5W High Reliable Green Medical On Board Type + + M F M - 0 5 series ■ ‧1.65"x0.88" compact size ‧Medical safety approved (2 x MOPP) accroding to ANSI/AAMI ES60601-1 and IEC/EN60601-1 ‧Suitable for BF application with appropriate system consideration ‧No load power consumption<0.075W ‧Extremely low leakage current ‧Wide operating temp. range -40 ~ +85℃ ‧EMI class B for class Ⅱ configuration ‧Protections: ■ ‧ ‧ ‧ ‧ Short circuit / Overload / Over voltage / Over temperature ‧ ‧52K hours ‧3 years warranty ■ Ⅱ μ ■ File Name:MFM-05-SPEC 2017-08-08 5W High Reliable Green Medical On Board Type MFM-05 series SPECIFICATION MFM-05-3.3 MODEL DC VOLTAGE RATED POWER 3.3V 1.25A Note.2 0 ~ 1.25A 1.38A 4.1W PEAK LOAD(10sec.) Note.3 RATED CURRENT CURRENT RANGE PEAK CURRENT OUTPUT 4.6W VOLTAGE TOLERANCE Note.5 ±2.5% LOAD REGULATION ±0.3% ±0.5% SETUP, RISE TIME 1000ms, 30ms/230VAC HOLD UP TIME (Typ.) 40ms/230VAC LINE REGULATION VOLTAGE RANGE INPUT MFM-05-5 RIPPLE & NOISE (max.) Note.4 100mVp-p MFM-05-12 MFM-05-15 MFM-05-24 5V 1A 0 ~ 1A 1.1A 5W 12V 0.42A 0 ~ 0.42A 0.46A 5W 15V 0.33A 0 ~ 0.33A 0.36A 5W 24V 0.23A 0 ~ 0.23A 0.25A 5.5W 5.5W 100mVp-p ±2.5% ±0.3% ±0.5% 5.5W 150mVp-p ±2.5% ±0.3% ±0.5% 5.4W 150mVp-p ±2.5% ±0.3% ±0.5% 6W 180mVp-p ±2.5% ±0.3% ±0.5% 81% 82% 1000ms, 30ms/115VAC at full load 12ms/115VAC at full load Note.6 80 ~ 264VAC FREQUENCY RANGE 47 ~ 440Hz EFFICIENCY (Typ.) 74% 80% 0.2A/115VAC 0.1A/230VAC COLD START 25A/115VAC 45A/230VAC AC CURRENT (Typ.) INRUSH CURRENT (Typ.) 80% LEAKAGE CURRENT (max.) Note.7 Touch current <80μA/264VAC PROTECTION OVERLOAD 110% ~ 180% rated output power Protection type : Hiccup mode, recovers automatically after fault condition is removed OVER VOLTAGE 13.8 ~ 16.2V 3.8 ~ 5V 5.75 ~ 6.8V Protection type : Shut off o/p voltage, clamping by zener diode OVER TEMPERATURE Protection type : Shut down o/p voltage, recovers automatically after temperature goes down WORKING TEMP. -40 ~ +85℃ (Refer to "Derating Curve") 20 ~ 90% RH non-condensing WORKING HUMIDITY ENVIRONMENT STORAGE TEMP., HUMIDITY TEMP. COEFFICIENT SOLDERING TEMPERATURE 17.3 ~ 20.3V 27.6 ~ 32.4V -40 ~ +100℃, 10 ~ 95% RH non-condensing ±0.03%/℃ (0 ~ 60℃) 260℃±5℃/10sec.max. VIBRATION 10 ~ 500Hz, 5G 10min./1cycle, period for 60min. each along X, Y, Z axes OPERATING ALTITUDE Note.8 5000 meters SAFETY STANDARDS IEC60601-1, EN60601-1, UL ANSI/AAMI ES60601-1(3.1 version), CAN/CSA-C22 3rd Edition approved ; Design refer to EN60335-1 ISOLATION LEVEL WITHSTAND VOLTAGE Primary-Secondary: 2xMOPP ISOLATION RESISTANCE I/P-O/P:100M Ohms / 500VDC / 25℃/ 70% RH Parameter Standard Test Level / Note Conducted EN55011 (CISPR11) Class B Radiated EN55011 (CISPR11) Class B Harmonic Current EN61000-3-2 Class A Voltage Flicker EN61000-3-3 ----- Parameter Standard Test Level / Note ESD EN61000-4-2 Level 3, 8KV air ; Level 4, 8KV contact Radiated EN61000-4-3 Level 3 EFT / Burst EN61000-4-4 Level 3 Surge EN61000-4-5 Level 2, 1KV/Line-Line Conducted EN61000-4-6 Level 3 Magnetic Field EN61000-4-8 Level 4 Voltage Dips and Interruptions EN61000-4-11 100% dip 1 periods, 30% dip 25 periods, 100% interruptions 250 periods EMC EMISSION SAFETY & EN55024 , EN60601-1-2, EN61204-3 EMC (Note 9) EMC IMMUNITY MTBF OTHERS NOTE I/P-O/P:4KVAC DIMENSION 1799.5Khrs min. MIL-HDBK-217F (25℃) 42*22.3*20.5mm (L*W*H) or 1.65"*0.88"0.80" inch PACKING 0.018Kg; 270pcs/5.8Kg/0.97CUFT ℃ μ μ ℃ File Name:MFM-05-SPEC 2017-08-08 MFM-05 5W High Reliable Green Medical On Board Type series Block Diagram fosc : 100KHz RECTIFIERS & FILTER EMI FILTER I/P PWM CONTROL O.L.P. Derating Curve +V RECTIFIERS & FILTER POWER SWITCHING -V DETECTION CIRCUIT Output Derating VS Input Voltage 100 100 90 80 70 LOAD (%) LOAD (%) 70 40 20 -40 -30 -20 -10 0 10 20 30 40 50 60 AMBIENT TEMPERATURE (℃) 70 80 85 (HORIZONTAL) 60 50 40 80 95 100 115 120 140 160 180 200 220 240 264 INPUT VOLTAGE (VAC) 60Hz File Name:MFM-05-SPEC 2017-08-08 MFM-05 5W High Reliable Green Medical On Board Type Mechanical Specification series Unit: mm 22.3 42 AC/N 1.9 8 FS1 1.9 10.75 +V AC/L 1.75 -V 1.75 20.5 16 TOP VIEW 1.8 1.05 SIDE VIEW 4.5 3.5 Installation Manual Please refer to : http://www.meanwell.com/manual.html File Name:MFM-05-SPEC 2017-08-08