XP Power IML0215S15 Dc-dc converter Datasheet

IML02 Series
DC-DC Converter
2 Watts
xxx Series
• World Wide Medical Approvals
• Single and Dual Outputs
• SIP7 Package
• -40 °C to +85 °C Operation
• Full Load at 85 °C Ambient
• 4000 VAC Isolation, 1 x MOPP at 250 VAC
• 2 µA Patient Leakage Current
• MTBF 2.5 MHrs
• 3 Year Warranty
Dimensions:
IML02:
0.77 x 0.39 x 0.49” (19.5 x 9.8 x 12.5 mm)
Models & Ratings
Input Voltage
4.5-5.5 V
10.8-13.2 V
13.5-16.5V
21.6-26.4V
Output Voltage
Output Current
3V3
5V
9V
12 V
15 V
±3V3
±5 V
±9 V
±12 V
±15 V
3V3
5V
9V
12 V
15 V
±3V3
±5 V
±9 V
±12 V
±15 V
3V3
5V
9V
12 V
15 V
±3V3
±5 V
±9 V
±12 V
±15 V
3V3
5V
9V
12 V
15 V
±3V3
±5 V
±9 V
±12 V
±15 V
600 mA
400 mA
222 mA
167 mA
133 mA
±300 mA
±200 mA
±111 mA
±83 mA
±66 mA
600 mA
400 mA
222 mA
167 mA
133 mA
±300 mA
±200 mA
±111 mA
±83 mA
±66 mA
600 mA
400 mA
222 mA
167 mA
133 mA
±300 mA
±200 mA
±111 mA
±83 mA
±66 mA
600 mA
400 mA
222 mA
167 mA
133 mA
±300 mA
±200 mA
±111 mA
±83 mA
±66 mA
Input current
No Load
Full Load
50
60
60
60
60
50
60
60
60
60
25
30
30
30
30
25
30
30
30
30
25
30
30
30
30
25
30
30
30
30
20
20
20
20
20
20
20
20
20
20
mA
mA
mA
mA
mA
mA
mA
mA
mA
mA
mA
mA
mA
mA
mA
mA
mA
mA
mA
mA
mA
mA
mA
mA
mA
mA
mA
mA
mA
mA
mA
mA
mA
mA
mA
mA
mA
mA
mA
mA
533 mA
533 mA
500 mA
482 mA
488 mA
519 mA
513 mA
500 mA
482 mA
488 mA
217 mA
208 mA
208 mA
200 mA
203 mA
214 mA
208 mA
208 mA
200 mA
203 mA
174 mA
170 mA
167 mA
167 mA
167 mA
170 mA
170 mA
167 mA
167 mA
167 mA
111 mA
106 mA
104 mA
104 mA
104 mA
107 mA
106 mA
104 mA
104 mA
104 mA
Notes
Input currents measured at nominal input voltage.
1
www.xppower.com
Max. capacitive load
1000
470
470
220
220
±470
±220
±220
±100
±100
1000
470
470
220
220
±470
±220
±220
±100
±100
1000
470
470
220
220
±470
±220
±220
±100
±100
1000
470
470
220
220
±470
±220
±220
±100
±100
µF
µF
µF
µF
µF
µF
µF
µF
µF
µF
µF
µF
µF
µF
µF
µF
µF
µF
µF
µF
µF
µF
µF
µF
µF
µF
µF
µF
µF
µF
µF
µF
µF
µF
µF
µF
µF
µF
µF
µF
Efficiency
Model Number
75%
78%
80%
83%
82%
77%
78%
80%
83%
82%
76%
80%
80%
83%
82%
78%
80%
80%
83%
82%
76%
78%
80%
80%
80%
78%
78%
80%
80%
80%
75%
78%
80%
80%
80%
77%
78%
80%
80%
80%
IML0205S3V3
IML0205S05
IML0205S09
IML0205S12
IML0205S15
IML0205D03
IML0205D05
IML0205D09
IML0205D12
IML0205D15
IML0212S3V3
IML0212S05
IML0212S09
IML0212S12
IML0212S15
IML0212D03
IML0212D05
IML0212D09
IML0212D12
IML0212D15
IML0215S3V3
IML0215S05
IML0215S09
IML0215S12
IML0215S15
IML0215D03
IML0215D05
IML0215D09
IML0215D12
IML0215D15
IML0224S3V3
IML0224S05
IML0224S09
IML0224S12
IML0224S15
IML0224D03
IML0224D05
IML0224D09
IML0224D12
IML0224D15
IML02 Series
DC-DC Converter
Input
Characteristic
Minimum
Typical
Maximum
Input Voltage Range
10.8
13.2
13.5
16.5
21.6
Input Reflected Ripple Current
Units
5.5
4.5
VDC
26.4
20
Input Surge
16.5
12 V nominal
15 V nominal
24 V nominal
mA pk-pk
5.5
13.2
Notes & Conditions
5 V nominal
VDC for 100 ms
26.4
Through 12 µH inductor and 47 µF capacitor
5 V models
12 V models
15 V nominal
24 V nominal
Output
Characteristic
Minimum
Output Voltage
3.3
Typical
Initial Set Accuracy
Minimum Load
Maximum
Units
Notes & Conditions
30
VDC
See Models and Ratings table
±5
%
10
At full load
%
Minimum load required to meet specified regulation
Output changes by max of 1.2% for each 1% change in
input voltage
Line Regulation
±1.2
%/1%
