Astrodyne AHP6301-09 63w medical power adapters for health care application Datasheet

AHP6301
63W Me d i c a l P o wer A dapt ers f or H ealt h Ca re Ap p lica tio n s
series
Wide 80-275VAC Input
Medical Safety IEC60601-1 3rd Edition; 2MOPP
medical
Choice of IEC320-C14, C8 or C6 Input Receptacles
Single Outputs from
12 to 48 VDC
HPU63B
series
63
The HPU63B
Efficiency
Level series
V of AC/DC switching mode power supplies
provide 63 Watts of continuous output power . All supplies are
min compliant. All models meet FCC Part-18 class B and
ClassUL94V-1
I or Class
II
CISPR-11 EN55011 class B emission Limits and are designed to
comply with ANSI/AAMI ES 60601-1: 2005(UL/cUL 3rd Edition), EN
60601-1:2006 (TUV/T-mark 3rd Edition) and new CE requirements.
All units are 100% burned in and tested.
RoHS2
2011/65/EU
Model Number
Output Watts Output Voltage Output Amps
Efficiency Total Regulation
*AHP6301(A,B,C)-05
*AHP6301(A,B,C)-06
*AHP6301(A,B,C)-07
*AHP6301(A,B,C)-08
*AHP6301(A,B,C)-09
*AHP6301(A,B,C)-10
*AHP6301(A,B,C)-11
63
63
63
63
63
63
63
84%
±5%
85%
±5%
85%
±5%
85%
±3%
86%APPROVALS:
±3%
86%
±3%
86%
±3%
* Note:
12VDC
15VDC
18VDC
24VDC
30VDC
36VDC
48VDC
5.25
4.20
3.50
2.62
2.10
1.75
1.31
* Wide O
* IEC-32
* Single
* Over V
* Medic
* Input t
* Energy
* Class II
* 3 year
APPLIC
* Home
* Patien
* Blood
* Portab
* ECG m
GENER
CB
EN 60601-1
IEC 60601-1
PS
E
Electrical Characteristics:
Symbol
FEATU
* Short C
* Coolin
* Flamm
* Protec
* Safety
(TUV/T-m
Characteristic
Safety
Approval
Input Voltage
Range
Safety Approval
Specification i
Vins (A)
The Asterisk * denotes a choice of receiving this product with an IEC-320-C14 3 Pole Input Receptacle
or with
an IEC-320-C8
2 Pole
Input Receptacle
(B) or &
with
an IEC-320-C6 3 Pole Input Receptacle (C). See the Mechanical files at the end of this specification
Simply add
the letter
“A”, “B” or “C”
afterto“AHP6301
Input Operate
Voltage
Range
Detail
see Fig.1 in
(Derate linearly
Vin for clarity.
the model number (i.e. “AHP6301A-05” for C14 or “AHP6301B-05” for C8 or “AHP6301C-05” forFiC6) when
ordering
to receive the desired version.
Input
Frequency
Sine wave
WWW.ASTRODYNETDI.COM
Po
Output Power Range
See Rating Chart
Iil
Low Line Input Current
Full Load, Vin=100VAC
Iih
High Line Input Current
Full Load, Vin=240VAC
Irl
Low Line Input Inrush Current
Full Load, 25°C, Cool start, Vin=1
Irh
High Line Input Inrush Current
Full Load, 25°C, Cool start, Vin=2
Efficiency
Full Load, Vin=230VAC, Detail to
△Voi
Line Regulation
Full Load, Vin=100~120VAC or 20
OVP
Over Voltage Protection
OLP
Over Load Protection
Recovers automatically after fau
ttr
Time of Transient Response
Full Load, Vin=110VAC
thu
Hold-Up Time
Full Load, Vin=100VAC
ts
Start-up time
Full Load, Vin=100~240VAC
Tc
Temperature Coefficient
All Condition
HV
Dielectric Withstanding Voltage (P-S)
Primary to Secondary, limit curre
EMI
EMC Emission
Compliance to EN55011 (CISPR1
ASTRODYNE USA: 1-800-823-8082
ASTRODYNE PACIFIC: 886-2-26983458
Environmental:
Symbol
Characteristic
63W Me d i c a l P o wer A dapt ers f or H ealt h Ca re Ap p lica tio n s
INPUT SPECIFICATIONS
Input Voltage Range
Frequency Range
Input Current
Inrush Current, typ (Cold Start)
Leakage Current (240 Vin)
No Load Power Consumption
80-275 VAC (100-240 Nominal)
47-63 Hz
Low Line (100VAC)
1.62A max.
