NI X Series Multifunction Data Acquisition
Overview
NI X Series devices for USB, PCI Express, and PXI Express are the most advanced data acquisition devices ever designed by National Instruments. They feature significant
improvements in onboard timing and triggering and optimizations for use with multicore PCs. X Series devices integrate high-performance analog, digital, and counter/timer
functionality onto a single device, making them well-suited for a broad range of applications, from basic data logging to control and test automation.
Back to Top
Requirements and Compatibility
OS Information
Driver Information
PharLap
Software Compatibility
NI-DAQmx
ANSI C/C++
Real-Time OS
LabVIEW
Windows 7
LabVIEW Real-Time Module
Windows 7 64-bit
LabWindows/CVI
Windows Vista x64/x86
Measurement Studio
Windows XP
SignalExpress
Visual Basic
Visual Studio .NET
Back to Top
Comparison Tables
Bus
Model
Number
Analog
Inputs (AI)
Max AI Sampling
Rate (1-channel)
Analog
Outputs
(AO)
Max Total AI
Throughput
Max AO
Digital
Update Rate I/O Lines
Max Digital
Triggering
I/O Rate
PCI Express
6320
16
250 kS/s
250 kS/s
0
-
24
1 MHz
Digital
PCI Express
6321
16
250 kS/s
250 kS/s
2
900 kS/s
24
1 MHz
Digital
PCI Express
6323
32
250 kS/s
250 kS/s
4
900 kS/s
48
1 MHz
Digital
USB, PCI Express,
6341
PXI Express
16
500 kS/s
500 kS/s
2
900 kS/s
24
1 MHz
Digital
USB, PCI Express 6343
32
500 kS/s
500 kS/s
4
900 kS/s
48
1 MHz
Digital
USB, PCI Express 6351
16
1.25 MS/s
1.25 MS/s
2
2.86 MS/s
24
10 MHz
Analog,
Digital
USB, PCI Express 6353
32
1.25 MS/s
1.25 MS/s
4
2.86 MS/s
48
10 MHz
Analog,
Digital
1/18
www.ni.com
Bus
Model
Number
Analog
Inputs (AI)
Max AI Sampling
Rate (1-channel)
Analog
Outputs
(AO)
Max Total AI
Throughput
Max AO
Digital
Update Rate I/O Lines
Max Digital
Triggering
I/O Rate
USB, PCI Express,
6361
PXI Express
16
2 MS/s
2 MS/s
2
2.86 MS/s
24
10 MHz
Analog,
Digital
USB, PCI Express,
6363
PXI Express
32
2 MS/s
2 MS/s
4
2.86 MS/s
48
10 MHz
Analog,
Digital
USB, PXI Express 6356
8
simultaneous
1.25 MS/s/channel
10 MS/s
2
3.33 MS/s
24
10 MHz
Analog,
Digital
6358
16
simultaneous
1.25 MS/s/channel
20 MS/s
4
3.33 MS/s
48
10 MHz
Analog,
Digital
USB, PXI Express 6366
8
simultaneous
2 MS/s/channel
16 MS/s
2
3.33 MS/s
24
10 MHz
Analog,
Digital
16
simultaneous
2 MS/s/channel
32 MS/s
4
3.33 MS/s
48
10 MHz
Analog,
Digital
PXI Express
PXI Express
6368
Back to Top
Application and Technology
NI-STC3 Timing and Synchronization Technology
NI X Series multifunction DAQ devices include the NI-STC3, an ASIC designed by NI for advanced timing, triggering, and synchronization. This technology includes the following:
Four counter/timers with more functionality than ever before, such as the ability to create a finite pulse train with a single counter
A 100 MHz timebase for faster triggering response and more precise generation of analog and digital sample clocks
Independent analog and digital timing engines
Retriggerable measurement tasks for analog I/O, digital I/O, and counter/timers
Native PCI Express Interface
In contrast to a PCI-to-PCI Express bridge chip, which limits the bandwidth of the device to that of the PCI bus and introduces latency, PCI Express and PXI Express X Series
devices use a native x1 PCI Express interface that provides up to 250 MB/s in each direction. National Instruments has also optimized this interface for low latency in single-point
control applications. You can use X Series PCI Express boards in any PCI Express slot from x1 up to x16.
NI Signal Streaming
USB X Series devices include patented NI Signal Streaming, a technology that uses message-based instructions and device-side intelligence to ensure high-speed, bidirectional
data transfer over USB. With USB X Series, you can concurrently transfer analog, digital, and counter data in both directions. The total device throughput over USB is
PC-dependent; on some systems, up to 32 MB/s sustained transfers are possible.
