Mean Well GSM06U 6w ac-dc reliable green medical adaptor Datasheet

6W AC-DC Reliable Green Medical Adaptor
GSM06U
series
+ +
ANSI/AAMI ES60601-1
■ Features
•2 pole USA AC plug, Class Ⅱ power unit
•Medical safety approved (2 x MOPP) accroding to
ANSI/AAMI ES60601-1
•Extremely low leakage current
•No load power consumption< 0.3W
•Energy efficiency Level Ⅴ
•Protections: Short circuit / Overload / Over voltage
•Fully enclosed plastic case
•3 years warranty
■ Applications
•Blood glucose meter
•Blood pressure meter
•Nebulizer
•Inhaler
•Portable medical device
■ Description
GSM06U is a highly reliable, 6W wall-mounted style single-output green medical adaptor series. This
product is equipped with a 2-pin (no FG) standard USA AC power plug, adopting the input range from
85VAC to 264VAC. The entire series supplies different output voltages between 5VDC and 24VDC that
can satisfy the demands for various kinds of miniature medical devices. The circuitry design meets the
international medical standards (2 x MOPP), having an ultra low leakage current (<50μA), fitting the medical
devices in direct electrical contact with the patients.
With the efficiency up to 82% and the extreme low no-load power consumption below 0.3W. GSM06U is
compliant with USA EISA 2007. The supreme feature allows the adaptor to save the energy when it is either
under the operating mode or the standby mode. The entire series utilizes the 94V-0 flame retardant plastic
case, providing the double insulation that effectively prevents electrical shock. GSM06U is approved with
the international medical safety certificates.
■ Model Encoding
GSM 06 U 05 - P1J
DC plug type
{
P1J : Plug for standard model, 2.1ψ × 5.5ψ ×11mm, c+, tuning fork type
Other options available by customer requested (see Page 4~5)
Output voltage
USA 2 pin AC plug
Rated wattage
Series name
File Name:GSM06U-SPEC 2017-10-11
GSM06U
6W AC-DC Reliable Green Medical Adaptor
series
SPECIFICATION
GSM06U05-P1J GSM06U06-P1J GSM06U07-P1J GSM06U09-P1J GSM06U12-P1J GSM06U15-P1J GSM06U18-P1J GSM06U24-P1J
ORDER NO.
SAFETY MODEL NO.
DC VOLTAGE
OUTPUT
GSM06U05
Note.2 5V
GSM06U06
GSM06U07
GSM06U09
GSM06U12
GSM06U15
GSM06U18
GSM06U24
6V
7.5V
9V
12V
15V
18V
24V
0.25A
RATED CURRENT
1.2A
1.0A
0.8A
0.66A
0.5A
0.4A
0.33A
CURRENT RANGE
0 ~ 1.2A
0 ~ 1.0A
0 ~ 0.8A
0 ~ 0.66A
0 ~ 0.5A
0 ~ 0.4A
0 ~ 0.33A
0 ~ 0.25A
RATED POWER
6W
6W
6W
6W
6W
6W
6W
6W
RIPPLE & NOISE (max.) Note.3 50mVp-p
50mVp-p
80mVp-p
80mVp-p
100mVp-p
120mVp-p
150mVp-p
180mVp-p
VOLTAGE TOLERANCE Note.4 ±5.0%
±5.0%
±5.0%
±5.0%
±5.0%
±5.0%
±5.0%
±4.0%
LINE REGULATION
Note.5 ±0.5%
±0.5%
±0.5%
±0.5%
±0.5%
±0.5%
±0.5%
±0.5%
LOAD REGULATION
Note.6 ±5.0%
±5.0%
±5.0%
±5.0%
±3.0%
±3.0%
±3.0%
±2.0%
74%
76%
77%
79%
80%
82%
SETUP, RISE, HOLD UP TIME 1000ms, 50ms, 12ms at full load
85 ~ 264VAC 120 ~ 370VDC
VOLTAGE RANGE
INPUT
FREQUENCY RANGE
47 ~ 63Hz
EFFICIENCY (Typ.)
68%
AC CURRENT
INRUSH CURRENT (max.)
0.18A / 100VAC
LEAKAGE CURRENT(max.)
OVERLOAD
PROTECTION
OVER VOLTAGE
WORKING TEMP.
WORKING HUMIDITY
ENVIRONMENT STORAGE TEMP., HUMIDITY
TEMP. COEFFICIENT
VIBRATION
SAFETY STANDARDS
WITHSTAND VOLTAGE
ISOLATION RESISTANCE
SAFETY &
EMC
(Note. 7)
74%
Cold start 15A / 115VAC
30A / 230VAC
Touch current < 50μA/264VAC
>105% rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is removed
>120% rated output voltage
Protection type : Clamp by zener diode
0 ~ +50℃ (Refer to "Derating Curve")
20% ~ 90% RH non-condensing
-20 ~ +85℃, 10 ~ 95% RH non-condensing
±0.04% / ℃ (0 ~ 40℃)
10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes
ANSI/AAMI ES60601-1(3.1 version),CAN/CSA-C22 3rd edition approved
I/P-O/P:5656VDC
I/P-O/P:100M Ohms / 500VDC / 25℃/ 70% RH
Parameter
Standard
Test Level / Note
Conducted emission
FCC PART 15 & 18 / CISPR22, CAN ICES-3(B)/NMB-3(B)
Class B
Radiated emission
FCC PART 15 & 18 / CISPR22, CAN ICES-3(B)/NMB-3(B)
Class B
EMC EMISSION
OTHERS
CONNECTOR
MTBF
DIMENSION
500Khrs min. MIL-HDBK-217F(25℃)
32*66*42.5mm (L*W*H)
100g; 110pcs/12Kg/CARTON
PACKING
PLUG
See page 4~5 ; Other type available by customer requested
CABLE
See page 4~5 ; Other type available by customer requested
1.All parameters are specified at 115VAC input, rated load, 25℃ 70% RH ambient.
