NI NIPCI-6120 S series multifunction daq 12 or 16-bit,1 to 10 ms/s, 4 analog input Datasheet

S Series Multifunction DAQ 12 or 16-Bit,
1 to 10 MS/s, 4 Analog Inputs
NI 6120, NI 6115, NI 6110, NI 6111
• 2 or 4 analog inputs; dedicated
A/D converter per channel
• 1 to 10 MS/s per channel
maximum sample rate
• Analog and digital triggering
• AC or DC coupling
• 8 input ranges from ±200 mV
to ±42 V
• 2 analog outputs at 4 MS/s single
channel or 2.5 MS/s dual channel
• 8 digital I/O lines
(5 V TTL/CMOS)
• Two 24-bit counter/timers
• Measurement services that simplify
configuration and measurements
Operating Systems
• Windows 2000/NT/XP
• Mac OS X
• Linux
Recommended Software
• LabVIEW 7.x or higher
• LabWindows/CVI 7.x or higher
• Measurement Studio 7.x or higher
• Digital Waveform Editor
• SignalExpress 1.x or higher
Other Compatible Software
• VI Logger 2.x or higher
• Visual Studio .NET
• Visual Basic, C/C++, and C#
Measurement Services
Software (included)1
• NI-DAQmx driver software
• Measurement & Automation
Explorer configuration utility
• VI Logger Lite data-logging software
1Mac
OS X and Linux applications
must use NI-DAQmx Base driver software.
Calibration Certificate Available
Family
NI 6120
NI 6115
NI 6110
NI 6111
14
Bus
PCI, PXI
PCI, PXI
PCI
PCI
Analog
Inputs
4
4
4
2
Input
Resolution (bits)
16
12
12
12
Sampling
Rate (MS/s)
13
10
5
5
Input
Range (V)
±0.2 to ±42
±0.2 to ±42
±0.2 to ±42
±0.2 to ±42
Analog
Outputs
2
2
2
2
Max Output
Rate (MS/s)
41
41
41
41
Output
Range (V)
±10
±10
±10
±10
Digital
I/O
82
82
8
8
Counter/
Timers
2, 24-bit
2, 24-bit
2, 24-bit
2, 24-bit
Triggers
Analog, digital
Analog, digital
Analog, digital
Analog, digital
MS/s single channel; 2.5 MS/s on two channels 2Hardware-timed up to 10 MB/s 3800 kS/s with NI-DAQmx, 1 MS/s with additional download. Special conditions apply.
Table 1. Channel, Speed, and Resolution Specifications
Overview and Applications
Features
National Instruments S Series devices combine the latest in PC
technologies to deliver simultaneous-sampling, high-bandwidth
capabilities for low and high-channel-count data acquisition and control
systems. Use S Series devices in a variety of applications including:
• Stimulus/response applications where synchronization among highspeed analog input (AI), analog output (AO), and digital I/O is required
• 42 V automotive applications
• Radar, sonar, ultrasound
• Transient/ballistics measurements
• High-energy physics
• IF digitization
The NI high-performance S Series DAQ devices have the same
basic features of a typical multifunction device – eight digital I/O lines,
two counter/timers, two analog outputs, RTSI or PXI trigger bus for
multidevice synchronization, and PFI lines for software-controlled routing
of timing and triggering signals. S Series devices exceed the analog input
performance of traditional multiplexed devices with features including:
Family
NI 6120
NI 6115
NI 6110
NI 6111
1Filter
Onboard Memory
(samples)
Antialias Filters1
64 M
5-pole Bessel,
100 kHz
32 M or 64 M
3-pole Bessel
50 and 500 kHz
8,192
–
8,192
–
can be bypassed.
Table 2. Special Features
Signal
Coupling
AC and DC
Input
Configuration
Pseudodifferential
DIO
Hardware-timed
AC and DC
Pseudodifferential
Hardware-timed
DC
DC
Pseudodifferential
Pseudodifferential
Software-timed
Software-timed
S Series Multifunction DAQ 12 or 16-Bit,
1 to 10 MS/s, 4 Analog Inputs
Dedicated A/D Converters per Channel – help you achieve faster
sampling rates per channel, simultaneous sampling, and better
dynamic accuracy for transient or frequency measurements.
