Embedded Power for Business-Critical Continuity Rev.06.9.09_118 NFS40 Medical of 4 NFS40 Medical Series Single & Triple Output Total Power: 40 - 50W Input Voltage: 85 - 264 VAC 120 - 370 VDC # of Outputs: Single, triple Special Features • 5.0 x 3.0 x 1.2 inch package (1U applications) • Industry standard package • Overvoltage and short circuit protection • 40 W with free air convection • EN55022, EN55011 conducted noise level A • UL, VDE and CSA safety approvals • Available RoHS compliant • 2 year warranty Electrical Specifications Output Voltage adjustability: +5 V output on triples Vout on singles ± 5.0% ± 5.0% Line regulation: LL to HL, FL Main output Auxiliary outputs ± 0.2% ± 1.0% Load regulation: FL to NL Main output Auxiliary outputs ± 2.0% ± 5.0% Transient response: +5 V (1.5 - 3 A step) ± 120 mV max. dev. 500 µs recovery Temperature coefficient: All outputs ±0.02%/°C Overvoltage protection: +5 V output 6.25 ± 0.75 Vout Output power limit: Primary power limited 90 W input power limit Short circuit protection: Single outputs Multiple outputs Continuous Short term Input Safety • UL2601 • CSA22.2 No. 125 • IEC601/VDE0750 Input voltage range: 85 - 264 Vac 120 - 370 Vdc Input frequency range: 47 - 440 Hz Input surge current: 110 Vac, 60 Hz, cold start 230 Vac, 50 Hz, cold start 10 A 22 A Safety ground leakage current: 110 Vac, 60 Hz 230 Vac, 50 Hz 18 µA max. 28 µA max. Embedded Power for Business-Critical Continuity Rev.06.9.09_118 NFS40 Medical of 4 Electrical Specifications Continued All specifications are typical at nominal input, full load at 25 °C unless otherwise stated EMC Charateristics Conducted emissions: EN55022, FCC part 15 Level A Radiated emissions: EN55022, FCC part 15 Level A ESD air: EN61000-4-2, level 3 Perf. criteria 1 ESD contact: EN61000-4-2, level 4 Perf. criteria 1 Surge: EN61000-4-5, level 3 Perf. criteria 1 Fast transients: EN61000-4-4, level 3 Perf. criteria 1 Radiated immunity: EN61000-4-3, level 3 Perf. criteria 2 Conducted immunity: EN61000-4-6, level 3 Perf. criteria 2 Hold-up time: 110 Vac 230 Vac 18 ms 132 ms Efficiency: Isolation voltage: 110 Vac, 230 Vac 70% typical Input/output Input/chassis 4000 Vac 1500 Vac 20 - 110 kHz General Specifications Switching frequency: Approvals and standards: (See Note 12) Weight: MTBF demonstrated: VDE0750, IEC601, EN60601-1, UL2601, CSA C22.2 No. 125 270 g (9.6 oz) MIL-HDBK-217E 170,000 hours Environmental Specifications Thermal performance: Operating, see curve 0° C to +70 °C Non-operating -40 °C to +85 °C 0 - 50 °C ambient temperature, convection cooled 0 - 50 °C ambient temperature 40 W +50 °C to +70 °C ambient Derate linearly to 50% load Peak (30 seconds) 60 W Relative humidity: Non-condensing 5 to 95% RH Altitude: Operating 10,000 feet max. Non-operating 40,000 feet max. 5 - 500 Hz 0.75 G peak Vibration (See Note 11): 50 W @ 20 CFM Embedded Power for Business-Critical Continuity Rev.06.9.09_118 NFS40 Medical of 4 Ordering Information Ordering Information Output Voltage 4 5 6 7 8 9 10 11 12 13 14 Conv Max (1) 20 CFM Max. (2) Peak (3) Ripple (4) PK-PK Total Regulation (5) +5.1 V (VA) 3A 5A 7A 50 mV ± 2.0% +12 V (VB) 2A 2A 3A 120 mV ± 5.0% –12 V (6) 0.35 A 0A --- 120 mV ± 5.0% +5.1 V (VA) 3A 5A 7A 50 mV ± 2.0% Model Numbers (13, 14) NFS40-7908J NFS40-7910J +15 V (VB) 2A 2A 2.5 A 150 mV ± 10%/-3.0% -15 V (6) 0.35 A 0.5 A --- 150 mV ± 5.0% 12 V (7) 3.3 A 4A 5A 120 mV ± 2.0% NFS40-7912J 15 V (7) 2.6 A 3.3 A 4A 150 mV ± 2.0% NFS40-7915J 24 V 1.6 A 2A 2.5 A 240 mV ± 2.0% NFS40-7924J +5.1 V 4A 7A 5A 50 mV ± 2.0% NFS40-7928J +12 V 0.35 A 1A 0.5 A 120 mV ± 5.0% –12 V 0.35 A 1A 0.5 A 120 mV ± 5.0% Notes 1 2 3 Output Currents (7) Natural convection cooling, 40 W maximum. Forced air, 20 CFM at 1 atmosphere, 50 W maximum. Peak output current lasting less than 30 seconds with duty cycle less than 10%. During peak loading, outputs may go outside of total regulation limits. Peak total power must not exceed 60 W. 50 MHz bandwidth, peak-to-peak, measured differentially. Total regulation is defined as the static output regulation at 25 °C, including initial tolerance, load currents within stated limits, and