Keysight E5080A E5080a ena series network analyzer Datasheet

Keysight Technologies
E5080A ENA Series Network Analyzer
- 9 kHz to 4.5/6.5/9 GHz
E5092A Configurable Multiport Test Set
Data Sheet
02 | Keysight | E5080A ENA Series Network Analyzer, E5092A Configurable Multiport Test Set - Data Sheet
Options
This document provides technical specifications for the E5080A ENA network analyzer.
E5080A-245
E5080A-265
E5080A-295
E5080A-445
E5080A-465
E5080A-495
2-port test set, 9 kHz to 4.5 GHz, with bias tees
2-port test set, 9 kHz to 6.5 GHz, with bias tees
2-port test set, 9 kHz to 9 GHz, with bias tees
4-port test set, 9 kHz to 4.5 GHz, with bias tees
4-port test set, 9 kHz to 6.5 GHz, with bias tees
4-port test set, 9 kHz to 9 GHz, with bias tees
Calibration kits and ECal modules
This E5080A data sheet also provides technical specifications for the following
calibration kits and ECal modules. For models not listed in this data sheet, please
download the free Uncertainty Calculator from (http://www.keysight.com/find/na_
calculator)to generate the curves for your calibration kit and enable ENA setup.
85032F
85033E
85092C
85093C
Calibration kit
Calibration kit
Electronic calibration (ECal) module
Electronic calibration (ECal) module
Definitions
Specification (spec.):
Warranted performance. All specifications apply at 25 ºC (± 5 ºC), unless otherwise
stated, and 90 minutes after the instrument has been turned on. Specifications
include guard bands to account for the expected statistical performance distribution,
measurement uncertainties, and changes in performance due to environmental
conditions.
Typical (typ.):
Expected performance of an average unit which does not include guardbands. It is not
covered by the product warranty.
General characteristics:
A general, descriptive term that does not imply a level of performance.
03 | Keysight | E5080A ENA Series Network Analyzer, E5092A Configurable Multiport Test Set - Data Sheet
Corrected System Performance
The specifications in this section apply to measurements made with the Keysight E5080A
network analyzer under the following conditions:
–– No averaging applied to data
–– Environmental temperature of 23 °C (± 3 °C) with less than 1 °C deviation from the
calibration temperature
–– Response and isolation calibration performed
System dynamic range 1
9 k to 100 kHz
100 k to 50 MHz
50 M to 6 GHz
6 G to 8.5 GHz
8.5 G to 9 GHz
Specification (10 Hz IFBW)
Typical (3 Hz IFBW)
120 dB
129 dB
135 dB
132 dB
120 dB
137 dB
146 dB
152 dB
145 dB
135 dB
System dynamic range (specification and actual measurement data example)
1.
The test port dynamic range is calculated as the difference between the test port rms noise floor and
the source maximum output power. The effective dynamic range must take measurement uncertainty
and interfering signals into account.
04 | Keysight | E5080A ENA Series Network Analyzer, E5092A Configurable Multiport Test Set - Data Sheet
Corrected System Performance with Calibration Kit
Corrected system performance with type-N device connectors,
85032F calibration kit
Network analyzer : E5080A
Calibration kit
: 85032F (Type-N, 50 Ω)
Calibration
: full 2-port
IF bandwidth = 10 Hz, no averaging applied to data, environmental temperature =
23 °C (± 3 °C) with < 1 °C deviation from calibration temperature, isolation calibration
performed
Specification
Description
9 k to 300 kHz
300 k to 10 MHz
10 M to 3 GHz
3 G to 6 GHz
6 G to 8.5 GHz
8.5 G to 9 GHz
Directivity
49
49
46
40
38
38
Source match
41
41
40
36
35
35
Load match
48
49
46
40
37
36
Reflection tracking
± 0.011
± 0.011
± 0.021
± 0.032
± 0.054
± 0.054
Transmission tracking
± 0.045
± 0.015
± 0.018
± 0.056
± 0.088
± 0.113
Transmission uncertainty (Specification)1
Reflection uncertainty (Specification)1
1.
Applies to the units with Serial Number Prefix MY552/SG552 and above
05 | Keysight | E5080A ENA Series Network Analyzer, E5092A Configurable Multiport Test Set - Data Sheet
Corrected system performance with type-N device connectors,
85092C electronic calibration (ECal) module
Network analyzer : E5080A
Calibration kit
: 85092C (Type-N, 50 Ω) Electronic calibration (ECal) module
Calibration
: full 2-port
IF bandwidth = 10 Hz, no averaging applied to data, environmental temperature = 23 °C
(± 3 °C) with < 1 °C deviation from calibration temperature , isolation calibration is not
performed
Specification
Description
9 k to 300 kHz
300 k to 10 MHz
10 M to 3 GHz
3 G to 6 GHz
6 G to 8.5 GHz
Directivity
45
52
49
45
45
Source match
36
44
41
36
36
Load match
37
47
44
38
38
Reflection tracking
± 0.1
± 0.04
± 0.06
± 0.07
± 0.07
Transmission tracking
± 0.082
± 0.028
± 0.053
± 0.157
± 0.167
Transmission uncertainty (Specification)1
Reflection uncertainty (Specification)1
1.
