TECHNICAL DATASHEET ``` MDS Medical AC-DC Adapter 12Volt, 120Watt / MDS-150AAS12 F/B 150AAS12 F/B Highlights & Features – – – – – – – – Meet DOE Level VI and ErP Lot 7 Safety Approvals to IEC 60601-1 3.1rd ed. & IEC 60950-1 Compliant with IEC 60601-1-2 4th Ed. Requirements Low touch current (<0.1mA Normal & ,0.3mA single fault) Over-Voltage/Load/Temperature & Short Circuit protections 1 Million Hours MTBF 2 × MOPP isolation 3 years warranty Safety Standards CB Certified for worldwide use Model Number: MDS-150AAS12 F/B Unit Weight: 1.1kg (39.6 ounces) Dimensions (W x L x H): 85.0x170.0x40.0 mm 3.3x6.7x1.6 in General Description The MDS series of external power supply come with universal AC input at 90Vac to 264Vac. Other features include low touch current, risk management report available and the electric shock protection comply with 2 x MOPP. The MDS series is certified for EMC standards according to EN 55011 for industrial, scientific and medical (ISM) radio-frequency equipment and EN 55022 for Industrial Technology Equipment (ITE) radio-frequency equipment. In addition, only recognized Japanese capacitors are used. The MDS series come with both medical and ITE safety approvals including UL/cUL/CCC/CE and CB certification and are fully compliant with RoHS Directive 2011/65/EU for environmental protection. Model Information Medical AC-DC Adapter Model Number Input Voltage Range Output Voltage Output Current MDS-150AAS12 F/B 90-264Vac 12Vdc 10A Model Numbering MDS Delta Medical power Supply 1 - 150 AAS 12 Max wattage in the product series. Maybe lower at some voltage. 150 → 150W Family Code Output Voltage Single Output: 12 for 12V All parameters are specified at 25°C ambient unless otherwise noted. www.DeltaPSU.com (November 2016, Rev. 03) F/B B TECHNICAL DATASHEET ``` MDS Medical AC-DC Adapter 12Volt, 120Watt / MDS-150AAS12 F/B Specifications Input Ratings / Characteristics Nominal Input Voltage 100-240Vac Input Voltage Range 90-264Vac Nominal Input Frequency 50-60Hz Input Frequency Range 47-63Hz Input Current (max.) 4A @ 115Vac, 2A @ 230Vac Efficiency (typ.) 88.1%, Reference Fig.1 Average Efficiency (min) 88% @ 115Vac and 87% @ 230Vac Standby Power (max.) 0.21W @ 115Vac and 0.5W @ 230Vac Inrush Current (typ.) 50A @ 115Vac, 100A @ 230Vac Touch Current (max.) 0.1mA @ 264Vac NC , 0.3mA @ 264Vac SFC 1) 2) 1) NC: normal condition 2) SFC: single fault condition Fig.1 Efficiency versus output load Output Ratings / Characteristics Nominal Output Voltage 12Vdc Total Regulation ± 5% Output Current 10A Output Power 120W Line Regulation (max) ±0.5% Load Regulation (max) ±4.5% Ripple & Noise (typ.) 104.0mV pk-pk @ Full load, Reference Fig. 2, Start-up Time (max) 3000ms Hold-up Time (min) 20ms @ 115Vac&230Vac Dynamic Response (Overshoot & Undershoot O/P Voltage) ± 5% @ 50-100% load 2 All parameters are specified at 25°C ambient unless otherwise noted. www.DeltaPSU.com (November 2016, Rev. 03) TECHNICAL DATASHEET ``` MDS Medical AC-DC Adapter 12Volt, 120Watt / MDS-150AAS12 F/B (a) 115V (b) 230V Fig. 2 Ripple & Noise example, 20MHz BW Ripple & Noise measurement circuit PSU Load Scope 0.47uF Ceramic 10uF Electrolytic 5cm Test Fixture Mechanical Case Chassis PC Case Cover PC Dimensions(W x Lx H) 85.0x 170.0x40.0 mm Unit Weight 1.1kg (39.6 ounces) Indicator NA Cooling System Terminal NA Input Output C14 socket 4pin Din / Barrel type / Tuning fork type Environment Surrounding Air Temperature Operating Storage 0°C to +40°C -40°C to +85°C Operating Humidity 10-95% RH (Non-Condensing) Operating Altitude 3,000 meters (9842.52 feet) Shock Test (Non-Operating) 50G, 11ms, 3 shocks for each direction Vibration (Operating) 5-500Hz, 2.09Grms, 20 minute for each three axis 3 All parameters are specified at 25°C ambient unless