Keysight N432A Thermistor power meter Datasheet

Keysight Technologies
N432A
Thermistor Power Meter
Data Sheet
02 | Keysight | N432A Thermistor Power Meter – Data Sheet
Why Keysight’s Power Meters and Sensors?
Keysight’s only power meter that supports thermistor mount with
useful enhancements for metrology and calibration lab environments.
Key features
–– Frequency range: 100 kHz to
18 GHz (sensor dependent)
–– Power range: –30 dBm (1 μW) to
+10 dBm (10 mW)
–– High accuracy ± (0.1% of reading
+ 0.5 μW), excellent for 1 mW
transfer calibration
(with 478A-H75)
–– Built in 6.5 digit ADC eliminates
the need of external DMM
–– Selectable bridge resistance
(100/200/300/400 Ω)
Reliable, high-performing solutions
Every power meter and sensor from Keysight consistently delivers great results.
A sure investment for many years to come
Code-compatibility between power meters reduces the need for re-coding. Not only that,
all Keysight power meters are backward-compatible with most legacy power sensors.
One specific application: One right solution
Keysight offers a wide selection of power meters and sensors for practically all
application needs—wireless communications, radar pulse measurements, component
test, and more.
–– Remote interface flexibility
–– DC substitution measurement,
traceable to the U.S. National
Institute of Standards and
Technology (NIST)
Global network support
No matter where you are, Keysight is committed to giving you the 24-hour support you
need regarding our products, applications, or services.
Keysight’s power meters have long been recognized as the
industry standard for RF and microwave power measurements.
03 | Keysight | N432A Thermistor Power Meter – Data Sheet
We Listened to the Industry, Then Delivered a Better and Easier Replacement
Solution for the Legacy 432A Analog Power Meter
As today’s power measurements become
more complex, it is increasingly difficult
to make reliable, accurate power measurements. For more than 40 years, you’ve
depended on Keysight’s 432A analog
power meter, used with temperature
compensated thermistor sensors to
provide high accuracy over a wide
temperature range.
Today, the Keysight N432A digital
thermistor power meter, loaded with
enhancements, including a color digital
display, and intuitive front panel interface
will continue support your measurement
needs with greater capability and expanded
functionality. Best of all, you can get these
extras at just a fraction of the price of the
legacy 432A.
LXI Class-C compliant N432A
The N432A is a single-channel, average RF
power measurement meter that supports
thermistor sensors. The N432A has the
capability to measure and display average
power, RF bridge voltages (VRF0 and VRF1),
compensation bridge voltages (VCOMP0 and
VCOMP1), V0, and V1. It is also provided with
a set of features such as zeroing and a
built-in self-test.
Developed using LXI technology, the
N432A is an LXI Class-C compliant
instrument.1 The N432A basic configuration consists of two key sections: bridge
and meter logic. When a compatible
thermistor sensor is connected to the
N432A, the RF and compensation bridge
circuits are formed in the bridge section.
VRF, which is the voltage at the top of the
RF bridge, is responsive to both input RF
power and ambient temperature changes.
On the other hand, VCOMP, which is the
voltage at the top of the compensation
bridge, is responsive only to ambient
temperature changes. The VRF and VCOMP
values are used in calculating the RF
power. Meanwhile, the meter logic section
processes VRF and VCOMP to produce a
meter current proportional to RF power.
1.
In general, the N432A is an easy‑to‑use
instrument, especially with the availability
of an integrated Web browser that provides a convenient way to view and modify
the instrument network configuration.
Besides the LAN interface, the N432A also
supports industry standard GPIB and USB
interfaces for measurement automation.
LXI, an acronym for LAN eXtension for Instrumentation, is an instrument standard for devices that use
the Ethernet (LAN) as their primary communication interface.
04 | Keysight | N432A Thermistor Power Meter – Data Sheet
High accuracy—
no thermoelectric error
The high accuracy ± (0.1% of reading +
0.5 μW) over a wide temperature range
featured in the legacy 432A power meter
is also included in the N432A, making it
excellent for 1 mW transfer calibration
(with 478A-H75/H76*). Accuracy can be
maintained on even the most sensitive
range as the error due to thermoelectric
effect is reduced to a negligible level.
*
Sensor compatibility
The N432A is compatible with the Keysight 478A and 8478B thermistor sensors.
The following table lists the frequency range and operating resistance for these two
sensors:
Thermistor sensor model
Frequency range (GHz)
Operating resistance (Ω)
478A
0.01 to 10
200
478A Option H63
0.000001 to 1
200
478A Option H75
0.001 to 1
200
478A Option H76
0.001 to 1
200
8478B
0.01 to 18
200
478-H75/H76 has a maximum SWR of
1.05 at 50 MHz
For further information on the thermistor sensors, refer to the respective manuals.
