Keysight E1411B 5.5-digit multimeter, c-size Datasheet

Agilent E1411B
5.5-Digit Multimeter, C-Size
Data Sheet
• 1-Slot, C-size, register based
• DCV, ACV, 2- & 4-wire Ω, temperature
• 5.5-digit low-noise integrating A/D
• 13 kHz high-speed sampling A/D
• Balanced differential isolated inputs
• Software calibration
Description
The Agilent Technologies E1411B 5.5-Digit Multimeter is a
C-size,1-slot, register-based VXI module. It is identical in electrical design to the E1326B, differing only in size. You can use the
integrating A/D to make 5.5-digit, low-noise measurements,
or switch to the sampling A/D to make 14-bit readings at rates
up to 13 kHz.
When combined with any Agilent VXI relay or FET multiplexer,
you can create a multichannel scanning multimeter. For example,
by sending just one SCPI command to the E1406A, you can
program the multimeter and the channels of your multiplexers
all at one time. The E1411B provides flexible triggering with
built-in timer pacer, also.
Product functions for this DMM include Vdc/ac, 2- and 4-wire
Ω, offset-compensated Ω, thermocouples thermistors,and
RTDs. This autoranging DMM is especially well suited for data
acquisition and computer-aided test applications.
Refer to the Agilent Technologies Website for instrument driver
availability and downloading instructions, as well as for recent
product updates, if applicable.
Product Specifications
Reading Rate
Auto zero off, fixed range, default trigger delay, offset comp off,
Sample Source “TIMER” for rates >15 readings/s.
Maximum reading rate: 13 K
Typical Reading Rates (rdgs/s)
APERTURE
320 ms 267 ms 20 ms 16.7 ms 2.5 ms 100 µs 10 µs
dc voltage: 3
Four-wire
resistance: 3
ac voltage: 1.3
3.5
49
59
365
3125 13000
3.5
1.4
49
1.9
59
1.9
365
1.9
3125 13000
1.9
1.9
Resolution (bits/digit)
APERTURE
320 ms 267 ms 20 ms 16.7 ms 2.5 ms 100 µs 10 µs
Binary bits: ±22
Decimal
digits:
6.5
±22
±20
±20
±18
±15
±14
6.5
6
6
5.5
4.5
4
Noise Rejection (dB) for DC voltage and Resistance functions.
Noise Rejection Conditions: CMR measured with 1 kΩ in both High and Low leads with a 10% imbalance. Low connted to Common at source,
measured with respect to earth ground. NMR is for specified frequencies ±0.1%
APERTURE
320 ms
267 ms
20 ms
16.7 ms
2.5 ms
100 µs
10 µs
dc:
Common mode rejection
ac (0-400): Common mode rejection*
150 dB
70 dB
150 dB
70 dB
150 dB
70 dB
150 dB
70 dB
150 dB
70 dB
150 dB
70 dB
150 dB
70 dB
50 Hz:
Power line cycles (NPLC)
Normal mode rejection
Effective Common Mode ECMR*
16
84 dB
154 dB
–
0 dB
70 dB
1
60 dB
130 dB
–
0 dB
70 dB
–
0 dB
70 dB
–
0 dB
70 dB
–
0 dB
70 dB
60 Hz:
Power line cycles (NPLCs)
–
Normal mode (60 Hz) rejection
0 dB
Effective Common Mode Rejection* 70 dB
16
84 dB
154 dB
–
0 dB
70 dB
1
60 dB
130 dB
–
0 dB
70 dB
–
0 dB
70 dB
–
0 dB
70 dB
400 Hz:
Power line cycles (NPLCs)
Normal mode (400 Hz) rejection
Effective Common Mode ECMR*
–
0 dB
70 dB
8
84 dB
154 dB
–
0 dB
70 dB
1
60 dB
130 dB
–
0 dB
70 dB
–
0 dB
70 dB
* 64 & 300 volt ranges reduced by
128
84 dB
154 dB
36 dB
dc Voltage
Accuracy Conditions: Auto zero on,one hour warmup. Temperature within ±5°C of calibration temperature (module calibrated at 18-28°C).
