XP Power ISM0105S05 Dc-dc converter Datasheet

ISM01 Series
DC-DC Converter
1 Watt
xxx Series
• International Medical Safety Approvals
• Single & Dual Output
• Compact SMD Package
• 4000 VAC Isolation
• 2 x MOPP at 250 VAC
• 2 µA Patient Leakage Current
• Operating Temperature -40 °C to +95 °C
• Tape & Reel Package Available
• Optional Water Washable Versions
Dimensions:
• 3 Year Warranty
ISM01:
0.74 x 0.68 x 0.41” (18.9 x 17.2 x 10.5 mm)
Models & Ratings
Output current
Min
Max
Input voltage Output voltage
5V
12 V
15 V
±12 V
±15 V
5V
12 V
15 V
±12 V
±15 V
5V
12 V
15 V
±12 V
±15 V
4.5-5.5
10.8-13.2
21.6-26.4
4.00 mA
1.68 mA
1.36 mA
±0.84 mA
±0.66mA
4.00 mA
1.68 mA
1.36 mA
±0.84 mA
±0.66 mA
4.00 mA
1.68 mA
1.36 mA
±0.84 mA
±0.66 mA
Input current(1)
No load
Full load
200 mA
84 mA
68 mA
±42 mA
±33 mA
200 mA
84 mA
68 mA
±42 mA
±33 mA
200 mA
84 mA
68 mA
±42 mA
±33 mA
Maximum(2)
Reflected
capacitive load ripple current
265 mA
250 mA
245 mA
250 mA
235 mA
110 mA
105 mA
105 mA
105 mA
105 mA
55 mA
55 mA
55 mA
55 mA
50 mA
50 mA
35 mA
20 mA
220 µF
220 µF
220 µF
±100 µF
±100 µF
220 µF
220 µF
220 µF
±100 µF
±100 µF
220 µF
220 µF
220 µF
±100 µF
±100 µF
80 mA
40 mA
30 mA
Efficiency
Model number(3)
76%
80%
83%
80%
84%
76%
79%
80%
79%
80%
76%
80%
80%
80%
80%
ISM0105S05
ISM0105S12
ISM0105S15
ISM0105D12
ISM0105D15
ISM0112S05
ISM0112S12
ISM0112S15
ISM0112D12
ISM0112D15
ISM0124S05
ISM0124S12
ISM0124S15
ISM0124D12
ISM0124D15
Notes
1. Input currents measured at nominal input voltage.
2. Maximum capacitive load is per output.
3. For optional water washable version, add suffix ‘-P’ e.g. ISM0112S05-P.
Mechanical Details
0.74 (18.9)
0.6 (15.24)
0.5 (12.7)
0.07
(1.83)
0.1 (2.54)
9
14
8
0.54
(13.7)
Top View
6
1
ø0.06 (ø1.5)
0.16
(4.0)
0.68
(17.2)
0.22
(5.5)
Bottom View
0.54
(13.7)
0.64
(16.16)
7
0.04 (1.0)
0.41
0.40
(10.5)
(10.2)
S
SEATING
PLANE
°
0.1 (2.54)
1
0.04
(1.1)
0
0.6 (15.0)
0°
~4
0.01 (0.30)
0.06 (1.6)
0.15 S
www.xppower.com
0.1 (2.54)
0.73
(18.6)
ISM01 Series
DC-DC Converter
Input
Characteristic
Minimum
Typical
4.5
Input Voltage Range
Input Filter
Maximum
Units
5.5
10.8
13.2
21.6
26.4
Notes & Conditions
5 V nominal
VDC
12 V nominal
24 V nominal
Internal Capacitor
9
18
VDC for 1 s
VDC for 1 s
5 V models
12 V models
30
VDC for 1 s
24 V models
Maximum
Units
Notes & Conditions
30
VDC
See Models and Ratings table
±1.0
±3.0
%
±0.1
±1.0
Input Surge
Output
Characteristic
Minimum
Output Voltage
5.0
Typical
Initial Set Accuracy
Output Voltage Balance
Minimum Load
2
Line Regulation
±1.2
Load Regulation
Ripple & Noise
%
For dual output with balanced loads
%
See Models and Ratings table
±1.5
%/1%
10
%
100
Output changes by max of 1.5% for each 1% change in
input voltage
10% to 100%
mV pk-pk
Short Circuit Protection
20 MHz bandwidth. Measured using 0.47 µF ceramic
capacitor.
