Keysight M9290A Pxie x-series signal analyzer Datasheet

Keysight Technologies
M9290A CXA-m
PXIe X-Series Signal Analyzer
10 Hz to 3.0, 7.5, 13.6 or 26.5 GHz
Data Sheet
Table of Contents
Technical Specifications....................................................................3
Definitions and conditions.........................................................3
Recommended best practices in use.......................................3
Frequency and Time Specifications..........................................4
Amplitude Accuracy and Range Specifications.......................6
Dynamic Range Specifications.................................................9
PowerSuite Measurement Specifications............................... 12
Tracking Generator Specifications ......................................... 13
General Specifications............................................................ 14
Inputs and Outputs................................................................. 14
I/Q Analyzer............................................................................. 15
System Requirements............................................................. 15
Software.......................................................................................... 16
Related Literature............................................................................ 17
Deploy a smaller
microwave footprint
In test system development, one
of your crucial requirements is
doing more in less space—but this
often means tradeoffs between
footprint and precision in signal
analysis. That is, until now: we’ve
expanded the Keysight X-Series
with the CXA-m, a PXIe signal
analyzer that offers fully specified
performance up to 26.5 GHz. It
lets you handle RF and microwave
signals in four slots, and you
can leverage your existing code.
Choose the CXA-m and deploy a
smaller microwave footprint.
03 | Keysight | M9290A CXA-m PXIe X-Series Signal Analyzer - Data Sheet
Technical Specifications
Definitions and conditions
Temperatures referred to in this document are defined as follows:
–– Full temperature range = Individual module temperature of 5 to 68 °C, as reported by
the module, and environment temperature of 0 to 55 °C.
–– Controlled temperature range = Individual module temperature of 25 to 40 °C, as
reported by the module, and environment temperature of 20 to 30 °C.
Specifications describe the warranted performance of calibrated instruments. Data
represented in this document are specifications under the following conditions unless
otherwise noted.
–– It is within its calibration cycle
–– Under auto couple control, except when Auto Sweep Time Rules = Accy
–– The analyzer has been stored at an ambient temperature within the allowed
operating range for at least two hours before being turned on; if it had previously
been stored at a temperature range inside the allowed storage range, but outside
the allowed operating range
–– The analyzer has been turned on at least 30 minutes with Auto Align set to normal,
or, if Auto Align is set to off or partial, alignments must have been run recently
enough to prevent an Alert message; if the Alert condition is changed from Time and
Temperature to one of the disabled duration choices, the analyzer may fail to meet
specifications without informing the user
95th percentile values indicate the breadth of the population (approx. 2σ) of
performance tolerances expected to be met in 95 percent of the cases with a 95 percent
confidence, for any ambient temperature in the range of 20 to 30 °C. In addition to the
statistical observations of a sample of instruments, these values include the effects of
the uncertainties of external calibration references. These values are not warranted.
These values are updated occasionally if a significant change in the statistically observed
behavior of production instruments is observed.
Typical describes additional product performance information that is not covered by
the product warranty. It is performance beyond specifications that 80 percent of the
units exhibit with a 95 percent confidence level over the temperature range 20 to 30 °C.
Typical performance does not include measurement uncertainty.
Nominal values indicate expected performance, or describe product performance that is
useful in the application of the product, but are not covered by the product warranty.
Recommended best practices in use
–– Use slot blockers and EMC filler panels in empty module slots to ensure proper
operating temperatures. Keysight chassis and slot blockers optimize module
temperature performance and reliability of test.
–– Set chassis fan to high at environmental temperatures above 45°C.
Get more information
This CXA-m PXIe X-Series
signal analyzer data sheet is
a summary of the complete
specifications and conditions
available in the CXA-m PXIe
Signal Analyzer Specification
Guide. The CXA-m PXIe Signal
Analyzer Specification Guide can
be obtained on the web at:
www.keysight.com/find/cxa-m_
manuals
For ordering information, refer to
the CXA-m PXIe Signal Analyzer
Configuration Guide
(5992-0193EN).