RESOLUTION vs APERTURE
(Volts)
Range
Input Resistance
125 mV
1V
8V
64 V
300 V
>100 MΩ
>100 MΩ
>100 MΩ
10 MΩ + 5%
10 MΩ + 5%
2
90-DAY ACCURACY vs APERTURE
± (% of Reading + Volts)
20/16.7 ms
10 µs
20/16.7 ms
10 µs
120 nV
1.0 µV
7.6 µV
61 µV
488 µV
7.6 µV
61 µV
488 µV
3.9 mV
31 mV
0.023% + 5 µV
0.013% + 15 µV
0.01% + 50 µV
0.015% + 1 mV
0.015% + 5 mV
0.115% + 60 µV
0.1% + 200 µV
0.1% + 1.5 µV
0.1% + 20 µV
0.1% + 80 µV
Four-Wire Resistance
Accuracy Conditions: Auto zero on, one hour warmup. Temperature within ±5°C of calibration temperature (module calibrated at 18-28°C).
Range
256 Ω
2 kΩ
16 kΩ
131 kΩ
1 MΩ
Source
Current
488 µA
488 µA
61 µA
61 µA
7.6 µA
Maximum Open
Circuit Voltage
11.5 V
11.5 V
11.5 V
11.5 V
11.5 V
RESOLUTION vs APERTURE
(Ω)
20/16.7 ms
250 µΩ
2 mΩ
15 mΩ
125 mΩ
1Ω
10 µs
15 mΩ
125 mΩ
1Ω
8Ω
64 Ω
90-DAY ACCURACY vs APERTURE
± (% of Reading + Ω)
20/16.7 ms
0.035% + 10 mΩ
0.025% + 20 mΩ
0.025% + 200 mΩ
0.025% + 1 Ω
0.015% + 10 Ω
10 µs
0.12% + 50 mΩ
0.1% + 200 mΩ
0.1% + 2 Ω
0.1% + 16 Ω
0.1% + 120 Ω
NOTE: With offset compensation on, accuracy is the same as for the voltmeter alone.
True RMS ac Voltage (ac coupled)
Crest Factor: 7 at 0% full scale; 5 at full scale. Accuracy Conditions: Sine wave inputs >0% of full scale. dc component <0% of ac component.
Auto-zero on, one hour warmup. Temperature within ± 5°C of calibration temperature (module calibrated at 8-28°C).
Range
(RMS)
87.5 mV
Input
Impedance
>100 MΩ,
<100 pF
700 mV
>100 MΩ,
<100 pF
5.6 V
>100 MΩ,
<100 pF
44.8 V
10 MΩ ± 5%,
<100 pF
300 V
10 MΩ ± 5%,
<100 pF
RESOLUTION vs APERTURE
(Volts)
Frequency
20-50 Hz
50 Hz-1 kHz
5-10 kHz
20-50 Hz
50 Hz-1 kHz
1-5 kHz
5-10 kHz
20-50 Hz
50 Hz-1 kHz
1-5 kHz
5-10 kHz
20-50 Hz
50 Hz-1 kHz
1-5 kHz
5-10 kHz
20-50 Hz
50 Hz-1 kHz
1-5 kHz
5-10 kHz
320/267 ms
30 nV
10 µs
7.6 µV
0.24 µV
61 µV
2.0 µV
488 µV
15 µV
3.9 mV
122 µV
31 mV
Timing/Synchronization
Timer/pacer
Timer range:
Resolution:
Programmable delay
Delay range:
Resolution:
External trigger
Minimum pulse width:
Maximum trigger rate:
Isolation
90-DAY ACCURACY vs APERTURE
± (% of Reading + Volts)
320/267 ms
All other apertures
2.175% +200 µV
2.175% +1 mV
0.675% +200 µV
0.675% +200 µV
3.175% +200 µV
3.175% +200 µV
2.125% +1.5 mV
2.125% +8 mV
0.625% +1.5 mV
0.625% +1.5 mV
0.625% +1.5 mV
0.625% +1.5 mV
3.125% +1.5 mV
3.125% +1.5 mV
2.125% +15 mV
2.125% +80 mV
0.625% +15 mV
0.625% +15 mV
1.125% +15 mV
1.125% +15 mV
10.125% +15 mV 10.125% +15 mV
2.125% +100 mV 2.125% + 500 mV
0.625% +100 mV 0.625% +100 mV
1.125% +100 mV 1.125% +100 mV
10.125% +100 mV 10.125% +100 mV
2.125% + 500 mV
2.125% + 2.5 V
0.625% + 500 mV 0.625% + 500 mV
1.125% + 500 mV 1.125% + 500 mV
10.125% + 500 mV 10.125% + 500 mV
dc Voltage Accuracy with Relay Multiplexers
76 µs to 65.5 ms
2 µs
40 µs to 16 s
2 µs
100 ns
5 kHz (Trigger Condition, negative
edge; Fixed range,10 µs aperture)
Range
125 mV
1V
8V
64 V
300 V
20/16.7 ms
0.023%+9 µV
0.013%+19 µV
0.01%+54 µV
0.015%+1 mV
0.015%+5 mV
10 µs
20/16.7 ms
0.115%+64 µV 0.023%+55 µV
0.1%+204 µV 0.013%+65 µV
0.1%+1.5 mV 0.01%+100 µV
0.1%+20 mV 0.015%+1.05 mV
0.1%+80 mV 0.015%+5.05 mV
10 µs
0.115%+110 µV
0.1%+250 µV
0.1%+1.55 mV
0.1%+20 mV
0.1%+80 mV
Accuracy Conditions: Auto zero on, one hour warmup.
