Medical AC-DC Enclosed Power Supply 12V 200W 2”x4” High Density / MDS-200APB12 AA MDS-200APB12 AA Highlights & Features –Safety Approvals to IEC 60601-1 3.1rd ed. & IEC 60950-1 –Compliant with IEC 60601-1-2 4th Ed. Requirements –Power Good Signal –Up to 800K Hours MTBF –Up to 130W convection, 200W forced air Safety Standards CB Certified for worldwide use Model Number: MDS-200APB12 AA Unit Weight: 200g(0.441lb) Dimensions (W x L x H): 50.8 x 101.6 x 35.6mm (2x4x1.4 inch) General Description Delta’s new MDS-200APB12 AA internal open frame power supply comes with universal AC input range from 90Vac to 264Vac. Other features include low leakage, Type BF Patient Access Leakage Currents, and electric shock protection compliance with 2 x MOPP requirements. The MDS-200APB12 AA is certified for EMC standards according to EN 55011for industrial, scientific and medical (ISM) radio-frequency equipment; and, EN 55022 for Industrial Technology Equipment (ITE) radio-frequency equipment. In addition, only recognized Japanese capacitors are used to ensure long product life. The MDS-200APB12 AA comes with both medical and ITE safety approvals, including UL/CE/CCC (5000meters), and CB certification. Designs are compliant with RoHS Directive 2011/65/EU for environmental protection. Model Information Model Number Input Voltage MDS-200APB12AA Output Voltage 90-264Vac Convection Current Output 12Vdc Forced Air Current Output 0-10.83A 0-16.67A* *With 8.5CFM forced air Model Numbering MDS Delta Medical power Supply 1 200 Max wattage in the product Series. Maybe lower at some voltage. 200→200W A Family Code A~ Z P B 12 Product Type Input Type Code Output Voltage P: Open Frame B: 3pin Class I 12 for 12V All parameters are specified at 25°C ambient unless otherwise noted. (November 2016, Rev. 06) AA Revision code Medical AC-DC Enclosed Power Supply 12V 200W 2”x4” High Density / MDS-200APB12 AA Specifications Input Ratings / Characteristics Nominal Input Voltage 100-240Vac Input Voltage Range 90-264Vac Nominal Input Frequency 50-60Hz Input Frequency Range 47-63Hz Input Current (max) 2.5A Input Surge Voltage (max) 300Vac for 100ms Full load Efficiency (typ.) 91% @ 115Vac/60Hz 93% @ 230Vac/50Hz, Reference Fig.1 Standby Power (max) 0.5W @ 115Vac/60Hz, 230Vac/50Hz Inrush Current (max) 60A@230Vac, cold start Input-PE(protective earth) leakage current(max) 0.1mA @ NC, 0.3mA @ SFC1) Output-PE(protective earth) leakage current for Type BF application (max) 0.1mA @ NC, 0.5mA @ SFC 1) Power Factor (min) 0.95 @ 115V/50Hz, 230V/50Hz, full load 1) NC: normal condition, SFC: single fault condition Leakage Current Input-PE Leakage Current 100Vac/60Hz(Typ) Normal Condition 18.6uA Single Fault Condition 45.6uA 264Vac/60Hz(Typ) Delta Limit IEC60601-1 Limit 44.8uA 100uA max 5000uA max 151.9uA 300uA max 10000uA max 82.2uA 100uA max 100uA max 128.1uA 500uA max 500uA max Output-PE Leakage Current for Type BF application Normal Condition 39.2uA Single Fault Condition 44.1uA 100.000 90.000 Efficiency (%) 80.000 115Vac 70.000 230Vac 60.000 50.000 25% 50% 75% Output Load (200W) Fig.1 Efficiency versus output load 2 All parameters are specified at 25°C ambient unless otherwise noted. (November 2016, Rev. 06) 100% Medical AC-DC Enclosed Power Supply 12V 200W 2”x4” High Density / MDS-200APB12 AA Output Ratings / Characteristics Nominal Output Voltage (Vrated) 12V Total Regulation (max.) ±3% Output Power 200W 8.5CFM air, up to 130W convection air Line Regulation (max) ±0.5% Load Regulation (max) ±2.5% Ripple& Noise (typ.) 1%pk-pkVrated @ Full load, Reference Fig. 2 Start-up Time(max) 2000ms@115Vac Hold-up Time(min) 12ms@115Vac, tested with 130W load Dynamic Response(Overshoot & Undershoot O/P Voltage) ±5% @50-100% load Capacitive load (max) 1000uF @200W 1500uF@130W Rise time (max) 100ms (a) 115V (measured value=80mV) (b) 230V(measured value=81.6mV) Fig.2 Ripple & Noise example, 20MHz BW Ripple & Noise measurement circuit 3 All parameters are specified at 25°C ambient unless otherwise noted. (November 2016, Rev. 06) Medical AC-DC Enclosed Power Supply 12V 200W 2”x4” High Density / MDS-200APB12 AA Mechanical Case Cover NA Dimensions(W x L x H typ.) 50.8x101.6x35.6mm (2x4x1.4inch) Unit Weight 200g(0.441lb) Indicator NA Cooling System NA Environment Surrounding Air Temperature Operating Storage Absolute Max -10°C to+70°C, supported power linearly de-rate from 50°C to 50% rated up to 70°C Note: see power de-rating curve -40°C to+85°C Operating Humidity 5-95% RH (Non-Condensing) Operating Altitude 5,000 meters (16402 feet) Shock Test (Non-Operating) 50G, 11ms, 3 shocks for each