Fuji EDSX10-60M Microjet recorder (180mm) Datasheet

MICROJET RECORDER (180mm)
DATA SHEET
PHA
This recorder can record a maximum of 12 channels of
DC voltage, mA, thermocouples and resistance bulbs.
The adoption of an ink jet system makes it possible to
record measured data in analog trace mode or to print in
digital mode at a high speed. This 180mm-wide recorder
performs recording clearly in 6 different colors.
FEATURES
1. Compact size
Compact and lightweight design, 199mm in depth and
about 6kg in mass{weight}.
2. High quality recording
• Ink jet system is used for recording and printing
measured data in 6 different colors at a high speed.
Operating noise is minimized.
• Six and twelve continuous traces without pen offset
are possible with this compact size of recorder; a
unique recording system is used for the first time in
the industry.
• Scale of each channel is printed on chart paper,
elimi- nating the need for scales.
3. Easy setting of input signals
DC voltage input (5mV span, 50V max.), 12 kinds of
thermocouples (Type B, R, S, K, E, J, T, N, W, L, U, PN)
and resistance bulbs (Pt100) can be set for each channel.
4. Digital printing
Beside the analog recording of measured data, digital
printing is also available (periodic printing, list printing,
alarm printing, daily report printing, message printing).
• Periodic data printing: Channel No., date, time, unit,
chart speed, measured value
• List printing: Date, time, unit, recording range, scaling value, alarm set value, chart speed, Tag No.
• Alarm printing: Channel No., alarm type, on/off time,
output relay No.
• Daily report and totalized data printing: Printing of
maximum, minimum, average and total of data
measured during maximum 24 hours
• Message printing : 10 messages, 16-character userentered messages
5. Interactive key operation
Fluorescent indicator is used to clearly indicate alphanumeric characters and symbols.
Input mode, recording range, alarm value, chart speed,
etc., can be set according to the comments indicated
by the display and operating keys. No bothersome operation is required.
6. Easy handling
• A cartridge type recording device is used for easy
replacement.
• Chart paper can be loaded without drawing out the
internal unit of the recorder.
• Shortage of ink is detected in early stages and an
alarm is given to the operator.
• The end of chart paper is detected and indicated on
the front panel display.
• Shortage of ink and the end of chart paper alarm
output is possible.
7. Full variety of functions
• Alarm relay output/external control (record start/
stop, chart speed change, data printing, message
printing). This unit can easily be connected to the
recorder by user (option).
• Chart paper illumination lamp (option): The result of
printing can be checked even in lower light.
• Burnout function is provided as a standard.
• Various recording: Enlarged/reduced recording,
auto-range recording, zone recording.
• Calculation: Square root extraction, subtraction, engineering unit conversion, logarithm.
• Language: Selectable 3 languages in display and
printing.
• Passcode security is configurable.
• The message print and alarm print function are operational, even when the recording mode is off.
• All parameters of recording format, daily report,
totalize, message and periodic data printing can be
printed cut.
EDSX10-60m
Date
Nov. 24, 2011
PHA
Chart speeds:
SPECIFICATIONS
Input system
Input points:
6 or 12 continuous recording and intermittent recording
Input signal:
Thermocouple input…B, R, S, K, E, J, T,
N, W, L, U, PN
Resistance bulb input…Pt100Ω
DC voltage input…50mV, 500mV, 5V,
50V range
DC current input…4 to 20mA DC, 10 to 50mA DC
(Shunt resistor (option) need to be
connected to the terminal)
Max. input voltage:
• Thermocouple, resistance bulb and
DC voltage (50mV, 500mV range)
…±10V DC or less
• DC voltage input (5V, 50V range)
…±100V DC or less
Input signal setting and change:
Setting and change of input signal between thermocouple, resistance bulb
and DC voltage (50mV, 500mV, 5V, 50V
range) is possible for each channel by
the setting pin in the instrument.
Setting of recording range:
Setting is possible within the reference
range by using the keyboard.
Burnout function:When thermocouple or resistance bulb
input is disconnected, the recording is
deflected to 100%.
