Keysight DSOX6JITTER Jitter analysis option Datasheet

Keysight DSOX6JITTER
Jitter Analysis Option
For InfiniiVision 6000 X-Series Oscilloscopes
Data Sheet
Introduction
As data rates continue to increase in today’s state-of-the-art high-speed digital designs, timing
budgets are decreasing. Ensuring that high-speed clocks and serial data signals are valid and stable
when receivers sample data often requires an understanding of the effects of the various components
of jitter that may contribute to decreased valid data windows. With the DSOX6JITTER jitter
analysis option, the InfiniiVision 6000 X-Series oscilloscope can perform various types of jitter
analysis measurements with multiple views and quantization of jitter to provide you with valuable
insight into the source of unwanted timing errors. The Keysight Technologies, Inc. InfiniiVision 6000
X-Series oscilloscope with the DSOX6JITTER jitter analysis option is the lowest priced oscilloscope
jitter analysis test solution in the market today.
Features
–– Serial data TIE jitter measurements
–– Clock jitter measurements including: clock TIE, N period, period to period, +width to +width,
-width to -width, and duty cycle
–– Various clock recovery algorithms including: constant/fixed clock, 1st order phase-locked loop
(PLL) with loop bandwidth, and 2nd order PLL with loop bandwidth and damping factor
–– Various views/plots of jitter including: jitter histogram, jitter trend, jitter spectrum, and
jitter statistics
–– Color-graded real-time eye with automatic eye-opening measurements
03 | Keysight | DSOX6JITTER Jitter Analysis Option - Data Sheet
Types of Jitter Analysis
There are various types of jitter analysis
measurements that can be selected and
performed using the DSOX6JITTER jitter
analysis option. Jitter on clock signals can be
measured either relative to a software-based
recovered ideal clock or relative to selected
cycles of the clock signal itself. Jitter measurements on serial data signals, which are
sometimes referred to as Time Interval Error
(TIE), can be performed relative to either an
explicit reference clock or software-based
recovered clock.
Figure 1 shows the InfiniiVision 6000 X-Series
oscilloscope jitter analysis selection/setup
menu. The following table summarizes
the various types of jitter analysis that
can be selected and performed using
the DSOX6JITTER option licensed on an
InfiniiVision 6000 X-Series oscilloscope.
Figure 1. InfiniiVision 6000 X-Series oscilloscope jitter analysis setup menu.
Clock jitter measurements
Clock TIE
Measures timing error of selected edges (rising, falling, or both) of the input clock signal relative to a recovered
ideal clock.
N period
Measures the time-span of N consecutive periods of the clock based on the selected edge (rising or falling).
Then, it shifts ahead by one period and measures the time-span of the next N consecutive periods, and so on.
For example, if N = 3, then jitter = (T1 + T2 + T3), (T2 + T3 + T4), (T3 + T4 + T5), etc.
Period to period
Measures the time difference of successive periods (or groups of periods if N > 1) of the input clock signal on
the selected edge (rising or falling). If N = 1, which is sometimes referred to as cycle-to-cycle jitter,
then jitter = T2 – T1, T3 – T2, T4 – T3, etc.
If N = 3, then jitter = (T4 + T5 + T6) – (T1 + T2 + T3), (T5 + T6 + T7) – (T2 + T3 + T4), etc.
+Width to +width
Measures the time difference of successive positive pulse widths of the input clock signal (+PWn+1 – +PWn).
-Width to -width
Measures the time difference of successive negative pulse widths of the input clock signal (-PWn+1 – -PWn).
+Duty cycle
Measures the duty cycle of every cycle of the input clock signal.
Serial data jitter measurements
Data TIE
Measures timing error of all edges of the input serial data signal relative to either a recovered ideal clock or an
explicit input clock signal.
04 | Keysight | DSOX6JITTER Jitter Analysis Option - Data Sheet
Views/Plots of Jitter Measurements
Various views/plots of jitter, as well as
statistical results of jitter measurements,
can be selected to help provide insight into
the distribution and source of timing errors
including the following:
– Jitter histogram
– Jitter trend
Jitter statistics
– Jitter spectrum
– Jitter statistics
Jitter trend
– Real-time eye
Figure 2 shows an example of a serial data
TIE measurement with a jitter histogram plot
and jitter trend plot. The histogram shows
the probability distribution function (PDF)
of accumulated timing errors along with
numerical statistics of measured jitter. The
jitter trend plot shows timing error of each
serial data edge time-correlated to the input
data signal (yellow trace) that is under test.