Applies to the units with Serial Number Prefix MY552/SG552 and above
06 | Keysight | E5080A ENA Series Network Analyzer, E5092A Configurable Multiport Test Set - Data Sheet
Corrected system performance with 3.5 mm device connector type,
85033E calibration kit
Network analyzer : E5080A
Calibration kit
: 85033E (3.5 mm, 50 Ω)
Calibration
: full 2-port
IF bandwidth = 10 Hz, no averaging applied to data, environmental temperature =
23 °C (± 3 °C) with < 1 °C deviation from calibration temperature, isolation calibration
performed
Specification
Description
9 k to 300 kHz
300 k to 10 MHz
10 M to 3 GHz
3 G to 6 GHz
6 G to 8.5 GHz
8.5 G to 9 GHz
Directivity
46
46
44
38
38
38
Source match
43
43
40
37
36
36
Load match
46
46
44
38
38
38
Reflection tracking
± 0.006
± 0.006
± 0.007
± 0.009
± 0.010
± 0.010
Transmission tracking
± 0.043
± 0.015
± 0.020
± 0.058
± 0.079
± 0.099
Transmission uncertainty (Specification)1
Reflection uncertainty (Specification)1
1.
Applies to the units with Serial Number Prefix MY552/SG552 and above
07 | Keysight | E5080A ENA Series Network Analyzer, E5092A Configurable Multiport Test Set - Data Sheet
Corrected system performance with 3.5 mm device connector type,
85093C electronic calibration (ECal) module
Network analyzer : E5080A
Calibration kit
: 85093C (3.5 mm, 50 Ω) Electronic calibration (ECal) module
Calibration
: full 2-port
IF bandwidth = 10 Hz, no averaging applied to data, environmental temperature = 23 °C
(± 3 °C) with < 1 °C deviation from calibration temperature , isolation calibration is not
performed
Specification
Description
300 k to 10 MHz
10 M to 3 GHz
3 G to 6 GHz
6 G to 8.5 GHz
8.5 G to 9 GHz
Directivity
45
52
50
47
47
Source match
36
44
39
34
34
Load match
37
47
44
40
39
Reflection tracking
± 0.1
± 0.04
± 0.05
± 0.07
± 0.07
Transmission tracking
± 0.086
± 0.025
± 0.049
± 0.143
± 0.152
Transmission uncertainty (Specification)1
Reflection uncertainty (Specification)1
1.
Applies to the units with Serial Number Prefix MY552/SG552 and above
08 | Keysight | E5080A ENA Series Network Analyzer, E5092A Configurable Multiport Test Set - Data Sheet
Uncorrected System Performance
User correction: OFF, system error correction: ON
Specification
Description
9 k to 50 kHz
50 k to 300 kHz
300 k to3 GHz
3 G to 6 GHz
6 G to 8.5 GHz
8.5 G to 9 GHz
Directivity
20
20
25
20
15
15
Source match
20
20
25
20
15
15
Load match
8
12
17
12
10
8
Typical
Description
9 k to 50 kHz
50 k to 300 kHz
300 k to3 GHz
3 G to 6 GHz
6 G to 8.5 GHz
8.5 G to 9 GHz
Directivity
40
40
40
35
35
35
Source match
40
40
40
35
35
35
Load match
12
18
20
20
16
14
Transmission tracking
± 0.1
± 0.1
± 0.1
± 0.1
± 0.2
± 0.2
Reflection tracking
± 0.1
± 0.1
± 0.1
± 0.1
± 0.2
± 0.2
Typical
Description
9 k to 100 kHz
100 k to 50 MHz
50 M to 6 GHz
6 G to 8.5 GHz
8.5 G to 9 GHz
Crosstalk
132
141
147
140
130
Test Port Output (Source)
Test port output frequency
Description
Specification
Frequency range
Option 245, 445
Option 265, 465
Option 295, 495
9 k to 4.5 GHz
9 k to 6.5 GHz
9 k to 9 GHz
Resolution
1 Hz
CW accuracy
Standard
Option 1E5
± 7 ppm (23 ± 3 ºC)
± 0.45 ppm
Source stability
Standard
Option 1E5
Typical
± 3.0 ppm (25 ± 5 ºC)
± 7 ppm (0 to 40 ºC)
± 0.05 ppm (0 to 40 ºC)
± 0.4 ppm/year
09 | Keysight | E5080A ENA Series Network Analyzer, E5092A Configurable Multiport Test Set - Data Sheet
Test port output power 1
Description
Specification
Nominal power (preset power)
0 dBm
Range
9 k to 100 kHz
100 k to 50 MHz
50 M to 6 GHz
6 G to 8.5 GHz
8.5 G to 9 GHz
—90 to 10 dBm
—90 to 12 dBm
—90 to 15 dBm
—90 to 12 dBm
—90 to 8 dBm
—110 to 12 dBm
—110 to 14 dBm
—110 to 17 dBm
—110 to 14 dBm
—110 to 12 dBm
Max leveled power
9 k to 100 kHz
100 k to 50 MHz
50 M to 6 GHz
6 G to 8.5 GHz
8.5 G to 9 GHz
10 dBm
12 dBm
15 dBm
12 dBm
8 dBm
12 dBm
14 dBm
17 dBm
14 dBm
12 dBm
Resolution
0.01 dB
Level accuracy
Sweep type: stepped
9 k to 50 kHz
50 k to 9 GHz
Sweep type: auto
9 k to 9 GHz
± 2.0 dB
± 1.5 dB
Level linearity
—20 dBm to maximum power
Sweep type: stepped
Sweep type: auto
—110 dBm to —20 dBm
± 0.75 dB
Typical
2
Power sweep range
9 k to 100 kHz
100 k to 50 MHz
50 M to 6GHz
6 G to 8.5 GHz
8.5 G to 9 GHz
± 2.0 dB
± 1.0 dB
± 1.0 dB
—90 to 10 dBm
—90 to 12 dBm
—90 to 15 dBm
—90 to 12 dBm
—90 to 8 dBm
Test port output signal purity
Description
Specification
Typical
Harmonics (2nd or 3rd)
(at 0 dBm)
9 k to 2 GHz
2 G to 9 GHz
< –25 dBc
< –20 dBc
Non-harmonic spurious
(at 0 dBm)
