[ Obsolete product ] Agilent has a new name Keysight Technologies. Keysight Technologies Inc. is the world's leading electronic measurement company, transforming today's measurement experience through innovations in wireless, modular, and software solutions. With its HP and Agilent legacy, Keysight delivers solutions in wireless communications, aerospace and defense and semiconductor markets with world-class platforms, software and consistent measurement science. Alldatasheet.com Agilent E5071C ENA Network Analyzer • 9 kHz to 4.5/6.5/8.5 GHz • 100 kHz to 4.5/6.5/8.5 GHz (with bias tees) • 300 kHz to 14/20 GHz (with bias tees) E5091A Multiport Test Set E5092A Configurable Multiport Test Set Data Sheet Table of Contents Options ........................................................................................................................ 3 Definitions ................................................................................................................. 4 Boundary Conditions ............................................................................................. 4 Corrected System Performance........................................................................ 5 System dynamic range ..................................................................................................... 5 Corrected system performance with calibration kit.................................................... 7 Uncorrected System Performance ................................................................13 Test Port Output (Source) .................................................................................14 Test port output frequency ............................................................................................ 14 Test port output power .................................................................................................. 15 Test port output signal purity ........................................................................................ 17 Test Port Input ........................................................................................................18 Test port input levels ...................................................................................................... 18 Test port input (trace noise) .......................................................................................... 20 Test port input (stability) ............................................................................................... 22 Test port input (dynamic accuracy) .............................................................................. 23 Test port input (group delay) ......................................................................................... 25 General Information.............................................................................................27 System bandwidths ......................................................................................................... 27 Front panel information .................................................................................................. 27 Rear panel information ................................................................................................... 28 LXI compliance................................................................................................................. 30 EMC, safety and environment ....................................................................................... 31 Analyzer environmental specifications and dimensions .......................................... 32 Measurement Throughput Summary ............................................................36 Typical cycle time for measurement completion (ms).............................................. 36 Cycle time (ms) vs. number of points.......................................................................... 48 Cycle time (ms) vs. IF bandwidth ................................................................................. 49 Data transfer time (ms) .................................................................................................. 50 E5091A Multiport Test Set.................................................................................51 Test set input/output performance .............................................................................. 51 Option E5091A-009 port performance ......................................................................... 51 Option E5091A-016 port performance ......................................................................... 52 Front panel information .................................................................................................. 52 Rear panel information ................................................................................................... 52 Test set dimensions and block diagram ...................................................................... 53 E5092A Configurable Multiport Test Set .....................................................56 Test set input/output performance .............................................................................. 56 Option E5092A-020 port performance ......................................................................... 56 Control line ....................................................................................................................... 58 DC source.......................................................................................................................... 58 Operating storage environment .................................................................................... 58 Non-operating storage environment ............................................................................ 58 Front panel information .................................................................................................. 59 Rear panel information ................................................................................................... 59 Test set dimensions and block diagram ...................................................................... 59 Corrected System Performance for 75 Ω Measurements with 11852B 50 to 75 Ω Minimum-Loss Pads ................................................ 63 2 Options This document provides technical specifications for the E5071C ENA network analyzer and the E5091A multiport test set. E5071C-230 E5071C-235 E5071C-240 E5071C-245 E5071C-260 E5071C-265 E5071C-280 E5071C-285 E5071C-430 E5071C-435 E5071C-440 E5071C-445 E5071C-460 E5071C-465 E5071C-480 E5071C-485 E5071C-2D5 E5071C-4D5 E5071C-2K5 E5071C-4K5 E5091A E5092A 2-port test set, 9 kHz to 3 GHz without bias tees (Discontinued) 2-port test set, 100 kHz to 3 GHz with bias tees (Discontinued) 2-port test set, 9 kHz to 4.5 GHz without bias tees 2-port test set, 100 kHz to 4.5 GHz with bias tees 2-port test set, 9 kHz to 6.5 GHz without bias tees 2-port test set, 100 kHz to 6.5 GHz with bias tees 2-port test set, 9 kHz to 8.5 GHz without bias tees 2-port test set, 100 kHz to 8.5 GHz with bias tees 4-port test set, 9 kHz to 3 GHz without bias tees (Discontinued) 4-port test set, 100 kHz to 3 GHz with bias tees (Discontinued) 4-port test set, 9 kHz to 4.5 GHz without bias tees 4-port test set, 100 kHz to 4.5 GHz with bias tees 4-port test set, 9 kHz to 6.5 GHz without bias tees 4-port test set, 100 kHz to 6.5 GHz with bias tees 4-port test set, 9 kHz to 8.5 GHz without bias tees 4-port test set, 100 kHz to 8.5 GHz with bias tees 2-port test set, 300 kHz to 14 GHz with bias tees 4-port test set, 300 kHz to 14 GHz with bias tees 2-port test set, 300 kHz to 20 GHz with bias tees 4-port test set, 300 kHz to 20 GHz with bias tees Multiport test set Configurable multiport test set Calibration kits and ECal modules This E5071C data sheet also provides technical specifications for the following calibration kits and ECal modules. For models not listed in this data sheet, please download the free Uncertainty Calculator from www.agilent.com/find/na_calculator to generate the curves for your calibration kit and enable ENA setup. 85032F 85033E 85052D 85092C 85093C N4691B Calibration kit Calibration kit Calibration kit Electronic calibration (ECal) module Electronic calibration (ECal) module Electronic calibration (ECal) module 3 Definitions Specification (spec.): Warranted performance. All specifications apply at 23 ºC (± 5 ºC), unless otherwise stated, and 90 minutes after the instrument has been turned on. Specifications include guard bands to account for the expected statistical performance distribution, measurement uncertainties, and changes in performance due to environmental conditions. Supplemental information is intended to provide information that is helpful for using the instrument but that is not guaranteed by the product warranty. Typical (typ.): Describes performance that will be met by a minimum of 80% of all products. It is not guaranteed by the product warranty. Supplemental performance data (SPD): Supplemental performance data represents the value of a parameter that is most likely to occur; the expected mean or average. It is not guaranteed by the product warranty. General characteristics: A general, descriptive term that does not imply a level of performance. Boundary Conditions In this data sheet, boundary conditions are given for the specifications. For example, system dynamic range is 98 dB with the following boundary conditions. Option: 485 Frequency: 10 MHz IF bandwidth: 3 kHz If the same boundary conditions fall under more than one category in a table, apply the best value. 