Reference range:
Kind
Reference range
Thermocouple
Resistance
bulb
DC voltage
Note: N
W
L
U
PN
Pt100
B
R
S
K
E
J
T
N
W
L
U
PN
400
0
0
–200
–200
–200
–200
0
0
–200
–200
0
to
to
to
to
to
to
to
to
to
to
to
to
1760°C
1760°C
1760°C
1370°C
800°C
1100°C
400°C
1300°C
1760°C
900°C
400°C
1300°C
Pt100
–200
to
600°C
–50
–500
–5
–50
to
to
to
to
+50mV
+500mV
+5V
+50V
Reference range
752
32
32
–328
–328
–328
–328
32
32
–328
–328
32
to
to
to
to
to
to
to
to
to
to
to
to
3200°F
3200°F
3200°F
2498°F
1472°F
2012°F
752°F
2372°F
3200°F
1652°F
752°F
2372°F
–328
to
1112°F
Scaling is possible
within the range of
–32767 to +32767
(decimal point may
be put as necessary)
: NICROSIL-NISIL (IEC584)
: +side 5% Re, –side 26% Re.W (Hoskins Mfg. Co.,
U.S.A.)
: +side Fe, –side Cu.Ni alloy (DIN43710)
: +side Cu, –side Cu.Ni alloy (DIN43710)
: Platinel
: DIN IEC751
Recording system
Writing system: Ink jet system, 6 colors
Chart width:
180mm
Recording color: No. 1,7 channel (orange), No. 2.8 channel (green), No. 3, 9 channel (purple),
No. 4, 10 channel (red), No. 5, 11 channel (black), No. 6, 12 channel (blue)
Recording color can be assigned for
each channel.
Chart length:
Z fold 20m
2
Continuous recording type
5 to 300 mm/h, continuous recording
301 to 1500 mm/h, intermittent recording
Intermitter recording type
5 to 1500 mm/h
Each can be set in 1 mm/h steps.
Recording cycle: Intermitter recording…30sec/all points
Continuous recording…Depends on
chart speed.
<Calculation formula>
Recording
450
=
cycle[sec] Chart speed [mm/h]
(not faster than 3 seconds.)
Measuring cycle: Up to 3 inputs…160ms
6, 12 inputs…320ms
Service life of ink: (Depends on operating conditions)
About 6 months for 6 points of linear
recording at 25 mm/h of chart speed
Chart handling:
Tear off without disturbance of recording.
Indicating system
Indication:
Fluorescent indication (blue-green),
20 characters × 2 lines
Characters indicated:
5 × 7 dots, 5.0mm high, 3.5mm wide
Contents of indication:
(1) Measured value:
Temperature…1 digit below decimal
point;
Voltage…6 digits (including sign and
decimal point)
M easured value of No.1 channel to
No.6 or No.7 channel to No.12 can be
indicated simultaneously.
(2) Channel No.: 2 digit (1 to 12)
(3) Engineering unit:
Max. 7 digits (°C, °F, %, Pa, bar, ppm,
m3/h, etc.)
(4) Tag No….8 characters
(5) Time: Year, month, day, hour, minute
(6) Status indication:
R
ecord ON, chart end, battery alarm,
alarm, ink shortage alarm, burnout,
carriage failure
Configuration:
These can be set according to the
comments indicated by operating
keys as follows,
Passcode
Main chart speed
Sub chart speed
Alarm setting
Record mode (trend/logging)
Recording range
Input signal
List print request
Tag No.
Daily report setting
Totalize function
Communication parameter
Date and time setting
Ink monitor clear
Illumination on/off
Message definition
Measured value shift
Printing system
Writing system: Ink jet system, 6 colors
Periodic data printing:
Measured value, unit, date, time, time
line, chart speed, channel No.
List printing:
(1) Measured value list (date, time, channel No., measured value, unit)
(2) P
arameter list (date, time, channel No., recording range,
scaling, unit, alarm set value, chart speed, Tag No.)
(3) Test pattern (all characters and color patterns)
Message printing: 1 0 m e s s a g e s , 1 6 - c h a r a c t e r u s e rentered messages.
