MPD MG205MD12HI Mg200mhi sery Datasheet

MG200MHI Series
Compact MiniDIP, 2W
Medical Approved
DC/DC Converters
Electrical Specifications
Specifications typical @ +25°C, nominal input voltage & rated output current, unless otherwise noted. Specifications subject to change without notice.
Key Features:
• 2W Output Power
• 4 kVAC rms Isolation
Input
Parameter
Conditions
Input Voltage Range
• Reinforced Insulation
• Compact MiniDIP Case
Input Filter
Reverse Polarity Input Current
Leakage Current
• Single & Dual Outputs
Output
• 2.0 MH MTBF
Parameter
Output Voltage Accuracy
Output Voltage Balance
Line Regulation
Load Regulation, See Note 1
Ripple & Noise (20 MHz), See Note 2
Ripple & Noise (20 MHz)
Ripple & Noise (20 MHz)
Temperature Coefficient
Output Short Circuit
• Low Leakage Current
• Industry Standard Pin-Out
• EN 60950 Approved
• EN 60601 Approved
Min.
Typ.
5 VDC Input
4.5
5.0
12 VDC Input 10.8
12.0
24 VDC Input 21.6
24.0
Internal Capacitor
Conditions
Max.
5.5
13.2
26.4
Units
0.3
2.0
A
µA
Max.
±4.0
±1.0
±1.5
Units
%
%
%
150
200
15
±0.02
mV P - P
mV P - P
mV rms
%/°C
Typ.
Max.
15
80
20
100
Units
VAC rms
VPk
VAC
G⍀
pF
kHz
Typ.
Max.
+80
+90
+125
Units
°C
°C
°C
95
%
Min.
Typ.
±2.0
Dual Outputs, Balanced Loads
±0.1
Vin = Min to Max
±1.2
See Model Selection Guide
100
Over Line, Load & Temp.
±0.01
Momentary (0.5 Sec.)
VDC
General
Parameter
Isolation Voltage, Rated
Isolation Test Voltage
Reinforced Insulation Working Voltage
Isolation Resistance
Isolation Capacitance
Switching Frequency
Conditions
60 Seconds
Flash Tested For 1 Sec
500 VDC
100 kHz, 1V
Min.
4,000
6,000
300
10
50
Environmental
MicroPower Direct
292 Page Street
Suite D
Stoughton, MA 02072
USA
T: (781) 344-8226
F: (781) 344-8481
E: [email protected]
W: www.micropowerdirect.com
Parameter
Operating Temperature Range
Operating Temperature Range
Storage Temperature Range
Cooling
Humidity
Physical
Case Size
Case Material
Weight
Conditions
Ambient
Case
Min.
-25
-50
Free Air Convection
RH, Non-condensing
0.94 x 0.53 x 0.34 Inches (23.8 x 13.4 x 8.62 mm)
Non-Conductive Black Plastic (UL94-V0)
0.18 Oz (5.1g)
Reliability Specifications
Parameter
MTBF
Safety Approvals
Conditions
Min.
Typ.
Max.
MIL HDBK 217F, 25°C, Gnd Benign
2.0
UL 60601, UL 60950, EN 60601, EN 60950
Units
MHours
Absolute Maximum Ratings
Parameter
Conditions
Input Voltage Surge (1 Sec)
Lead Temperature
Internal Power Dissipation
5 VDC Input
12 VDC Input
24 VDC Input
1.5 mm From Case For 10 Sec
All Models
Min.
-0.7
-0.7
-0.7
Typ.
Max.
