Keysight Technologies B2900A Series Precision Source/Measure Unit Data Sheet 02 | Keysight | B2900A Series Precision Source/Measure Unit – Data Sheet Innovative SMU Provides Superior Performance and Rapid Measurement Results Single view Dual view The Keysight Technologies, Inc. B2900A Series of Precision Source/Measure Units are compact and cost-effective bench-top Source/Measure Units (SMUs) with the capability to source and measure both voltage and current. These capabilities make the B2900A Series ideal for a wide variety of IV (current versus voltage) measurement tasks that require both high resolution and accuracy. The Keysight B2900A Series of SMUs provide best-in-class performance for a modest price. They have broad voltage (±210 V) and current (± 3 A DC and ±10.5 A pulsed) sourcing capability, excellent precision (minimum 10 fA/100 nV sourcing and measuring resolution) and possess a superior color LCD graphical user interface (GUI). In addition, several task-based viewing modes dramatically improve productivity for test, debug and characterization. Roll view The Keysight B2900A Series offers unmatched measurement throughput and supports conventional SMU SCPI commands for easy test code migration. These features improve efficiency and lower the cost of ownership when integrating the SMUs into systems for production test. The Keysight B2900A Series consists of four models, the B2901A, B2902A, B2911A and B2912A, differentiated through their available features (number of digits displayed, measurement resolution, minimum timing interval, supported viewing modes, etc.) and by the number of SMU channels (one or two) they contain. This makes it easy to select the exact price/performance point you require to meet your testing needs. Feature Benefit Integrated 4-quadrant sourcing and measuring capabilities Easily and accurately measure current and voltage using a single instrument without the need to manually change any connections Measurement range: ±210 V, ±3 A (DC), ±10.5 A (pulsed) A single SMU product covers both high voltage and high current measurement needs, allowing for more standardization and simplifying inventory and support concerns. Source and measurement resolution down to 10 fA and 100 nV Can make low-level measurements using a low-cost bench-top SMU that were previously only possible using a more expensive semiconductor device analyzer. User-friendly front panel GUI with 4.3 inch color LCD display supports both graphical and numerical view modes Can quickly and easily perform measurements and display data on the front panel, thereby greatly speeding up interactive test, characterization and debug operations. 10 microsecond digitizing capability Can capture low frequency phenomena in addition to DC characteristics. Free PC-based control software Can make measurements remotely from a PC without the need to program. Supports both conventional and default SCPI commands Conventional SCPI commands provide some compatibility with older SMU code (such as Keithley 2400 series) to minimize code conversion work. Default SCPI commands support advanced B2900A Series features. Small form factor with USB2.0, LAN, GPIB and digital I/O interfaces Easy integration into rack and stack systems. 03 | Keysight | B2900A Series Precision Source/Measure Unit – Data Sheet The Best SMU Solution for a Broad Range of IV Measurements Keysight B2900A Series Precision Source/Measure Unit Testing semiconductors, discrete and passive components SMUs are popular and widespread instruments for performing IV measurements in many different fields and applications due to their integrated voltage and current sourcing and measurement capabilities. The Keysight B2900A Series provides superior performance and usability at a very reasonable price. In addition, the Keysight B2900A Series supports many functions to speed up production test and increase throughput. The versatile measurement capabilities of the B2900A Series SMUs make them an ideal choice for a variety of IV measurements such as semiconductor test, active/passive component test and general electronic device and material characterization. The B2900A Series has a broad application range that spans uses from R&D and education to industrial development, production test and automated manufacturing. Moreover, they work equally well as either standalone or system components. Diodes, laser diodes, LEDs Photodetectors, sensors Field effect transistors (FETs), bipolar junction transistors (BJTs) ICs (analog ICs, RFICs, MMICs, etc) Resistor, varistor, thermistors, switches Testing precision electronics and green energy devices Photovoltaic cells Power transistors, power devices Battery Automotive Medical instruments Power and DC bias source for circuit test Research and education New material investigations Nano devices characterization (e.g. CNT) Giant magnetic resistance (GMR) Organic devices Any precise voltage/current source and measurement Application literature is available on the Keysight web site. For more information please visit www.keysight.com/find/precisionSMU 04 | Keysight | B2900A Series Precision Source/Measure Unit – Data Sheet Integrated Source and Measurement Capabilities Simplify Difficult IV Measurement Tasks The Keysight B2900A Series reduces measurement complexity Performing IV measurements with conventional instruments such as voltage/current sources, arbitrary waveform generators (AWGs), switches, and voltage/current meters is complicated and confusing. It requires deep technical knowledge of both the measurement technique and the instrumentation to