In other words, the jitter trend waveform
plots timing error on the vertical axis (sec/
div) versus time on the horizontal axis.
Jitter histogram
Figure 2. Jitter analysis with the jitter histogram and jitter trend views of timing errors.
Figure 3 shows an example of a serial data
TIE measurement with the jitter spectrum
view along with a jitter trend plot. The jitter
spectrum is actually a Fast Fourier Transform
(FFT) waveform math operation performed
on the jitter trend waveform data. This view
of jitter can often times help identify periodic
components of jitter as illustrated in this
jitter measurement example. Also available
are frequency-domain search features to
automatically identify peak frequencies.
Note that although they not shown, all three
views of jitter analysis (histogram, trend, and
spectrum) and can be selected and viewed
simultaneously.
Jitter spectrum
Figure 3. Jitter analysis with the jitter spectrum and jitter trend views of timing errors.
05 | Keysight | DSOX6JITTER Jitter Analysis Option - Data Sheet
Views/Plots of Jitter Measurements (continued)
In addition to viewing jitter in histogram,
trend, and spectrum views, the InfiniiVision
6000 X-Series oscilloscopes can also display
a color-graded Real-Time Eye (RTE) of
serial data signals as shown in Figure 4. A
real-time eye is an overlay of all bits to show
worst-case timing (jitter) and worst-case
noise relative to a recovered or explicit clock.
With a real-time eye display, it is easy to
observe the valid data window present at the
input of receivers, along with automatic eye
opening measurements (eye height and
eye width).
Clock recovery for jitter
measurements
Figure 4. The color-graded real-time eye waveform shows worst-case data valid window (eye-opening).
Clock and serial data TIE jitter measurements
are performed on all captured edges based
on the selected transition(s) relative to either
an explicit input reference clock signal or a
software-based recovered clock (sometimes
referred to as clock data recovery (CDR)).
Creating real-time eye displays also requires
a recovered clock to use as a reference for
slicing the captured serial data waveform
into individual Unit Internals (UIs) for
overlapping/folding. Since most of today’s
serial bus signals don’t have explicit clocks
(clocks are typically embedded in the data
signal), software-based clock recovery is
required to extract the clock from the data.
Figure 5 shows a zoomed-in view of a serial
data input (yellow trace), a recovered clock
waveform (upper purple trace), and a TIE
trend waveform (lower purple trace). TIE
(jitter) measurements are performed on each
transition of the serial data signal relative to
the mid-point of the closest recovered clock,
and then these measurement results are
plotted as the TIE (jitter) trend waveform.
The DSOX6JITTER jitter analysis option
provides the following software-based clock
recovery algorithms:
– Constant/fixed clock
– 1 st order PLL clock with user-specified
loop bandwidth/cut-off frequency
– 2 nd order PLL clock with user-specified
loop bandwidth and damping factor
– Explicit clock (captured/not recovered)
Figure 5. Software-based clock recovery provides a reference waveform for performing TIE jitter measurements.
06 | Keysight | DSOX6JITTER Jitter Analysis Option - Data Sheet
Probing Differential Clock and Serial Bus Data Signals
Most of today’s higher-speed digital designs
are based on differential signaling. If direct
50-Ω cabling from the measurement source
to the oscilloscope is not possible, then a
differential probe must be used. Keysight
recommends the N2750A InfiniiMode Series
differential active probes (1.5 to 6 GHz bandwidth models) shown in Figure 6.
The InfiniiMode Series probe allows you to
view not only the differential signal, but you
can also select to view each side of a differential signal (high-side and low-side) bus
relative to ground, as well as the common
mode of the bus — without ever moving
probe connections.
The InfiniiMode differential active probe
comes standard with multiple probe heads
for various probing use-models including
solder-in, socketed, and browser. This probe
also has a high-intensity LED headlight to
make it easier to connect to finepitch devices.
Figure 6. Keysight’s InfiniiMode N2750A Series differential active probe.
