NI NIPXI-5124 150 mhz, 200 ms/s, 12-bit digitizer Datasheet

150 MHz, 200 MS/s, 12-Bit Digitizer
NI PXI-5124
• 2 channels simultaneously sampled
at 12-bit resolution
• 200 MS/s real-time and 4.0 GS/s
random interleaved sampling
• 150 MHz bandwidth
• 200 mV to 20 V input range
• >75 dBc SFDR
• 8, 32, 256, or 512 MB of memory
per channel
• Edge, window, hysteresis, video,
and digital triggering with
50 ps timestamping
Operating Systems
• Windows 2000/NT/XP
Recommended Software
• LabVIEW
• LabWindows/CVI
• Measurement Studio
Driver Software (included)
• NI-SCOPE driver
• LabVIEW Express VIs
• SCOPE Soft Front Panel
• NI Spectral Measurements Toolkit
(with 32, 256 and 512 MB models)
Calibration
• Gain, offset, frequency response
and timing self-calibration
• 2 year external calibration cycle
Overview
Triggering, Clocking, and Synchronization
The National Instruments PXI-5124 high-resolution digitizer
features two 200 MS/s simultaneously sampled inputs channels with
12-bit resolution, 150 MHz bandwidth, and up to 512 MB of
memory per channel in a compact 3U PXI module. The NI PXI-5124
uses the high-speed PCI bus and the scatter-gather bus mastering of
the NI MITE ASIC to move data to the computer at speeds up to 100
times faster than traditional instrument interfaces, thereby
dramatically decreasing overall test time. With the Synchronization
and Memory Core (SMC) architecture of the PXI-5124, you can
create mixed-signal systems using signal generators and digital
waveform generator/analyzers or build a high-channel-count
digitizer with subnanosecond synchronization between channels.
• Edge, window, hysteresis, video, digital, triggering with
50 ps timestamping
• Pre and posttrigger acquisition in single and multiple-record mode
• Internal 200 MHz clock or external clock from 50 to 210 MHz
• Phase lock to PXI 10 MHz reference or external reference from
1 to 20 MHz
Dual 200 MS/s, 12-Bit Input Channels
for Time and Frequency Analysis
• 150 MHz input bandwidth with anti-alias and noise filters
• >75 dBc spurious free dynamic range (SFDR)
• 4.0 GS/s equivalent time sampling for repetitive signals
• Independent channel selectable 200 mVpp to 20 Vpp input ranges
• Independent channel selectable 50 Ω or 1 MΩ input impedance
• 2-year calibration cycle and 0 to 55 ˚C operating temperature
Software
• IVI-compliant NI-SCOPE driver for
LabVIEW, LabWindows/CVI, and
Microsoft C++ and Visual Basic with
more than 50 built-in measurements
• SCOPE Soft Front Panel
for interactive control
• Spectral Measurements Toolkit
for sophisticated frequency-domain
measurements in communications,
signal intelligence, and
avionics applications
Applications
Communications
xDSL
Wireless communications
Baseband I & Q
Consumer Electronics
DVD, DVD-R, and PVR
Set top box
Gaming console
Biomedical and Scientific Research
Ultrasonic medical imaging
Mass spectrometry
Particle physics
Aerospace/Defense
Emulation of IC communications
Consumer Electronics
RADAR, SONAR, and LIDAR
Satellite
Signal intelligence
Ordering Information
Deep Onboard Memory
• 8, 32, 256, or 512 MB of memory per channel
• Capture more than 1 million triggered waveforms with multiple
record hardware rearm
• Stream data continuously from onboard memory to host memory
or disk
NI PXI-5124 ..............................................................778757-0M1
Includes NI-SCOPE driver and SCOPE Soft Front Panel. 32, 256, and 512 MB
models include NI Spectral Measurements Toolkit.