04 | Keysight | M9290A CXA-m PXIe X-Series Signal Analyzer - Data Sheet
Frequency and Time Specifications
Frequency range
Option F03
Option F07
Option F13
Option F26
Band
0
1
2
3
4
4
5
6
7
8
9
10
Frequency reference
Accuracy
Aging rate
Temperature stability
20 to 30 °C
Full temperature range
Achievable initial calibration accuracy
DC coupled
10 Hz to 3.0 GHz
10 Hz to 7.5 GHz
10 Hz to 13.6 GHz
10 Hz to 26.5 GHz
LO multiple (N)
1
2
2
2
2
4
4
4
4
4
4
4
AC coupled
10 MHz to 3.0 GHz
10 MHz to 7.5 GHz
10 MHz to 13.6 GHz
10 MHz to 26.5 GHz
10 Hz to 3.08 GHz
2.95 to 7.575 GHz
7.45 to 9.55 GHz
9.45 to 12.60 GHz
12.50 to 13.05 GHz
12.95 to 13.80 GHz
13.40 to 15.55 GHz
15.45 to 19.35 GHz
19.25 to 21.05 GHz
20.95 to 22.85 GHz
22.75 to 24.25 GHz
24.15 to 26.55 GHz
± [(time since last adjustment x aging rate) + temperature stability + calibration accuracy]
Option PFR
Standard
± 1 x 10-7 / year
± 1 x 10-6 / year
-7
± 1.5 x 10 / 2 years
Option PFR
Standard
± 1.5 x 10-8
± 2 x 10-6
-8
± 5 x 10
± 2 x 10-6
Option PFR
Standard
± 4 x 10-8
± 1.4 x 10-6
= ± (1 x 1 x 10-7 + 5 x 10-8 + 4 x 10-8)
Example frequency reference accuracy
(with Option PFR)
1 year after last adjustment
= ± 1.9 x 10-7
Residual FM
Option PFR
≤ 0.25 Hz p-p in 20 ms nominal
Standard
≤ 10 Hz p-p in 20 ms nominal
Frequency readout accuracy (start, stop, center, marker)
± (marker frequency x frequency reference accuracy + 0.25 % x span + 5 % x RBW + 2 Hz + 0.5 x horizontal resolution 1)
Marker frequency counter
Accuracy
± (marker frequency x frequency reference accuracy + 0.100 Hz)
Delta counter accuracy
± (delta frequency x frequency reference accuracy + 0.141 Hz)
Counter resolution
0.001 Hz
Frequency span (FFT and swept mode)
Range
0 Hz (zero span), 10 Hz to maximum frequency of instrument
Resolution
2 Hz
Accuracy
Swept
± (0.25 % x span + horizontal resolution)
FFT
± (0.10 % x span + horizontal resolution)
1.
Horizontal resolution is span/(sweep points – 1)
05 | Keysight | M9290A CXA-m PXIe X-Series Signal Analyzer - Data Sheet
Frequency and Time Specifications (Continued)
Sweep time and triggering
Range
Accuracy
Trigger
Trigger delay
Time gating
Gate methods
Gate length range (except method = FFT)
Gate delay range
Gate delay jitter
Sweep (trace) point range
All spans
Resolution bandwidth (RBW)
Range (–3.01 dB bandwidth)
Bandwidth accuracy (power)
Bandwidth accuracy (–3.01 dB)
RBW range
Selectivity (–60 dB/–3 dB)
EMI bandwidth (CISPR compliant)
EMI bandwidth (MIL STD 461E compliant)
Analysis bandwidth 1
Maximum bandwidth
Video bandwidth (VBW)
Range
Accuracy
Measurement speed 2
Local measurement and display update rate
Remote measurement and LAN transfer rate
Marker peak search
Center frequency tune and transfer
Measurement/mode switching
1.
Span = 0 Hz
Span ≥ 10 Hz
Span ≥ 10 Hz, swept
Span ≥ 10 Hz, FFT
Span = 0 Hz
Free run, video, external, RF burst, periodic timer
Span = 0 Hz or FFT
Span ≥ 10 Hz, swept
Resolution
1 μs to 6000 s
1 ms to 4000 s
± 0.01 % nominal
± 40 % nominal
± 1 % nominal
–150 to +500 ms
1 μs to 500 ms
0.1 μs
Gated LO; gated video; gated FFT
100.0 ns to 5.0 s
0 to 100.0 s
33.3 ns p-p nominal
1 to 40001
1 Hz to 3 MHz (10 % steps), 4, 5, 6, 8 MHz
1 Hz to 750 kHz
820 kHz to 1.2 MHz
1.3 to 2.0 MHz
2.2 to 3 MHz
4 to 8 MHz
1 Hz to 1.3 MHz
± 1.0 % (± 0.044 dB) nominal
± 2.0 % (± 0.088 dB) nominal
± 0.13 dB nominal
± 0.22 dB nominal
± 0.32 dB nominal
± 2 % nominal
4.1:1 nominal
200 Hz, 9 kHz, 120 kHz, 1 MHz
10 Hz, 100 Hz, 1 kHz, 10 kHz, 100 kHz, 1 MHz
(Option EMC)
(Option EMC)
Option B25
Standard
25 MHz
10 MHz
1 Hz to 3 MHz (10 % steps), 4, 5, 6, 8 MHz, and wide open (labeled 50 MHz)
± 6 % nominal
11 ms (90/s) nominal
6 ms (167/s) nominal
5 ms nominal
22 ms nominal
75 ms nominal
Analysis bandwidth is the instantaneous bandwidth available around a center frequency over which the input signal can be digitized for further analysis or
processing in the time, frequency, or modulation domain.
