4-Channel, 14 S/s, 24-Bit, ±80 mV C Series Thermocouple Input Module
NI 9211
4 thermocouple or ±80 mV analog inputs
Hot-swappable operation
-40 to 70 °C operating range
NIST-traceable calibration
24-bit resolution; 50/60 Hz noise rejection
Overview
The NI 9211 C Series thermocouple input module for use with NI CompactDAQ and CompactRIO chassis includes a 24-bit delta-sigma ADC, antialiasing filters,
open-thermocouple detection, and cold-junction compensation for high-accuracy thermocouple measurements. It contains NIST-traceable calibration and a channel-to-earth
ground double-isolation barrier for safety, noise immunity, and high common-mode voltage range.
Back to Top
Requirements and Compatibility
OS Information
Driver Information
Software Compatibility
Real-Time OS
NI-DAQmx
LabVIEW
Windows
NI-RIO
LabWindows/CVI
Measurement Studio
SignalExpress
Visual Basic
Visual C#
Visual C++
Visual Studio
Visual Studio .NET
Back to Top
Comparison Tables
Thermocouple Module
Channels
Sample Rate
Resolution
Feature
NI 9213
16
1,200 S/s
24-bit
Lowest cost/channel
NI 9219
4
50 S/s/ch
24-bit
Channel-to-channel isolation
NI 9211
4
14 S/s
24-bit
Low-channel count
Back to Top
Application and Technology
1/9
www.ni.com
Application and Technology
High-accuracy C Series analog input modules for NI CompactDAQ and CompactRIO provide high-performance measurements for a wide variety of industrial, in-vehicle, and
laboratory sensors and signal types. Each module includes built-in signal conditioning and an integrated connector with screw terminal or cable options for flexible and low-cost
signal wiring. All modules feature CompactRIO Extreme Industrial Certifications and Ratings.
C Series Compatibility
The C Series hardware family features more than 50 measurement modules and several chassis and carriers for deployment. With this variety of modules, you can mix and match
measurements such as temperature, acceleration, flow, pressure, strain, acoustic, voltage, current, digital, and more to create a custom system. Install the modules in one of
several carriers to create a single module USB, Ethernet, or Wi-Fi system, or combine them in chassis such as NI CompactDAQ and CompactRIO to create a
mixed-measurement system with synchronized measurements. You can install up to eight modules in a simple, complete NI CompactDAQ USB data acquisition system to
synchronize all of the analog output, analog input, and digital I/O from the modules. For a system without a PC, CompactRIO holds up to eight modules and features a built-in
processor, RAM, and storage for an embedded data logger or control unit. For higher-speed control, CompactRIO chassis incorporate a field-programmable gate array (FPGA)
that you can program with NI LabVIEW software to achieve silicon-speed processing on I/O data from C Series modules.
Advanced Features
When used with CompactRIO, C Series analog input modules connect directly to reconfigurable I/O (RIO) FPGA hardware to create high-performance embedded systems. The
reconfigurable FPGA hardware within CompactRIO provides a variety of options for custom timing, triggering, synchronization, filtering, signal processing, and high-speed
decision making for all C Series analog modules. For instance, with CompactRIO, you can implement custom triggering for any analog sensor type on a per-channel basis using
the flexibility and performance of the FPGA and the numerous arithmetic and comparison function blocks built into the LabVIEW FPGA Module.
Key Features
High-accuracy, high-performance analog measurements for any CompactRIO embedded system, R Series expansion chassis, or NI CompactDAQ chassis
Screw terminals, BNC, D-Sub, spring terminals, strain relief, high voltage, cable, solder cup backshell, and other connectivity options
Available channel-to-earth ground double-isolation barrier for safety, noise immunity, and high common-mode voltage range
CompactRIO Extreme Industrial Certifications and Ratings
Built-in signal conditioning for direct connection to sensors and industrial devices
Visit ni.com/compactrio or ni.com/compactdaq for up-to-date information on module availability, example programs, application notes, and other developer tools.
Connectivity Accessories
NI CompactDAQ and CompactRIO systems are designed to provide flexible options for low-cost field wiring and cabling. Most C Series modules have a unique connector block
option that offers secure connections to your C Series system. Table 2 contains all of the connector blocks available for C Series I/O modules.
Table 3 lists the recommended connector block accessories for each C Series analog input module.