05 | Keysight | N432A Thermistor Power Meter – Data Sheet
Take a Closer Look
N432A front panel
High-resolution color LCD display
for easy viewing of test results
Soft keys provide
menu selection
Arrow keys enable navigation of parameter entry
screens, values selection, and entering of text
Display key allows
selection of the display
format for the active
window (single/split
screen)
Display key allows
selection of the display
mode (windowed,
expanded, or fullscreen) of numeric
measurements
Numeric keypad
for easy entry of
numeric values
Power reference
of 1 mW (0 dBm)
Hard keys provide quick access to the
most frequently used functions, such as
System, Trigger, and Acquisition, etc.
Channel sensor connector
N432A back panel
Ground connector
DC recorder outputs (DC voltage corresponds
to the power level of the channel input)
Line power accepts
voltage with automatic
range selection
VRF and VCOMP output
terminals for calibrating
the N432A and for
precision power
measurements
Go beyond GPRB with USB and LAN/LXI-C interface
06 | Keysight | N432A Thermistor Power Meter – Data Sheet
Specifications and Characteristics
Specifications describe the instrument’s
warranted performance and apply after
a 30 minute warm-up.
General specifications
These specifications are valid over its
operating/environmental range unless
otherwise stated and after performing
a zero and calibration procedure.
Compatible power sensors
Supplemental characteristics (shown in
italics) are intended to provide additional
information, useful in applying the instrument by giving typical (expected), but
not warranted performance parameters.
These characteristics are shown in italics
or labeled as “typical,” “nominal” or
“approximate.”
Frequency range
100 kHz to 18 GHz, sensor dependent
Power range
–30 to +10 dBm (1 μW to 10 mW), thermistor-sensor dependent
–– Keysight 478A thermistor sensor (100 kHz to 10 GHz,
with option H63)
–– Keysight 8478B thermistor sensor (10 MHz to 18 GHz)
Meter power accuracy
Power absolute accuracy
± (0.1% of reading + 0.5 μW)
Meter voltage accuracy
(1-year reference specifications)
VRF and VCOMP
V0 and V1
23 °C ± 5 °C: ± (0.0035% + 50 μV) [reading + range]
23 °C ± 5 °C: ± (0.0040% + 25 μV) [reading + range]
Bridge resistance
100, 200, 300, or 400 Ω (user selectable)
Display units
Power
Relative
VRF, VCOMP, V0, and V1
Bridge resistance
Watts (W) or dBm
Percent (%) or dB
VDC and mVDC
Ohm
Display resolution
Power
Default resolution
Voltage
Bridge resistance
Selectable resolution of: 1.0, 0.1, 0.01 and 0.001 dB in logarithmic
mode, or 1, 2, 3, and 4 significant digits in linear mode
0.01 dB in logarithmic mode or three digits in linear mode
6.5-digit resolution
6.5-digit resolution
07 | Keysight | N432A Thermistor Power Meter – Data Sheet
Specifications and Characteristics (continued)
Power meter functions
Accessed by key entry
Either hard keys, or soft key menu, and programmable
Zero
Zeros the meter. (Power reference calibrator is switched off during zeroing)
Frequency
Entered frequency range is used to interpolate the calibration factors table. Frequency range from 1 kHz to 999.9 GHz
Bridge resistance type
Setting the bridge resistance, user-selectable Rmeasure (factory set) or Ruser (measured by user)
Cal factor
Sets the cal factor versus frequency for the calibration factor for power sensor. Range: 1.00% to 150.00%, in 0.01%
increments. Cal factor can be set from either CF table or single entry CF
Relative
Displays all successive measurements relative to the last displayed value
Offset
Allows power measurements to be offset by –100 to +100 dB, settable in 0.001 dB increments, to compensate for
external loss or gain. Offset table can be set from either CF table or single entry CF
Save/recall
Store up to 10 instrument states via the save/recall menu
Measurement averaging
Selectable from 4 to 128
Duty cycle
Duty cycle values between 0.001 to 99.999%, in 0.001% increments, can be entered to display a peak power
representation of measured power. The following equation is used to calculate the displayed peak power value:
peak power = measured power/duty cycle
Limits
High and low limits can be set in the range –150.000 to +230.000 dBm, in 0.001 dBm increments
Preset default values
dBm mode, relative off, power reference off, duty cycle off, offset off, frequency 50 MHz, measurement averaging 16,
free run
Display
Color display with selectable single and split screen formats are available. User selectable on digital measurement
type or analog scale presentation
1 mW (0 dBm) power reference
1.0 mW (0.0 dBm), 50 MHz from type N (f) connector on the front panel, for power meter/sensor function check.