2. Sweep points = 101.
06 | Keysight | M9290A CXA-m PXIe X-Series Signal Analyzer - Data Sheet
Amplitude Accuracy and Range Specifications
Amplitude range
Measurement range
Preamp off
Preamp on
Input attenuator range
Standard
Option FSA
Displayed average noise level (DANL) to +30 dBm
Displayed average noise level (DANL) to +23 dBm
0 to 70 dB in 10 dB steps
0 to 70 dB in 2 dB steps, 7.5 GHz
0 to 70 dB in 10 dB steps, 7.5 to 26.5 GHz
Maximum safe input level
Average total power
+30 dBm (1 W)
+30 dBm (1 W)
Input attenuation ≥ 10 dB, preamp off
Input attenuation ≥ 20 dB, preamp on
+50 dBm (100 W)
< 10 μs pulse width, < 1 % duty cycle, and input attenuation ≥ 30 dB
Peak pulse power
DC volts
AC coupled
DC coupled
Display range
Log scale
± 50 Vdc
± 0.2 Vdc
0.1 to 1 dB/division in 0.1 dB steps
1 to 20 dB/division in 1 dB steps (10 display divisions)
Linear scale
10 divisions
Scale units
dBm, dBmV, dBμV, dBmA, dBμA, V, W, A
Frequency response
Specification
(10 dB input attenuation, 20 to 30 °C, σ = nominal standard deviation)
9 kHz to 10 MHz
± 0.50 dB
10 MHz to 3 GHz
± 0.65 dB
3 to 13.6 GHz
± 1.30 dB
13.6 to 19.3 GHz
± 1.50 dB
19.3 to 24.2 GHz
± 2.20 dB
24.2 to 26.5 GHz
± 2.50 dB
Preamp on (Option P03, P07, P13, P26)
100 kHz to 10 MHz
10 MHz to 3 GHz
3 to 7.5 GHz
7.5 to 13.6 GHz
13.6 to 21 GHz
21 to 24.2 GHz
24.2 to 26.5 GHz
Input attenuation switching uncertainty
Specifications
Attenuation > 2 dB, preamp off
50 MHz (reference frequency)
± 0.3 dB
Relative to 10 dB
100 kHz to 3.0 GHz
(reference setting)
3.0 to 7.5 GHz
7.5 to 26.5 GHz
95th percentile (≈ 2σ)
± 0.4 dB
± 0.5 dB
± 0.8 dB
± 1.0 dB
± 1.3 dB
± 1.3 dB
± 0.5 dB
± 1.0 dB
± 1.2 dB
± 1.0 dB
± 1.2 dB
± 1.8 dB
± 2.4 dB
Additional information
± 0.15 dB typical
± 0.30 dB nominal
± 0.50 dB nominal
± 0.70 dB nominal
07 | Keysight | M9290A CXA-m PXIe X-Series Signal Analyzer - Data Sheet
Amplitude Accuracy and Range Specifications (Continued)
Total absolute amplitude accuracy
(10 dB attenuation, 20 to 30 °C, 1 Hz ≤ RBW ≤ 1 MHz, input signal –10 to –50 dBm, all settings auto-coupled except Auto Swp Time = Accy, any
reference level, any scale, σ = nominal standard deviation)
At 50 MHz
± 0.40 dB
At all frequencies
± (0.40 dB + frequency response)
100 kHz to 3 GHz
± 0.60 dB (95th Percentile ≈ 2σ)
Preamp on (Option P03/P07/P13/P26)
± (0.36 dB + frequency response) (95th percentile)
Precision amplitude accuracy (Option PAA)
Preamp off (10 dB attenuation)
Specification
Additional information
9 kHz to 10 MHz
± 0.70 dB
± 0.31 dB typical
10 MHz to 3 GHz
± 0.85 dB
± 0.40 dB typical
3 to 7.5 GHz
± 0.95 dB
± 0.44 dB typical
7.5 to 9.55 GHz
± 0.90 dB
± 0.38 dB typical
9.55 to 12.55 GHz
± 0.95 dB
± 0.44 dB typical
12.55 to 13.6 GHz
± 0.90 dB
± 0.32 dB typical
13.6 to 15.55 GHz
± 0.95 dB
± 0.35 dB typical
15.55 to 22.8 GHz
± 1.10 dB
± 0.50 dB typical
22.8 to 24.2 GHz
± 1.20 dB
± 0.50 dB typical
24.2 to 26.5 GHz
± 1.30 dB
± 0.57 dB typical
Preamp on (0 dB attenuation)