< -30 dBc
1. Source output performance on all ports.
2. Power calibration using an external power meter improves level accuracy of the test port output power.
3. Level linearity given is relative to 0 dBm.
10 | Keysight | E5080A ENA Series Network Analyzer, E5092A Configurable Multiport Test Set - Data Sheet
Test Port Input
Description
Specification
+27 dBm or ± 35 VDC
Damage Level
Test port noise floor
9 k to 100 kHz
100 k to 50 MHz
50 M to 6 GHz
6 G to 8.5 GHz
8.5 G to 9 GHz
1
Compression level
(@+10dBm input)
Magnitude
9 k to 6 GHz
6 G to 9 GHz
Phase
9 k to 6 GHz
6 G to 9 GHz
0.1 dB compression input level
9 k to 100 k
100 k to 50 MHz
50 M to 6 GHz
6 G to 8.5 GHz
8.5 G to 9 GHz
1.
Typical
IFBW=1 Hz, FOM is off.
—120 dBm
—127 dBm
—130 dBm
—130 dBm
—122 dBm
—130 dBm
—137 dBm
—140 dBm
—136 dBm
—128 dBm
0.207 dB
0.207 dB
0.04 dB
0.08 dB
0.503 deg
0.503 deg
0.3 deg
0.6 deg
10 dBm
12 dBm
15 dBm
12 dBm
10 dBm
11 | Keysight | E5080A ENA Series Network Analyzer, E5092A Configurable Multiport Test Set - Data Sheet
Trace noise
Description
Specification
Typical
0.003 dBrms
0.0015 dBrms
0.0015 dBrms
0.002 dBrms
0.001 dBrms
0.0005dBrms
0.0005 dBrms
0.0006 dBrms
0.004 dBrms
0.003 dBrms
0.003 dBrms
0.0015 dBrms
0.002 dBrms
0.002 dBrms
0.001 dBrms
0.001 dBrms
0.0005 dBrms
0.0006 dBrms
0.03 degrms
0.02 degrms
0.01 degrms
0.01 degrms
0.02 degrms
0.015 degrms
0.008 degrms
0.003 degrms
0.003 degrms
0.006 degrms
0.06 degrms
0.035 degrms
0.02 degrms
0.02 degrms
0.01 degrms
0.02 degrms
0.02 degrms
0.01 degrms
0.005 degrms
0.005 degrms
0.003 degrms
0.006 degrms
Specification
Typical
(test port input level = maximum power in
Specification)
Magnitude
(Transmission)
9 k to 30 kHz, 1 kHz IFBW
30 k to 10 MHz, 1 kHz IFBW
10 M to 6 GHz, 10 kHz IFBW
6 G to 9 GHz, 10 kHz IFBW
(Reflection)
9 k to 30 kHz, 1 kHz IFBW
30 k to 10 MHz, 1 kHz IFBW
10 M to 1.5 GHz, 10 kHz IFBW
1.5 G to 6 GHz, 10 kHz IFBW
6 GHz to 9 GHz, 10 kHz IFBW
Phase
(Transmission)
9 k to 30 kHz, 1 kHz IFBW
30 k to 100 kHz, 1 kHz IFBW
100 k to 10 MHz, 1 kHz IFBW
10 M to 6 GHz, 10 kHz IFBW
6 GHz to 9 GHz, 10 kHz IFBW
(Reflection)
9 k to 30 kHz, 1 kHz IFBW
30 k to 100 kHz, 1 kHz IFBW
100 k to 10 MHz, 1 kHz IFBW
10 M to 1.5 GHz, 10 kHz IFBW
1.5 G to 6 GHz, 10 kHz IFBW
6 GHz to 9 GHz, 10 kHz IFBW
Stability
Description
Magnitude
9 k to 3 GHz
3 G to 6 GHz
6 G to 9 GHz
Phase
9 k to 3 GHz
3 G to 6 GHz
6 G to 9 GHz
± 0.005 dB/°C
± 0.01 dB/°C
± 0.04 dB/°C
± 0.1 deg/°C
± 0.2 deg/°C
± 0.8 deg/°C
12 | Keysight | E5080A ENA Series Network Analyzer, E5092A Configurable Multiport Test Set - Data Sheet
Dynamic accuracy 1, 2
Description
Specification
Magnitude
10 dBm
5 dBm
0 dBm
—5 dBm
—10 dBm (Ref.)
—15 dBm
—20 dBm
—30 dBm
—40 dBm
—50 dBm
—60 dBm
—70 dBm
—80 dBm
—90 dBm
—100 dBm
± 0.207 dB
± 0.071 dB
± 0.027 dB
± 0.012 dB
± 0.006 dB
± 0.009 dB
± 0.011 dB
± 0.014 dB
± 0.018 dB
± 0.021 dB
± 0.029 dB
± 0.072 dB
± 0.116 dB
± 0.159 dB
± 0.202 dB
Phase
10 dBm
5 dBm
0 dBm
—5 dBm
—10 dBm (Ref.)