200 mV to 42 V Input Ranges – configure each analog input channel
to read from a very wide variety of voltage ranges. The input
impedance is lower for 20 and 42 V ranges, preserving the bandwidth
for signals with high frequencies and large amplitudes.
AC or DC Coupling – configure analog inputs for AC or DC coupling
on a per-channel basis. AC coupling removes the DC offset for
applications only analyzing signals in the frequency domain.
Antialias Filters (NI 6115 and NI 6120) – the antialias filters are turned
on from software on a per-channel basis. NI 6115 devices have 50 or
500 kHz filters for each channel, and NI 6120 devices have 100 kHz
filters for each channel.
NI LabVIEW Sound and Vibration Toolkit
This toolkit includes a full set of frequency-domain analysis and
presentation tools required for sound and vibration applications.
These tools include:
• Distortion, gain, phase, cross-talk, and dynamic range measurements
• Octave analysis
• Swept-sine analysis
• Linear, exponential, and peak hold measurements
• Baseband and zoom FFT, power, FFT, power spectral density (PSD)
• Limit testing
• Short-time Fourier transform and shock response spectrum
• Weighting filters
• System calibration
• Waterfall/colormap display
Large Onboard Memory (NI 6115 and NI 6120) – NI 6120 devices
feature a 64 MS onboard memory, and NI 6115 devices feature a
64 or 32 MS onboard memory. With the extra memory, which
requires no extra programming, you can capture high-bandwidth
signals over long periods of time, even if multiple devices are sharing
the same PCI or PXI bus.
Hardware-Timed Digital I/O – the digital I/O lines can synchronize
with the analog input, analog output, counter/timer I/O, or an
external clock for mixed-signal measurement applications. Using the
onboard FIFO, you can achieve reliable data transfers to 10 MB/s.
Phase Locking (PXI only) – synchronize multiple devices to a
common PXI backplane clock to preserve the phase between
measurements across multiple device analog input channels.
Professional Analysis and
Presentation Toolkits
There are several analysis and presentation add-on toolkits available
from National Instruments that help make you more productive
when developing custom applications with your S Series devices.
Turn your S Series device into a transient recorder, spectrum
analyzer, or one of several other instruments with these toolkits.
2
National Instruments • Tel: (800) 813 3693 • [email protected] • ni.com
Figure 1. Sound and Vibration Toolkit
NI LabVIEW Signal Processing Toolkit
This toolkit is a suite of software tools, example programs, and
utilities for time-frequency analysis and digital filter design. The
toolkit includes:
• Wavelet and filter-bank design
• Digital filter design of FIR and IIR filters
• Joint time-frequency analysis (JTFA) with the Gabor spectrograph
• High-frequency resolution model-based spectral estimation for
small datasets
S Series Multifunction DAQ 12 or 16-Bit,
1 to 10 MS/s, 4 Analog Inputs
NI Spectral Measurements Toolkit
This toolkit includes all of the analysis and presentation tools to
convert your S Series device into a full-featured spectrum analyzer.
The toolkit includes:
• Measurements such as in-band power and adjacent-channel power
• 3D spectrogram
• Analog modulation and demodulation
• I-Q data for digital demodulation
• Highly optimized FFT processing
High-Performance Data Throughput
Using NI-DAQmx driver software and S Series, you can take full
advantage of your PC’s throughput potential, which is useful in
many applications including transient recording.
Flexible Data Transfer
With NI-DAQmx, you can choose where to accumulate data – on the
onboard memory, PC RAM memory, or PC hard drive memory.
Thus, you can balance your PC application requirements and
stream-to-disk/memory capabilities. In any case, NI-DAQmx
optimizes data transfers by using DMA (direct memory access) and
transferring large chunks of data, while ensuring no data loss. You
can also choose the format of your data, further optimizing
throughput. Choices include raw (binary), scaled, or the high-level
waveform data type. Using these techniques, you can achieve rates of
more than 40 MS/s sustained over the PCI/PXI bus to hard drive
(depending on the PC).