Improved Mechanical Enclosure
USB X Series devices introduce a redesigned, extruded aluminum enclosure with an easy-access magnetic lid. This lid keeps signal wiring secured and shielded and can be
opened easily when needed. The underside of the lid has a device-specific pinout label so that you can quickly determine the corresponding screw terminals for a given channel.
The enclosure also includes a lockable USB port to prevent accidental removal during operation and a security slot that can be used with ordinary laptop locks to secure the
device to a desk or workstation.
Board-Only Form Factor for OEM Applications
USB X Series devices are also available in a board-only form factor for high-volume OEM applications. These devices include standard 34-pin and 50-pin IDC connectors for easy
mating to a daughter board, which can contain custom signal conditioning and connectivity.
Software Enhancements
2/18
www.ni.com
PCI Express and PXI Express X Series devices are compatible with NI-DAQmx Version 9.0 or later driver software. USB X Series devices require NI-DAQmx Version 9.2 or later.
More than a basic driver, NI-DAQmx includes the NI Measurement & Automation Explorer (MAX) configuration utility, the DAQ Assistant for rapid development of basic
applications, and hundreds of example programs for NI LabVIEW and text-based languages. NI-DAQmx also includes LabVIEW SignalExpress LE basic data-logging software.
NI-DAQmx 9.0 introduces the ability to synchronize multiple PCI Express or PXI Express X Series devices with a single NI-DAQmx task, which previously took several tasks and
manual routing of clocks and triggers. This version also introduces the fastest, easiest way to acquire measurement data to disk in the Technical Data Management Streaming
(TDMS) format with the new Configure Logging VI. NI-DAQmx 9.2 introduces the ability to log acquired data to TDMS files within the DAQ Assistant Express VI.
With NI-DAQmx and intuitive LabVIEW graphical programming, you can easily develop applications that take advantage of today's multicore systems so you can perform
acquisition, signal processing, and data logging on different CPU cores.
Simultaneous Sampling X Series
Some X Series devices for USB and PXI Express offer simultaneous sampling, with the same channel counts and connectivity as multiplexed devices.
Unlike multiplexed devices that reduce sampling rates as you add more channels, you can use simultaneous sampling devices to maintain sampling rates as you expand the
number of channels. Simultaneous sampling X Series devices are available with up to 16 differential channels per device, and with PXI Express, you can sample more than 200
channels simultaneously.
Simultaneous X Series devices for USB include 32 or 64 MS onboard memory to ensure the transfer of finite acquisitions, even in the presence of heavy USB traffic.
Applications
Acquisition and Visualization
X Series devices include analog, digital, and counter circuitry for the most common types of static and waveform measurements. With LabVIEW, you can easily acquire the data
and view it on a variety of graphs and displays. You can use configuration-based wizards called Express VIs to take measurements and perform signal processing with minimal
programming.
Data Logging
Whether you are validating a new hardware design, monitoring conditions on a factory floor, or recording temperature changes during a scientific experiment, you need to take
measurements, visualize your data, and often log it to disk. With X Series multifunction DAQ, you can develop a user-defined measurement system by using intuitive graphical
programming software and incorporating the exact visualization, analysis, and data-logging capabilities your application requires.
Control Systems
If you need to control the temperature of a room, the speed of a motor, or the pressure of hydraulic fluids, you can use X Series DAQ hardware to connect sensors and actuators
to your computer and build the control system that meets your exact application needs. The low-latency PCI Express bus improves single-point I/O performance, and with
LabVIEW software and NI-DAQmx driver software, you can easily take sensor measurements, compare values to a setpoint, and update output signals. X Series devices also
have four counter/timers for performing quadrature encoder measurements, pulse-width modulation, pulse train generation, frequency measurements, and much more, making
them ideal for basic motor control.
Due to the inherent higher latency of USB as compared to PCI Express, National Instruments recommends that you use PCI Express or PXI Express X Series devices for
applications that require single-point control or deterministic operation.
Test Automation
X Series DAQ hardware provides analog inputs, analog outputs, hardware-timed digital I/O, and four counter/timers on a single device, making it a cost-effective option for basic
device under test characterization and test automation. With NI-DAQmx software, you can easily synchronize acquisition or generation on multiple subsystems, such as an analog
input and analog output channel. In addition, you can easily synchronize two or more X Series devices for further expansion by using a RTSI cable for PCI Express devices or
over the PXI Express backplane for PXI Express modules. It is possible to synchronize two or more USB X Series devices by exporting sample or reference clocks from the
master device to the slave device, and using external wiring.
Compatible Accessories
PCI Express and PXI Express X Series devices use either a single or dual-stack 68-pin VHDCI female connector, depending on the number of analog and digital channels on the
device. National Instruments offers several options for cables, from 0.5 m to 10 m and from low-cost to high-performance with shielding. Connector blocks are available with screw
terminal, BNC, or custom connector types.
