Hamamatsu H8249-011 Side-on pmt Datasheet

Side-on PMT
Photosensor Modules H8249 Series
The H8249 series photosensor modules incorporate a 28-mm (1-1/8")
diameter side-on photomultiplier tube, a high-voltage power supply circuit and
a low noise amplifier. Two types of amplifiers are available with a current-tovoltage conversion coefficient of 1 V/µA or 0.1 V/µA and a frequency
bandwidth of DC to 20 kHz or DC to 200 kHz. Three types of photomultiplier
tubes are provided as standard lineups for general applications in the visible
range, UV to visible range requiring low-noise, and UV to near IR range.
Product Variations
Type No. Spectral Response Current-to-Voltage
Features
Conversion Factor Frequency Bandwidth
For general applications in visible range
H8249-001 300 nm to 650 nm
For general applications in UV to visible range
H8249-011 185 nm to 680 nm
1 V/µA
DC to 20 kHz
High sensitivity in UV to near IR range. Uses photomultiplier
H8249-101 185 nm to 900 nm
tube with meshless grid for excellent uniformity.
For general applications in visible range
H8249-002 300 nm to 650 nm
For general applications in UV to visible range
H8249-012 185 nm to 680 nm
0.1 V/µA
DC to 200 kHz
High sensitivity in UV to near IR range. Uses photomultiplier
H8249-102 185 nm to 900 nm
tube with meshless grid for excellent uniformity.
Specifications
Anode
Anode
Cathode
Parameter
Suffix
Input Voltage
Max. Input Voltage
Max. Input Current
Max. Control Voltage
Recommended Control Voltage Adjustment Range
Effective Area
Sensitivity Adjustment Range
Peak Sensitivity Wavelength
Min.
Luminous Sensitivity
Typ.
Blue Sensitivity Index (CS 5-58)
Red/White Ratio
Radiant Sensitivity *1
Suffix (with internal 20 kHz amp)
Min.
Luminous Sensitivity *2
Typ.
Radiant Sensitivity *1 *2
Voltage Output Depending Typ.
on PMT Dark Current *2 *3 Max.
Max. Output Signal Voltage
Current-to-Voltage Conversion Factor
Suffix (with internal 200 kHz amp)
Min.
Luminous Sensitivity *2
Typ.
Radiant Sensitivity *1 *2
Voltage Output Depending Typ.
on PMT Dark Current *2 *3 Max.
Max. Output Signal Voltage
Current-to-Voltage Conversion Factor
Offset Voltage *2
Typ.
Ripple Noise *2 *4 (peak to peak) Max.
Settling Time *5
Operating Ambient Temperature
Storage Temperature
Weight
-001 / -002
30
60
7.1
—
60
-001
5.0 × 107
6.0 × 108
600
5
50
-002
5.0 × 106
6.0 × 107
60
0.5
5
H8249 Series
-011 / -012
±11.5 to ±15.5
±18
+50/-1
+1.2 (Input Impedance 100 kΩ)
+0.3 to +1.1
4 × 20
1: 104
400
40
60
6.4
—
60
-011
2.0 × 108
4.0 × 108
400
0.1
2
+10 (Load resistance 10 kΩ)
1
-012
2.0 × 107
4.0 × 107
40
0.01
0.2
+10 (Load resistance 10 kΩ)
0.1
H8249 series
±3
4
0.2
+5 to +45
-20 to +50
Approx. 220
-101 / -102
140
250
8.0
0.3
74
-101
4.0 × 108
2.5 × 109
740
3
50
µA/lm
—
—
mA/W
V/lm
V/nW
mV
V
V/µA
-102
4.0 × 107
2.5 × 108
74
0.3
5
*1: Measured at the peak sensitivity wavelength
*2: Control voltage = +1.0 V
*3: After 30 minute storage in darkness
*4: Cable RG-174/U, Cable length 450 mm, Load resistance = 1 MΩ, Load capacitance = 22 pF
*5: The time required for the output to reach a stable level following a change in the control voltage from +1.0 V to +0.5 V.