Data Compaction
With the release of NI-DAQmx 7.4 or later, data compaction is
available for stream-to-disk applications. The NI-DAQmx
compaction feature strips unused or unwanted bits of data from
each sample taken from your S Series device before it is permanently
recorded to hard disk. This is helpful because hard drives are
typically the throughput bottleneck compared to the PCI/PXI bus
and PC RAM memory read/write speed. Using data compression,
users with standard hard drives can see up to 20 percent higher
throughput when streaming-to-disk.
Signal Conditioning
With the release of NI-DAQmx 7.4 or later, all S Series devices are
compatible with parallel-mode SCXI modules.
Model
SCXI-1520
SCXI-1125
SCXI-1141/2/3
Description
Universal Strain Gage Input
Isolation Amplifier
8-Pole Lowpass Filters
Table 3. Most Popular Parallel-Mode SCXI Modules
Channels
8
8
8
SCXI is limited to 100 kS/s for 16-bit accuracy and 333 kS/s for
12-bit accuracy when used in multiplexed mode. However, in
parallel mode, the sample rate is limited by the maximum scan rate
of the data acquisition device. When SCXI is used in parallel mode
with S Series, all channels are simultaneously sampled and the
sampling rate is limited by the S Series device, from 250 kS/s per
channel to 10 MS/s per channel.
Measurement Services Software
National Instruments measurement services software, built
around NI-DAQmx driver software, includes intuitive application
programming interfaces, configuration tools, I/O assistants, and
other tools designed to reduce system setup, configuration, and
development time.
Automatic Code Generation – DAQ Assistant is an interactive guide
that helps you navigate through configuring, testing, and programming
measurement tasks and automatically generates the necessary code
for LabVIEW, LabWindows/CVI, or Measurement Studio.
Cleaner Code Development – Basic and advanced software functions
have been combined into one easy-to-use yet powerful set to help
you build cleaner code and move from basic to advanced
applications without replacing functions.
High-Performance Driver Engine – NI-DAQmx delivers maximum
I/O system throughput with a multithreaded driver.
Test Panels – With the Measurement & Automation Explorer
configuration utility, you can test all of your module functionality
before you begin development.
Scaled Channels – Easily scale your voltage data into the proper
engineering units using the NI-DAQmx Measurement Ready virtual
channels by choosing from a list of common sensors and signals or
creating your own custom scale.
Data-Logging Software – VI Logger Lite is configuration-based
software designed specifically for data logging. Features include
easy logging and viewing of data, data extraction to Microsoft
Excel, and code generation in LabVIEW.
LabVIEW Integration – All NI-DAQmx functions create the
waveform data type, which carries acquired data and timing
information directly into more than 400 LabVIEW built-in analysis
routines for display of results in engineering units on a graph.
NI-DAQmx Base Driver
NI-DAQmx Base (available at ni.com/downloads) offers Mac OS X
and Linux users a programming interface similar to NI-DAQmx. It
includes ready-to-use LabVIEW VIs and C function features similar
to those included in NI-DAQmx driver software.