For measurements requiring signal conditioning, you can use X Series with SCXI signal conditioning modules.
You can purchase DIN-rail or panel mount kits for USB X Series devices as well as replacement power supplies and latching USB cables. See the model page for ordering
information.
Upgrading
Because X Series DAQ devices use the same NI-DAQmx driver software as NI M Series DAQ devices, upgrading is easy. In addition, PCI Express and PXI Express X Series
devices use the same VHDCI connector as PCI and PXI M Series. You can reuse your code and preserve your investment in accessories. The pinouts for X Series devices are
backward-compatible with M Series devices.
Back to Top
Ordering Information
For a complete list of accessories, visit the product page on ni.com.
Products
Part Number
Recommended Accessories
Part Number
NI PCIe-6341
NI PCIe-6341
Requires: 1 Cables , 1 Connector Blocks ;
781046-01
Cables: Shielded - SHC68-68-EPM Cable (2m)
**Also Available: [Unshielded]
Connector Blocks: Spring-Screw_Terminals - SCB-68A
**Also Available: [BNC_Terminals]
192061-02
782536-01
NI PXIe-6341
3/18
www.ni.com
NI PXIe-6341
Requires: 1 Cable , 1 Connector Block ;
781052-01
Cable: Shielded - SHC68-68-EPM Cable (2m)
**Also Available: [Unshielded]
192061-02
Connector Block: Spring-Screw_Terminals - SCB-68A
**Also Available: [BNC_Terminals, Custom]
782536-01
Back to Top
Software Recommendations
LabVIEW Professional
Development System for
Windows
NI LabWindows™/CVI for
Windows
SignalExpress for Windows
Advanced software tools for large project
development
Quickly configure projects without
programming
Automatic code generation using DAQ
Assistant and Instrument I/O Assistant
Control over 400 PC-based and stand-alone
instruments
Tight integration with a wide range of
hardware
Log data from more than 250 data acquisition
devices
Advanced measurement analysis and digital
signal processing
Perform basic signal processing, analysis,
and file I/O
Open connectivity with DLLs, ActiveX, and
.NET objects
Scale your application with automatic
LabVIEW code generation
Capability to build DLLs, executables, and
MSI installers
Create custom reports or easily export data
to LabVIEW, DIAdem or Microsoft Excel
NI Measurement Studio
Professional Edition
Real-time advanced 2D graphs and charts
Complete hardware compatibility with IVI,
VISA, DAQ, GPIB, and serial
Customizable graphs and charts for WPF,
Windows Forms, and ASP.NET Web Forms
UI design
Analysis libraries for array operations, signal
generation, windowing, filters, signal
processing
Analysis tools for array manipulation, signal
processing statistics, and curve fitting
Simplified cross-platform communication with
network variables
Measurement Studio .NET tools (included in
LabWindows/CVI Full only)
The mark LabWindows is used under a
license from Microsoft Corporation.
Hardware integration support with native
.NET data acquisition and instrument control
libraries
Automatic code generation for all NI-DAQmx
data acquisition hardware
Intelligent and efficient data-logging libraries
for streaming measurement data to disk
Support for Microsoft Visual Studio .NET
2012/2010/2008
Back to Top
Support and Services
System Assurance Programs
NI system assurance programs are designed to make it even easier for you to own an NI system. These programs include configuration and deployment services for your NI PXI,
CompactRIO, or Compact FieldPoint system. The NI Basic System Assurance Program provides a simple integration test and ensures that your system is delivered completely
assembled in one box. When you configure your system with the NI Standard System Assurance Program, you can select from available NI system driver sets and application
development environments to create customized, reorderable software configurations. Your system arrives fully assembled and tested in one box with your software preinstalled.
When you order your system with the standard program, you also receive system-specific documentation including a bill of materials, an integration test report, a recommended
maintenance plan, and frequently asked question documents. Finally, the standard program reduces the total cost of owning an NI system by providing three years of warranty
coverage and calibration service. Use the online product advisors at ni.com/advisor to find a system assurance program to meet your needs.
Calibration
NI measurement hardware is calibrated to ensure measurement accuracy and verify that the device meets its published specifications. To ensure the ongoing accuracy of your
measurement hardware, NI offers basic or detailed recalibration service that provides ongoing ISO 9001 audit compliance and confidence in your measurements. To learn more
about NI calibration services or to locate a qualified service center near you, contact your local sales office or visit ni.com/calibration.
Technical Support
Get answers to your technical questions using the following National Instruments resources.
Support - Visit ni.com/support to access the NI KnowledgeBase, example programs, and tutorials or to contact our applications engineers who are located in NI sales
offices around the world and speak the local language.
Discussion Forums - Visit forums.ni.com for a diverse set of discussion boards on topics you care about.
Online Community - Visit community.ni.com to find, contribute, or collaborate on customer-contributed technical content with users like you.