4 Corrected System Performance The specifications in this section apply to measurements made with the Agilent E5071C network analyzer under the following conditions: • No averaging applied to data • Environmental temperature of 23 °C (±5 °C) with less than 1 °C deviation from the calibration temperature • Response and isolation calibration performed System dynamic range Table 1. Option 230/235/240/245/260/265/280/285/430/435/440/445/ 460/465/480/485 Description Specification SPD System dynamic range1, 2 9 kHz to 300 kHz 300 kHz to 10 MHz 72 dB IF bandwidth = 3 kHz 82 dB 10 MHz to 6 GHz 98 dB 6 GHz to 8.5 GHz 92 dB 9 kHz to 300 kHz 97 dB 115 dB 300 kHz to 10 MHz 107 dB 115 dB 123 dB 130 dB 6 GHz to 7 GHz 117 dB 128 dB 7 GHz to 8 GHz 117 dB 126 dB 8 GHz to 8.5 GHz 117 dB 124 dB 10 MHz to 6 GHz IF bandwidth = 10 Hz System dynamic range [dB] 160 150 140 130 120 110 E5071C specification 100 90 80 1.E+05 1.E+09 2.E+09 3.E+09 4.E+09 5.E+09 6.E+09 7.E+09 8.E+09 Test frequency [Hz] Figure 1. System dynamic range (specification and actual measurement data example, IF bandwidth 10 Hz) 1. The test port dynamic range is calculated as the difference between the test port rms noise floor and the source maximum output power. The effective dynamic range must take measurement uncertainty and interfering signals into account. 2. The specification might not be met at 5 MHz or 50 MHz. 5 System dynamic range (continued) Table 2. Option 2D5/2K5/4D5/4K5 Description Specification SPD System dynamic range1, 2 300 kHz to 1 MHz 70 dB 1 MHz to 10 MHz 82 dB 10 MHz to 100 MHz 95 dB 100 MHz to 6 GHz IF bandwidth = 3 kHz 98 dB 6 GHz to 8.5 GHz 92 dB 8.5 GHz to 10.5 GHz 80 dB 10.5 GHz to 15 GHz 75 dB 15 GHz to 20 GHz 71 dB 300 kHz to 1 MHz 95 dB 105 dB 1 MHz to 10 MHz 107 dB 115 dB 10 MHz to 100 MHz 120 dB 129 dB 123 dB 130 dB 6 GHz to 8 GHz 117 dB 129 dB 8 GHz to 8.5 GHz 117 dB 127 dB 8.5 GHz to 10.5 GHz 105 dB 115 dB 10.5 GHz to 15 GHz 100 dB 111 dB 15 GHz to 20 GHz 96 dB 105 dB 100 MHz to 6 GHz IF bandwidth = 10 Hz 160 System dynamic range[dB] 150 140 130 120 110 100 E5071C-2D5/2K5/4D5/4K5 specification 90 80 0.0E+00 5.0E+09 1.0E+10 1.5E+10 2.0E+10 Frequency [Hz] Figure 2. System dynamic range (specification and actual measurement data example, IF bandwidth 10 Hz) 1. The test port dynamic range is calculated as the difference between the test port’s rms noise floor and the source’s maximum output power. Effective dynamic range must take measurement uncertainty and interfering signals into account. 2. The specification might not be met at 5 MHz or 50 MHz. 6 Corrected system performance with calibration kit Option 230/235/240/245/260/265/280/285/430/435/440/445/460/465/480/485 Table 3. Corrected system performance with type-N device connectors, 85032F calibration kit Network analyzer: Calibration kit: Calibration: E5071C 85032F (Type-N, 50 Ω) full 2-port IF bandwidth = 10 Hz, no averaging applied to data, environmental temperature = 23 °C (±5 °C ) with < 1 °C deviation from calibration temperature, isolation calibration performed Specification (dB) Description 9 kHz to 300 kHz 300 kHz to 10 MHz 10 MHz to 3 GHz 3 GHz to 6 GHz 6 GHz to 8.5 GHz Directivity 49 49 46 40 38 Source match 41 41 40 36 35 Load match 49 49 46 40 37 Reflection tracking ±0.011 ±0.011 ±0.021 ±0.032 ±0.054 Transmission tracking ±0.027 ±0.015 ±0.018 ±0.056 ±0.088 Transmission uncertainty (specification) Magnitude 9 KHz to 300 KHz 300 KHz to 10 MHz 10 MHz to 3 GHz 3 GHz to 6 GHz 6 GHz to 8.5 GHz 1 0.1 0.01 10 Phase 100 Uncertainty (degrees) Uncertainty (dB) 10 9 KHz to 300 KHz 300 KHz to 10 MHz 10 MHz to 3 GHz 3 GHz to 6 GHz 6 GHz to 8.5 GHz 10 1 S 1 1 = S 2 2 = 0 ; C a l p o w e r = -1 0 d B m ; M e a s p o w e r = -1 0 d B m 0 –10 –20 –30 –40 –50 –60 S 1 1 = S 2 2 = 0 ; C a l p o w e r = -1 0 d B m ; M e a s p o w e r = -1 0 d B m 0.1 –70 –80 –90 10 Transmission coefficient (dB) 0 -10 -20 -30 -40 -50 -60 -70 -80 -90 Transmission coefficient (dB) Reflection uncertainty (specification) Magnitude 10 9 KHz to 300 KHz, 300 KHz to 10 MHz 10 MHz to 3 GHz 3 GHz to 6 GHz 6 GHz to 8.5 GHz 0.04 Uncertainty (degrees) Uncertainty (linear) 0.05 0.03 0.02 0.01 Phase 9 KHz to 300 KHz, 300 KHz to 10 MHz 10 MHz to 3 GHz 3 GHz to 6 GHz 6 GHz to 8.5 GHz 8 6 4 2 S21 = S12 = 0; Cal power = -10 dBm; Meas power = -10 dBm S21 = S12 = 0; Cal power = -10 dBm; Meas power = -10 dBm 0 0 0 0.2 0.4 0.6 0.8 1 0 0.2 Reflection coefficient (linear) 0.4 0.6 Reflection coefficient (linear) 7 0.8 1 Table 4. Corrected system performance with type-N device connectors, 85092C electronic calibration (ECal) module Network analyzer: Calibration module: E5071C 85092C (Type-N, 50 Ω) Electronic calibration (ECal) module full 2-port Calibration: IF bandwidth = 10 Hz, no averaging applied to data, environmental temperature = 23 °C (±5 °C ) with < 1 °C deviation from calibration temperature, isolation calibration performed Specification (dB) Description 300 kHz to 10 MHz 10 MHz to 3 GHz 3 GHz to 6 GHz 6 GHz to 8.5 GHz Directivity 45 52 52 47 Source match 36 44 41 36 Load match 41 47 44 39 Reflection tracking ±0.100 ±0.040 ±0.060 ±0.070 Transmission tracking ±0.055 ±0.039 ±0.068 ±0.136 Transmission uncertainty (specification) Magnitude Phase 10 100 3 0 0 K H z to 1 0 M H z 3 0 0 kH z to 1 0 M H z 1 0 M H z to 3 G H z 1 0 M H z to 3 G H z Uncertainty (degrees) 3 G H z to 6 G H z Uncertainty (dB) 6 G H z to 8 .5 G H z 1 0 .1 3 G H z to 6 G H z 6 G H z to 8 .5 G H z 10 1 S 11 = S 22 = 0 ; Cal power = -10 dBm; Meas power = -10 dBm S 11 = S 22 = 0 ; Cal power = -10 dBm; Meas power = -10 dBm 0 .1 0 .0 1 10 0 -1 0 -2 0 -3 0 -4 0 -5 0 -6 0 -7 0 -8 0 10 -9 0 0 -1 0 -2 0 -3 0 -4 0 -5 0 -6 0 -7 0 -8 0 -9 0 Transmission coefficient (dB) Transmission coefficient (dB) Reflection uncertainty (specification) Magnitude Phase 0 .0 5 10 3 0 0 kH z to 1 0 M H z 3 0 0 kH z to 1 0 M H z 1 0 M H z to 3 G H z 3 G H z to 6 G H z 6 G H z to 8 .5 G H z 1 0 M H z to 3 G H z 8 3 G H z to 6 G H z 6 G H z to 8 .5 G H z Uncertainty (linear) Uncertainty (linear) 0 .0 4 S 2 1 = S 1 2 = 0 ; Cal power = -10 dBm; Meas power = -10 dBm 0 .0 3 0 .0 2 0 .0 1 S 2 1 = S 1 2 = 0 ; Cal power = -10 dBm; Meas power = -10 dBm 6 4 2 0 0 0 0 .2 0 .4 0 .6 0 .8 1 0 0 .2 Reflection coefficient (linear) 0 .4 0 .6 Reflection coefficient (linear) 8 0 .8 1 Table 5. Corrected system performance with 3.5 mm device connector type, 85033E calibration kit Network analyzer: Calibration kit: Calibration: E5071C 85033E (3.5 mm, 50 Ω) full 2-port IF bandwidth = 10 Hz, no averaging applied to data, environmental temperature = 23 °C (±5 °C ) with < 1 °C deviation from calibration temperature, isolation calibration performed Specification (dB) 9 kHz to 300 kHz Description 300 kHz to 10 MHz 10 MHz to 3 GHz 3 GHz to 6 GHz 6 GHz to 8.5 GHz Directivity 46 46 44 38 38 Source match 43 43 40 37 36 Load match 46 46 44 38 38 Reflection tracking ±0.006 ±0.006 ±0.007 ±0.009 ±0.010 Transmission tracking ±0.026 ±0.015 ±0.020 ±0.058 ±0.079 Transmission uncertainty (specification) Magnitude 9 KHz to 300 KHz 300 KHz to 10 MHz 10 MHz to 3 GHz 3 GHz to 6 GHz 6 GHz to 8.5 GHz 1 0.1 S11 = S22 = 0; Cal power = -10 dBm; Meas power = -10 dBm 0 -10 -20 -30 -40 -50 -60 -70 -80 9 KHz to 300 KHz 300 KHz to 10 MHz 10 MHz to 3 GHz 3 GHz to 6 GHz 6 GHz to 8.5 GHz 10 1 S11 = S22 = 0; Cal power = -10 dBm; Meas power = -10 dBm 0.1 0.01 10 Phase 100 Uncertainty (degrees) Uncertainty (dB) 10 10 -90 Transmission coefficient (dB) 0 -10 -20 -30 -40 -50 -60 -70 -80 -90 Transmission coefficient (dB) Reflection uncertainty (specification) Magnitude 10 Phase 9 KHz to 300 KHz, 300 KHz to 10 MHz 10 MHz to 3 GHz 3 GHz to 6 GHz 6 GHz to 8.5 GHz 0.04 Uncertainty (degrees) Uncertainty (linear) 0.05 0.03 0.02 0.01 0 0.2 0.4 0.6 Reflection coefficient (linear) 0.8 8 6 4 2 S21 = S12 = 0; Cal power = -10 dBm; Meas power = -10 dBm S21 = S12 = 0; Cal power = -10 dBm; Meas power = -10 dBm 0 9 KHz to 300 KHz, 300 KHz to 10 MHz 10 MHz to 3 GHz 3 GHz to 6 GHz 6 GHz to 8.5 GHz 0 1 0 0.2 0.4 0.6 Reflection coefficient (linear) 9 0.8 1 Table 6. Corrected system performance with 3.5 mm device connector type, 85093C electronic calibration (ECal) module Network analyzer: Calibration module: E5071C 85093C (3.5 mm, 50 Ω) electronic calibration (ECal) module full 2-port Calibration: IF bandwidth = 10 Hz, no averaging applied to data, environmental temperature = 23 °C (±5 °C ) with < 1 °C deviation from calibration temperature, isolation calibration performed Specification (dB) Description 300 kHz to 10 MHz 10 MHz to 3 GHz 3 GHz to 6 GHz 6 GHz to 8.5 GHz Directivity 45 52 51 47 Source match 36 44 39 34 Load match 41 47 44 40 Reflection tracking ±0.100 ±0.040 ±0.050 ±0.070 Transmission tracking ±0.055 ±0.049 ±0.068 ±0.117 Transmission uncertainty (specification) Phase Magnitude 10 100 Uncertainty (degrees) Uncertainty (dB) 3 0 0 K Hz to 1 0 MHz 1 0 MHz to 3 G Hz 3 G Hz to 6 G Hz 6 G Hz to 8 .5 G Hz 1 0 .1 3 0 0 kH z to 1 0 M H z 1 0 M H z to 3 G H z 3 G H z to 6 G H z 6 G H z to 8 .5 G H z 10 1 S 11 = S 22 = 0 ; Cal power = -10 dBm; Meas power = -10 dBm S 11 = S 22 = 0 ; Cal power = -10 dBm; Meas power = -10 dBm 0 .1 0 .0 1 10 0 -1 0 -2 0 -3 0 -4 0 -5 0 -6 0 -7 0 -8 0 10 -9 0 0 -1 0 -2 0 -3 0 -4 0 -5 0 -6 0 -7 0 -8 0 -9 0 Transmission coefficient (dB) Transmission coefficient (dB) Reflection uncertainty (specification) Phase Magnitude 10 0 .0 5 3 0 0 kH z to 1 0 M H z 1 0 M H z to 3 G H z 3 G H z to 6 G H z 6 G H z to 8 .5 G H z 3 0 0 kH z to 1 0 M H z 1 0 M H z to 3 G H z Uncertainty (degrees) Uncertainty (linear) 8 3 G H z to 6 G H z 0 .0 4 6 G H z to 8 .5 G H z S 2 1 = S 1 2 = 0 ; Cal power = -10 dBm; Meas power = -10 dBm 0 .0 3 0 .0 2 S 2 1 = S 1 2 = 0 ; Cal power = -10 dBm; Meas power = -10 dBm 6 4 2 0 .0 1 0 0 0 0 .2 0 .4 0 .6 0 .8 1 0 0 .2 0 .4 0 .6 Reflection coefficient (linear) Reflection coefficient (linear) 10 0 .8 1 Option 2D5/2K5/4D5/4K5 Table 7. Corrected system performance with 3.5 mm device connectors, 85052D calibration kit Network analyzer: Calibration kit: Calibration: E5071C 85052D (3.5 mm, 50 Ω) full 2-port IF bandwidth = 10 Hz, no averaging applied to data, environmental temperature = 23 °C (±5 °C ) with < 1 °C deviation from calibration temperature, isolation calibration performed Specification (dB) Description 300 kHz to 500 MHz 500 MHz to 2 GHz 2 GHz to 6 GHz 6 GHz to 20 GHz Directivity 42 42 38 36 Source match 37 37 31 28 Load match 42 42 38 36 Reflection tracking ±0.003 ±0.003 ±0.004 ±0.008 Transmission tracking ±0.068 ±0.034 ±0.100 ±0.208 Transmission uncertainty (specification) Phase Magnitude 100 3 0 0 K H z to 5 0 0 M H z 3 0 0 K H z to 5 0 0 M H z 5 0 0 M H z to 2 G H z 5 0 0 M H z to 2 G H z Uncertainty (Degrees) Uncertainty (dB) 10 2 G H z to 6 G H z 6 G H z to 2 0 G H z 1 0 .1 2 G H z to 6 G H z 6 G H z to 2 0 G H z 10 1 S11 = S22 = 0; Cal power = -10 dBm; Meas power = -10 dBm S11 = S22 = 0; Cal power = -10 dBm; Meas power = -10 dBm 0 .1 0 .0 1 10 0 -1 0 -2 0 -3 0 -4 0 -5 0 -6 0 -7 0 -8 0 10 -9 0 0 -1 0 -2 0 -3 0 -4 0 -5 0 -6 0 -7 0 -8 0 -9 0 T rans mis s ion C oefficient (dB ) Transmission Coefficient (dB) Reflection uncertainty (specification) Phase Magnitude 10 0 .0 5 3 0 0 K H z to 5 0 0 M H z, 5 0 0 M H z to 2 G H z 3 0 0 K H z to 5 0 0 M H z, 5 0 0 M H z to 2 G H z 2 G H z to 6 G H z 2 G H z to 6 G H z 8 Uncertainty (degrees) Uncertainty (linear) 0 .0 4 6 G H z to 2 0 G H z 0 .0 3 0 .0 2 6 G H z to 2 0 G H z S 2 1 = S 1 2 = 0 ; Cal power = -10 dBm; Meas power = -10 dBm 6 4 2 0 .0 1 S 2 1 = S 1 2 = 0 ; Cal power = -10 dBm; Meas power = -10 dBm 0 0 0 0 .2 0 .4 0 .6 0 .8 1 0 0 .2 0 .4 0 .6 Reflection coefficient (linear) Reflection coefficient (linear) 11 0 .8 1 Table 8. Corrected system performance with 3.5 mm device connectors, N4691B electronic calibration (ECal) module Network analyzer: Calibration module: E5071C N4691B (3.5 mm, 50 Ω) electronic calibration (ECal) module full 2-port Calibration: IF bandwidth = 10 Hz, no averaging applied to data, environmental temperature = 23 °C (±5 °C ) with < 1 °C deviation from calibration temperature, isolation calibration performed Specification (dB) Description 10 MHz to 500 MHz 500 MHz to 2 GHz 2 GHz to 6 GHz 6 GHz to 20 GHz Directivity 46 56 54 48 Source match 41 47 45 44 Load match 38 41 39 36 Reflection tracking ±0.050 ±0.020 ±0.030 ±0.040 Transmission tracking ±0.087 ±0.086 ±0.097 ±0.130 Transmission uncertainty (specification) Phase Magnitude 100 10 1 0 M H z to 5 0 0 M H z 1 0 M H z to 5 0 0 M H z 5 0 0 M H z to 2 G H z Uncertainty (Degrees) 5 0 0 M H z to 2 G H z Uncertainty (dB) 2 G H z to 6 G H z 6 G H z to 2 0 G H z 1 0 .1 S 11 = S 22 = 0 ; Cal power = -10 dBm; Meas power = -10dBm 2 G H z to 6 G H z 6 G H z to 2 0 G H z 10 1 S 11 = S 22 = 0 ; Cal power = -10 dBm; Meas power = -10dBm 0 .1 0 .0 1 10 0 -1 0 -2 0 -3 0 -4 0 -5 0 -6 0 -7 0 -8 0 10 -9 0 0 -1 0 -2 0 -3 0 -4 0 -5 0 -6 0 -7 0 -8 0 -9 0 Transmission Coefficient (dB) Transmission Coefficient (dB) Reflection uncertainty (specification) Phase Magnitude 10 0 .0 5 1 0 M H z to 5 0 0 M H z 1 0 M H z to 5 0 0 M H z 5 0 0 M H z to 2 G H z 5 0 0 M H z to 2 G H z 2 G H z to 6 G H z 0 .0 4 Uncertainty (degrees) 6 G H z to 2 0 G H z Uncertainty (linear) 2 G H z to 6 G H z 8 S 2 1 = S 1 2 = 0 ; Cal power = -10 dBm; Meas power = -10 dBm 0 .0 3 0 .0 2 6 G H z to 2 0 G H z S 2 1 = S 1 2 = 0 ; Cal power = -10 dBm; Meas power = -10 dBm 6 4 2 0 .0 1 0 0 0 0 .2 0 .4 0 .6 0 .8 1 0 0 .2 0 .4 0 .6 Reflection coefficient (linear) Reflection coefficient (linear) 12 0 .8 1 Uncorrected System Performance Table 9. Option 230/235/240/245/260/265/280/285/430/435/440/445/ 460/465/480/485 User correction: OFF, system correction: ON Specification (dB) Description 9 kHz to 300 kHz 300 kHz to 3 GHz 3 GHz to 6 GHz 6 GHz to 8.5 GHz Directivity 20 dB 25 dB 20 dB 15 dB Source match 20 dB 25 dB 20 dB 15 dB Load match 12 dB 17 dB 12 dB 10 dB Transmission tracking ±1.5 dB ±1.0 dB ±1.0 dB ±1.0 dB Reflection tracking ±1.5 dB ±1.0 dB ±1.0 dB ±1.0 dB Table 10. Option 2D5/2K5/4D5/4K5 User correction: OFF, system correction: ON Specification (dB) Description 300 kHz 1 MHz to 1 GHz to 3 GHz to 6 GHz to to 1 MHz 1 GHz 3 GHz 6 GHz 8.5 GHz 8.5 GHz to 11 GHz to 11 GHz 20 GHz Directivity 20 dB 25 dB 25 dB 20 dB 15 dB 15 dB 15 dB Source match 20 dB 25 dB 25 dB 20 dB 15 dB 15 dB 15 dB Load match 17 dB 15 dB 11 dB 9 dB 8 dB 7 dB Transmission ±1.0 dB tracking ±1.0 dB ±1.0 dB ±1.0 dB ±1.0 dB ±1.0 dB ±1.0 dB Reflection tracking ±1.0 dB ±1.0 dB ±1.0 dB ±1.0 dB ±1.0 dB ±1.0 dB 9 dB ±1.0 dB 13 Test Port Output (Source) Test port output frequency Table 11. Option 230/235/240/245/260/265/280/285/430/435/440/ 445/460/465/480/485/2D5/2K5/4D5/4K5 Description Specification Frequency range Option 230/430 Option 240/440 Option 260/460 Option 280/480 Option 235/435 Option 245/445 Option 265/465 Option 285/485 Option 2D5/4D5 Option 2K5/4K5 9 kHz to 3 GHz 9 kHz to 4.5 GHz 9 kHz to 6.5 GHz 9 kHz to 8.5 GHz 100 kHz to 3 GHz 100 kHz to 4.5 GHz 100 kHz to 6.5 GHz 100 kHz to 8.5 GHz 300 kHz to 14 GHz 300 kHz to 20 GHz Resolution 1 Hz Source stability standard Option 1E5 CW accuracy standard Option 1E5 Typical ±5 ppm (5 to 40 ºC) ±0.05 ppm (5 to 40 ºC), ±0.5 ppm/year ±5 ppm ±1 ppm 14 Test port output power1 Table 12. Option 230/235/240/245/260/265/280/285/430/435/440/ 445/460/465/480/485 Description Specification Nominal power (preset power) 0 dBm Level accuracy2, 6 (stepped sweep mode) ±0.650 dB (at 0 dBm, 50 MHz absolute) ±1.0 dB Level accuracy2 (swept sweep mode) Level linearity3, 5, 6 (stepped sweep mode) 9 kHz to 5 GHz 5 GHz to 6 GHz 6 GHz to 7 GHz 7 GHz to 8.5 GHz Typical ±2.5 dB ±0.75 dB (–20 to 10 dBm) ±0.75 dB (–20 to 9 dBm ) ±0.75 dB (–20 to 8 dBm ) ±0.75 dB (–20 to 7 dBm ) Level linearity5 (swept sweep mode)4 9 kHz to 5 GHz 5 GHz to 6 GHz 6 GHz to 7 GHz 7 GHz to 8.5 GHz ±1.5 dB (at –20 to 10 dBm) ±1.5 dB (at –20 to 9 dBm) ±1.5 dB (at –20 to 8 dBm) ±1.5 dB (at –20 to 7 dBm) Range5, 6 9 kHz to 5 GHz 5 GHz to 6 GHz 6 GHz to 7 GHz 7 GHz to 8.5 GHz –55 to 10 dBm –55 to 9 dBm –55 to 8 dBm –55 to 7 dBm Sweep range5, 6 9 kHz to 5 GHz 5 GHz to 6 GHz 6 GHz to 7 GHz 7 GHz to 8.5 GHz –55 to 10 dBm –55 to 9 dBm –55 to 8 dBm –55 to 7 dBm Level resolution 0.05 dB Description Specification SPD power5, 6 Max leveled (Option 230/235/240/245/260/ 265/280/285) 9 kHz to 8.5 GHz 10 dBm (Option 430/435/440/445/460/ 465/480/485) 9 kHz to 6 GHz 6 GHz to 7 GHz 7 GHz to 8.5 GHz 10 dBm 9 dBm 8 dBm 1. 2. 3. 4. 5. 6. Source output performance on port 1 only. Other port output performance is typical. Level accuracy is taken at 0 dBm, relative to 50 MHz reference unless otherwise stated. Level linearity given is relative to 0 dBm. The specification might not be met at 5 MHz or 50 MHz. The level accuracy specification needs to be taken into account for test port output power level. Power calibration using an external power meter improves level accuracy of the test port output power. Proper power meters/sensors, and the 82357B USB-GPIB interface are required to conduct power calibration. 15 Test port output power7 (continued) Table 13. Option 2D5/2K5/4D5/4K5 Description Specification Nominal power (preset power) –5 dBm Level accuracy6 (stepped sweep mode)1 300 kHz to 1 MHz 1 MHz to 5 MHz 5 MHz to 8.5 GHz 8.5 GHz to 20 GHz ±0.650 dB +2.0 dB, –6.0 dB ±2.0 dB ±1.0 dB ±2.5 dB Level accuracy (swept sweep mode)2 300 kHz to 1 GHz 1 GHz to 8.5 GHz 8.5 GHz to 20 GHz Level linearity5, 6 (stepped sweep mode)3 300 kHz to 1 MHz 1 MHz to 6 GHz 6 GHz to 8 GHz 8 GHz to 10.5 GHz 10.5 GHz to 15 GHz 15 GHz to 20 GHz ±5.0 dB ±2.5 dB +5.0 dB, –7.0 dB ±0.75 dB (–25 to 8 dBm) ±0.75 dB (–25 to 10 dBm) ±0.75 dB (–25 to 9 dBm ) ±0.75 dB (–25 to 7 dBm ) ±0.75 dB (–25 to 3 dBm ) ±0.75 dB (–25 to 0 dBm ) Level linearity5 (swept sweep mode)3 300 kHz to 1 MHz 1 MHz to 6 GHz 6 GHz to 8 GHz 8 GHz to 10.5 GHz 10.5 GHz to 15 GHz 15 GHz to 20 GHz Range5, 6 300 kHz to 1 MHz 1 MHz to 6 GHz 6 GHz to 8 GHz 8 GHz to 10.5 GHz 10.5 GHz to 15 GHz 15 GHz to 20 GHz Typical ±1.5 dB (–25 to 8 dBm) ±1.5 dB (–25 to 10 dBm) ±1.5 dB (–25 to 9 dBm ) ±1.5 dB (–25 to 7 dBm ) ±1.5 dB (–25 to 3 dBm ) ±1.5 dB (–25 to 0 dBm ) –85 to 8 dBm –85 to 10 dBm –85 to 9 dBm –85 to 7 dBm –85 to 3 dBm –85 to 0 dBm Sweep range4, 5, 6 300 kHz to 1 MHz 1 MHz to 6 GHz 6 GHz to 8 GHz 8 GHz to 10.5 GHz 10.5 GHz to 15 GHz 15 GHz to 20 GHz (Source attenuator = 0 dB) –25 to 8 dBm –25 to 10 dBm –25 to 9 dBm –25 to 7 dBm –25 to 3 dBm –25 to 0 dBm Level resolution 0.05 dB 1. 