High Line (240VAC)
0.72A max.
Low Line (100VAC)
37A max.
High Line (240VAC)
74A max.
0.1mA
0.5 Watts
See Selection Charts
0 to FL Rating
±5%, max. 60%L, changing ±10%
±1%, max. changing line ±10%
110-150%, Auto-Recovery
112-132%
Auto Recovery
110Vin, FL
4mS
±0.04%/°C, typ
100mV Pk-Pk max.
100~240Vin, FL
12mS min.
100~240Vin, FL 0.3S typ., 2S max.
GENERAL SPECIFICATIONS
Input-Out Isolation
Input-GND
Isolation Resistance
Efficiency
Safety
Protection Class
series
PHYSICAL SPECIFICATIONS
OUTPUT SPECIFICATIONS
Voltage and Current
Output Power Range
Load Regulation
Line Regulation (FL)
Over Load Protection*
Over Voltage Protection*
Short Circuit Protection
Transient Response
Temperature Coefficient
Ripple/Noise (Note 1)
Hold Up Time (Note 2)
Start Up Time
AHP6301
4000 VAC, Limit Current <10mA
2800 VDC
50MΩ
(FL, 230Vin) See Selection Chart
ANSI/AAMI ES 60601-1:2005
(UL/cUL 3rd Edition),
EN 60601-1:2006
(TUV/T-mark 3rd Edition)
AHP6301A, 6301C: Class I
AHP6301B: Double Insulated
Class II
All specifications are typical at nominal input, full load, and 25°C unless otherwise noted
WWW.ASTRODYNETDI.COM
Size
Construction
Weight
Flammability Rating
5.17” x 2.18” x 1.46”
(131 x 55.4 x 37mm)
Desktop Style
13.4oz (380g)
UL94V-1
ENVIRONMENTAL SPECIFICATIONS
Oper. Temperature
Cooling
Storage Temperature
Operating Humidity
EMC
MTBF (MIL-HDBK-217F)
Surge Voltage
ESD - Air
ESD - Contact
Operating Altitude (Elevation)
Vibration
Drop Test
-10 to +40°C (FL) *
50% Load @ 70°C
(See Derate Curve Last Page)
Free Air Convection
-40 to +85°C, 10~95% RH *
0% to 95%, non-cond *
Compliance to EN55011 (CISPR11),
EN61000-3-2, -3 Class B
100,000 Hrs
2kV
IEC61000-4-2, 8kV
IEC61000-4-2, 6kV
3000M
5G, 10 ~ 500Hz, 10min./1cycle,
60min. each along X, Y, Z axes
120cm, 1 ea. on six sides
NOTES
1. Ripple & noise is measured by using 20MHz bandwidth limited
oscilloscope and terminated each output with a 0.47uF capacitor at
rated load and nominal line.
2. Hold up time is measured from the end of the last charging pulse to
the time which the main output drops down to low limit of main output at
rated load and nominal line.
* These are stress ratings. Exposure of the devices to any of these conditions
may adversely affect long term reliability. Proper operation under conditions
other than the standard operating conditions is neither warranteed nor implied.
