JHL03 Series DC-DC Converter 3 Watts xxx Series • International Medical Approvals • 4000 VAC Reinforced Insulation • Meets IEC60601-1, 3rd Edition • 2 MOPP Isolation at 250 VAC • 2 µA Patient Leakage Current • DIP24 Package • EN55011 Level A With No External Components • 3 Year Warranty Dimensions: JHL03: 1.25 x 0.80 x 0.40” (31.15 x 20.32 x 10.20 mm) Models & Ratings Input Voltage Output Voltage 10-17 V 20-30 V Input Current No Load(1) Full Load(2) Output Current Maximum Capacitive Load(3) Efficiency(4) Model Number 5.0 V 600 mA 56 mA 325 mA 720 µF 76% JHL0312S05 12.0 V 250 mA 72 mA 316 mA 300 µF 78% JHL0312S12 15.0 V 200 mA 67 mA 315 mA 240 µF 78% JHL0312S15 ±12.0 V ±125 mA 43 mA 304 mA ±140 µF 81% JHL0312D12 ±15.0 V ±100 mA 56 mA 303 mA ±120 µF 80% JHL0312D15 5.0 V 600 mA 38 mA 167 mA 720 µF 74% JHL0324S05 12.0 V 250 mA 37 mA 165 mA 300 µF 78% JHL0324S12 15.0 V 200 mA 23 mA 146 mA 240 µF 82% JHL0324S15 ±12.0 V ±125 mA 29 mA 150 mA ±140 µF 80% JHL0324D12 ±15.0 V ±100 mA 42 mA 166 mA ±120 µF 80% JHL0324D15 Notes 1. Input current measured at nominal input voltage. 2. Input current measured at lowest input voltage. 3. Maximum capacitive load is per output. 4. Typical values. Mechanical Details 1.25 (31.75) 14 0.80 (20.32) Pin Connections 0.10 (2.54) 0.20 (5.08) 16 22 23 0.60 (15.24) Bottom View 11 10 0.18 (4.57) 0.40 (10.20) 9 3 2 0.90 (22.86) Notes: 1. Pin diameter: 0.02 ±0.002 (0.5 ±0.05) 2. Pin pitch tolerance: ±0.014 (±0.35) 3. Case Tolerance: ±0.02 (±0.5) 0.12 (3.05) Min. 1 www.xppower.com Pin Single Dual 2 -Vin -Vin 3 -Vin -Vin 9 No Pin Common 10 Trim Trim 11 No Pin -Vout 14 +Vout +Vout 16 -Vout Common 22 +Vin +Vin 23 +Vin +Vin Notes 1. All dimensions are in inches (mm) 2. Weight: 0.04 lbs (20 g) approx. 3. Pin diameter: 0.02 ±0.002 (0.5 ±0.05) 4. Pin pitch tolerance: ±0.014 (±0.35) 5. Case tolerance: ±0.02 (±0.5) JHL03 Series DC-DC Converter Input Characteristic Input Voltage Range Minimum Typical Units 17 VDC 12 V nominal 20 30 VDC 24 V nominal 25 A 2 µA Input Current See Models and Ratings table Inrush Current Input Filter At 30 VDC input Pi type Patient Leakage Current Undervoltage Lockout Notes & Conditions Maximum 10 On at >8.8 V. Off <8.3 V 12 V models On at >17.5 V. Off <17.0 V 24 V models Input Surge 25 VDC 12 V models for 3 s 50 VDC 24 V models for 3 s Maximum Units Notes & Conditions 30 V See Models and Ratings table ±10 % Via external resistors, see Application Notes ±1 % on V1 ±2 % on V2 of dual output models A No minimum load required Output Characteristic Minimum Output Voltage 5 Typical Output Voltage Trim Initial Set Accuracy Minimum Load 0 Start Up Delay 5 Start Up Rise Time 2 ms ms Line Regulation ±0.3 % Load Regulation ±1 % 0 - 100% load On dual output models with one output set to 50% load and the other varied from 10% to 100% load (D05 20% to 100%) Cross Regulation ±4 % Transient Response 4 % deviation Ripple & Noise 1 % pk-pk Short Circuit Protection Recovery to within 1% in <500 µs for a 50% load change at 0.25 A/µs rate 20 MHz bandwidth Trip & Restart (hiccup mode), auto recovery Overload Protection 120 200 % Overvoltage Protection 115 140 % 0.03 %/ ˚C Maximum Units Temperature Coefficient Trip & Restart (hiccup mode) General Characteristic Minimum Efficiency Isolation Typical 80 % Patient Leakage Current 2 Input to Output Capacitance 20 Switching Frequency pF kHz Power Density 7.5 Weight For 1 min. Double/reinforced with a working voltage of 250 VAC. Meets 2 x MOPP per 3rd edition of IEC60601-1 5000 VAC for 10 ms in accordance with IEC60664-1 µA 250 Mean Time Between Failure See Models and Ratings table VAC 4000 Notes & Conditions W/in3 >1 MHrs 0.04 (20.0) lb (g) MIL-HDBK-217F, +25 °C GB Environmental Characteristic Minimum Operating Temperature -20 Storage Temperature -40 Case Temperature Humidity 5 Typical Maximum Units Notes & Conditions +80 °C See derating curve +100 °C +100 °C 90 %RH Cooling Non-condensing Natural convection Shock ±3 shocks in each plane, total 18 shocks of 30 g : 11 ms halfsine. Conforms to EN60068-2-27 & EN60068-2-47 Vibration 10-500 Hz at 2 g sweep and endurance at resonance in all 3 planes. Conforms to EN60068-2-6 www.xppower.com 2 JHL03 Series DC-DC Converter EMC: Emissions Phenomenon Standard Test Level Conducted EN55011 Level A Radiated EN55011 Level A Notes & Conditions EMC: Immunity Standard Test Level Criteria Immunity Phenomenon IEC60601-1-2 Ed 4.0: 2014 As Below ESD Immunity EN61000-4-2 ±8 kv Contact, ±15 kv Air A Radiated Immunity EN61000-4-3 10 V/m A EFT/Burst EN61000-4-4 2 A Surges EN61000-4-5 1 A Conducted Immunity EN61000-4-6 3 Vm A Magnetic Fields EN61000-4-8 30 A/m A Safety Approvals Notes & Conditions 80 MHz - 2.7 GHz plus discrete communication proximity field frequencies ANSI/AMMI ES60601-1 3rd Edition, CSA-22.2 No.60601-1:2008, IEC60601-1 3rd Edition Safety Approvals Safety Agency Safety Standard Notes & Conditions CB Report IEC60601-1 Including Risk Management Medical UL ANSI/AAMI ES60601-1 3rd Ed. & CSA C22.2, No.60601-1:2008 Medical EN EN60601-1 Medical Application Notes Derating Curve 100 1 2 3 Load (%) 80 60 40 20 0 -20 -10 0 10 20 30 40 50 60 70 80 90 Ambient Temperature (°C) External Output Trim Trim + Vout Trim Up -Vout 3 Trim Down Trim For 5 V output: Trim +10%, R = 3.4 k typical Trim -10%, R = 1.1 k typical For ±12 V output: Trim +10%, R = 12.8 k typical Trim -10%, R = 9.5 k typical For 12 V output: Trim +10%, R = 5.9 k typical Trim -10%, R = 11.3 k typical For ±15 V output: Trim +10%, R = 18 k typical Trim -10%, R = 14.8 k typical For 15 V output: Trim +10%, R = 8.4 k typical Trim -10%, R = 10.4 k typical www.xppower.com 10 Mar 16