Powerbox OBN01055 150-200 watts single & multiple output ac/dc medical Datasheet

MEDICAL & MILITARY
Medline 200 Series
1 5 0 - 2 0 0 W AT T S S I N G L E & M U LT I P L E O U T P U T A C / D C M E D I C A L
Features
• Low safety ground leakage current
• Wide input range 90 to 260Vac
• Input surge current protection
• Overvoltage protection
• Overcurrent protection
• Power fail detection (PFD) signal
• Optional cover and fan assembly
• 100% burn-in at full rated load
• Remote inhibit - TTL high disables output
• Short curcuit protection
Specifications
PFD signal
TTL logic high for normal operation and TTL logic low
upon loss of input power. This signal appears at least
1ms prior to +5V output dropping 5% below its
nominal value. This signal also provides a minimum
delay of 100ms after +5V is within regulation
Efficiency
70% min on all models
Hold up time
20ms min at 110VAC
Line regulation
± 0.5% max at full load
INPUT
Voltage range
90-260VAC
Frequency
47-63Hz.
Input Current
3.2A (rms) at 115VAC
1.6A (rms) at 230VAC
Leakage current
90µA max at 115VAC, 60Hz
150A max at 230VAC, 50Hz
Inrush Current
20A @ 115V or 40A at 230V at 25°C and cold start
ENVIRONMENTAL
Isolation
Input – Output: 4000VAC
Input – Ground: 1500VAC
Output – Ground: 500VAC
Operating Temperature
0°C to 70°C
Storage temperature
–40°C to +85°C
Humidity
5% to 95% non-condensing
Derating
Derate from 100% at +50°C linearly to 50% at
+70°C.
Cooling
25CFM forced air, provided on “C” version, to be
provided for “B” version by user
OUTPUT
Power
200W
Voltage
See table
Current
See table
Ripple & Noise
2% p-p max
GENERAL
Overvoltage protection
Provided on output #1 only, set at 112-132% of it
nominal output voltage
Switching frequency
88-112kHz.
MTBF
350,000 hours at full load 25°C ambient ,
calculated per MIL-HDBK-217F
Overcurrent Protection
The output protected to short circuit conditions
Temperature coefficient
0.04%°C all outputs
STANDARDS AND APPROVALS
Transient response
Max excursion of 4% or better on all models,
recovering to 1% of final value within 500us after
a 25% step load change
Safety standards
IEC60601-1 TUV EN60601-1, EN60601-1
C-Tick
AS/NZS CISPR11 Group 1, Class A
EMC standards
EN55011 Class B, FCC Class B,VCII, EN61000-3-2, -3,
EN60001-4-2, -3, -4, -5, -6, -8, -11
Technical Illustration
MECHANICAL
Dimensions
See drawing
Weight
0.82kg for “B” version
0.96kg for version “C” version
Notes:
1. Dimensions shown in inches (mm)
2. Tolerance 0.02 (0.5) maximum
3. Input connector P1 is Dinkle DT-35-B01W-09 screws are M3, nickle plated
4. Connector P3 mates with Molex housing 22-01-1043 and Molex 40445 series
crimp terminal
5. Connector P4/P5 mates with Molex housing 22-01-1023 and Molex 40445
series crimp terminal
6. P4 is for DC fan rated 24V/0.2A (models PM201/27), or 5V/0.38A (models
PM201-40-3), and 12V/0.2A (other models, Pin1 +V and Pin2 COM
7. Weight:
0.82 Kgs (1.83 Lbs) approx for “B” version
0.96 Kgs (2.14 Lbs) approx for “C” version
Your dependable power partner – www.powerbox.com.au
345
MEDICAL & MILITARY
Medline 200 Series
1 5 0 - 2 0 0 W AT T S S I N G L E & M U LT I P L E O U T P U T A C / D C M E D I C A L
Selection Table
MAX.
