20 Storage Temperature -40 Case Temperature Humidity Typical +100 °C +125 °C +105 °C 95 %RH Cooling Notes & Conditions Derate linearly from 100% load at +60 °C to 0% at +100 °C Non-condensing Natural convection www.xppower.com 2 IMM01 Series DC-DC Converter EMC: Emissions Phenomenon Standard Test Level Notes & Conditions Conducted EN55011 Class A See Application Note Radiated EN55011 Class A EMC: Immunity Standard Test Level Criteria Medical Device EMC Phenomenon IEC60601-1-2 Ed 4.0:2014 As below ESD Immunity EN61000-4-2 ±15 kV A Air Discharge A 80% mod, 80 MHz - 2.7 GHz plus discrete communcation proximity field frequencies Radiated Immunity EN61000-4-3 10 Vrms Notes & Conditions EFT/Burst EN61000-4-4 2 kV A External input filter required, see applications note Surge EN61000-4-5 2 kV A External input filter required, see applications note Conducted Immunity EN61000-4-6 10 V rms A Magnetic Fields EN61000-4-8 30 A/m A Safety Approvals Safety Agency Safety Standard Notes & Conditions ANSI/AMMI ES60601-1 UL CSA CSA C22.2 No. 60601-1 TUV EN60601-1 CB IEC60601-1 Mechanical Details 0.76 (19.5) 0.02 (0.50) 0.36 (9.20) Printed Face 0.42 (10.6) 0.44 (11.1) 0.12 (3.0) 1 2 3 5 6 Bottom View 0.36 (9.20) 0.02 (0.50) 7 0.02 (0.50) Pin Connections 0.10 (2.54) Pin 1 2 0.6 (15.24) Notes 3 4. Pin pitch tolerance: ±0.014 (±0.35) 5. Case tolerance: ±0.02 (±0.5) 1. All dimensions are in inches (mm) 2. Weight: 0.008 lbs (3.6 g) approx. 3. Pin diameter: 0.02±0.002 (0.5±0.05) External Filter for Surge and EFT -Vin On/Off Remote On/Off +Vout -Vout No Pin +Vout Common -Vout + Vin L1 IMM01 LRemote + Vout + Vout D1 Dual -Vin +Vin EMI Filter Conducted Emissions + Vin C1 Single -Vin +Vin 5 6 7 Application Note Load C1 C2 IMM01 -Vin -Vout C1 is 220 µF, 100 V electrolytic capacitor D1 is 18 V, 3 kW TVS for 5 V input or 28 V, 3 kW TVS for 12 V input 3 0.01 (0.25) 0.13 (3.20) Load -Vout C1 & C2 are 10 µF, 35 V multilayer ceramic chip capacitors, placed as close as possible to the input pins L1 is 12 µH inductor www.xppower.com 12 April 16