100 kHz to 10 MHz
± 0.30 dB typical
10 MHz to 3 GHz
± 0.65 dB typical
3 to 7.5 GHz
± 0.73 dB typical
7.5 to 9.55 GHz
± 0.49 dB typical
9.55 to 12.55 GHz
± 0.71 dB typical
12.55 to 19.3 GHz
± 0.65 dB typical
19.3 to 21 GHz
± 0.80 dB typical
21 to 22.8 GHz
± 1.20 dB typical
22.8 to 24.2 GHz
± 1.39 dB typical
24.2 to 26.5 GHz
± 1.66 dB typical
Input voltage standing wave ratio (VSWR) (≥ 10 dB attenuation)
10 MHz to 3 GHz
< 1.2 nominal
3 to 7.5 GHz
< 1.4 nominal
7.5 to 13.6 GHz
< 1.6 nominal
13.6 to 24.2 GHz
< 1.8 nominal
24.2 to 26.5 GHz
< 2.2 nominal
Resolution bandwidth switching uncertainty (referenced to 30 kHz RBW)
1 Hz to 3 MHz RBW
± 0.15 dB
4, 5, 6, 8 MHz RBW
Reference level
Range
Log scale
Linear scale
Accuracy
Display scale switching uncertainty
Switching between linear and log
Log scale/div switching
± 1.0 dB
–170 to +23 dBm in 0.01 dB steps
Same as log (707 pV to 3.16 V)
0 dB
0 dB
0 dB
08 | Keysight | M9290A CXA-m PXIe X-Series Signal Analyzer - Data Sheet
Amplitude Accuracy and Range Specifications (Continued)
Display scale fidelity
–80 dBm ≤ input mixer level < –10 dBm
± 0.15 dB total
Trace detectors
Normal, peak, sample, negative peak, log power average, RMS average, and voltage average
Preamplifier (Option P03/P07/P13/P26)
Frequency range
Option P03
Option P07
Option P13
Option P26
Gain
100 kHz to 26.5 GHz
Noise figure
10 MHz to 3 GHz
3 to 26.5 GHz
100 kHz to 3.0 GHz
100 kHz to 7.5 GHz
100 kHz to 13.6 GHz
100 kHz to 26.5 GHz
+20 dB nominal
10 dB nominal
DANL + 176.24 dB nominal
09 | Keysight | M9290A CXA-m PXIe X-Series Signal Analyzer - Data Sheet
Dynamic Range Specifications
1 dB gain compression (two-tone)
Preamp off
Preamp on
10 MHz to 7.5 GHz
7.5 to 26.5 GHz
10 MHz to 7.5 GHz
7.5 to 26.5 GHz
Total power at input mixer
+6 dBm nominal
+4 dBm nominal
–15 dBm nominal
–19 dBm nominal
Displayed average noise level (DANL)
(Input terminated, sample or average detector, averaging type = Log, 0 dB input attenuation, IF Gain = High, 20 to 30 °C)
Parentheses indicate typical performance
Preamplifier OFF
Preamplifier ON
10 Hz
–95 dBm nominal
100 Hz
–110 dBm nominal
1 kHz
–115 dBm nominal
9 kHz to 1 MHz
(–125) dBm
1 to 10 MHz
–144, (–148) dBm
–154, (–158) dBm
10 MHz to 1.5 GHz
–148, (–150) dBm
–160, (–163) dBm
1.5 to 4.5 GHz
–146, (–149) dBm
–160, (–163) dBm
4.5 to 7 GHz
–141, (–145) dBm
–157, (–161) dBm
7 to 9.5 GHz
–144, (–147) dBm
–158, (–160) dBm
9.5 to 13 GHz
–136, (–140) dBm
–156, (–160) dBm
13 to 14.5 GHz
–142, (–145) dBm
–158, (–161) dBm
14.5 to 19.3 GHz
–132, (–138) dBm
–153, (–157) dBm
19.3 to 23 GHz
–134, (–139) dBm
–152, (–157) dBm
23 to 24 GHz
–132, (–137) dBm
–150, (–155) dBm
24 to 26.5 GHz
–128, (–133) dBm
–144, (–149) dBm
Spurious responses
Residual response
200 kHz to 26.5 GHz (swept)
–90 dBm
(Input terminated and 0 dB attenuation) Zero span or FFT or other
–100 dBm nominal
frequencies
Tuned frequency (f)
Excitation Freq
Mixer level
Image responses (Second mixer)
10 MHz to 7.5 GHz