2. 3. 4. 5. 6. Level accuracy is taken at –5 dBm, relative to 50 MHz reference unless otherwise stated. Level accuracy is taken at –5 dBm, relative to 50 MHz reference. Level linearity given is relative to –5 dBm. The sweep range shifts based on the selected source attenuator value (0 dB to 60 dB, 10 dB step). The level accuracy specification needs to be taken into account for test port output power level. Power calibration using an external power meter improves level accuracy of the test port output power. Proper power meters/sensors, and the 82357B USB-GPIB interface are required to conduct power calibration. 7. Source output performance on port 1 only. Other port output performance is typical. 16 Test port output power6 (continued) Table 13. Option 2D5/2K5/4D5/4K5 Description Specification SPD power3, 4 Max leveled 300 kHz to 1 MHz 1 MHz to 10 GHz 10 GHz to 13 GHz 13 GHz to 15 GHz 15 GHz to 18 GHz 18 GHz to 20 GHz 9 dBm 10 dBm 9 dBm 7 dBm 5 dBm 4 dBm Test port output signal purity Table 14. Option 230/235/240/245/260/265/280/285/430/435/440/ 445/460/465/480/485 Description Specification Typical Harmonics (2nd or 3rd) 9 kHz to 2 GHz 2 GHz to 8.5 GHz < –25 dBc (at 5 dBm) < –20 dBc (at 5 dBm) Non-harmonic spurious 9 kHz to 8.5 GHz < –30 dBc (at 5 dBm) Table 15. Option 2D5/2K5/4D5/4K5 Description Harmonics (2nd to 300 kHz to 1 GHz Specification Typical 5th)1 1 GHz to 20 GHz Non-harmonic spurious2 300 kHz to 20 GHz < –12 dBc (at maximum output power)5 < –15 dBc (at maximum output power)5 < –30 dBc (at –5 dBm) 1. 2. 3. 4. This includes 6th and 7th harmonics when the test frequency is from 1 MHz to 1 GHz. The carrier ±1/8th spurious is excluded from 8.76 GHz to 17.52 GHz. The level accuracy specification needs to be taken into account for test port output power level. Power calibration using an external power meter improves level accuracy of the test port output power. Proper power meters/sensors, and the 82357B USB-GPIB interface are required to conduct power calibration. 5. Maximum output power is the maximum power of "Range" in Table 13 Test port output power. 6. Source output performance on port 1 only. Other port output performance is typical. 17 Test Port Input Test port input levels Table 16. Option 230/235/240/245/260/265/280/285/430/435/440/ 445/460/465/480/485 Description Specification Typical Maximum test port input level 9 kHz to 5 GHz 5 GHz to 6 GHz 6 GHz to 7 GHz 7 GHz to 8.5 GHz +10 dBm +9 dBm +8 dBm +7 dBm Damage level 9 kHz to 8.5 GHz +26 dBm ±35 VDC Crosstalk1 9 kHz to 300 kHz 300 kHz to 10 MHz 10 MHz to 3 GHz 3 GHz to 6 GHz 6 GHz to 8.5 GHz –100 dB –110 dB –120 dB –110 dB –100 dB Description Specification SPD Test port noise floor 9 kHz to 10 MHz 10 MHz to 5 GHz 5 GHz to 6 GHz 6 GHz to 7 GHz 7 GHz to 8 GHz 8 GHz to 8.5 GHz –115 dBm/Hz –130 dBm/Hz –131 dBm/Hz –130 dBm/Hz –129 dBm/Hz –127 dBm/Hz 1. The specification might not be met at 5 MHz or 50 MHz. 18 Test port input levels (continued) Table 17. Option 2D5/2K5/4D5/4K5 Description Specification Typical Maximum test port input level 300 kHz to 20 GHz +10 dBm Damage level 300 kHz to 20 GHz +26 dBm or ±35 VDC Crosstalk1 300 kHz to 1 MHz 1 MHz to 5 MHz 5 MHz to 10 MHz 10 MHz to 45 MHz 45 MHz to 4 GHz 4 GHz to 6 GHz 6 GHz to 8.5 GHz 8.5 GHz to 15 GHz 15 GHz to 20 GHz –68 dB –70 dB –100 dB –110 dB –118 dB –123 dB –120 dB –112 dB –106 dB Description Specification SPD Test port noise floor 300 kHz to 1 MHz 1 MHz to 10 MHz 10 MHz to 100 MHz 100 MHz to 8.5 GHz 8.5 GHz to 15 GHz 15 GHz to 20 GHz –110 dBm/Hz –115 dBm/Hz –129 dBm/Hz –130 dBm/Hz –118 dBm/Hz –115 dBm/Hz 1. The specification might not be met at 5 MHz or 50 MHz. 19 Test port input (trace noise) Table 18. Option 230/235/240/245/260/265/280/285/430/435/440/ 445/460/465/480/485 Description Specification SPD Compression level Magnitude 9 kHz to 5 GHz 5 GHz to 8.5 GHz (maximum test port input level = +10 dBm) 0.08 dB 0.1 dB Phase 9 kHz to 3 GHz 3 GHz to 5 GHz 5 GHz to 8.5 GHz (maximum test port input level = +10 dBm) 0.3 deg 0.6 deg 1.0 deg Trace noise magnitude1 9 kHz to 30 kHz (IFBW = 3 kHz) (maximum test port input level = +10 dBm) 0.004 dBrms 0.001 dBm 30 kHz to 100 kHz (IFBW = 3 kHz) (maximum test port input level = +10 dBm) 0.003 dBm 0.001 dBm 100 kHz to 10 MHz (IFBW = 3 kHz) (maximum test port input level = +10 dBm) 0.003 dBrms 0.0005 dBm 10 MHz to 4.38 GHz (IFBW = 70 kHz) (maximum test port input level = +10 dBm) 0.004 dBrms 0.001 dBrms 4.38 GHz to 5 GHz (IFBW = 70 kHz) (maximum test port input level = +10 dBm) 0.006 dBrms 0.0012 dBm 5 GHz to 6 GHz (IFBW = 70 kHz) (maximum test port input level = +9 dBm) 0.006 dBrms 0.0012 dBm 6 GHz to 7 GHz (IFBW = 70 kHz) (maximum test port input level = +8 dBm) 0.006 dBrms 0.0012 dBm 7 GHz to 8.5 GHz (IFBW = 70 kHz) (maximum test port input level = +7 dBm) 0.006 dBrms 0.0012 dBm Trace noise phase2 9 kHz to 30 kHz (IFBW = 3 kHz) (maximum test port input level = +10 dBm) 0.035 deg rms 30 kHz to 10 MHz (IFBW = 3 kHz) (maximum test port input level = +10 dBm) 0.020 deg rms 10 MHz to 4.38 GHz (IFBW = 70 kHz) (maximum test port input level = +10 dBm) 0.035 deg rms 4.38 GHz to 5 GHz (IFBW = 70 kHz) (maximum test port input level = +10 dBm) 0.050 deg rms 5 GHz to 6 GHz (IFBW = 70 kHz) (maximum test port input level = +9 dBm) 0.050 deg rms 6 GHz to 7 GHz (IFBW = 70 kHz) (maximum test port input level = +8 dBm) 0.050 deg rms 7 GHz to 8.5 GHz (IFBW = 70 kHz) (maximum test port input level = +7 dBm) 0.050 deg rms 1. The specification might not be met at the following frequencies: 333.333 kHz, 406.25 kHz, 857.143 kHz, 928.571 kHz, 1.3 MHz, 2.4 MHz and 4.333333 MHz. 2. The specification might not be met at 5 MHz or 50 MHz. 20 Test port input (trace noise) (continued) Table 19. Option 2D5/2K5/4D5/4K5 Description Specification SPD Compression level Magnitude 300 kHz to 10 MHz 0.3 dB 10 MHz to 10 GHz 0.182 dB 10 GHz to 15 GHz 0.712 dB 15 GHz to 20 GHz 1.87 dB (maximum test port input level = +10 dBm) Phase 300 kHz to 5 GHz 2.3 deg 5 GHz to 10 GHz 4.3 deg 10 GHz to 15 GHz 17.3 deg 15 GHz to 20 GHz 20.3 deg (maximum test port input level = +10 dBm) Trace noise magnitude1, 3 300 kHz to 1 MHz (IFBW = 3 kHz) 0.006 dBrms 0.0009 dBrms 1 MHz to 10 MHz (IFBW = 3 kHz) 0.003 dBrms 0.0005 dBrms 10 MHz to 4.38 GHz (IFBW = 70 kHz) 0.004 dBrms 0.0010 dBrms 4.38 GHz to 8.5 GHz (IFBW = 70 kHz) 0.006 dBrms 0.0012 dBrms 8.5 GHz to 13.137 GHz (IFBW = 70 kHz) 0.009 dBrms 0.0024 dBrms 13.137 GHz to 17 GHz (IFBW = 70 kHz) 0.013 dBrms 0.0040 dBrms 17 GHz to 20 GHz (IFBW = 70 kHz) 0.023 dBrms 0.0065 dBrms 300 kHz to 1 MHz (IFBW = 3 kHz) 0.040 deg rms 0.0120 deg rms 1 MHz to 10 MHz (IFBW = 3 kHz) 0.020 deg rms 0.0025 deg rms 10 MHz to 4.38 GHz (IFBW = 70 kHz) 0.035 deg rms 0.0075 deg rms 4.38 GHz to 8.5 GHz (IFBW = 70 kHz) 0.050 deg rms 0.0150 deg rms 8.5 GHz to 13.137 GHz (IFBW = 70 kHz) 0.064 deg rms 0.0250 deg rms 13.137 GHz to 17 GHz (IFBW = 70 kHz) 0.095 deg rms 0.0320 deg rms 17 GHz to 20 GHz (IFBW = 70 kHz) 0.165 deg rms 0.0520 deg rms (at maximum output power level of sweep range) Trace noise phase2, 3 (at maximum output power level of sweep range) 1. The specification might not be met at the following frequencies: 406.25 kHz, 666.667 kHz, 722.222 kHz, 857.143 kHz, 928.571 kHz, 1.444444 MHz, 1.714286 MHz, 1.8 MHz, 1.857143 MHz, 1.95 MHz, 2.4375 MHz, 2.571429 MHz, 3.714286 MHz, 4.8 MHz, 5 MHz, 5.416667 MHz, 7.583333 MHz, 10 MHz, 10.833333 MHz, 12.5 MHz. 2. The specification might not be met at 5 MHz or 50 MHz. 3. Trace noise SPD is defined with transmission measurements only. 21 Test port input (stability)1 Table 20. Option 230/235/240/245/260/265/280/285/430/435/440/ 445/460/465/480/485 Description Specification Typical Stability magnitude 9 kHz to 3 GHz ±0.005 dB/°C 3 GHz to 6 GHz ±0.01 dB/°C 6 GHz to 8.5 GHz ±0.04 dB/°C Stability phase 9 kHz to 3 GHz ±0.1 deg/°C 3 GHz to 6 GHz ±0.2 deg/°C 6 GHz to 8.5 GHz ±0.8 deg/°C Table 21. Option 2D5/2K5/4D5/4K5 Description Specification Typical Stability magnitude 300 kHz to 3 GHz ±0.005 dB/°C 3 GHz to 6 GHz ±0.01 dB/°C 6 GHz to 20 GHz ±0.04 dB/°C Stability phase 300 kHz to 3 GHz 3 GHz to 6 GHz 6 GHz to 20 GHz ±0.1 deg/°C ±0.2 deg/°C ±0.8 deg/°C 1. Stability is defined as a ratio measurement at the test port. 22 Test port input (dynamic accuracy) Table 22. Option 230/235/240/245/260/265/280/285/430/435/440/ 445/460/465/480/485 Accuracy of the test port input power reading is relative to –10 dBm reference input power level. Description Specification Typical Dynamic accuracy magnitude 10 dBm ±0.21 dB –30 dBm ±0.05 dB –100 dBm ±2.01 dB Dynamic accuracy phase 10 dBm ±5 deg –30 dBm ±0.3 deg –100 dBm ±15.1 deg Specification Typical Magnitude Phase 100 Accuracy [degrees] Accuracy [dB] 10 1 0.1 0.01 10 0 -10 -20 -30 -40 -50 -60 -70 -80 -90 10 1 0.1 -100 Test port power [dBm] 10 0 -10 -20 -30 -40 -50 -60 Test port power [dBm] ±3.0 dB (at –110 dBm, Ref.