Amphenol HTR400 Operates on standard line voltage Datasheet

®
Kaye ValProbe Series
Temperature Calibration and Reference
Features
• Operates on standard line voltage
• Positions for two IRTD standards
Fluid Baths
• Wide operating range covers most common
application requirements
• CTR-40 stability to ±0.005°C or CTR-80 stability
to ±0.01°C
• Rapid cool down from ambient to –40°C
• Very low noise
• Modest footprint with floor cart available for
portability
• Quick drain spout simplifies fluid change
Dry Wells
• HTR and LTR Series dry wells
• Rapid response time with no oils or fluids
• 120 minute cool-down from ambient to –80°C
• Stability of ±0.02°C to ±0.05°C for temperatures
exceeding 300°C
• Mounted on casters for portability
• Lightweight yet rugged design for portability
Amphenol
Advanced Sensors
Temperature Calibration
Temperature Calibration
CTR–40
Stable Uniform Heat Sources
The CTR–40 is a portable temperature bath designed
to meet the calibration and validation needs of the
ValProbe system. Combined with the Intelligent RTD
(IRTD) and ValProbe software, the CTR–40 provides prestudy and post-study verifications, as well as ValProbe
temperature sensor calibration.
Kaye temperature references are designed for
easy operation while delivering the highest level of
temperature stability possible. These stable uniform
heat sources combine rapid heat-up and cooldown with large sensor capacity to minimize overall
calibration time. Multiple calibration set points are
programmed via the easy-to-use operator panel and
displayed (set point or well temperature) to 0.01 degree
accuracy. These references provide fully automated
sensor calibration when used with Kaye’s Validator®
2000 and traceable IRTD temperature standard.
The advanced design combines excellent temperature
stability and uniformity with a broad temperature range
–40°C to 150°C to address most ValProbe applications.
The generous nine liter tank and specially designed
ValProbe immersion basket accommodate up to eight
ValProbe loggers, making calibration or verification a
quick and easy process.
HTR and LTR Series Dry Wells
The portable tabletop design easily fits onto a benchtop
without consuming precious space. An optional
floor cart, with locking casters, raises the unit to a
convenient operating height and allows easy transport
within your facility.
The HTR and LTR Series dry wells are specifically
designed for calibrating sensors used for process
validation. These are the most advanced reference
units on the market, featuring fast heat-up and cooldown, large well capacity to accommodate 18 to 24
thermocouples, and they use no messy oils or fluids.
Intelligent RTD Standard
The HTR 400 is ideal for high-temperature applications
such as autoclaves, dry heat ovens and sterilizer
tunnels. The LTR models offer low-temperature
performance for applications including freezers, cold
rooms, incubators and autoclaves. The LTR model
selection should be based on the application’s lowtemperature point.
The IRTD Temperature Standard (IRTD 400) is a
National Industry of Standards and Technology (NIST)
traceable instrument calibrated from –195°C to 420°C
with 0.025°C accuracy over the entire range. This
completely self-contained measurement system serves
as the secondary standard providing traceability for
ValProbe calibration or verification. By interfacing
with the ValProbe software, the IRTD 400 eliminates
the potential for human error, ensuring accurate and
traceable measurements.
ValProbe immersion basket shown with IRTD
Standard (not included).
CTR–80 Cryo Temperature Bath
Operating from –80°C to 30°C, the CTR delivers fast
response, high stability, and automated sensor
calibration for the most severe cold-temperature
applications. A generous 3.7 liter tank is heated and
cooled quickly and quietly by a two-stage refrigeration
system (R507 and R508B). The CTR–80 is the ideal unit
for calibrating temperature sensors used in freeze
dryer, freezer, and cryo unit validation.
LTR Series
ValProbe Series Specifications
CTR-40
Temperature Range
Fault Protection
Ambient Operating Range
•
•
•
•
Set-Point Accuracy
CTR-80
-40°C to 150°C
15°C to 25°
Oven temperature limits (user can set)
Low voltage cutout
Automatic refrigeration turn off
Electrical fuse
0.5°C
Temperature Range
Temperature Stability
-80°C to 100°C
• ±0.005°C at -40°C
• ±0.005°C at -25°C
• ±0.005°C at -150°C
Ambient Operating Range
Temperature Uniformity
Set-Point Accuracy
±0.01°C
Typical Cool-Down Time
25°C to -40°C, 110 minutes
Access Opening
94 mm x 172 mm with positions for IRTD and up to eight
ValProbe data loggers with immersion basket
Display
LED with 0.01°C resolution
Computer Interface
15°C to 25°
0.5°C
Temperature Stability
±0.01°C
Temperature Uniformity
±0.012°C
Typical Cool-Down Time
25°C to -80°C, 20 minutes
Access Opening
RS232
86 mm x 114 mm with positions for (two) IRTDs and
(three) 11 mm diameter x 203 mm deep calibration wells
Off Cart Dimensions (h x w x d)
Display
584 mm x 305 mm x 622 mm
On Cart Dimensions (h x w x d)
819 mm x 305 mm x 622 mm
Volume
9 liters
Recommended Bath Fluids
• -40°C to 130°C, Silicone oil type 200
(five centistoke viscosity)
• -30°C to 150°C, Silicone oil type 200
(10 centistoke viscosity)
LED with 0.01°C resolution
Computer Interface
RS232
Dimensions (h x w x d)
762 mm x 305 mm x 610 mm
Weight
57 kg
Power
115 VAC 60 Hz, 16 A or 230 VAC 50 Hz, 8 A 1700 W
Weight
32 kg
Power
115 VAC 60 Hz, 16 A or 230 VAC 50 Hz, 8 A 1700 W
Fault Protection
•
•
•
•
Oven temperature limits (user can set)
Low voltage cutout
Automatic refrigeration turn off
Electrical fuse
ValProbe Series Specifications
IRTD-400
Temperature Range
-195°C to 420°C
Accuracy Over Range
0.025°C
Accurate for one year, 0°C to 60°C ambient. Includes
calibration certificate with traceability to NIST.