—15 dBm
—20 dBm
—30 dBm
—40 dBm
—50 dBm
—60 dBm
—70 dBm
—80 dBm
—90 dBm
—100 dBm
± 5.02 deg
± 0.67 deg
± 0.14 deg
± 0.06 deg
± 0.04 deg
± 0.05 deg
± 0.06 deg
± 0.08 deg
± 0.10 deg
± 0.13 deg
± 0.18 deg
± 0.47 deg
± 0.76 deg
± 1.05 deg
± 1.34 deg
Typical
1. Accuracy of the test port input power reading is relative to –10 dBm reference input power level. IFBW is equal to or lower than 10 Hz.
2. Applies to the units with Serial Number Prefix MY552/SG552 and above
13 | Keysight | E5080A ENA Series Network Analyzer, E5092A Configurable Multiport Test Set - Data Sheet
Group delay 1
Description
Specification
Aperture (selectable)
(frequency span)/(number of points - 1)
Maximum aperture
25% of frequency span
Typical
Minimum delay
Limited to measuring no more than 180° of phase
change within the minimum aperture.
Accuracy
See graph below (typical)
1.
Group delay is computed by measuring the phase change within a specified step (determined by the frequency span and the number of points per sweep).
The following graph shows group delay accuracy with type-N connectors, full 2-port
calibration and a 10 Hz IF bandwidth.
–– Calibration kit (85032F).
–– Insertion loss is assumed to be < 2 dB.
In general, the following formula can be used to determine the accuracy, in seconds, of a
specific group delay measurement:
± phase accuracy (degrees)/[360 x aperture (Hz)]
14 | Keysight | E5080A ENA Series Network Analyzer, E5092A Configurable Multiport Test Set - Data Sheet
General Information
Description
IF band width
Range
General characteristic
1 Hz to 1.5 MHz
Nominal settings are:
1, 2, 3, 4, 5, 7, 10, 15, 20, 30, 40, 50, 70, 100,
150, 200, 300, 400, 500, 700, 1 k, 1.5 k, 2 k, 3 k,
4 k, 5 k, 7 k, 10 k, 15 k, 20 k, 9 k, 40 k, 50 k, 70 k,
100 k, 150 k, 200 k, 300 k, 400 k, 500 k, 700 k,
1M, 1.5 M
Front panel
Description
Typical
General characteristic
Test Ports
Type –N, female, 50 Ω (nominal)
Display
Type
Resolution
12.1 inch TFT color LCD with touch screen
WXGA (1280 x 800) 1
Universal serial bus jack, Type A configuration;
female; provides connection to mouse, key board,
printer, ECal module, USB power sensor, or USB/
GPIB interface
USB host port
Rear panel
Description
Typical
External trigger input connector
Type
Input level
BNC, female
Low threshold voltage: 0.5 V
High threshold voltage: 2.1 V
Input level range: 0 to + 5 V
≥ 2 μsec
Positive or negative
Pulse width
Polarity
External trigger output connector
Type
Maximum output current
Output level
BNC, female
50 mA
Low level voltage: 0 V
High level voltage: 5 V
1 μsec
Positive or negative
Pulse width
Polarity
External reference signal input connector
Type
Input frequency
Input level
Internal reference signal output connector
Type
Output frequency
Signal Type
Output level
Output Impedance
1.
General characteristic
BNC, female
10 MHz ± 10 ppm
—3 to +10 dBm
BNC, female
10 MHz ± 7 ppm
Sinewave
0 dBm ± 3 dB into 50 Ω
50 Ω
Valid pixels are 99.99% and more. Below 0.02% of fixed points of black, blue, green or red are not regarded as failure.
15 | Keysight | E5080A ENA Series Network Analyzer, E5092A Configurable Multiport Test Set - Data Sheet
Description
Typical
Internal reference signal oven connector
(option 1E5)
Type
Output frequency
Output level
10 MHz ± 0.45 ppm
0 dBm minimum
Bias tee input connector
Type
Damage level
Max. current
(no degradation in RF specifications)
Over current protection
General characteristic
BNC, female
BNC, female
± 35 V, 500 mA DC
± 300 mA at 100 kHz to 9 GHz
500 mA (with Fuse)
Video output
15-pin mini D-Sub; female; drives WXGA
compatible monitors
GPIB
24-pin D-Sub (Type D-24), female; compatible
with IEEE-488
USB host port
Universal serial bus jack, Type A configuration;
female; provides connection to mouse, key board,
printer, ECal module, USB power sensor, or USB/
GPIB interface
USB (USBTMC 1) interface port
Universal serial bus jack, Type B configuration
(4 contacts inline); female; provides connection
to an external PC; compatible with USBTMCUSB488 and USB 2.0.LA
LAN
10/100BaseT Ethernet, 8-pin configuration; auto
selects between the two data rates
Handler I/O port
36-pin centronics, female; provides connection
to handler system
Line power 2
Frequency
Power consumption 3
47 Hz to 63 Hz
2-port: 135 W
4-port: 165 W
90-264 VAC (Vpeak > 120 V)
300 VA max.
Voltage
VA max
1. USB Test and Measurement Class (TMC) interface that communicates over USB, complying with the IEEE 488.1 and IEEE 488.2 standards.
