GSM15 Medical 15 Watt Global Performance Medical Switcher GLOBAL PERFORMANCE SWITCHERS FEATURES: DC Power Supplies • Industry’s smallest 15 W medically approved switcher • Compact size (3.00” x 2.10” x 0.92”) • Wide-range ac input: 90-264 Vac • Less than 75 μA leakage current @ 120 Vac • Approved to UL2601-1, EN60601-1 • EMI to FCC, CISPR 11 Class B • Overvoltage protection standard • RoHS compliant models available (G suffix) • marked to LVD SPECIFICATIONS Ac Input 90-264 Vac, 47-63 Hz single phase. Class I or class II grounding. Temperature Coefficient 0.03% / °C typical. Input Current Maximum input current at 90 Vac, 60 Hz with full rated output load not to exceed 0.6 A. EMI/EM Compliance All models include built-in EMI filtering to meet the following EMC requirements of IEC601-1-2. Input Protection Internal ac fuse provided on all units. Designed to blow only if a catastrophic failure occurs in the unit -- Fuse does not blow on unsustained overload or short circuit. Performance Requirement EMC Standard Typical Margin Conducted Emissions EN55011, Class B; FCC Class B 2 dB Class II Gnd 6 dB Class I Gnd Surge Discharge EN61000-4-2, Level 3 2 kV RF Field Susceptibility EN61000-4-3, Level 3 2V Fast Transients/Bursts EN61000-4-4, Level 3 500 V Efficiency 69-85% depending on model. Surge Susceptibility EN61000-4-5, Level 3 500 V Overload Protection Fully protected against short circuit and output overload. Short circuit protection is cycling type power limit. Factory set to begin power limiting at 23 W. Conducted RF Susceptibility EN61000-4-6 25% Voltage Sags & Surges EN61000-4-11 5% Inrush Current Inrush is limited by internal thermistors. The inrush at 240 Vac, averaged over the first ac half-cycle under cold start conditions will not exceed 37 A. Overvoltage Protection Built in OVP on all models. Approximately 120-140% of output voltage. Output Noise 0.5% rms, 1% Pk-Pk, 20 MHz Bandwidth, differential mode. Measured with noise probe directly across output terminals of the power supply. Medical Safety Approvals All models are Certified to be in compliance with the applicable requirements of UL2601-1, IEC60601-1, CSA-C22.2 No. 601-1, EN60601-1. Leakage Current The maximum leakage current for GSM15 series will be as follows; 132Vac/60Hz UL2601-1 test method Transient Response Main Output - 500 µs max. response time for return to within 0.5% of final value for a 50% load step change, ∆i/ ∆t< 0.2 A/μs. Maximum voltage deviation is 3.5%. Hold-up Time 10 ms minimum from loss of ac input voltage at full load, nominal line (120 Vac). GND Connection Normal Single Fault Class I 75 μA 105 μA Class II 39 μA 54 μA GND Connection Normal Single Fault Class I 128 μA 180 μA Class II 66 μA 94 μA 264Vac/50Hz IEC60601-1 test method All specifications are typical at nominal input, full load at 25OC unless otherwise stated Medical Model Voltage Output Min. Normal (A) Peak (B) Initial Set Point OVP Setpoint Total Regulation Ripple and Noise GSM15-5 5.1 V 0A 2.35 A 3A 2.5% 7.2 V 2% 1% GSM15-12 12 V 0A 1.25 A 1.5 A 2.5% 16 V 2% 1% GSM15-15 15 V 0A 1.0 A 1.2 A 2.5% 21 V 2% 1% GSM15-24 24 V 0A 0.625 A 0.75 A 2.5% 32 V 2% 1% GSM15-28 28 V 0A 0.54 A 0.64 A 2.5% 280 V 2% 1% Notes: A. Rating with unrestricted convection cooling. B. Peak Power for 60 sec. 10% duty cycle or continuous rating with 150 LFM of airflow. C. Output voltages preset at factory, not user adjustable. D. Add “G” suffix to model nuber for RoHS compliant model. GSM15 MECHANICAL SPECIFICATIONS INPUT: J1 AMP P/N 640456-4 PIN 1) AC LINE PIN 2) N/C PIN 3) N/C PIN 4) AC NEUTRAL GND: 0.098 DIA. THRU HOLE OUTPUT: J2 AMP P/N 640456-4 PIN 1) COMMON Return PIN 2) COMMON Return PIN 3) OUTPUT #1 + Vout PIN 4) OUTPUT#1 +Vout MATING CONNECTOR AMP P/N MTA – 100 Recepticle NOTE: 3A MAXIMUM RECOMMENDED CURRENT PER CONNECTOR PIN Overall Dimensions: 3.00 x 2.10 x .92 inches 76.20mm x 53.34mm x 23.37mm Weight: 0.25 LBS. [.113 kg] MAX. ENVIRONMENTAL SPECIFICATIONS OPERATING NON-OPERATING Temperature (A) 0 to 50° C -40 to +85°C Humidity (A) 0 to 95% RH 0 to 95% RH Shock (B) 20 gpk 40 gpk Altitude -500 to 10,000 ft -500 to 40,000 ft Vibration (C) 1.5 g rms’ 0.003 g2/Hz 5 g rms’ 0.026 g2/Hz A. Units should be allowed to warm up/operate under non-condensing conditions before application of power. B. Shock testing—half-sinusoidal, 10 ± 3 ms duration, ± direction, 3 orthogonal axes, total 6 shocks. C. Random vibration—10 to 2000Hz, 6dB/octave roll-off from 350 to 2000Hz, 3 orthogonal axes. Tested for 10 min./axis operating and 1 hr./axis non-operating. SL Power Electronics Corp., 6050 King Drive, Bldg. A, Ventura, CA 93003, USA. Phone:(805) 486 4565 Fax:(805) 487 8911 www.slpower.com Rev. 1/07. Data Sheet © 2007 SL Power Electronics Corp. The information and specifications contained in this data sheet are believed to be correct at time of publication. However, Condor accepts no responsibility for consequences arising from reproduction errors or inaccuracies. Specifications are subject to change without notice.