MPS HPU12A-103 12w wall mount type medical power supplies for home health care application Datasheet

HPU12A SERIES
12W Wall mount type medical power supplies For Home health care applications
Description:
The HPU12A series of AC/DC switching mode power supplies
provide 12 watts of continuous output power.
This series is suitable for use in Home health care and portable
applications. All models meet FCC Part-18 class B and CISPR11 EN55011 class B emission Limits and are designed to
comply with UL/c-UL(UL 60601-1) ,TUV/T-mark(EN 60601-1)
and new CE requirements.
Features:
Wide Input Voltage 90 to 264 VAC,47 to 63 Hz
2 Prong Plug-In Mains Connector
Output Voltage Available From 5VDC Thru 27VDC
Optional Output Connector (See appendix)
Single Output
Safety Approvals:
Class II
CEC level V and Engergy Star 2.0 Compliance
Over Load protection.
2 year warranty
Electrical Characteristics:
Sym.
Parameter
Vin
fin
Po
Input Voltage
Input Frequency
Output Power Range
Vo
Io
Iil
Iih
Output Voltage Range
Output Current Range
Input Current (Low Line)
Input Current (High Line)
Irl
Low Line Inrush Current
Irh
High Line Inrush Current
Eff
Efficiency
Test Conditions
Operating Voltage
Vin=90 to 264VAC
Io=Full load,
Io=Full load,
Io=Full load,
Vin=115VAC
Io=Full load,
Vin=230VAC
Io=Full Load,
Vin=115VAC
Vin=230VAC
25 c, Cool start,
Min.
Unit
264
63
12
VAC
Hz
W
See rating chart
See rating chart
0.25 0.35
0.17 0.22
V
A
A
A
14
16
A
28
30
A
25 c, Cool start,
Vin=230VAC
Typ. Max.
90
47
0
77.7
%
REG-i Line Regulation
Io=Full Load
1
%
REG-o Load Regulation
Vin=230VAC
5
%
OCP
Over Current Protection
Ttr
Time of Transient Response Io=Full Load to Half Load, Vin=100VAC
Thold Hold-Up Time
110
Io=Full Load, Vin=110VAC
10
Ts
Start Up Time
Io=Full Load, Vin=100VAC
0.3
Vrn
Ripple & Noise(Peak to Peak)
Full Load, Vin=90VAC
TC
Temperature Coefficient
All output
The parameter is not subject
to production test-verified
by design/characterization
of integrated controller.
Tjsd
1
Thermal Shutdown ○
by
Junction Temperature
Controller
150
%
4
mS
mS
1
2
S
1
%
-0.04
0.04
%/ c
-25
130
c
1 As long as faulty conditions have been removed , the adaptor will automatically power up as usual.
○
Environmental :
Sym.
Parameter
Toper
Operating Temperature
Tstg
Storage Temperature
Hr
Relative Humidity
T e s t Conditions
Min. Typ. Max. Unit
No-Condensing
MTBF
Operating Temperature at 25 C, Calculated per MIL-HDBK-217F
Pd
Derate linearly from 100% load at 40 C to 50% load at 70 C
40
70
C
-40
85
C
5
95
0
0.1M
%
Hrs
HPU12A SERIES
12W Wall mount type medical power supplies For Home health care applications
Safety Specifications:
Parameter
Sym.
Vps Dielectric Withstanding Voltage
for Primary to secondary
CISPR EMI requirements for CISPR-11
FCC EMI requirements for FCC PART-18
Test Conditions
Primary to secondary
Vin=220VAC
Vin=110VAC
Min. Typ. Max.
Unit
5600
VDC
B
B
CLASS
CLASS
Output Voltage And Current Rating Chart ( Single Output ) :
*
*
*
*
*
Model Number
Output Voltage
Output Current
Total
Regulation
Maximum
Output Power
HPU12A-102
5 ~ 6 VDC
2.00 ~ 1.66 A
5%
10W
HPU12A-103
6 ~ 8 VDC
2.00 ~ 1.50 A
5%
12W
HPU12A-104
8 ~ 11 VDC
1.50 ~ 1.09 A
5%
12W
HPU12A-105
11 ~ 13 VDC
1.09 ~ 0.92 A
5%
12W
HPU12A-106
13 ~ 16 VDC
0.92 ~ 0.75 A
5%
12W
HPU12A-107
16 ~ 21 VDC
0.75 ~ 0.57 A
3%
12W
HPU12A-108
21 ~ 27 VDC
0.57 ~ 0.45 A
3%
12W
Mark “
*
” means approved by CEC level V.
The total regulation on 102~104 required to use AWG#18/6FT output cable.
The total regulation on 105~108 required to use AWG#20/6FT output cable.
The regulation and efficiency will be changed by modified output cable.
Mechanical Specifications:( USA Type)
PO W E R
Note:
1. Dimensions are shown in mm.
2. Weight: 130gs approx.
3. Optional output connector:
See page Appendix.