MORNSUN G0512S-2WR2 Sip package Datasheet

DC/DC Converter
G_S-2WR2 & H_S-2WR2 Series
FEATURES
2W, Fixed input voltage, 4200VAC or 6000VDC isolated
& unregulated positive-negative dual/single output




SIP package
High efficiency up to 84%
Reinforced insulation
The patient leakage current: Max 2µA
 Isolation voltage: 4200VAC or 6000VDC




Patent Protection RoHS
Operating temperature range:-40℃~+85℃
Internal surface mounted design
International standard pin-out
Meets EN60601-1, ANSI/AAMI ES60601-1
standards(Pending)
(1xMOPP/2xMOOP)
G_S-2WR2 & H_S-2WR2 series meet reinforced insulation requirements. It is specially designed for applications where require compact size,
high isolation, low isolation capacitor and low leakage current power. It is widely used in medical, electricity, IGBT driver and so on. It is
suitable for:
1.Where the voltage of the input power supply is stable (voltage variation: ±10%Vin);
2.Where isolation is necessary between input and output (isolation voltage ≤4200VAC or 6000VDC);
3.Where do not has high requirement of line regulation and the ripple & noise of the output voltage;
Such as: Medical collection and isolation, High voltage collection circuit, IGBT-driven circuits, etc.
Selection Guide
Input Voltage (VDC)
Certification
Part No.
Efficiency
(%,Min./Typ.)
Max. Capacitive
Load*
@ Full Load
(µF)
Output Voltage(VDC)
Output Current
(mA)(Max./Min.)
G0505S-2WR2
±5
±200/±20
74/78
470
G0509S-2WR2
±9
±111/±12
74/78
470
±12
±83/±9
74/78
220
±15
±67/±7
76/80
220
5
400/40
73/77
1000
H0512S-2WR2
12
167/17
75/79
470
H0515S-2WR2
15
133/14
75/79
470
G1205S-2WR2
±5
±200/±20
74/78
470
G1209S-2WR2
±9
±111/±12
78/82
470
±12
±83/±9
78/82
220
±15
±67/±7
76/80
220
5
400/40
73/77
1000
12
167/17
76/80
470
G0512S-2WR2
G0515S-2WR2
H0505S-2WR2
UL/CE
(Pending)
Nomina
l(Range)
Output
G1212S-2WR2
G1215S-2WR2
H1205S-2WR2
5
(4.5-5.5)
12
(10.8-13.2)
H1212S-2WR2
H1215S-2WR2
15
133/14
78/82
470
G2405S-2WR2
±5
±200/±20
75/79
470
G2409S-2WR2
±9
±111/±12
77/81
470
±12
±83/±9
78/82
220
±15
±67/±7
77/81
220
H2405S-2WR2
5
400/40
75/79
1000
H2412S-2WR2
12
167/17
78/82
470
H2415S-2WR2
15
133/14
80/84
470
G2412S-2WR2
G2415S-2WR2
24
(21.6-26.4)
Note:* The capacitive loads of positive and negative outputs are identical.
2016.03.07-A/0
Page 1 of 5
MORNSUN Guangzhou Science & Technology Co., Ltd. reserves the copyright and right of final interpretation
DC/DC Converter
G_S-2WR2 & H_S-2WR2 Series
Input Specifications
Item
Operating Conditions
Input Current (no-load/full load)
Surge Voltage (1sec. max.)
Min.
Typ.
Max.
5V input
--
35/520
80/--
12V input
--
15/217
40/--
24V input
--
10/106
25/--
5V input
-0.7
--
9
12V input
-0.7
--
18
24V input
-0.7
--
30
--
0.2
--
Reflected Ripple Current
Input Filter
Unit
mA
VDC
A
Capacitor filter
Output Specifications
Item
Operating Conditions
Min.
Output Voltage Accuracy
Typ.
Max.
Unit
See tolerance envelope graph (Fig. 1)
Line Regulation
Input voltage change: ±1%
Load Regulation
10%-100% load
--
--
±1.2
5VDC output
--
--
20
9VDC output
--
--
15
12VDC output
--
--
15
15VDC output
--
--
15
--
%
Ripple & Noise*
20MHz bandwidth
--
100
150
mVp-p
Temperature Drift Coefficient
100% full load
--
±0.02
--
%/℃
--
--
3
s
Output Short Circuit Protection**
Note: *Ripple and noise tested with “parallel cable” method, please see DC-DC Converter Application Notes for specific operation methods.
**Supply voltage must be discontinued at the end of short circuit duration which less than 3s.
General Specifications
Item
Operating Conditions
Isolation Voltage
Input-output, with the test time of 1 minute
Patient Leakage Current
250VAC,50/60Hz
Isolation Resistance
Input-output, isolation voltage 500VDC
Isolation Capacitance
Input-output, 100KHz/0.1V
Operating Temperature
Storage Temperature
Min.
Typ.
Max.
Unit
4200
--
--
VAC
6000
--
--
VDC
--
--
2
μA
1000
--
--
MΩ
--
5
--
pF
-40
--
85
-55
--
125
Casing Temperature Rise
Pin Welding Resistance
Temperature
Ta=25℃
--
25
--
Welding spot is 1.5mm away from the casing, 10 seconds
--
--
300
Storage Humidity
Non-condensing
--
--
95
%
Switching Frequency
100% load, nominal input voltage
--
100
--
KHz
MTBF
MIL-HDFK-217F@25℃
K hours
3500
--
--
Transformer Creepage
5
--
--
Transformer Clearance
5
--
--
5.5
--
--
PCB Creepage & Clearance
℃
mm
Note:1、Patient leakage current and reinforced insulation is based on a 250 VAC, 50/60 Hz system input voltage.
2、The UL certification (ANSI/AAMI ES60601-1, File No. E347375) of G_S-2WR2 & H_S-2WR2 series is pending, G_S-2WR2 & H_S-2WR2 series meets
1xMOPP/2xMOOP when system input voltage is with 250VAC , 50/60Hz.