= –10 dBm, typical) 23 -70 -80 -90 -100 Test port input (dynamic accuracy) (continued) Table 23. Option 2D5/2K5/4D5/4K5 Accuracy of the test port input power reading is relative to –10 dBm reference input power level. Description Specification Typical Dynamic accuracy magnitude 10 dBm ±2.4972 dB –30 dBm ±0.05 dB –100 dBm ±2.01 dB Dynamic accuracy phase 10 dBm ±20.66 deg –30 dBm ±0.3 deg –100 dBm ±15.1 deg Specification Typical Phase Magnitude 100 Accuracy [degrees] Accuracy [dB] 10 1 0.1 0.01 10 0 10 1 0.1 -10 -20 -30 -40 -50 -60 -70 -80 -90 -100 10 0 -10 -20 -30 -40 -50 -60 -70 -80 -90 -100 Test port power [dBm] Test port power [dBm] ±3.0 dB (at –110 dBm, Ref.= –10 dBm, typical) 24 Test port input (group delay)1 Table 24. Option 230/235/240/245/260/265/280/285/430/435/440/ 445/460/465/480/485 Description Specification Aperture (selectable) (frequency span)/(number of points - 1) Supplemental information Maximum aperture 25% of frequency span Minimum delay Limited to measuring no more than 180 °C of phase change within the minimum aperture. Accuracy See graph below, typical The following graph shows group delay accuracy with type-N connectors, full 2-port calibration and a 10 Hz IF bandwidth. Calibration kit (85032F). Insertion loss is assumed to be < 2 dB. 100 Accuracy (nSec) 10 1 0.1 0.01 0.001 0.01 0.1 1 10 100 Aperture (MHz) In general, the following formula can be used to determine the accuracy, in seconds, of a specific group delay measurement: ±phase accuracy (degrees) / [360 x aperture (Hz)] 1. Group delay is computed by measuring the phase change within a specified step (determined by the frequency span and the number of points per sweep). 25 Test port input (group delay)1 (continued) Table 25. Option 2D5/2K5/4D5/4K5 Description Specification Aperture (selectable) (frequency span)/(number of points - 1) Supplemental information Maximum aperture 25% of frequency span Minimum delay Limited to measuring no more than 180 °C of phase change within the minimum aperture. Accuracy See graph below, typical The following graph shows group delay accuracy with 3.5 mm (male) connectors, full 2-port calibration and a 10 Hz IF bandwidth. Calibration kit (85052D). Insertion loss is assumed to be < 2 dB. 100 Accuracy (nSec) 10 1 0 .1 0 .0 1 Frequency = 1 G H z C al pow er = -10 dBm ; M eas pow er = -10 dBm ; Electrical Length = 10 m 0 .0 0 1 0 .0 1 0 .1 1 10 100 Aperture (MHz) In general, the following formula can be used to determine the accuracy, in seconds, of a specific group delay measurement: ±phase accuracy (degrees) / [360 x aperture (Hz)] 1. Group delay is computed by measuring the phase change within a specified step (determined by the frequency span and the number of points per sweep). 26 General Information Table 26. System bandwidth Description General characteristics IF bandwidth settings Range 10 Hz to 500 kHz Nominal settings are: 10, 15, 20, 30, 40, 50, 70, 100, 150, 200, 300, 400, 500, 700, 1 k, 1.5 k, 2 k, 3 k, 4 k, 5 k, 7 k, 10 k, 15 k, 20 k, 30 k, 40 k, 50 k, 70 k, 100 k, 150 k, 200 k, 300 k, 400 k, 500 kHz Table 27. Front panel information Description Typical General characteristics RF connectors Type-N, female, 50 Ω Option 230/235/240/ 245/260/265/280/285/ 430/435/440/445/460/ 465/480/485 Option 2D5/2K5/4D5/4K5 3.5 mm (male), 50 Ω nominal Probe power Connector Voltage & maximum current 3 terminal connector x 2 +15 V ±2% (400 mA) –12.6 V ±5% (300 mA) (Combined load for both probe connections) Display Type 10.4 in TFT color LCD with touch screen Resolution XGA (1024 x 768)1 1. Valid pixels are 99.99 % and more. Below 0.01 % (approx. 30 points) of fixed points of black, blue, green or red are not regarded as failure. 27 Table 28. Rear panel information Description Typical General characteristics External trigger input connector Type BNC, female Input level Low threshold voltage: 0.5 V High threshold voltage: 2.1 V Input level range: 0 to + 5 V Pulse width ≥ 2 μsec Polarity Positive or negative External trigger output connector Type BNC, female Maximum output current 50 mA Output level Low level voltage: 0 V High level voltage: 5 V Pulse width 1 μsec Polarity Positive or negative External reference signal input connector Type BNC, female Input frequency 10 MHz ±10 ppm Input level –3 to + 10 dBm Internal reference signal output connector Type BNC, female Output frequency 10 MHz ±5 ppm Signal type Sinewave Output level 0 dBm ±3 dB into 50 Ω Output impedance 50 Ω Internal reference signal oven connector Type BNC, female Output frequency 10 MHz ±1 ppm Output level 0 dBm minimum 28 Table 28. Rear panel information (continued) Description Typical General characteristics Bias tee input connector Type BNC, female (for each port) Maximum voltage ±35 VDC Maximum current (no degradation in RF specifications) Option 230/235/240/245/260/265/280/ 285/430/435/440/445/460/465/480/485 ±50 mA (1 MHz < test frequency < 500 MHz) ±200 mA (test frequency ≤ 1 MHz, ≥ 500 MHz) Option 2D5/2K5/4D5/4K5 ±200 mA Maximum current (damage level) ±500 mA Fuse 500 mA, bi-pin style Video output 15-pin mini D-Sub; female; drives XGA compatible monitors GPIB 24-pin D-Sub (Type D-24), female; compatible with IEEE-488 USB-host port Universal serial bus jack, Type A configuration (4 contacts inline, contact 1 on left); female; provides connection to printer, ECal module, USB/GPIB interface or multiport test set Contact 1 Vcc: 4.75 to 5.25 VDC, 500 mA, maximum Contact 2 – Data Contact 3 + Data Contact 4 Ground 29 Table 28. Rear panel information (continued) Description Typical General characteristics USB (USBTMC1) interface port Universal serial bus jack, Type B configuration (4 contacts inline); female; provides connection to an external PC; compatible with USBTMC-USB488 and USB 2.0. LAN 10/100BaseT Ethernet, 8-pin configuration; auto selects between the two data rates Handler I/O port 36-pin centronics, female; provides connection to handler system Line power2 Frequency 47 Hz to 63 Hz Voltage 90 to 132 VAC, or 198 to 264 VAC (automatically switched) VA max 350 VA max. Description Specification General characteristics AUX input connector Type BNC, female x 2 Input range Accuracy ±1 V or ±10 V selectable 1 % + 1 mV for ±1 V input 1 % + 10 mV for ±10 V input Table 29. LXI compliance Description General characteristics Class C (only applies to units that are shipped with firmware revision A.08.00 or later) 1. USB Test and Measurement Class (TMC) interface that communicates over USB, complying with the IEEE 488.1 and IEEE 488.2 standards. 2. A third-wire ground is required. 30 Table 30. EMC, safety and environment Description General characteristics EMC European Council Directive 2004/108/EC IEC 61326-1:2005 EN 61326-1:2006 CISPR 11:2003+A1:2004 EN 55011:2007 Group 1, Class A IEC 61000-4-2:1995 +A2:2000 EN 61000-4-2:1995 +A2:2001 4 kV CD / 8 kV AD IEC 61000-4-3:2006 EN 61000-4-3:2006 1-3 V/m, 80-1000 MHz / 1.4 GHz - 2.7 GHz, 80% AM IEC 61000-4-4:2004 EN 61000-4-4:2004 1 kV power lines / 0.5 kV signal lines IEC 61000-4-5:2005 EN 61000-4-5:2006 0.5 kV line-line / 1 kV line-ground IEC 61000-4-6:2003 + A1:2004+ A2:2006 EN 61000-4-6:2007 3 V, 0.15-80 MHz, 80% AM IEC 61000-4-11:2004 EN 61000-4-11:2004 0.5-300 cycle, 0% / 70% ICES-001:2006 Group 1, Class A AS/NZS CISPR11:2004 Group 1, Class A Safety European Council Directive 2006/95/EC IEC 61010-1:2001 / EN 61010-1:2001 Measurement Category I Pollution Degree 2 Indoor Use CAN/CSA C22.2 No. 61010-1-04 Measurement Category I Pollution Degree 2 Indoor Use Environment This product complies with the WEEE Directive (2002/96/EC) marking requirements. The affixed label indicates that you must not discard this electrical/electronic product in domestic household waste. Product Category: With reference to the equipment types in the WEEE Directive Annex I, this product is classed as a “Monitoring and Control instrumentation” product. Do not dispose in domestic household waste. To return unwanted products, contact your local Agilent office, or see www.agilent.com/environment/product/ for more information. 