FFI SYSTEM OPTICAL TEMPERATURE TRANSMITTER DATA SHEET FUK···3, FUL···3, FUM···1 The Model FUK 3/FUL 3/FUM 1 is Optical Temperature Transmitter for thermocouple or platinum resistance bulb used for measuring the temperature of different fluids. The transmission unit incorporates a microprocessor that carries out digital processing of signals for ensuring a high accuracy and an intelligent operation. A fiber optical cable used for the signal transmission line forms an optical field instrumentation system together with an optical star coupler and a master station. FUK FUL MODEL TYPE CONFIGURATION 1) Thermocouple temperature transmitter (integrated type), Model type: FUK 2) Resistance bulb temperature transmitter (integrated type), Model type: FUL 3) Temperature transmitter (separate type), Model type: FUM FUM FEATURES 1. Resistive to noise and lightning Optical signal ensures a reliable signal transmission, because it is not be affected by external noise and inductive lightning. Use of a nonmetallic optical (fiber) cable prevents propagation of inductive lightning through the cable, so a signal transmission immune to lightning can be realized. 2. Reliability due to redundant configuration Host system can be duplicated by using two optical cable trunk lines (between an optical star coupler and host system). This enhances reliability of users' systems. 3. Intrinsic safety type explosion-proof Each equipment with a built-in battery can be constructed so as to be an intrinsic safety type individually (intrinsic safety type barrier unnecessary). SPECIFICATIONS -200 to +400°C -200 to +400°C -200 to +400°C -200 to +400°C J thermocouple E thermocouple K thermocouple R thermocouple T thermocouple Resistance bulb Pt100 -200 to +750°C -200 to +800°C -200 to +1200°C 0 to +1600°C -200 to +350°C -200 to +600°C Measuring span Minimum Maximum 100°C 100°C 150°C 200°C 50°C 80°C 750°C 800°C 1200°C 1600°C 350°C 600°C Operating pressure: Flange type ... Rated pressure of flange Screw-in type ... 4 MPa or less Output: Optical digital signal Burnout: If sensor is burnt out, last output is held, and alarm is delivered (detecting unit failure) Self-diagnosis: Displayed on indication unit (option) and transmitted to master station. Diagnosis item Host system Indication unit Detecting unit failure Amplifier abnormal Fluids measured: Liquid, gas or steam Measuring range, measuring span: (1) Integrated type (FUK, FUL) J thermocouple E thermocouple K thermocouple Resistance bulb Pt100 Measuring range Measuring range abnormal Functional specifications Measuring range (2) Separate type (FUM) Battery voltage Battery voltage low alarm Measuring span Minimum Maximum 50°C 50°C 50°C 50°C 400°C 400°C 400°C 400°C EDS2-17 Date May 20, 2004 FUK···3, FUL···3, FUM···1 Remote Control Function: (Items readable and setting from hand-held communicator) Item Maximum range Measuring range Real scale indication Battery voltage Error indication Measured value Adjustment Reading Setting Maximum measuring range of equipment Actual measuring range Indication in industrial value Battery voltage of equipment Errors of detection unit and amplifier Measured data Zero and span adjustment Output signal: Power supply: Optical digital signal Built-in lithium battery (expected life: about 4 years) Optical cable: Code set type, silica fiber ... core/clad diameter 100/140 µm Optical connector: FC connector Transmission distance: 1.5 km max. (when transmission loss of optical cable is 4 dB/km) Zero point shift: Minimum temperature to 80% of maximum temperature of measuring range Explosion-proof: Intrinsic safety type, JIS ib IIC T3 Ambient temperature: –30 to +70°C –10 to +60°C for intrinsic safety explosion-proof type –20 to +70°C when provided with indicator Storage temperature: –40 to +80°C Performance specifications 1. Detecting unit (for FUK, FUL) Accuracy rating: Thermocouple ... IEC 584, class 2 Resistance bulb ... IEC751, class B Response time (time constant): For protective pipe Ø4.8 mm ... About 8 sec For protective pipe Ø12 mm ... About 70 sec Temperature range (integrated type) J thermocouple -140 to +400°C (separate type) Less than -140°C ∗ (separate type) J thermocouple -100 to +750°C Less than -100°C ∗ (integrated type) E thermocouple -200 to +400°C (separate type) (separate type) E thermocouple -80 to +800°C Less than -80°C ∗ (integrated type) K thermocouple -150 to +400°C (separate type) Less than -150°C ∗ (separate type) K thermocouple 0 to +1200°C Less than 0°C ∗ (separate type) R thermocouple 0 to +1600°C (separate type) T thermocouple -200 to +350°C (integrated type) Pt100 -200 to +400°C (separate type) (separate type) -200 to +600°C Model type Description Note: For operation of the "3" type transmitter ("3" at the 8th digit of product code), a hand-held communicator is required to have a version 1.6 or higher, but a communicator before version 1.6 can be operated with memory data updated. (Refer to the instruction manual of temperature transmitter.) 2 2. Transmission unit Accuracy rating FUK1 FUM A FUM B FUK2 FUM C FUM D FUK3 FUM E FUM F FUM FUM FUL1 FUM FUM G H L K M FUM N (separate type) FUM P (separate type) Sensor type JPt100 JPt100 -200 to +400°C -200 to +500°C Accuracy rating ±0.6°C ±1°C ±1.1°C ±1.9°C ±0.6°C ±1.2°C ±2°C ±0.6°C ±1°C ±1.8°C ±3°C ±4°C ±0.8°C ±0.6°C ±0.9°C ±0.6°C ±0.7°C Note 1: Accuracy is for when segment correction is made. Note 2: Reference junction compensation error is not included. Reference junction compensation accuracy: ± 1°C (for thermocouple) Ambient temperature effect: (shift between -30 to +70°C) ± 0.5% of maximum measurement temperature or, if temperature range above is marked "∗", ± 1% Allowable wiring resistance: (For separate type) Thermocouple ... 100Ω or less. Resistance bulb ... 10Ω or less. Physical specifications Environmental protection: Meets JIS C0920, immersion-proof (equivalent to IEC IP67 or NEMA 6/6P) Optical cable connection: G1/2 Mounting method: (1) FUK, FUL: Screw-in or flange mounting ... As specified by code symbol (2) Separate type (FUM): Mounted on 50A (2B) pipe with U-bolt or on a wall Finish: Epoxy-polyurethane double coat, silver (blue for amplifier case cover). Mass: (1) FUK, FUL ... 3 to 5 kg. Varies with sensor unit length. (2) Separate type (FUM) ... About 2.4 kg Insertion length: (FUK, FUL sensor unit length) 100 to 2000 mm ... See Code symbols Minimum bend radius: (Allowable bend of FUK or FUL sensor unit) 14.4 mm ... Sheath type (Ø4.8 mm) only External dimensions: See OUTLINE DIAGRAM. Material: (FUK, FUL sensor unit protective pipe) SUS304 or SUS316 ... As specified by code symbol Optional specifications Indication unit: 5-digit LCD indication, % or real scale indication Operating temperature range: -20 to +70°C SCOPE OF DELIVERY Instrument body and pipe fixture (for separate type) Relation between velocity of fluid measured and insertion length When measuring a temperature upon insertion in the fluid piping, the temperature detector is subjected to a stress by fluid flow. So, the mechanical strength of a protective pipe to put in must be taken into account. Study the insertion length for a thermometer to use referring to the relation between velocity and insertion length in Fig. 1. If a sheath type is inserted in fluid, provide a protective pipe separately. Insertion length L (mm) 1300 1000 Air Water 100 10 0.1 1 10 100 Velocity (m/s) Fig. 1 Relation between velocity and insertion length (for Ø12 mm protective pipe) 3 FUK···3, FUL···3, FUM···1 SYSTEM BLOCK DIAGRAM 1:1 system Host system Master station Control room Field 1 to 5V DC 4 to 20mA DC Optical fiber type electric converter (Type; PSN) Star coupler Transmitter Transmitter Field equipment (max. 8 units) Transmission distance: 1.5 km max. (between master station and field equipment) 4 Transmission distance: 4 km max. CODE SYMBOLS (1) Thermocouple temperature transmitter 1 2 3 4 5 6 7 8 F UK 3 9 10111213 Description Measuring range (4th digit) J thermocouple -200 to +400°C E thermocouple -200 to +400°C K thermocouple -200 to +400°C 1 2 3 Protective pipe shape (mm) (5th digit) Ø12 Ø4.8 (sheath type) A B Protective pipe material (6th digit) SUS304 (cannot be specified when 5th digit is B) SUS316 S W Indicator (7th digit) Not provided Digital, % indication Digital, real scale indication A L P Mounting dimension (9th digit) Flange JIS 10K-25ARF Flange JIS 20K-25ARF Flange JIS 30K-25ARF Flange JIS 63K-25ARF Screw-in G3/4B Screw-in R3/4 