Arch MSC-5S Switching power modules for pcb mounting Datasheet

VER : D_5
update : 2016.09.01
AC-DC Medical
MSC SERIES
40 Watts
KEY FEATURES
 Switching Power Modules for PCB Mounting
 Fully Encapsulated Plastic Case
 Universal Input Range 90-264VAC, 47-440 Hz
 Regulated Output
 Low Ripple and Noise
 Isolation Class II
 CE, CB, UL, cUL, Approvals
 3-Year Product Warranty
ELECTRICAL SPECIFICATIONS
Model No.
( Single Output )
Max output wattage
Output
(W)
MSC-5S
MSC-9S
MSC-12S
MSC-15S
MSC-24S
26.4W
40W
40W
40W
40W
40W
Voltage (V.DC.)
3.3V
5V
9V
12V
15V
24V
Current (mA) max
Voltage set accuracy
8000
±2%
8000
±2%
4444
±2%
3333
±2%
2666
±2%
1667
±2%
Line regulation (typ.)
0.5%
0.5%
0.5%
0.5%
0.5%
0.5%
1%
1%
1%
1%
1%
1%
Load regulation (1-100%) (typ.)
Minimum load
1%
1%
1%
1%
1%
1%
Efficiency (typ.)
74%
79%
82%
83%
83%
83%
70000 uF
33000 uF
10000 uF
4000 uF
3000 uF
1200 uF
Maximum Capacitive Load (at 230 VAC)
Model No.
MSC-3.3S
( Dual Output )
Max output wattage
MSC-5D
(W)
40W
40W
Output 1
Output 2
Output 1
Output 2
Voltage (V.DC.)
+5V
-5V
+12V
-12V
+15V
-15V
Current (mA) max
4000
4000
1666
1666
1333
±2%
Line regulation (typ.)
±2%
1333
±2%
0.5%
0.5%
0.5%
1% (symmetric load)
1% (symmetric load)
1% (symmetric load)
Cross regulation (typ.)
5%
5%
5%
Minimum load
10%
10%
10%
Efficiency (typ.)
80%
83%
83%
±6000 uF
±1000 uF
±2000 uF
Load regulation (10-100%) (typ.)
( Dual Output )
Max output wattage
MSC-5S12S
(W)
Voltage (V.DC.)
Current (mA) max
Voltage set accuracy
Output
40W
Output 2
Maximum Capacitive Load (at 230 VAC)
Model No.
MSC-15D
Output 1
Voltage set accuracy
Output
MSC-12D
Line regulation (typ.)
Load regulation (10-100%) (typ.)
Cross regulation (typ.)
MSC-5S24S
40W
40W
Output 1
Output 2
Output 1
Output 2
5V
12V
5V
24V
5000
±3%
1250
±5%
5000
±3%
625
±5%
0.5%
5%
0.5%
5%
2% (symmetric load)
6% (symmetric load)
2% (symmetric load)
6% (symmetric load)
1%
7%
1%
7%
Minimum load
25%
25%
Efficiency (typ.)
80%
80%
15000 uF / 750 uF
15000 uF / 200 uF
Maximum Capacitive Load (at 230 VAC)
http://www.archcorp.com.tw
TEL: +886-2-26989508 FAX: +886-2-26981319
-1-
VER : D_5
update : 2016.09.01
AC-DC Medical
MSC SERIES
40 Watts
ELECTRICAL SPECIFICATIONS
Model No.
( Triple Output )
Max output wattage
Output
MSC-5S12D
MSC-5S15D
40W
40W
(W)
Output 2
Output 3
Output 1
Output 2
Voltage (V.DC.)
+12
-12
5
+15
Output 3
-15
Current (mA) max
600
600
500
500
Voltage set accuracy
±5%
5000
±3%
Line regulation (typ.)
Load regulation (25-100%) (typ.)
(symmetric load)
Cross regulation (typ.)
5%
0.5%
5%
7%
3%
7%
7%
Minimum load
3%
25%
Efficiency (typ.)
±150 uF
7%
25%
79%
Maximum Capacitive Load (at 230 VAC)
±5%
79%
18000 uF
±400 uF
ELECTRICAL SPECIFICATIONS
All specifications valid at normal input voltage, full load and +25°C after warm-up time unless otherwise stated.
Input
Output
Protection
Isolation
Environment
Physical
Voltage
90-264 VAC or 100-375 VDC
Frequency (Hz)
47-440 Hz
Current (Full load)
860 mA max. (115 VAC) / 460 mA max. (230 VAC)
Inrush current (<2ms)
55 A max. (115 VAC) / 95 A max. (230 VAC)
Leakage Current
< 0.1mA / 264 VAC (Touch Current)
External fuse (recommend)
3.15 A slow blow type
Ripple & Noise
3.3V:50mV of Vout
Hold-up time
18 ms min.
Over current protection
Above 105% rated output power
Over voltage protection
Zener diode clamp
Short circuit protection
Hiccup mode, indefinite (automatic recovery)
OTP
Input-Output (V.AC)
4000V
Operating temperature
-25°C…+70°C (Case Temperature max. +95°C)
Storage temperature
-40°C…+85°C
Temperature coefficient
0.01%/°C
Humidity
95% RH
100°C
MTBF
200,000 h~400,000 h @ 25°C
Dimension (L x W x H)
3.5 x 2.5 x 1.06 Inches
Case Material
Plastic resin (flammability to UL 94V-0)
( 89.0 x 63.5 x 27.0 mm ) Tolerance 0.5 mm
Weight
280 g
Cooling method
Free air convection
Class II
To IEC / EN 60536
cUL / UL / CB /
ANSI/AAMI ES 60601-1: 2005, 1st Edition and CAN/CSA-C22.2 No. 60601-1:08, 2nd Edition,
2XMOPP
EN 55011 class B
Approval
EMI
Safety & EMC
others:1% of Vout
EMC Standards
EN60601-1-2
ESD Susceptibility
EN60601-1-2
Radiated Susceptibility
EN60601-1-2
EFT / Burst
EN60601-1-2
Surge
EN60601-1-2
Conducted Susceptibility
EN60601-1-2
Safety Approvals
EN60601-1
http://www.archcorp.com.tw
TEL: +886-2-26989508 FAX: +886-2-26981319
-2-
VER : D_5
update : 2016.09.01
AC-DC Medical
MSC SERIES
40 Watts
NOTE
1. Ripple & Noise are measured at 20MHz of bandwidth with 0.1uF & 47uF parallel capacitor.
2. It's recommended to add Varistor 14S471K at L / N input side in parallel.
3. Please refer to our PDF file "AC-DC Application" on our website: www.archcorp.com.tw
MECHANICAL DIMENSION
(Top View)
5S/12S
PIN#
Single
Dual
Triple
5S/24S
BLOCK DIAGRAM
http://www.archcorp.com.tw
TEL: +886-2-26989508 FAX: +886-2-26981319
1
AC IN (L)
AC IN (L)
AC IN (L)
AC IN (L)
2
AC IN (N)
AC IN (N)
AC IN (N)
AC IN (N)
3
+DC OUT
+DC OUT
+OUT2
+DC OUT
4
NO PIN
NO PIN
+OUT1
+5V
5
-DC OUT
COMMON
+OUT2 RTN
COMMON
6
NO PIN
NO PIN
+OUT1 RTN
+5 RTN
7
NOCONNECT
-DC OUT
NO PIN
-DC OUT
DERATING
-3-