MENB1030 Universal 30 Watt Series Medical Switch-Mode Power Supply 3 Year Warranty •100-240VAC Universal Input •Meets EISA2007, CEC Efficiency Level V, EU (EC) No 278/2009 Phase II •Certified to UL/EN60601-1 •5V to 48V outputs, up to 32W •Modified and Custom Designs Available •Regulated Output with Low Ripple •Impact-Resistant Polycarbonate Enclosure •No load Power Consumption <0.3W •Desktop & Wall Plug Styles RoHS Specifications All Specifications are typical at nominal input, full load at 25°C unless otherwise stated. AC Input Input Current 100-240VAC, -/-10%, 47-63 Hz, 1∅ 100VAC: 0.9A Inrush Current Input Fuse Efficiency Output Voltage 60A peak, 264Vac, cold start 3.15A Internal Primary Current Fuse is provided Meets EISA2007, CEC Efficiency Level V, and EU(EC) No. 278/2009 Phase II requirements See chart Output Power See chart Ripple and Noise 1% pk-pk max., 20MHz BW Line & Load Voltage Regulation Transient Response Line: +/- 1%, Load: +/-5% 500µs max., 50% load step, typical Minimum Load Not required Case Material: Black 94V0 Polycarbonate Case Dimensions: Output Connector: 95 x 54 x 31mm. See outline drawing Weight = 200g 2.5mm barrel type (Ault #3) connector, center contact (+). Other options available MTBF >100,000 hours calculated Hold-up Time 18 ms min. @ 115VAC, 60 msec @ 230VAC Turn-on Time 2 seconds, max. Topology Switching – Fixed Frequency Flyback Overload Protection Hiccup Mode Short Circuit Protection Hiccup Mode EMC See chart below Safety Standards Dielectric Withstand Operating Temperature Storage Temperature Relative Humidity EN/IEC/CSA/UL60601-1 Input-Output: 5,656VDC, Input-GND: 1,500 Vac, Output-GND: 500Vdc 0° to 40°C, no derating -30 to +85°C 5% to 95%, non-condensing Altitude Output Cable: 0 to 10,000 ft 5V – 9V models: #18AWG (SPT-1), 1,500mm, 2 conductor 12V – 24V models: #18AWG (UL1185), 1,500mm, 2 conductor EMC Specifications Conducted Emissions Radiated Emissions Line Frequency Harmonics Voltage Fluctuations/Flicker Static Discharge Immunity Radiated RF Immunity EFT/Burst Immunity Line Surge Immunity Conducted RF Immunity Power Frequency Magnetic Field Immunity Voltage Dip Immunity EN55011 Class B, FCC Part 15, Class B. EN55011 Class B, FCC Part 15, Class B. EN61000-3-2, Class A EN61000-3-3 EN61000-4-2, 6kV Contact Discharge, 8kV air discharge EN61000-4-3, 3V/m. EN61000-4-4, 2kV/5kHz. EN61000-4-5, 1kV differential, 2kV common-mode EN61000-4-6, 3Vrms EN61000-4-8, 3A/m EN61000-4-11, Criteria A ______________________________________________________________________________________________________________________________ 1 of 2 SL Power Electronics Corp • 6050 King Drive Ventura, CA 93003 • Phone 805.486.4565 • Fax 858.712.2040 • Email: [email protected] • www.slpower.com MENB1030 Universal 30 Watt Series Medical Switch-Mode Power Supply Model Number 3 Year Warranty Volts (V) Output Current (max) Max Watts Ripple (Vp-p max) 5V 7.5 V 9V 12 V 14 V 15 V 18 V 24 V 48 V 4.00 A 3.00 A 3.00 A 2.50 A 2.10 A 2.00 A 1.67 A 1.33 A 0.63 A 20.0 W 22.5 W 27.0 W 30.0 W 29.4 W 30.0 W 30.0 W 31.9 W 30.0 W 50 mV 75 mV 90 mV 120 mV 140 mV 150 mV 180 mV 240 mV 480 mV MENB1030A0503F01 MENB1030A0703F01 MENB1030A0903F01 MENB1030A1203F01 MENB1030A1403F01 MENB1030A1503F01 MENB1030A1803F01 MENB1030A2403F01 MENB1030A4803F01 Notes: Part numbers above include #3 output connector and IEC320 C14 grounded input receptacle. See below for other options. Model Number Key MENB 1 030 A VV 03 F 01 Model Input Connector: Output Connector: Output Voltage: Model Output Pow er: # of Outputs Product Fam ily: "01" = Standard. "02" and higher indicates a m odified m odel. "F" = IEC320 C14 grounded, Other options available, see below "03" = 2.5m m Barrel Type Connector. Other options available, contact factory. "05" = 5Vdc, "12" = 12Vdc, "24" = 24Vdc, etc. "A" = Original Configuration "030" = 30 Watts "1" = Single Output "M" = Medical, "E" = External, "NB" = Model Series Designator AC Input Receptacle Options Desktop IEC320 - C14 Class I Grounded IEC320 - C18 Class II Ungro unde d IEC320 - C8 Class II "Shaver" (F) (Q) (N) Notes: Output Connector Wall-Plug N. Am erica Japan N. Am erica Japan Int e rc ha nga ble Fixed (C) (B) Europe Australia Fixed United Kingdom Fixed (M) (G) (H) Fixed 1. For desktop options, choose applicable letter above. Notes: 2. For Wall-Plug options, choose the applicable letter above and the unit w ill be fitted w ith the receptacle chosen. The N. A. blade version (C) w ill be an interchangeable blade. The other options (B), (M), (G), and (H) w ill be fixed blades m olded in the case. 3. Blade Kit is available w hich w ill include one each of a EU, UK, and Australian blade. Kit part # is KT1027K. These can be used w ith the (C) version, to allow blades to be changed. 1. Ault #3 Connector, center contact (+). 2. Connector is Sw itchcraft 760 plug or equivalent. Mating connector is Sw itchcraft 712A or equivalent. 3. Other connector options are available. Contact your local SL Pow er Representative for details. Outline Drawings DATE CODE DATE CODE Desktop Style Wall-plug Style (“B” Option Shown) Data Sheet © 2011 SL Power Electronics Corp. The information and specifications contained herein are believed to be correct at the time of publication. Rev. 5-12-2011 However, SL Power accepts no responsibility for consequences arising from reproduction errors or inaccuracies. Specifications are subject to change without notice. ______________________________________________________________________________________________________________________________ 2 of 2 SL Power Electronics Corp • 6050 King Drive Ventura, CA 93003 • Phone 805.486.4565 • Fax 858.712.2040 • Email: [email protected] • www.slpower.com