SL Power MINT3110A1908K01 110w triple output medical sery Datasheet

Features
2” x 4” x 1.3” Package
Suitable for 1U Applications
Class I and Class II versions
110W w/air, 80W convection cooled
Universal Input 90-264Vac
Efficiency 87% typical
Approved to CSA/EN/IEC/UL60601-1, 3rd
Edition
2 x MOPP Isolation
Power Fail Signal
Remote Sense (optional)
3 Year Warranty
RoHS Compliant
Description
The MINT3110 Series are medical grade, high efficiency, triple output power supplies in a small 2” x 4” size. The
MINT3110 family is ideal for medical applications in surgical and laboratory instruments requiring low leakage, multiple
rd
outputs and small footprint. Surpassing the strict requirements of IEC 60601-1 3 edition, the MINT3110 models feature
outputs of 5/±12V, 5/±15V, 5/±24V, Power Fail/DC OK signal, very low earth leakage current of 200µA, Class B
Conducted EMI, 2 MOPP (means of patient protection) isolation, and are available in Class I and Class II configurations.
Model Selection
Model
Number
Volts*
V1
MINT3110A0508K01
MINT3110A1708K01
MINT3110A1908K01
Notes:
Output Current**
w/200LFM air
Convection
Ripple &
Noise***
Total
Regulation
OVP
Threshold
±2%
7.5V max.
5V
14.0A
10.0A
1.0% pk-pk
V2
12V
6.0A
4.5A
1.0% pk-pk
±3%
115%-135%
V3
-12V
1.0A
1.0A
2.0% pk-pk
±10%
115%-135%
V1
5V
14.0A
10.0A
1.0% pk-pk
±2%
7.5V max.
V2
15V
4.5A
3.5A
1.0% pk-pk
±3%
115%-135%
V3
-15V
1.0A
1.0A
2.0% pk-pk
±10%
115%-135%
V1
5V
12.0A
8.0A
1.0% pk-pk
±2%
7.5V max.
V2
24V
4.0A
3.0A
1.0% pk-pk
±3%
115%-135%
V3
-24V
1.0A
1.0A
2.0% pk-pk
±10%
115%-135%
* 5V output is adjustable with +/-10% range
** Total convection power is 80 Watts.
*** Measured with noise probe directly across output terminals, and load terminated with 0.1µF ceramic and 10µF low ESR capacitors. Ripple & Noise of V2 at no
load is 2% maximum.
General Specifications
AC Input
Input Current
100-240Vac, ±10%, 47-63Hz, 1∅
120–370Vdc
Turn On Time
Less than 2 sec. @115Vac (inversely
proportional to input voltage and thermistor
temperature)
115Vac: 1.5A, 230Vac: 0.75A
Hold-up Time
16mS typical at 110W, 120Vac input
MINT3110
110W Triple Output
Medical Series
General Specifications (continued)
Inrush Current
Input Fuses
Earth Leakage
Current
Efficiency
Output Power
Transient
Response
Signals
264Vac, cold start: will not exceed 45A
Overload
Protection
F1, F2: 2.5A, 250Vac fuses provided on all
models
Short Circuit
Protection
<290µA@264Vac, 60Hz, NC
Overvoltage
Protection
87% typical at 230Vac
110W continuous with 200 lfm airflow, 80W
convection cooled – See chart for specific
voltage model ratings.
Switching
Frequency
500µS typ. for return to within 0.5% of
nominal, 50% load step. ∆i/∆t <0.2A/µS.
Max Volt Deviation = 3%
Isolation
AC Power Fail, DC OK
150%-300% above rating for V2 & V3, 110%200% for V1. Hiccup Mode
Provided - no damage will occur if the output is
shorted.
See models chart for trip range.
PFC: Variable 30-400kHz.
Main Converter: Variable 35-180kHz, 65-70kHz
at full load.
Input-Output: 4000Vac, 2 x MOPP
Input-Ground: 1800Vac, 1 x MOPP
Output-Ground: 500Vac
Ripple and Noise
See models chart
Operating
Temperature
-10°C to +70°C
Output Voltage
See models chart
Temperature
Derating
Derate output power linearly above 50°C to 50%
at 70°C
+/-10% from nominal on 5V output
Storage
Temperature
-40°C to +85°C
Voltage
Adjustability
Minimum Load
Total Regulation
Vibration
Dimensions
Weight
Altitude
Not required
Relative Humidity
See models chart
Operating: 0.003g²/Hz, 1.5grms overall, 3
axes, 10 min/axis
Non-Operating: 0.026g²/Hz, 5.0grms overall,
3 axes, 1 hr/axis
Shock
Safety Standards
W: 2.0” x L: 4.0” x H: 1.3”
MTBF
200g
Operating: -500 to 10,000 ft.
Non-operating: -500 to 40,000 ft.
5% to 95%, non-condensing
Operating: Half-sine, 20gpk, 10ms, 3 axes, 6
shocks total
Non-Operating: Half-sine, 40 gpk, 10 ms, 3
axes, 6 shocks total
EN/CSA/UL/IEC 60601-1, 3rd Edition
245,000 hours, 25°C Ambient, 110Vac input
Auxiliary Signals
AC Power Fail:
During normal operation, stays HIGH. Signal
goes LOW with at least 6mS warning before
loss of DC output from AC failure.
Remote Sense:
(5V output, optional) Will compensate for 0.5V
drop min. Will operate without remote sense
connected. Reverse connection protected.
DC OK:
Open collector logic signal goes and stays HIGH
100mS to 500mS after main output reaches
regulation.