Screw-in 3/4-14NPT Flange ANSI 150LB, 1B Flange ANSI 300LB, 1B Flange ANSI 600LB, 1B 0 1 2 3 4 5 6 7 8 9 Cable lead-in port (10th digit) G1/2 1/2-14NPT S T Explosion-proof (11th digit) Non-explosion proof Intrinsic safety, JIS 0 1 0 0 0 0 0 0 0 0 0 1 1 1 1 1 1 1 1 1 1 2 1 2 3 4 5 6 7 8 9 0 1 2 3 4 5 6 7 8 9 0 Insertion length (mm) (12th and 13th digits) 100 200 300 400 500 600 700 800 900 1000 1100 1200 1300 1400 1500 1600 1700 Cannot be specified when 1800 5th digit is A 1900 2000 5 FUK···3, FUL···3, FUM···1 CODE SYMBOLS (2) Resistance bulb temperature transmitter 1 2 3 4 5 6 7 8 FUL 3 9 10111213 Description Measuring range (4th digit) 3 wire system Pt100 (1997) -200 to +400°C 1 Protective pipe shape (mm) (5th digit) Ø12 Ø4.8 (sheath type) A B Protective pipe material (6th digit) SUS304 (cannot be specified when 5th digit is B) SUS316 S W Indicator (7th digit) Not provided Digital, % indication Digital, real scale indication A L P Mounting dimension (9th digit) Flange JIS 10K-25ARF Flange JIS 20K-25ARF Flange JIS 30K-25ARF Flange JIS 63K-25ARF Screw-in G3/4B Screw-in R3/4 Screw-in 3/4-14NPT Flange ANSI 150LB, 1B Flange ANSI 300LB, 1B Flange ANSI 600LB, 1B 0 1 2 3 4 5 6 7 8 9 Cable lead-in port (10th digit) G1/2 1/2-14NPT S T Explosion-proof (11th digit) Non-explosion proof Intrinsic safety, JIS 0 1 0 0 0 0 0 0 0 0 0 1 1 1 1 1 1 1 1 1 1 2 6 1 2 3 4 5 6 7 8 9 0 1 2 3 4 5 6 7 8 9 0 Insertion length (mm) (12th and 13th digits) 100 200 300 400 500 600 700 800 900 1000 1100 1200 1300 1400 1500 1600 1700 Cannot be specified when 1800 5th digit is A 1900 2000 (3) Temperature transmitter (separate type) 1 2 3 4 5 6 7 8 F UM 1 S T Description Cable lead-in port (4th digit) G1/2 1/2-14NPT Sensor type, measuring range (5th digit) J thermocouple -200 to +400°C J thermocouple -200 to +750°C E thermocouple -200 to +400°C E thermocouple -200 to +800°C K thermocouple -200 to +400°C K thermocouple -200 to +1200°C R thermocouple 0 to +1600°C T thermocouple -200 to +350°C 3 wire system Pt100 (1989) -200 to +400°C 3 wire system Pt100 (1989) -200 to +600°C 3 wire system Pt100 (1997) -200 to +400°C 3 wire system Pt100 (1997) -200 to +500°C 3 wire system JPt100 -200 to +400°C 3 wire system JPt100 -200 to +500°C Explosion-proof (6th digit) For general application (Non-explosion proof) Intrinsic safety, JIS A B C D E F G H J K L M N P A G A L P Indicator (7th digit) Not provided Digital, % indication Digital, real scale indication OUTLINE DIAGRAM (Unit : mm) 1) Thermocouple temperature transmitter, Resistance bulb temperature transmitter 0 3– 1 7 28 39 L 222 49 t Øg ØD Ød P Ø102 FUK FUL 241 f 54 9th digit of code symbol Flange t ØD ØC Øh 0 JIS 10K-25A RF 14 125 90 19 1 JIS 20K-25A RF 16 125 90 19 2 JIS 30K-25A RF 20 130 95 19 3 JIS 63K-25A RF 27 140 100 23 7 ANSI 150lb,1B 14.5 108 79.4 16 8 ANSI 300lb,1B 17.5 124 88.9 20 9 ANSI 600lb,1B 17.5 124 88.9 20 70 121 8 80 15 ØC 4-Øh Indicator (option) "B" Optical cable lead-in port View from arrow P f Øg 1.0 70 1.6 50.8 15 3 Ø15 FUK FUL E Details of "B" 4 3– 5 6 L 223 20 Type 5th digit 10th digit E A S G1/2 A T 1/2-14NPT B S G1/2 B T Ød L 12 100 to 1300 (as specified) 100 to 2000 4.8 (as specified) (sheath type) 1/2-14NPT Ød Ø102 38 M 54 9th digit of code symbol M 4 G3/4 5 R3/4 6 3/4-14NPT 7 FUK···3, FUL···3, FUM···1 OUTLINE DIAGRAM (Unit : mm) 2) Temperature transmitter (separate type) Mounting pipe JIS 50A (2B) 241 80 108 70 U-bolt (M8) Terminal block (for sensor connection) 70 Ø102 8 225 266 54 15 "B" Optical cable and sensor cable lead-in port Indicator (option) 27 47 15 3 Ø15 M3 screw A 6.5 Details of "B" Type A FUMS G1/2 FUMT 1/2-14NPT A + B – B E Resistance bulb E Thermocouple Connection diagram ORDERING INFORMATION 1. 2. 3. 4. Model type Measuring range Indication scale for real scale specification Others Caution on Safety *Before using this product, be sure to read its instruction manual in advance. Head Office 6-17, Sanbancho, Chiyoda-ku, Tokyo 102-0075, Japan http://www.fesys.co.jp/eng Sales Div. International Sales Dept. No.1, Fuji-machi, Hino-city, Tokyo, 191-8502 Japan Phone: 81-42-585-6201, 6202 Fax: 81-42-585-6187 http: //www.fic-net.jp/eng Information in this catalog is subject to change without notice. Printed in Japan