f + 1645 MHz
−10 dBm
7.5 GHz to 19.3 GHz
f - 1645 MHz
−10 dBm
19.3 GHz to 21 GHz
f + 1645 MHz
−10 dBm
21 GHz to 22.8 GHz
f - 1645 MHz
−10 dBm
22.8 GHz to 24.2 GHz
f + 1645 MHz
−10 dBm
24.2 GHz to 26.5 GHz
f - 1645 MHz
−10 dBm
LO-related spurious
10 MHz to 26.5 GHz
−10 dBm
Other spurious responses
Mixer level
Response
IF feedthrough
−10 dBm
−75 dBc (−80 dBc typical)
First RF order (f ≥ 10 MHz from carrier) −10 dBm
−70 dBc (−80 dBc nominal)
High RF order (f ≥ 10 MHz from carrier) −30 dBm
−70 dBc (−80 dBc nominal)
Second harmonic distortion (SHI)
Source frequency
SHI (nominal)
10 MHz to 3.75 GHz
+50 dBm
3.75 to 13.25 GHz
+62 dBm
Response
−70 dBc (−80 dBc typical)
−70 dBc (−80 dBc typical)
−70 dBc (−80 dBc typical)
−70 dBc (−80 dBc typical)
−70 dBc (−80 dBc typical)
−70 dBc (−80 dBc typical)
−64 dBc typical
10 | Keysight | M9290A CXA-m PXIe X-Series Signal Analyzer - Data Sheet
Dynamic Range Specifications (Continued)
Third-order intermodulation distortion (TOI)
Parentheses indicate typical performance
Preamp off
(Two –20 dBm tones at input mixer
spaced by 100 kHz, 0 dB attenuation,
20 to 30 °C
Option P03/P07/P13/P26
Preamp on
(Two –45 dBm tones at the preamp input,
spaced by 100 kHz, 0 dB attenuation,
20 to 30 °C)
10 MHz to 2 GHz
2 to 3 GHz
3 to 7.5 GHz
7.5 to 13.6 GHz
13.6 to 26.5 GHz
10 MHz to 26.5 GHz
+12 dBm, (+16) dBm
+12 dBm, (+17) dBm
+12 dBm, (+16) dBm
+11 dBm, (+15) dBm
+10 dBm, (+14) dBm
–8 dBm nominal
Figure 1. Nominal dynamic range for Band 0,
for second and third order distortion,
10 MHz to 3 GHz
Figure 2. Nominal dynamic range, for
second and third order distortion, 3 GHz to
26.5 GHz
11 | Keysight | M9290A CXA-m PXIe X-Series Signal Analyzer - Data Sheet
Dynamic Range Specifications (Continued)
Phase noise
Offset
Noise sidebands (20 to 30 °C, CF = 1 GHz)
100 Hz
1 kHz
10 kHz
100 kHz
1 MHz
10 MHz
Specification
Typical
–102 dBc/Hz
–106 dBc/Hz
–108 dBc/Hz
–130 dBc/Hz
–90 dBc/Hz nominal
–105 dBc/Hz
–110 dBc/Hz
–110 dBc/Hz
–132 dBc/Hz
–145 dBc/Hz nominal
Figure 3. Nominal phase noise at different
center frequencies
12 | Keysight | M9290A CXA-m PXIe X-Series Signal Analyzer - Data Sheet
PowerSuite Measurement Specifications
Channel power
Amplitude accuracy, W-CDMA or IS95
(20 to 30 °C, attenuation = 10 dB)
Occupied bandwidth
Frequency accuracy
Adjacent channel power
Accuracy, W-CDMA (ACLR)
(at specific mixer levels and ACLR ranges)
MS
BTS
Dynamic range (typical)
± 1.23 dB (± 0.62 dB 95th percentile)
± [span/1000] nominal
Without noise correction
With noise correction
1 to 6
Up to 12
Adjacent
Alternate
± 0.21 dB
± 0.89 dB
± 0.25 dB
± 0.67 dB
–68 dB
–73 dB
–72 dB
–78 dB
Offset channel pairs measured
Multiple number of carriers measured
Power statistics CCDF
Histogram resolution
0.01 dB
Harmonic distortion
Maximum harmonic number
10th
Results
Fundamental power (dBm), relative harmonics power (dBc), total harmonic distortion in %
Intermod (TOI)
Measure the third-order products and intercepts from two tones
