XP Power ALM85US12 Ac-dc power supply Datasheet

ALM85 Series
AC-DC Power Supplies
85 Watts
• Medical & IT Safety Approvals
• Energy Efficiency Level VI & EU CoC Tier 2 Compliant
• 4th Edition Medical EMC
• IP32 Environmental Rating
• Class I and Class II Versions
• <0.15 W Standby Power
• 0 °C to 60 °C Operation
• Low Earth Leakage Current
• 3 Year Warranty
Dimensions:
The ALM85 series of medical external power supplies is fully approved to international medical & IT
safety standards. It has been designed with very high efficiency and low standby power, enabling it to
meet the latest environmental legislation. The unit has a fully sealed enclosure complying with IP32
and a smooth surface finish making it easier to wipe down in a clinical setting.
ALM85:
5.315 x 2.441 x 1.457” (135.0 x 62.0 x 37.0 mm)
Models & Ratings
Output Power
Output Voltage
Output Current
12.0 V
15.0 V
19.0 V
24.0 V
80 W
85 W
6.67
5.33
4.47
3.54
Total Regulation
A
A
A
A
Model Number(1,2)
ALM85US12
ALM85US15
ALM85US19
ALM85US24
±5%
Notes
1. For class II versions, add suffix 'C2-8' to the end of the part number e.g.
ALM85US24C2-8.
2. For optional input connector retention clip add suffix ‘-A’ to the model number,
e.g. ALM85US24-A (not available for C2 versions)
Mechanical Details
AC Input
IEC 320-C14
-A Versions
5.31 (135.0)
1.58 (40.2)
35.43 ±0.79 (900.0 ±20.0)
1.5 (38.4)
AC Input
IEC 320-C14
Standard Version
1.46
(37.0)
2.44
(62.0)
1.46
(37.0)
12 &K15 V Versions
KPPX-4P
(Non-Locking)
ø0.685 (17.4)
1.95 ±0.039
(50.0 ±10.0)
0.17 (4.4)
AC Input
IEC 320-C8
C2-8 Versions
0.37 ±0.012
(9.5 ±0.3)
Pin
1
2
3
4
Connection
+Vout
+Vout
RTN
RTN
0.039 (1.0)
1.89 (48.0)
19 & 24 V Versions
4.76 (121.0)
Return (-)
3.04 (77.2)
1.46
(37.0)
Output (+)
1.58 (40.2)
Notes
1. All dimensions shown in inches (mm). Tolerance is 0.02 (0.5) maximum, except
output cable length.
2. Weight: 0.78 lbs (355 g) approx.
3. For European mains lead order part EU-MAINS-IEC for C14 versions,
or EU-MAINS-8 for C8 versions.
4. For UK mains lead order part UK-MAINS-IEC for C14 versions, or UK-MAINS-8
for C8 versions.
1
5. For US mains lead order part US-MAINS-IEC for C14 versions, or US-MAINS-8
for C8 versions.
6. Output connector: 12 V & 15 V: 4 pin power din with pin 1 & 2 - posiitive and pin
3 & 4 - return, equivalent to KPPX-4P (non-locking). 19 V & 24 V: DC barrel jack;
5.5 outer diameter, 2.5 mm inner diameter with centre positive.
www.xppower.com
ALM85 Series
AC-DC Power Supplies
Input
Characteristic
Minimum
Input Voltage
80
Input Frequency
Input Current
Inrush Current
Power Factor
Earth Leakage Current
No Load Input Power
Input Protection
47
Typical
Units
Notes & Conditions
264
VAC
Derate linearly from 100% load at 90 VAC to 80% load
at 80 VAC. 300 VAC/5 s maximum
63
Hz
A
A
Maximum
1.7/0.9
60/120
250
0.15
T3.15A/250 VAC internal fuse in both line & neutral
Measured at 115/230 VAC
115/230 VAC, cold start at 25 °C
EN61000-3-2 Class A
264 VAC, 60 Hz
µA
W
Output
Characteristic
Output Voltage
Initial Set Accuracy
Minimum Load
Start Up Delay
Start Up Rise Time
Hold Up Time
Line Regulation
Total Regulation
Min.
12
Typ.
Max.
24
±2
Units
VDC
%
2
150
±0.5
±5
s
ms
ms
%
%
4
%
Ripple and Noise
1.5
% pk-pk
Overshoot
Overload Protection
Overvoltage Protection
Short Circuit Protection
Thermal Protection
Temperature Coefficient
Patient Leakage Current
10
175
150
%
%
%
95
%/°C
µA
264 VAC, 60 Hz
Maximum
+60
Units
°C
Notes & Conditions
Derate from 100% load at 40 °C to 60% load at 60 °C
10/20
Transient Response
115
Trip and restart (hiccup), auto resetting
Measured internally, auto resetting
0.02
Notes & Conditions
See Models and Ratings table
At 50% load
No minimum load required
Full load and 115/230 VAC
Maximum deviation, recovering to less than 1% within
500 µs for 25% step load
20 MHz bandwidth, measured with 20 MHz Bandwidth and
10 µF electrolytic in parallel with 0.1 µF ceramic capacitor.