Repair
While you may never need your hardware repaired, NI understands that unexpected events may lead to necessary repairs. NI offers repair services performed by highly trained
technicians who quickly return your device with the guarantee that it will perform to factory specifications. For more information, visit ni.com/repair.
Training and Certifications
4/18
www.ni.com
Training and Certifications
The NI training and certification program delivers the fastest, most certain route to increased proficiency and productivity using NI software and hardware. Training builds the skills
to more efficiently develop robust, maintainable applications, while certification validates your knowledge and ability.
Classroom training in cities worldwide - the most comprehensive hands-on training taught by engineers.
On-site training at your facility - an excellent option to train multiple employees at the same time.
Online instructor-led training - lower-cost, remote training if classroom or on-site courses are not possible.
Course kits - lowest-cost, self-paced training that you can use as reference guides.
Training memberships and training credits - to buy now and schedule training later.
Visit ni.com/training for more information.
Extended Warranty
NI offers options for extending the standard product warranty to meet the life-cycle requirements of your project. In addition, because NI understands that your requirements may
change, the extended warranty is flexible in length and easily renewed. For more information, visit ni.com/warranty.
OEM
NI offers design-in consulting and product integration assistance if you need NI products for OEM applications. For information about special pricing and services for OEM
customers, visit ni.com/oem.
Alliance
Our Professional Services Team is comprised of NI applications engineers, NI Consulting Services, and a worldwide National Instruments Alliance Partner program of more than
700 independent consultants and integrators. Services range from start-up assistance to turnkey system integration. Visit ni.com/alliance.
Back to Top
Detailed Specifications
Specifications listed below are typical at 25 °C unless otherwise noted. Refer to the X Series User Manual for more information about NI PCIe-6341/6343, NI PXIe-6341, and NI
USB-6341/6343 devices.
Analog Input
Number of channels
NI 6341
8 differential or 16 single ended
NI 6343
16 differential or 32 single ended
ADC resolution
16 bits
DNL
No missing codes guaranteed
INL
Refer to the AI Absolute Accuracy Table
Sampling rate
Maximum
500 kS/s single channel, 500 kS/s multi-channel (aggregate)
Minimum
No minimum
50 ppm of sample rate
Timing accuracy
10 ns
Timing resolution
DC
Input coupling
±10 V, ±5 V, ±1 V, ±0.2 V
Input range
±11 V of AI GND
Maximum working voltage for analog inputs (signal + common mode)
100 dB
CMRR (DC to 60 Hz)
Input impedance
Device on
>10 GΩ in parallel with 100 pF
AI+ to AI GND
>10 GΩ in parallel with 100 pF
AI– to AI GND
Device off
1200 Ω
AI+ to AI GND
1200 Ω
AI– to AI GND
5/18
www.ni.com
±100 pA
Input bias current
Crosstalk (at 100 kHz)
–75 dB
Adjacent channels
–90 dB
Non-adjacent channels
1.2 MHz
Small signal bandwidth (–3 dB)
4,095 samples
Input FIFO size
4,095 entries
Scan list memory
Data transfers
DMA (scatter-gather), programmed I/O
NI PCIe/PXIe-6341/6343
USB Signal Stream, programmed I/O
NI USB-6341/6343
Overvoltage protection (AI <0..31>, AI SENSE, AI SENSE 2)
±25 V for up to two AI pins
Device on
±15 V for up to two AI pins
Device off
±20 mA max/AI pin
Input current during overvoltage condition
Settling Time for Multichannel Measurements