Burst power
Methods
Power above threshold, power within burst width
Results
Single burst output power, average output power, maximum power, minimum power within burst, burst
width
Spurious emission
W-CDMA (1 to 3.0 GHz) table-driven spurious signals; search across regions
Dynamic range
75.0 dB
(79.1 dB typical)
Absolute sensitivity
−82.5 dBm
(−86.5 dBm typical)
Spectrum emission mask (SEM)
cdma2000 ® (750 kHz offset)
Relative dynamic range (30 kHz RBW)
73.0 dB
(78.1 dB typical)
Absolute sensitivity
−99.7 dBm
(−102.7 dBm typical)
Relative accuracy
± 0.11 dB
3GPP W-CDMA (2.515 MHz offset)
Relative dynamic range (30 kHz RBW)
77.5 dB
(82.4 dB typical)
Absolute sensitivity
−97.7 dBm
(−101.7 dBm typical)
Relative accuracy
± 0.15 dB
13 | Keysight | M9290A CXA-m PXIe X-Series Signal Analyzer - Data Sheet
Tracking Generator Specifications
Output frequency
Frequency range
Option T03
Option T07
Option T13
Option T26
Resolution
Output power level
Range
2 MHz to 10 GHz
10 to 20 GHz
20 to 26.5 GHz
Resolution
Absolute accuracy (at 50 MHz, –15 dBm)
Output flatness (Referenced to 50 MHz, –15 dBm)
2 MHz to 7.5 GHz
7.5 to 13.6 GHz
13.6 to 23 GHz
23 to 26 GHz
26 to 26.5 GHz
Level accuracy
2 MHz to 7.5 GHz
7.5 to 13.6 GHz
13.6 to 23GHz
23 to 26 GHz
26 to 26.5GHz
Output power sweep
Range
2 MHz to 10 GHz
10 to 20 GHz
20 to 26.5 GHz
Resolution
Maximum safe reverse level
Average total power
AC coupled
Phase noise
Noise sidebands (Center Frequency = 1 GHz)
2 MHz to 3 GHz
2 MHz to 7.5 GHz
2 MHz to 13.6 GHz
2 MHz to 26.5 GHz
10 Hz
—35 to 0 dBm
—35 to —5 dBm
—40 to —12 dBm
0.1 dB
± 1.0 dB typical
95th percentile (≈ 2σ)
± 1.0 dB
± 1.2 dB
± 1.8 dB
± 2.5 dB
± 0.3 dB nominal
± 0.3 dB
± 0.3 dB
± 0.6 dB
± 1.2 dB
± 2.3 dB
Nominal
± 0.8 dB
± 0.9 dB
± 1.5 dB
± 1.8 dB
± 2.9 dB
—35 to 0 dBm
—35 to —5 dBm
—40 to —12 dBm
0.1 dB
+ 30 dBm (1 W)
± 50 Vdc
Offset
10 kHz
100 kHz
1 MHz
Typical
—100 dBc/Hz
—108 dBc/Hz
—122 dBc/Hz
Dynamic range
Maximum output power – displayed average 110 dBc nominal
Output VSWR
2 MHz to 7 GHz
7 to 23 GHz
23 to 26.5 GHz
< 1.7:1 nominal
< 2.5:1 nominal
< 3.5:1 nominal
14 | Keysight | M9290A CXA-m PXIe X-Series Signal Analyzer - Data Sheet
General Specifications
Temperature range
Operating
0 to 55 °C
Storage
–40 to 70 °C
EMC
Complies with European EMC Directive 2004/108/EC
– IEC/EN 61326-1
– CISPR Pub 11 Group 1, class A
– AS/NZS CISPR 11
– ICES/NMB-001
This ISM device complies with Canadian ICES-001
Cet appareil ISM est conforme à la norme NMB-001 du Canada
Environmental stress
Samples of this product have been type tested in accordance with the Keysight Environmental Test Manual and verified to be robust against the
environmental stresses of storage, transportation, and end-use; those stresses include, but are not limited to, temperature, humidity, shock, vibration,
altitude, and power line conditions; test methods are aligned with IEC 60068-2 and levels are similar to MILPRF-28800F Class 3.