1M (memory per channel): 1 (8 MB), 2 (32 MB), (256 MB), 4 (512 MB)
Recommended PXI Switch
NI PXI-2593 ..................................................................778793-01
150 MHz, 200 MS/s, 12-Bit Digitizer
Specifications
These specifications are valid for 0 to 55 ˚C, unless otherwise stated.
RMS Noise (20 MHz filter enabled)
Acquisition System
Number of channels......................................... 2 simultaneously sampled
Resolution......................................................... 12 bits
Bandwidth (-3 dB)
Full Scale Input Range
50 Ω
400 mV, 1 V, 2 V, 5 V, 10 V, 20 V
150 MHz
200 mV
85 MHz
1 MΩ
145 MHz1
75 MHz
Bandwidth limit filters
(software selectable) ....................................... 20 MHz noise (2-pole Bessel)
60 MHz antialias (4-pole elliptical)
Maximum sampling rate .................................. 200 MS/s real-time, 4 GS/s random interleave sampling
Onboard sample memory ................................. 8, 32, 256, or 512 MB per channel
(4, 16, 128, 256 million samples)
Pre and posttrigger data points2 ...................... 0 to 100% of full record length
Memory per
channel (MB)
8
32
256
512
Multiple records acquisition
(0 to 100% pre and posttrigger data)
Maximum number
of records
32,768
131,072
1,048,576
2,097,152
Input impedance............................................... 50 Ω and 1 M Ω || 25 pF, software selectable
Full-scale input range....................................... 50 Ω: 200 mV, 400 mV, 1 V, 2 V, 4 V, 10 V
1 MΩ: 200 mV, 400 mV, 1 V, 2 V, 4 V, 10 V, 20 V
Vertical offset ranges....................................... ±50% of full scale input range
Maximum input overload ................................. 50 Ω: 7 Vrms with peaks ≤ 10 V
1 MΩ: peaks ≤ 42 V
Input coupling................................................... AC, DC, GND (AC coupling on 1 MΩ only)
AC coupling cutoff frequency (-3 dB)............... 12 Hz
Accuracy
DC accuracy (0 V offset setting)
50 Ω and 1 MΩ
±0.65% of Input ±1.3 mV
±0.65% of Input ±1.5 mV
±0.65% of Input ±10.0 mV
Full Scale Input Range
200 mV, 400 mV
1 V, 2 V
4 V, 10 V, 20 V
50 Ω
94 µVrms
188 µVrms
470 µVrms
940 µVrms
1.88 mVrms
4.7 mVrms
–
Full Scale Input Range
200 mV
400 mV
1V
2V
4V
10 V
20 V (1 MΩ only)
1 MΩ
104 µVrms
192 µVrms
480 µVrms
960 µVrms
1.92 mVrms
4.8 mVrms
9.4 mVrms
Intermodulation distortion3 (IMD) .................... -75 dBc
Phase noise density (10 MHz input) ................ <-100 dBc/Hz at 100 Hz
<-120 dBc/Hz at 1 kHz
<-130 dBc/Hz at 10 kHz
Acquisition Modes
Real-time sampling rate................................... 200 MS/s to 3.052 kS/s sample rate
Random interleave sampling (RIS)................... 4 GS/s to 400 MS/s sample rate (repetitive signals only)
Timebase System
Total sample clock jitter4 ................................. ≤1 psrms
Internal
Internal sample clock frequency ...................... 200 MS/s sampling rate with
decimation by n where 1≤ n ≤ 65,535
Timebase accuracy ........................................... ±25 ppm (±0.0025%)
External
External clock sources...................................... CLK IN (SMB connector), PXI star
External clock range......................................... 50 MHz to 210 MHz, variable with
decimation by n where 1≤ n ≤ 65,535
External reference sources .............................. CLK IN (SMB connector), PXI backplane 10 MHz
External reference range.................................. 1 to 20 MHz in 1 MHz increments
External clock/reference amplitude................. Sine wave: 0.65 to 2.8 Vpp (0 to 13 dBm)
Square wave: 0.2 to 2.8 Vpp
External clock/reference impedance ............... 50 Ω, AC coupled
Trigger System
AC amplitude accuracy (50 kHz) ...................... 50 Ω: ±0.06 dB
1 MΩ: ±0.09 dB