Alarm printing:
Channel No., alarm type (H, L, RH,
RL), output relay No., on/off time
Burnout printing: Burnout channel No. and time
Other:
Ink shortage message, automatic
range selection mark, recording start
mark, chart speed change mark
Note: Printing is not available for more than 301 mm/h (continuous
recording), or more than 51 mm/h (intermittent recording).
Performance and characteristics
Accuracy and resolution:
Performance under reference condition
(23±2°C, 65±10%RH, power voltage and
frequency variation ±1%, warm-up time
30 minutes or more, vertical mounting,
free from the effect of external noise)
Resistance
bulb
DC
voltage
Physical data
Mounting:
Panel (may be inclined up to 30° backwards from the vertical.)
Indication (digital) Recording
Input
Thermocouple
Power supply terminal - ground
…2000V AC, 1 minute
Input terminal - ground
…500V AC, 1 minute
Power terminal - input terminal
…500V AC, 1 minute
Between alarm terminals
…750V AC, 1 minute
(Leakage current 5mA or less, however, if the power supply is 24V DC,
Leakage current of “Power supply
terminal - ground” is 10mA or less)
Reference junction compensation accuracy:
K, E, J, T, N, L, U, PN ............. ±0.5°C
(In case of minus input measurement:
±1.2°C)
R, S, B, W .............................. ±1°C
(In case of minus input measurement:
±2.4°C)
Common mode noise rejection:
120dB at 50, 60Hz ±0.1Hz
Series mode noise rejection:
30dB at 50, 60Hz ±0.1Hz
Accuracy
B
R
S
K
E
J
T
N
W
L
U
PN
±(0.15%
+1 digit)
(without
reference
junction
compensation
error)
Pt100
±(0.15%
+1 digit)
–50 to +50mV
±(0.15%
–500 to +500mV +1 digit)
–5 to 5V
–50 to 50V
Resolution
0.1°C
0.1°C
0.1°C
0.1°C
0.1°C
0.1°C
0.1°C
0.1°C
0.1°C
0.1°C
0.1°C
0.1°C
Accuracy
Material:
Indication
accuracy,
±0.25%
of recording span
0.1mm
minimum
0.1°C
10 µV
100 µV
1mV
10mV
Note: Indication accuracy is in % of reference range.
Indication accuracy of B type TC is ±(0.36%+1digit) between
400°C to 600°C.
Indication accuracy of all type TC is ±(0.36%+1digit) between
–200°C to –100°C.
Input resistance:
=90° to 60°
Resolution
Thermocouple:>10MΩ
50mV range: >10MΩ
500mV range: >100kΩ
5V and 50V range: >1MΩ
Chart speed accuracy:
±0.1% (expansion and contraction of
paper is not included)
Clock accuracy:
±50ppm or less (monthly error; about
2 minutes)
Insulation resistance:
100MΩ or more (between each terminal and earth, at 500V DC)
Withstand voltage:
Input terminal - input terminal
…500V AC, 1 minute
Case...Steel plate
Front door frame…Polycarbonate with
glass wool
Mass{weight}:
Approx. 6 kg (without option)
Approx. 7 kg (with option)
Case size:
Bezel 288x288mm
Depth 199mm
Cutout 281x281mm
Finish color:
Case…Black, Front door frame…Black
External terminals: Screw terminal (M4 screw)
Power requirement
1) S upply voltage 100 to 240V AC products (9digit of
code symbol = “E”)
• Power supply voltage:
100V (-15%) to 240V (+10%) AC (Free
power supply)
• Supply frequency:
50/60Hz both employable
• Power consumption:
1 00V AC with all options approximately 37VA
2 40V AC with all options approximately 56VA
2) Supply voltage 24V DC products (9digit of code symbol = “L”)
• Power supply voltage:
24V (±10%) DC
• Power consumption:
2 6.4V DC with all options 37VA or
less
3
PHA
Operating environment
(for devices to operate continuously)
Temperature limits:
0 to 50°C
Humidity limits:
20 to 80%RH, non condensing is required
(temperature x humidity<3200)
Vibration:
10 to 60Hz, 0.2m/s2{0.02G} or less
Mounting position:
Front inclination 0°, rear inclination 30°,
left/right inclination 0°
Signal source resistance:
Thermocouple input…1kΩ or less
Voltage input…Less than 0.1% of input resistance
Resistance bulb input…Less than 10Ω
per line (Resistance of each wire of
3-wire system should be balanced
with others.)