Power requirement
Power drawn from chassis
≤ 65 W
Weight
Net
1.9 kg (4.2 lbs)
Shipping
4.2 kg (9.3 lbs)
Dimensions
Height
132 mm (5.2 in)
Width
82 mm (3.2 in)
Length
202 mm (8.0 in)
Warranty
The CXA-m PXIe signal analyzer is supplied with a three-year warranty
Calibration cycle
The recommended calibration cycle is one year; calibration services are available through Keysight service centers
Inputs and Outputs
RF input
Connector
RF output
Connector
10 MHz in
Connector
10 MHz out
Connector
Trigger in
Connector
Trigger out
Connector
Analog out
Connector
Noise source
3.5mm female, 50 Ω nominal
3.5mm female, 50 Ω nominal
SMB male, 50 Ω nominal
SMB male, 50 Ω nominal
SMB male, 10 kΩ nominal
SMB male, 50 Ω nominal
SMB male, 50 Ω nominal
For use with Keysight 346 and SNS series noise sources
15 | Keysight | M9290A CXA-m PXIe X-Series Signal Analyzer - Data Sheet
I/Q Analyzer
Frequency
Frequency span
Standard instrument
10 Hz to 10 MHz
Option B25
10 Hz to 25 MHz
Resolution bandwidth (spectrum measurement)
Range
Overall
100 MHz to 3 MHz
Span = 1 MHz
50 Hz to 1 MHz
Span = 10 kHz
1 Hz to 10 kHz
Span = 100 Hz
100 MHz to 100 Hz
Window shapes
Flat top, Uniform, Hanning, Gaussian, Blackman, Blackman-Harris, Kaiser Bessel (K-B 70 dB, K-B 90 dB and K-B 110 dB)
Analysis bandwidth
Standard instrument
10 Hz to 10 MHz
Option B25
10 Hz to 25 MHz
IF frequency response (standard 10 MHz IF path)
IF frequency response (demodulation and FFT response relative to the center frequency, 20 to 30 °C)
Center frequency (GHz)
Span (MHz)
Max. error
RMS (nominal)
≤ 3.0
≤ 10
± 0.40 dB
0.03 dB
3.0 < f ≤ 26.5
≤ 10
0.10 dB
IF phase linearity (deviation from mean phase linearity, nominal)
Center frequency (GHz)
Span (MHz)
Peak-to-peak
RMS
≤ 3.0
≤ 10
0.5 °
0.2 °
3.0 < f ≤ 7.5
≤ 10
0.5 °
0.4 °
Data acquisition (standard 10 MHz IF path)
Time record length
4,000,000 IQ sample pairs
Sample rate
90 MSa/s
ADC resolution
16 Bits
Option B25 25 MHz analysis bandwidth
IF frequency response (demodulation and FFT response relative to the center frequency, 20 to 30 °C)
Center frequency (GHz)
Span (MHz)
Max. error
RMS (nominal)
≤ 3.0
10 to ≤ 25
± 0.45 dB
0.03 dB
3.0 < f ≤ 26.5
10 to ≤ 25
0.20 dB
IF phase linearity (deviation from mean phase linearity, nominal)
Center frequency (GHz)
0.02 ≤ f < 3.0
3.0 < f ≤ 7.5
Data acquisition (B25 IF path)
Time record length
IQ analyzer
Sample rate
ADC resolution
Span (MHz)
10 to ≤ 25
10 to ≤ 25
Peak-to-peak
1°
1°
RMS
0.3 °
0.5 °
4,000,000 IQ sample pairs
90 MSa/s
16 Bits
System Requirements
Operating system
Processor speed
Available memory
Available disk space
Video
Browser
Windows 7 (32 & 64 bit)
1.86 GHz minimum
2.4 GHz recommended
4 GB minimum
8 GB recommended
4 GB
Support for DirectX 9 graphics with 128 MB graphics recommended (SuperVGA supported)
Microsoft Internet Explorer 7.0 or greater
16 | Keysight | M9290A CXA-m PXIe X-Series Signal Analyzer - Data Sheet
Software
Instrument connection software
Keysight IO library The IO library suite offers a single entry point for connection to
the most common instruments including AXIe, PXI, GPIB, USB,
Ethernet/LAN, RS-232, and VXI test instruments from Keysight and
other vendors. It automatically discovers interfaces, chassis, and
instruments. The graphical user interface allows you to search for,
verify, and update IVI instrument and soft front panel drivers for
modular and traditional instruments. The IO suite safely installs in
side-by-side mode with NI I/O software.
Module setup and usage
Keysight soft
The CXA-m includes a soft front panel (SFP), a software-based
|front panel
graphical user interface (GUI) which enables the instrument’s
capabilities from your PC.
Module management
Keysight connection expert
Connection expert is the graphical user interface included in the
IO libraries suite that allows you to search for, verify and update IVI
instrument and soft front panel drivers for modular and traditional
instruments
Programming
Driver
lVI-COM
IVI-C
LabVIEW
MATLAB
Programming assistance
Development environments
Visual Studio (VB .NET, C#, C/C++), VEE, LabVIEW, LabWindows/
CVI, MATLAB
Free software download at
www.keysight.com/find/
iosuite
Included on CD-ROM shipped
with module or online
Free software download at
www.keysight.com/find/
iosuite
Included on CD-ROM shipped
with module.