Channel-to-channel crosstalk .......................... ≤-85 dB at 10 MHz
Edge, Hysteresis, Window, Video, Digital, Immediate, Software
CH 0, CH 1, TRIG, PXI_Trig <0:6>, PFI <0:1>, PXI Star, Software
Rising or falling
Fully programmable
Negative sync of NTSC, PAL, and SECAM standards
Any line, specific line, specific field
50 kHz software selectable
50 kHz software selectable
CH0 and CH1: 2.5% FS up to 50 MHz
decreasing to 10% FS at 150 MHz
TRIG: 2.5% up to 100 MHz decreasing to 10% at 200 MHz
Level accuracy .................................................. CH0, CH1: ±4.7% FS up to 10 MHz
TRIG: ±3.5% FS up to 10 MHz
Time resolution................................................. 50 ps with time-to-digital converter enabled
Holdoff5 ............................................................ 2 µs to 171.79 s, software selectable
Spectral Characteristics (typical)
External Trigger Channel (TRIG)
Dynamic Performance (50 Ω input impedance with 10 MHz, -1 dBFS input signal)
Impedance ........................................................ 1 MΩ || 22 pF
Vertical Range .................................................. ±5 V
Coupling............................................................ AC, DC
Passband flatness (referenced at 50 kHz)
Full-scale input range
Filters Off
400 mV, 1 V, 2 V, 5 V, 10 V, 20 V
200 mV
Antialias Filter On
Full Scale Input Range
200 mV
400 mV
1V
2V
4V
10 V
All ranges
SFDR
75 dB
75 dB
72 dB
72 dB
65 dB
65 dB
50 Ω and 1 MΩ
±0.5 dB, DC to 20 MHz
±1 dB, 20 MHz to 50 MHz
±1.7 dB, 50 MHz to 100 MHz
±0.6 dB, DC to 20 MHz
±1.5 dB, 20 MHz to 40 MHz
-1 dB to +2 dB, DC to 55 MHz
THD
-74 dBc
-74 dBc
-72 dBc
-72 dBc
-63 dBc
-63 dBc
SNR
57 dB
58 dB
58 dB
58 dB
–
–
SINAD
57 dB
58 dB
58 dB
58 dB
–
–
Modes...............................................................
Source...............................................................
Slope.................................................................
Hysteresis.........................................................
Video trigger.....................................................
Video trigger types ...........................................
High frequency reject filter ..............................
Low frequency reject filter...............................
Sensitivity.........................................................
Power Requirements (typical)
+3.3 VDC
1.3 A
+5 VDC
1.7 A
+12 VDC
130 mA
-12 VDC
270 mA
Total Power
17.6 W
Environment
Dynamic Performance (1 MΩ input impedance with 10 MHz, -1 dBFS input signal)
Full Scale Input Range
200 mV
400 mV
1V
2V
4V
10 V
20 V
SFDR
70 dB
70 dB
70 dB
70 dB
67 dB
60 dB
60 dB
THD
-68 dBc
-68 dBc
-68 dBc
-67 dBc
-66 dBc
-58 dBc
-58 dBc
SFDR = Spurious-free dynamic range
THD = Total harmonic distortion
SNR = Signal-to-noise ratio, excluding distortion (antialias filter enabled)
SINAD = Signal-to-noise and distortion (antialias filter enabled)
2
SNR
53 dB
55 dB
57 dB
57 dB
56 dB
–
–
SINAD
53 dB
55 dB
57 dB
57 dB
56 dB
–
–
Operating temperature6 ................................... 0 to 55 ºC (Meets IEC-60068-2-1 and IEC-60068-2-2)
Storage temperature ........................................ -40 to 71 ºC (meets IEC-60068-2-1 and 60068-2-2)
Relative humidity ............................................. 10 to 90 %, noncondensing (meets 60068-2-56)
Calibration
Self-calibration................................................. Gain, offset, frequency response,
triggering, and timing for all input ranges
External calibration interval............................. 2 years
Certification and Compliances
CE Mark compliance
For detailed specifications on power, environmental, safety, and physical dimensions see PXI-5124
detailed specifications.