OUTPUT #1
MODEL
OUTPUT #2
VNOM IMIN
IMAX
OUTPUT #3
TOL
VNOM
OUTPUT #4
IMIN IMAX TOL
VNOM
IMIN
OUTPUT
NUMBER
VNOM
IMIN
IMAX
TOL
IMAX
TOL
POWER
OBN01016
5.1V
0A
35A
2%
175W
OBN01013
3.3V
0A
46A
3%
150W
OBN01037
12V
0A
16.7A
2%
200W
OBN01046
15V
0A
13.4A
2%
200W
OBN01055
24V
0A
8.4A
2%
200W
OBN01070
30V
0A
6.7A
2%
200W
OBN01204
48V
0A
4.2A
2%
200W
OBN01207
+5.1V
3.0A
30A
2%
+12V
0A
8A
4%
200W
OBN01210
+5.1V
3.0A
30A
2%
+15V
0A
6A
4%
200W
OBN012A5
+5.1V
3.0A
30A
2%
+24V
0A
4A
4%
200W
OBN01310
+12V
1.0A
8A
2%
+24V
0A
4A
4%
OBN01313
+5.1V
3.0A
30A
2%
+12V
0A
8A
4%
-5V
0A
6A
4%
200W
OBN01316
+5.1V
3.0A
30A
2%
+12V
0A
8A
4%
-12V
0A
4A
4%
200W
OBN01315
+5.1V
3.0A
30A
2%
+15V
0A
6A
4%
-15V
0A
4A
4%
200W
OBN01319
+5.1V
3.0A
30A
2%
+15V
0A
6A
4%
-12V
0A
4A
4%
200W
OBN01404
+5.1V
3.0A
30A
2%
+12V
0A
8A
4%
-24V
0A
4A
4%
200W
OBN01403
+5.1V
3.0A
30A
2%
+12V
0A
8A
4%
-12V
0A
4A
4%
5V
0A
6A
4%
200W
OBN01407
+5.1V
3.0A
30A
2%
+15V
0A
6A
4%
-15V
0A
4A
4%
24V
0A
4A
4%
200W
OBN01410
+5.1V
3.0A
30A
2%
+12V
0A
8A
4%
-12V
0A
4A
4%
12V
0A
4A
4%
200W
OBN01420
+5.1V
3.0A
30A
2%
+12V
0A
8A
4%
-15V
0A
4A
4%
15V
0A
4A
4%
200W
OBN01413
+3.3V
3.0A
30A
3%
+5.1V
0A
8A
4%
-12V
0A
4A
4%
12V
0A
4A
4%
175W
200W
NOTES: (1) Add suffix “B” for U-bracket format and “C” for enclosed format.
(2) Peak current is 12A on +12V, 9A on +15V and 6A on +24V.
(3) Output #4 is floating. It can be connected externally for positive or negative output.
(4) 200 watts for “C” version with a cover and fan assembly 100 watts for “B” version without moving air (maximum current of ouput #1 & #2 derated to 50%),
or 200 watts with 25 CFM forced air provided by user.
(5) When the remote Sense facility is not used, +Sense must be connected to +V, and -Sense to Return, on P2 connector.
(6) All models may be operated at no-load. At no-load, output voltage tolerance increases to 10%.
Pin Chart
CONNECTOR
P1
MODEL/PIN
OBN01016
OBN01031
OBN01046
OBN01070
OBN01013
OBN0PM201-13
OBN01055
OBN01204
OBN01210
OBN01207
OBN01310
OBN01316
OBN01313
OBN01315
OBN01319
OBN01404
OBN01407
OBN01420
346
OBN01403
OBN01410
OBN14113
P2
P3
P4
1
2
3
4
5
-Sense
RET.
RET.
RET
RET
Output Output Output +Sense FAN
#1
#1
#1
RET
Output #1
Output COM
#1
RET.
COM
RET.
COM
RET.
Output
#2
Output #1
Output COM
#1
RET.
COM
RET.
COM
RET.
Output Output
#2
#3
Output #1 Output COM
#1
RET.
COM
RET.
COM
RET.
Output
#2
COM
RET.
COM
RET.
COM
RET.
Output #1
Output
#1
6
7
8
9
1
2
3
4
1
2
COM COM. PFD INHIBIT+ INHIBITRET.
FAN
COM COM
RET. RET.
PFD INHIBIT+ INHIBIT-
FAN
COM COM
RET. RET.
PFD INHIBIT+ INHIBIT-
Output Output
#3
#3
RETURN
FAN
COM COM
RET. RET.
PFD INHIBIT+ INHIBIT-
Output Output Output Output
#2
#2
#3
#4
RETURN
FAN
COM COM
RET. RET.
PFD INHIBIT+ INHIBIT-
N.C.
N.C.
N.C.
N.C.
N.C.
N.C.
Similar pages