Notes:
1. Specifications are for convection rating at factory settings at 115 Vac input, 25°C ambient unless otherwise stated.
2. For DC input an external DC safety rated fuse must be used.
www.slpower.com
MINT3110 REV 1.4, 19-SEP-14
2
MINT3110
110W Triple Output
Medical Series
Connector Information
Input Connector
J100
Ground
J101
PIN 1) AC NEUTRAL
PIN 2) EMPTY
PIN 3) AC LINE
0.187” FASTON TAB
Mating Connector:
Molex 09-50-3031
Pins= 08-52-0072
AMP #640250-3
Pins= 3-640706-1
Mating Connector:
Molex 01-90020001
DC Output Connector
J200
PIN 1) +V1
PIN 2) +V1
PIN 3) GND
PIN 4) GND
PIN 5) GND
PIN 6) GND
Signal Connector
J300
PIN 1) Power Fail/DC OK
PIN 2) GND
PIN 3) +Remote Sense
PIN 4) –Remote Sense
PIN 7) V2
PIN 8) V3
Mating Connector:
Amp #640250-8
Pins=Amp #3-640706-1
Mating Connector:
Amp #1375820-4
Pin= Amp #1375819
Isolation Specifications
Parameter
Conditions/Description
Insulation Safety Rating
Input/Ground
Input/Output
Output/Ground
Electric Strength Test Voltage
Input/Ground
Input/Output
Output/Ground
Min
Nom
Max
Units
Basic (1 MOPP)
Reinforced (2 MOPP)
Operational
1800
4000
500
-
Vac
Vac
Vac
-
Leakage Current
Parameter
Conditions/Description
Max
Earth Leakage Current
Normal Condition (NC)
Single Fault Condition (SFC)
290µA
420µA
Touch Current
Normal Condition (NC)
Single Fault Condition (SFC)
90µA
170µA
Input Specifications
All specifications apply over specified input voltage, output load, and temperature range, unless otherwise noted.
Parameter
Conditions/Description
Input Voltage
Input Frequency
Input Current
Min
Nom
Max
Units
90
115/230
264
Vac
47
50/60
115Vac/max load
63
Hz
1.5
A
Input Current
230Vac/max load
0.75
A
Inrush Current
264Vac, cold start, 25°C
-
-
45
A
Efficiency
Vi nom, Io nom
MINT3110A0508K01
MINT3110A1708K01
MINT3110A1908K01
-
87%
-
%
Parameter
Conditions/Description
Min
Nom
Max
Units
Output Current V1
Output Current V2
Output Current V3
MINT3110A0508K01
0
0
0
10.0
4.5
1
14.0
6.0
1
ADC
Output Current V1
Output Current V2
Output Current V3
MINT3110A1708K01
0
0
0
10.0
3.5
1
14.0
4.5
1
ADC
Output Current V1
Output Current V2
Output Current V3
MINT3110A1908K01
0
0
0
8.0
3.0
1
12.0
4.0
1
ADC
Output Specifications
www.slpower.com
MINT3110 REV 1.4, 19-SEP-14
4
MINT3110
110W Triple Output
Medical Series
Output Specifications (continued)
Parameter
Conditions/Description
Min
Nom
Max
Units
Static Line Regulation V1
Vi min-Vi max, Vi nom, 0-100% Io1 max
-2
-
2
% Vo nom
Static Line Regulation V2
Vi min-Vi max, Vi nom, 0-100% Io2 max
-3
-
3
% Vo nom
Static Line Regulation V3
Vi min-Vi max, Vi nom, 0-100% Io3 max
-10
-
10
% Vo nom
Static Load Regulation V1
(Droop Characteristic)
Vi min-Vi max, Vi nom, 0-100% Io1 max
-2
-
2
% Vo nom
Static Load Regulation V2
(Droop Characteristic)
Vi min-Vi max, Vi nom, 0-100% Io2 max
-3
-
3
% Vo nom
Static Load Regulation V3
(Droop Characteristic)
Vi min-Vi max, Vi nom, 0-100% Io3 max
-10
-
10
% Vo nom
Hold-Up Time
Vin = 120Vac, Po = 110W
16
-
-
mS
Dynamic Load Regulation V1, V2, V3
Load change =50%, di/dt = 0.2A/µS
0
3
% Vo nom
Start-Up Time
Vin = 115Vac, Io nom
0
2
S
Ripple & Noise V1
20MHz bandwidth
0
1%
% Vo nom
Ripple & Noise V2
20MHz bandwidth
0
1%
% Vo nom
Ripple & Noise V3
20MHz bandwidth
0
2%
% Vo nom
Max
Units
-
Protection
All specifications apply over specified input voltage, output load, and temperature range, unless otherwise noted.
Parameter
Conditions/Description
Min
Input Fuse
T2.5A/250V internal fuse in both line & neutral
Input Transient Protection
2KV(CM) and 1KV(DM) surge
Nom
Not user accessible
2
Short Circuit Protection
KV (CM)
Hiccup Mode
Overload Protection
Hiccup Mode
Overvoltage Protection
Latching Type, recycle AC input to reset
See models chart for trip ranges
Characteristic Curves
Output vs. Temperature
80W convection cooled and 110W continuous with 200 LFM airflow. Derate output power to 50% at 70°C.
120
Output Power (W)
100
80
60
40
20
0
-10
0
10
20
30
40
50
60
70
Temperature (°C)
www.slpower.com
Convection
200LFM Cooling
MINT3110 REV 1.4, 19-SEP-14
5