Continuous, with auto recovery
Maximum Capacitive Load
See Models and Ratings table
Temperature Coefficient
0.02
%/ ˚C
Maximum
Units
General
Characteristic
Minimum
Typical
Efficiency
Isolation: Input to Output
80
20
Switching Frequency
55
Power Density
Mean Time Between Failure
Moisture Sensitivity Level
Ω
pF
kHz
4.8
4.7
Weight
Reinforced, 2 x MOPP at 250 VAC working voltage
µA
2
109
Isolation Capacitance
Notes & Conditions
See Models and Ratings table
VAC
4000
Leakage Current
Isolation Resistance
%
W/in3
MHrs
0.01 (4.2)
At 500 VDC
MIL-HDBK-217F, +25 °C GB
lb (g)
Level 2
IPC/JEDEC J-STD-020D.1
Environmental
Characteristic
Minimum
Operating Temperature
-40
Storage Temperature
-50
Typical
Case Temperature
Humidity
Maximum
Units
Notes & Conditions
+95
°C
See Derating Curve.
+125
°C
+105
°C
95
%RH
Cooling
Case Flammability
Non-condensing
Natural convection
UL 94V-0 Rated
Non conductive black plastic
Lead-Free Reflow Solder Process
IPC/JEDEC J-STD-020D.1
Safety Approvals
Safety Agency
Safety Standard
Notes & Conditions
CB
IEC60601-1
Medical
UL
ANSI/AAMI ES60601-1
Medical
www.xppower.com
2
ISM01 Series
DC-DC Converter
EMC: Emissions
Phenomenon
Standard
Test Level
Conducted
EN55011
Class A
Notes & Conditions
EMC: Immunity
Test Level
Criteria
ESD
EN61000-4-2
±15 kV air discharge, ±8 kV contact
A
Radiated
EFT/Burst
EN61000-4-3
EN61000-4-4
10 V/m
±2 kV
A
A
A
Phenomenon
Standard
Surge
EN61000-4-5
±1 kV
Conducted
EN61000-4-6
10 V rms
A
Magnetic Fields
EN61000-4-8
30 A/m
A
Notes & Conditions
Mechanical Details
Connecting Pin Pattern
0.74 (18.9)
0.6 (15.24)
0.5 (12.7)
0.07
(1.83)
0.1 (2.54)
9
14
0.54
(13.7)
6
0.16
(4.0)
8
Top View
1
ø0.06 (ø1.5)
0.68
(17.2)
0.22
(5.5)
Bottom View
0.54
(13.7)
0.64
(16.16)
7
0.73
(18.6)
0.04 (1.0)
0.41
0.40
(10.5)
(10.2)
S
SEATING
PLANE
°
0.1 (2.54)
0.04
(1.1)
0.6 (15.0)
0.06 (1.6)
0°
~4
0.01 (0.30)
0.15 S
Pin Connections
Pin
1
6
7
8
9
14
0
Notes
1. All dimensions are in inches (mm)
2. Weight: 0.01 lbs (4.2 g) approx.
0.1 (2.54)
3. Tolerance: X.XX±0.02 (X.X±0.5)
X.XXX±0.01 (X.XX±0.25)
4. Pin Tolerance: ±0.002 (±0.05)
Single
-Vin
No Connection
No Connection
+Vout
-Vout
+Vin
Dual
-Vin
Common
-Vout
+Vout
Common
+Vin
Application Notes
Derating Curve
Output Power (%)
100
0
-40
3
Ambient Temperature (°C)
85
www.xppower.com
95
23 Feb 18