The NI 9932 kit provides strain relief and operator protection from high-voltage signals for any 10-position screw-terminal module.
2/9
www.ni.com
The NI 9936 consists of 10-position screw-terminal plugs for any 10-position screw-terminal module.
Visit ni.com/compactrio or ni.com/compactdaq for up-to-date information on accessories.
Back to Top
Ordering Information
For a complete list of accessories, visit the product page on ni.com.
Products
Part Number
Recommended Accessories
Part Number
No accessories required.
Back to Top
Software Recommendations
LabVIEW Professional Development System for Windows
Advanced software tools for large project development
Automatic code generation using DAQ Assistant and Instrument I/O Assistant
Tight integration with a wide range of hardware
Advanced measurement analysis and digital signal processing
Open connectivity with DLLs, ActiveX, and .NET objects
Capability to build DLLs, executables, and MSI installers
Back to Top
Support and Services
System Assurance Programs
NI system assurance programs are designed to make it even easier for you to own an NI system. These programs include configuration and deployment services for your NI PXI,
CompactRIO, or Compact FieldPoint system. The NI Basic System Assurance Program provides a simple integration test and ensures that your system is delivered completely
assembled in one box. When you configure your system with the NI Standard System Assurance Program, you can select from available NI system driver sets and application
development environments to create customized, reorderable software configurations. Your system arrives fully assembled and tested in one box with your software preinstalled.
When you order your system with the standard program, you also receive system-specific documentation including a bill of materials, an integration test report, a recommended
maintenance plan, and frequently asked question documents. Finally, the standard program reduces the total cost of owning an NI system by providing three years of warranty
coverage and calibration service. Use the online product advisors at ni.com/advisor to find a system assurance program to meet your needs.
3/9
www.ni.com
Calibration
NI measurement hardware is calibrated to ensure measurement accuracy and verify that the device meets its published specifications. To ensure the ongoing accuracy of your
measurement hardware, NI offers basic or detailed recalibration service that provides ongoing ISO 9001 audit compliance and confidence in your measurements. To learn more
about NI calibration services or to locate a qualified service center near you, contact your local sales office or visit ni.com/calibration.
Technical Support
Get answers to your technical questions using the following National Instruments resources.
Support - Visit ni.com/support to access the NI KnowledgeBase, example programs, and tutorials or to contact our applications engineers who are located in NI sales
offices around the world and speak the local language.
Discussion Forums - Visit forums.ni.com for a diverse set of discussion boards on topics you care about.
Online Community - Visit community.ni.com to find, contribute, or collaborate on customer-contributed technical content with users like you.
Repair
While you may never need your hardware repaired, NI understands that unexpected events may lead to necessary repairs. NI offers repair services performed by highly trained
technicians who quickly return your device with the guarantee that it will perform to factory specifications. For more information, visit ni.com/repair.
Training and Certifications
The NI training and certification program delivers the fastest, most certain route to increased proficiency and productivity using NI software and hardware. Training builds the skills
to more efficiently develop robust, maintainable applications, while certification validates your knowledge and ability.
Classroom training in cities worldwide - the most comprehensive hands-on training taught by engineers.
On-site training at your facility - an excellent option to train multiple employees at the same time.
Online instructor-led training - lower-cost, remote training if classroom or on-site courses are not possible.
Course kits - lowest-cost, self-paced training that you can use as reference guides.
Training memberships and training credits - to buy now and schedule training later.
Visit ni.com/training for more information.
Extended Warranty
NI offers options for extending the standard product warranty to meet the life-cycle requirements of your project. In addition, because NI understands that your requirements may
change, the extended warranty is flexible in length and easily renewed. For more information, visit ni.com/warranty.
OEM
NI offers design-in consulting and product integration assistance if you need NI products for OEM applications. For information about special pricing and services for OEM
customers, visit ni.com/oem.
Alliance
Our Professional Services Team is comprised of NI applications engineers, NI Consulting Services, and a worldwide National Instruments Alliance Partner program of more than
700 independent consultants and integrators. Services range from start-up assistance to turnkey system integration. Visit ni.com/alliance.
Back to Top
Detailed Specifications
The following specifications are typical for the range –40 to 70 °C unless otherwise noted.