32
Unit
—
V
V
mA
V
V
mm
—
nm
V/lm
V/nW
mV
V
V/µA
mV
mV
s
°C
°C
g
Voltage Output Type Photosensor Modules
Characteristics (Anode radiant sensitivity, PMT gain)
1
-101
-102
H8249-002
-012
-102
-011
-012
1
0.1
-001
-002
0.1
0.01
0.01
100
200
300
400
500
600
700
800
108
TPMOB0158EA
107
-101
-102
106
PMT GAIN
PMT GAIN: 105
TPMOB0155EA
ANODE RADIANT SENSITIVITY (V/nW)
ANODE RADIANT SENSITIVITY (V/nW)
H8249-001
-011 10
-101
-001
-002
105
-011
-012
104
103
102
0.25 0.3
0.001
900
0.4
WAVELENGTH (nm)
0.5
0.7
1.0 1.2
1.5
2.0
CONTROL VOLTAGE (V)
Sensitivity Adjustment Method
VOLTAGE PROGRAMMING
PHOTOSENSOR MODULE
RESISTANCE PROGRAMMING
POWER SUPPLY
POWER SUPPLY
PHOTOSENSOR MODULE
SIGNAL OUTPUT
COAXIAL CABLE
RED (LOW VOLTAGE INPUT)
GREEN (LOW VOLTAGE INPUT)
BLACK (GND)
BLUE (Vref OUTPUT)
WHITE (Vcontrol INPUT)
SIGNAL OUTPUT
COAXIAL CABLE
RED (LOW VOLTAGE INPUT)
GREEN (LOW VOLTAGE INPUT)
BLACK (GND)
BLUE (Vref OUTPUT)
WHITE (Vcontrol INPUT)
+15 V
-15 V
GND
CW
+15 V
-15 V
GND
+0.3 V to
+1.1 V
• Adjust the control voltage to
adjust the sensitivity.
• Electrically insulate the
reference voltage output.
MONITOR
POTENTIOMETER (10 kΩ)
When using a potentiometer, adjust sensitivity
while monitoring the control voltage.
GND
TPMOC0162ED
Dimensional Outlines (Unit: mm)
50.0 ± 0.5
+1.0
2.5 -0
2 × 2-M3 DEPTH: 10
C1
REAR VIEW
WINDOW
+2
EFFECTVE
AREA
(4 × 20)
20 -0
8
PHOTOCATHODE
WINDOW
TOP VIEW
SIDE VIEW
+0.5
1.5 -0
55.0 ± 0.5
95.0 ± 1.0
85.0 ± 0.5
34.0 ± 0.8
SIGNAL OUTPUT
BNC-R
4.5 ± 0.3
R3
1: GND
2: Vcc +15 V
3: Vcont +0.3 V to +1.1 V
4: Vref
5: GND
6: Vee -15 V
17.2 ± 0.5
4-M3
DEPTH: 10
FRONT VIEW
55.0 ± 0.5
24.0 +0.5
-0
87.0 ± 0.5
19.0 ± 0.5
8.0 +0.5
-0
POWER INPUT
HIROSE HR10A-7R-6P
10.0 ± 0.5
23.0 ± 0.8
EFFECTIVE
AREA
(4 × 20)
5.0 ± 0.5
38.0 ± 0.5
27.0 ± 0.5
FRONT VIEW
TPMOC0123EC
Power cable with connector (HIROSE HR10A-7P-6S) is supplied with the H8249 series
700 ± 20
BLACK:
RED:
WHITE:
BLUE:
GREEN:
GND
LOW VOLTAGE INPUT (+15 V)
Vcont INPUT (+0.3 V to +1.1 V)
Vref OUTPUT (+1.2 V)
LOW VOLTAGE INPUT (-15 V)
TPMOA0022EB
33