Keysight N8834A-AFP Multiscope application Datasheet

Keysight N8834A
MultiScope Application
Achieve Greater Insight and Productivity – Infiniium 9000,
S-Series, 90000A, 90000 X-, 90000 Q-, V-Series and Z-Series
InfiniiVision 2000 X-, 3000T X-, 3000A X-, 4000 X-, 6000 X-,
6000 and 7000B Series
Data Sheet
Introduction
Introducing the new MultiScope application
Need more than four simultaneous oscilloscope channels? Ever placed two oscilloscopes
together to see eight channels simultaneously? The Keysight Technologies, Inc. N8834A
MultiScope application enables Infiniium and InfiniiVision real-time oscilloscope users to
bring multiple oscilloscopes together for measurement on a single timebase. Capture, view,
and analyze up to 40 time-correlated oscilloscope channels. If your need for more than four
simultaneous channels goes away, each oscilloscope can be used independently and then
brought back together when there are future needs for more than four channel measurements.
Are you working on applications like power rails, high-speed multi-lane serial buses, optical
networking, or DDR memory systems? The N8834A MultiScope software allows connection
of 2 to 10 oscilloscopes to achieve up to 40 channels on a single timebase. Each oscilloscope
is daisy chained via cables and power splitters to the first oscilloscope, called the leader.
Automated calibration is available to allow channel correlation across frames down to less
than 1 ps. All oscilloscopes connect to a control PC via LAN or USB. The PC runs the Infiniium
application (user interface) and allows users to change settings on all oscilloscopes. The
Infiniium application on the PC shows all waveforms and measurements and offers analysis in
addition to controlling the oscilloscope settings. The leader can also work as the controller in
the absence of a control PC.
03 | Keysight | N8834A MultiScope Application - Data Sheet
Features
–– Save costs by connecting up to 10 of your existing Infiniium
Series oscilloscopes to achieve 40 channel acquisitions in a
single timebase.
–– Data is presented to the user on a control PC or the leader
scope as if all oscilloscope channels existed on a single
frame.
–– Oscilloscopes can be separated and used independently
when four channels or fewer are needed.
–– Application supports multiple frames using the same family
of oscilloscopes (e.g. all V-Series scopes) or a heterogeneous
combination of different oscilloscope families (e.g. one 9000
Series and one S-Series). Time-correlation precision depends
on which frames are connected together.
Increase your productivity and efficiency
The N8834A MultiScope software works well with the new PChosted mode, which allows you to see live signals acquired from
the oscilloscopes on a PC installed with the N8900A Infiniium
Offline oscilloscope analysis software. Users can test lanes on
the MultiScope from the comfort of their own workbench. Users
can remotely login to the MultiScope application from the PC and
observe live traffic. The Infiniium Offline oscilloscope analysis
software application allows users to save the setup and waveform
files in a composite format (.osc) to be analyzed offline on the PC,
freeing up the MultiScope connection for others.
Analyze data on PC. User can either connect to live data for online
analysis via hosted mode or save the setup file to be viewed offline via
offline mode.
MultiScope connection of two S-Series oscilloscopes.
Heterogeneous combination: One S-Series oscilloscope and one 90000
X-Series oscilloscope.
04 | Keysight | N8834A MultiScope Application - Data Sheet
MultiScope Software Architecture
Control PC
The leader and the followers
The N8834A MultiScope application runs on a Windows 7 PC.
All the scopes are connected to the control PC via LAN or USB.
A control PC must be used if InfiniiVision scopes exist in the
multiscope configuration.
–– PC supports both hosted/online and offline analysis.
–– Scope front panels are disabled during MultiScope operation.
MultiScope is controlled by a PC installed with the N8900A
Infiniium Offline analysis software or the leader scope in the
absence of a PC.
–– Display up to 16 grids per window, up to 8 windows.
–– Full support of all measurements, functions, and analysis on
all channels.
–– Test at the comfort of your workbench.
The leader oscilloscope connects to the control PC via LAN or
USB or can act as the controller in the absence of the PC. The
one leader oscilloscope can connect two to nine followers. The
connection between oscilloscopes is daisy chained with cables
and splitters.
–– Support edge or glitch trigger on leader (other triggering
modes (including InfiniiScan) are not supported).
–– One-time deskew eliminates frame-to-frame skew created by
trigger daisy chain connection.
–– Achieve inter-scope intrinsic jitter of 150 fs with precision
calibration on the 90000 Q-Series and Z-Series Infiniium
oscilloscopes.
–– 10-MHz reference clock aligns sample for deep memory
synchronization.
Daisy chained connection with power splitter and BNC cables.