Input Characteristics
Number of channels
4 thermocouple channels, 1 internal autozero channel, 1 internal cold-junction
compensation channel
ADC resolution
24 bits
Type of ADC
Delta-Sigma
Sampling mode
Scanned
Voltage measurement range
±80 mV
Works over temperature ranges defined by NIST (J, K, T, E, N, B, R, S
thermocouple types)
70 ms per channel; 420 ms total for all channels including the autozero and
cold-junction channels
Temperature measurement ranges
Conversion time
Common-mode voltage range
Channel-to-COM
±1.5 V
COM-to-earth ground
±250 V
Common-mode rejection ratio (0 to 60 Hz)
Channel-to-COM
95 dB
4/9
www.ni.com
COM-to-earth ground
>170 dB
Input bandwidth (–3 dB)
15 Hz
Noise rejection (at 50 and 60 Hz)
85 dB min
Overvoltage protection
±30 V between any input and COM
Differential input impedance
20 MΩ
Input current
50 nA
Input noise
1 μVrms
Gain error
0.05% max at 25 °C, 0.06% typ at –40 to 70 °C, 0.1% max at –40 to 70 °C
Offset error (with autozero channel on)
15 μV typ, 20 μV max
Gain error from source impedance
Add 0.05 ppm per Ω when source impedance >50 Ω
Offset error from source impedance
Add 0.05 μV typ, 0.07 μV max per Ω source impedance >50 Ω
Cold-junction compensation sensor accuracy
0 to 70 °C
0.6 °C typ, 1.3 °C max
–40 to 70 °C
1.7 °C max
633,012 hours at 25 °C; Bellcore Issue 2, Method 1, Case 3, Limited Part
Stress Method
MTBF
Note Contact NI for Bellcore MTBF specifications at other temperatures or for MIL-HDBK-217F specifications.
Temperature Measurement Accuracy
Measurement sensitivity 1
With autozero channel on
Types J, K, T, E, N
<0.07 °C
Types B
<0.25 °C
Types R, S
<0.60 °C
With autozero channel off
Types J, K, T, E, N
<0.05 °C
Types B
<0.20 °C
Types R, S
<0.45 °C
The following figures show the typical and maximum errors for each thermocouple type when used with the NI 9211 over the full temperature range. The figures also display the
maximum errors for the thermocouple types with the NI 9211 at room temperature, which is 15 to 35 °C. The figures account for gain errors, offset errors, differential and integral
nonlinearity, quantization errors, noise errors, and isothermal errors. The figures do not account for the accuracy of the thermocouple itself.
Thermocouple Types J and N Errors
Thermocouple Type K Errors
5/9
www.ni.com
Thermocouple Types T and E Errors
Thermocouple Type B Errors
Thermocouple Types R and S Errors
6/9
www.ni.com
Power Requirements
Power consumption from chassis
Active mode
170 mW max
Sleep mode
4 mW max
Thermal dissipation (at 70 °C)
Active mode
170 mW max
Sleep mode
4 mW max
Physical Characteristics
If you need to clean the module, wipe it with a dry towel.
Note For two-dimensional drawings and three-dimensional models of the C Series module and connectors, visit ni.com/dimensions and search by module number.
Screw-terminal wiring
12 to 24 AWG wire with 10 mm (0.39 in.) of insulation stripped from the end
Torque for screw terminals
0.5 to 0.6 N · m (4.4 to 5.3 lb · in.)
Ferrules
0.25 mm2 to 2.5 mm2
Weight
150 g (5.3 oz)
Safety
Safety Voltages
Connect only voltages that are within the following limits.
Channel-to-COM
±30 V max
Isolation
Channel-to-channel
None
Channel-to-earth ground
Continuous
250 Vrms, Measurement Category II
Withstand
2,300 Vrms, verified by a 5 s dielectric withstand test
Measurement Category II is for measurements performed on circuits directly connected to the electrical distribution system (MAINS 2 ). This category refers to local-level electrical
distribution, such as that provided by a standard wall outlet (for example, 115 AC voltage for U.S. or 230 AC voltage for Europe). Examples of Measurement Category II are
measurements performed on household appliances, portable tools, and similar hardware.
Caution Do not connect the NI 9211 to signals or use for measurements within Measurement Categories III or IV.
