MINT1045 45 W Single Output Series Medical & ITE – Universal Input Power Supply 3 Year Warranty •45 W Convection Cooled •Universal Input 80--270 Vac •Very Small Size: 2” W x 3” L x 1.0” H •88% Efficiency Typical •Class I or Class II AC Input Applications rd •CSA/EN/IEC/UL60601-1 3 Edition for Medical, Type BF Capable, Ultra-low Leakage Current nd •CSA/EN/IEC/UL60950-1 2 Specifications All Specifications are typical at nominal input, full load at 25°C unless otherwise stated. AC Input 100-240Vac +/- 10%, 47-63 Hz 1∅ rated 80-270 Vac, 47-440 Hz operational 0.75A @ 120 Vac, 0.45 A @ 240 Vac Input Current Inrush Current 37A max., 264 Vac Input Fuses F1, F2: 1.6A, 250Vac fuses provided internally Power Factor Complies with EN61000-3-2, Class A Efficiency Output Power Edition for ITE 88% typical 45 W continuous. See chart for operation above 50°C Peak Output Power 50 W for 1 minute, 10% duty cycle Ripple and Noise 1% pk-pk max., 20 MHz BW, differential mode Measured with noise probe directly across output terminals Output Voltage See chart Voltage Adjustability Not adjustable Factory set with fixed resistors for higher reliability Minimum Load Not required Total Regulation Transient Response Switching Frequency See chart 500µs max. to 1%, 50% load step 0.2A/µs, 3.5% deviation. typical Fixed 65 kHz, typical Turn On Time 500 ms max at 120 Vac Hold-up Time 25 ms minimum at 45 W, nominal input (120Vac) Overvoltage Protection 120-150%, auto-reset Overload Protection Hiccup Mode, 58W typical Short Circuit Protection Hiccup Mode, 58W typical Overtemperature Protection Self-Recovering Medical Safety Standards CSA/EN/IEC/UL 60601-1, 3rd Edition Methods Of Protection: 2 MOPP for Type BF Applications ITE Safety Standards CSA/EN/IEC/UL 60950-1, IEC 2nd Edition See page 2 for detailed safety agency standards Isolation Input-Output: 4500Vac, Input-Ground: 1900Vac Output-Ground: 1900Vac Leakage Current 12V-28V models: 120Vac: 30µA NC/70µA SFC 264Vac: 75µA NC/130µA SFC 48V model: 120Vac: 90µA NC/120µA SFC 264Vac: 190µA NC/220µA SFC Operating Temperature -20° to 70°C See derating chart Output Power Derating Derate output power by 2.5% per °C above 50°C, up to 70°C max Storage Temperature -40 to +85°C Relative Humidity 5% to 95%, non-condensing MTBF 300K hours per Telcordia Model Number Key MINT 1 045 A 12 75 K 01 01 = Standard Model, 02 and higher indicates a m odified m odel. "K" = Input Connector - 3 pin header "75" = Output Connector - 4 pin header Output Voltage: "12" = 12V output, "24" = 24V output, etc. Model Configuration: "A" = Class I Input; "B" = Class II Input (no ground) Output Pow er (Watts) - "045" = 45W # of Outputs Product Fam ily: "M" = Medical, "I" = Internal, "NT" = New Technology ______________________________________________________________________________________________________________________________ 1 of 3 SL Power Electronics Corp • 6050 King Drive Ventura, CA 93003 • Phone 805.486.4565 • Fax 858.712.2040 • Email: [email protected] • www.slpower.com MINT1045 45 W Single Output Series Medical & ITE – Universal Input Power Supply 3 Year Warranty Output Parameters Model Number Volts Output Current Ripple & Noise Total Regulation Initial Set-point Tolerance ±2% MINT1045A1275K01 12 V 3.75 120mV ±0.5% MINT1045A1575K01 15 V 3.00 150mV ±0.5% ±2% MINT1045A2475K01 24 V 1.90 240mV ±0.5% ±2% MINT1045A2875K01 28 V 1.60 280mV MINT1045A4875K01 48 V 0.95 480mV Notes: 1. Derate output power by 2% per volt AC for input voltage <90Vac. ±0.5% ±2% ±0.5% ±2% Consult factory for availability Specifications (continued) Random vibration per MIL-STD-810E, Method 514.4, Cat. 1, Figure 514.4-1, 1 hr in each of three axes Enclosure Dimensions See outline drawing Vibration Weight 113 grams Shock Operating Altitude Half-sine, 40 gpk, 10 msec duration, +/- in each of three axes, total of 6 shocks -500 to 10,000 ft Non-operating Altitude -500 to 40,000 ft EMI/EMC Compliance Conducted Emissions EN55011/22 Class B, FCC Part 15, Class B, 6db margin Radiated Emissions Static Discharge Immunity EN55011/22 Class A, FCC Part 15, Class A, 10db margin To meet class B for grounded input (Class I), unit should be mounted on a metal plate. EN61000-4-2, 6kV Contact Discharge, 8kV air discharge Radiated RF Immunity EN61000-4-3, 3V/m. EFT/Burst Immunity EN61000-4-4, 2kV/5kHz. Line Surge Immunity EN61000-4-5, 1kV differential, 2kV common-mode Conducted RF Immunity EN61000-4-6, 3Vrms Power Frequency Magnetic Field Immunity Convection Cooled EN61000-4-8, 3A/m Voltage Dip Immunity EN61000-4-11 100Vac, 95% dip/0.5 cycle (Criteria A), 60%/5cycles (Criteria A at 30% load), 30%/25 cycles (Criteria A). Complies with EN61000-3-2, EN60601-2-1 Line Frequency Harmonics Safety Agency Compliance • • • Medical Certification to the New 3rd Edition: Demko certified to EN 60601-1:2005, 3rd Ed. CB certified to IEC 60601-1:2005, 3rd Edition UL Recognized to ANSI AAMI ES60601-1:2005 & CAN/CSA C22.2, No. 60601-1-08 ITE Certification to the New 2nd Edition: • Demko certified to EN60950-1:2006+ A11:2009,2nd Ed. • CB certified to IEC 60950-1:2005, 2nd Edition • UL Recognized to UL 60950-1:2007, 2nd Edition & CAN/CSA C22.2, No. 60950-1-07, 2nd Edition ______________________________________________________________________________________________________________________________ 2 of 3 SL Power Electronics Corp • 6050 King Drive Ventura, CA 93003 • Phone 805.486.4565 • Fax 858.712.2040 • Email: [email protected] • www.slpower.com MINT1045 45 W Single Output Series Medical & ITE – Universal Input Power Supply 3 Year Warranty Outline Drawing [4] 0.122 4 PLCS MTG HOLES Mating Connectors: J1 = AMP 640250-3, J2 = 640250-4, Contacts = AMP 770461-1 Note: For Class II (No earth ground) applications, all mounting hardware must be non-conductive. Data Sheet © 2011 SL Power Electronics Corp. The information and specifications contained herein are believed to be correct at the time of publication. Rev. 6-7-2011 However, SL Power accepts no responsibility for consequences arising from reproduction errors or inaccuracies. Specifications are subject to change without notice. ______________________________________________________________________________________________________________________________ 3 of 3 SL Power Electronics Corp • 6050 King Drive Ventura, CA 93003 • Phone 805.486.4565 • Fax 858.712.2040 • Email: [email protected] • www.slpower.com