31 Table 31. Analyzer environmental specifications and dimensions Description General characteristics Operating environment Temperature +5 °C to +40 °C Error-corrected temperature range 23 °C (±5 °C) with < 1 °C deviation from calibration temperature Humidity 20% to 80% at wet bulb temperature < +29 °C (non-condensation) Altitude 0 to 2,000 m (0 to 6561 feet) Vibration 0.21 G maximum, 5 Hz to 500 Hz Non-Operating storage environment Temperature –10 °C to +60 °C Humidity 20% to 90% at wet bulb temperature < +40 °C (non-condensation) Altitude 0 to 4572 m (0 to 15,000 feet) Vibration 0.5 G maximum, 5 Hz to 500 Hz Dimensions See next page. Weight (net) Option 230/240/260/280 (2-port) Option 235/245/265/285 (2-port) Option 430/480/440/460 (4-port) Option 435/445/465/485 (4-port) Option 2D5/2K5 (2-port) Option 4D5/4K5 (4-port) 18.2 kg 18.3 kg 19.9 kg 20.0 kg 19.8 kg 21.8 kg 32 22 19 22 222 426 27 58 20 51 122 37 Figure 3. Dimensions (front view, E5071C with Option 230/235/240/245/260/265/280/285, in millimeters) 22 19 22 222 426 27 58 20 51 41 41 41 Figure 4. Dimensions (front view, E5071C with Option 430/435/440/445/460/465/480/485, in millimeters) 33 37 26 19 22 222 426 27 58 20 51 122 Figure 5. Dimensions (front view, E5071C with Option 2D5/2K5, in millimeters) Figure 6. Dimensions (front view, E5071C with Option 4D5/4K5, in millimeters) 34 37 424 422 128 1 96 21 18 27 25 25 290 25 58 109 148 25 144 146 159 179 190 215 1 25 25 25 165 Figure 7. Dimensions (rear view with Option 1E5, in millimeters) 472 23.4 417 13 18 235 215 222 32 15 Figure 8. Dimensions (side view, E5071C Option 230/235/240/245/260/265/280/285/430/435/440/445/460/465/480/485, in millimeters) 472 417 23 13 18 215 222 235 32 14 Figure 9. Dimensions (side view, E5071C with Option 2D5/2K5/4D5/4K5, in millimeters) 35 Measurement Throughput Summary Cycle time for measurement completion1, 2 (ms) Table 32. Option 230/235/240/245/260/265/280/285/430/435/440/445/ 460/465/480/485 Sweep mode: Swept, analyzer display turned off with: DISP:ENAB OFF, number of traces = 1, system error correction: OFF Number of points 51 201 401 1601 Start 1 GHz, stop 1.2 GHz, 500 kHz IF bandwidth Uncorrected 3.4 4.9 7.1 20 2-port cal 6.3 9.3 14 41 4-port cal 13 19 28 92 Start 1 GHz, stop 1.2 GHz, 100 kHz IF bandwidth Uncorrected 3.7 5.2 7.3 20 2-port cal 6.6 10 15 42 4-port cal 13 21 32 94 Start 1 GHz, stop 1.2 GHz, 1 kHz IF bandwidth Uncorrected 54 202 397 1563 2-port cal 107 401 792 3125 4-port cal 213 800 1581 6247 Start 100 kHz, stop 3 GHz, 500 kHz IF bandwidth Uncorrected 8.7 11 12 25 2-port cal 17 21 24 48 4-port cal 34 42 49 102 Start 100 kHz, stop 3 GHz, 100 kHz IF bandwidth Uncorrected 8.8 11 13 26 2-port cal 17 22 24 54 4-port cal 34 42 49 118 Start 100 kHz, stop 3 GHz, 1 kHz IF bandwidth Uncorrected 56 206 404 1583 2-port cal 112 410 805 3163 4-port cal 222 818 1608 6324 1. Supplemental performance data. 2. Measured with firmware version A.09.30. 36 Table 32. Option 230/235/240/245/260/265/280/285/430/435/440/445/ 460/465/480/485 (continued) Number of points 51 201 401 1601 Start 100 kHz, stop 8.5 GHz, 500 kHz IF bandwidth Uncorrected 15 18 19 24 2-port cal 28 35 38 48 4-port cal 55 69 74 104 Start 100 kHz, stop 8.5 GHz, 100 kHz IF bandwidth Uncorrected 15 18 20 26 2-port cal 28 36 38 52 4-port cal 56 70 75 117 Start 100 kHz, stop 8.5 GHz, 1 kHz IF bandwidth Uncorrected 57 207 404 1583 2-port cal 114 412 806 3163 4-port cal 225 821 1611 6324 37 Cycle time for measurement completion1, 2 (ms) Table 33. Option 230/235/240/245/260/265/280/285/430/435/440/445/ 460/465/480/485 Sweep mode: Swept, analyzer display turned off with: DISP:ENAB OFF, number of traces = 1, system error correction: ON Number of Points 51 201 401 1601 Start 1 GHz, stop 1.2 GHz, 500 kHz IF bandwidth Uncorrected 3.5 5.5 8.0 24 2-port cal 6.4 9.8 15 49 4-port cal 13 20 29 108 Start 1 GHz, stop 1.2 GHz, 100 kHz IF bandwidth Uncorrected 3.6 5.4 8.0 23 2-port cal 6.6 11 16 48 4-port cal 13 22 33 107 Start 1 GHz, stop 1.2 GHz, 1 kHz IF bandwidth Uncorrected 54 202 397 1563 2-port cal 107 401 792 3125 4-port cal 213 800 1581 6247 Start 100 kHz, stop 3 GHz, 500 kHz IF bandwidth Uncorrected 8.7 11 13 28 2-port cal 18 21 24 53 4-port cal 34 42 49 112 Start 100 kHz, stop 3 GHz, 100 kHz IF bandwidth Uncorrected 8.8 11 13 28 2-port cal 17 22 25 57 4-port cal 34 42 50 122 Start 100 kHz, stop 3 GHz, 1 kHz IF bandwidth Uncorrected 56 206 404 1583 2-port cal 112 410 805 3163 4-port cal 222 818 1608 6324 1. Supplemental performance data. 2. Measured with firmware version A.09.30. 38 Table 33. Option 230/235/240/245/260/265/280/285/430/435/440/445/ 460/465/480/485 (continued) Number of points 51 201 401 1601 Start 100 kHz, stop 8.5 GHz, 500 kHz IF bandwidth Uncorrected 15 18 20 27 2-port cal 28 35 38 52 4-port cal 55 69 74 111 Start 100 kHz, stop 8.5 GHz, 100 kHz IF bandwidth Uncorrected 15 18 20 27 2-port cal 28 36 38 56 4-port cal 56 70 75 120 Start 100 kHz, stop 8.5 GHz, 1 kHz IF bandwidth Uncorrected 57 207 404 1583 2-port cal 114 412 806 3163 4-port cal 225 821 1611 6324 39 Cycle time for measurement completion1, 2 (ms) Table 34. Option 230/235/240/245/260/265/280/285/430/435/440/445/ 460/465/480/485 Sweep mode: Stepped, analyzer display turned off with: DISP:ENAB OFF, number of traces = 1, system error correction: ON Number of Points 51 201 401 1601 Start 1 GHz, stop 1.2 GHz, 500 kHz IF bandwidth Uncorrected 4.4 7.7 12 29 2-port cal 8.2 15 22 56 4-port cal 16 29 45 126 Start 1 GHz, stop 1.2 GHz, 100 kHz IF bandwidth Uncorrected 4.8 9.2 15 41 2-port cal 9.0 18 28 80 4-port cal 18 35 56 163 Start 1 GHz, stop 1.2 GHz, 1 kHz IF bandwidth Uncorrected 54 202 397 1563 2-port cal 107 401 792 3125 4-port cal 213 800 1581 6247 Start 100 kHz, stop 3 GHz, 500 kHz IF bandwidth Uncorrected 6.7 13 18 48 2-port cal 13 24 36 94 4-port cal 26 47 70 190 Start 100 kHz, stop 3 GHz, 100 kHz IF bandwidth Uncorrected 7.1 14 22 60 2-port cal 14 27 42 119 4-port cal 27 53 82 237 Start 100 kHz, stop 3 GHz, 1 kHz IF bandwidth Uncorrected 56 206 404 1583 2-port cal 112 410 805 3163 4-port cal 222 818 1608 6324 1. Supplemental performance data. 2. Measured with firmware version A.09.30. 40 Table 34. Option 230/235/240/245/260/265/280/285/430/435/440/445/ 460/465/480/485 (continued) Number of points 51 201 401 1601 Start 100 kHz, stop 8.5 GHz, 500 kHz IF bandwidth Uncorrected 7.5 13 19 48 2-port cal 15 26 37 94 4-port cal 29 50 73 189 Start 100 kHz, stop 8.5 GHz, 100 kHz IF bandwidth Uncorrected 7.9 15 22 60 2-port cal 16 29 43 120 4-port cal 30 56 85 237 Start 100 kHz, stop 8.5 GHz, 1 kHz IF bandwidth Uncorrected 57 207 404 1583 2-port cal 114 412 806 3163 4-port cal 225 821 1611 6324 41 Cycle time for measurement completion1, 2 (ms) Table 35. Option 2D5/2K5/4D5/4K5 Sweep mode: Swept, analyzer display turned off with: DISP:ENAB OFF, number of traces = 1, system error correction: OFF Number of points 51 201 401 1601 Start 300 kHz, stop 20 GHz, 500 kHz IF bandwidth Uncorrected 23 36 39 43 2-port cal 45 71 77 85 4-port cal 88 142 155 169 Start 300 kHz, stop 20 GHz, 100 kHz IF bandwidth Uncorrected 23 37 40 43 2-port cal 45 72 78 86 4-port cal 89 144 156 171 Start 300 kHz, stop 20 GHz, 1 kHz IF bandwidth Uncorrected 60 212 410 1591 2-port cal 120 421 817 3180 4-port cal 237 840 1632 6358 Start 8 GHz, stop 18 GHz, 500 kHz IF bandwidth Uncorrected 17 22 22 23 2-port cal 34 42 44 45 4-port cal 67 83 86 102 Start 8 GHz, stop 18 GHz, 100 kHz IF bandwidth Uncorrected 18 22 23 24 2-port cal 34 42 44 47 4-port cal 67 84 87 106 Start 8 GHz, stop 18 GHz, 1 kHz IF bandwidth Uncorrected 58 207 405 1583 2-port cal 115 413 807 3163 4-port cal 229 823 1612 6324 1. Supplemental performance data. 2. Measured with firmware version A.09.30. 42 Table 35. Option 2D5/2K5/4D5/4K5 (continued) Number of points 51 201 401 1601 Start 11 GHz, stop 12 GHz, 500 kHz IF bandwidth Uncorrected 3.0 3.2 4.9 17 2-port cal 5.6 6.1 9.9 37 4-port cal 11 14 23 84 Start 11 GHz, stop 12 GHz, 100 kHz IF bandwidth Uncorrected 3.0 3.4 5.5 18 2-port cal 5.6 6.4 11 38 4-port cal 11 14 24 85 Start 11 GHz, stop 12 GHz, 1 kHz IF bandwidth Uncorrected 52 200 396 1566 2-port cal 104 399 790 3130 4-port cal 206 795 1578 6258 43 Cycle time for measurement completion1, 2 (ms) Table 36. Option 2D5/2K5/4D5/4K5 Sweep mode: Swept, analyzer display turned off with: DISP:ENAB OFF, number of traces = 1, system error correction: ON Number of points 51 201 401 1601 Start 300 kHz, stop 20 GHz, 500 kHz IF bandwidth Uncorrected 23 36 39 43 2-port cal 45 71 77 85 4-port cal 88 142 155 170 Start 300 kHz, stop 20 GHz, 100 kHz IF bandwidth Uncorrected 23 37 40 43 2-port cal 45 72 78 86 4-port cal 89 144 156 171 Start 300 kHz, stop 20 GHz, 1 kHz IF bandwidth Uncorrected 60 212 410 1591 2-port cal 120 421 817 3180 4-port cal 237 840 1632 6358 Start 8 GHz, stop 18 GHz, 500 kHz IF bandwidth Uncorrected 17 22 22 23 2-port cal 34 42 44 48 4-port cal 67 83 86 107 Start 8 GHz, stop 18 GHz, 100 kHz IF bandwidth Uncorrected 18 22 23 24 2-port cal 34 42 44 49 4-port cal 67 83 87 110 Start 8 GHz, stop 18 GHz, 1 kHz IF bandwidth Uncorrected 58 207 405 1583 2-port cal 115 413 807 3163 4-port cal 229 823 1612 6324 1. Supplemental performance data. 2. Measured with firmware version A.09.30. 44 Table 36. Option 2D5/2K5/4D5/4K5 (continued) Number of points 51 201 401 1601 Start 11 GHz, stop 12 GHz, 500 kHz IF bandwidth Uncorrected 3.0 3.4 5.9 20 2-port cal 5.6 6.6 12 43 4-port cal 11 14 26 98 Start 11 GHz, stop 12 GHz, 100 kHz IF bandwidth Uncorrected 3.0 3.5 5.8 20 2-port cal 5.6 6.8 12 43 4-port cal 11 15 26 98 Start 11 GHz, stop 12 GHz, 1 kHz IF bandwidth Uncorrected 52 200 396 1566 2-port cal 104 399 790 3130 4-port cal 206 795 1578 6258 45 Cycle time for measurement completion1, 2 (ms) Table 37. Option 2D5/2K5/4D5/4K5 Sweep mode: Stepped, analyzer display turned off with: DISP:ENAB OFF, number of traces = 1, system error correction: ON Number of points 51 201 401 1601 Start 300 kHz, stop 20 GHz, 500 kHz IF bandwidth Uncorrected 11 18 24 56 2-port cal 21 35 48 112 4-port cal 41 69 94 222 Start 300 kHz, stop 20 GHz, 100 kHz IF bandwidth Uncorrected 11 20 28 69 2-port cal 22 38 54 137 4-port cal 42 75 107 272 Start 300 kHz, stop 20 GHz, 1 kHz IF bandwidth Uncorrected 60 212 410 1591 2-port cal 120 421 817 3180 4-port cal 237 840 1632 6358 Start 8 GHz, stop 18 GHz, 500 kHz IF bandwidth Uncorrected 8.4 14 19 48 2-port cal 17 27 38 94 4-port cal 32 52 74 188 Start 8 GHz, stop 18 GHz, 100 kHz IF bandwidth Uncorrected 8.8 15 22 60 2-port cal 17 30 44 120 4-port cal 34 58 86 237 Start 8 GHz, stop 18 GHz, 1 kHz IF bandwidth Uncorrected 58 207 405 1583 2-port cal 115 413 807 3163 4-port cal 229 823 1612 6324 1. Supplemental performance data. 2. Measured with firmware version A.09.30. 