9.0
18.0
30.0
260
650
Caution: Exceeding Absolute Maximum Ratings may damage the module. These are not continuous operating ratings.
www.micropowerdirect.com
Units
VDC
°C
mW
www.micropowerdirect.com
Model Selection Guide
Model
Number
MG205MS-05HI
MG205MS-12HI
MG205MS-15HI
MG205MD-12HI
MG205MD-15HI
MG212MS-05HI
MG212MS-12HI
MG212MS-15HI
MG212MD-12HI
MG212MD-15HI
MG224MS-05HI
MG224MS-12HI
MG224MS-15HI
MG224MD-12HI
MG224MD-15HI
Input
Voltage (VDC)
Current (mA)
Voltage
Nominal
Range
Full-Load No-Load (VDC)
5
4.5 - 5.5
606
60
5.0
5
4.5 - 5.5
600
60
12.0
5
4.5 - 5.5
605
60
15.0
5
4.5 - 5.5
553
60
±12.0
5
4.5 - 5.5
542
60
±15.0
12
10.8 - 13.2
253
30
5.0
12
10.8 - 13.2
250
30
12.0
12
10.8 - 13.2
252
30
15.0
12
10.8 - 13.2
224
30
±12.0
12
10.8 - 13.2
220
30
±15.0
24
21.6 - 26.4
126
15
5.0
24
21.6 - 26.4
125
15
12.0
24
21.6 - 26.4
126
15
15.0
24
21.6 - 26.4
112
15
±12.0
24
21.6 - 26.4
110
15
±15.0
Notes:
1. Load regulation is measured for a load change of 20% to 100%.
2. When measuring output ripple, it is recommended that an external 0.47 µF ceramic capacitor
be placed from the +Vout pin to the -Vout pin for single output units and from each output to
common for dual output units. For noise sensitive applications, the use of 3.3 µF capacitors will
reduce the output ripple.
3. Operation at no-load will not damage these units. However, they may not meet all specifications.
4. For dual output units, the maximum capacitive load is given for each output.
5. Dual output units may be connected to provide a 24 VDC or 30 VDC output. To do this, connect
the load across the positive (+Vout) and negative (-Vout) outputs and float the output common.
6. The converter should be connected to a low ac-impedance source. An input source with a highly
inductive impedance may affect the stability of the converter. In applications where the converter
output loading is high and input power is supplied over long lines, it may be necessary to use
a capacitor on the input to insure start-up.
In this case, it is recommended that a low ESR (ESR <1.0 at 100 kHz) capacitor be mounted
close to the converter. For 5V input units a 2.2 µF is recommended, for 12V input units, a
1.0 µF; and for 24V units a 0.47 µF.
7. It is recommended that a fuse be used on the input of a power supply for protection. See the
table above for the correct rating.
Output
Current
Current
(mA, Max)
(mA, Min)
400.0
165.0
133.0
±83.0
±66.0
400.0
165.0
133.0
±83.0
±66.0
400.0
165.0
133.0
±83.0
±66.0
8.0
3.0
2.5
±1.5
±1.0
8.0
3.0
2.5
±1.5
±1.0
8.0
3.0
2.5
±1.5
±1.0
Capacitive
Load
Rating
Efficiency Fuse
Slow-Blow
Load
Regulation
(%, Typ)
(mA)
(µF, Max)
(%, Max)
330
330
330
±100
±100
330
330
330
±100
±100
330
330
330
±100
±100
12
10
10
10
10
12
10
10
10
10
12
10
10
10
10
66
66
66
72
73
66
66
66
74
75
66
66
66
74
75
1,000
1,000
1,000
1,000
1,000
500
500
500
500
500
200
200
200
200
200
Derating Curve
Mechanical Dimensions
Pin Connections
Pin
Single
Dual
1
-Vin
-Vin
7
NC
NC
8
NC
Common
9
+Vout
+Vout
10
-Vout
-Vout
16
+Vin
+Vin
NC = No Connection
Mechanical Notes:
• All dimensions are typical in inches (mm)
• Tolerance x.xx = ±0.01 (±0.25)
MicroPower Direct
We Power Your Success - For Less!
292 Page Street Ste D Stoughton, MA 02072
•
TEL: (781) 344-8226
•
FAX: (781) 344-8481
•
E-Mail: [email protected]