Accuracy, full scale step, all ranges
2 μs convert interval
±90 ppm of step (±6 LSB)
3 μs convert interval
±30 ppm of step (±2 LSB)
5 μs convert interval
±15 ppm of step (±1 LSB)
None
Analog triggers
Typical Performance Graph
Analog Output
Number of channels
2
NI 6341
4
NI 6343
16 bits
DAC resolution
±1 LSB
DNL
16 bit guaranteed
Monotonicity
Maximum update rate
900 kS/s
1 channel
840 kS/s per channel
2 channels
775 kS/s per channel
3 channels
719 kS/s per channel
4 channels
6/18
www.ni.com
50 ppm of sample rate
Timing accuracy
10 ns
Timing resolution
±10 V
Output range
DC
Output coupling
0.2 Ω
Output impedance
±5 mA
Output current drive
±15 V
Overdrive protection
15 mA
Overdrive current
±20 mV
Power-on state
Power-on/off glitch
2 V for 500 ms
NI PCIe/PXIe-6341/6343
1.5 V for 1.2 s 1
NI USB-6341/6343
8,191 samples shared among channels used
Output FIFO size
Data transfers
NI PCIe/PXIe-6341/6343
DMA (scatter-gather), programmed I/O
NI USB-6341/6343
USB Signal Stream, programmed I/O
AO waveform modes:
Non-periodic waveform
Periodic waveform regeneration mode from onboard FIFO
Periodic waveform regeneration from host buffer including dynamic update
Settling time, full scale step 15 ppm (1 LSB)
6 μs
Slew rate
15 V/μs
Glitch energy
Magnitude
100 mV
Duration
2.6 μs
Calibration (AI and AO)
Recommended warm-up time
15 minutes
Calibration interval
1 year
AI Absolute Accuracy Table
Nominal Range
Positive
Full
Scale
Negative
Full
Scale
Residual Gain Error
(ppm of Reading)
Gain
Tempco
(ppm/°C)
Reference
Tempco
(ppm/°C)
Residual Offset
Error (ppm of
Range)
Offset Tempco
(ppm of
Range/°C)
INL Error
(ppm of
Range)
Random
Noise, σ (μV
)
rms
Absolute Accuracy
at Full Scale* (μV)
10
–10
65
7.3
5
13
23
60
270
2260
5
–5
72
7.3
5
13
23
60
135
1170
1
–1
78
7.3
5
17
26
60
28
250
0.2
– 0.2
105
7.3
5
27
39
60
9
60
Absolute Accuracy Formulas
AbsoluteAccuracy = Reading · (GainError) + Range · (OffsetError) + NoiseUncertainty
GainError = ResidualGainError + GainTempco · (TempChangeFromLastInternalCal) + ReferenceTempco · (TempChangeFromLastExternalCal)
OffsetError = ResidualOffsetError + OffsetTempco · (TempChangeFromLastInternalCal) + INL_Error
NoiseUncertainty = (RandomNoise · 3) / √ 10,000, for a coverage factor of 3 σ and averaging 10,000 points.
*
Absolute accuracy at full scale on the analog input channels is determined using the following assumptions:
TempChangeFromLastExternalCal = 10 °C
TempChangeFromLastInternalCal = 1 °C
7/18
www.ni.com
number_of_readings = 10,000
CoverageFactor = 3 σ
For example, on the 10 V range, the absolute accuracy at full scale is as follows:
GainError = 65 ppm + 7.3 ppm · 1 + 5 ppm · 10
GainError = 122 ppm
OffsetError = 13 ppm + 23 ppm · 1 + 60 ppm
OffsetError = 96 ppm
NoiseUncertainty = (270 μV · 3) / √ 10,000
NoiseUncertainty = 8.1 μV
AbsoluteAccuracy = 10 V · (GainError) + 10 V · (OffsetError) + NoiseUncertainty
AbsoluteAccuracy = 2,190 μV
Accuracies listed are valid for up to one year from the device external calibration.
AO Absolute Accuracy Table
Nominal Range
Positive
Full Scale
Negative
Full Scale
10
–10
Residual Gain Error
(ppm of Reading)
80
Gain
Tempco
(ppm/°C)
11.3
Reference
Tempco
(ppm/°C)
5
Residual Offset Error
(ppm of Range)
Offset Tempco
(ppm of Range/°C)
53
4.8
INL Error
(ppm of
Range)
128
Absolute Accuracy at
Full Scale* (μV)
3,271
*
Absolute Accuracy at full scale numbers is valid immediately following internal calibration and assumes the device is operating within 10 °C of the last external calibration.
Accuracies listed are valid for up to one year from the device external calibration.