Hazardous Locations
U.S. (UL)
Class I, Division 2, Groups A, B, C, D, T4; Class I, Zone 2, AEx nC IIC T4
Canada (C-UL)
Class I, Division 2, Groups A, B, C, D, T4; Class I, Zone 2, Ex nC IIC T4
Europe (DEMKO)
EEx nC IIC T4
Safety Standards
This product is designed to meet the requirements of the following standards of safety for electrical equipment for measurement, control, and laboratory use:
IEC 61010-1, EN 61010-1
UL 61010-1, CSA 61010-1
Note For UL and other safety certifications, refer to the product label or the Online Product Certification section.
7/9
www.ni.com
Electromagnetic Compatibility
This product is designed to meet the requirements of the following standards of EMC for electrical equipment for measurement, control, and laboratory use:
EN 61326 (IEC 61326): Class A emissions; Industrial immunity
EN 55011 (CISPR 11): Group 1, Class A emissions
AS/NZS CISPR 11: Group 1, Class A emissions
FCC 47 CFR Part 15B: Class A emissions
ICES-001: Class A emissions
Note For the standards applied to assess the EMC of this product, refer to the Online Product Certification section.
Note For EMC compliance, operate this device with double-shielded cables.
CE Compliance
This product meets the essential requirements of applicable European Directives, as amended for CE marking, as follows:
2006/95/EC; Low-Voltage Directive (safety)
2004/108/EC; Electromagnetic Compatibility Directive (EMC)
Note For the standards applied to assess the EMC of this product, refer to the Online Product Certification section.
Online Product Certification
Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for this product, visit
ni.com/certification, search by module number or product line, and click the appropriate link in the Certification column.
Shock and Vibration
To meet these specifications, you must panel mount the system and either affix ferrules to the ends of the terminal wires or use the NI 9932 backshell kit to protect the
connections.
Operating vibration
Random (IEC 60068-2-64)
5 grms, 10 to 500 Hz
Sinusoidal (IEC 60068-2-6)
5 g, 10 to 500 Hz
Operating shock (IEC 60068-2-27)
30 g, 11 ms half sine, 50 g, 3 ms half sine, 18 shocks at 6 orientations
Environmental
National Instruments C Series modules are intended for indoor use only but may be used outdoors if installed in a suitable enclosure. Refer to the manual for the chassis you are
using for more information about meeting these specifications.
Operating temperature (IEC 60068-2-1, IEC 60068-2-2)
–40 to 70 °C
Storage temperature (IEC 60068-2-1, IEC 60068-2-2)
–40 to 85 °C
Ingress protection
IP 40
Operating humidity (IEC 60068-2-56)
10 to 90% RH, noncondensing
Storage humidity (IEC 60068-2-56)
5 to 95% RH, noncondensing
Maximum altitude
2,000 m
Pollution Degree (IEC 60664)
2
Environmental Management
National Instruments is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous
substances from our products is beneficial not only to the environment but also to NI customers.
For additional environmental information, refer to the NI and the Environment Web page at ni.com/environment. This page contains the environmental regulations and directives
with which NI complies, as well as other environmental information not included in this document.
Waste Electrical and Electronic Equipment (WEEE)
EU Customers At the end of their life cycle, all products must be sent to a WEEE recycling center. For more information about WEEE recycling centers and National
Instruments WEEE initiatives, visit ni.com/environment/weee.htm.
Calibration
You can obtain the calibration certificate for this device at ni.com/calibration.
Calibration interval
1 year
8/9
www.ni.com
1
Measurement sensitivity represents the smallest change in temperature that a sensor can detect. It is a function of noise. The values assume the full measurement range of the
standard thermocouple sensor according to ASTM E230-87.
2
MAINS is defined as a hazardous live electrical supply system that powers hardware. Suitably rated measuring circuits may be connected to the MAINS for measuring purposes.
Back to Top
©2010 National Instruments. All rights reserved. CompactRIO, CVI, FieldPoint, LabVIEW, Measurement Studio, National Instruments, National Instruments Alliance Partner, NI, ni.com, NI CompactDAQ, and
SignalExpress are trademarks of National Instruments. The mark LabWindows is used under a license from Microsoft Corporation. Windows is a registered trademark of Microsoft Corporation in the United States
and other countries. Other product and company names listed are trademarks or trade names of their respective companies. A National Instruments Alliance Partner is a business entity independent from National
Instruments and has no agency, partnership, or joint-venture relationship with National Instruments.
My Profile | RSS | Privacy | Legal | Contact NI © 2012 National Instruments Corporation. All rights reserved.
9/9
www.ni.com
Similar pages