Resolution
0.001°C
Sensor Element
200 Ω platinum RTD sensor
Sheath Material
Inconel 600
HTR 400
LTR –25/140
LTR –40/140
Temperature
Range
25°C above
ambient to 400°C
–25°C to 140°C
–40°C to 140°C
Ambient
Operating
Range
5°C to 50°C
5°C to 50°C
5°C to 50°C
Set-Point
Accuracy
0.2°C to 300°C
0.2°C
0.2°C
Temperature
Stability
0.02°C to 300°C
0.02°C
0.02°C
Transfer
Calibration
Accuracy*
50°C to 150°C:
±0.1°C
–25°C to 80°C:
±0.1°C
–40°C to –25°C:
±0.15°C
IRTD Standard to
Thermocouples
50°C to 250°C:
±0.2°C
80°C to 130°C:
±0.15°C
–25°C to 80°C:
±0.1°C
250°C to 350°C:
±0.3°C
130°C to 140°C:
±0.18°C
80°C to 130°C:
±0.15°C
Immersion Depth (Minimum)
101.6 mm
Calibration
0.3°C to 400°C
0.05°C to 400°C
350°C to 400°C:
±0.4°C
Ambient to 90°C:
5 minutes
Ambient to 80°C:
6 minutes
Ambient to 80°C:
6 minutes
90°C to 125°C:
3 minutes
Ambient to 140°C:
14 minutes
Ambient to 140°C:
14 minutes
Reference wells
(2): 6.7 mm
diameter x 127
mm deep
Reference wells
(2): 6.7 mm
diameter x 155
mm deep
Reference wells
(2): 6.7 mm
diameter x 155
mm deep
Calibration
wells (8): 9 mm
diameter x 155
mm) deep
Calibration
wells (6): 9 mm
diameter x 155
mm) deep
Calibration
wells (6): 9 mm
diameter x 155
mm) deep
Display
LED w/0.01°C
resolution
LED w/0.01°C
resolution
LED w/0.01°C
resolution
One reading per second
Computer
Interface
RS232
RS232
RS232
Environmental
Dimensions
343 mm x 198
mm x 317.5 mm
343 mm x 198
mm x 317.5 mm
343 mm x 198
mm x 317.5 mm
Weight
8.2 kg
13.6 kg
13.6 kg
Power
115 VAC 60 Hz, 6
A or 230 VAC 50
Hz, 3 A
700 watts
115 VAC 60 Hz, 3
A or 230 VAC 50
Hz, 1.5 A
350 watts
115 VAC 60 Hz, 3
A or 230 VAC 50
Hz, 1.5 A
350 watts
Fault Protection
Sensor burnout
protection, over
temperature
thermal cutout,
electrical fuse
Sensor burnout
protection, over
temperature
thermal cutout,
electrical fuse
Sensor burnout
protection, over
temperature
thermal cutout,
electrical fuse
±0.01°C
Kaye provides a re-certification service for calibrating
the temperature standard.
Typical Heat-Up
Time
130°C to 140°C:
±0.18°C
Power Probe
• Unregulated DC, 10 to 25 V
• 850 mW at 15 V for first probe
• 550 mW for each additional probe
350°C: 25
minutes
Well
Configuration
Power Supply
115 VAC US-style adaptor or 230 VAC VDE-approved
adaptor
Power supply is not required for use with the ValProbe
system.
Measurement Rate
• Ambient temperature range 0°C to 60°C
• Humidity 0 to 95% non-condensing
Overall Dimension Length
603 mm
Grip
89 mm x 32 mm
Sensor Sheath
457 mm x 6.35 mm
Amphenol
Advanced Sensors
* Transfer calibration accuracy is the difference between the thermocouple tip
and the sensor of the IRTD temperature standard. This accuracy includes well
to well uniformity.
www.amphenol-sensors.com
© 2014 Amphenol Corporation. All Rights Reserved. Specifications are subject to change without notice.
Other company names and product names used in this document are the registered trademarks or
trademarks of their respective owners.
AAS-920-237B-03/2014
Similar pages