AbsoluteAccuracy = OutputValue · (GainError) + Range · (OffsetError)
GainError = ResidualGainError + GainTempco · (TempChangeFromLastInternalCal) + ReferenceTempco · (TempChangeFromLastExternalCal)
OffsetError = ResidualOffsetError + OffsetTempco · (TempChangeFromLastInternalCal) + INL_Error
Digital I/O/PFI
Static Characteristics
Number of channels
NI 6341
24 total,
8 (P0.<0..7>)
16 (PFI <0..7>/P1,
PFI <8..15>/P2)
NI 6343
48 total,
32 (P0.<0..31>)
16 (PFI <0..7>/P1,
PFI <8..15>/P2)
Ground reference
D GND
Direction control
Each terminal individually programmable as input or output
Pull-down resistor
50 kΩ typical,
20 kΩ minimum
±20 V on up to two pins
Input voltage protection 2
Waveform Characteristics (Port 0 Only)
Terminals used
Port 0 (P0.<0..7>)
NI 6341
Port 0 (P0.<0..31>)
NI 6343
Port/sample size
Up to 8 bits
NI 6341
Up to 32 bits
NI 6343
2,047 samples
Waveform generation (DO) FIFO
255 samples
Waveform acquisition (DI) FIFO
8/18
www.ni.com
DO or DI Sample Clock frequency
0 to 1 MHz, system and bus activity dependent
NI PCIe/PXIe-6341/6343
0 to 1 MHz, system and bus activity dependent
NI USB-6341/6343
Data transfers
DMA (scatter-gather), programmed I/O
NI PCIe/PXIe-6341/6343
USB Signal Stream, programmed I/O
NI USB-6341/6343
160 ns, 10.24 μs, 5.12 ms, disable
Digital line filter settings
PFI/Port 1/Port 2 Functionality
Static digital input, static digital output, timing input, timing output
Functionality
Many AI, AO, counter, DI, DO timing signals
Timing output sources
90 ns, 5.12 μs, 2.56 ms, custom interval, disable; programmable high and
low transitions; selectable per input
Debounce filter settings
Recommended Operation Conditions
Level
Min
Max
Input high voltage (VIH)
2.2 V 5.25 V
Input low voltage (VIL)
0V
0.8 V
—
—
–24 mA
–16 mA
—
—
24 mA
16 mA
Output high current (IOH)
P0.<0..31>
PFI <0..15>/P1/P2
Output low current (IOL)
P0.<0..31>
PFI <0..15>/P1/P2
Electrical Characteristics
Level
Min
Max
Positive-going threshold (VT+)
—
2.2 V
Negative-going threshold (VT–)
0.8 V —
Delta VT hysteresis (VT+ – VT–) 0.2 V —
IIL input low current (Vin = 0 V)
—
–10 μA
IIH input high current (Vin = 5 V)
—
250 μA
Digital I/O Characteristics
9/18
www.ni.com
General-Purpose Counter/Timers
4
Number of counter/timers
32 bits
Resolution
Edge counting, pulse, pulse width, semi-period, period, two-edge separation
Counter measurements
Position measurements
X1, X2, X4 quadrature encoding with Channel Z reloading; two-pulse
encoding
Output applications
Pulse, pulse train with dynamic updates, frequency division, equivalent time
sampling
Internal base clocks
100 MHz, 20 MHz, 100 kHz
External base clock frequency
NI PCIe/USB-6341/6343
0 MHz to 25 MHz
NI PXIe-6341
0 MHz to 25 MHz; 0 MHz to 100 MHz on PXIe-DSTAR<A, B>
Base clock accuracy
50 ppm
Inputs
Gate, Source, HW_Arm, Aux, A, B, Z, Up_Down, Sample Clock
Routing options for inputs
NI PCIe-6341/6343
Any PFI, RTSI, many internal signals
NI PXIe-6341
Any PFI, PXIe-DSTAR<A, B>, PXI_TRIG, PXI_STAR, many internal signals
NI USB-6341/6343
Any PFI, many internal signals
127 samples per counter
FIFO
Data transfers
Dedicated scatter-gather DMA controller for each counter/timer,
programmed I/O
NI PCIe/PXIe-6341/6343
USB Signal Stream, programmed I/O
NI USB-6341/6343
Frequency Generator
1
Number of channels
20 MHz, 10 MHz, 100 kHz
Base clocks
1 to 16
Divisors
50 ppm
Base clock accuracy
Output can be available on any PFI or RTSI terminal.
Phase-Locked Loop (PLL)
1
Number of PLLs
Reference clock locking frequencies
10/18
www.ni.com
Reference Signal
Locking Input Frequency (MHz)
PCIe
PXIe
USB
PXIe_DSTAR<A, B> —
10, 20, 100
—
PXI_STAR
—
10, 20
—
PXIe_CLK100
—
100
—
PXI_TRIG <0..7>
—
10, 20
—
RTSI <0..7>
10, 20
—
—
PFI <0..15>
10, 20
10, 20
10
100 MHz Timebase; other signals derived from 100 MHz Timebase
including 20 MHz and 100 kHz Timebases
Output of PLL
External Digital Triggers
Source
NI PCIe-6341/6343
Any PFI, RTSI
NI PXIe-6341
Any PFI, PXIe-DSTAR<A, B>, PXI_TRIG, PXI_STAR
NI USB-6341/6343
Any PFI
Polarity
Software-selectable for most signals
Analog input function
Start Trigger, Reference Trigger, Pause Trigger, Sample Clock, Convert
Clock, Sample Clock Timebase
Analog output function
Start Trigger, Pause Trigger, Sample Clock, Sample Clock Timebase
Counter/timer functions
Gate, Source, HW_Arm, Aux, A, B, Z, Up_Down, Sample Clock
Digital waveform generation (DO) function
Start Trigger, Pause Trigger, Sample Clock, Sample Clock Timebase
Digital waveform acquisition (DI) function
Start Trigger, Reference Trigger, Pause Trigger, Sample Clock, Sample
Clock Timebase
Device-To-Device Trigger Bus
Input source
NI PCIe-6341/6343
RTSI <0..7> 3
NI PXIe-6341
PXI_TRIG <0..7>, PXI_STAR, PXIe-DSTAR<A, B>
NI USB-6341/6343
None
Output destination
RTSI <0..7> 3
NI PCIe-6341/6343
PXI_TRIG <0..7>, PXIe-DSTARC
NI PXIe-6341
None
NI USB-6341/6343
10 MHz Clock, frequency generator output, many internal signals
Output selections
90 ns, 5.12 μs, 2.56 ms, custom interval, disable; programmable high and
low transitions; selectable per input
Debounce filter settings
Bus Interface
NI PCIe-6341/6343
x1 PCI Express, specification v1.1 compliant
Form factor
Slot compatibility
x1, x4, x8, and x16 PCI Express slots 4
DMA channels
8, analog input, analog output, digital input, digital output, counter/timer 0,
counter/timer 1, counter/timer 2, counter/timer 3
NI PXIe-6341
x1 PXI Express peripheral module, specification rev 1.0 compliant
Form factor
x1 and x4 PXI Express or PXI Express hybrid slots
Slot compatibility
8, analog input, analog output, digital input, digital output, counter/timer 0,
counter/timer 1, counter/timer 2, counter/timer 3
DMA channels
11/18
www.ni.com
All NI PXIe-6341 devices may be installed in PXI Express slots or PXI Express hybrid slots.