1.Bandwidth on the 1 MΩ input is 145 MHz for 0 to 40 ºC and 135 MHz for 40 to 55 ºC
2. NI PXI-5124 is also capable of continous acquisition
3. Measured on ranges up to 2 V on 50 Ω input with two tones at 10.2 MHz and 11.2 MHz, each at -7 dBFS
4. Includes effects of converter aperture and clock circuitry jitter from 100 Hz to 100 kHz
5. Time-digital converter disabled
6. 0 to 45 ºC in PXI-101x and 1000/B chassis
National Instruments • Tel: (800) 433-3488 • Fax: (512) 683-9300 • [email protected] • ni.com
NI Services and Support
NI has the services and support to meet your
needs around the globe and through the
application life cycle – from planning
and development through deployment
and ongoing maintenance. We offer
services and service levels to meet
customer requirements in research,
design, validation, and manufacturing.
Visit ni.com/services.
Local Sales and Technical Support
In offices worldwide, our staff is local to the country, giving you
access to engineers who speak your language. NI delivers industryleading technical support through online knowledge bases, our
applications engineers, and access to 14,000 measurement and
automation professionals within NI Developer Exchange forums.
Find immediate answers to your questions at ni.com/support.
We also offer service programs that provide automatic upgrades to
your application development environment and higher levels of
technical support. Visit ni.com/ssp.
Training and Certification
NI training is the fastest, most certain route to productivity with our
products. NI training can shorten your learning curve, save
development time, and reduce maintenance costs over the
application life cycle. We schedule instructor-led courses in cities
worldwide, or we can hold a course at your facility. We also offer a
professional certification program that identifies individuals who
have high levels of skill and knowledge on using NI products.
Visit ni.com/training.
Hardware Services
NI Factory Installation Services
NI Factory Installation Services (FIS) is the fastest and easiest way to
use your PXI or PXI/SCXI™ combination systems right out of the
box. Trained NI technicians install the software and hardware and
configure the system to your specifications. NI extends the standard
warranty by one year on hardware components (controllers, chassis,
modules) purchased with FIS. To use FIS, simply configure your
system online with ni.com/pxiadvisor.
Professional Services
Our Professional Services Team is comprised of NI applications
engineers, NI Consulting Services, and a worldwide NI Alliance
Partner Program of more than 600 independent consultants and
integrators. Services range
from start-up assistance to
turnkey system integration.
Visit ni.com/alliance.
Calibration Services
NI recognizes the need to maintain properly calibrated devices for
high-accuracy measurements. We provide manual calibration
procedures, services to recalibrate your products, and automated
calibration software specifically designed for use by metrology
laboratories. Visit ni.com/calibration.
Repair and Extended Warranty
OEM Support
We offer design-in consulting and product integration assistance
if you want to use our products for OEM applications. For
information about special pricing and services for OEM customers,
visit ni.com/oem.
NI provides complete repair services for our products. Express repair
and advance replacement services are also available. We offer
extended warranties to help you meet project life-cycle requirements.
Visit ni.com/services.
ni.com • (800) 433-3488
National Instruments • Tel: (512) 683-0100 • Fax: (512) 683-9300 • [email protected]
© 2004 National Instruments Corporation. All rights reserved. LabVIEW, LabWindows/CVI, Measurement Studio and ni.com are trademarks
of National Instruments. Other product and company names listed are trademarks or trade names of their respective companies.
Similar pages