2. A third-wire ground is required.
3. At preset.
Description
AUX input connector
Type
Input range
Accuracy
Specification
Typical
BNC, female x 2 ports
± 1 V or ± 10 V selectable
1 % ± 1 mV for 1 V input range
1 % ± 10 mV for 10 V input range
16 | Keysight | E5080A ENA Series Network Analyzer, E5092A Configurable Multiport Test Set - Data Sheet
EMC, safety, environment and compliance
Description
General characteristics
EMC
European Council Directive 2004/108/EC
IEC 61326-1:2012
EN 61326-1:2013
CISPR 11:2009 +A1:2010
EN 55011: 2009 +A1:2010
Group 1, Class A
IEC 61000-4-2:2008
EN 61000-4-2:2009
4 kV CD / 8 kV AD
IEC 61000-4-3:2006 +A1:2007 +A2:2010
EN 61000-4-3:2006 +A1:2008 +A2:2010
3 V/m, 80-1000 MHz, 1.4 - 2.0 GHz /1V/m, 2.0 - 2.7 GHz, 80% AM
IEC 61000-4-4:2004 +A1:2010
EN 61000-4-4:2004 +A1:2010
1 kV power lines / 0.5 kV signal lines
IEC 61000-4-5:2005
EN 61000-4-5:2006
0.5 kV line-line / 1 kV line-ground
IEC 61000-4-6:2008
EN 61000-4-6:2009
3 V, 0.15-80 MHz, 80% AM
IEC 61000-4-8:2009
EN 61000-4-8:2010
30A/m, 50/60Hz
IEC 61000-4-11:2004
EN 61000-4-11:2004
0.5-300 cycle, 0% / 70%
ICES-001:2006 Group 1, Class A
AS/NZS CISPR11:2004 Group 1, Class A
KN11, KN61000-6-1 and KN61000-6-2
Group 1, Class A
Safety
European Council Directive 2006/95/EC
IEC 61010-1:2001/EN 61010-1:2001
Measurement Category I
Pollution Degree 2
Indoor Use
CAN/CSA C22.2 No. 61010-1-12
Measurement Category I
Pollution Degree 2
Indoor Use
Environment
This product complies with the WEEE Directive (2002/96/EC) marking requirements. The affixed label indicates that you must not discard this electrical/electronic product in domestic
household waste.
Product Category: With reference to the equipment types in the WEEE Directive Annex I, this
product is classed as a “Monitoring and Control instrumentation” product.
Do not dispose in domestic household waste. To return unwanted products, contact your local
Keysight office, or see http://www.keysight.com/environment/product/ for more information.
Compliance
Class C
17 | Keysight | E5080A ENA Series Network Analyzer, E5092A Configurable Multiport Test Set - Data Sheet
Analyzer environmental specifications and dimensions
Description
General characteristic
Operating environment
Temperature
Error-corrected temperature range
Humidity
Altitude
Vibration
0 °C to +40 °C
23 °C (± 3 °C) with < 1 °C deviation from calibration temperature
20% to 80 % at wet bulb temperature < +29 °C (non-condensation)
0 to 2,000 m (0 to 6561 feet)
0.21 G maximum, 5 Hz to 500 Hz
Non-operating environment
Temperature
Humidity
Altitude
Vibration
—10 °C to +60 °C
20 % to 90 % at wet bulb temperature < +40 °C (non-condensation)
0 to 4,572 m (0 to 15,000 feet)
0.5 G maximum, 5 Hz to 500 Hz
Weight (net)
Option 245/265/295: 20.3 kg
Option 445/465/495: 22.7 kg
Dimensions
Dimensions (front view, E5080A with option 245/265/295, in millimeters)
18 | Keysight | E5080A ENA Series Network Analyzer, E5092A Configurable Multiport Test Set - Data Sheet
Dimensions (front view, E5080A with option 445/465/495, in millimeters)
Dimensions (rear view with option 1E5, in millimeters)
19 | Keysight | E5080A ENA Series Network Analyzer, E5092A Configurable Multiport Test Set - Data Sheet
Dimensions (rear view with option 1E5, in millimeters)
Measurement Throughput Summary
Measurement throughput data is supplemental performance data.