NI USB-6341/6343
USB compatibility
USB 2.0 Hi-Speed or full-speed 5
USB Signal Stream
8, can be used for analog input, analog output, digital input, digital output,
counter/timer 0, counter/timer 1, counter/timer 2, counter/timer 3
Power Requirements
NI PCIe-6341/6343
Without disk drive power connector installed
+3.3 V
1.4 W
+12 V
8.6 W
With disk drive power connector installed
+3.3 V
1.4 W
+12 V
3W
+5 V
15 W
NI PXIe-6341
+3.3 V
1.6 W
+12 V
19.8 W
Caution NI USB-6341/6343 devices must be powered with NI offered AC adapter or a National Electric Code (NEC) Class 2 DC source that meets the power
requirements for the device and has appropriate safety certification marks for country of use.
NI USB-6341/6343
Power supply requirements
11 to 30 VDC, 30 W, 2 positions 3.5mm pitch pluggable screw terminal with
screw locks similar to Phoenix Contact MC 1,5/2-STF-3,5 BK
Power input mating connector
Phoenix Contact MC 1,5/2-GF-3,5 BK or equivalent
Current Limits
Caution Exceeding the current limits may cause unpredictable behavior by the device and/or PC/chassis.
NI PCIe-6341/6343
Without disk drive power connector installed
1 A max
P0/PFI/P1/P2 and +5 V terminals combined
With disk drive power connector installed
+5 V terminal (connector 0)
1 A max 6
+5 V terminal (connector 1)
1 A max 6
P0/PFI/P1/P2 combined
1 A max
NI PXIe-6341
+5 V terminal (connector 0)
1 A max 6
+5 V terminal (connector 1)
1 A max 6
P0/PFI/P1/P2 and +5 V terminals combined
2 A max
NI USB-6341/6343
+5 V terminal
1 A max 6
P0/PFI/P1/P2 and +5 V terminals combined
2 A max
Physical Requirements
Printed circuit board dimensions
12/18
www.ni.com
9.9 × 16.8 cm (3.9 × 6.6 in.) (half-length)
NI PCIe-6341/6343
Standard 3U PXI
NI PXIe-6341
Enclosure dimensions (includes connectors)
26.4 × 17.3 × 3.6 cm (10.4 × 6.8 × 1.4 in.)
NI USB-6341/6343
Enclosure dimensions (without connectors)
175.3 x 162.6 mm (6.9 x 6.4 in.)
NI USB-6341/6343 OEM
Weight
104 g (3.6 oz)
NI PCIe-6341
114 g (4.0 oz)
NI PCIe-6343
157 g (5.5 oz)
NI PXIe-6341
1.42 kg (3 lb 2 oz)
NI USB-6341/6343
156 g (5.4 oz)
NI USB-6341 OEM
167 g (5.8 oz)
NI USB-6343 OEM
I/O connector
NI PCIe/PXIe-6341
1 68-pin VHDCI
NI PCIe/PXIe-6343
2 68-pin VHDCI
NI USB-6341
64 screw terminals
NI USB-6343
128 screw terminals
NI PCIe/PXIe-6341/6343 mating connectors:
68-Pos Right Angle Single Stack PCB-Mount VHDCI (Receptacle), MOLEX 71430-0011
68-Pos Right Angle Dual Stack PCB-Mount VHDCI (Receptacle), MOLEX 74337-0016
68-Pos Offset IDC Cable Connector (Plug) (SHC68-*), MOLEX 71425-3001
NI PCIe-6341/6343 disk drive power connector
Standard ATX peripheral connector (not serial ATA)
NI USB-6341/6343 screw terminal wiring
16–24 AWG
Maximum Working Voltage 7
11 V, Measurement Category I
Channel to earth
Caution Do not use for measurements within Categories II, III, or IV.