perform an accurate measurement. The Keysight B2900A SMU integrates many different source and measurement capabilities into one compact form factor. It can operate as a seamless 4-quadrant precision voltage/current Voltage source source, an electrical load, an accurate voltage/current meter, a pulse generator and an AWG. Its versatile all-in-one integrated source and measurement capabilities allow it to perform a wide variety of measurements from DC to low frequency AC without the need to change connections or use additional equipment. Moreover, the availability of 2-channel models supports the testing of devices with up to three terminals (as long as one terminal can be tied to the circuit common). If you wish to learn more about the advantages of using SMUs to make IV measurements, then please refer to the section at the back of this brochure entitled “Overview: Why use an SMU?” Current meter Current source A AWG Switch Volt meter V DUT Rack & stack solution: Multiple instruments connected together with no easy means to coordinate them. device under test Same measurement using an SMU High DUT device under test Low SMU solution: Integrated 4-quadrant voltage and current sourcing and measurement (including AWG function). 05 | Keysight | B2900A Series Precision Source/Measure Unit – Data Sheet Wide Voltage and Current Coverage for Testing a Variety of Devices Test up to 210 V and 3 A (DC) or 10.5 A (pulsed) with a single instrument The B2900A SMUs can source and measure voltages of ±210 V and currents of ±3 A (DC) or ±10.5 A (pulsed). This versatility allows you to standardize on a single SMU model and minimize support costs. These capabilities are present on both single and dual channel versions, since on the 2-channel versions both channels can be operated completely independently. Integrated sweep and arbitrary waveform measurement functionality The Keysight B2900A Series has capabilities that allow it to perform more than just simple DC and pulsed measurements. The B2900A SMUs have a built-in sweep capability that supports all of the standard sweep parameters such as linear and logarithmic modes, single and double sweep functions and constant and pulsed sweep operation. The B2900A GUI fully supports the sweep measurement function, thereby allowing sweep measurements to be made and displayed quickly from the instrument front panel. Of course, the user can also make the same sweep measurements just as efficiently on the B2900A SMUs under remote control using SCPI commands. This integrated sweep measurement capability improves efficiency and reduces measurement setup time. In addition to its resident sweep functionality, the Keysight B2900A Series also supports arbitrary waveform generation (AWG) and list sweep capabilities. The AWG and list sweep functions allow you to create waveforms with up to 2500 steps for maximum flexibility, and enable you to specify a waveform of arbitrary shape using familiar spreadsheet compatible data-entry formats. The AWG and list sweep features are especially useful when characterizing devices where the response varies greatly depending upon the applied voltage or current, since they give you the flexibility to “zoom in” on areas of interest. Built-in functions provide flexible waveform generation capabilities 200 μs Constant DC Pulse You can create current pulses of up to 10.5 amps, which helps to minimize device self-heating effects. Linear sweep Load: open Pulsed linear sweep 210 V Custom wave forms ( list sweep) 0.5 ms You can ramp up to a voltage of 200 V in 0.5 millisecond, which is useful for evaluating high-power components. Logarithmic sweep 06 | Keysight | B2900A Series Precision Source/Measure Unit – Data Sheet Unmatched Bench-top SMU Measurement Performance Source and measurement resolution down to 10 fA and 100 nV 4-wire measurement capability permits accurate low resistance measurement The Keysight B2900A Series consists of four models primarily differentiated by number of channels (one or two) and measurement and sourcing resolution. The B2901A (single channel) and B2902A (dual channel) versions possess 100 fA and 100 nV measurement resolution and 1 pA and 1 μV sourcing resolution. The B2911A (single channel) and B2912A (dual channel) precision versions possess 10 fA and 100 nV of resolution for both measurement and sourcing. All members of the Keysight B2900A Series support popular banana jack style inputs for cost-effective and flexible connectivity; for low-current measurements below 1 nA, banana jack to triaxial adapters are available. When measuring small resistances the innate cable resistance can create serious measurement error. To solve this, the Keysight B2900A Series supports a 4-wire (also known as a remote sense or Kelvin connection) measurement function. In the 4-wire scheme two of the connectors force current and the other two connectors measure voltage. Since the connectors measuring voltage do not have any current flowing through them, they can accurately sense the actual voltage at the DUT. Capture transient phenomena effortlessly The Keysight B2900A Series supports a high speed sampling measurement function that permits the capture and display of low frequency transient phenomena. The Keysight B2901A and B2902A support a 20 μs (50,000 points/s) sampling rate and the Keysight B2911A and B2912A support a 10 μs (100,000 points/s) sampling rate. Of course, the maximum achievable sampling rate depends on many factors including signal level, ambient noise and desired resolution. You can make and display accurate low-current measurements directly on the B2900A front panel. Measure large capacitive loads without oscillation Large capacitive loads can sometimes cause SMUs to oscillate. To mitigate this, the B2900A Series supports a high capacitance measurement mode. The high capacitance mode enables the measurement of large capacitive loads without having to worry about SMU oscillation. Roll view mode allows you to capture low-frequency transient phenomena. 