Physical Specifications
Casing Material
Package Dimensions
Black flame-retardant and heat-resistant plastic (UL94 V-0)
19.50*9.80*12.50 mm
Weight
4.2g(Typ.)
Cooling Method
Free air convection
2016.03.07-A/0
Page 2 of 5
MORNSUN Guangzhou Science & Technology Co., Ltd. reserves the copyright and right of final interpretation
DC/DC Converter
G_S-2WR2 & H_S-2WR2 Series
EMC Specifications
EMI
EMS
Conducted emission
CISPR22/EN55022 CLASS B (see Fig. 5 for recommended circuit)
Radiated emission
CISPR22/EN55022 CLASS B (see Fig. 5 for recommended circuit)
Electrostatic discharge
IEC/EN61000-4-2
Contact ±8KV
perf. Criteria B
Product Characteristic Curve
Other output
5VDC output
Tolerance Envelope Cur ve
+5%
0
-5%
Max .
Ty p .
M in.
+2.5%
-2.5%
-7.5%
-10%
10% 20%
+15%
Output Voltage Accuracy
+10%
40%
80%
60%
Output Current Percent
(Nominal Input Voltage)
+10%
Max .
+5%
Ty p .
0
+2.5%
M in.
-5%
-2.5%
-7.5%
-10%
10% 20%
100%
40%
60%
80%
100%
Output Current Percent
(Nominal Input Voltage)
Fig. 1
Output Power Percentage(%)
Output Voltage Accuracy
+15%
Tolerance Envelope Cur ve
Temperature Curve
120
100
80
60
Safe Operating Area
40
20
0
-40
0
40
85
120
Operating Temperature(℃)
Fig. 2
2016.03.07-A/0
Page 3 of 5
MORNSUN Guangzhou Science & Technology Co., Ltd. reserves the copyright and right of final interpretation
DC/DC Converter
G_S-2WR2 & H_S-2WR2 Series
Design Reference
1. Typical application
If it is required to further reduce input and output ripple, a filter capacitor can be connected to the input and output terminals, see Fig.3.
Moreover, choosing suitable filter capacitor is very important, start-up problems may be caused by too large capacitance. To ensured the
modules running well, the recommended capacitive load values as shown in Table 1.
The simplest device for output voltage regulation, over-voltage and over-current protection is a linear voltage regulator with overheat
protection that is connected to the input or output end in series (see Fig. 4).
S in g l e
Single
+ Vo
V in
DC
C in
GN D
DC
Vin
C ou t
REG
GND
0V
Dual
REG
+Vo
DC DC
0V
Dual
V in
DC
C in
GN D
C ou t
DC
C o ut
Fig. 3
+ Vo
Vin
0V
GND
REG
-Vo
REG
+Vo
REG
-Vo
DC DC
0V
Fig. 4
Recommended capacitive load value table (Table 1)
Single
Dual
Vin
Cin
Cout
Cout
Vout
Vout
(VDC)
(µF)
(µF)
(µF)
(VDC)
(VDC)
5
10
5
10
±5
4.7
12
4.7
12
2.2
±9
2.2
24
2.2
15
1
±12/±15
1
2. EMC typical recommended circuit (CLASS B)
LD M
V in
+
DC Power
S u p p ly
-
C1
D C /D C
C2
GND
- Vo
(0 V )
Recommended typical circuit parameters:
Input voltage (V)
+ Vo
C3
LO A D
EMI
5/12
24
C1,C2
4.7µF /50V
C3
Refer to the Cout in Fig.3
LDM
6.8µH
15µH
Fig. 5
3. Output load requirements
In order to ensure the converter can work reliably with high efficiency, the minimum load should not less than 10% rated load when it is used.
If the needed power is indeed small, please parallel a resistor on t the output side ( The sum of the efficient power and resistor consumption
power is not less than 10%).
4. For more information please find the application notes on www.mornsun-power.com
2016.03.07-A/0
Page 4 of 5
MORNSUN Guangzhou Science & Technology Co., Ltd. reserves the copyright and right of final interpretation
DC/DC Converter
G_S-2WR2 & H_S-2WR2 Series
Dimensions and Recommended Layout
Notes:
1.
Packing Information please refer to 'Product Packing Information'. Packing bag number: 58200013;
2.
If the product is operated under the min. required load, the product performance cannot be guaranteed to comply with all
performance indexes in this datasheet;
3.
The max. capacitive load should be tested within the input voltage range and under full load conditions;
4.
Unless otherwise specified, data in this datasheet should be tested under the conditions of Ta=25°C, humidity<75% when inputting
nominal voltage and outputting rated load;
5.
All index testing methods in this datasheet are based on our Company’s corporate standards;
6.
We can provide product customization service, and please directly contact our technicians for specific information;
7.
Specifications of this product are subject to changes without prior notice.
MORNSUN Guangzhou Science & Technology Co., Ltd.
Address: No. 5, Kehui St. 1, Kehui Development Center, Science Ave., Guangzhou Science City, Luogang District, Guangzhou, P. R. China
Tel: 86-20-38601850-8801
Fax: 86-20-38601272
E-mail: [email protected]
2016.03.07-A/0
Page 5 of 5
MORNSUN Guangzhou Science & Technology Co., Ltd. reserves the copyright and right of final interpretation
Similar pages