National Instruments • Tel: (800) 813 3693 • [email protected] • ni.com
3
S Series Multifunction DAQ 12 or 16-Bit,
1 to 10 MS/s, 4 Analog Inputs
+
CH0
Amplifier
–
AntiAliasing
Filter
CH0
12-Bit
ADC
12
CH0
Latch
Data (16)
AI CH1
Mux
+
CH1
Amplifier
–
AntiAliasing
Filter
CH1
12-Bit
ADC
12
CH1
Latch
Data (16)
AI CH2
Mux
+
CH2
Amplifier
–
AntiAliasing
Filter
CH2
12-Bit
ADC
12
CH2
Latch
Data (16)
+
CH3
Amplifier
–
AntiAliasing
Filter
AI CH0
Mux
CH0–
CH1+
CH1–
CH2+
CH2–
I/O Connector
CH3+
AI CH3
Mux
CH3–
Control
ADC
FIFO
CH3
12-Bit
ADC
12
CH3
Latch
Calibration
Mux
Address/Data
Data (16)
AI Control
EEPROM
IRQ
Trigger Level
DACs
Trigger
2
Analog
Trigger
Circuitry
DMA
PFI / Trigger
Timing
STC DIO (8)
Digital I/O (8)
DIO
Mux
Trigger
Analog Input
Timing/Control
DMA/IRQ
Counter/
Timing I/O
DAQ - STC
Bus
Interface
Digital I/O
Analog Output
Timing/Control
RTSI Bus
Interface
DIO FIO
AO Control
FPGA Digital I/O (8)
DAC0
DAC
FIFO
Data (12)
Data (32)
DAC1
RTSI Bus
Calibration
DACs
Figure 2. NI 6120 and NI 6115 Hardware Block Diagram
4
Data (32)
Generic Mini
PCI
Bus
Bus
MITE
Interface
Interface
National Instruments • Tel: (800) 813 3693 • [email protected] • ni.com
Analog Input
Control
EEPROM
Control
DMA
Interface
DAQ-STC
Bus Interface
FPGA
DIO
FIFO
Analog
Output Control
DIO
Control
I/O Bus
Interface
PXI/CompactPCI Bus
CH0+
S Series Multifunction DAQ 12 or 16-Bit,
1 to 10 MS/s, 4 Analog Inputs
Voltage
REF
Calibration
DACs
Gain
Amplifier
ACH0+
ACH0-
AI CH0
Mux
+
CH0
12-Bit ADC
ACH1+
ACH1-
AI CH1
Mux
+
-
CH1
12-Bit ADC
ACH2+
ACH2-
AI CH21
Mux
+
CH2
12-Bit ADC
AI CH31
Mux
+
CH3
12-Bit ADC
-
Control
Generic
Bus
Interface
ADC
FIFO
MITE
PCI
Bus
Interface
Address/Data
Not available on PCI-6111
Trigger Level
DACs
AI Control
IRQ
DMA
Analog
Trigger
Circuitry
Trigger
EEPROM
Analog Input
Timing/Control
DMA/
Interrupt
Request
Counter/
Timing I/O
DAQ - STC
Bus
Interface
Digital I/O
Analog Output
Timing/Control
RTSI Bus
Interface
PFI / Trigger
Trigger
Timing
Digital I/O (8)
Analog
Input
Control
EEPROM
Control
DAQ-STC
Bus
Interface
MIO
Interface
Analog
Output
Control
DMA
Interface
I/O
Bus
Interface
PCI Bus
Calibration
Mux
Address (5)
1
I/O Connector
-
Data (32)
ACH3+
ACH3-
-
AO Control
DAC0
DAC
FIFO
Data (32)
DAC1
Calibration
DACs
RTSI Bus
Figure 3. NI 6110 and 6111 Hardware Block Diagram
Ordering Information
PCI
PXI
NI PCI-6120 ..................................................................778397-01
NI PCI-6115 with
32 MS ........................................................................778534-01
64 MS ........................................................................778554-01
NI PCI-6111 ..................................................................777530-01
NI PCI-6110 ..................................................................777475-01
NI PXI-6115 with
32 MS ........................................................................778194-01
64 MS ........................................................................778204-01
NI PXI-6120 ..................................................................778396-01
Includes data acquisition driver software.
BUY NOW!
For complete product specifications, pricing, and accessory
information, call (800) 813 3693 (U.S. only) or go to
ni.com/dataacquisition.
National Instruments • Tel: (800) 813 3693 • [email protected] • ni.com
5
S Series Multifunction DAQ 12 or 16-Bit,
1 to 10 MS/s, 4 Analog Inputs
Specifications
These specifications are typical for 25 °C unless otherwise noted.
Timing I/O
Analog Input
Number of channels
Up/down counter/timers ...........................
Frequency scaler........................................