46 Table 37. Option 2D5/2K5/4D5/4K5 (continued) Number of points 51 201 401 1601 Start 11 GHz, stop 12 GHz, 500 kHz IF bandwidth Uncorrected 2.8 6.4 11 31 2-port cal 5.1 13 21 61 4-port cal 9.6 24 41 131 Start 11 GHz, stop 12 GHz, 100 kHz IF bandwidth Uncorrected 3.4 8.0 14 43 2-port cal 5.9 16 27 86 4-port cal 12 30 53 171 Start 11 GHz, stop 12 GHz, 1 kHz IF bandwidth Uncorrected 52 200 396 1566 2-port cal 104 399 790 3130 4-port cal 206 795 1578 6258 47 Cycle time (ms)1, 2 vs. number of points Table 38. Option 230/235/240/245/260/265/280/285/430/435/440/445/ 460/465/480/485 Start 1 GHz, stop 1.2 GHz, 500 kHz IF bandwidth, error correction: OFF, display update: OFF, number of traces = 1 Number of points Sweep mode: Fast swept, system error correction: OFF Sweep mode: Fast swept, system error correction: ON Sweep mode: Std stepped, system error correction: ON 3 3.1 3.1 3.4 11 3.3 3.3 3.3 51 3.5 3.5 4.4 101 3.9 4.1 5.6 201 4.9 5.5 7.7 401 7.1 8.0 12 801 12 14 18 1601 20 24 29 Cycle time (ms)1, 2 vs. number of points Table 39. Option 2D5/2K5/4D5/4K5 Start 11 GHz, stop 12 GHz, 500 kHz IF bandwidth, error correction: OFF, display update: OFF, number of traces = 1 Number of points Sweep mode: Fast swept, system error correction: OFF Sweep mode: Fast swept, system error correction: ON Sweep mode: Std stepped, system error correction: ON 51 3.0 3.0 2.8 201 3.2 3.4 6.5 401 4.9 5.9 11 1601 17 20 31 1. Supplemental performance data. 2. Measured with firmware version A.09.30. 48 Cycle time (ms)1, 2 vs. IF bandwidth Table 40. All options Sweep mode: Swept, analyzer display turned off with: DISP:ENAB OFF, number of traces = 1, system error correction: OFF, NOP = 201 Option 230/235/240/245/260/265/280/ 285/430/435/440/445/460/465/480/485 Option 2D5/2K5/4D5/4K5 Frequency [Hz] IFBW [Hz] Cycle time [ms] Frequency [Hz] IFBW [Hz] Cycle time [ms] 4G 500000 5.0 10 G 500000 2.8 4G 400000 5.0 10 G 400000 2.8 4G 300000 4.9 10 G 300000 2.8 4G 200000 4.9 10 G 200000 2.8 4G 150000 4.9 10 G 150000 2.8 4G 100000 5.0 10 G 100000 3.0 4G 70000 5.7 10 G 70000 3.8 4G 50000 6.9 10 G 50000 4.9 4G 40000 7.8 10 G 40000 5.9 4G 30000 9.4 10 G 30000 7.5 4G 20000 13 10 G 20000 11 4G 15000 16 10 G 15000 14 4G 10000 23 10 G 10000 21 4G 7000 31 10 G 7000 29 4G 5000 42 10 G 5000 40 4G 4000 52 10 G 4000 50 4G 3000 68 10 G 3000 66 4G 2000 100 10 G 2000 98 4G 1500 133 10 G 1500 132 4G 1000 198 10 G 1000 196 4G 700 278 10 G 700 276 4G 500 391 10 G 500 389 4G 400 487 10 G 400 485 4G 300 648 10 G 300 646 4G 200 970 10 G 200 968 4G 150 1291 10 G 150 1289 4G 100 1935 10 G 100 1933 4G 70 2739 10 G 70 2739 4G 50 3864 10 G 50 3869 4G 40 4829 10 G 40 4837 4G 30 6437 10 G 30 6449 4G 20 9653 10 G 20 9671 4G 15 12870 10 G 15 12892 4G 10 19302 10 G 10 19330 1. Supplemental performance data. 2. Measured with firmware version A.09.30. 49 Data transfer time1, 2 (ms) Table 41. All options Number of points 51 SCPI over 201 401 1601 GPIB3 64-bit floating point 4 12 25 88 32-bit floating point 3 7 13 45 ASCII 30 108 223 841 REAL 64 1 1 2 4 REAL 32 1 1 1 2 ASCII 15 59 117 457 SCPI over 100 Mbps LAN (Telnet)3 SCPI over 100 Mbps LAN (SICL-LAN)3 REAL 64 3 3 4 6 REAL 32 3 3 3 5 4 8 14 47 2 2 2 3 ASCII SCPI over USB (SICL-USB)3 REAL 64 REAL 32 2 2 2 3 ASCII 4 13 24 91 REAL 64 8 15 21 68 REAL 32 7 10 13 37 ASCII 75 290 579 2313 1 1 1 1 SCPI over GPIB/USB (82357B) COM4 Variant type 1. Supplemental performance data. 2. Measured with firmware version A.09.30. 3. Measured using a VEE Pro 7.0 program running on a 3.2 GHz Pentium 4 DELL Precision 370, Transferred complex S11 data, using :CALC{1-36}:DATA:FDAT?. 4. Measured using an E5071C VBA macro running inside the analyzer. Transferred complex S11 data. 50 E5091A Multiport Test Set The section provides port performance of test set without calibration. Table 42. Test set input/output performance Description Specification Range Damage level 50 MHz to 8.5 GHz Typical 20 dBm, ±25 VDC (typical) Table 43. Option E5091A-009 port performance Specification 50 MHz to 300 MHz 300 MHz to 1.3 GHz 1.3 GHz to 3 GHz 3 GHz to 6 GHz 6 GHz to 8.5 GHz A, T2, R1+, R1– 19 dB 20 dB 18 dB 12 dB 10 dB T1, R2+, R2–, R3+, R3– 15 dB 17 dB 15 dB 11 dB 8 dB Description Load match Test port selected Test port unselected A, T2, R1+, R1–, R3+, R3– 23 dB 25 dB 19 dB 12 dB 11 dB T1, R2+, R2– 18 dB 20 dB 16 dB 12 dB 9 dB 19 dB 19 dB 17 dB 13 dB 9 dB A, T2, R1+, R1– 3 dB 3 dB 4 dB 5 dB 6 dB T1, R2+, R2–, R3+, R3– 5 dB 5 dB 7 dB 8 dB 9.5 dB Stability, typical 0.005 dB/° C 0.005 dB/° C 0.005 dB/° C 0.01 dB/° C 0.015 dB/° C –100 dB –100 dB –100 dB –100 dB –90 dB Interconnect port, typical P1, P2, P3, P4 Insertion loss Test port Isolation Over arbitrarily test ports 51 Table 44. Option E5091A-016 port performance Description Specification 50 MHz to 300 MHz 300 MHz to 1.3 GHz 1.3 GHz to 3 GHz 3 GHz to 6 GHz 6 GHz to 8.5 GHz 17 dB 15 dB 9 dB 8 dB Load match Test port selected A, T4, R1+, R1–, R2+, R2–, 15 dB R3+, R3–, R4+, R4– T1, T2, T3 12 dB 14 dB 14 dB 8 dB 6 dB Test port unselected A, T4, T2, R1+, R1–, R2+, R2– 18 dB R3+, R3–, R4+, R4–, R4– 20 dB 16 dB 10 dB 9 dB T1, T2, T3 13 dB 15 dB 14 dB 8 dB 6 dB Interconnect port, typical P1, P2, P3, P4 12 dB 12 dB 12 dB 9 dB 7 dB Test port A, T4, R1+, R1–, R2+, R2–, 6 dB R3+, R3–, R4+, R4– 6 dB 7 dB 8 dB 9.5 dB T1, T2, T3 6 dB Stability per switch, typical 0.005 dB/° C 9 dB 0.005 dB/° C 10.5 dB 12 dB 0.005 dB/° C 0.01 dB/° C 14.5 dB 0.015 dB/° C –100 dB –100 dB –80 dB Insertion loss Isolation Over arbitrarily test ports –100 dB Table 45. Front panel information Description RF connectors Interconnect ports Test ports (Option E5091A-009) Test ports (Option E5091A-016) Control line General characteristics Type-N, female, 50 Ω, nominal Number of ports 4 ports Type-N, female, 50 Ω, nominal Number of ports 9 ports SMA, female, 50 Ω, nominal Number of ports 25 ports (includes configurable switch port) 15 pin D-sub, female Table 46. Rear panel information Description USB port Line power 1 Frequency Voltage VA max General characteristics Type B-receptacle, provide connection to the E5071C 47 to 63 Hz 90 to 132 VAC, or 198 to 264 VAC (automatically switched) 150 VA max. 1. A third-wire ground is required. 52 –100 dB Table 47. Test set dimensions and block diagram Description Dimensions Option E5091A-009 Option E5091A-016 Weight Option E5091A-009 Option E5091A-016 Block diagram Option E5091A-009/016 General characteristics See Figure 10, 12, and 13 See Figure 11, 12, and 15 6 kg 7 kg See Figure 15 Figure 10. Dimensions (front view, with Option E5091A-009, in millimeters, nominal) Figure 11. Dimensions (front view, with Option E5091A-016, in millimeters, nominal) 53 Figure 12. Dimensions (rear view, in millimeters, nominal) Figure 13 Dimensions (side view, with Option E5091A-009, in millimeters, nominal) Figure 14. Dimensions (side view, with Option E5091A-016, in millimeters, nominal) 54 Figure 15. Block diagram 55 E5092A Configurable multiport test set The section provides test input/output performance without calibration by the E5071C. Table 48. Test set input/output performance Description Specification Frequency range 50 MHz to 20 GHz Damage level Typical 20 dBm, ±35 VDC Table 49. Option E5092A-020 port performance Description Specification Typical Load match (selected port) SPDT switch1 50 MHz to 2 GHz 2 GHz to 4 GHz 4 GHz to 8 GHz 8 GHz to 10 GHz 10 GHz to 18 GHz 18 GHz to 20 GHz 17 dB 11 dB 8 dB 7 dB 4 dB 4 dB SP4T switch2 50 MHz to 2 GHz 2 GHz to 3 GHz 3 GHz to 8 GHz 8 GHz to 10 GHz 10 GHz to 18 GHz 18 GHz to 20 GHz 17 dB 11 dB 8 dB 7 dB 4 dB 4 dB Load match (unselected port) SPDT switch1 50 MHz to 3 GHz 3 GHz to 10 GHz 10 GHz to 16 GHz 16 GHz to 18 GHz 18 GHz to 20 GHz 17 dB 11 dB 8 dB 6 dB 4 dB SP4T switch2 50 MHz to 3 GHz 3 GHz to 10 GHz 10 GHz to 16 GHz 16 GHz to 18 GHz 18 GHz to 20 GHz 17 dB 11 dB 8 dB 6 dB 4 dB 1. SPDT: Single-pole-double-throw switches. Applies to SW5, SW6, SW7, SW8, SW9 and SW10 in the E5092A. (See Figure 20.) 2. SP4T: Single-pole-four-throw switches. Applies to SW1, SW2, SW3 and SW4 in the E5092A. (See Figure 20.) 