Temperature within ±5°C of calibration temperature
(module calibrated at 8-28°C).
450 Vpk between any terminal
and chassis.
3
True RMS ac Voltage (ac coupled)with Relay Multiplexers
Four-Wire Resistance with Relay Multiplexers
1-5 kHz and 5-10 kHz frequencies (all apertures) when using Relay
Multiplexers (E1343A, E1345A, E1346A, or E1347A).
Add 0.2% to the ac Voltage specifications.
Accuracy Conditions: Auto zero on, one hour warmup, temperature
within ±5°C of calibration temperature (module calibrated at 18-28°C).
NOTE: With offset compensation on, accuracy is the same as for
the voltmeter alone.
Strain Measurements with Strain Relay Multiplexers
90-DAY ACCURACY vs APERTURE
±(% of reading + Ω)
E1326B & E1345A/47A
All measurements are made using the MEAS command.
NOTE: The Agilent E1406A command module and embedded controllers
Range
20/16.7 ms
10 µs
provide units conversion; if the E1411B is register programmed, your program
must make the units conversion.
256 Ω
2 kΩ
16 kΩ
131 kΩ
1 MΩ
0.035%+18.2 mΩ
0.025%+28.2 mΩ
0.025%+266 m Ω
0.025%+1.1 Ω
0.025%+10.5 Ω
0.12%+58.2 mΩ
0.1%+208 m Ω
0.1%+2.1 Ω
0.1%+16.1 Ω
0.1%+121 Ω
Vs = 5 V Power Supply/Gage Factor = 2
18-20°C
Relays
FETs
Quarter
Half
Full
Quarter
Half
Full
µe%
25.8
33.2
16
353
169
83.8
%e
0.023
0.023
0.023
0.023
0.023
0.023
Temp. Coefficient
µe%
1.96
0.23
0.053
103
50.7
25.3
%e
0.006
0.006
0.006
0.006
0.006
0.006
Vs = 1 V Power Supply/Gage Factor = 2
18-20°C
Relays
FETs
Quarter
Half
Full
Quarter
Half
Full
µe%
25.8
5.39
2.07
52.9
18.9
8.85
%e
0.023
0.023
0.023
0.023
0.023
0.023
Temp. Coefficient
µe%
1.96
0.23
0.053
12
5.27
2.57
%e
0.006
0.006
0.006
0.006
0.006
0.006
Vs = 0.1 V Power Supply/Gage Factor = 2
18-20°C
Relays
FETs
4
Quarter
Half
Full
Quarter
Half
Full
µe%
81.3
33.2
16
353
169
83.8
%e
0.023
0.023
0.023
0.023
0.023
0.023
Temp. Coefficient
µe%
1.96
0.23
0.053
103
50.7
25.3
%e
0.006
0.006
0.006
0.006
0.006
0.006
NOTE: Accuracy data includes all errors contributed by the multimeter, analog
bus ribbon cables, multiplexer, and transducer linearization (if applicable). The
accuracies do not include transducer accuracy errors.
Temperature
The temperature accuracy graphs (below) include instrument and firmware
linearization errors. The linearization algorithm used is based on the
ITS-90 standard transducer curves. Add your transducer accuracy
to determine total measurement error.
NOTE: The E1406A command modules and Agilent embedded VXI controllers
provide units conversion; if the E1411B is register-programmed, your program
must make the necessary units conversion.