Resolution
Up/down counter/timers ...........................
Frequency scaler........................................
Compatibility ....................................................
Base clocks available
Up/down counter/timers ...........................
Frequency scaler........................................
Base clock accuracy .........................................
Maximum source frequency
Up/down counter/timers ...........................
Minimum source pulse duration ......................
Minimum gate pulse duration..........................
Data transfers ..................................................
Input Characteristics
Number of channels
NI 6110, NI 6115, NI 6120 ........................ 4 pseudodifferential
NI 6111 ...................................................... 2 pseudodifferential
ADC resolution
NI 6110, NI 6111, NI 6115 ........................ 12 bits, 1 in 4,096
NI 6120 ...................................................... 16 bits, 1 in 65,536
ADC pipeline
NI-6110, NI 6111 ....................................... 3
NI 6115 ...................................................... 4
NI 6120 ...................................................... 0
Maximum sampling rate
NI 6110, NI 6111 ....................................... 5 MS/s
NI 6115 ...................................................... 10 MS/s
NI 6120 ...................................................... 1 MS/s1
Minimum sampling rate
NI 6110, NI 6111 ....................................... 1 kS/s
NI 6115 ...................................................... 20 kS/s
NI 6120 ...................................................... No minimum
Input signal ranges
(selectable by channel).............................. ±42, ±20, ±10, ±5, ±2, ±1, ±0.5, ±0.2 V
Input coupling................................................... AC or DC
Max working voltage for all analog input channels
Overvoltage protection
(ACH+, ACH-) ............................................. ±42 V
Input FIFO buffer
NI 6110, NI 6111 ....................................... 8,192 samples
NI 6115, NI 6120 ....................................... 16 Msamples or 32 Msamples
Data transfers .................................................. DMA (scatter-gather), interrupts, programmed I/O
Analog filter type
NI 6115 ...................................................... 3-pole Bessel
NI 6120 ...................................................... 5-pole Bessel
Analog filter frequency (-3 dB)
NI 6115 ...................................................... 50 and 500 kHz
NI 6120 ...................................................... 100 kHz
1800
kS/s with NI-DAQmx, 1 MS/s with additional download. Special conditions apply.
Analog Output
Output Characteristics
Number of channels.........................................
Resolution
NI 6110, NI 6111, NI 6120 ........................
NI 6115 ......................................................
Max update rate
1 channel ..................................................
2 channel ...................................................
Output FIFO buffer size
NI 6110, NI 6111 .......................................
NI 6115, NI 6120 .......................................
Data transfers ..................................................
2 voltage
16 bits, 1 in 65,536
12 bits, 1 in 4,096
20 MHz, 100 kHz
10 MHz, 100 kHz
±0.01%
20 MHz
10 ns, edge-detect mode
10 ns, edge-detect mode
DMA (scatter-gather), interrupts, programmed I/O
Triggers
Analog Trigger
Purpose
Analog input ..............................................
Analog output ............................................
General-purpose counter/timers ...............
Source...............................................................
Level
Internal source, ACH<0..3> .......................
External source, PFI0/TRIG1......................
Slope.................................................................
Resolution
NI 6110, NI 6111, NI 6115 ........................
NI 6120 ......................................................
Hysteresis.........................................................
Bandwidth (-3 dB)
Internal source, ACH<0..3> .......................
External source, PFI0/TRIG1......................
Start and stop trigger, gate, clock
Start trigger, gate, clock
Source, gate
All analog input channels, PFI0/TRIG1
±Full-scale
±10 V
Positive or negative; software-selectable
8 bits, 1 in 256
12 bits, 1 in 4,096
Programmable
5 MHz
5 MHz
Digital Trigger
Purpose
Analog input ..............................................
Analog output ............................................
General-purpose counter/timers ...............
External Sources ..............................................
Compatibility ....................................................
Response ..........................................................
Pulse width.......................................................