07 | Keysight | B2900A Series Precision Source/Measure Unit – Data Sheet Ultra Fast Throughput Lowers Cost-of-test Best-in-class measurement throughput SCPI commands provide compatibility and versatility Although an excellent R&D tool, the Keysight B2900A Series is also well-suited for production test. It can achieve excellent accuracy and repeatability at even short integration times. The B2900A Series possesses the fastest measurement speed of any SMU in its class. Standard Commands for Programmable Instruments (SCPI) are a popular and easy-to-understand instrument control protocol. The Keysight B2900A Series supports two SCPI command sets, conventional and default, to provide both code compatibility and flexibility. The conventional command set has a large number of its commands compatible with those of older SMUs (such as the Keithley 2400) to minimize code conversion work. The default command set supports advanced Keysight B2900A Series features and they should be used to fully utilize its performance and capabilities. Program memory and trace buffer features improve throughput To reduce bus communication time, the B2900A Series supports a program memory feature that allows long strings of SCPI commands to be stored on the instrument. These code sequences can be executed by sending a single command across the communication bus, greatly improving throughput for frequently executed command strings.In addition, the B2900A Series supports a trace buffer that can store up to 100,000 data points. This allows the results from multiple measurements to be downloaded at once, thereby reducing data transfer time and also improving overall throughput. Maximum sweep operation reading rate per second (Source/measure to GPIB at 50 Hz) FAST! Keysight B2900A Series SMU Conventional system SMU-A Conventional system SMU-B Conventional bench-top SMU-C 0 2500 5000 7500 10000 12500 15000 Readings per second 08 | Keysight | B2900A Series Precision Source/Measure Unit – Data Sheet Overview: Why Use an SMU? Resource integration reduces measurement error Limit (compliance) feature prevents device damage An SMU is an instrument that combines the capabilities of a current source, a voltage source, a current meter and a voltage meter (along with the capability to switch easily between these various functions). Because the source and measurement circuitry is closely integrated, the user can achieve far better measurement performance with less measurement error than would be possible by using various independent instruments to make the same measurement. SMUs also possess a voltage and current limit (compliance) feature that allows the user to set limits and to protect devices from damage caused by excessive voltage or current. Although the SMU normally continues to function when it reaches the user-specified limit value, it can also be set to shutdown just like the over current protection (OCP) and over voltage protection (OVP) functions do on a power supply. Accurate timing control of source and measurement resources Feedback mechanism stabilizes voltage and current sourcing Since SMUs have the ability to very accurately measure their own current and voltage output, they have many advantages over conventional power supplies. All SMUs have internal feedback loops that provide instantaneous feedback to the sourcing circuitry, which in-turn allows the SMU output to remain accurate and stable even if the load conditions change unexpectedly. The integration of the source and measurement resources in an SMU allows much tighter synchronization than would be possible with separate instruments. Moreover, the B2900A Series provides very flexible triggering options that allow the measurements points to be defined independently from the sourced current or voltage waveform. On two channel units you can operate both channels in synchronization or independently, and under remote control you can trigger multiple units in unison using a group trigger signal. Feedback High A V Low Source block: Precision voltage and current source/sink with sweep, pulse and AWG capabilities. Voltage/current limit capability is also available. Simplified block diagram of the Keysight B2900A Series Measurement block: Precision voltage and current meter with high speed digitizing capability. 09 | Keysight | B2900A Series Precision Source/Measure Unit – Data Sheet Innovative GUI and 4.3" Color LCD Display Facilitate Fast Bench-top Test, Debug and Characterization The B2900A’s front panel has many features that make interactive use fast and friendly. These include a 4.3” color LCD display, a USB2.0 memory I/O port, an assist key, an alphanumeric keypad and a rotary knob. The 4.3” color LCD display supports both graphical and numerical view modes, and enables test setup and check test results quickly. The USB2.0 memory port supports easy data storing and porting. The Innovative graphical user interfaces, such as single view, dual view, graph view, roll view and zoom, improves usability and productivity of bench-top tests, debug and characterizations dramatically. Single view: Dual view: Single view provides basic and advanced settings and display