Common condition for the measurement throughput data:
–– Analyzer display turned off with: DISPlay:ENABle OFF, DISPlay:VISible OFF
–– Number of traces = 1
–– firmware version: A.11.00
Cycle time for measurement completion
Sweep type: Auto
System error correction: OFF
Sweep type: Stepped
System error correction: ON
Number of Points
51
201
401
1601
51
201
401
1601
Start 800 MHz, stop 1 GHz, 500 kHz IF bandwidth
Uncorrected
1.3
1.8
2.5
7.0
1.5
2.8
4.3
13
2-port cal
2.2
3.0
5.2
17
2.6
5.1
8.2
24
4-port cal
4.0
6.2
12
31
4.7
11
18
51
Start 800 MHz, stop 1 GHz, 1 kHz IF bandwidth
Uncorrected
51
196
389
1546
51
196
389
1546
2-port cal
101
391
777
3092
101
391
777
3092
4-port cal
201
781
1554
6187
201
781
1554
6187
Start 100 kHz, stop 4.5 GHz, 500 kHz IF bandwidth
Uncorrected
3.1
5.4
7.7
12
3.1
5.4
7.8
20
2-port cal
5.7
11
15
24
5.8
11
15
39
4-port cal
11
21
31
50
11
21
31
81
Unit: ms
20 | Keysight | E5080A ENA Series Network Analyzer, E5092A Configurable Multiport Test Set - Data Sheet
Start 100 kHz, stop 4.5 GHz, 1 kHz IF bandwidth
Uncorrected
53
199
393
1554
53
199
393
1554
2-port cal
104
396
784
3107
104
396
784
3107
4-port cal
207
792
1569
6217
207
792
1569
6217
Start 100 kHz, stop 9 GHz, 500 kHz IF bandwidth
Uncorrected
3.8
6.1
8.4
18
3.8
6.1
8.4
20
2-port cal
7.0
12
17
35
7.0
12
17
40
4-port cal
14
23
33
73
14
23
34
83
Start 100 kHz, stop 9 GHz, 1 kHz IF bandwidth
Uncorrected
53
199
393
1554
53
199
393
1554
2-port cal
105
397
785
3108
105
397
785
3108
4-port cal
210
794
1571
6219
210
794
1571
6219
Unit: ms
Cycle time vs. number of points
Condition: Start = 800 MHz, Stop = 1 GHz, 500 kHz IF bandwidth
Number of points
Sweep type: Auto
System error correction: OFF
Sweep type: Stepped
System error correction: ON
3
1.1
1.0
11
1.1
1.1
51
1.3
1.5
101
1.4
2.0
201
1.8
2.8
401
2.5
4.3
801
3.9
7.0
1601
7.0
13
Unit: ms
Cycle time vs. IF bandwidth
Condition: Frequency = 4 GHz, NOP = 201, system error correction: OFF
IF BW
(Hz)
Cycle
time (ms)
IF BW
(Hz)
Cycle
time (ms)
IF BW
(Hz)
Cycle
time (ms)
IF BW
(Hz)
Cycle
time (ms)
IF BW
(Hz)
Cycle
time (ms)
IF BW
(Hz)
Cycle
time (ms)
10
19299
100
1932
1000
195
10000
21
100000
3.1
1000000
1.8
1500000
1.8
15
12867
150
1288
1500
131
15000
14
150000
2.4
20
9651
200
967
2000
98
20000
11
200000
2.1
30
6434
300
645
3000
66
30000
7.6
300000
1.9
40
4826
400
484
4000
50
40000
6.0
400000
1.8
50
3861
500
388
5000
40
50000
5.0
500000
1.8
70
2736
700
275
7000
29
70000
3.9
700000
1.8
21 | Keysight | E5080A ENA Series Network Analyzer, E5092A Configurable Multiport Test Set - Data Sheet
Data transfer time (ms) 1
Number of Points
51
201
401
1601
64-bit floating point
2.7
6.8
13
45
32-bit floating point
1.9
4.0
6.8
23
ASCII
6.1
21
37
143
1.4
1.4
2.0
SCPI over GPIB
2
SCPI over 1 Gbps LAN (Socket) 2
REAL 64
1.4
REAL 32
1.4
1.4
1.4
1.6
ASCII
4.3
13
23
86
SCPI over 1 Gbps LAN (SICL-LAN) 2
REAL 64
1.9
1.8
1.8
2.6
REAL 32
2.0
1.9
1.9
2.1
ASCII
2.8
4.3
6.4
20
2.3
2.5
2.5
3.0
SCPI over USB (SICL-USB) 2
REAL 64
REAL 32
2.3
2.3
2.5
2.7
ASCII
2.3
5.2
8.2
28
SCPI over GPIB/USB (82357B)
REAL 64
5.9
9.5
15
44
REAL 32
5.3
7.1
9.6
25
ASCII
36
133
265
1061
1. Supplemental performance data. Data transfer time varies depending on the type of PC and control software.
2. Transferred LogMag S11 data, using :CALC{ch}:MEAS{tr}:DATA:FDAT?.
22 | Keysight | E5080A ENA Series Network Analyzer, E5092A Configurable Multiport Test Set - Data Sheet
E5092A Configurable multiport test set
The section provides test input/output performance without calibration by the E5080A.
Table 46. Test set input/output performance
Description
Specification
Frequency range
50 MHz to 20 GHz
Damage level
Typical
20 dBm, ± 35 VDC
Table 47. Option E5092A-020 port performance
Description
Specification
Typical
Load match (selected port)
SPDT switch 1
50 MHz to 2 GHz
2 GHz to 4 GHz
4 GHz to 8 GHz
8 GHz to 10 GHz
10 GHz to 18 GHz
18 GHz to 20 GHz
17 dB
11 dB
8 dB
7 dB
4 dB
4 dB
SP4T switch 2
50 MHz to 2 GHz
2 GHz to 3 GHz
3 GHz to 8 GHz
8 GHz to 10 GHz
10 GHz to 18 GHz
18 GHz to 20 GHz
17 dB
11 dB
8 dB
7 dB
4 dB
4 dB
Load match (unselected port)
SPDT switch 1
50 MHz to 3 GHz
3 GHz to 10 GHz
10 GHz to 16 GHz
16 GHz to 18 GHz
18 GHz to 20 GHz
17 dB
11 dB
8 dB
6 dB
4 dB
SP4T switch 2
50 MHz to 3 GHz
3 GHz to 10 GHz
10 GHz to 16 GHz
16 GHz to 18 GHz
18 GHz to 20 GHz
17 dB
11 dB
8 dB
6 dB
4 dB
Load match (common port)
