Astrodyne MMK150S-12 Medical grade 150 watt enclosed switching power supply Datasheet

mmk150s
M e d i c a l G r a d e 1 50 Watt Enclosed Switching Power Supplies
series
Universal 88-264VAC Auto Range Input
Compact Footprint
High Efficiency
Regulated Outputs
4000V Isolation
Built in EMI Filter
3.5”W x 6”L x 1.75”H
Model Number
Output Voltage
Output Amps (max) Ripple/Noise pp,typ
MMK150S-5
MMK150S-12
MMK150S-15
MMK150S-24
5 VDC
12 VDC
15 VDC
24 VDC
30
12.5
10
6
100mv
100mv
100mv
100mv
Order On-Line!
WWW.ASTRODYNE.COM
astrodyne.com
300 Myles Standish Blvd., Taunton, MA 02780
Ph: 1-800-823-8082
Fax: 508-823-8181
M e d i c a l G r a d e 1 50 Watt Enclosed Switching Power Supplies
INPUT SPECIFICATIONS
Input Voltage Range
Frequency Range
Inrush Current, typ:
series
ENVIRONMENTAL SPECIFICATIONS
88-132/176-264 VAC
Auto Voltage Select
47-63 Hz
20A @ 100VAC *
40A @ 200VAC *
Oper. Temperature
Storage Temperature
Relative Humidity
EMI (Designed to Meet)
MTBF
See Selection Chart
± 0.5%, typ
± 1%, typ
± 6%, typ
± 0.03%/°C
100mV, typ
130%, Latching, Re-power *
160%, Indefinite
20mS, typ (Nom I/P, FL)
≤ 1.0 sec, typ
≤ 20 msec, typ
< 10% of FS
≤ 0.3 mA @ 240VAC, 50 Hz
PHYSICAL SPECIFICATIONS
OUTPUT SPECIFICATIONS
Voltage and Current
Line/Load Regulation (10%-FL)
Voltage Tolerance
DC Voltage Adjust (typ)
Temperature Coefficient
Ripple/Noise
Over Voltage Protection
Short Circuit Protection
Hold Up Time
Turn On Time
Rise Time
Over/Under Shoot
Leakage Current
mmk150s
Size
Construction
Weight
0 to +60°C
(See Derate Curve)
-25 to +85°C *
10% to 90%, non-cond *
EN55011, Class B
100,000 Hrs
Mil Std 217, 25°C
6” x 3.5” x 1.75”
Enclosed
1.37 lb, (630g)
GENERAL SPECIFICATIONS
Isolation
Efficiency
Switching Frequency
Safety I/P-O/P: 4000VAC
I/P-Ground: 1500VAC
O/P-Ground: 500 VAC
78%, typ
100Khz, (fixed, typical)
UL60601-1, EN60601-1, IEC60601-1
All specifications are typical at nominal input, full load, and 25°C unless otherwise noted
* These are stress ratings. Exposure of the devices to any of these conditions
may adversely affect long term reliability. Proper operation under conditions
other than the standard operating conditions is neither warranteed nor implied.
WWW.ASTRODYNE.COM
Astrodyne products are not authorized or warranteed for use
as critical components in life support systems, equipment used
in hazardous environments, nuclear controls systems, or other
mission-critical applications.
300 Myles Standish Blvd., Taunton, MA 02780
Ph: 1-800-823-8082
Fax: 508-823-8181
M e d i c a l G r a d e 1 50 Watt Enclosed Switching Power Supplies
mmk150s
series
MECHANICAL DIMENSIONS
Pin #
Outputs
1
2
3
4
5
6
7
ACN
ACL
GND
+ Output
+ Output
- Output
- Output
PLEASE NOTE:
To improve specifications and safety isolation, connection
designations may have changed from prior revisions. Please
verify connections per product labels on input and output
connections prior to applying power to the unit
OUTPUT DERATING CURVE
(Forced air cooling direction: Straight down from top of unit)
22CFM Forced Air - All Outputs (Recommended)
1
2
3
12/15/24VDC Outputs, Convection Cooled
5VDC Output, Convection Cooled
120
1
100
Load, %
80
60
2
40
3
20
0
0
5
10
15
20
25
30
35
40
45
50
55
60
Ambient Temp, C
WWW.ASTRODYNE.COM
300 Myles Standish Blvd., Taunton, MA 02780
Ph: 1-800-823-8082
Fax: 508-823-8181
Similar pages