Accuracy (for two years)
± 0.4% (25 ± 10 ºC)
± 1.2% (0 to 45 ºC)
SWR
1.05 (typical),
1.08 (0 °C to 45 °C)
08 | Keysight | N432A Thermistor Power Meter – Data Sheet
Specifications and Characteristics (continued)
Power meter general specifications
Dimensions
The following dimensions exclude
front and rear protrusions:
212.6 mm W x 88.5 mm H x 348.3 mm D
(8.5 in x 3.5 in x 13.7 in)
Weight
Model
N432A
Net
3.6 kg (8.0 lb)
Shipping
8.2 kg (18.1 lb)
Rear panel connectors
VRF output
VRF BNC terminal outputs the RF bridge voltage, used to connect to external DMM for more precise power
measurement
VCOMP output
VCOMP BNC terminal outputs the compensation bridge voltage, used to connect to external DMM for more
precise power measurement
Recorder output
Analog 0 to 1 V, 1 kΩ output impedance, BNC connector
GPIB, USB 2.0 and 10/100BaseT LAN
Interfaces to allow communication with an external controller
Ground
Binding post, accepts 4 mm plug or bare wire connection
Line power
Input voltage range
Input frequency range
Power requirement
100 – 240 VAC, automatic selection
220 – 240 V ±10%
50 – 60 Hz, 400 Hz
400 Hz (100 – 120 Vac)
70 VA
09 | Keysight | N432A Thermistor Power Meter – Data Sheet
Specifications and Characteristics (continued
Environmental characteristics
Electromagnetic compatibility
Complies with the essential requirements of EMC Directive (2004/108/EC) as follows:
–– IEC61326-1:2005 / EN61326-1:2006
–– CISPR11:2003 / EN55011:2007 (Group 1, Class A)
The product also meets the following EMC standards:
–– Canada: ICES/NMB-001:2004
–– Australia/New Zealand: AS/NZS CISPR 11:2004
Product safety
This product conforms to the requirements of the following safety standards:
–– IEC 61010-1:2001 / EN 61010-1:2001
–– CAN/CSA-C22.2 No.61010-1-04
–– ANSI/UL61010-1:2004
Low voltage directive
This product conforms to the requirements of European Council Directive “2006/95/EC”
Operating environment
Temperature
Maximum humidity
Minimum humidity
Maximum altitude
0 °C to 45 °C
95% at 40 °C (non-condensing)
15% at 40 °C (non-condensing)
4,600 m (15,000 ft)
Storage conditions
Non-operating storage temperature
Non-operating maximum humidity
Non-operating maximum altitude
–40 °C to +70 °C
90% at 65 °C (non-condensing)
4,600 m (15,000 ft)
Remote programming
Interface
Command language
GPIB compatibility
GPIB, USB, and LAN interfaces operates to IEEE 488.2 standard
SCPI standard interface commands
SH1, AH1, T6, TE0, L4, LE0, SR1, RL1, PP1, DC1, DT1, C0
10 | Keysight | N432A Thermistor Power Meter – Data Sheet
Ordering Information
Standard product
N432A
Thermistor power meter
Standard shipped accessories
Thermistor sensor adaptor cable 1.5 m* (5 ft)
Power cord (country dependant)
USB adaptor cable
Keysight N432A thermistor power meter product reference CD-ROM
Envelope—calibration certificate
IO libraries media suite
Warranty
Standard 3-year, return-to-Keysight warranty and service plan
3 months for standard shipped accessories
Power sensors, accessories, calibration, and documentation options
*
478A-H13
Frequency range 100 kHz to 1 GHz, with maximum SWR 1.8 to 300 KHz, 1.3 to 1 GHz. Include 13 frequency points
Standard Lab calibration
478A-H55
Frequency range 1 MHz to 1 GHz, with maximum SWR 1.3
478A-H63
Frequency range 100 kHz to 1 GHz, max SWR 1.8 to 300 kHz, 1.3 to 1 GHz
478A-H72
Frequency range 1 MHz to 1 GHz, with maximum SWR 1.2
478A-H73
Frequency range 1 MHz to 100 MHz, with maximum SWR 1.1, except 1.05 at 50 MHz
478A-H75
Frequency range 1 MHz to 1 GHz, max SWR 1.3 except 1.05 at 50 MHz
478A-H76
Frequency range 1 MHz to 1 GHz, max SWR 1.3 except 1.05 at 50 MHz. Standard lab calibration at 50 MHz
478A-H83
Frequency range 1 MHz to 1 GHz, with maximum SWR 1.3, except 1.05 at 50 MHz Include 8 frequency points
Standard Lab calibration
478A-H93