SPDT switch 1
50 MHz to 2 GHz
2 GHz to 4 GHz
4 GHz to 8 GHz
8 GHz to 10 GHz
10 GHz to 20 GHz
16 dB
11 dB
8 dB
7 dB
4 dB
SP4T switch 2
50 MHz to 1.3 GHz
1.3 GHz to 4 GHz
4 GHz to 8 GHz
8 GHz to 10 GHz
10 GHz to 20 GHz
16 dB
11 dB
8 dB
7 dB
4 dB
1. SPDT: Single-pole-double-throw switches. Applies to SW5, SW6, SW7, SW8, SW9 and SW10 in the E5092A. (See Figure 20.)
2. SP4T: Single-pole-four-throw switches. Applies to SW1, SW2, SW3 and SW4 in the E5092A. (See Figure 20.)
23 | Keysight | E5080A ENA Series Network Analyzer, E5092A Configurable Multiport Test Set - Data Sheet
Table 48. Option E5092A-020 port performance (continued)
Description
Specification
Typical
Insertion loss
SPDT switch 1
50 MHz to 100 MHz
100 MHz to 2 GHz
2 GHz to 3 GHz
3 GHz to 4 GHz
4 GHz to 6 GHz
6 GHz to 8 GHz
8 GHz to 10 GHz
10 GHz to 14 GHz
14 GHz to 18 GHz
18 GHz to 20 GHz
4 dB
3.5 dB
4.5 dB
5 dB
5.5 dB
7 dB
8 dB
8.5 dB
10 dB
11.5 dB
SP4T switch 2
50 to 100 MHz
100 MHz to 2 GHz
2 GHz to 3 GHz
3 GHz to 4 GHz
4 GHz to 6 GHz
6 GHz to 8 GHz
8 GHz to 10 GHz
10 GHz to 14 GHz
14 GHz to 18 GHz
18 GHz to 20 GHz
4 dB
3.5 dB
4.5 dB
5.5 dB
6 dB
7.5 dB
8.5 dB
9.5 dB
10.5 dB
12 dB
Stability per switch
Condition: Enviroment Temperature +23 °C ± 3 °C and Internal DC source: ≤ 100 mA (Sum of 4 channels), no heat source and no wall close to the unit.
0.003 dB/°C (SPD)
0.005 dB/°C (SPD)
0.008 dB/°C (SPD)
50 M to 4 GHz
4 G to 12 GHz
12 G to 20 GHz
Condition: besides the above
50 M to 4 GHz
4 G to 12 GHz
12 G to 20 GHz
Isolation 3
50 MHz to 500 MHz
500 MHz to 1 GHz
1 GHz to 2 GHz
2 GHz to 6 GHz
6 GHz to 10 GHz
10 GHz to 18 GHz
18 GHz to 20 GHz
0.007 dB/°C (SPD)
0.012 dB/°C (SPD)
0.017 dB/°C (SPD)
65 dB
80 dB
85 dB
90 dB
85 dB
75 dB
65 dB (Over arbitrarily test ports)
1. SPDT: Single-pole-double-throw switches. Applies to SW5, SW6, SW7, SW8, SW9 and SW10 in the E5092A. (See Figure 20.)
2. SP4T: Single-pole-four-throw switches. Applies to SW1, SW2, SW3 and SW4 in the E5092A. (See Figure 20.)
3. This specification is defined when all ports are terminated with a 50 ohm load.
24 | Keysight | E5080A ENA Series Network Analyzer, E5092A Configurable Multiport Test Set - Data Sheet
Table 49. Control line
Description
Specification
Number of groups
4
Group A: 8 bits
Group B,C,D: 4 bits
Input voltage range1
0 V to +5 V (positive input)
—5 V to 0 V (negative input)
Maximum current
Group A, B: 50 mA in total of each group
Group C, D: 500 uA in total of each group
Impedance
Typical
Group A, B: < 10 ohm
Group C, D: < 200 ohm
Table 50. DC source
Description
Specification
Number of sources
4
Output voltage range
Output voltage accuracy
0 V to +5.2 V (nominal) 2
± 3 % of setting (+1 V to +5 V) at 1 M ohm load impedance
Voltage resolution
Maximum current
Typical
10 mV (nominal) 3
150 mA for each source
Output impedance
< 5 ohm
Table 51. Operating storage environment
Description
General characteristics
Temperature
+5 °C to +40 °C
Humidity
20 to 80 % at wet bulb temperature < +29 °C (non-condensing)
Altitude
0 to 2,000m (0 to 6,561 feet)
Vibration
0.21 G max., 5 to 500 Hz
Table 52. Non-operating storage environment
Description
General characteristics
Temperature
—10 °C to +60 °C
Humidity
20 to 90 % at wet bulb temperature < +40 °C (non-condensing)
Altitude
0 to 4,572 m (0 to 15,000 feet)
Vibration
0.5 G max., 5 Hz to 500 Hz
Table 53. Front panel information
Description
General characteristics
RF connectors
SMA (female)
Test ports
38 ports
Control line
15-pin D-sub, female
25-pin D-sub, female
1. Input voltage will be clipped at about ± 5.2 V when over this range.
2. The output voltage can be set in this range.
3. The output voltage resolution becomes effective between 0 V to 5.2 V.
25 | Keysight | E5080A ENA Series Network Analyzer, E5092A Configurable Multiport Test Set - Data Sheet
Table 54. Rear panel information
Description
General characteristics
USB port
Type B-receptacle, provide connection to the E5080A
Line power
Frequency
Voltage
1
47 to 63 Hz
90 to 132 VAC, or 198 to 264 VAC (automatically switched)
300 VA max.
VA max
1.