Start and stop trigger, gate, clock
Start trigger, gate, clock
Source, gate
PFI<0..9>, RTSI<0..6>
5 V/TTL
Rising or falling edge
10 ns minimum
PXI Trigger Bus (PXI only)
RTSI Bus (PCI only)
Trigger lines...................................................... 7
2,048 samples
16 or 32 Msamples
DMA (scatter-gather), interrupts, programmed I/O
Bus Interface
PXI or PCI.......................................................... Master, slave
Physical
Ranges.............................................................. ±10 V
Output coupling ................................................ DC
Digital I/O
6
24 bits
4 bits
TTL/CMOS
Trigger lines...................................................... 7
4 MS/s
2.5 MS/s
Voltage Output
Number of channels.........................................
Compatibility ....................................................
Power-on state .................................................
Data transfers
NI 6110, NI 6111 .......................................
NI 6115, NI 6120 .......................................
Transfer rate
NI 6115, NI 6120 .......................................
Input buffer size
NI 6115, NI 6120 .......................................
Output buffer size
NI 6115, NI 6120 .......................................
2
1
8 input/output
TTL/CMOS
Input (high-impedence)
Programmed I/O
DMA (scatter-gather), interrupts, programmed I/O
10 Mbytes/s
2,048 bytes
2,048 bytes
National Instruments • Tel: (800) 813 3693 • [email protected] • ni.com
Dimensions (not including connectors)
PCI ............................................................ 31.2 by 10.6 cm (12.3 by 4.2 in.)
PXI ............................................................ 16 by 10 cm (6.3 by 3.9 in.)
I/O connector.................................................... 68-pin male SCSI II type
S Series Multifunction DAQ
Cables and Accessories
Recommended Configurations
I/O Configuration
BNC shielded
Front-mount shielded screw terminals (PXI only)
Noise-rejecting shielded screw terminals
Low-cost unshielded screw terminals
Accessory
BNC-2110 (777643-01)
TB-2705 (778241-01)
SCB-68 (776844-01)
CB-68LPR (777145-02)
Cable
SH68-68-EP (184749-01)
–
SH68-68-EP (184749-01)
R6868 (777643-01)
SH68-68-EP Noise-Rejecting Shielded Cable
SH68-68-EP Shielded Cable
The SH68-68-EP is a shielded 68-conductor cable terminated with two 68-pin female
0.050 series D-type connectors. It features individually shielded analog twisted pairs
for reduced cross-talk with high-speed devices.
1 m ....................................................................................................................184749-01
2 m ....................................................................................................................184749-02
R6868 Low-Cost, Ribbon Cable
R6868 Ribbon Cable
The R6868 is a 68-conductor flat ribbon cable terminated with two 68-pin female
0.050 series D-type connectors.
1 m ....................................................................................................................182482-01
BNC-2110 Noise-Rejecting BNC I/O Connector Block
The BNC-2110 is a shielded connector block with signal-labeled BNC connectors for
easy connectivity of your analog input, digital I/O, and counter/timer signals.
BNC-2110 ........................................................................................................777643-01
Dimensions – 20.3 by 11.2 by 5.5 cm (8.0 by 4.4 by 2.2 in.)
TB-2705 Noise-Rejecting Front-Mount I/O Connector Block
BNC-2110 I/O Connector Block
The TB-2705 is a screw terminal block for your PXI DAQ devices. It latches to the
front of your PXI modules with locking screws and provides strain relief and easy access
to your analog, digital, trigger, and counter/timer signals through screw terminals.
TB-2705 ............................................................................................................778241-01
Dimensions – 8.43 by 10.41 by 2.03 cm (3.32 by 4.1 by 0.8 in.)
SCB-68 Noise-Rejecting Shielded I/O Connector Block
TB-2705 Terminal Block
The SCB-68 is a shielded I/O connector block for rugged, very-low-noise signal
termination for connecting 68-pin DAQ devices. Silk-screened component locations
provide an easy addition of simple signal-conditioning circuitry for your analog input
channels. It also includes a general-purpose breadboard area as well as an IC
temperature sensor for cold-junction compensation in temperature measurements.