View of 6 channels and 3 eye diagrams on a single window with 8 grids.
Leader scope is used to control and display multichannel waveforms.
05 | Keysight | N8834A MultiScope Application - Data Sheet
Calibration Methods
There are three calibration methods available with the N8834A
MultiScope software. Each calibration method provides a
different level of channel-to-channel skew value. Manual
calibration is available for users who are interested in viewing
multi-channel signals but not concerned about the skew factor
across channels. External trigger helps users achieve 100 ps
channel-to-channel skew. The most accurate measurements can
be achieved with the drift correction method with skew in the
range of 200 fs.
1. Manual
Manual calibration requires no cabling and users can achieve
skew correction by manually dialing in the right skew values
between the channels. InfiniiVision scopes supports only manual
calibration.
2. Basic
The trigger out process provides the most basic level of interoscilloscope time synchronization. This process connects the
trigger output signal (Trig Out) from the leader oscilloscope
through each follower oscilloscope (Aux Trig) in a daisy chain
configuration. The followers are each armed in descending order
from furthest from the leader (F{N}) to nearest to the leader (F{1}).
The leader’s trigger is armed last. Once the leader triggers, each
follower will trigger in turn, down through the chain.
3. Precision
The horizontal position of channels within different oscilloscope
frames can drift with temperature (inter-oscilloscope drift). The
drift correction synchronization process corrects for this drift by
monitoring the inter-oscilloscope drift on a separate, dedicated
synchronization channel. The idea is that any “time corrections”
that keep the synchronization channels aligned will also keep
the signals aligned. Precision calibration will consume one of the
channels on each oscilloscope and is only available on 90000 X-,
90000 Q-, V-Series, and Z-Series oscilloscopes. On the Infiniium
90000 X- and V-Series, a maximum of two oscilloscopes can
be connected to perform precision calibration. A sync port is
available for precision calibration on 90000 Q- and Z-Series.
06 | Keysight | N8834A MultiScope Application - Data Sheet
MultiScope Application Setup
1.
Set up a MultiScope connection with cables and splitters
to create a daisy chain configuration.
3. Perform calibration using one of the calibration methods.
2.
Connect all the scopes to the control PC using LAN or USB
cables or use the leader scope as the controller.
4.
Customize the display using the Infiniium Offline
oscilloscope analysis software on a PC or the Infiniium
analysis software on the leader scope.
Add waveform
Add window
07 | Keysight | N8834A MultiScope Application - Data Sheet
N8834A Infiniium MultiScope Configuration Table
90000 Q-Series / Z-Series
Max bandwidth
90000 X-Series/
V-Series
DSOX95004Q/
DSOX92004Q/ DSOV084A,
DSOZ504A,
DSOZ204A,
DSOX91304A/
DSOX96204Q/
DSOX92504Q/ DSOV134A,
DSOZ634A
DSOZ254A,
DSOX91604A/
DSOX93304Q/ DSOV164A,
DSOZ334A
DSOX92004A/
DSOV204A,
DSOX92504A/
DSOV254A,
DSOX92804A/
DSOV284A,
DSOX93304A/
DSOV334A
50 to 63 GHz 33 GHz 20 to 33 GHz
20 to 33 8 to 16
GHz
GHz
160 GSa/s
80 GSa/s 80 GSa/s
80 GSa/s 40
GSa/s
2
4
4
2
4
Max sample
rate
Max channels/
frame
Max frames
10
Sync port
Not available Available Available
External
100 MHz
reference
frequency
Calibration modes
Manual
Yes
Basic
Yes
(inter-scope
(10 ps)
intrinsic jitter 2)
Precision 1
Yes
(inter-scope
(0.15 ps)
intrinsic jitter 2)
Cable kit to
N2105A
connect to two
frames
S-Series
DSOS054A,
DSOS104A,
DSOS204A,
DSOS254A,
DSOS404A,
DSOS604A,
DSOS804A
6 to 8
GHz
20
GSa/s
2
0.5 to 4
GHz
10
GSa/s
4
90000A
9000 Series
Series
DSO90254A, DSO9064A
DSO90404A,
DSO90504A,
DSO90804A,
DSO91204A,
DSO91304A
2.5 to 13
GHz
40 GSa/s
4
600 MHz
10
GSa/s
2
1 to 4 GHz
5
GSa/s
4
10
Not available
10 MHz
10
Not available
10 MHz
10
10
Not available Not available
10 MHz
10 MHz
Yes
Yes
(10 ps)
Yes
Yes
(10 ps)
Yes
Yes
(5 ps)
Yes
Yes
(5 ps)
Yes
Yes
(N/A)
Yes
(0.5 ps)
Yes
(0.5 ps)
No
No
No
N2124A
N2124A
N2124A
N2123A
1. Precision calibration is available on 90000 Q- and Z-Series for up to 10 scopes and on 90000 X- and V-Series for up to 2 scopes.