otherwise noted. www.DeltaPSU.com (November 2016, Rev. 03) TECHNICAL DATASHEET ``` MDS Medical AC-DC Adapter 12Volt, 120Watt / MDS-150AAS12 F/B Protections Overvoltage (max) 150%, Latch Mode Over load / Over current (max) 130% of rated load current, Hiccup Mode, (Non-Latching, Auto-Recovery) Over Temperature Hiccup Mode, (Non-Latching, Auto-Recovery) Short Circuit Hiccup Mode, (Non-Latching, Auto-Recovery) Protection Against Shock Class I for F series Class II with functional earth for B series Reliability Data MTBF (typ.) 1 Million Hours based on Telecordia SR-332 Safety Standards / Directives rd rd Medical Safety IEC60601-1 3 and 3.1 edition CB report IEC60601-1 edition 3.1rd (2012), EN60601-1 (2006) + A11 + A1 + A12, CAN/CSA-C22.2 NO. 60601-1:14, ANSI/AAMI ES60601-1:2005/(R)2012 ITE Safety IEC60950-1 (Ed.2,2005), GB4943.1-2011, GB92542008, GB17625.1-2003 CE MDD Directive 93/42/EEC Material and Parts RoHS Directive 2011/65/EU Compliant Galvanic Isolation Input to Output 4000 Vac Input to Ground 1500 Vac Output to Ground 4 500 Vac All parameters are specified at 25°C ambient unless otherwise noted. www.DeltaPSU.com (November 2016, Rev. 03) TECHNICAL DATASHEET ``` MDS Medical AC-DC Adapter 12Volt, 120Watt / MDS-150AAS12 F/B EMC (Compliant with IEC 60601-1-2 4th Ed. Requirements) EMC / Emissions Harmonic Current Emissions EN55011/EN55022,FCC Title 47:Class B IEC61000-3-2 Meet Class A limit Immunity to Voltage Flicker IEC61000-3-3 Electrostatic Discharge IEC61000-4-2 Level 4 Criteria A Air Discharge: 15kV Contact Discharge: 8kV Radiated Field IEC61000-4-3 Electrical Fast Transient / Burst IEC61000-4-4 Criteria B 80MHz-2700MHz, 10V/m AM modulation 385MHz-5785MHz, 28V/m Pulse mode and other modulation 1) Level 3 Criteria A :2kV Surge IEC61000-4-5 Level 3 Criteria A 3) Common Mode : 2kV 4) Differential Mode : 1kV Conducted IEC61000-4-6 Level 2 Criteria A 150kHz-80MHz, 3Vrms, 6Vrms at ISM bands and 1) 2) 1) 1) Amateur radio bands Power Frequency Magnetic Fields Voltage Dips IEC61000-4-8 IEC61000-4-11 1) Criteria A Magnetic field strength 30A/m 1) Criteria A 0% UT, 0.5 cycle (10ms) o o o o o o o o o , 0 /45 /90 /135 /180 /225 /270 /315 /360 1) Criteria A o 0% UT, 1 cycle (20ms), 0 1) Criteria A o 70% UT, 25 cycle (500ms), 0 2) Criteria B o 0% UT, 250 cycle (5000ms), 0 1) Criteria A: Normal performance within the specification limits 2) Criteria B: Output out of regulation, or shuts down during test. Automatically restore to normal operation after test. 3) Asymmetrical: Common mode (Line to earth) 4) Symmetrical: Differential mode (Line to line) 5 All parameters are specified at 25°C ambient unless otherwise noted. www.DeltaPSU.com (November 2016, Rev. 03) TECHNICAL DATASHEET ``` MDS Medical AC-DC Adapter 12Volt, 120Watt / MDS-150AAS12 F/B Block Diagram Dimensions W x L x H: 85.0 x 170.0 x 40.0 mm mm Item Device Description 6 1 Cover 2 Chassis 3 Power Cord 4 Socket All parameters are specified at 25°C ambient unless otherwise noted. www.DeltaPSU.com (November 2016, Rev. 03) TECHNICAL DATASHEET ``` MDS Medical AC-DC Adapter 12Volt, 120Watt / MDS-150AAS12 F/B Functions Start-up Time The time required for the output voltage (Vo) to reach 90% of its set value, after the input AC voltage is applied. Rise Time The time required for the output voltage (Vo) to change from 10% to 90% of its steady state value. Hold-up Time Hold up time is the time when the AC input collapses and output voltage retains regulation for a certain period of time. The time required for the output to reach 95% of its set value, after the input voltage is removed. ■ Graph illustrating the Start-up Time, Rise Time, and Hold-up Time 7 All parameters are specified at 25°C ambient unless otherwise noted. www.DeltaPSU.com (November 2016, Rev. 03) TECHNICAL DATASHEET ``` MDS Medical AC-DC Adapter 12Volt, 120Watt / MDS-150AAS12 F/B Dynamic Response Overload & Over current Protections The power supply output voltage will remain within ±5% of its steady state value, when subjected to a dynamic load change from 50 to 100% of its rated current. 