capabilities for the selected channel from the front panel of the instrument. No additional controller or software is required. Dual view provides the basic settings and display capabilities for both channels 1 and 2. Up to 6. digits can be displayed in dual view mode. This mode is available only for the B2902A/B2912A. Graph view: Roll view: Graph View displays measurement results on XY graphs (such as I-V and I-t/V-t curves) on up to 2 channels. This is useful for quick evaluation of device characteristics, especially those obtained from sweep measurements. Roll view draws I-t or V-t curves similar to the curves drawn by a strip chart recorder. Up to 1000 acquired data points can be displayed and updated while the measurement is still in progress. Roll view’s continuous measurement capability is especially useful for monitoring low frequency phenomena. Roll view is available only for the B2911A/B2912A. 10 | Keysight | B2900A Series Precision Source/Measure Unit – Data Sheet Multiple Remote Control Choices Optimize Performance for Different Applications The Keysight B2900A Series of SMUs offers multiple options for instrument remote control at little or no cost. Four solutions are available: BenchVue, B2900A Graphical Web Interface, B2900A Quick IV Measurement Software, and EasyEXPERT group+.These multiple software control options allow you to choose the solution that best fits your particular application BenchVue BenchVue allows you to control the B2900A SMUs as voltage/current sources from a PC without the need to do any programming. In addition, because BenchVue supports a wide variety of Keysight instruments (oscilloscopes, meters, etc.) it is a good choice when you need to integrate together many different types of instruments on a benchtop. Graphical Web Interface The Keysight B2900A has a built-in LXI compliant web server that allows any Java enabled web browser (such as Internet Explorer) to control it over the LAN. The graphical web interface supports all of the basic measurement functions such as spot measurements, sweep measurements and pulsed source measurements. Since no special software is required this feature is convenient for making quick measurements on the fly. 11 | Keysight | B2900A Series Precision Source/Measure Unit – Data Sheet Quick I/V Measurement Software Keysight B2900A Quick I/V Measurement Software permits easy measurement setup and execution on a Windows-based PC. It has a user-friendly GUI that can be used with all of the B2900A’s interfaces (LAN, USB and GPIB). It can control up to four SMU channels in any configuration of single and dual channel units. EasyEXPERTgroup+ Keysight EasyEXPERT group+ provides a powerful IV parametric characterization solution for a wide range of devices and materials. It has an intuitive mouse and keyboard driven graphical user interface that simplifies common characterization tasks such as test setup and execution, data analysis, data management/protection, etc. Moreover, since it can support up to eight SMU channels (four 2-channel SMU units) EasyEXPERT group+ makes it easy to characterize multi-terminal devices. 12 | Keysight | B2900A Series Precision Source/Measure Unit – Data Sheet Available Accessories Facilitate Special Test Needs Easy banana jack connectivity with various accessories The Keysight B2900A uses convenient and low-cost banana jack terminals, which support a variety of cables, adaptors and accessories. Banana to triaxial adaptor for low current measurement Since banana jacks cannot support low current measurement (i.e. measurements of 1 nA and below), a banana jack to triaxial adapter is available to permit the use of high-performance triaxial cables. This makes it easy to connect to both triaxial-based test fixtures and wafer probers. Of course, both 2-wire and 4-wire triaxial adapters are available. Test fixtures for testing packaged devices The Keysight N1295A Device/Component Test Fixture provides a low-cost solution to quickly and easily test packaged devices and components. It has four triaxial inputs and supports voltage and current measurements of up to 42 V and 1 A. For more advanced packaged testing needs, the Keysight 16442B test fixture provides more capabilities. It offers support for higher pin count devices, more flexible connectivity and an interlock feature for safely applying voltages greater than 42 V. Keysight can supply adapters to use the 16442B interlock with the B2900A’s digital output. The Keysight N1295A Device/Component Test Fixture provides a low-cost solution to quickly and easily test packaged devices and components. Banana jack to triaxial adapters are available in both 2-wire and 4-wire versions. Using the available accessories and furnished software you can create a low-cost PC-based component testing solution. 13 | Keysight | B2900A Series Precision Source/Measure Unit – Data Sheet Specifications Specification conditions Current [A] 10.5 Current [A] Temperature 23 °C ± 5 °C Humidity 30% to 80% RH After 60 minutes warm-up Ambient temperature change less than ± 3 °C after self-calibration execution 3.03 -210 Calibration period 1 year Measurement speed 1 PLC (power line cycle) -21 -6 1.515 0.1 Voltage [V] 21 -0.1 6 210 1.515 Voltage [V] -200 -1.515 DC or pulsed 1 Pulsed only1 210 V 0.105 A 21 V 1.515 A 2 6V 3.03 A 2 200 V 1.515 A 6V 10.5 A 200 -10.5 Pulsed only1 DC or pulsed1 Max current 6 -1.515 -3.03 Maximum voltage and current Max voltage -6 1. See “Maximum pulse width and duty cycle” in Pulse Source Supplemental Characteristics for applicable maximum voltage and current. 