Command expert
Assists in finding the right instrument commands and setting correct Free software download at
parameters. A simple interface includes documentation, examples,
www.keysight.com/find/
syntax checking, command execution, and debug tools to build
commandexpert
sequences for integration in Excel, MATLAB, Visual Studio, and VEE.
X-Series
measurement
applications
Provides measurements for analog demodulation, noise figure,
phase noise, vector signal analysis, and others.
Licensed software. For more
information, visit
www.keysight.com/find/xseries_apps
89600 VSA
89600 VSA software sees through the complexity of emerging and
existing industry standards, serving as your window into complex
signal interactions.
Licensed software. For more
information, visit
www.keysight.com/find/vsa
SystemVue
SystemVue is a system-level EDA platform for designing
communications and defense systems. Used with the M9290A,
SystemVue enables you to create model-based design validation
tests to ensure consistency from design to manufacturing.
Licensed software. For more
information, visit
www.keysight.com/find/
systemvue
Signal analysis software
17 | Keysight | M9290A CXA-m PXIe X-Series Signal Analyzer - Data Sheet
Related Literature
Literature
M9290A CXA-m PXIe Signal Analyzer - Product Fact Sheet
M9290A CXA-m PXIe Signal Analyzer - Configuration Guide
M9018A PXIe 18 slot Chassis - Data Sheet
M9037A PXIe High Performance Embedded Controller - Data Sheet
M9036A PXIe Embedded Controller - Data Sheet
Web
For more information or literature resources please visit the web:
Product page: www.keysight.com/find/M9290A
X-Series measurement applications: www.keysight.com/find/X-Series_Apps
X-Series signal analyzers: www.keysight.com/find/X-Series
Pub number
5992-0044EN
5992-0193EN
5990-6583EN
5991-3661EN
5990-8465EN
18 | Keysight | M9290A CXA-m PXIe X-Series Signal Analyzer - Data Sheet
myKeysight
www.keysight.com/find/mykeysight
A personalized view into the information most relevant to you.
Three-Year Warranty
www.keysight.com/find/ThreeYearWarranty
Keysight’s commitment to superior product quality and lower total cost
of ownership. The only test and measurement company with three-year
warranty standard on all instruments, worldwide.
Keysight Assurance Plans
www.keysight.com/find/AssurancePlans
Up to five years of protection and no budgetary surprises to ensure your
instruments are operating to specification so you can rely on accurate
measurements.
www.keysight.com/go/quality
Keysight Technologies, Inc.
DEKRA Certified ISO 9001:2008
Quality Management System
Keysight Infoline
www.keysight.com/find/service
Keysight’s insight to best in class information management. Free access to
your Keysight equipment company reports and e-library.
Keysight Channel Partners
www.keysight.com/find/channelpartners
Get the best of both worlds: Keysight’s measurement expertise and product
breadth, combined with channel partner convenience.
www.keysight.com/find/CXA-m
For more information on Keysight
Technologies’ products, applications or
services, please contact your local Keysight
office. The complete list is available at:
www.keysight.com/find/contactus
Americas
Canada
Brazil
Mexico
United States
(877) 894 4414
55 11 3351 7010
001 800 254 2440
(800) 829 4444
Asia Pacific
Australia
China
Hong Kong
India
Japan
Korea
Malaysia
Singapore
Taiwan
Other AP Countries
1 800 629 485
800 810 0189
800 938 693
1 800 11 2626
0120 (421) 345
080 769 0800
1 800 888 848
1 800 375 8100
0800 047 866
(65) 6375 8100
Europe & Middle East
Austria
Belgium
Finland
France
Germany
Ireland
Israel
Italy
Luxembourg
Netherlands
Russia
Spain
Sweden
Switzerland
United Kingdom
0800 001122
0800 58580
0800 523252
0805 980333
0800 6270999
1800 832700
1 809 343051
800 599100
+32 800 58580
0800 0233200
8800 5009286
800 000154
0200 882255
0800 805353
Opt. 1 (DE)
Opt. 2 (FR)
Opt. 3 (IT)
0800 0260637
For other unlisted countries:
www.keysight.com/find/contactus
(BP-04-23-15)
This information is subject to change without notice.
© Keysight Technologies, 2014 - 2015
Published in USA, June 11, 2015
5992-0032EN
www.keysight.com
Similar pages