Environmental
Operating temperature
0 to 50 °C
NI PCIe-6341/6343
0 to 55 °C
NI PXIe-6341
0 to 45 °C
NI USB-6341/6343
–40 to 70 °C
Storage temperature
10 to 90% RH, noncondensing
Humidity
2,000 m
Maximum altitude
2
Pollution Degree (indoor use only)
Shock and Vibration (NI PXIe-6341 Only)
30 g peak, half-sine, 11 ms pulse (Tested in accordance with
IEC-60068-2-27. Test profile developed in accordance with
MIL-PRF-28800F.)
Operational shock
Random vibration
5 to 500 Hz, 0.3 grms
Operating
5 to 500 Hz, 2.4 grms (Tested in accordance with IEC-60068-2-64.
13/18
www.ni.com
Nonoperating
Nonoperating test profile exceeds the requirements of MIL-PRF-28800F,
Class 3.)
Safety Standards
This product is designed to meet the requirements of the following standards of safety for electrical equipment for measurement, control, and laboratory use:
IEC 61010-1, EN 61010-1
UL 61010-1, CSA 61010-1
Note For UL and other safety certifications, refer to the product label or the Online Product Certification section.
Electromagnetic Compatibility
This product meets the requirements of the following EMC standards for electrical equipment for measurement, control, and laboratory use:
EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity
EN 55011 (CISPR 11): Group 1, Class A emissions
AS/NZS CISPR 11: Group 1, Class A emissions
FCC 47 CFR Part 15B: Class A emissions
ICES-001: Class A emissions
Note For EMC declarations and certifications, refer to the Online Product Certification section.
Caution When operating this product, use shielded cables and accessories.
CE Compliance
This product meets the essential requirements of applicable European Directives, as amended for CE marking, as follows:
2006/95/EC; Low-Voltage Directive (safety)
2004/108/EC; Electromagnetic Compatibility Directive (EMC)
Online Product Certification
To obtain product certifications and the DoC for this product, visit ni.com/certification, search by module number or product line, and click the appropriate link in the Certification
column.
Environmental Management
NI is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our
products is beneficial not only to the environment but also to NI customers.
For additional environmental information, refer to the NI and the Environment Web page at ni.com/environment. This page contains the environmental regulations and directives
with which NI complies, as well as other environmental information not included in this document.
Waste Electrical and Electronic Equipment (WEEE)
EU Customers At the end of the product life cycle, all products must be sent to a WEEE recycling center. For more information about WEEE recycling centers, National
Instruments WEEE initiatives, and compliance with WEEE Directive 2002/96/EC on Waste Electrical and Electronic Equipment, visit ni.com/environment/weee.htm.
1
Typical behavior. Time period may be longer due to host system USB performance. Time period will be longer during firmware updates.
2
Stresses beyond those listed under Input voltage protection may cause permanent damage to the device.
3
In other sections of this document, RTSI refers to <0..7> for NI PCIe-6341/6343 or PXI_TRIG <0..7> for NI PXIe-6341.
4
Some motherboards reserve the x16 slot for graphics use. For PCI Express guidelines, refer to ni.com/pciexpress.
5
Operating on a full-speed bus will result in lower performance and you might not be able to achieve maximum sampling/update rates.
6
Has a self-resetting fuse that opens when current exceeds this specification.
7
Maximum working voltage refers to the signal voltage plus the common-mode voltage.
Back to Top
14/18
www.ni.com
Pinouts/Front Panel Connections
NI PCIe/PXIe-6341 Pinout
15/18
www.ni.com
NI USB-6341 Pinout
16/18
www.ni.com
NI PCIe-6343 Pinout
17/18
www.ni.com
NI USB-6343 Pinout
Back to Top
©2012 National Instruments. All rights reserved. CompactRIO, CVI, FieldPoint, LabVIEW, Measurement Studio, National Instruments, NI, ni.com, NI-DAQ, RTSI, SCXI, and SignalExpress are trademarks of
National Instruments. The mark LabWindows is used under a license from Microsoft Corporation. Windows is a registered trademark of Microsoft Corporation in the United States and other countries. Other
product and company names listed are trademarks or trade names of their respective companies. A National Instruments Alliance Partner is a business entity independent from National Instruments and has
no agency, partnership, or joint-venture relationship with National Instruments.
My Profile | RSS | Privacy | Legal | Contact NI © 2012 National Instruments Corporation. All rights reserved.