2. Max current limitation: For 21 V/1.515 A and 6 V/3.03 A ranges, total max current is limited by the table below for using 2 channels. Max current is not limited for using 1 channel only. Maximum current limitation Ch1 voltage Ch2 voltage Max total current limitation of Ch1 and Ch2 ±(0 V < V ≤ 6 V) ±(0 V < V ≤ 6 V) Ch1 current + Ch2 current ≤ 4 A ±(0 V < V ≤ 6 V) ±(6 V < V ≤ 21 V) Ch1 current + Ch2 current x 1.6 ≤ 4 A ±(6 V < V ≤ 21 V) ±(0 V < V ≤ 6 V) Ch1 current + Ch2 current x 0.625 ≤ 2.5 A ±(6 V < V ≤ 21 V) ±(6 V < V ≤ 21 V) Ch1 current + Ch2 current ≤ 2.5 A Voltage source specifications Range B2901A/B2902A B2911A/B2912A Accuracy (% reading + offset) ±200 mV 1 μV 100 nV ±(0.015 % + 225 μV) ≤ 10 μV ±210 mV ±2 V 10 μV 1 μV ±(0.02 % + 350 μV) ≤ 20 μV ±2.1 V ±20 V 100 μV 10 μV ±(0.015 % + 5 mV) ≤ 200 μV ±21 V ±200 V 1 mV 100 μV ±(0.015 % + 50 mV) ≤ 2 mV ±210 V 1. Supplemental characteristics Programming resolution Noise (peak to peak) 0.1 Hz to 10 Hz1 Max voltage (over range) 14 | Keysight | B2900A Series Precision Source/Measure Unit – Data Sheet Current source specifications Range Programming resolution B2901A/B2902A B2911A/B2912A Accuracy (% reading + offset) Noise (peak to peak) 0.1 Hz to 10 Hz 1 Max current (over range) — 10 fA ±(0.10 % + 50 pA) ≤ 1 pA ±10.5 nA ±100 nA 1 pA 100 fA ±(0.06 % + 100 pA) ≤ 2 pA ±105 nA ±1 μA 10 pA 1 pA ±(0.025 % + 500 pA) ≤ 25 pA ±1.05 μA ±10 nA 2 ±10 μA 100 pA 10 pA ±(0.025 % + 1.5 nA) ≤ 60 pA ±10.5 μA ±100 μA 1 nA 100 pA ±(0.02 % + 25 nA) ≤ 2 nA ±105 μA ±1 mA 10 nA 1 nA ±(0.02 % + 200 nA) ≤ 6 nA ±1.05 mA ±10 mA 100 nA 10 nA ±(0.02 % + 2.5 μA) ≤ 200 nA ±10.5 mA ±100 mA 1 μA 100 nA ±(0.02 % + 20 μA) ≤ 600 nA ±105 mA ±1 A 10 μA 1 μA ±(0.03 % + 1.5 mA) ≤ 70 μA ±1.05 A ±1.5 A 10 μA 1 μA ±(0.05 % + 3.5 mA) ≤ 100 μA ±1.515 A ±3 A 100 μA 10 μA ±(0.4 % + 7 mA) ≤ 120 μA ±3.03 A ±10 A 3 100 μA 10 μA ±(0.4 % + 25 mA)4 1. 2. 3. 4. ±10.5 A Supplemental characteristics 10 nA range is not available for B2901A/B2902A. 10 A range is available only for pulse mode, not available for DC mode. Measurement speed: 0.01 PLC Source supplemental characteristics Temperature coefficient (0 to 18°C and 28 to 50°C) ± (0.1 x accuracy) /°C Max output power and source/sink limits 31.8 W ±6 V @ ±3.03 A, ±21 V @ ±1.515 A, ±210 V @ ±105 mA, four quadrant source or sink operation Current limit/compliance Accuracy is same as current source. Minimum value is 1 % of range, or 1 nA in 10 nA range. Voltage limit/compliance Accuracy is same as voltage source. Minimum value is 1 % of range, or 20 mV in 200 mV range Over range 101 % of source range for 1.5 A and 3 A ranges. 105 % of source range other than 1.5 A and 3 A ranges. No over range for 200 V range with current exceeding 105 mA pulse only condition. Over temperature protection Output turns off then resets at over temperature sensed internally Voltage output settling time Time required to reach within 0.1 % of final value at open load condition. Step is 10 % to 90 % range 200 mV, 2 V ranges < 50 μs 20 V range < 110 μs 200 V range < 700 μs Slew rate ≤ 0.36 V/μs, 20 V and 10 mA ranges, 10 MΩ load resistance Current output settling time Time required to reach within 0.1 % (0.3 % for 3 A range) of final value at short condition. Step is 10 % to 90 % range 10 nA, 100 nA ranges < 10 ms 1 μA range < 500 μs 10 μA, 100μA ranges < 250 μs 1 mA to 3 A ranges < 80 μs Noise 10 Hz to 20 MHz (V source) 3 mVrms, 20 V range V source overshoot < ±(0.1 % + 10 mV). Step is 10 % to 90 % range, resistive load I source overshoot < ±0.1 % (< ±0.3 % for 3 A range). Step is 10 % to 90 % range, resistive load Voltage source range change overshoot ≤ 250 mV. 100 kΩ load, 20 MHz bandwidth Current source range change overshoot ≤ 250 mV/R load, 20 MHz bandwidth 15 | Keysight | B2900A Series Precision Source/Measure Unit – Data Sheet Pulse source supplemental characteristics Minimum programmable pulse width 50 μs Pulse width programming resolution 1 μs Pulse width definition The time from 10 % leading to 90 % trailing edge as follows Pulse width Peak level 90% Base level 10% Maximum pulse width and duty cycle Pulsed DC or pulsed Pulsed only 1. DC Max voltage Max peak current Max base current Pulse width Max duty cycle Max voltage Max current 210 V 0.105 A 0.105 A 50 μs to 99999.9 s 99.9999 % 210 V 0.105 A 21 V 1.515 A1 1.515 A1 50 μs to 99999.9 s 99.9999 % 21 V 1.515 A1 6V 3.03A1 3.03 A1 50 μs to 99999.9 s 99.9999 % 6V 3.03 A1 200 V 1.515A 50 mA 50 μs to 2.5 ms 2.5 % 180 V 1.05 A 50 mA 50 μs to 10 ms 2.5 % 6V 10.5 A 0.5 A 50 μs to 1 ms 2.5 % Max current limitation: For 21 V/1.515 A and 6 V / 3.03 A ranges, total max current is limited by the table in page 12 for using 2 channels. Max current is not limited for using 1 channel only. Minimum pulse width at the given voltage, current and settling conditions Source value Limit value Load Source settling (% of range) Min pulse width 200 V 1.5 A 200 Ω 0.1% 1 ms 6V 10.5 A 0.6 Ω 0.1% 0.2 ms 1.5 A 200 V 65 Ω 0.1% 2.5 ms 10.5 A 6V 0.5 Ω 0.1% 0.2 ms 16 | Keysight | B2900A Series Precision Source/Measure Unit – Data Sheet Voltage measurement specifications Range Measurement resolution B2901A/B2902A B2911A/B2912A Accuracy (% reading + offset) ±200 mV 100 nV 100 nV ±(0.015 % + 225 μV) ±2 V 1 μV 1 μV ±(0.02 % + 350 μV) ±20 V 10 μV 10 μV ±(0.015 % + 5 mV) ±200 V 100 μV 100 μV ±(0.015 % + 50 mV) 1. Supplemental characteristics Current measurement specifications Range Measurement resolution B2901A/B2902A B2911A/B2912A Accuracy (% reading + offset) ±10 nA1 — 10 fA ±(0.10 % + 50 pA) ±100 nA 100 fA 100 fA ±(0.06 % + 100 pA) ±1 μA 1 pA 1 pA ±(0.025 % + 500 pA) ±10 μA 10 pA 10 pA ±(0.025 % + 1.5 nA) ±100 μA 100 pA 100 pA ±(0.02 % + 25 nA) ±1 mA 1 nA 1 nA ±(0.02 % + 200 nA) ±10 mA 10 nA 10 nA ±(0.02 % + 2.5 μA) ±100 mA 100 nA 100 nA ±(0.02 % + 20 μA) ±1 A 1 μA 1 μA ±(0.03 % + 1.5 mA) ±1.5 A 1 μA 1 μA ±(0.05 % + 3.5 mA) ±3 A 10 μA 10 μA ±(0.4 % + 7 mA) 10 μA 10 μA ±(0.4 % + 25 mA)3 ±10 A 2 1. 