output voltages adjusted to their factory settings. Also, 0.25 ² IA / IB ² 5.0 to maintain stated regulation. A minimum load of 0.5 A is required on the +5 V output to obtain full current from the negative output. Single output models have floating outputs which may be referenced as either positive or negative. Derating curve is application specific for ambient temperatures > 50 °C, for optimum reliability no part of the heatsink should exceed 120 °C and no semiconductor case temperature should exceed 130 °C. Caution: Allow a minimum of 1 second after disconnecting the power when making thermal measurements. Although the minimum output current of the NFS40-79XXJ is 0 A, a 4 W minimum load is required to achieve design MTBF. Three orthogonal axes, sweep at 1 octave/min, 5 minute dwell at four major resonances. This product is only for inclusion by professional installers within other equipment and must not be operated as a stand alone product. The ‘J’ suffix indicates that these parts are Pb-free (RoHS 6/6) compliant. NOTICE: Some models do not support all options. Please contact your local Emerson Network Power representative or use the on-line model number search tool at http://www.PowerConversion.com to find a suitable alternative. AC mating connector Molex 09-50-3031 or equivalent with Molex 08-50-0105 or equivalent crimp terminal. DC mating connector Molex 09-91-0600 or equivalent with Molex 08-50-0164 or equivalent crimp terminal. Pin Connections J1 -7908J, -7928J -7910J SINGLES Pin 1 AC Line AC Line AC Line Pin 2 AC Neutral AC Neutral AC Neutral J2 Pin 1 +12 V +15 V +Vout Pin 2 +5.1 V +5.1 V +Vout Pin 3 +5.1 V +5.1 V +Vout Pin 4 Return Return Return Pin 5 Return Return Return Pin 6 -12 V -15 V Return P1 Pin 1 Safety Earth Ground DERATING CURVE Output Power (Watts) 50W 20 CFM FORCED AIR COOLING 40W NATURAL CONVECTION COOLING 0C 10C 20C 30C 40C 25W 20W 50C 60C 70C Embedded Power for Business-Critical Continuity Americas Mechanical Drawing 0.156 (3.96) Dia. (4 Places) NON-METALLIC STAND-OFF (SEE NOTE A) J1 3.00 (76.2) 5810 Van Allen Way Carlsbad, CA 92008 USA Telephone: +1 760 930 4600 Facsimile: +1 760 930 0698 5.00 (127.00) C5 Europe (UK) MAXIMUM COMPONENT HEIGHT 1.2 inches Waterfront Business Park Merry Hill, Dudley West Midlands, DY5 1LX United Kingdom Telephone: +44 (0) 1384 842 211 Facsimile: +44 (0) 1384 843 355 Q1 1 ~ N ~ L F1 2.550 (64.8) T1 J2 F 3.15A, H 250VAC P1 (SEE NOTE C) 0.22 (5.59) 0.22 (5.59) 4.55 (115.8) ALL DIMENSIONS IN INCHES (mm) Mechanical Notes Rev.06.9.09_118 NFS40 Medical of 4 A In order to meet safety requirements, a non-metallic stand-off is mandatory for one hole as specified in the mechanical drawing above. B The ground pad of the mounting hole near P1 allows system grounding through a metal stand-off. C To improve conducted noise, the ground pad of the mounting hole near the output connector should be connected with the ground pad of the mounting hole near P1. Use metal stand-offs attached to a common metal chassis. This connection also significantly attenuates common mode noise. D A standard L-bracket and enclosure kit is available for mounting which contains all screws, connectors and necessary mounting hardware. Order part number NFS40CJ. Asia (HK) 14/F, Lu Plaza 2 Wing Yip Street Kwun Tong, Kowloon Hong Kong Telephone: +852 2176 3333 Facsimile: +852 2176 3888 For global contact, visit: www.PowerConversion.com techsupport.embeddedpower @emerson.com While every precaution has been taken to ensure accuracy and completeness in this literature, Emerson Network Power assumes no responsibility, and disclaims all liability for damages resulting from use of this information or for any errors or omissions. Emerson Network Power. The global leader in enabling business-critical continuity. AC Power Connectivity DC Power Embedded Computing Embedded Power Monitoring Outside Plant Power Switching & Controls Precision Cooling Racks & Integrated Cabinets Services Surge Protection EmersonNetworkPower.com Emerson Network Power and the Emerson Network Power logo are trademarks and service marks of Emerson Electric Co. ©2009 Emerson Electric Co.