Features and characteristics
Data jitter measurements
Data TIE
Clock jitter measurements
Clock TIE
N period
Period to period
+Width to +width
-Width to –width
Duty cycle
Jitter views
Histogram
Trend
Spectrum
Statistics
Real time eye (color-graded)
Clock recovery
Explicit
Constant/fixed
1st order PLL with user-specified loop bandwidth/cut-off frequency
2nd order PLL with user-specified loop bandwidth/cut-off frequency and damping factor
07 | Keysight | DSOX6JITTER Jitter Analysis Option - Data Sheet
Ordering Information
The DSOX6JITTER jitter analysis option is compatible on all InfiniiVision 6000 X-Series oscilloscope models with bandwidths ranging from
1 to 6 GHz and real-time sampling up to 20 GSa/s. Refer to the InfiniiVision 6000 X-Series oscilloscope data sheet for ordering information about
specific oscilloscope models and other licensed options.
Related Literature
Publication title
Publication type
Publication number
InfiniiVision 6000 X-Series Oscilloscopes
Data sheet
5991-4087EN
N2750A InfiniiMode Series Differential Active Probe
Data sheet
5991-0560EN
Jitter Analysis using Keysight’s InfiniiVision 6000 X-Series and Infiniium Series Oscilloscopes
Application note
5991-4000EN
Finding Sources of Jitter with Real Time Jitter Analysis
Application note
5988-9740EN
To download these documents, insert the publication number in the URL: http://literature.cdn.keysight.com/litweb/pdf/xxxx-xxxxEN.pdf
Product web site
For the most up-to-date and complete application and product information, please visit our product Web site at:
www.keysight.com/find/6000X-Series
08 | Keysight | DSOX6JITTER Jitter Analysis Option - Data Sheet
myKeysight
www.keysight.com/find/mykeysight
A personalized view into the information most relevant to you.
www.axiestandard.org
AdvancedTCA® Extensions for Instrumentation and Test (AXIe) is an
open standard that extends the AdvancedTCA for general purpose and
semiconductor test. Keysight is a founding member of the AXIe consortium.
ATCA®, AdvancedTCA®, and the ATCA logo are registered US trademarks of
the PCI Industrial Computer Manufacturers Group.
www.lxistandard.org
LAN eXtensions for Instruments puts the power of Ethernet and the Web
inside your test systems. Keysight is a founding member of the
LXI consortium.
www.pxisa.org
PCI eXtensions for Instrumentation (PXI) modular instrumentation delivers a
rugged, PC-based high-performance measurement and automation system.
Three-Year Warranty
www.keysight.com/find/ThreeYearWarranty
Keysight’s commitment to superior product quality and lower total cost
of ownership. The only test and measurement company with three-year
warranty standard on all instruments, worldwide.
Keysight Assurance Plans
www.keysight.com/find/AssurancePlans
Up to five years of protection and no budgetary surprises to ensure
your instruments are operating to specification so you can rely on
accurate measurements.
www.keysight.com/go/quality
Keysight Technologies, Inc.
DEKRA Certified ISO 9001:2008
Quality Management System
Keysight Channel Partners
www.keysight.com/find/channelpartners
Get the best of both worlds: Keysight’s measurement expertise and product
breadth, combined with channel partner convenience.
For more information on Keysight
Technologies’ products, applications or
services, please contact your local Keysight
office. The complete list is available at:
www.keysight.com/find/contactus
Americas
Canada
Brazil
Mexico
United States
(877) 894 4414
55 11 3351 7010
001 800 254 2440
(800) 829 4444
Asia Pacific
Australia
China
Hong Kong
India
Japan
Korea
Malaysia
Singapore
Taiwan
Other AP Countries
1 800 629 485
800 810 0189
800 938 693
1 800 112 929
0120 (421) 345
080 769 0800
1 800 888 848
1 800 375 8100
0800 047 866
(65) 6375 8100
Europe & Middle East
Austria
Belgium
Finland
France
Germany
Ireland
Israel
Italy
Luxembourg
Netherlands
Russia
Spain
Sweden
Switzerland
United Kingdom
0800 001122
0800 58580
0800 523252
0805 980333
0800 6270999
1800 832700
1 809 343051
800 599100
+32 800 58580
0800 0233200
8800 5009286
800 000154
0200 882255
0800 805353
Opt. 1 (DE)
Opt. 2 (FR)
Opt. 3 (IT)
0800 0260637
www.keysight.com/find/6000X-Series
For other unlisted countries:
www.keysight.com/find/contactus
(BP-09-23-14)
This information is subject to change without notice.
© Keysight Technologies, 2014
Published in USA, September 25, 2014
5991-4061EN
www.keysight.com