Powerbox DLM80-37E 3000-4000 watts programmable dc supply Datasheet

LABORATORY
DLM Series
3000-4000 WATTS PROGRAMMABLE DC SUPPLY
Features
•
•
•
•
•
•
•
High power density
Preview push button
Remote voltage sense & remote operation
Calibration without removing covers
Regulatory compliance
Parrallel or series operation
IEEE-488.2 and RS-232 interface options
Specifications
OPERATING
INPUT
Input Voltage
3KW:180-264VAC, single or three phase
(<200VAC range limited to 40°C maximum) 4KW:180264VAC, three phase (<200VAC range limited to 40°C
maximum)
Frequency
47-63Hz for both single and three phase input
Power Factor
0.95 typical with three phase input
0.98 typical with single phase input
ENVIRONMENTAL
Efficiency
5-8V Models: 82% typical
16-80V Models: 87% typical
150-600V Models: 85% typical
Remote sense
The maximum allowed sense line drop is 2V for 5V, 8V
and 16V models and 5V for all other models
Remote sense protection
Remote programming
Voltage, current (0-100%) and OVP (5-110%) of full
scale can be programmed by selectable 0-5VDC, 010VDC, or 0-5kΩ
Remote monitoring
Voltage or current can be monitored with userselectable ranges, 0-5VDC or 0-10VDC
Operating temperature 0°C to 50° (No De-rating)
Storage temperature
-40°C to 65°C
Cooling
Internal fan with over-temperature protection
OUTPUT
Voltage adjustment
Zero to rated output
Current adjustment
Zero to rated output
Ripple & Noise
See table
Regulation
(Line or Load)
Voltage: 0.05% of maximum rated output +2mV
Current: 0.1% of maximum rated output
Transient response
Typically recovers in 1ms to within 1% of steady-state
output voltage (greater than 50% Vmax) for a 70100% or 100-70% load change
Stability
+/-0.05% of maximum voltage or current over 8 hours
after 15 minute warm-up time at fixed line, load and
temperature
Over-voltage protection Crowbar type adjustable from 5-110% rated output
using front panel control
(local or remote program selectable)
Displays and indicators Two 3.5 digital LED displays indicate output voltage,
output current and over-voltage setting. LED indication
for power on, shut-down, remote, overvoltage, overtemperature and front panel lockout, constant voltage
and constant current modes. IEEE-488.2 indicators
include error, SRQ and address
Software
LabVIEW® driver for M9E/M85
(Contact Powerbox)
Built-in protection
Over-voltage (resets by cycling the Enable/standby
switch), over-temperature (will automatically reset)
Master/slave parallel operation, up to 2 units can be
connected in parallel with active current sharing
control to within 10% of each supply. Series operation,
up to 3 units of the same model type can be connected
in series (consult manual). Negative terminal rated at
150Vmax Above ground
Operational features
Temperature coefficient 0.02%/°C of maximum output voltage
0.03%/°C of maximum output current
Regulatory compliance
488
Unit will not be damaged due to misconnection of the
remote sense leads
CE Mark on all models
LABORATORY
DLM Series
3000-4000 WATTS PROGRAMMABLE DC SUPPLY
Selection Tables
OUTPUT RATING
REGULATION LINE AND LOAD
METER ACCURACY
VOLTAGE
VOLTAGE
CURRENT
CURRENT
CURRENT
(0.05% OF VMAX
(0.5% OF VMAX (0.75%OF IMAX
(ADC)
(0.1% OF IMAX)
+2MV)
+1 COUNT)
+1 COUNT)
MODEL
VOLTAGE
(VDC)
DLM 5-350E
0-5
0-350
5mV
350mA
0.04V
4A
DLM 5-450E
0-5
0-450
5mV
450mA
0.04V
5A
DLM 8-350E
0-8
0-350
6mV
350mA
0.05V
4A
DLM 8-450E
0-8
0-450
6mV
450mA
0.05V
5A
DLM 16-185E
0-16
0-185
10mV
185mA
0.09V
3A
DLM 16-250E
0-16
0-250
10mV
250mA
0.09V
3A
DLM 32-95E
0-32
0-95
18mV
95mA
0.3V
0.8A
DLM 32-125E
0-32
0-125
18mV
125mA
0.3V
1A
DLM 40-75E
0-40
0-75
22mV
75mA
0.3V
0.7A
DLM 40-100E
0-40
0-100
22mV
100mA
0.3V
0.9A
DLM 60-50E
0-60
0-50
32mV
50mA
0.4V
0.5A
DLM 60-66E
0-60
0-66
32mV
66mA
0.4V
0.6A
DLM 80-37E
0-80
0-37
42mV
37mA
0.5V
0.4A
DLM 80-50E
0-80
0-50
42mV
50mA
0.5V
0.5A
DLM 150-20E
0-150
0-20
77mV
20mA
0.9V
0.3A
DLM 150-26E
0-150
0-26
77mV
26mA
0.9V
0.3A
DLM 300-10E
0-300
0-10
152mV
10mA
1.6V
0.09A
DLM 300-13E
0-300
0-13
152mV
13mA
1.6V
0.11A
DLM 600-5E
0-600
0-5
302mV
5mA
3.1V
0.05A
DLM 600-6.6E
0-600
0-6.6
302mV
7mA
3.1V
0.06A
PREVIEW ACCURACY
MODEL
OVP
VOLTAGE CURRENT ADJUSTMENT
(0.5% OF (1.0% OF RANGE (5%
VMAX +1 IMAX +1 TO 110% OF
VMAX)
COUNT) COUNT)
RIPPLE AND NOISE
RIPPLE
(RMS)*
NOISE
(P-P)
STABILITY
TEMP. COEFF.