Load Regulation
10
%
From 10% to full load, see application note
%
On dual output models, when one output is at 25% load
and other is varied from 10% load to full load
mV pk-pk
20 MHz bandwidth. Measured using 10 µF electrolytic in
parallel with 0.1 µF ceramic capacitor
Cross Regulation
±4
Ripple & Noise
150
Short Circuit Protection
Continuous
Maximum Capacitive Load
See Models and Ratings table
Temperature Coefficient
0.03
%/ ˚C
Maximum
Units
General
Characteristic
Minimum
Efficiency
Isolation: Input to Output
Typical
80
4000
VAC
Patient Leakage Current
Isolation Resistance
Switching Frequency
10
50
Power Density
Mean Time Between Failure
2
µA
20
100
pF
kHz
13.6
W/in3
109
Isolation Capacitance
2.5
Weight
Notes & Conditions
%
Ω
MHrs
0.009 (4.2)
1 x MOPP at 250 VAC working voltage, 2 x MOPP at
125 VAC working voltage
MIL-HDBK-217F, +25 °C GB
lb (g)
Environmental
Characteristic
Minimum
Operating Temperature
-40
Storage Temperature
-40
Case Temperature
Humidity
2.5
Typical
Maximum
Units
+85
°C
+125
°C
+100
°C
95
%RH
Cooling
Notes & Conditions
No derating
Non-condensing
Natural convection
www.xppower.com
2
IML02 Series
DC-DC Converter
EMC: Emissions
Phenomenon
Standard
Test Level
Notes & Conditions
Conducted
EN55011
Class B
See Application Note
Radiated
EN55011
Class B
EMC: Immunity
Phenomenon
Standard
Test Level
Criteria
ESD Immunity
EN61000-4-2
±15 kV
A
Notes & Conditions
Radiated Immunity
EN61000-4-3
10 Vrms
A
EFT/Burst
EN61000-4-4
2 kV
A
External components required,
see application notes
Surge
EN61000-4-5
2 kV
A
External components required,
see applications note
Conducted Immunity
EN61000-4-6
10 V rms
A
Magnetic Fields
EN61000-4-8
30 A/m
A
Air Discharge
Safety Approvals
Safety Agency
Safety Standard
Notes & Conditions
ANSI/AMMI ES60601-1
UL
CSA
CSA C22.2 No. 60601-1
TUV
EN60601-1
CB
IEC60601-1
Mechanical Details
0.02
(0.5)
0.77 (19.5)
0.47
(12.0)
0.02
(0.5)
0.39
(9.8)
1
0.10
(2.54)
2
5
6
7
0.12
(3.0)
0.39 (9.8)
Printed
Face
0.09
(2.4)
0.49
(12.5)
0.01
(0.25)
0.08
(2.13)
0.02
(0.5)
0.60 (15.24)
Pin Connections
Notes
1. All dimensions are in inches (mm)
2. Weight: 0.009 lbs (4.2 g) approx.
3. Pin diameter: 0.02±0.002 (0.5±0.05)
3
4. Pin pitch tolerance: ±0.014 (±0.35)
5. Case tolerance: ±0.02 (±0.5)
www.xppower.com
Pin
1
2
5
6
7
Single
+Vin
-Vin
-Vout
No Pin
+Vout
Dual
+Vin
-Vin
-Vout
Common
+Vout
3
IML02 Series
DC-DC Converter
Application Note
Regulation
Hig
hl
Typ
ica
Low
imit
l Lo
ad L
ine
Lim
it
EFT and Surge Filter
Input components (C1, C2, D1) are used to help meet EFT and surge test requirements for the module.
IML02
C1
C2
D1
IML0205XXXX
1000 µF/35 V
330 µF/50 V
SMDJ9.0A
IML0212XXXX
1000 µF/35 V
330 µF/50 V
SMDJ13A
IML0215XXXX
1000 µF/35 V
330 µF/50 V
SMDJ18A
IML0224XXXX
1000 µF/35 V
330 µF/50 V
SMDJ28A
EMI Filter
Input filter components (C1,C2 and L) are used to help meet conducted emissions requirements for the module. These components should be
mounted as close as possible to the module, and all leads should be minimised to decrease radiated noise.
C1
IML02
C2
L
IML0205XXXX
1206, 4.7 µF/ 50 V
IML0212XXXX
1206, 4.7 µF/ 50 V
1206, 4.7 µF/ 50 V
6.8 µH
IML0215XXXX
1206, 4.7 µF/ 50 V
1206, 4.7 µF/ 50 V
6.8 µH
IML0224XXXX
1206, 4.7 µF/ 50 V
1206, 4.7 µF/ 50 V
6.8 µH
www.xppower.com
6.8 µH
21 July 16
Similar pages