2. Inter-scope intrinsic jitter value is used to calculate MultiScope delta-time measurement accuracy.
DSO9104A,
DSO9254A,
DSO9404A
20
GSa/s
2
10
10
GSa/s
4
08 | Keysight | N8834A MultiScope Application - Data Sheet
Recommended Oscilloscopes
The N8834A Multiscope software is compatible with Infiniium Series Oscilloscope with operating software revision 5.60 or higher. For
oscilloscopes with earlier revisions, free upgrade software is available at: www.keysight.com/find/scope-apps-sw.
Ordering Information for Infiniium Oscilloscope Multiscope Configuration
Software options
Application
License type
Infiniium analysis software
(optional)
Multiscope software –
connect 2 scopes
Floating
Multiscope software –
connect up to 5 scopes
Fixed
Floating
Multiscope software –
connect up to 10 scopes
Fixed
Floating
Multiscope software
(upgrade to connect from
2 to up to 5 scopes)
Multiscope software
(upgrade to connect from
5 to up to 10 scopes)
Fixed
Floating
Fixed
Floating
Fixed
Floating
Transportable
Server-based
User-installed
Transportable
Server-based
User-installed
Transportable
Server-based
User-installed
Transportable
Server-based
User-installed
Transportable
Server-based
User-installed
Transportable
Server-based
Infiniium
V/Z-Series
N8900A-001
Infiniium
S-Series
N8900A-001
N8834A-AFP
N8834A-ATP
N8834A-AFP
N8834A-ATP
N8834A-BFP
N8834A-BTP
N8834A-BFP
N8834A-BTP
N8834A-CFP
N8834A-CTP
N8834A-CFP
N8834A-CTP
N8834A-DFP
N8834A-DTP
N8834A-DFP
N8834A-DTP
N8834A-EFP
N8834A-ETP
N8834A-EFP
N8834A-ETP
Infiniium
90000 Series
N8900A-001
Infiniium
9000 Series
N8900A-001
N8834A-AFP
N8834A-ATP
N5435A-085
N8834A-BFP
N8834A-BTP
N5435A-086
N8834A-CFP
N8834A-CTP
N5435A-090
N8834A-DFP
N8834A-DTP
—
N8834A-EFP
N8834A-ETP
—
N8834A-AFP
N8834A-ATP
—
N8834A-BFP
N8834A-BTP
N8834A-CFP
N8834A-CTP
N8834A-DFP
N8834A-DTP
N8834A-EFP
N8834A-ETP
Note: Transportable license is required for viewing the multiscope channels on a PC via Hosted mode.
Ordering Information for InfiniiVision Oscilloscope Multiscope Configuration
Application
Infiniium analysis software (required)
License type
Floating
Multiscope software – connect 2 scopes
Floating
Multiscope software – connect up to 5 scopes
Floating
Multiscope software – connect up to 10 scopes
Floating
Multiscope software (upgrade to connect from 2 to up to 5 scopes)
Floating
Multiscope software (upgrade to connect from 5 to up to 10 scopes)
Floating
Server-based
Cable kits
N2123A
N2124A
N2105A
Infiniium 90000 X- and V-Series multi-frame cable kit for two frames
Infiniium 9000, 90000A, and S-Series multi-frame cable kit for two frames
Infiniium 90000 Q- and Z-Series multi-frame base kit for stacking two frames
Transportable
Server-based
Transportable
Server-based
Transportable
Server-based
Transportable
Server-based
Transportable
Server-based
Transportable
—
InfiniiVision Series
N8900A-001
—
N8834A-ATP
N5435A-085
N8834A-BTP
N5435A-086
N8834A-CTP
N5435A-090
N8834A-DTP
—
N8834A-ETP
—
09 | Keysight | N8834A MultiScope Application - Data Sheet
Performance Characteristics
Control PC requirement
Leader and followers
oscilloscope
requirement 1
Channels supported
Windows 7 64 bit, minimum 4 G RAM (recommended 8 G), 1 G free hard drive
space Infiniium software 5.50 and above
Note: Control PC is not required for Infiniium Series scopes if leader scope is used
as the controller. A control PC must be used if InfiniiVision scopes exist in the
multiscope configuration
Infiniium software 5.50 and above 2
Analog channel 1 to 4
Note: Digital channels on all MSO models are not supported
Trigger modes 3
Edge
Triggers on a specified slope (rising, falling, or alternating between rising and
falling) and voltage level on any channel or auxiliary trigger
Glitch
Triggers on glitches narrower than the other pulses in your waveform by
specifying a width less than your narrowest pulse and a polarity
Note: All other trigger modes are not supported
1.