Warm-up time:
30 min or more
Shock:
No external shock
Environmental protection:
IEC IP50(Front) / 20(Terminal)
Installation category: Ⅱ
Pollution degree: 2
Operating altitude: 2000m max.
Operating environment influence
Effect of power supply fluctuation :
(1) Supply voltage 100 to 240V AC products (9digit of
code symbol = “E”)
Voltage variation: 85 to 264V AC
(frequeucy: 50/60Hz) 100VAC base
Change in indication… ±(0.1%+1
digit) max.
Change in recording… ±0.2% of
recording span, max.
Frequency variation: 47 to 63Hz
(power voltage: 100V AC) 50Hz base
Change in indication… ±(0.1%+1
digit) max.
Change in recording… ±0.2% of
recording span, max.
(2) S upply voltage 24V DC products (9digit of code
symbol = “L”)
With 21.6 to 26.4V DC fluctuation,
24V DC base
Change in indication....±(0.1% + 1
digit) max.
Change in recording....±0.2% of
recording span, max.
Input signal source resistance or wiring resistance influence:
Thermocouple…10 µV per 100Ω
Vo l t a g e i n p u t … Va r i a t i o n o f 0 . 1 %
change of resistance
Change in indication… ±(0.1%+1
digit) max.
Change in recording… ±0.2% of
recording span, max.
Reistance bulb…Variation of resistance with changes in 10Ω per wire
Change in indication… ±(0.1%+1
digit) max.
Change in recording… ±0.2% of recording span, max.
(3 wires should be balanced.)
4
Temperature influence:
Change in indication… ±(0.3%+1
digit)/10°C, max.
Change in recording… ±0.5%/10°C,
max.
Mounting position influence:
Inclination within 30°
Change in indication… ±(0.1%+1
digit) max.
Change in recording… ±0.2% of
recording span, max.
Vibration influence:
Linear vibration with 10 to 60Hz of
frequency and 0.2m/s 2 {0.02G} of
acceleration is applied to each of 3
directions for 2 hours.
Change in indication… ±(0.1%+1
digit) max.
Change in recording… ±0.2% of
recording span, max.
Effect of external noise:
Normal mode noise (50,60Hz±0.1Hz)
…30dB or more
ommon mode noise (50,60Hz±0.1Hz)
C
…120dB or more
Chart paper influence:
Standard temperature/humidity: 20°C,
65%RH
Expansion at 85%RH…0.4% max.
Contraction at 35%RH…0.5% max.
Alarm
Setting method: Setting from keyboard
Number of alarm levels:
Max. 4 levels for each channel
Alarm type:
High(H), Low(L), High-rate of
change(RH), Low-rate of change(RL)
Alarm action indication:
Kind of alarm and output relay No.
are indicated for each channel at occurrence of alarm.
Printing:
Channel No., kind of alarm, output relay No. and on/off time are printed on
chart paper.
Output:
See optional specifications.
Hysteresis:
Approx. 0.5% of recording span
Alarm timing:
Recognition; 1 second (worst case)
Action; additional 1 second (worst case)
Alarm latch:
Hold the alarm display and alarm output.
Others:
Shortage of ink and the end of chart
paper alarm output is possible.
Transportation/storage condition
(Detach a recording head from the main unit before
transportation)
Temperature limits:
–10 to +60°C
Humidity limits:
5 to 90%RH, non condensing is required
Vibration:
10 to 60Hz, 2.45m/s2{0.25G}
Shock:
294m/s2{30G} or less
Optional specifications
1. Chart illumination:
LED
2. Alarm output/3-point external control:
This unit can be mounted from the
rear side of the recorder.
(1) Alarm output (DO):
6 or 12 points of relay contact N.O. (1a)
output for individual channel operation or common operation
Maximum contact voltage 240V AC,
30V DC
Maximum contact current 3A
(resistive load)
(2) External control (DI):
The following control is possible with
external contact signal.