Thermocouple: (E1411B Multimeters and E1347A / E1476A TC MUX)
16 ms aperture (1 PLC):
100 µs aperture:
Thermistors: (E1411B Multimeters and E1345A/E1347A/E1476A MUXs)
4-wire Ω:
16 ms aperture (1 PLC):
100 µs aperture:
5
RTDs: (E1411B Multimeters and E1345A / E1476A MUXs)
4-wire Ω:
16 ms aperture (1 PLC):
100 µs aperture:
Functions
Idc:
Iac:
Frequency:
Period:
Temp.:
Module Current
—
—
—
—
Tm Tc RTD
Ipm
0.2
0.55
0
0
0
0
0
General Specifications
+5 V:
+12 V:
-12 V:
+24 V:
-24 V:
-5.2 V:
-2 V:
VXI Characteristics
Cooling Slot
VXI device type:
Data transfer bus:
Size:
Slots:
Connectors:
Shared memory:
Watts/slot:
8.50
∆P mm H2O:
0.14
Air Flow liter/s:
0.71
VXI buses:
Register based
Not specified
C
1
P1/2
Yes, shared memory available
with E1406A SCPI driver
TTL Trigger Bus
Instrument Drivers
Command module firmware:
Command module firmware rev:
I-SCPI Win 3.1:
I-SCPI Series 700:
C-SCPI LynxOS:
C-SCPI Series 700:
Panel Drivers:
VXIplug&play Win Framework:
VXIplug&play Win 95/NT Framework:
VXIplug&play HP-UX Framework:
6
Ordering Information
Description
See the Agilent Technologies Website
(http://www.agilent.com/find/inst_drivers)
for driver availability and downloading
Downloadable
A.02
Yes
Yes
Yes
Yes
Yes
Yes
Yes
No
Idm
0.01
0.01
0
0
0
0
0
Product Number
5.5-Digit Multimeter, High-Accuracy, C-Size E1411B
Service Manual
E1411B 0B3
Japan - Japanese Localization
E1411B ABJ
ANSI Z540 Compliant Calibration
E1411B A6J
Agilent Technologies’ Test and Measurement Support,
Services, and Assistance
By internet, phone, or fax, get assistance with all
your test & measurement needs.
Agilent Technologies aims to maximize the value you receive, while minimizing your risk and problems. We
strive to ensure that you get the test and measurement capabilities you paid for and obtain the support you
need. Our extensive support resources and services can help you choose the right Agilent products for
your applications and apply them successfully. Every instrument and system we sell has a global warranty.
Support is available for at least five years beyond the production life of the product. Two concepts underlie
Agilent’s overall support policy: "Our Promise" and "Your Advantage."
Online assistance:
www.agilent.com/find/assist
Our Promise
"Our Promise" means your Agilent test and measurement equipment will meet its advertised performance
and functionality. When you’re choosing new equipment, we will help you with product information,
including realistic performance specifications and practical recommendations from experienced test engineers. When you use Agilent equipment, we can verify that it works properly, help with product operation,
and provide basic measurement assistance for the use of specified capabilities, at no extra cost upon
request. Many self-help tools are available.
Your Advantage
"Your Advantage" means that Agilent offers a wide range of additional expert test and measurement services, which you can purchase according to your unique technical and business needs. Solve problems efficiently and gain a competitive edge by contracting with us for calibration, extra-cost upgrades, out-of-warranty repairs, and on-site education and training, as well as design, system integration, project management, and other professional services. Experienced Agilent engineers and technicians worldwide can help
you maximize your productivity, optimize the return on investment of your Agilent instruments and systems, and obtain dependable measurement accuracy for the life of those products.
Phone or Fax
United States:
(tel) 1 800 829 4444
Canada:
(tel) 1 877 894 4414
(fax) (905) 282 6495
China:
(tel) 800 810 0189
(fax) 800 820 2816
Europe:
(tel) (31 20) 547 2323
(fax) (31 20) 547 2390
Japan:
(tel) (81) 426 56 7832
(fax) (81) 426 56 7840
Korea:
(tel) (82 2) 2004 5004
(fax) (82 2) 2004 5115
Latin America:
(tel) (305) 269 7500
(fax) (305) 269 7599
Taiwan:
(tel) 0800 047 866
(fax) 0800 286 331
www.agilent.com/find/emailupdates
Get the latest information on the products and applications you select.
For more assistance with your test & measurement needs or
to find your local Agilent office go to:
www.agilent.com/find/assist
Other Asia Pacific Countries:
(tel) (65) 6375 8100
(fax) (65) 6836 0252
(e-mail) [email protected]
Data Subject to Change
© Agilent Technologies, Inc. 2004
Printed in the U.S.A. May 1, 2004
5965-5562E