10 nA range is not available for B2901A/B2902A. 2. 10 A range is available only for pulse mode, not available for DC mode. 3. Measurement speed: 0.01 PLC Measurement supplemental characteristics Temperature coefficient (0 to 18°C and 28 to 50°C) ± (0.1 x accuracy) /°C Over range 102 % of measurement range for 1.5 A and 3 A ranges 106 % of measurement range other than 1.5 A and 3 A ranges Voltage measurement range change overshoot < 250 mV. 100 kΩ load, 20 MHz bandwidth Current measurement range change overshoot < 250 mV/R load, 20 MHz bandwidth Derating accuracy for measurement speed less than 1 PLC Add % of range using the following table for measurement with PLC < 1 Derating accuracy with PLC setting< 1 PLC Voltage range Current range 0.2 V 2 V to 200 V 10 nA 100 nA 1 μA to 100 mA 1 A to 3 A 0.1 PLC 0.01% 0.01% 0.1% 0.01% 0.01% 0.01% 0.01 PLC 0.05% 0.02% 1% 0.1% 0.05% 0.02% 0.001 PLC 0.5% 0.2% 5% 1% 0.5% 0.2% 17 | Keysight | B2900A Series Precision Source/Measure Unit – Data Sheet Timer and triggering specification Timer Triggering1 1. Time stamp TIMER value automatically saved when each measurement is triggered Trigger timing resolution 1 μs to 100 ms Accuracy ± 50 ppm Arm/trigger delay 0 μs to 100,000 s Arm/trigger interval B2901A/B2902A : 20 μs to 100,000 s B2911A/B2912A : 10 μs to 100,000 s Arm/trigger event 1 to 100,000 Digital I/O Trigger in to trigger out ≤ 5 μs Digital I/O Trigger in to source change ≤ 5 μs LXI Trigger in to source change Minimum 100 μs, Typical 200 μs, Maximum Unknown LXI Trigger in to measurement Minimum 100 μs, Typical 200 μs, Maximum Unknown Internal event to external LXI trigger output Minimum 100 μs, Typical 200 μs, Maximum Unknown LXI event send/receive latency Unknown Minimum trigger interval 10 μs Supplemental characteristics Other supplemental characteristics Output characteristics Sensing Modes 2-wire or 4-wire (Remote-sensing) connections Low terminal connection Chassis grounded or floating Output connectors Banana jack. Triaxial connections are recommended for sourcing and measuring less than 1 nA. A banana jack to triaxial adapter is available for low current measurement. Output location Channel 1 at front, and channel 2 at rear Maximum load Normal mode: 0.01 μF High Capacitance mode: 50 μF DC loating voltage Max ±250 V DC between low force and chassis ground Guard offset voltage (V source) < 4 mV Remote sense operation range Max voltage between High Force and High Sense = 3 V Max voltage between Low Force and Low Sense = 3 V Common mode isolation > 1 GΩ, < 4500 pF Maximum sense lead resistance: 1 kΩ for rated accuracy Sense input impedance > 10 GΩ 18 | Keysight | B2900A Series Precision Source/Measure Unit – Data Sheet High capacitance mode The high capacitance mode permits the measurement of devices with capacitances greater than the normal mode maximum load value of 0.01 μF. In high capacitance mode the maximum allowed value of the load capacitance is 50 μF. Time required to reach within 0.1 % of final value with 4.7 μF capacitive load on a fixed range at specified current range and limit value Voltage output settling time 200 mV, 2 V ranges 600 μs, at 1 A limit 20 V range 1.5 ms, at 1 A limit 200 V range 20 ms, at 100 mA limit Time required to reach within 0.1 % of final value after voltage source is stabilized on fixed range. Vout is 5 V unless noted. Current measurement settling time Mode change delay Delay into high cap mode Delay out of high cap mode 1 μA range 230 ms 10 μA, 100 μA ranges 23 ms 1 mA, 10 mA ranges 0.23 ms 100 mA to 3 A ranges 100 μs 1 μA range 230 ms 10 μA, 100 μA ranges 23 ms 1 mA to 3 A ranges 1 ms All ranges 10 ms Noise 10 Hz to 20 MHz (20 V range) 4.5 mVrms Voltage source range change overshoot (20 V range or below) < 250 mV, 20 MHz bandwidth High Capacitance mode working conditions V/I mode Voltage source mode only Range Current measurement range is limited to fixed range only. 10 nA and 100 nA ranges are not available. Current limit ≥ 1 μA Other supplemental characteristics, continued Resistance measurement Resistance measurement can be performed under either auto or manual test conditions. Auto resistance measurement is performed in current source and voltage measurement mode. The total auto resistance measurement error is shown in the table below. The total error of a manual resistance measurement can be calculated using the voltage andcurrent accuracy information as shown below. Source I mode, manual ohm measurement (4-wire) Total error = Vmeas/Isrc = R reading x (gain error % of V range + gain error % of I range + offset error of I range/Isrc value %) + (offset error of V range/Isrc value) Source V mode, manual ohm measurement (4-wire) Total error = Vsrc/Imeas = 1/ [1/R reading x (gain error % of I range + gain error % of V range + offset error of V range/Vsrc value %) + (offset error of I range/Vsrc value)] Measurement speed 1 PLC Applicable for temperature 23 °C ± 5 °C Example of total error calculation I source value = 1 mA at 1 mA range V measure range = 2 V range Total error (% reading + offset) = (0.02 % + 0.02 % + 200 nA/1 mA) + (350 μV/1 mA) = 0.06 % + 0.35 Ω 19 | Keysight | B2900A Series Precision Source/Measure Unit – Data Sheet Typical performance of auto resistance measurement (4-wire), 