direction Vibration (Operating) 5-500Hz, 2Grms, 15 minute for each three axis Power De-rating curve 4 All parameters are specified at 25°C ambient unless otherwise noted. (November 2016, Rev. 06) Medical AC-DC Enclosed Power Supply 12V 200W 2”x4” High Density / MDS-200APB12 AA Protections Overvoltage (max) 125% of rated voltage, Latch Mode Overload / Overcurrent (max) Main output 160% of rated current Hiccup Mode(Non-Latching, Auto-Recovery) Over Temperature Latch Mode Short Circuit Hiccup Mode, (Non-Latching, Auto-Recovery) Reliability MTBF(Minimum) at 115Vac, 130W, 35°C, Convection Air Flow 800 kHrs based on Telecordia SR-332 Operating life at 115Vac, 130W, ambient 25 °C, Convection Air Flow 26,280Hrs Safety Standards / Directives nd rd Medical Safety IEC60601-1 2 and 3.1 edition CB report rd IEC60601-1 edition 3.1 (2012), EN60601-1 (2006) + A11 + A1 + A12, CAN/CSA-C22.2 NO. 60601-1:14, ANSI/AAMI ES60601-1:2005/(R)2012 ITE Safety IEC60950-1 CB report TUV60950-1 UL60950-1+CAN/CSA60950-1 GB4943.1-2011, GB9254-2008, GB17625.1-2003 CE MDD Directive 93/42/EEC Environmental RoHS Directive 2011/65/EU Compliant Galvanic Isolation 1) 5 Input to/Output (2XMOPP) 4000 Vac 1) Input to/Ground(1XMOPP) 1500Vac Output to/Ground(1XMOPP) 1500Vac ( Type BF application rated) PSU can support PoE applications with Primary to FG 2500Vac test. All parameters are specified at 25°C ambient unless otherwise noted. (November 2016, Rev. 06) Medical AC-DC Enclosed Power Supply 12V 200W 2”x4” High Density / MDS-200APB12 AA EMC (Compliant with IEC 60601-1-2 4th Ed. Requirements) EMC / Emissions Harmonic Current Emissions EN55011/EN55022,FCC Title 47:Class B IEC61000-3-2 Meet Class D limit Immunity to Voltage Flicker IEC61000-3-3 Electrostatic Discharge IEC61000-4-2 Level 4 Criteria A Air Discharge: 15kV Contact Discharge: 8kV Radiated Field IEC61000-4-3 Electrical Fast Transient / Burst IEC61000-4-4 Criteria A 80MHz-2700MHz, 10V/m AM modulation 385MHz-5785MHz, 28V/m Pulse mode and other modulation 1) Level 3 Criteria A :2kV Surge IEC61000-4-5 Level 3 Criteria A 3) Common Mode : 2kV 4) Differential Mode : 1kV Conducted IEC61000-4-6 Level 2 Criteria A 150kHz-80MHz, 3Vrms, 6Vrms at ISM bands and 1) 1) 1) 1) Amateur radio bands Power Frequency Magnetic Fields Voltage Dips IEC61000-4-8 IEC61000-4-11 1) Criteria A Magnetic field strength 30A/m 1) Criteria A 0% UT, 0.5 cycle (10ms) o o o o o o o o o , 0 /45 /90 /135 /180 /225 /270 /315 /360 2) Criteria B o 0% UT, 1 cycle (20ms), 0 2) Criteria B o 70% UT, 25 cycle (500ms), 0 2) Criteria B o 0% UT, 250 cycle (5000ms), 0 1) Criteria A: Normal performance within the specification limits 2) Criteria B: Output out of regulation, or shuts down during test. Automatically restore to normal operation after test. 3) Asymmetrical: Common mode (Line to earth) 4) Symmetrical: Differential mode (Line to line) 6 All parameters are specified at 25°C ambient unless otherwise noted. (November 2016, Rev. 06) Medical AC-DC Enclosed Power Supply 12V 200W 2”x4” High Density / MDS-200APB12 AA Dimensions W x L x H: 50.8 x 101.6 x 35.6 mm Mechanical drawing (3Pin input type) Notes ― Dimensions are in mm Input Connector CN1 Pin 1 Neutral Pin 2 Line CN1 mates with Molex housing 26033031 and Molex series 6838 crimp terminals. Input Line can also be connected to pin 1, and Input Neutral can be connected to pin 2 Pin 1 Pin 2 Signal Connector CN102 Power Good -(DC RTN) Power Good + CN102 mates with Molex housing 0874390200 and Molex 874210000 crimp terminals. CN3: PINGOOD : JP-13T or equivalent mate with KST :FDFNYD1187 or other applicable connectors Output Connector CN101 Pin1 DC RTN Pin2 DC RTN Pin3 Vo Pin4 Vo CN101 mates with JST housing VHR-4N and JST terminal SVH-41T-P1.1. Two mounting points in mechanical drawing need to be connected to system earth case together, Protective bonding conductor from the end product protective earth terminal (if any) can be tied to CN3 for open frame model. 