At turn on / turn off
Recycle mains to reset
Environmental
Characteristic
Operating Temperature
Cooling
Operating Humidity
Storage Temperature
Operating Altitude
Shock
Vibration
Minimum
0
Natural convection
5
-20
Typical
95
%RH
+80
°C
5000
m
IEC68-2-27, 30 g, 11 ms half sine, 3 times in each of 6 axes
IEC68-2-6, 10-500 Hz, 2 g 10 mins/sweep, 60 mins for each of 3 axes
Non-condensing
Derating Curve
Output Power (%)
100%
80%
60%
40%
20%
0
20
Ambient Temperature (ºC)
www.xppower.com
40
60
2
ALM85 Series
AC-DC Power Supplies
General
Characteristic
Minimum
Efficiency
Typical
Maximum
Notes & Conditions
Units
%
88
Switching Frequency
Power Density
Mean Time Between Failure
Weight
See Models and Ratings table and curves.
4000
VAC
2 x MOPP
1500
VAC
1 x MOPP (Class I versions only)
500
VAC
Class I versions only
65 KHz
kHz
PWM
4.55
W/in3
Isolation: Input to Output
Input to Ground
Output to Ground
>300
kHrs
0.78 (355)
lb (g)
MIL-HDBK-217F at 25 °C GB
Efficiency Curves
ALM85US12
100.00%
Efficiency
90.00%
80.00%
115V
230V
70.00%
60.00%
10%
20%
30%
40%
50%
60%
70%
80%
90%
100%
Load
ALM85US24
100.00%
Efficiency
90.00%
80.00%
115V
230V
70.00%
6
60.00%
10%
20%
30%
40%
50%
60%
70%
80%
90%
100%
Load
EMC: Emissions
Phenomenon
Emissions
Harmonic Current
Voltage Flicker
Standard
EN55022
EN61000-3-2
EN61000-3-3
Test Level
Level B
Class A
www.xppower.com
Notes & Conditions
Conducted & Radiated
3
ALM85 Series
AC-DC Power Supplies
EMC: Immunity
Phenomenon
ESD
Radiated
EFT/Burst
Surge
Conducted
Magnetic Fields
Test Level
3
10 V/m
3
Installation Class 3
10 V
4
Dip: 30% 500 ms
Dip: 60% 200 ms
Dip: 80% 5000 ms
Int: 100% 10 ms
Int: 100% 20 ms
Int: 100% 5000 ms
Dip: 30% 25 AC Cycles
Dip: 60% 5 AC Cycles
Int: 100% 0.5 AC Cycles
Int: 100% 1.0 AC Cycles
Int: 100% 250 AC Cycles
Standard
EN61000-4-2
EN61000-4-3
EN61000-4-4
EN61000-4-5
EN61000-4-6
EN61000-4-8
EN61000-4-11
Dips and Interruptions
EN60601-1-2
Criteria
A
A
A
A
A
A
A/B
A/B
B
A
B
B
A
A
A
A
B
Notes & Conditions
±15 kV air/±8 kV contact
80-2700 MHz. IEC60601-1-2 Ed.4 at other frequencies
High Line/Low Line
High Line/Low Line
230 VAC 100% load, 100 VAC 80%
230 VAC 100% load, 100 VAC 30%
230 VAC 100% load, 100 VAC 50%
Safety Approvals
Safety Agency
Safety Standard
UL
Notes & Conditions
UL60950-1 2nd Edition
TUV
EN60950-1
CB
IEC60950-1
CE
LVD
UL
ANSI/AAMI ES 60601-1
CSA
CSA C22.2 No. 60601
TUV
EN60601-1
CB
IEC60601-1
Information Technology
Medical
Mechanical Details
AC Input
IEC 320-C14
-A Versions
5.31 (135.0)
1.58 (40.2)
35.43 ±0.79 (900.0 ±20.0)
1.5 (38.4)
AC Input
IEC 320-C14
Standard Version
1.46
(37.0)
2.44
(62.0)
1.46
(37.0)
12 & 15 V Versions
KPPX-4P
K
(Non-Locking)
ø0.685 (17.4)
1.95 ±0.039
(50.0 ±10.0)
0.17 (4.4)
AC Input
IEC 320-C8
C2-8 Versions
0.37 ±0.012
(9.5 ±0.3)
Pin
1
2
3
4
Connection
+Vout
+Vout
RTN
RTN
0.039 (1.0)
1.89 (48.0)
19 & 24 V Versions
4.76 (121.0)
3.04 (77.2)
Return (-)
1.46
(37.0)
Output (+)
1.58 (40.2)
Notes
1. All dimensions shown in inches (mm). Tolerance is 0.02 (0.5) maximum, except
output cable length.
2. Weight: 0.78 lbs (355 g) approx.
3. For European mains lead order part EU-MAINS-IEC for C14 versions,
or EU-MAINS-8 for C8 versions.
4. For UK mains lead order part UK-MAINS-IEC for C14 versions, or UK-MAINS-8
for C8 versions.
5. For US mains lead order part US-MAINS-IEC for C14 versions, or US-MAINS-8
for C8 versions.
6. Output connector: 12 V & 15 V: 4 pin power din with pin 1 & 2 - posiitive and pin
3 & 4 - return, equivalent to KPPX-4P (non-locking). 19 V & 24 V: DC barrel jack;
5.5 outer diameter, 2.5 mm inner diameter with centre positive.