1A third-wire ground is required.
Table 55. Test set dimensions and block diagram
Description
General characteristics
Weight
E5092A Option 020
9 kg
Dimensions
Figure 10. Dimensions (front view, with Option E5092A-020, in millimeters, nominal)
Figure 11. Dimensions (pitch between switches, with Option E5092A-020, in millimeters, nominal)
26 | Keysight | E5080A ENA Series Network Analyzer, E5092A Configurable Multiport Test Set - Data Sheet
37
51
83
133
129
Figure 11. Dimensions (pitch between switches, with Option E5092A-020, in millimeters, nominal)
22
13
18
32
133
146
472
417
144
126
23
Figure 13. Dimensions (side view, with Option E5092A-020, in millimeters, nominal)
Figure 14. Switch configuration (E5092A-020)
27 | Keysight | E5080A ENA Series Network Analyzer, E5092A Configurable Multiport Test Set - Data Sheet
Control Line 1 (Group A)
DC Source (Group A)
Max Current
Output
Voltage
DC Source Output
Control Line 1 (Group A)
Control Line 2 (Group A)
Positive Input
(High)
Control Line 3 (Group A)
Output
Impedance
:
Negative Input
(Low)
DC Source (Group B)
DC Source Output
Impedance
Control Line 2 (Group B)
Positive Input (High)
Negative Input (Low)
DC Source (Group C)
DC Source Output
Control Line 2 (Group C)
Positive Input (High)
Negative Input (Low)
DC Source (Group D)
DC Source Output
Control Line 2 (Group D)
Positive Input (High)
Negative Input (Low)
Figure 15. DC control line (E5092A-020)
Figure 16. Control line pin assignment (E5092A-020)
:
Control Line 4 (Group A)
:
Control Line 8 (Group A)
Control Line 1 (Group B)
Control Line 2 (Group B)
Control Line 3 (Group B)
Control Line 4 (Group B)
Control Line 1 (Group C)
Control Line 2 (Group C)
Control Line 3 (Group C)
Control Line 4 (Group C)
Control Line 1 (Group D)
Control Line 2 (Group D)
Control Line 3 (Group D)
Control Line 4 (Group D)
28 | Keysight |E5080A ENA Series Network Analyzer, E5092A Configurable Multiport Test Set - Data Sheet
From Hewlett-Packard through Agilent to Keysight
For more than 75 years, we‘ve been helping you unlock measurement insights. Our
unique combination of hardware, software and people can help you reach your next
breakthrough. Unlocking measurement insights since 1939.
1939
THE FUTURE
myKeysight
www.keysight.com/find/mykeysight
A personalized view into the information most relevant to you.
www.lxistandard.org
LAN eXtensions for Instruments puts the power of Ethernet and the
Web inside your test systems. Keysight is a founding member of the LXI
consortium.
Three-Year Warranty
www.keysight.com/find/ThreeYearWarranty
Keysight’s committed to superior product quality and lower total cost of
ownership. Keysight is the only test and measurement company with a
three-year warranty standard on all instruments, worldwide. And, we provide
a one-year warranty on many accessories, calibration devices, systems and
custom products.
Keysight Assurance Plans
www.keysight.com/find/AssurancePlans
Up to ten years of protection and no budgetary surprises to ensure your
instruments are operating to specification, so you can rely on accurate
measurements.
Keysight Infoline
www.keysight.com/find/service
Keysight’s insight to best in class information management. Free access to
your Keysight equipment company reports and e-library.
Keysight Channel Partners
www.keysight.com/find/channelpartners
Get the best of both worlds: Keysight’s measurement expertise and product
breadth, combined with channel partner convenience.
For more information on Keysight
Technologies’ products, applications or
services, please contact your local Keysight
office. The complete list is available at:
www.keysight.com/find/contactus
Americas
Canada
Brazil
Mexico
United States
(877) 894 4414
55 11 3351 7010
001 800 254 2440
(800) 829 4444
Asia Pacific
Australia
China
Hong Kong
India
Japan
Korea
Malaysia
Singapore
Taiwan
Other AP Countries
1 800 629 485
800 810 0189
800 938 693
1 800 11 2626
0120 (421) 345
080 769 0800
1 800 888 848
1 800 375 8100
0800 047 866
(65) 6375 8100
Europe & Middle East
Austria
Belgium
Finland
France
Germany
Ireland
Israel
Italy
Luxembourg
Netherlands
Russia
Spain
Sweden
Switzerland
United Kingdom
0800 001122
0800 58580
0800 523252
0805 980333
0800 6270999
1800 832700
1 809 343051
800 599100
+32 800 58580
0800 0233200
8800 5009286
800 000154
0200 882255
0800 805353
Opt. 1 (DE)
Opt. 2 (FR)
Opt. 3 (IT)
0800 0260637
For other unlisted countries:
www.keysight.com/find/contactus
(BP-01-01-16)
www.keysight.com/go/quality
Keysight Technologies, Inc.
DEKRA Certified ISO 9001:2008
Quality Management System
This information is subject to change without notice.
© Keysight Technologies, 2015, 2016
Published in USA, January 19, 2016
5992-0291EN
www.keysight.com
Similar pages