Front panel control is disabled during MultiScope operation. Oscilloscope control is handled by the Infiniium
Offline software on PC or the leader scope.
2. Software 4.30 or above requires Windows 7. N2753A Infiniium Windows XP to 7 OS upgrade kit (oscilloscope
already has M890 motherboard). N2754A Infiniium Windows XP to 7 OS and M890 motherboard upgrade kit
(oscilloscope without M890 motherboard). Verify the M890 motherboard using the procedure found in the
Windows 7 upgrade kit data sheet with the publication number 5990-8569EN.
3. If leader scope is used as the controller, all trigger modes supported on the leader scope will be available in
the multiscope configuration.
10 | Keysight | N8834A MultiScope Application - Data Sheet
www.axiestandard.org
AdvancedTCA® Extensions for Instrumentation and Test (AXIe) is an
open standard that extends the AdvancedTCA for general purpose and
semiconductor test. Keysight is a founding member of the AXIe consortium.
ATCA®, AdvancedTCA®, and the ATCA logo are registered US trademarks of
the PCI Industrial Computer Manufacturers Group.
www.lxistandard.org
LAN eXtensions for Instruments puts the power of Ethernet and the
Web inside your test systems. Keysight is a founding member of the LXI
consortium.
www.pxisa.org
PCI eXtensions for Instrumentation (PXI) modular instrumentation delivers a
rugged, PC-based high-performance measurement and automation system.
11 | Keysight | N8834A MultiScope Application - Data Sheet
From Hewlett-Packard through Agilent to Keysight
For more than 75 years, we‘ve been helping you unlock measurement insights. Our
unique combination of hardware, software and people can help you reach your next
breakthrough. Unlocking measurement insights since 1939.
1939
THE FUTURE
myKeysight
www.keysight.com/find/mykeysight
A personalized view into the information most relevant to you.
Three-Year Warranty
www.keysight.com/find/ThreeYearWarranty
Keysight’s committed to superior product quality and lower total cost
of ownership. Keysight is the only test and measurement company with
three-year warranty standard on all instruments, worldwide. And, we provide
a one-year warranty on many accessories, calibration devices, systems and
custom products.
Keysight Assurance Plans
www.keysight.com/find/AssurancePlans
Up to ten years of protection and no budgetary surprises to ensure your
instruments are operating to specification, so you can rely on accurate
measurements.
Keysight Infoline
www.keysight.com/find/service
Keysight’s insight to best in class information management. Free access to
your Keysight equipment company reports and e-library.
Keysight Channel Partners
www.keysight.com/find/channelpartners
Get the best of both worlds: Keysight’s measurement expertise and product
breadth, combined with channel partner convenience.
www.keysight.com/find/n8834a
For more information on Keysight
Technologies’ products, applications or
services, please contact your local Keysight
office. The complete list is available at:
www.keysight.com/find/contactus
Americas
Canada
Brazil
Mexico
United States
(877) 894 4414
55 11 3351 7010
001 800 254 2440
(800) 829 4444
Asia Pacific
Australia
China
Hong Kong
India
Japan
Korea
Malaysia
Singapore
Taiwan
Other AP Countries
1 800 629 485
800 810 0189
800 938 693
1 800 11 2626
0120 (421) 345
080 769 0800
1 800 888 848
1 800 375 8100
0800 047 866
(65) 6375 8100
Europe & Middle East
Austria
Belgium
Finland
France
Germany
Ireland
Israel
Italy
Luxembourg
Netherlands
Russia
Spain
Sweden
Switzerland
United Kingdom
0800 001122
0800 58580
0800 523252
0805 980333
0800 6270999
1800 832700
1 809 343051
800 599100
+32 800 58580
0800 0233200
8800 5009286
800 000154
0200 882255
0800 805353
Opt. 1 (DE)
Opt. 2 (FR)
Opt. 3 (IT)
0800 0260637
For other unlisted countries:
www.keysight.com/find/contactus
(BP-02-10-16)
DEKRA Certified
ISO9001 Quality Management System
www.keysight.com/go/quality
Keysight Technologies, Inc.
DEKRA Certified ISO 9001:2015
Quality Management System
This information is subject to change without notice.
© Keysight Technologies, 2015 - 2016
Published in USA, May 5, 2016
5992-0511EN
www.keysight.com