MAXIMUM
CURRENT TOTAL
VOLTAGE CURRENT VOLTAGE
(0.03%/ REMOTE
(0.05% OF (0.05% (0.02%/°C
SENSE
°C OF
VMAX)
OFIMAX) OF VMAX)
DROP
IMAX)
DLM 5-350E
0.04V
5A
0.3-5.5V
12mV
100mV
3mV
175mA
1mV
105mA
2V
DLM 5-450E
0.04V
6A
0.3-5.5V
12mV
100mV
3mV
225mA
1mV
135mA
2V
DLM 8-350E
0.05V
5A
0.4-8.8V
12mV
100mV
4mV
175mA
1.6mV
105mA
2V
DLM 8-450E
0.05V
6A
0.4-8.8V
12mV
100mV
4mV
225mA
1.6mV
135mA
2V
DLM 16-185E
0.09V
3A
0.8-17.6V
10mV
100mV
8mV
93mA
3.2mV
55mA
2V
DLM 16-250E
0.09V
4A
0.8-17.6V
10mV
100mV
8mV
125mA
3.2mV
75mA
2V
DLM 32-95E
0.3V
1.1A
1.6-35V
10mV
100mV
16mV
48mA
6mV
30mA
5V
DLM 32-125E
0.3V
1.4A
1.6-35V
10mV
100mV
16mV
63mA
6mV
38mA
5V
DLM 40-75E
0.3V
0.9A
2-44V
10mV
100mV
20mV
38mA
8mV
23mA
5V
DLM 40-100E
0.3V
1.1A
2-44V
10mV
100mV
20mV
50mA
8mV
30mA
5V
DLM 60-50E
0.4V
0.6A
3-66V
15mV
100mV
30mV
25mA
12mV
15mA
5V
DLM 60-66E
0.4V
0.8A
3-66V
15mV
100mV
30mV
33mA
12mV
19.8mA
5V
DLM 80-37E
0.5V
0.5A
4-88V
15mV
120mV
40mV
19mA
16mV
12mA
5V
DLM 80-50E
0.5V
0.6A
4-88V
15mV
120mV
40mV
25mA
16mV
15mA
5V
DLM 150-20E
0.9V
0.3A
7.5-165V
30mV
200mV
75mV
10mA
30mV
6mA
5V
DLM 150-26E
0.9V
0.4A
7.5-165V
30mV
200mV
75mV
13mA
30mV
7.8mA
5V
DLM 300-10E
1.6V
0.11A
15-330V
60mV
300mV
150mV
5mA
60mV
3mA
5V
DLM 300-13E
1.6V
0.14A
15-330V
60mV
300mV
150mV
6.5mA
60mV
3.9mA
5V
DLM 600-5E
3.1V
0.06A
30-660V
100mV
500mV
300mV
2.5mA
120mV
1.5mA
5V
DLM 600-6.6E
3.1V
0.08A
30-660V
100mV
*RMS ripple typical from 20Hz to 300kHz
Dimensions: 2RU or 88mm(H)x482mm(W)x508mm(D) ; Weight: 18.2kg
500mV
300mV
3.3mA
120mV
2mA
5V
Your dependable power partner – www.powerbox.com.au
489
LABORATORY
DLM Series
3000-4000 WATTS PROGRAMMABLE DC SUPPLY
OPTIONS & ACCESSORIES
J3 PROGRAM AND SENSE
INPUT VOLTAGE OPTIONS
M9E
IEEE-488.2 and RS 232 Interfaces
3
Remote Shutdown Return(-)
15
Remote OVP Programing Input
16
M13
Locking shafts (front panel potentiometers)
4
Remote Programming Indicator 17
Remote Shutdown Input (+), Pos or
Neg true logic selection with S1
+5 VDC Auxillary Output
1 mACurrent Source for OVP
Programming
OVP Status Indicator
M51A
Optically Isolated analog programming
5
Operating Mode Indicator
18
Overtemperature Shutdown Indicator
M85
Multichannel Slave Interface
6
Status Indicator Return (-)
19
DC Voltage Monitor Output
7
Current Monitor Output
N/C
20
M1
400 VAC input (4 kW only)
M2
480 VAC input (4 kW only
REMOTE INTERFACE OPTIONS
PARALLEL CABLE
5361969-01 Paralleling Cable, one cable per slave unit
RACK SLIDE KIT
105-330-26 Rack slide kit
Technical Illustration
490
1
2
Remote Output Enable
14
10
Current Programing Unit
23
Remote/Local Voltage Control Select
1 mA Current Source for Voltage
Programming
1 mA Current Source for Current
Programming
Remote/Local Current Control Select
11
N/C
24
N/C
12
Programming/Monitor Return
25
N/C
13
N/C
8
9
Voltage Programing Input
21
22
Similar pages