2 V range Range Resolution Test current Current range Total error (% reading + offset) 2Ω 1 μΩ 1A 1A 0.2 % + 0.00035 Ω 20 Ω 10 μΩ 100 mA 100 mA 0.06 % + 0.0035 Ω 200 Ω 100 μΩ 10 mA 10 mA 0.065 % + 0.035 Ω 2 kΩ 1 mΩ 1 mA 1 mA 0.06 % + 0.35 Ω 20 kΩ 10 mΩ 100 μA 100 μA 0.065 % + 3.5 Ω 200 kΩ 100 mΩ 10 μA 10 μA 0.06 % + 35 Ω 2 MΩ 1Ω 1 μA 1 μA 0.095 % + 350 Ω 20 MΩ 10 Ω 100 nA 100 nA 0.18 % + 3.5 kΩ 200 MΩ 100 Ω 10 nA 100 nA 1.08 % + 35 kΩ System speeds Maximum sweep operation reading rates (reading/second) for 50 Hz Measure speed Measure to memory Measure to GPIB Source measure to memory Source measure to GPIB < 0.001 PLC 20000 12500 19500 12500 0.01 PLC 4500 3950 4500 3950 0.1 PLC 500 490 500 490 1 PLC 49 49 49 49 Operation reading rate varies by number of sweep steps. Number of sweep steps is specified. Environmental specifications Environment For use in indoor facilities Operating 0 °C to +55 °C, 30 % to 80 % non-condensing Storage -30 °C to 70 °C, 10 % to 90 % non-condensing Altitude Operating: 0 m to 2000 m, Storage: 0 m to 4600 m Power supply 90 V to 264 V, 47 Hz to 63 Hz, 250 VA maximum EMC IEC61326-1/EN61326-1, AS/NZS CISPR 11, KC: RRA Notification amending Radio Waves Act Article 58-2 Safety IEC61010-1/EN61010-1, CAN/CSA-C22.2 No. 61010-1-04, C/US Certifications CE, cCSAus, C-Tick, KC Warm-up Dimensions Weight 1 hour Case 88 mm (2U) x 213 mm (half width) x 450 mm Working 180 mm x 260 mm x 480 mm (with handle & feet) Net 5.0 kg (B2901A, B2911A), 6.4 kg (B2902A, B2912A) Shipping 9.0 kg (B2901A, B2902A, B2911A, B2912A) 20 | Keysight | B2900A Series Precision Source/Measure Unit – Data Sheet Front panel operation Front panel interface 4.3” TFT color display (480x272, with LED backlight) with keypads and rotary knob View mode Single view, Dual view, Graph view and Roll view Hardkeys Single Trigger and Auto Trigger control, 10-key, Rotary Knob and Cursors, Channel on/off, View, Cancel/Local Softkeys Function, System and Input Assist Keys Indicators Channel (measurement) status, System status Source/measurement capabilities Sweep measurement Number of steps 1 to 2,500 Sweep mode Linear, logarithmic (log) or list Sweep direction Single or double Type DC, or pulse Min programmable value to create list sweep waveform B2901A/B2902A: Min 20 μs with 1 μs resolution B2911A/B2912A: Min 10 μs with 1 μs resolution Digitizing/sampling measurement Min trigger interval (Max speed of measurement) B2901A/B2902A: 20 μs (50,000 points/s) B2911A/B2912A: 10 μs (100,000 points/s) Data buffers Max buffer size 100,000 points / channel 104 mm 260.5 mm 480 mm 87.7 mm 212.8 mm 450 mm 21 | Keysight | B2900A Series Precision Source/Measure Unit – Data Sheet Input/Output connectivity GP-IB IEEE-488.2 Ethernet 100BASE-T / 10BASE-T USB USB 2.0 host controller (front) USB 2.0 device interface (rear) Digital I/O 1. Connector type 25-pin female D. Input/output pins 14 open drain I/O bits Absolute max input voltage 5.25 V Absolute min input voltage - 0.25 V Max logic L input voltage 0.8 V Min logic H input voltage 2.0 V Max source current 1 mA @ Vout = 0 V Max sink current 50 mA @ Vout = 5 V 5 V power supply pin Limited to 500 mA, solid state fuse protected Safety interlock pin: One active high pin and one active low pin. Activation of both pin enables output voltage > 42 V Maximum number of simultaneously triggered units (using Digital I/O)1: 8 Supplemental characteristic USB 2.0 memory I/O 4.3” color LCD display Dual view Assist key, Channel 1 numeric /alpha terminal keys and rotary knob Channel 2 terminal GPIB Digital I/O Port USB2.0/LAN Interface Program, software and drivers Programming SCPI Program memory 100 kB (2500 lines typical) LXI compliance LXI Core 2011 Software available EasyEXPERT group+, Quick I/V Measurement Software, Graphical Web Interface, BenchVue Drivers available IVI-C, IVI-COM drivers, LabVIEW drivers 22 | Keysight | B2900A Series Precision Source/Measure Unit – Data Sheet Software prerequisites EasyEXPERTgroup+ Quick IV BenchVue Operating system Microsoft Windows Vista Business SP2 or later (32bit), Microsoft Windows 7 Professional SP1 or later (32bit/64bit), Microsoft Windows 8.1 Professional or later (32bit/64bit) Supporting language English (US) .NET Framework Microsoft .NET, Framework 3.5 SP1 IO Libraries Keysight IO Libraries Suite 16.2, 16.3, 17.1 update 1 or later (for the Online execution mode) Memory 2 GB memory Display XGA 1024 x 768 (SXGA 1280 x 1024 recommended) HDD Installation: 1GB free disk space on the C drive, Test setup/result data storage: Free disk space more than 30GB is recommended Recommended GPIB I/F Keysight 82350B/C (PCI)1, 82351B(PCIe)1, 82357A (USB) 2, 3, 82357B (USB) 2, 3 National Instrument: GPIB-USB-HS (USB)2,1 Operating system Windows 7 (64 bit/32 bit) Supporting language English (US) .NET Framework Microsoft .NET, Framework 4.0 or later IO Libraries 16.0 or later Interfaces USB, GP-IB, LAN Operating system Windows 10 32-bit and 64-bit (Professional, Enterprise, Education, Home versions) Windows 8 32-bit and 64-bit (Professional, Enterprise, Core) Windows 7 SP1 and later 32-bit and 64-bit (Professional, Enterprise, Ultimate, Home Basic, Home Premium) HDD Processor: 1 GHz or faster (2 GHz or greater recommended) RAM: 1 GB (32-bit) or 2 GB (64-bit) (3 GB or greater recommended) Display resolution 1024 x 768 minimum for single instrument view (higher resolutions are recommended for multiple instrument view) Interfaces USB, GPIB, LAN, RS-232 1. A PCI or PCIe card is highly recommended because of stability and speed. 