• Recording start/stop;
Recording start/stop is effected
by contact signal. Recording is
started when contact is closed and
stopped when contact is open.
• Chart speed change;
Selection between normal and remote chart speeds is effected by
contact signal. Remote chart speed
is selected when contact is closed
and normal when contact is open.
• Measured value printing;
Measured value list printing (date,
time, channel No., measured value,
unit) is effected by contact signal.
Printing is started when contact is
closed.
• Message printing
Periodic data
printing function
Time, date, chart speed, measured value and unit
can be printed at fixed intervals.
Printing can be enabled/disabled from keyboard.
Message
printing
Maximum 10 messages, 16- character userentered messages can be printed.
Alarm printing
function
Time, channel No., kind of alarm, and output relay No. can be printed when alarm is on or off.
Unit indication
Engineering units such as °C, °F, %, mV, mA,
Pa,L, etc., are indicated (setting from key board).
Scaling function
Scaling with DC voltage input is possible.
(Setting of decimal point is also possible within
the range of –32767 to +32767).
Subtract function
Difference between any channels is recorded
(channel is set from keyboard).
Logarithm
Measured value can be displayed and printed by
10n power
Auto-range
recording
Recording range is automatically changed for recording in the event of overrange or underrange
(setting with keyboard).
This function is not available for combination of
zone recording and enlarged/reduced recording.
Zone recording
Recording area is divided into a maximum of 4
zones for recording.
This function is not available for combination of
automatic range selection and enlarged/reduced
recording.
Enlarged/reduced
recording
A part of recording area of each channel is expanded or contracted for recording.
This function is not available for combination of
automatic range selection and zone recording.
Square-root extraction function
Square-root extraction of DC voltage input is possible.
Daily report function
Measured value of every hour for maximum one
day (24 data) in each channel is stored for printing. Maximum, minimum and average values are
also printed at the same time.
ON-OFF operation, ON-OFF of each channel and
operation start time/stop time can be set from
keyboard.
Totalize function
Integrated value of every hour for maximum one day (24 data) in each channel is stored for printing (integration in 1 sec steps). Possible to print
total value only.
Total value is also printed at the same time.
ON-OFF operation, ON-OFF of each channel and
operation start time/stop time can be set from
keyboard.
Measured value
shift
Shift the zero point and inclination of the measured value so that the measured value can be
adjusted according to other instruments.
Memory backup
Set data and clock function are protected by
built-in lithium battery (expected battery life,
about 10 years under normal temperature).
Input filter
Response is delayed according to sudden changes in input of each channel (1st order lag filter).
Time constant setting range: 0 to 900 sec (setting
from keyboard).
Burnout function
When thermocouple or resistance bulb input is
disconnected, it is deflected 100%. Also, it is
indicated and printed at the same time.
Passcode
4 digits passcode security is available.
Language
English, German, or French is selectable for display and printing.
Alarm latch
function
The alarm display and alarm output are held even
after the cause of alarming was gone. ON-OFF
operation can be set from keyboard. Cancellation of the held alarm can be made from external
control (DI).
Parameter
copy
Set parameters on any channel can be copied to
any other channels.
Note: For external control, use a dry contact.
Contact capacity: 12V DC, 0.05A,
N.O. (1a) contact
3. Transmission function:
T-link interface for transmitting measured value and receiving the condition of setting.
Transmission method
Half-duplex bit serial
Transmission speed
500kbps
Transmission distance
Max. 500m
I/O frame
8w or 16w
Message frame
Available to set/change parameters etc.
FUNCTIONS
Function
Description
Recording range can be set for each channel.
Input setting
Any input can be set for each channel.
Skip function
Used to skip recording, indication and alarm at
any measuring point.
List printing function
Range setting
Measured
value list
Date, time, and measured value unit can be printed.
Parameter
list
Date, time, recording range, scaling, unit, kind of
input, alarm set value, chart speed, and Tag No.
can be printed.
Test pattern
All characters and color patterns can be printed.
