HP G4290B Agilent 1220 infinity ii hplc system for cannabinoid potency testing Datasheet

AGILENT 1220 INFINITY II HPLC
SYSTEM FOR CANNABINOID
POTENCY TESTING
APPLICATION OVERVIEW
Introduction
The Agilent 1220 Infinity II HPLC system offers reliable and robust
profiling and quantification tool for eleven of the most commonly targeted
cannabinoids found in marijuana (Cannabis Sativa) and products such as
edible foodstuffs derived from the plant material. The method defines the
analytical column, mobile phase modifiers and full system parameters to
separate the cannabinoids in under 5 minutes with baseline resolution of
1.86 for the critical pair.
Agilent 1220 Infinity II
Target Cannabinoids
Cannabidivarin (CBDV)
Tetrahydrocannabivarin (THCV)
Δ9-tetrahydrocannabinol (THC)
Δ8-tetrahydrocannabinol (THC)
Cannabidiol (CBD)
Cannabigerol (CBG)
Tetrahydrocannabinolic acid (THCA)
Cannabidiolic Acid (CBDA)
Cannabinol (CBN)
Cannabigerolic acid (CBGA)
Cannabichromene (CBC)
Full method parameters are included to separate
the above eleven cannabinoids.
Agilent offer solutions, proven over many years, to assess the
potency and safety of botanical commodities.
•
Certify the identity of a commodity and ensure accurately labeled.
•
Assess cannabis quality, and efficacy.
•
Choose from our broad portfolio of testing solutions for both
medicinal and recreational cannabis.
•
Buy or lease the best analytical equipment.
•
Move ahead with our method development and consulting services
The Agilent 1220 Infinity II for cannabinoid
potency testing configuration:
Solution
Part Number
1220 Infinity II LC Gradient System
o Includes gradient pump (max.
pressure 600 bar) with
integrated degassing unit, auto
sampler, column oven and
variable wavelength detector
and standard flow cell.
G4290B
OpenLAB 2.2 data acquisition,
analysis and reporting software
M8413AA
Poroshell Bonus RP 3.0 mm x 50
mm, 2.7 um particle size
Ammonium formate and formic acid
mobile phase modifiers
699968-301
THC
THCA
G1946-85021
G2453-85060
CBD
Technical brief with detailed method
parameters
Contact your Agilent representative for a quote.
Approximately 90% of the total THC in fresh marijuana plant
material occurs as THCA 1. Therefore, it is often desired to
measure the total THC content of C. sativa: Total THC = THC
+ 0.877 x (THCA). The Agilent 1220 Infinity II for cannabinoid
potency testing, easily measures total THC content and
provides a qualitative and quantitative profile of all the
targeted cannabinoids for potency, and efficacy
determinations.
For more information: Contact your
local Agilent representative or visit:
www.agilent.com/chem/cannabis
Agilent products and solutions are intended to be used for
cannabis quality control and safety testing in laboratories
where such use is permitted under applicable state law.
This information is subject to change without notice.
© Agilent Technologies, Inc. 2017
Published August 17, 2017
5991-8418EN
Similar pages