56 Table 49. Option E5092A-020 port performance (continued) Description Specification Typical Load match (common port) SPDT switch1 50 MHz to 2 GHz 2 GHz to 4 GHz 4 GHz to 8 GHz 8 GHz to 10 GHz 10 GHz to 20 GHz 16 dB 11 dB 8 dB 7 dB 4 dB SP4T switch2 50 MHz to 1.3 GHz 1.3 GHz to 4 GHz 4 GHz to 8 GHz 8 GHz to 10 GHz 10 GHz to 20 GHz 16 dB 11 dB 8 dB 7 dB 4 dB Insertion loss SPDT switch1 50 MHz to 100 MHz 100 MHz to 2 GHz 2 GHz to 3 GHz 3 GHz to 4 GHz 4 GHz to 6 GHz 6 GHz to 8 GHz 8 GHz to 10 GHz 10 GHz to 14 GHz 14 GHz to 18 GHz 18 GHz to 20 GHz 4 dB 3.5 dB 4.5 dB 5 dB 5.5 dB 7 dB 8 dB 8.5 dB 10 dB 11.5 dB SP4T switch2 50 to 100 MHz 100 MHz to 2 GHz 2 GHz to 3 GHz 3 GHz to 4 GHz 4 GHz to 6 GHz 6 GHz to 8 GHz 8 GHz to 10 GHz 10 GHz to 14 GHz 14 GHz to 18 GHz 18 GHz to 20 GHz 4 dB 3.5 dB 4.5 dB 5.5 dB 6 dB 7.5 dB 8.5 dB 9.5 dB 10.5 dB 12 dB Stability 50 MHz to 6 GHz 6 GHz to 20 GHz 0.003 dB/°C 0.01 dB/°C (Stability per switch) Isolation3 50 MHz to 500 MHz 500 MHz to 1 GHz 1 GHz to 2 GHz 2 GHz to 6 GHz 6 GHz to 10 GHz 10 GHz to 18 GHz 18 GHz to 20 GHz 65 dB 80 dB 85 dB 90 dB 85 dB 75 dB 65 dB (Over arbitrarily test ports) 1. SPDT: Single-pole-double-throw switches. Applies to SW5, SW6, SW7, SW8, SW9 and SW10 in the E5092A. (See Figure 20.) 2. SP4T: Single-pole-four-throw switches. Applies to SW1, SW2, SW3 and SW4 in the E5092A. (See Figure 20.) 3. This specification is defined when all ports are terminated with a 50 ohm load. 57 Table 50. Control line Description Specification Number of groups 4 Group A: 8 bits Group B,C,D: 4 bits Input voltage range1 0 V to +5 V (positive input) –5 V to 0 V (negative input) Maximum current Group A, B: 50 mA in total of each group Group C, D: 500 uA in total of each group Typical Impedance Group A, B: < 10 ohm Group C, D: < 200 ohm Table 51. DC source Description Specification Number of sources 4 Typical 0 V to +5.2 V (nominal)2 Output voltage range Output voltage accuracy ±3 % of setting (+1 V to +5 V) at 1 M ohm load impedance 10 mV (nominal)3 Voltage resolution Maximum current 150 mA for each source Output impedance < 5 ohm Table 52. Operating storage environment Description General characteristics Temperature +5 °C to +40 °C Humidity 20 to 80% at wet bulb temperature < +29 °C (non-condensing) Altitude 0 to 2,000m (0 to 6,561 feet) Vibration 0.21 G max., 5 to 500 Hz Table 53. Non-operating storage environment Description General characteristics Temperature –10 °C to +60 °C Humidity 20 to 90 % at wet bulb temperature < +40 °C (non-condensing) Altitude 0 to 4,572 m (0 to 15,000 feet) Vibration 0.5 G max., 5 Hz to 500 Hz 1. Input voltage will be clipped at about ±5.2 V when over this range. 2. The output voltage can be set in this range. 3. The output voltage resolution becomes effective between 0 V to 5.2 V. 58 Table 54. Front panel information Description General characteristics RF connectors SMA Test ports 38 ports Control line 15-pin D-sub, female 25-pin D-sub, female Table 55. Rear panel information Description General characteristics USB port Type B-receptacle, provide connection to the E5071C Line power1 Frequency Voltage VA max 47 to 63 Hz 90 to 132 VAC, or 198 to 264 VAC (automatically switched) 300 VA max. Table 56. Test set dimensions and block diagram Description General characteristics Dimensions E5092A Option 020 See Figures 16, 17, 18 and 19 Weight E5092A Option 020 9 kg 426 192 133 18 129 Figure 16. Dimensions (front view, with Option E5092A-020, in millimeters, nominal) 1. 1A third-wire ground is required. 59 40.6 40.6 36.8 17 40.6 27 23.38 132.72 27 27 27 27.8 13.5 27 65.34 27 18.57 23.38 27 69.8 Figure 17. Dimensions (pitch between switches, with Option E5092A-020, in millimeters, nominal) 37 51 83 133 129 Figure 18. Dimensions (rear view, with Option E5092A-020, in millimeters, nominal) 472 417 22 133 146 32 13 18 144 126 23 Figure 19. Dimensions (side view, with Option E5092A-020, in millimeters, nominal) 60 SW 1 PORT 1 1A 1B 1C 1D SW 5 5A 5B PORT 2 SW 2 SW 3 6A 6B 6C O M 5C O M 7A 7B SW 7 SW 4 3A 3B 3C 3D 2A 2B 2C 2D SW 6 PORT 3 SW 8 7C O M 8A 8B SW 9 4A 4B 4C 4D 9A 9B 9C O M 8C O M PORT 4 S W 10 10A 10B 10C O M Figure 20. Switch configuration (E5092A-020) Control Line 1 (Group A) DC Source (Group A) Max Current DC Source Output Control Line 1 (Group A) Control Line 2 (Group A) Positive Input (High) Control Line 3 (Group A) Output Impedance Output Voltage Control Line 4 (Group A) : Negative Input (Low) DC Source (Group B) DC Source Output Impedance DC Source Output Control Line 1 (Group B) Control Line 2 (Group B) Control Line 3 (Group B) Control Line 4 (Group B) Control Line 2 (Group C) Control Line 1 (Group C) Control Line 2 (Group C) Control Line 3 (Group C) Control Line 4 (Group C) Positive Input (High) Negative Input (Low) DC Source (Group D) DC Source Output : Control Line 8 (Group A) Control Line 2 (Group B) Positive Input (High) Negative Input (Low) DC Source (Group C) : Control Line 2 (Group D) Control Line 1 (Group D) Control Line 2 (Group D) Control Line 3 (Group D) Control Line 4 (Group D) Positive Input (High) Negative Input (Low) Figure 21. DC control line (E5092A-020) 61 23 22 21 20 19 18 17 16 Positive input DC source GND Negative input Positive input DC source GND Negative input 12 11 GROUP B 10 9 8 7 6 5 GROUP C 12 11 10 9 GND Negative input Positive input DC source Reserved Reserved Reserved Line 8 Line 7 Line 6 Line 5 Line 4 Line 3 Line 2 Line1 8 7 6 5 4 3 2 1 GROUP C 4 3 Line 1 13 Line 2 Line 3 Line 4 Line 1 GROUP B Line 2 Line 3 Line 4 Line 1 Line 2 Line 3 Line 4 GND 14 15 14 DC source 24 Negative input 13 15 Positive input 25 GND D-SUB 15 pin assignment GROUP A GROUP A D-SUB 25pin assignment GROUP D 2 1 GROUP D Figure 22. Control line pin assignment (E5092A-020) 62 Corrected System Performance for 75 Ω Measurements with 11852B 50 to 75 Ω Minimum-Loss Pads (Supplemental Information) Option 230/235/240/245/260/265/280/285/430/435/440/445/ 460/465/480/485 Table 55. Corrected system performance with type-N 75 Ω device connectors, 85036E calibration kit Network analyzer: calibration kit: 50 to 75 Ω adapters: calibration: E5071C 85036E (type-N 75 Ω) 11852B full 2-port IF bandwidth = 10 Hz, no averaging applied to data, environmental temperature = 23° C ±5° C with < 1° C deviation from calibration temperature, isolation calibration performed Description Typical (dB) 10 MHz to 3 GHz Directivity Source match Load match Reflection tracking Transmission tracking 37 33 39 ±0.015 ±0.019 Transmission uncertainty 10 MHz to 3 GHz (typical) Phase Magnitude Transmission using 85036E and 11852B 100 1 10 Phase (deg) Mag (dB) Transmission using 85036E and 11852B 10 0.1 1 10 MHz to 3 GHz 10 MHz to 3 GHz S11 = S22 = 0; Source power = –10 dBm 0.01 10 0 S11 = S22 = 0; Source power = –10 dBm 0.1 10 –10 –20 –30 –40 –50 –60 –70 –80 –90 0 –10 –20 –30 –40 –50 –60 –70 –80 –90 Transmission coefficient (dB) Transmission coefficient (dB) Reflection uncertainty 10 MHz to 3 GHz (typical) Phase Magnitude Reflection using 85036E and 11852B Reflection using 85036E and 11852B 10 0.1 Phase (deg) Mag (linear) 10 MHz to 3 GHz 8 10 MHz to 3 GHz 0.08 S21 = S12 = 0; Source power = –10 dBm 0.06 0.04 S21 = S12 = 0; Source power = –10 dBm 6 4 2 0.02 0 0 0 0.2 0.4 0.6 0.8 1 0 0.2 Reflection coefficient (linear) 0.4 0.6 0.8 Reflection coefficient (linear) 63 1 www.agilent.com www.agilent.com/find/ena Agilent Email Updates www.agilent.com/find/emailupdates Get the latest information on the products and applications you select. Agilent Direct www.agilent.com/find/agilentdirect Quickly choose and use your test equipment solutions with confidence. www.lxistandard.org LXI is the LAN-based successor to GPIB, providing faster, more efficient connectivity. Agilent is a founding member of the LXI consortium. Remove all doubt Our repair and calibration services will get your equipment back to you, performing like new, when promised. You will get full value out of your Agilent equipment throughout its lifetime. Your equipment will be serviced by Agilent-trained technicians using the latest factory calibration procedures, automated repair diagnostics and genuine parts. You will always have the utmost confidence in your measurements. For information regarding self maintenance of this product, please contact your Agilent office. Agilent offers a wide range of additional expert test and measurement services for your equipment, including initial start-up assistance, onsite education and training, as well as design, system integration, and project management. For more information on repair and calibration services, go to: www.agilent.com/find/removealldoubt For more information on Agilent Technologies’ products, applications or services, please contact your local Agilent office. The complete list is available at: www.agilent.com/find/contactus Americas Canada Latin America United States (877) 894-4414 305 269 7500 (800) 829-4444 Asia Pacific Australia China Hong Kong India Japan Korea Malaysia Singapore Taiwan Thailand 1 800 629 485 800 810 0189 800 938 693 1 800 112 929 0120 (421) 345 080 769 0800 1 800 888 848 1 800 375 8100 0800 047 866 1 800 226 008 Europe & Middle East Austria 01 36027 71571 Belgium 32 (0) 2 404 93 40 Denmark 45 70 13 15 15 Finland 358 (0) 10 855 2100 France 0825 010 700* *0.125 €/minute Germany 07031 464 6333 Ireland 1890 924 204 Israel 972-3-9288-504/544 Italy 39 02 92 60 8484 Netherlands 31 (0) 20 547 2111 Spain 34 (91) 631 3300 Sweden 0200-88 22 55 Switzerland 0800 80 53 53 United Kingdom 44 (0) 118 9276201 Other European Countries: www.agilent.com/find/contactus Revised: March 24, 2009 Product specifications and descriptions in this document subject to change without notice. © Agilent Technologies, Inc. 2007-2009 Printed in USA, July 30, 2009 5989-5479EN