ASTRODYNE USA: 1-800-823-8082
ASTRODYNE PACIFIC: 886-2-26983458
AHP6301
63W Me d i c a l P o wer A dapt ers f or H ealt h Ca re Ap p lica tio n s
series
MECHANICAL DIMENSIONS - AHP6301A
Dimensions are In mm [inches].
DC OUTPUT CONNECTIONS
Coaxial Plug:
-Vout
+Vout
5.5mm outer diameter
2.1mm inner diameter, female
11mm length, center “+”
AHP6301x-05~07 Output Cord Length 48”, 16AWG
AHP6301x-08~10 Output Cord Length 72”, 18AWG
Modifying the output cable will effect Regulation and
Efficiency
WWW.ASTRODYNETDI.COM
ASTRODYNE USA: 1-800-823-8082
ASTRODYNE PACIFIC: 886-2-26983458
AHP6301
63W Me d i c a l P o wer A dapt ers f or H ealt h Ca re Ap p lica tio n s
series
MECHANICAL DIMENSIONS - AHP6301B
Dimensions are In mm [inches].
DC OUTPUT CONNECTIONS
Coaxial Plug:
-Vout
+Vout
5.5mm outer diameter
2.1mm inner diameter, female
11mm length, center “+”
AHP6301x-05~07 Output Cord Length 48”, 16AWG
AHP6301x-08~10 Output Cord Length 72”, 18AWG
Modifying the output cable will effect Regulation and
Efficiency
WWW.ASTRODYNETDI.COM
ASTRODYNE USA: 1-800-823-8082
ASTRODYNE PACIFIC: 886-2-26983458
AHP6301
63W Me d i c a l P o wer A dapt ers f or H ealt h Ca re Ap p lica tio n s
series
MECHANICAL DIMENSIONS - AHP6301C
Dimensions are In mm [inches].
DC OUTPUT CONNECTIONS
Coaxial Plug:
-Vout
+Vout
5.5mm outer diameter
2.1mm inner diameter, female
11mm length, center “+”
AHP6301x-05~07 Output Cord Length 48”, 16AWG
AHP6301x-08~10 Output Cord Length 72”, 18AWG
Modifying the output cable will effect Regulation and
Efficiency
WWW.ASTRODYNETDI.COM
ASTRODYNE USA: 1-800-823-8082
ASTRODYNE PACIFIC: 886-2-26983458
AHP6301
series
DERATE CURVES
63W External Medical Grade Power Supply
110
100
LOAD (%)
90
80
70
60
50
80
100
-10
0
120
140
160
180
200
220
INPUT VOLTAGE (VAC)
(FIG.1) INPUT VOLTAGE DERATING CURVE
240
260
280
110
100
90
80
70
60
50
40
30
20
10
0
10
20
30
40
50
TEMPERATURE (°C)
(FIG.2) TEMPERATURE DERATING CURVE
60
70
80
OUTPUT CABLE RECOMMEND :
1. Selected output connectors and wire, please refer to Appendix.
2. HPU63B-105~107 are required to use AWG#16 / 4FT output cable.
3. HPU63B-108~111 are required to use AWG#18/ 6FT output cable.
4. The regulation and efficiency will be changed by modified output cable.
PACKING :
1. Net weight: 330~380g approx.
2. Optional output connectors available contact sales for details.
Ripple & Noise
Total Regulation
Typ. Efficiency
No Load
Consumption
Hold-Up Time
(W)
(mVp-p)
(%)
(%)
(W)
(ms)
WWW.ASTRODYNETDI.COM
±5
63
100
85
63
100
±5
84
0.5
12
Protection Mode
Maximum
Output Power
ent
63W Me d i c a l P o wer A dapt ers f or H ealt h Ca re Ap p lica tio n s
OLP
0.5
12
OLP
63
100
±5
85
0.5
12
OLP
63
100
±3
85
0.5
12
OLP
0315
ASTRODYNE USA: 1-800-823-8082
ASTRODYNE PACIFIC: 886-2-26983458