SCB-68 ..............................................................................................................776844-01
Dimensions – 19.5 by 15.2 by 4.5 cm (7.7 by 6.0 by 1.8 in.)
CB-68LP and CB-68LPR Low-Cost I/O Connector Blocks
SCB-68 Shielded I/O Connector Block
The CB-68 and CB-68LP are low-cost termination accessories with 68 screw
terminals for easy connector of field I/O signals to 68-pin DAQ devices. The
connector blocks include standoffs for use on a desktop or for mounting in a custom
panel. The CB-68LP has a vertical-mounted 68-pin connector. The CB-68-LPR has a
right-angle mounted connector, and it is used with the CA-1000.
CB-68LP ............................................................................................................777145-01
Dimensions – 14.35 by 10.74 cm (5.65 by 4.23 in.)
CB-68LPR ........................................................................................................777145-02
Dimensions – 7.62 by 16.19 cm (3.00 by 6.36 in.)
CB-68LP and CB-68LPR I/O Connector Blocks
National Instruments • Tel: (800) 813 3693 • [email protected] • ni.com
7
NI Services and Support
NI has the services and support to meet your
needs around the globe and through the
application life cycle – from planning
and development through deployment
and ongoing maintenance. We offer
services and service levels to meet
customer requirements in research,
design, validation, and manufacturing.
Visit ni.com/services.
Local Sales and Technical Support
SERVICE
NEEDS
In offices worldwide, our staff is local to the country, giving you
access to engineers who speak your language. NI delivers industryleading technical support through online knowledge bases, our
applications engineers, and access to 14,000 measurement and
automation professionals within NI Developer Exchange forums.
Find immediate answers to your questions at ni.com/support.
We also offer service programs that provide automatic upgrades to
your application development environment and higher levels of
technical support. Visit ni.com/ssp.
Training and Certification
NI training is the fastest, most certain route to productivity with our
products. NI training can shorten your learning curve, save
development time, and reduce maintenance costs over the
application life cycle. We schedule instructor-led courses in cities
worldwide, or we can hold a course at your facility. We also offer a
professional certification program that identifies individuals who
have high levels of skill and knowledge on using NI products.
Visit ni.com/training.
Hardware Services
NI Factory Installation Services
NI Factory Installation Services (FIS) is the fastest and easiest way to
use your PXI or PXI/SCXI combination systems right out of the box.
Trained NI technicians install the software and hardware and
configure the system to your specifications. NI extends the standard
warranty by one year on hardware components (controllers, chassis,
modules) purchased with FIS. To use FIS, simply configure your
system online with ni.com/pxiadvisor.
Professional Services
Our Professional Services Team is comprised of NI applications
engineers, NI Consulting Services, and a worldwide National Instruments
Alliance Partner program of more than 600 independent consultants
and integrators. Services
range from start-up assistance
to turnkey system integration.
Visit ni.com/alliance.
Calibration Services
NI recognizes the need to maintain properly calibrated devices for
high-accuracy measurements. We provide manual calibration
procedures, services to recalibrate your products, and automated
calibration software specifically designed for use by metrology
laboratories. Visit ni.com/calibration.
Repair and Extended Warranty
We offer design-in consulting and product integration assistance
if you want to use our products for OEM applications. For
information about special pricing and services for OEM customers,
visit ni.com/oem.
NI provides complete repair services for our products. Express repair
and advance replacement services are also available. We offer
extended warranties to help you meet project life-cycle requirements.
Visit ni.com/services.
ni.com • (800) 813 3693
National Instruments • [email protected]
© 2005 National Instruments Corporation. All rights reserved. CVI, DAQ-STC, LabVIEW, Measurement Studio, MITE, National Instruments, National Instruments Alliance Partner, NI,
ni.com, NI-DAQ, RTSI, SCXI, and SignalExpress are trademarks of National Instruments. Other product and company names listed are trademarks or trade names of their respective
companies. A National Instruments Alliance Partner is a business entity independent from NI and has no agency, partnership, or joint-venture relationship with NI.
2005_4961_301_101_D_6120_611x
OEM Support
Similar pages