7 All parameters are specified at 25°C ambient unless otherwise noted. (November 2016, Rev. 06) Medical AC-DC Enclosed Power Supply 12V 200W 2”x4” High Density / MDS-200APB12 AA Functions Start-up Time The time required for the output voltage to reach 90% of its set value, after the input voltage is applied. Rise Time The time required for the output voltage to change from 10% to 90% of its set value. Hold-up Time Hold up time is the time when the AC input collapses and output voltage retains regulation for a certain period of time. The time required for the output to reach 90% of its set value, after the input voltage is removed. ■Graph illustrating the Start-up Time, Rise Time, and Hold-up Time 8 All parameters are specified at 25°C ambient unless otherwise noted. (November 2016, Rev. 06) Medical AC-DC Enclosed Power Supply 12V 200W 2”x4” High Density / MDS-200APB12 AA Dynamic Response Short Circuit Protection The power supply output voltage will remain within ±5% of its steady state value, when subjected to a dynamic load 50 to 100% of its rated current. ■ 50 to 100% Load The power supply’s output OLP/OCP function also provides protection against short circuits. When a short circuit is applied, the output current will operate in “Hiccup mode”, as shown in the illustration in the OLP/OCP section on this page. The power supply will return to normal operation after the short circuit is removed. Inrush Current Inrush current is the peak, instantaneous, input current measured and, occurs when the input voltage is first applied. For AC input voltages, the maximum peak value of inrush current will occur during the first half cycle of the applied AC voltage. This peak value decreases exponentially during subsequent cycles of AC voltage. Overload & Overcurrent Protections The power supply’s Overload (OLP) and Overcurrent (OCP) Protections will be activated before output current under 160% of I O (Max load). Upon such occurrence, V O will start to drop. Once the power supply has reached its maximum power limit, the protection will be activated and the power supply will go into “Hiccup mode” (Auto-Recovery). The power supply will recover once the fault condition causing the OLP and OCP is removed and I O is back within the specified limit. Additionally, if the Iout is <160% but >100% for a prolong period of time (depending on the load), the Over Temperature Protection (OTP) will be activated due to high temperature on critical components. Then, the power supply will be latched off, and require recycling of input voltage to restart it. Over Temperature Protection Overvoltage Protection The power supply’s overvoltage circuit will be activated when its internal feedback circuit fails. The output voltage shall not exceed its specifications defined on Page 6 under “Protections”. Power supply will latch off, and require removal/re-application of input AC voltage in order to restart. 9 As mentioned above, the power supply also has Over Temperature Protection (OTP). This is activated when the overload condition persists for an extended duration and the output current is below the overload trigger point but >100% load. In the event of a higher operating temperature condition at 100% load, the power supply will run into OTP when the surrounding air temperature is higher than the operating temperature. When activated, the output voltage will go into latch mode until the input voltage is removed; then, reapplied, and the surrounding air temperature drops to its normal operating temperature. All parameters are specified at 25°C ambient unless otherwise noted. (November 2016, Rev. 06) Medical AC-DC Enclosed Power Supply 12V 200W 2”x4” High Density / MDS-200APB12 AA Power Good Power Good+ pin is an open collector transistor (40V/600mA rating). A resistor (suggested value 20Kohm, 1/8W) can be added between output pin (or, other available pull-up voltage that is no greater than 30V) and the Power Good+ pin (refer to figure below). Value of pull-up resistor may have to be adjusted, depending on voltage used, and other end-use conditions of the Power Good+ pin connection to the product. When AC input is on, Power Good+ pin will be high. When AC input is off, Power Good+ pin will be low. There will be a minimum of 5 milliseconds (with 120W convection output power) between the time the power good goes to low level, and the time when the output reaches 90% of its rated value. 10 All parameters are specified at 25°C ambient unless otherwise noted. (November 2016, Rev. 06) Medical AC-DC Enclosed Power Supply 12V 200W 2”x4” High Density / MDS-200APB12 AA Certificate All Delta Medical Power products conform to the European directive 2011/65/EU. RoHS is the abbreviation for “Restriction of the use of certain hazardous substances Delta has been certified as meeting the requirement of ISO 13485: 2003 and EN ISO 13485:2012 for the design and manufacture of switching power supply and adaptor for medical device. In addition to a UL Total Certification Program (TCP) approved client laboratory for IEC60950 and IEC60065. Delta also has participated UL Client Test Data Program (CDTP) for IEC 60601 11 All parameters are specified at 25°C ambient unless otherwise noted. (November 2016, Rev. 06)