Frequency range 1 MHz to 1 GHz, with maximum SWR 1.3, except 1.05 at 50 MHz Include 2 frequency points
Standard Lab calibration
8478B-H01
Frequency range 10 MHz to 18 GHz, with maximum SWR 1.05 at 50 MHz
8478B-H27
Frequency range 10 MHz to 18 GHz. Include Standard Lab calibration at 9 MHz
N4998A
Thermistor sensor adaptor cable 1.5 m (5 ft)
N4998B
Thermistor sensor adaptor cable 3 m (10 ft)
N4998C
Thermistor sensor adaptor cable 6.1 m (20 ft)
N432A-908
Rackmount kit with 1 unit with blank filler included
N432A-909
Rackmount kit with 2 units side-by-side
N432A-OBK
English language user guide and English programming guide
N432A-ABJ
Japanese user guide and English programming guide
N432A-0B1
English language user guide and installation guide
N432A-0BF
English language programming guide
N432A-A6J
Certificate of compliance calibration - ANSI/NCSL Z540
N432A-1A7
Compliant calibration test data - ISO17025
The 1.5 m (5 ft) standard cable can be replaced with a 3 or 6.1 m (10 or 20 ft) cable, charges apply
11 | Keysight | N432A Thermistor Power Meter – Data Sheet
myKeysight
www.keysight.com/find/mykeysight
A personalized view into the information most relevant to you.
www.axiestandard.org
AdvancedTCA® Extensions for Instrumentation and Test (AXIe) is an
open standard that extends the AdvancedTCA for general purpose and
semiconductor test. Keysight is a founding member of the AXIe consortium.
ATCA®, AdvancedTCA®, and the ATCA logo are registered US trademarks of
the PCI Industrial Computer Manufacturers Group.
www.lxistandard.org
LAN eXtensions for Instruments puts the power of Ethernet and the
Web inside your test systems. Keysight is a founding member of the LXI
consortium.
www.pxisa.org
PCI eXtensions for Instrumentation (PXI) modular instrumentation delivers a
rugged, PC-based high-performance measurement and automation system.
Three-Year Warranty
www.keysight.com/find/ThreeYearWarranty
Keysight’s commitment to superior product quality and lower total cost
of ownership. The only test and measurement company with three-year
warranty standard on all instruments, worldwide.
Keysight Assurance Plans
www.keysight.com/find/AssurancePlans
Up to five years of protection and no budgetary surprises to ensure your
instruments are operating to specification so you can rely on accurate
measurements.
www.keysight.com/go/quality
Keysight Technologies, Inc.
DEKRA Certified ISO 9001:2008
Quality Management System
Keysight Channel Partners
www.keysight.com/find/channelpartners
Get the best of both worlds: Keysight’s measurement expertise and product
breadth, combined with channel partner convenience.
www.keysight.com/find/N432A
For more information on Keysight
Technologies’ products, applications or
services, please contact your local Keysight
office. The complete list is available at:
www.keysight.com/find/contactus
Americas
Canada
Brazil
Mexico
United States
(877) 894 4414
55 11 3351 7010
001 800 254 2440
(800) 829 4444
Asia Pacific
Australia
China
Hong Kong
India
Japan
Korea
Malaysia
Singapore
Taiwan
Other AP Countries
1 800 629 485
800 810 0189
800 938 693
1 800 112 929
0120 (421) 345
080 769 0800
1 800 888 848
1 800 375 8100
0800 047 866
(65) 6375 8100
Europe & Middle East
Austria
Belgium
Finland
France
Germany
Ireland
Israel
Italy
Luxembourg
Netherlands
Russia
Spain
Sweden
Switzerland
United Kingdom
0800 001122
0800 58580
0800 523252
0805 980333
0800 6270999
1800 832700
1 809 343051
800 599100
+32 800 58580
0800 0233200
8800 5009286
0800 000154
0200 882255
0800 805353
Opt. 1 (DE)
Opt. 2 (FR)
Opt. 3 (IT)
0800 0260637
For other unlisted countries:
www.keysight.com/find/contactus
(BP-09-04-14)
This information is subject to change without notice.
© Keysight Technologies, 2014
Published in USA, November 21, 2014
5990-5740EN
www.keysight.com