100% Load The power supply’s Overload (OLP) and Over current (OCP) Protections will be activated when output current is between 110% and 130% of IO (Max load). Upon such an occurrence, VO will start to drop. Once the power supply has reached its maximum power limit, the protection will be activated. and the power supply will go into “Hiccup mode” (Auto-Recovery). The power supply will recover once the fault condition causing the OLP and OCP is removed and IO is back within the specified limit. Output Voltage (Vo) Load > 130% (Output Hiccup) 50% Load 50ms 50ms 50ms 50ms T (sec) 0% Load Inrush Current Inrush current is the input current that occurs when the input voltage is first applied. For AC input voltages, the maximum peak value of inrush current will occur during the first half cycle of the applied AC voltage. This peak value decreases exponentially during subsequent cycles of AC voltage. 130% Load T (sec) Additionally, if the IO is <130% but >110% for a prolong period of time (depending on the load), the Over Temperature Protection (OTP) will be activated due to high temperature on critical components. The power supply will then go into hiccup mode until the fault is removed; and, the input voltage is removed, then reapplied. Over Temperature Protection Overvoltage Protection As mentioned above, the power supply also has Over Temperature Protection (OTP). This is activated when the overload condition persists for an extended duration and the output current is below the overload trigger point but >100% load. In the event of a higher operating condition at 100% load, the power supply will run into OTP when the surrounding air temperature is higher than the operating temperature. When activated, the output voltage will go into hiccup mode until the input voltage is removed; then, reapplied, and the surrounding air temperature drops to its normal operating temperature. The power supply’s overvoltage circuit will be activated when its internal feedback circuit fails. The output voltage shall not exceed its specifications defined on Page 3 under “Protections”. Power supply will latch off, and require removal/re-application of input AC voltage in order to restart. Short Circuit Protection The power supply’s output OLP/OCP function also provides protection against short circuits. When a short circuit is applied, the output current will operate in “Hiccup mode”, as shown in the illustration in the OLP/OCP section on this page. The power supply will return to normal operation after the short circuit is removed. 8 All parameters are specified at 25°C ambient unless otherwise noted. www.DeltaPSU.com (November 2016, Rev. 03) TECHNICAL DATASHEET ``` MDS Medical AC-DC Adapter 12Volt, 120Watt / MDS-150AAS12 F/B Certificate . All Delta Medical Power products conform to the European directive 2011/65/EU. RoHS is the abbreviation for “Restriction of the use of certain hazardous substances. Delta has been certified as meeting the requirement of ISO 13485: 2003 and EN ISO 13485:2012 for the design and manufacture of switching power supply and adaptor for medical device. Delta is approved for the UL Total Certification Program (TCP) approved client laboratory for IEC60950 and IEC60065. Delta also has participated UL Client Test Data Program (CTDP) for IEC 60601. Meet DOE Level VI and ErP Lot 7 Efficiency Requirement 9 All parameters are specified at 25°C ambient unless otherwise noted. www.DeltaPSU.com (November 2016, Rev. 03)