5
PHA
SCOPE OF DELIVERY
CODE SYMBOLS
1 2 3 4 5 6 7 8 - 9 10111213
PHA
0 0 4 6
7
8
9
V
Description
Recording points
6 continuous recording
6 intermittent recording
12 intermittent recording
12 continuous recording
6
6
8
8
E
L
A
B
2
Y
T
Note: Ink cartridge is not mounted on the recorder at the time of delivery.
Power Supply
100 to 240V AC 50/60Hz
24V DC
Spare parts
Chart paper illumination
Without
With
Ink cartridge
PHZH1002
Chart paper (0 to 100,
100 uniform division)
PEX00BL1-1000B
Alarm output/external control
Without
6-point alarm output/3-point
external control
12-point alarm output/3-point
external control
0
1
Recorder, panel mounting bracket, accessories (ink cartridge (1), fuse (1), chart paper (1), ink absorption cloth (1)).
Instruction manual (1).
Transmission function
Without
With T-Link
Remarks: Input signal
Setting prior to delivery is as follows.
• Thermocouple K: 0 to 1200 °C
Note: Contact Fuji Electric for additional features
not listed such as Flow integration record
and Calculation of input signals.
Item
Part No.
Unit of quantity for sale
1 pc
1 box (6 charts)
Other (optional items)
Item
Type
Specification
Shunt resistor
PHZT8101
For 10Ω ±0.1%
Alarm output/
external control
unit
PHZK8601
With 6-point alarm output/3-point
external control
PHZK8201
With 12-point alarm output/3-point
external control
The product conforms to the requirements of the
Electromagnetic compatibility Directive 89/336/EEC as
detailed within the technical construction file number
TN510405. The applicable standards used to demonstrate
compliance are:EN 55011 : 1991 CLASS A
Conducted and Radiated emissions
EN 50082-1 :-1992
Radiated immunity, ESD and FBT
PHA
6
OUTLINE DIAGRAMS (Unit:mm)
Alarm unit(option)
(Alarm 7 to 12)
2
Alarm/external control
unit(option)
(Alarm 1 to 6, DI 1 to 3)
Transmission terminal
(option)
T 30
t
Mounting
bracket
Panel
Input terminal 1
(Input 1 to 6)
288
280
Input terminal 2
(Input 7 to 12)
Power terminal
39
280
288
160
199
312
18
360 min.
Panel cutout
320 min.
281+20
281+20
CONNECTION DIAGRAMS
Input terminal 2
Terminal
No.
Thermocouple
Resistance Voltage
+
buld
–
Input terminal 1
71
72
Input 7
73
81
82
+
–
+
–
11
12
Input 2
31
32
33
Input 9
41
102
42
Input 10
103
43
111
51
123
Transmission
terminal
(option)
21
12
22
13
23
DI 1
DI 2
DI 3
TRX1
TRX2
SHLD
–
+
Power terminal
23
101
122
11
Input 1
13
21
22
Input 8
91
112
113
121
Alarm/external control
unit (option)
(Alarm 1 to 6, DI 1 to 3)
Terminal
No.
83
92
93
Alarm unit
(option)
(Alarm 7 to 12)
Input 11
Input 12
Input 3
Input 4
Input 5
+
52
53
61
–
62
Input 6
63
31
41 Alarm 7
14
24 Alarm 1
32
42 Alarm 8
15
25 Alarm 2
33
43 Alarm 9
16
26 Alarm 3
34
44 Alarm 10
17
27 Alarm 4
35
45 Alarm 11
18
28 Alarm 5
36
46 Alarm 12
19
29 Alarm 6
DI 1
DI 2
DI 3
AC
AC
G
Power source
100 to 240V AC
50/60Hz
+
-
G
Power source
24V DC
Record start or message print 1
Chart speed change or message print 2
Measured value printing or alarm latch cancel
or message print 3
7
PHA
Caution on Safety
*Before using this product, be sure to read its instruction manual in advance.
International Sales Div
Sales Group
Gate City Ohsaki, East Tower, 11-2, Osaki 1-chome,
Shinagawa-ku, Tokyo 141-0032, Japan
http://www.fujielectric.com
Phone: 81-3-5435-7280, 7281 Fax: 81-3-5435-7425
http://www.fujielectric.com/products/instruments/
Information in this catalog is subject to change without notice.
Printed in Japan