2.DC voltage: The output voltage set at point measure by plug terminal & 50% load.
3.Ripple & noise are measured at 20MHz by using a 12" twisted pair terminated with a 0.1μf & 47μf capacitor.
4.Tolerance: includes set up tolerance, line regulation, load regulation.
NOTE
5.Line regulation is measured from low line to high line at rated load.
6.Load regulation is measured from 10% to 100% rated load.
7.The power supply is considered as an independent unit, but the final equipment still need to re-confirm that the whole system complies with the
EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.”
(as available on http://www.meanwell.com)
File Name:GSM06U-SPEC 2017-10-11
GSM06U
6W AC-DC Reliable Green Medical Adaptor
Derating Curve
series
Static Characteristics
100
100
LOAD (%)
LOAD (%)
90
70
80
70
60
50
40
0
10
20
30
40
AMBIENT TEMPERATURE (℃)
50
85
100
110
135
155
230 264
INPUT VOLTAGE (V) 60Hz
Mechanical Specification
12.7
Name Plate
18.2
6.3
1.5
24.2
2 Pole USA Plug
UL2468 22AWG 1220 +150/-0mm for 5~9V
UL2468 24AWG 1830 +150/-0mm for 12~24V
17.1±0.5
48±1
42.5±1
66±1
32±1
File Name:GSM06U-SPEC 2017-10-11
GSM06U
6W AC-DC Reliable Green Medical Adaptor
series
DC output plug
◎ Standard plug: P1J
Unit:mm
P1J
Pin Assignment
C"+"
11 ± 0.5mm
5.5
2.1
Outside
Inside
◎ Optional DC plug:
Tuning Fork Style
Type No.
C
(Straight)
A
C
B
(Right-angled)
Barrel Style
Type No.
C
A
(Straight)
C
B
(Right-angled)
P2I
P2J
P2L
P2M
P2IR
P2JR
P2LR
P2MR
Type No.
Lock Style
Locking
A
P1I
P1L
P1M
P1IR
P1JR
P1LR
P1MR
C
B
SWITCHCRAFT original or equivalent
P2S(S761K)
P2K(761K)
P2C(S760K)
P2D(760K)
Type No.
Min. Pin Style
C
A
B
EIAJ equivalent
P3A
P3B
P3C
A
OD
5.5
5.5
5.5
5.5
5.5
5.5
5.5
A
OD
5.5
5.5
5.5
5.5
5.5
5.5
5.5
5.5
A
OD
5.53
5.53
5.53
5.53
A
OD
2.35
B
ID
2.1
2.5
2.5
2.1
2.1
2.5
2.5
B
ID
2.1
2.1
2.5
2.5
2.1
2.1
2.5
2.5
B
ID
2.03
2.54
2.03
2.54
B
ID
0.7
C
L
9.5
9.5
11.0
9.5
11.0
9.5
11.0
C
L
9.5
11.0
9.5
11.0
9.5
11.0
9.5
11.0
C
L
12.06
12.06
9.52
9.52
C
L
11.0
4.0
1.7
11.0
4.75
1.7
11.0
File Name:GSM06U-SPEC 2017-10-11
GSM06U
6W AC-DC Reliable Green Medical Adaptor
C
A
B
D
EIAJ equivalent
B
ID
C
L
Center Pin
P4A
5.5
3.4
P4B
6.5
4.4
11.0
11.0
1.0
1.4
P4C
7.4
Min. DIN 3 Pin with Lock (male)
Type No.
1
2
3
R6B
KYCON KPPX - 3P equivalent
2 3
1 4
R7B
KYCON KPPX - 4P equivalent
Type No.
Min. DIN 4 Pin with Lock (female)
R7BF
23
14
KYCON KPJX-CM-4S equivalent
Stripped and tinned leads
Type No.
1
2
L1
Length of Land L1 by request
( MW’s standard length, L: 25 mm, L1: 10 mm)
11.0
0.6
5.1
Pin Assignment
Output
PIN No.
+Vo
1
2
-Vo
+Vo
Pin Assignment
Output
PIN No.
+Vo
1
-Vo
2
3
-Vo
4
+Vo
Pin Assignment
Output
PIN No.
+Vo
1
-Vo
2
3
-Vo
4
+Vo
Pin Assignment
Output
PIN No.
3
Type No.
Min. DIN 4 Pin with Lock (male)
D
A
OD
Type No.
Center Pin Style
series
1
( Ribbed )
+Vo
2
( Letter )
-Vo
by customer
Installation Manual
Please refer to : http://www.meanwell.com/manual.html
File Name:GSM06U-SPEC 2017-10-11
Similar pages