2. USB GPIB interfaces might cause serial poll error intermittently due to the intrinsic communication scheme differences. It is reported that using an even GPIB address sometimes significantly decreases the chance of the error. The NI GPIB-USB-HS is recommended for stability, and the Keysight 82357x is recommended for speed. 3. EasyEXPERT software prohibits to set the odd GPIB address to prevent the issue above. Keysight B2900A family The B2900A series is a member of B2900 Precision Instrument Family, which provide a variety of precision measurement solutions with both sourcing and measurement capabilities. The B2960A series of Low Noise Power Sources have up to 6.5-digit voltage/current sourcing resolution and a 10 μVrms noise floor. The B2980A series of picoammeter/ electrometer have down to 0.01fA and up to 10 PΩ. Further information on the B2900 Precision Instrument Family is available at http://www.keysight.com/find/b2900a. B2960A Low Noise Power Source B2980A Picoammeter and Electrometer/High R meter 23 | Keysight | B2900A Series Precision Source/Measure Unit – Data Sheet Furnished Accessories Power cable, USB cable, Quick Reference (English), CD-ROMs (including PDF manuals, Quick I/V Measurement Software and drivers, EasyEXPERT group+ install media), Keysight I/O Library Suite Ordering Information Model number B2901A Precision Source/Measure Unit, 1ch, 100 fA, 210 V, 3 A DC/10.5 A pulse B2902A Precision Source/Measure Unit, 2ch, 100 fA, 210 V, 3 A DC/10.5 A pulse B2911A Precision Source/Measure Unit, 1ch, 10 fA, 210 V, 3 A DC/10.5 A pulse B2912A Precision Source/Measure Unit, 2ch, 10 fA, 210 V, 3 A DC/10.5 A pulse Options 0B0 Download the Product Manual from the Keysight website ABA User Guide English for B2900 Series ABJ User Guide Japanese for B2900 Series AB0 User Guide Traditional Chinese for B2900 Series AB2 User Guide Simplified Chinese for B2900 Series A6J ANSI Z540-1-1994 Calibration UK6 Commercial Calibration Certificate with Test Data B2980A-1CM Rack Mount Kit Accessories N1294A-001 Banana - Triax adapter for 2-wire (non Kelvin) connection N1294A-002 Banana - Triax adapter for 4-wire (Kelvin) connection N1294A-011 Interlock cable for 16442B (1.5 m) N1294A-012 Interlock cable for 16442B (3.0 m) N1294A-031 GPIO-BNC trigger adapter N1294A-032 Digital I/O trigger cable for multiple unit control 16494A-001 Low leakage triax cable (1.5 m) 16494A-002 Low leakage triax cable (3.0 m) 16494A-005 Low leakage triax cable (4.0 m) Upgrade kit B2901AU B2901A Software Upgrade Package, Extension Support and Subscription B2902AU B2902A Software Upgrade Package, Extension Support and Subscription B2911AU B2911A Software Upgrade Package, Extension Support and Subscription B2912AU B2912A Software Upgrade Package, Extension Support and Subscription Keysight precision current-voltage analyzer series and power device analyzer series www.keysight.com/find/analyzer 24 | Keysight | B2900A Series Precision Source/Measure Unit – Data Sheet From Hewlett-Packard through Agilent to Keysight For more than 75 years, we‘ve been helping you unlock measurement insights. Our unique combination of hardware, software and people can help you reach your next breakthrough. Unlocking measurement insights since 1939. 1939 THE FUTURE myKeysight www.keysight.com/find/mykeysight A personalized view into the information most relevant to you. Three-Year Warranty www.keysight.com/find/ThreeYearWarranty Keysight’s committed to superior product quality and lower total cost of ownership. Keysight is the only test and measurement company with three-year warranty standard on all instruments, worldwide. And, we provide a one-year warranty on many accessories, calibration devices, systems and custom products. Keysight Assurance Plans www.keysight.com/find/AssurancePlans Up to ten years of protection and no budgetary surprises to ensure your instruments are operating to specification, so you can rely on accurate measurements. Keysight Infoline www.keysight.com/find/service Keysight’s insight to best in class information management. Free access to your Keysight equipment company reports and e-library. Keysight Channel Partners www.keysight.com/find/channelpartners Get the best of both worlds: Keysight’s measurement expertise and product breadth, combined with channel partner convenience. http://www.keysight.com/find/b2900a For more information on Keysight Technologies’ products, applications or services, please contact your local Keysight office. The complete list is available at: www.keysight.com/find/contactus Americas Canada Brazil Mexico United States (877) 894 4414 55 11 3351 7010 001 800 254 2440 (800) 829 4444 Asia Pacific Australia China Hong Kong India Japan Korea Malaysia Singapore Taiwan Other AP Countries 1 800 629 485 800 810 0189 800 938 693 1 800 11 2626 0120 (421) 345 080 769 0800 1 800 888 848 1 800 375 8100 0800 047 866 (65) 6375 8100 Europe & Middle East Austria Belgium Finland France Germany Ireland Israel Italy Luxembourg Netherlands Russia Spain Sweden Switzerland United Kingdom 0800 001122 0800 58580 0800 523252 0805 980333 0800 6270999 1800 832700 1 809 343051 800 599100 +32 800 58580 0800 0233200 8800 5009286 800 000154 0200 882255 0800 805353 Opt. 1 (DE) Opt. 2 (FR) Opt. 3 (IT) 0800 0260637 For other unlisted countries: www.keysight.com/find/contactus (BP-02-10-16) DEKRA Certified ISO9001 Quality Management System www.keysight.com/go/quality Keysight Technologies, Inc. DEKRA Certified ISO 9001:2015 Quality Management System This information is subject to change without notice. © Keysight Technologies, 2013 - 2016 Published in USA, April 25, 2016 5990-7009EN www.keysight.com